78
.v,^v.::^...j.^....aEl: by Elediic Means 1911 4

High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

.v,^v.::^...j.^....aEl: by Elediic Means

1911

4

Page 2: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 3: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 4: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

Digitized by the Internet Archive

in 2013

http://archive.org/details/hightemperaturemOOcarr

Page 5: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

HIGH TEMPERATURE MEASUREMENT

ELECTRIC MEANS ^'^^

BY

MAURICE LeROY CARRB. S. UNIVERSITY OF ILLINOIS, 1905

THESIS

Submitted in Partial Fulfillment of the Requirements for the

Degree of

ELECTRICAL ENGINEER

m

THE GRADUATE SCHOOL

OF THE

UNIVERSITY OF ILLINOIS

1911

Page 6: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 7: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

nil

UNIVERSITY OF ILLINOIS

THE GRADUATE SCHOOL

HEREBY RECOMMEND THAT THE THESIS PREPARED UNDER MY SUPERVISION BY

^^^^^ ^^^^«rr^^ ^^^^^-r-lr^rX,

BE ACCEPTED AS FULFILLING THIS PART OF THE REQUIREMENTS FOR THE

DEGREE OF

Recommendation concurred in:

Committee

on

Final Examination

197576

Page 8: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

uiuc

Page 9: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

HIGH TSMPERATURK MEASUREMENT

by

ELECTRIC MEANS.

CONTENTS.

I. INTRODUCTION

II. GENERAL

III. RESISTANCE PYROMETRY

In the thesis of Mr. Karr one or two more slight

would, I think, increase the clearness. On pap:e 2 a he of

refers to manufacturers in the United States, would it not

to add this.

On page 3, paragraph marked B, which has no real

Very truly yours.

Professor of Mi!ning Engineering.

rxeuj.sxon oi Measurement 263. Protection of Element 274. Operation 275. Recording Instruments 276. Permanent Installations 277. Cost of Apparatus • 288. Cost of Up Keep 289. Tabular Comparison of Methods 29

IX. CONCLUSION 30

X. BIBLIOGRAPHY 31

XI. MANUFACTURERS OF PYROMETRIC APPARATUS 32

Page2

3

d.

additions

course

be well

anteoedant

.

Page 10: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 11: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

HIGH TEMPERATURE MEASUREilENT

by

ELECTRIC MEANS,

CONTENTS.

PageI, INTRODUCTION 2

II. GENERAL 3

III. RESISTANCE PYROMETRY 41, Theory 42, Apparatus 6

IV. THERMO-ELECTRIC PYROMETRY 121. Theory 122. Apparatus -• ^~ 15

V. OPTICAL ANT) RADIATION PYROMETERS • 19

VI. INDUSTRIAL APPLICATIONS 211. Resistance Method 212. Thermo-Elcotric Method 223. Optical and Radiation Method 22

VII. CALIBRATION OP PYROMETERS 23

VIII. COMPARISON OF METHODS 261. Temperature Range 262. Precision of Measurement 263. Protection of Element 274. Operation 275. Recording Instruments 276. Permanent Installations 277. Cost of Apparatus -^ 288. Cost of Up Keep 289. Tabular Comparison of Methods 29

IX. CONCLUSION 30

X. BIBLIOGRAPHY 31

XI. MANUFACTURERS OF PYROMETRIC APPARATUS 32

Page 12: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 13: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

I. INTRODUCTION.

The accurate measurement of high temperatures hascome to be a very essential feature of many industrial pro-cesses. The pyrometer makes it possible to reproduce withprecision any temperature that has been found to yield thebest results in operations requiring careful heat treatment.In the making of tinted glass, for example, the proper regu-lation of the temperature of the furnace is necessary to pre-vent burning out the color. Instead of depending upon theworkman to judge of the condition of the material with hiseye, the manufacturer gives instructions to maintain the heatat a certain temperature, knowing then that the output willbe of a certain uniform grade. In this manner the variablehuman element is eliminated and waste prevented. The temper-ing of tools, the burning of ceramic products, the operationof gas producers, the manufacture of iron and steel and theannealing of castings are a few examples of processes thatmay be carried on to better advantage with the aid of refinedmethods of measuring high temperatures. The recording pyro-meter is rapidly coming into use as an efficient means of su-pervising processes requiring heat treatment.

