High Range Illuminance Meter Calibration_044high

Embed Size (px)

DESCRIPTION

High Range Illuminance Meter Calibration

Citation preview

  • Ttulo del trabajo Title of paper Autor/ es Author/s

    Afiliacin/ es del autor/ es Affiliation/s of the author/s

    Direccin principal Mail adress

    Telfono, fax. e-mail de la persona de contacto Phone, Fax number and e-mail adress of the contact person

    High Range Illuminance Meter Calibration

    Using Substitution Method

    National Metrology Centre, Standards, Productivity and Innovation

    Board

    XU Gan, LIU Yuanjie, TAN Hwee Lang and WANG Xiaojuan

    1 Science Park Drive, Singapore 118221

    Tel: 65 67739874, Fax: 65 67739804

    e-mail: [email protected]

  • Abstract

    The measuring range of many commercial illuminance meters is well above 10,000 lx.

    Calibration at such high illuminance levels cannot be done directly through the inverse

    square law by using a normal luminous intensity standard lamp, which can only produce

    an illuminance level up to a few thousand lx at the minimum acceptable distance (~ 0.5

    metre). Some laboratories use lamps of very high power (up to 10 kW) for such

    calibrations but this requires additional space and proper heat isolation, in addition to

    the much larger power supply needed by such lamps.

    A simple and low-cost method has been developed for high range illuminance meter

    calibration based on a substitution-superposition technique. In this method, a luminous

    intensity standard lamp is first set up at a known distance to calibrate the meter at a

    point of low illuminance level. The light from the standard lamp is then blocked and the

    meter is illuminated by an auxiliary lamp, the position and orientation of which are

    adjusted so that the meter reading remains unchanged. The meter is then calibrated at

    twice the initial level by using illumination from both the standard and auxiliary lamps

    at the same time. Once this is done, the standard lamp is blocked and the auxiliary lamp

    is again adjusted to give the same meter reading at twice the initial level, thereby

    substituting the illuminance level of the standard lamp. By repeating the steps, the meter

    can be calibrated at any desired level, limited only by the capacity of the auxiliary lamp

    As the auxiliary lamp is used only as a substitute, similar to a capacitor in a holding

    circuit in electronics, there is no special requirement on the type of lamp used except for

    its stability. By this method, high illuminance can be achieved by using a focused beam

    from a tungsten halogen lamp of only 50W at 12V dc with a collimator. Together with a

    normal standard lamp, calibration of up to 50,000 lx can be performed. The relative

    calibration uncertainty is similar to that of any low range illuminance meter calibration.

  • Introduction

    Illuminance meters are normally calibrated using luminous intensity standard lamps

    placed at several known distances for required illuminance levels according to inverse

    square law. For inverse square law to be valid the standard lamps must fulfil the

    conditions of a point source. i.e. the dimensions of the lamp filament shall be negligible

    as compared with the distance between the illuminance meter and the standard lamp.

    Such requirement sets a limit to the minimum distance (and hence the maximum

    illuminance level) that can be used for illuminance meter calibration. For example, for a

    1200 W tungsten halogen lamp of a filament size 10x10 mm running at 2856 K, the

    maximum illuminance level it can produce is only about 7,000 lx at a minimum distance

    of 0.5 m.

    However the measuring range of many commercial illuminance meters is well above

    10,000 lx. Calibration at such high illuminance levels cannot be done directly through

    the inverse square law by using a normal luminous intensity standard lamp. To

    overcome this problem, some laboratories have developed special standard lamps of

    very high power (up to 10 kW), but this requires additional space and proper heat

    isolation, in addition to the much larger power supply needed by such lamps. Others use

    photometer-based method together with a special and complex source system that has a

    spectral power distribution close to the CIE Illuminance A. Both methods require

    substantial resources and the costs of establishment and maintenance of such calibration

    facilities are not practical for most laboratories.

    This paper presents a simple and low-cost method for high range illuminance meter

    calibration based on a substitution-superposition technique. The calibration setup,

    measurement procedure and working principle are illustrated. The additional

    measurement uncertainties associated with this method are discussed.

    Calibration setup

    The setup for the proposed high range illuminance meter calibration is illustrated in

    Figure 1.

    Figure 1 Setup for high range illuminance meter calibration

  • The illuminance meter to be calibrated and a luminous intensity standard lamp (Model :

    6895Z made by Philips, rated 1200W tungsten halogen lamp running at a correlated

    colour temperature of 2856K) are set up on the main optical bench with a known

    distance of 0.7 m between the measurement plane of the illuminance meter and the

    reference plane of the standard lamp. This is exactly the same as one would use a

    luminous intensity standard lamp to calibrate an illuminance meter by the inverse square

    law.

