13
HCA – Liquids User Exercise Intrepid3.0

HCA - Liquids v1 0 (04 03 2014)[1]

Embed Size (px)

DESCRIPTION

HCA

Citation preview

  • HCA Liquids

    User Exercise

    Intrepid 3.0

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 2 of 13

    Performing a HCA Liquids Analysis

    Step 1: If needed, add required layers

    Double-check to ensure that you have the following layers added to your MXD

    document:

    Spill Points: Required for Potential Method

    Spill Plumes: Required for Potential Method

    HCA Layers: Required for the graphical representation of the HCA analysis

    results

    HCA Polygons: The layers to analyze. These layers are required for all

    methods.

    For more information about layers, see Exercise 5.4: Manipulating Map Layers and the

    other exercises in the Map Layer Exercises section of the Intrepid 3.2 User Manual.

    Step 2: Click the Run Analysis workflow

    In in the Workflow Manager tab, click DOT Compliance > HCA Analysis > HCA

    Liquid Analysis > Run Analysis.

    The HCA Options Liquid Pipelines window displays:

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 3 of 13

    Step 3: Select the HCA layers to analyze

    1) Click the Add Layer button

    The Add Layer button is located in the HCA Layers to Analyze section:

    When you click it, the Pick a Layer window displays:

    2) Select the layers to use

    Select the required HCA layers to use in the analysis by clicking their checkboxes and

    then click OK.

    When you click OK, the selected layers load. Once loaded, their names display in the

    HCA Layers to Analyze section:

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 4 of 13

    Step 4: Select the lines to analyze

    1) Click the Add Lines button

    The Add Lines button is located in the Segments to Analyze section:

    The Select window displays.

    2) Select the lines to analyze

    Navigate to the line(s) to analyze, put a checkmark next to all lines to include, and

    then click the Select button. The names of the selected lines now display in the

    Segments to Analyze section.

    Step 5: Choose an analysis method

    There are three possible methods of analysis:

    Direct: This method determines where a centerline (and its optional buffer)

    overlaps an HCA and creates a ranged HCA event for the overlapped portion of

    the route:

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 5 of 13

    Indirect: This method determines where a HCA (and its optional buffer) overlaps

    a centerline and creates a ranged HCA event for the overlapped route section:

    Potential: This method determines where a spill plume overlaps a HCA and

    creates a ranged HCA event centered on the spill point that generated the spill

    plume. The ranged HCA event that is created has a length equal to twice the

    buffer distance:

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 6 of 13

    Select the chosen option from the Method drop-down menu and click Select:

    Step 6: Configure the HCA Options window for the method chosen

    Configure the options for the analysis method being used:

    A. Configure options for the Direct method

    1) (Optional) Click the Buffer Centerline checkbox

    The buffer is an extra amount of space added on either side of the centerline to

    provide an additional margin of safety when performing the calculations.

    Note: Setting a centerline buffer is recommended in order to compensate for any

    potential measurement inaccuracies. Forty (40) feet is a commonly used value

    for the buffer.

    The Buffer Centerline checkbox is located in the Analysis Options section:

    2) (Optional) Select a buffering method

    Note: This field is not optional if the Buffer Centerline checkbox is checked.

    Click the drop-down menu, located directly below the Buffer Centerline checkbox,

    and select the buffering method to use:

    Calculate from Spill Volume

    Use HVL Heat Explosion Buffer

    Use HVL Toxicity Buffer

    User Specified Buffer Distance

    For the purposes of this exercise, select User Specified Buffer Distance.

    Note: If you select one of the first three options, the calculations are based on values

    stored in the Line Segment window:

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 7 of 13

    For more information, see Exercise 3.8: Viewing, Editing, and Deleting

    Organizational Information of the Intrepid PODS 3.2 User Manual.

    3) (Optional) Set the buffer value

    Note: This field is not optional if the Buffer Centerline checkbox is checked.

    In the test field located directly to the right of the drop-down menu, type in a buffer

    value to set it.

    Note: If you are unsure what to enter, 40 is a commonly recommended value.

    4) Select the Analyze multiple crossings of the same polygon individually

    checkbox

    Click the Analyze multiple crossings of the same polygon individually checkbox

    to select it.

    Note: Selecting the above box ensures that when the centerline crosses the same

    polygon more than once, HCA events are created separately for each

    crossing. Otherwise, only one HCA event would be created. This event would

    have a begin station at the first crossing and the end station at the end

    crossing.

    B. The Indirect method

    1) (Optional) Click the Buffer HCA Areas checkbox

    The buffer is an extra amount of space added on either side of the HCA areas to

    provide an additional margin of safety when performing the calculations.

    Note: Setting an HCA buffer is recommended in order to compensate for any

    potential measurement inaccuracies. A commonly used value for such a buffer

    is 1320 ft.

