22
High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner, In collaboration with Volker Dangendorf from PTB Braunschweig, Neutron Radiography group

High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Embed Size (px)

Citation preview

Page 1: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

High resolution imaging with MCP detectors using delay line anodes

Ottmar Jagutzki, Uwe Spillmann, Achim Czasch

Horst Schmidt-Böcking, Reinhard Dörner, …

In collaboration with

Volker Dangendorf from PTB Braunschweig, Neutron Radiography group

Page 2: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,
Page 3: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

The helical wire delay-line anode

Fast timing electronics high rate, multi-hit 5(7) electronic channels only easy read-outCenter-of-mass averaging high position resolution

time resolution: < 1 ns

Sobottka and Williams 1988

DLD40

80 mm “Hexanode” delay-line detector (patented)

Our typical resolution: 40-50 μm rms

Page 4: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,
Page 5: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Electronic readout:

Page 6: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Time-to-Digital-Converter:

- (old) Wilkinson-type analog-digital conversion (like a TAC): high resolution but a slow digital conversion

- “counter”-type fast (no!) conversion: resolution limited by clock rate Pretty high resolution and fast read-out

TDC8HP: PC-controlled multi-hit TDC with 8 channels, 25 psec LSB, stableread-out speed: 400 kHz now, up to 2 MHz planned (< 10 ns dead-time between hits)

“TDCHP” chip developed by CERN

Page 7: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

How to detect visible/near-UV photons with that ?

Page 8: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Photek 75 mm image intensifier with RoentDek DL80 anode

NASA test mask

0.15 mm FWHM

0.075 mm pixel (rms)

Now: < 40 micron rms

5 mm grid constant, 1 mm obstacle

Page 9: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Counting/Imaging near-UV and visible photons with delay-line read-out

DL40-PMT

Quartz windowPhoto cathode

MCP stack

DL40 anode(or bigger)

works fine, but

- difficult to build- not very robust (?)

- size up to 150 mm possible?

Page 10: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Counting/Imaging near-UV and visible photons with delay-line read-out

via image charge pick-up: the RS-PMT

Quartz windowPhoto cathode

MCP stack

Resistive screen

pickup electrode(e.g. delay-line)

Patented technique

- easy, robust design of detector head- size up to 40 mm, may be bigger- standard (multi-layer) PCB pickup electrodes

Image charge read-out: Battistoni et al. 1982

Page 11: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Counting/Imaging near-UV and visible photons with reconfigurable read-out

Quartz windowPhoto cathode

MCP stack

Resistive screen

otherpickup electrodes:

Wedge&Strip, Pixel,…

Patented technique

Read-out technique can easily be reconfigured using the same detector head.

Page 12: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Counting/Imaging VUV photons or particles with reconfigurable read-out

“open RS-PMT”

MCP stack

Resistive screen

pickup electrodes:

Delay-line, Wedge&Strip, Pixel,…

Patented technique

Image charge read-out has some advantages over charge collection for some anodes: Beneficial even for open-face detectors

Page 13: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Delay-line read-out of RS-PMT:

Multilayer PCB

Front and rear side “antennas” (Eland 1994)

connected to delay-line:

(Berkeley group) LC-delay-line

Page 14: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

LC-DL50 on open RS-PMT40, irradiated with α – particles, mesh obstacle 80 micron

Page 15: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

LC-DL50 on open RS-PMT40, irradiated by ions with 500 kHz, mesh obstacle 80 micron

Page 16: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

LC-DL50 on open RS-PMT40, irradiated by ions with 500 kHz, mesh obstacle 80 micron

8 x 8 mm linear scale log scale

Page 17: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

50 mm LC Hex-DL on a 4-layer PCB

Hexanode for RS-PMT:

- multi-hit (simultaneous photon pair detection is possible): Compton telescope?- intrinsic linearity correction (no test mask necessary): 3 combination for determining x- and y-coordinates from u, v, w coordinate system

Example for x-layer: x = u = - (v + w) (over-determination) (vector addition)

u

v w

x = u

y =

(w

- v

)/√3

But how can this help correcting non-linearity ?

Page 18: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Linearity correction of LC-delay-line: (no mask needed)

of RS-PMT: 16 18 20 220

20

40

60

80

100

120

140

rela

tive

char

ge /

%

position / mm

Experiment 4 (d=1.0 mm)

Gauss Fit X

c=18.9 mm

=0.86 mm

2,0 mm

Page 19: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,
Page 20: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Summary:

- Large (open-face) MCP delay-line detector with up to 3000x3000 pixel

- Resistive screen PMT or open MCP detector for counting photons (might qualify for space)

- redundant triple-layer delay-line anode read-out with intrinsic linearity correction

- read-out anodes can easily by swapped

work to be done:

- position resolution needs to be improved for the 40 mm and 25 mm formats

- no experience with flight mission

- not much experience with photon counting at all !

Page 21: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Applications:- FLIM (fluorescence life-time microscopy) - NEURRAD (element sensitive neutron radiography) in cooperation with Volker Dangendorf, PTB Braunschweig

Thanks to Jürgen Barnstedt from Tübingen for borrowing us his RS-PMT!

Obj

ect

Page 22: High resolution imaging with MCP detectors using delay line anodes Ottmar Jagutzki, Uwe Spillmann, Achim Czasch Horst Schmidt-Böcking, Reinhard Dörner,

Projection of grid mask 70/600micron, resolution 35 micron (1:700)

Results of Barnstedt et al. (1998)

25mm RS-PMT (Proxitronic) - bi-alkali photo-cathode - 4-jaw Wedge&Strip anode