The methods of measuring high temperatures have beengreatly improved in the last few years and the making of pyro-meters and accessories engages the attention of about a dozenmanufacturers. The development of the science of pyrometryhas been confined largely to the refinement of a few methodsrather than to the evolution of new ones. Leaving out ofconsideration the expansion thermometer, the ranges of use-fulness of present day methods of measuring temperatures maybe defined as follows:-

Absolute zero to 1600°C. (3000°P) Electrical MethodsAbove 900^0. (I5000p) Radiation and Optical Methods.

While the discussion of the electrical methods con-stitutes the principal subject matter of this thesis, briefreference is made to a number of the radiation and opticalmethods, for the reason that electricity is indispensable intheir application.

Page 14: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 15: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

II. GENERAL

Electricity furnishes two valuable means of measur-ing high temperatures. One method turns to account the in-crease in resistance that most substances undergo when heatedand the other method depends upon the fact that an electromo-tive is developed when heat is applied to the junction of twodissimilar metals. Both methods are suitable for use oververy wide temperature ranges extending from the region of ab-solute zero up nearly to the melting point of the substancesemployed. These methods may be designated the

Resistance Method, or Resistance Pyrometry.Thermo-Electric Method, or Thermo-Electric Pyrometry.

The resistance method involves the measurement of thechange in resistance of a properly protected platinum wiresubjected to the temperature to be measured. Platinum ispractically the only substance used for this purpose for thereason that its melting point is high and its other propertiesare satisfactory. Resistance thermometers are suitable foruse at all tem.peratures from - 200OC. to 12000G. (2200°?) .

The method is capable of great refinement and furnishes themost exact means of temperature measurement, or control, knowntoday. The bolometer, an instrument involving the resis-tance principle, is capable of indicating a change in tempera-ture of one ten millionth of a degree G.

The thermo-electric method involves the measurementof the minute electro motive forces developed at the hotjunctions of the thermo-electric couples. Platinum and itsalloys when properly protected are the most satisfactory me-tals, every thing considered, for measuring- temperatures bythis means. Their useful temperature range is 300^0.- 1600°C, Other metals and alloys extend the range down to theregion of absolute zero and up to 21000C, (38000P).

Both methods outlined require more or less delicateapparatus and careful protection of the element. Hot gasesexert marked deteriorating effects upon the metals employedwhich destroys the value of calibrations and makes measure-ments subject to large errors.

The radiation pyrometer is a combinetion of somemeans for concentrating radiant energy and a delicate resis-tance thermometer, or a thermo-couple. The optical pyro-meters are essentially photometers. The light from thebody, or region, the temperature which is to be measured, iscompared with that from a standard electric lamp.

Page 16: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 17: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

III. RESISTANCE PYROMETRY

1. Theory,

The suitability of a substance for use as the re-sistance element of a pyrometer depends upon the constancyof the rate of change of its resistance when subjected torepeated heating and cooling and upon the temperature rangeover which it may be used. Platinum is practically the on-ly material fulfilling all of the requirements. Even plat-inum fails to meet the needs of the situation, unless it beproperly protected. Other metals with high melting pointsare unsuitable because their resistance is not constant ata given temperature. Instruments made of such materialswould not give identical readings twice when subjected tothe same temperature, if allowed to cool between observa-tions.

The work of numerous investigators shows that therelation between temperatures measured upon the scale of theplatinum thermometer and those measured upon the scale ofthe gas thermometer may be expressed by the following com-paratively simple, empirical formula:-

T - t - 5 (t/IOOT- t/100.

In this expression, T, is the temperature upon the scale ofthe gas thermometer, t, the temperature upon the scale ofthe platinum thermometer and e,a constant which is differentfor different samples of platinum, depending upon their pu-rity. Its value is about 1.5 for chemically pure metal.

