Upload
carlos-nieves
View
15
Download
5
Tags:
Embed Size (px)
DESCRIPTION
HCA
Citation preview
HCA Liquids
User Exercise
Intrepid 3.0
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 2 of 13
Performing a HCA Liquids Analysis
Step 1: If needed, add required layers
Double-check to ensure that you have the following layers added to your MXD
document:
Spill Points: Required for Potential Method
Spill Plumes: Required for Potential Method
HCA Layers: Required for the graphical representation of the HCA analysis
results
HCA Polygons: The layers to analyze. These layers are required for all
methods.
For more information about layers, see Exercise 5.4: Manipulating Map Layers and the
other exercises in the Map Layer Exercises section of the Intrepid 3.2 User Manual.
Step 2: Click the Run Analysis workflow
In in the Workflow Manager tab, click DOT Compliance > HCA Analysis > HCA
Liquid Analysis > Run Analysis.
The HCA Options Liquid Pipelines window displays:
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 3 of 13
Step 3: Select the HCA layers to analyze
1) Click the Add Layer button
The Add Layer button is located in the HCA Layers to Analyze section:
When you click it, the Pick a Layer window displays:
2) Select the layers to use
Select the required HCA layers to use in the analysis by clicking their checkboxes and
then click OK.
When you click OK, the selected layers load. Once loaded, their names display in the
HCA Layers to Analyze section:
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 4 of 13
Step 4: Select the lines to analyze
1) Click the Add Lines button
The Add Lines button is located in the Segments to Analyze section:
The Select window displays.
2) Select the lines to analyze
Navigate to the line(s) to analyze, put a checkmark next to all lines to include, and
then click the Select button. The names of the selected lines now display in the
Segments to Analyze section.
Step 5: Choose an analysis method
There are three possible methods of analysis:
Direct: This method determines where a centerline (and its optional buffer)
overlaps an HCA and creates a ranged HCA event for the overlapped portion of
the route:
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 5 of 13
Indirect: This method determines where a HCA (and its optional buffer) overlaps
a centerline and creates a ranged HCA event for the overlapped route section:
Potential: This method determines where a spill plume overlaps a HCA and
creates a ranged HCA event centered on the spill point that generated the spill
plume. The ranged HCA event that is created has a length equal to twice the
buffer distance:
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 6 of 13
Select the chosen option from the Method drop-down menu and click Select:
Step 6: Configure the HCA Options window for the method chosen
Configure the options for the analysis method being used:
A. Configure options for the Direct method
1) (Optional) Click the Buffer Centerline checkbox
The buffer is an extra amount of space added on either side of the centerline to
provide an additional margin of safety when performing the calculations.
Note: Setting a centerline buffer is recommended in order to compensate for any
potential measurement inaccuracies. Forty (40) feet is a commonly used value
for the buffer.
The Buffer Centerline checkbox is located in the Analysis Options section:
2) (Optional) Select a buffering method
Note: This field is not optional if the Buffer Centerline checkbox is checked.
Click the drop-down menu, located directly below the Buffer Centerline checkbox,
and select the buffering method to use:
Calculate from Spill Volume
Use HVL Heat Explosion Buffer
Use HVL Toxicity Buffer
User Specified Buffer Distance
For the purposes of this exercise, select User Specified Buffer Distance.
Note: If you select one of the first three options, the calculations are based on values
stored in the Line Segment window:
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 7 of 13
For more information, see Exercise 3.8: Viewing, Editing, and Deleting
Organizational Information of the Intrepid PODS 3.2 User Manual.
3) (Optional) Set the buffer value
Note: This field is not optional if the Buffer Centerline checkbox is checked.
In the test field located directly to the right of the drop-down menu, type in a buffer
value to set it.
Note: If you are unsure what to enter, 40 is a commonly recommended value.
4) Select the Analyze multiple crossings of the same polygon individually
checkbox
Click the Analyze multiple crossings of the same polygon individually checkbox
to select it.
Note: Selecting the above box ensures that when the centerline crosses the same
polygon more than once, HCA events are created separately for each
crossing. Otherwise, only one HCA event would be created. This event would
have a begin station at the first crossing and the end station at the end
crossing.
B. The Indirect method
1) (Optional) Click the Buffer HCA Areas checkbox
The buffer is an extra amount of space added on either side of the HCA areas to
provide an additional margin of safety when performing the calculations.
Note: Setting an HCA buffer is recommended in order to compensate for any
potential measurement inaccuracies. A commonly used value for such a buffer
is 1320 ft.