    The Buffer Centerline checkbox is located in the Analysis Options section:

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 8 of 13

    2) (Optional) Select User Specified Buffer Distance

    Note: The Select Buffer Distance field is required if Buffer HCA Areas is checked.

    Click the drop-down menu located directly below the Buffer Centerline checkbox,

    and select the buffering method to use:

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 9 of 13

    Calculate from Spill Volume

    Use HVL Heat Explosion Buffer

    Use HVL Toxicity Buffer

    User Specified Buffer Distance

    For the purposes of this exercise, select User Specified Buffer Distance.

    Note: If you select one of the first three options, the calculations are based on values

    stored in the Line Segment window:

    For more information, see Exercise 3.8: Viewing, Editing, and Deleting

    Organizational Information of the Intrepid PODS 3.2 User Manual.

    3) (Optional) Set the buffer value

    Note: The Select Buffer Distance field is required if Buffer HCA Areas is checked.

    In the test field located directly to the right of the drop-down menu, type in a buffer

    value to set it.

    Note: If you are unsure what to enter, 1320 is a commonly recommended value.

    4) Select the Analyze multiple crossings of the same polygon individually

    checkbox

    Click the Analyze multiple crossings of the same polygon individually checkbox

    to select it.

    Note: Selecting the above box ensures that when the centerline crosses the same

    polygon more than once, HCA events are created separately for each

    crossing. Otherwise, only one HCA event would be created. This event would

    have a begin station at the first crossing and the end station at the end

    crossing.

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 10 of 13

    C. The Potential method

    1) Click the Extend HCA Results Along the Centerline checkbox

    Note: This buffer is different than the ones used in the Direct and Indirect Methods

    because it does not affect when an HCA event needs to be created, but rather

    how large the HCA event is once it needs to be created. This buffer is not

    optional.

    When using the Potential Method, a new ranged HCA event is created on the

    centerline when the analysis determines that a calculated spill plume polygon

    intersects with a HCA polygon (see the illustration above). To create the HCA

    event on the line, this buffer amount is subtracted from the station of the spill

    plume point to set the new HCA events begin point. To set the end point, this

    buffer value is added to the spill plume points station.

    For example, an HCA event needs to be created because a spill plume from

    station 10+00 intersects with an HCA area. If the buffer is 500 feet, this means

    that the resulting HCA event that is created on the centerline runs from 5+00

    to 15+00.

    The Extend HCA Results Along the Centerline checkbox is located in the Analysis

    Options section:

    2) Select User Specified Buffer Distance

    Click the drop-down menu located directly below the Buffer Centerline checkbox,

    and select the buffering method to use:

    Calculate from Spill Volume

    Use HVL Heat Explosion Buffer

    Use HVL Toxicity Buffer

    User Specified Buffer Distance

    For the purposes of this exercise, select User Specified Buffer Distance.

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 11 of 13

    Note: If you select one of the first three options, the calculations are based on values

    stored in the Line Segment window:

    For more information, see Exercise 3.8: Viewing, Editing, and Deleting

    Organizational Information of the Intrepid PODS 3.2 User Manual.

    3) Set the buffer value

    Note: The buffer value is typically equal to the interval between spill plume points.

    In the test field located directly to the right of the drop-down menu, type in a buffer

    value to set it.

    4) Check the Measure Horizontal checkbox

    5) Check the Utilize Spill Plume Analysis Layer checkbox

    If the Utilize Spill Plume Analysis Layer checkbox is unchecked, put a check in it

    now.

    6) Click the Select Points button

    The Select Points button is located in the bottom of the Analysis Options section:

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 12 of 13

    The Pick a Layer window displays:

    7) Select the layer containing the spill point data, then click OK

    The layers name now displays next to the Select Points button.

    8) Click the Select Plumes button

    The Select Plumes button is located immediately below the Select Points button:

    When you click it, the Pick a Layer window displays again.

    9) Select the layer containing the spill plume data, then click OK

    The layers name now displays next to the Select Plumes button.

    Step 7: (Optional) Save the settings for future use

  • HCA Liquids User Exercises

    HCA Liquids User Exercises v1.0 (04.03.2014) Page 13 of 13

    You can save your current settings for future use by using the buttons in the Default

    Settings section:

    Enter a name for your settings and then click the Save/Update button. The name you

    chose now appears in the Default Settings list.

    Note: You can modify a saved setting by double-clicking on it (click Yes on the

    message window that appears), making your changes, and clicking the

    Save/Update button. To delete a saved setting, click it and then click the Delete

    button.

    Step 8: Start the Analysis

    1) Click Run Analysis

    Click the Run Analysis button, located in the extreme bottom-right of the window, to

    start the analysis process.

    The Enter HCA Analysis Name window displays:

    2) Specify a name for the HCA analysis

    Enter a name for this particular HCA analysis and then click OK.

    The Status processing window displays.

    3) Click OK

    When the analysis is completed, a confirmation message displays. Click OK to

    confirm the completion of the process.