Since this expression is a parabolic function,readings of the instrument at three known temperatures aresufficient to establish the scale of any given platinum thermometer. The three temperatures recommended for this useare 0^, 100° and 444.6 G. , or in other words, the meltingand boiling points of water and the boiling point of sul-phur. The relation of temperatures upon the platinum scaleto the change in resistance of a platinum conductor is shownin the expression given below.

t - (R - 100

tf is the temperature upon the platinum scale, the resistance of the wire at 0°C. , R,„at lOO^c and R, at the tempera-ture to be measured. An inspection of this formula showsthat temperatures upon the platinum scale are directly pro-portional to the resistance of the thermometer, the incre-ment in resistance from 0° to 100° C, being taken as 100

Page 18: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 19: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

5

degrees. Combining the two expressions given we have,

T - (r - ^y^,,r 100 = 6 (t/ioo)2- t/ioo

The relations indicated above hold only when the plat-inum is protected from the contaminating effects of hot gases.If not shielded, the element increases in resistance after heat-ing. This increase in resistance is due to chemical changesin the platinum. Fortunately, it is possible to eliminatethese effects by enclosing the element in a porcelain, or aplatinum tube. Externally glazed porcelain tubes are mostcommonly used, platinum being too expensive.

The range of the platinum thermometer extends fromthe region of - 200°C (- 350^F,) up to 1200OC. (2200°P,) Atabout 1200°C,, platinum begins to volatilize and readingstaken above this temperature are untrustworthy. Betweenthe limits mentioned, the resistance method affords the mostprecise means known for measuring, or controlling tempera-tures. The precision attainable is dependent largely uponthe precision with which the resistance measurements can bemade.

2. Apparatus.

The resistance method of measuring temperatures iseasily applied in a laboratory. Its application involvessimply the measurement of the change in resistance of a prop-erly protected platinum wire, the constant of which caneasily be determined.

For industrial applications, suitable portable in-struments and rugged platinum thermometers are necessary.At least two companies. The Cambridge Scientific Co., ofCambridge, England, represented in this country by the Tay-lor Instrument Companies, Rochester, N. Y. and the Leedsand Northrup Co. of Philadelphia, have placed resistance py-rometers upon the market.

The Cambridge company was the first concern tomanufacture satisfactory apparatus of this kind for commer-cial uses. Both companies make platinum thermometers andindicating and recording instruments for use with the same.Substantially the same principles are applied by each maker.(See drawing of theoretical circuits.)

The resistance thermometer of the Cambridge Companyconsists of a coil of fine platinum wire wound upon a micaframe and enclosed in a porcelain tube, (See cut). Theterminals of the coil are connected with suitable leads to

Page 20: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 21: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

6

CIRCUITS OF RESISTANCE PYROMETERS

THREE LEAD TYPE

Page 22: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 23: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 24: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 25: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

Leeds and Horthrup Indicator.

Page 26: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 27: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

9

binding posts upon the handle of the instrument. It will beobserved that there are four posts. Two of these are theterminals of leads ^riiich serve to compensate for temperaturechanges. These compensation leads are similar in all re-spects to the leads coming from the coil. They are connect-ed together at a point so that the resistance of the loopthey form in the tube is equal to that of the leads of thecoil. The purpose of this compensation loop is seen by aninspection of the drawing. Its use eliminates the necessi-ty for corrections for changes in the resistance of the leadsconnecting the thermometer with the instrument. The connec-tion leads consist of four parallel, identical conductorsprovided with suitable armor for mechanical protection.

The Leeds and Northrup resistance thermometerresembles the Cambridge Company's instrument in principle ofoperation and in general appearance. It, however, requiresbut three leads, except in the case of thermometers for thehighest temperatures, in which the practice of the Englishconcern is followed. The coil in these thermometers ismade of platinum wire heavy enough to render the use of amica support unnecessary. Both companies supply metalsheaths for protection of thermometers from mechanical in-jury. For the highest temperatures, these sheaths must bedispensed with and reliance placed upon the naked, porcelaintube.

The instruments used in connection with platinumthermometers are modified forms of the Wheat stone bridge.The Cambridge Company's instrument is known as the Whippleindicator. It consists of a wooden box about one foot on aside and weighing nearly 20 pounds. (See cut). Electric-ally, it is a ?/heatstone bridge with a very long slide wirewound upon a revoluble drum turned by the disk, H. Thescale beside the wire travels with it and is laid off in de-grees. To make a reading it is necessary to press the key,P, and turn the drum until the galvanometer, 3, shows no de-flection. The temperature is then read off the scale whichshows at the window, A.