    An auxiliary tungsten halogen lamp is set up off-axis on a mini-bench at a distance of

    about 0.5 m away from the head of the illuminance meter to be calibrated. This

    auxiliary lamp has a rated power of 50W at 12V dc (Model : DECOSTAR 51, 41870 SP

    made by OSRAM) and has a dichroic reflector which collimates its light beam to a

    divergence angle of about 10 degree. The dichroic reflector reflects visible light but

    allows most of the infrared radiation to pass. This reduces the heating effect of the

    tungsten halogen lamp on the head of the illuminance meter to be calibrated.

    The auxiliary lamp is mounted on a rotating holder with which the direction of the

    collimated light beam can be adjusted freely. With the collimating reflector the 50 W

    auxiliary lamp is capable of producing a high illuminace reading of more than 50,000 lx

    when its light beam is pointed directly at the head of the illuminance meter.

    Calibration principle and procedure

    Set the calibration system up as illustrated in Figure 1. Arrange the auxiliary lamp and

    its baffle properly so that they do not block the light from the standard lamp reaching

    the head of the illuminance meter during the calibration. The principle and detailed

    calibration procedure of substitution-superposition technique is explained in Figure 2.

    The first step is to calibrate the illuminance meter at low illuminance level produced by

    the standard lamp. This is done by block the direct illumination of the auxiliary lamp

    and illuminate the head of the illuminance meter with the luminous intensity standard

    lamp alone, and take the meter reading Y1. Y1 is the illuminance meter reading for

    illuminance level of x + D, where x is the net direct illuminance value from the

    standard lamp calculated based on inverse square law and D is the dark signal

    (including stray light) when the direct illuminance from both lamps are blocked.

    Next is to substitute the illuminance from the standard lamp with an equivalent

    illuminance from the auxiliary lamp. This is done by block the standard lamp and

    illuminate the meter only with the auxiliary lamp, and adjust the angle of the light beam

    of the auxiliary lamp so that the illuminance meter reading exactly equals Y1, indicating

    that the auxiliary lamp now provides a substitute illuminance equal to the illuminance

    produced by the standard lamp (in terms of the illuminance meter response).

  • Figure 2 Calibration procedure using substitution-superposition technique

  • Strictly speaking this equivalence of substitution is valid only if the stray light level

    remains the same after the adjustment of the pointing angle of the auxiliary lamp. This

    condition can be verified by blocking both the lamps and measure the dark signal

    again after the adjustment of the auxiliary lamp. Experiments showed that the dark

    signal varies less than 0.1% of x under the calibration setup described in the last section.

    This variation can be treated as an uncertainty source in the calibration uncertainty

    budget.

    Once the substitution is established, let both the standard lamp and the auxiliary lamp

    illuminate the head of the illuminance meter simultaneously and take the meter reading

    Y2. Y2 is the illuminance meter reading for illuminance level of 2x + D. This is the

    superposition process.

    Repeat these steps n times will increase the illuminance to the desired value:

    nx + D, n =1,2,..

    The validity of this substitution-superposition step requires that the dark signal

    remains constant and the output of the auxiliary lamp is stable. Under the calibration

    setup, a 1200 W standard lamp with a luminous intensity of 1715 cd placed at a distance

    of 0.7 m away from the illuminance meter, x will be 3500 lx, and the calibration level of

    nx + D will reach 50,000 lx with a n of less than 15.

    Uncertainty analysis

    As can be seen from the calibration setup, the auxiliary lamp has a collimating reflector

    and is placed off-axis. The illuminance it produces on the detector head of the

    illuminance meter is therefore not uniform. And because of the dichroic reflector the

    spectral power distribution of the illumination from the auxiliary lamp cannot be treated

    as CIE Illuminant A. These are two important factors to be considered for normal

    illuminance meter calibration. However they are not critical in the proposed method

    because the illumination from the auxiliary lamp is used only to produce a response on

    the illuminance meter that is equal to the response of the same illuminance meter to the

    known illuminace produced by the luminous intensity standard lamp. Although the

    actual average illuminance produced by the auxiliary lamp might be different from the

    illuminance produced by the standard lamp, they still can be used to substitute to each

    other as long as the responses produced by them on the illuminance meter to be

    calibrated are the same.

    There are two factors affecting the validity of this substitution-superposition technique.