The Buffer Centerline checkbox is located in the Analysis Options section:
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 8 of 13
2) (Optional) Select User Specified Buffer Distance
Note: The Select Buffer Distance field is required if Buffer HCA Areas is checked.
Click the drop-down menu located directly below the Buffer Centerline checkbox,
and select the buffering method to use:
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 9 of 13
Calculate from Spill Volume
Use HVL Heat Explosion Buffer
Use HVL Toxicity Buffer
User Specified Buffer Distance
For the purposes of this exercise, select User Specified Buffer Distance.
Note: If you select one of the first three options, the calculations are based on values
stored in the Line Segment window:
For more information, see Exercise 3.8: Viewing, Editing, and Deleting
Organizational Information of the Intrepid PODS 3.2 User Manual.
3) (Optional) Set the buffer value
Note: The Select Buffer Distance field is required if Buffer HCA Areas is checked.
In the test field located directly to the right of the drop-down menu, type in a buffer
value to set it.
Note: If you are unsure what to enter, 1320 is a commonly recommended value.
4) Select the Analyze multiple crossings of the same polygon individually
checkbox
Click the Analyze multiple crossings of the same polygon individually checkbox
to select it.
Note: Selecting the above box ensures that when the centerline crosses the same
polygon more than once, HCA events are created separately for each
crossing. Otherwise, only one HCA event would be created. This event would
have a begin station at the first crossing and the end station at the end
crossing.
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 10 of 13
C. The Potential method
1) Click the Extend HCA Results Along the Centerline checkbox
Note: This buffer is different than the ones used in the Direct and Indirect Methods
because it does not affect when an HCA event needs to be created, but rather
how large the HCA event is once it needs to be created. This buffer is not
optional.
When using the Potential Method, a new ranged HCA event is created on the
centerline when the analysis determines that a calculated spill plume polygon
intersects with a HCA polygon (see the illustration above). To create the HCA
event on the line, this buffer amount is subtracted from the station of the spill
plume point to set the new HCA events begin point. To set the end point, this
buffer value is added to the spill plume points station.
For example, an HCA event needs to be created because a spill plume from
station 10+00 intersects with an HCA area. If the buffer is 500 feet, this means
that the resulting HCA event that is created on the centerline runs from 5+00
to 15+00.
The Extend HCA Results Along the Centerline checkbox is located in the Analysis
Options section:
2) Select User Specified Buffer Distance
Click the drop-down menu located directly below the Buffer Centerline checkbox,
and select the buffering method to use:
Calculate from Spill Volume
Use HVL Heat Explosion Buffer
Use HVL Toxicity Buffer
User Specified Buffer Distance
For the purposes of this exercise, select User Specified Buffer Distance.
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 11 of 13
Note: If you select one of the first three options, the calculations are based on values
stored in the Line Segment window:
For more information, see Exercise 3.8: Viewing, Editing, and Deleting
Organizational Information of the Intrepid PODS 3.2 User Manual.
3) Set the buffer value
Note: The buffer value is typically equal to the interval between spill plume points.
In the test field located directly to the right of the drop-down menu, type in a buffer
value to set it.
4) Check the Measure Horizontal checkbox
5) Check the Utilize Spill Plume Analysis Layer checkbox
If the Utilize Spill Plume Analysis Layer checkbox is unchecked, put a check in it
now.
6) Click the Select Points button
The Select Points button is located in the bottom of the Analysis Options section:
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 12 of 13
The Pick a Layer window displays:
7) Select the layer containing the spill point data, then click OK
The layers name now displays next to the Select Points button.
8) Click the Select Plumes button
The Select Plumes button is located immediately below the Select Points button:
When you click it, the Pick a Layer window displays again.
9) Select the layer containing the spill plume data, then click OK
The layers name now displays next to the Select Plumes button.
Step 7: (Optional) Save the settings for future use
HCA Liquids User Exercises
HCA Liquids User Exercises v1.0 (04.03.2014) Page 13 of 13
You can save your current settings for future use by using the buttons in the Default
Settings section:
Enter a name for your settings and then click the Save/Update button. The name you
chose now appears in the Default Settings list.
Note: You can modify a saved setting by double-clicking on it (click Yes on the
message window that appears), making your changes, and clicking the
Save/Update button. To delete a saved setting, click it and then click the Delete
button.
Step 8: Start the Analysis
1) Click Run Analysis
Click the Run Analysis button, located in the extreme bottom-right of the window, to
start the analysis process.
The Enter HCA Analysis Name window displays:
2) Specify a name for the HCA analysis
Enter a name for this particular HCA analysis and then click OK.
The Status processing window displays.
3) Click OK
When the analysis is completed, a confirmation message displays. Click OK to
confirm the completion of the process.