Page 2: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

What is anaesthesia ?

* Art or Science * Loss of Awareness * Loss of all Sensation (Pain, Temp., Position) * The process should be reversible * Modern general anaesthesia (TRIAD)

Unconsciousness

BalancedAnaesthesia

Analgesia Muscle Relaxation

Page 3: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Automatic Control in Anaesthesia

The main problem : measurement of clinicalsigns for on-line input to the system

* Muscle Relaxation: direct measurement: EMG control algorithm: PID, GPC, FLC, etc.

* Unconsciousness (Depth of Anaesthesia): no direct measurement

Page 4: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Indirect measurement:

(a) SAP, HR (b) SW, LA, PR, MO (c) End Tidal of AA, MAC, (d) Plasma concentration (e) Heart Rate Variability (f) Brain Signals EEG, AEP, SEP Etc.

Page 5: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Control algorithm: (a) PID, GPC (b) Fuzzy Logic (c) Neural Networks Etc.

Interpretative algorithm: (a) Aperiodic Analysis (b) Fourier Transform Analysis (c) Auto-Regression Algorithm (d) Fuzzy Logic (e) Neural Networks Etc.

Page 6: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

* Analgesia (Pain Relief)

no direct measurement subjective indirect measurement brain signals (i.e. SEP)

Postoperative conditions Cancer pain PCA (Patient-Controlled Analgesia)

Page 7: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Fuzzy logic features:

* Can reason with imprecise data * Leads to "soft computing" * Concept of "machine IQ" * Cope with non-linear, complex, unknown processes * Link with neural networks * Link with genetic algorithms * Lack of stability theory

Page 8: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Anaesthetists use “rules of thumb” “imprecise, personal rules”Ex: IF T1% is greater than the set point by a LARGE AMOUNT THEN set the atracurium infusion rate to a HIGH LEVE

This rules contains imprecise terms: a LARGE AMOUNT a HIGH LEVEL

Page 9: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

In Clinical Engineering:

1988 Linkens & Mahfouf (UK) Muscle Relaxation (Simulation) 1988 Sheppard & Ying (USA) MAP, SNP (Sodium Nitroprusside) 1992 Hacisalihzade et al. (Switzerland) MAP, Isoflurane 1994 Tsutsui & Arita (Japan) SAP, Enflurane 1995 Shieh et al. (UK) DOA, Propofol & Isoflurane

Page 10: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

1995 Zbinden & Hacisalihzade (Switzerland) MAP, Isoflurane, Human1996 Schaublin & Zbinden (Switzerland) End Tidal CO2, Ventilation (Fre., Vol.)1996 Curatolo & Zbinden (Switzerland) Fi (Iso.) & O2 conc., min. flow 1996 Mason et al. (UK) Muscle relaxation (Atracurium)1996 Shieh et al. (ROC) Muscle Relaxation (Atr.)1997 Mason et al. (UK) Muscle relaxation (Atracurium), SOFLC1997 Shieh et al. (ROC) Muscle Relaxation (Miv.)

Page 11: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

1998 Shieh et al. (ROC) Unconsciousness (Desflurane)

2000 Shieh et al. (ROC) Muscle Relaxation (Roc.)

Page 12: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Computer Monitoring and Controlin Muscle Relaxation

Page 13: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Introduction:

Short-acting non-depolarizing relaxants

Advance of modern computer technology

The purpose of this approach:

Small, handy and easy to use

Page 14: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Clinical Methods of Nerve Stimulation:

• Single-twitch Stimulation• Train-of-four Stimulation • Double-burst Stimulation • Tetanic Stimulation

Page 15: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 16: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 17: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 18: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Recording of Evoked Responses:

• Mechanical Responses• Electromyographic Responses • Accelerative Responses

Page 19: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Mechanical Responses:

Page 20: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Electromyographic Responses:

Page 21: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 22: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Accelerative Responses:

Page 23: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 24: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Requirements for the ideal neuromuscular blocking agent: 1. Non-depolarizing mechanism of action 2. Rapid onset of action 3. Short duration of action 4. Rapid recovery 5. Non cumulative 6. No cardiovascular side effects 7. No histamine release 8. Reversible by cholinesterase inhibitors 9. High potency 10. Pharmacologically inactive metabolites.

Page 25: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 26: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 27: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

The characteristics of Atracurium, Mivacurium and Rocuronium : 1. An aqueous for intravenous injection 2. Conc.: 10 (A), 2 (M), 10 (R) mg/ml 3. Non-depolarizing neuromuscular blocking drug 4. Onset time: 1.5 (A), 2.5 (M), 1 (R) min 5. Duration of action: Intermediate (A, R), Short(M) 6. Recovery: Intermediate (A, R), Rapid (M) 7. Cardiovascular effect: Yes(A, M), No (R) 8. Histamine release: Yes (A, M), No (R) 9. Metabolites.