    The first is the variation of the stray light level when the angle of the auxiliary lamp is

    altered. This variation can happen at random directions or at the same direction during

    the process of the substitution-superposition technique. In the worse case if all the steps

    the variations are at the same direction, they will be accumulated. E.g. if the stray light

    increases by 0.1%x at each step of substitution-superposition, then the total increase of

    stray light after n steps will be 0.1%x multiplied by n. Since the calibrated illuminance

  • level is also increased to nx after n steps, the relative increases in stray light will still be

    0.1%.

    The second factor affecting the calibration uncertainty of the substitution-superposition

    technique is the stability of the output of the auxiliary lamp. The extent of this factor

    depends on the time required for the whole calibration process. Experiments shown that

    the total time required for a 15-step calibration is less than 30 minutes. The maximum

    drift of the auxiliary lamp during a 30 minutes time period is assessed to be less than

    0.1%.

    Most of the uncertainty components associated with the first-step calibration, i.e. the

    direct illuminance calibration by the standard lamp based on reverse square law, will be

    accumulated in absolute values. But their relative values against the total illuminance

    level will remain unchanged just as the worse case of the stray light variations

    discussed.

    Table 1 shows the uncertainty budget of the high range illuminance meter calibration

    using the substitution-superposition technique under the calibration setup shown in

    Figure 1. Among the 12 uncertainty components, u(1) to u(10) are uncertainty

    components associated with the first-step calibration, u(11) and u(12) are additional

    uncertainty components due to the substitution-superposition process. The relative

    expanded uncertainty for the direct illuminance meter calibration at low range using the

    standard lamp is 1.11%. The relative expanded uncertainty for the high range

    illuminance meter calibration using the substitution-superposition technique is 1.13%.

    Conclusion

    A new approach based on a substitution-superposition technique is proposed to

    calibrated illuminance meter at levels much higher than what can be achieved by

    commonly available luminous intensity standard lamps. The advantages of this

    approach are its simplicity and low cost. No specially designed, complex and high

    power source is required. By using a 50W tungsten halogen lamp as an auxiliary source,

    which is commercially available at very low cost, illuminance meters can be calibrated

    up to 50,000 lx illuminance level. The relative calibration uncertainty is similar to that

    of any low range illuminance meter calibration. More than one auxiliary lamp can be

    used if even higher range is required, and the same principle still applies.

    References 1. CIE Publication No. 53, Methods of Characterizing the Performance of Radiometers and

    Photometers (1982)

    2. CIE Publication No. 69, Methods of Characterizing Illuminance Meters and Luminance

    Meters (1987)

  • 3. Ohno, Y., High Illuminance Calibration Facility and Procedures, Paper for IESNA Annual Conference (1997)

  • Uncertainty components

    u(xi)

    Description Value of u(xi)

    (%)

    Sensitivity factor Ci

    Probability Distribution

    Multiplier k

    Standard uncertainty ui(y) (%)

    Degree of freedom

    u(1) Calibration uncertainty of the luminous intensity standard lamp (type B)

    0.8 1 normal 2 0.40

    u(2) Long-term stability of the standard lamp (B)

    0.5 1 rectangular 1.73 0.29

    u(3) Calibration uncertainty of multimeter for lamp current measurement (B)

    0.005 6.25 rectangular 1.73 0.02

    u(4) Calibration uncertainty of standard shunt resistor for lamp current (B)

    0.005 6.25 rectangular 1.73 0.02

    u(5) Uncertainty due to temperature change of standard shunt resistor (B)

    0.01 6.25 rectangular 1.73 0.04

    u(6) Uncertainty of distance measurement, 1mm for a distance of 0.7 m (B)

    0.14 2 rectangular 1.73 0.16

    u(7) Alignment (cosine) error, 2 degree from the normal (B)

    0.06 1 rectangular 1.73 0.03

    u(8) Uncertainty due to the resolution of the meter under test

    0.1 1 rectangular 1.73 0.06

    u(9) Repeatability of illuminance meter readings (A)

    0.1 1 normal 1 0.10 n-1

    u(10) Uncertainty due to departure from inverse square law, 1mm for 0.7m (B)

    0.14 2 rectangular 1.73 0.16

    u(11) Uncertainty due to extra stray light 0.1 1 rectangular 1.73 0.06

    u(12) Uncertainty due to short-term drift of the auxiliary lamp

    0.1 1 rectangular 1.73 0.06

    u(y) Combined standard uncertainty 0.565

    U(Y) Expanded uncertainty 2 1.13

    Table 1 Uncertainty budget of high range illuminance meter calibration. u(1) to u(10) are uncertainty components associated with

    the first-step calibration, u(11) and u(12) are additional uncertainty components due to the substitution-superposition process.