Page 28: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Atracurium: Laudanosine (Central effects) (Cisatracurium: 1/3 Laudanosine)

Mivacurium: Enzyme Rocuronium : Liver, Kidney (Vecuronium: causing prolonged block) The Wellcome Foundation (A, M): J. Savarese Organon Teknika ( R )

Page 29: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 30: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 31: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Hierarchical Monitoring of EMG via filters * Built-in filter: Noise High Frequency Disturbance * Pharmacological filter: T4/T1 T2/T1, T3/T2, T4/T3 * Median filter x1, x2, x3,x4,x5 => Median Value Example: T1% 7, 17, 11, 10, 8 => 10

Page 32: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Hierarchical Monitoring and Fuzzy Logic Control

EmergencyTable

CoarseTable

Self-Tuning

FLC(Fine Table)

MedianFilter

PharmacologicalFilter

InstrumentFilter

EMGPatient EMG (F)S.P.+ E,CE

-

Page 33: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 34: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 35: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

(a)

020406080

100

0 360 720 1080

Time (Samples of 10 s)

T1(%

)

(b)

0

20

40

60

80

100

0 360 720 1080

Time (Samples of 10 s)

T1(%

, F

)

Manual Control for Atracurium:

Pt. Weight: 54 kg;Pt. Age: 34 yr;Sex: Female;Clinical Diagnosis: Lipoma retroperitoneal tumourOperation: Debaking

Page 36: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

0

50

100

0 180 360 540 720

Time (Sec.*10)

T1 (F

)

00.5

11.5

2

0 180 360 540 720

Time (Sec.*10)

MIR

-0.2-0.100.10.2

STP

01020304050

0 180 360 540 720

Time (Sec.*10)

IR01020304050

DC

Automatic Control for Mivacurium:

Pt. Weight = 58.5 kg;Pt. Age = 45 yr;Sex : FemaleClinical Diagnosis: CPS;Operation: FESS

Page 37: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Summary: (Muscle Relaxation) Clinically, this system was useful:

* Provided stable surgical operating conditions

* Minimized the amount of neuromuscular

blocker required by each patient

* Reduced the need for the anaesthetist to

spend time controlling neuromuscular block

* Allowed reliable antagonism of

neuromuscular block at the end of surgery

Page 38: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Automatic Control of Anaesthesiawith Desflurane Using Hierarchical

Structure

Page 39: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Unconsciousness (Depth of Anaesthesia ) :

no direct measurement indirect measurement:

SAP, HR SW, LA, PR, MO End Tidal of AA, MAC Plasma concentration Brain Signals (i.e., EEG, AEP, SEP) Etc.

Page 40: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

PATIENT

COMPUTER

SAP, HR, AA

Supervision by Anaesthetist

Datex AS/3

Control Box

Stepping Motor

Vaporizer

An Automatic Control System for Inhalational Anaesthesia

Page 41: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 42: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 43: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 44: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 45: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

A Clinical Trial (Unconsciousness)

Pt. Weight: 52 kg;Pt. Age: 19 yr;Sex: Female;Clinical Diagnosis: CPSOperation: FESS

0

50

100

150

200

0 50 100 150 200 250

Time (Sec.*30)

BP

(mm

Hg)

0

25

50

75

100

0 50 100 150 200 250

Time (Sec.*30)

Con

tuol

ler O

utpu

t

0

4

8

12

0 50 100 150 200 250

Time (Sec.*30)A

A (%

)

Page 46: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Modelling of Anaesthesia for Unconsciousness

Neural Networks

PatientModel

FiaaGender

AgeWeight

HRSAP

BIS

FuzzyController

Etaa

Rule-basedFuzzyModel

Neural NetworksVaporizer Model

FuzzyModel

FuzzyModel

SDOA

PDOA

DOA

Page 47: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Automatic Monitoring and Control in the Operating Theatre

AS/3 MonitorVaporizer ControlNoteBook

Graseby 3500

Aspect 1050

Page 48: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Monitoring EEG Signals of AwakeningDuring Propofol Anaesthesia

Page 49: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Problem: Detecting awareness In Paralyzed patient remained unsolved. Cardiovascular signs are not always reliable.

The aim of this study: To investigate the changes of different waves in EEG signals during different anaesthetic stages To identify the parameter for early detection awareness

Page 50: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 51: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 52: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 53: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 54: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 55: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Summary: (Awareness, EEG)

* The Beta, Alpha, Sub-alpha and Delta waves have significant differences during different stages

* The mean(SD) of the mean percentage of beta wave in 8 patients’ EEG signals during induction, maintenance, and recovery stages were 73.43(4.84)%, 42.26(8.31)% and 72.14(7.19)%, respectively.

Page 56: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Detection of Awareness

Method I: Questions asked during structured interview 1. What was last thing you remember before you went to sleep for your operation ? 2. What was the first thing you remember after your operation ? 3. Can you remember anything in between these two periods ? 4. Did you dream during your operation ? 5. What was the worst thing about your operation ?

Page 57: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Method II:Tape-recorder using earphones(church bells, farmyard noises, light orchestral music, piano music, market voices, bird song, pop music and choir music).

Method III:Forearm Technique

Page 58: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

EEG signals in Drowsy States

Page 59: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Fundamental EEG Signals Research in Drowsy States

EEG electrode positions on the scalp

Page 60: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Harmonie Digital EEG Systems- An Awake State

Page 61: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Harmonie Digital EEG Systems- A Sleep State

Page 62: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Brain Signal Processing and Analysis

1. Aperiodic Theory

2. 95% Spectrum Edge Frequency

3. Variety Analysis

N

k

nk

nk

nk

nk

n Nt

WWWWC

12

)1()()()1()( )(

)()(

unit: 0.01 μV/

N : sampling number

n : segment number

: variety

: brain signals potential

nkW

2sec

)(nC

Page 63: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

A Genetic Fuzzy System Applied in Analysis of Brain Signals

Signal P.

Rules

CCD

Crossover

Inf. Eng.Fuzz. Defuzz.

Reproduction

Mutation

Encode

Decode

Output E

R

R

O

R

A Best Group

of Rules

Yes

Volunteers

Image

Brain Signals

Experts

CL

No

Page 64: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Commercial Products for Mechanical, Sound, and Electrical Stimulation for Preventing Drowsy States

(a) Pulse massage (b) Sound wave that move your mind (c) Nod suppresser

Page 65: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

The Prototype of Monitoring and Controlling Brain Signals for Deterring Driver Drowsiness

Page 66: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Pain Control and Evaluation using Fuzzy Logic Control

& Patient Controlled Analgesiafor Shock Wave Lithotripsy

Page 67: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Problem: Detecting pain Subjective & no direct measurement Clinical: Visual Analogue Scale (VAS)

Where will cause the pain: Endoscope Operating room:ESWL; Prostate Post OP.: PCA 300/2000 per month $ 4,500 , nearly 1.5 million NT ICU Cancer Pain

Page 68: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

How to study the Pain: Provide a constant pain Not too long or too short for experiment Easy communication with the patient

Why do we choose the ESWL to study the Pain ? Provide a constant pain using ultrasonic waves Used for destroying calculi in the upper urinary tract and gallstones (OP time < 1 hr) Patients are consciousness

Page 69: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Current Drugs:

1) Pain killer: fentanyl (0.0785 mg/ml) Preventing vomit: droperidol (2.5 mg/ml) 2) Loading dose: Fentanyl: 2 ml; droperidol: 0.25 ml 3) NA to add further dose of fentanyl if the patient complain. 4) If NA can not handle, call anaesthetist.

Page 70: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

PCA (Patient-Controlled Analgesia) : Management of pain in (1) postoperative patients (2) cancer patients Function: (1) administer small bolus doses of pain-control drugs (2) at fixed intervals (3) controlled by the patient with the push of a button

Pain Control Using PCA

Page 71: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

PCA in ESWL Drug: alfentanil Conc.: 0.5 mg/ml Loading dose: 0.5 ml Bolus if needed: 0.4 ml Lockout time: 1 min Infusion rate: 120 ml/hr

Page 72: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Fuzzy Logic Control:– bolus + continuous infusion

– pain feedback controlled by patient

– input variable : pain, chan_pain

– output variable: chan_inf

Pain Control Using Fuzzy Logic Control

pain: BP, SP, ZPchan_pain: NB, NS, ZR, PS, PBchan_inf: BI, SI, ZO, SD, BD

Page 73: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Syringe Pump

Notebook(Monitoring & Control)

Catheter

Pain signals controlled by the patient

Patient

Patient-Controlled Analgesia

Page 74: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 75: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 76: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 77: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 78: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Clinical Evaluation of the Pain

0 10 20 30 40 50 60 70 80 90

0

10

20

30

40

50

60

70

80

90

100(e)

Drug Conc. (ng/ml)

Pai

n I

nte

nsi

ty

0 10 20 30 40 50 60 70

0

10

20

30

40

50

60

70

80

90

100(e)

Drug Conc. (ng/ml)

Pai

n I

nte

nsi

ty

Current Commercial Machine Fuzzy Control Machine

Page 79: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 80: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia

Prostate Research

Page 81: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 82: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 83: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 84: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 85: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 86: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia
Page 87: Hierarchical Monitoring and Fuzzy Logic Control in Anaesthesia