A Leeds and Northrup indicating instrument isshown in the accompanying illustration. It is also made ina number of other forms, all of which employ the bridge prin-ciple, however. The type of instrument shown is admirablyadapted for use in temperature control, the workman beingable to tell by a glance whether his temperature is too high,or too low. The apparatus shown in the lower part of thepicture may be located in the superintendents' office, orany other place out of the workman's reach. The company alsomakes a portable balance indicator of about the size andweight of an ordinary voltmeter.

Page 28: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

t.

Page 29: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

10

Page 30: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 31: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 32: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 33: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

The stationary form of instrument used in con-nection with the Cambridge Company's resistance thermometersis known as the Callendar recorder, (See cut). It is acomplicated mechanism consisting of a -ATieatstone bridge inwhich the movable contact is shifted along the slide wire byclockwork that is controlled by the galvanometer. The con-tact carries the pen which makes a record upon a wide stripof paper moved by a clock. The leads from the thermometerare connected in opposing arms of the bridge so that anychange in the resistance of the thermometer coil destroys thebalance. This causes a deflection of the galvanometer whichstarts a clock that drags the contact point and pen in the proper direction to restore the balance. The device is a veryingenious piece of mechanism and upon the whole, reliable inoperation. It is adapted for recording any physical phenom-enon that is manifest by a change in resistance. Thus itmay be used as a recording voltmeter, a recording ammeter, ora recording wattmeter, provided the proper accessories are em-ployed.

The recorder manufactured by the Leeds and Nor-thrup Co., shown in the cut, operates upon substantially thesame principle as the Callendar recorder.

IV. THERMO-ELECTRIC PYROMETRY.

1. Theory,

Thermo-electric pyrometry is made possible bythe fact that the thermo-electromotive produced at the junc-tion of two dissimilar metals is proportional to the tempera-ture of the junction. Strictly speaking, the electromotiveforce is the algebraic sum of the e. m. fs. of all of thejunctions in the circuit. In practice, except for precisemeasurements the temperatures of the colder junctions may bedisregarded, provided they are low in comparison with thetemperature to be measured and are constant. Such a combi-nation of metals is known as a thermo-electric couple, ormore simply, a couple. Not all metals, however, are suita-ble for use in couples. Some are unfit because of theirlow fusing points, others because of their small thermo-elec-tric power and others because of the production of parasiticelectromotive forces. Platinum and its alloys of iridiumand rhodium, silver, copper and a few alloys of base metalsare practically the only substances used for ordinary thermo-electric couples. Couples are sometimes made of iridiumand an alloy of iridium and ruthenium. Temperatures as highas 2100OC. (SeOO^F.) can be measured with couples of thiskind, but they are not suitable for use outside of a labora-tory.

Page 34: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 35: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

13

Of all of the metals mentioned, platinum and analloy consisting of 90 percent platinum and 10 percent rhodi-um are the most satisfactory for temperatures up to 1600^0,(2900°F.) This combination is known as the LeChateliercouple. Its useful range does not extend much below 300^0.(600OF.)

Platinum and its alloys require careful handlingand protection to secure the best results. Protection ofthe heated portion from the contaminating effects of hotgases, particularly in a reducing atmosphere, is as essentiala feature of thermo-electric pyrometry as it is of resistancepyrometry. All of the volatile metals, such as silver, cop-per, zinc and antimony attack platinum readily producingchanges that alter the constants of couples, Silicon in areducing atmosphere is also detrimental to thermo-couples.The only way to avoid the injurious effects, is to insertthe couple in a porcelain, or platinum tube.

The relation between the electromotive force ofthe platinum-rhodium couple and temperature is expressed bythe following equation:

-

e = m (T"- t)

is the electromotive force, m is a constant ,T is the temp-erature of the hot junction and t., the temperature of thecold junctions. If the cold junctions are at 0°G,, the ex-pression reduces to,

e = m T"

which, upon taking logarithms, becomes

log e = n log T f log m

This is a straight line, necessitating the determination ofthe e.m.fo of the couple at but two points to standardize it.This formula is accurate to within 2^C, and is suitable forall but the most precise measurements.

Couples of copper and an alloy known as constan-tan, iron and constantan, gold and platinum are suitable forthe range from the region of absolute zero to 600 degrees C.(llCOOp.) (See Curves).

Couples made of such base metals as iron, nickeland chromium and their alloys are now used extensively assubstitutes for platinum and its alloys for temperatures upto 1200 degrees C. , or higher. In general they yield higherelectromotive forces and their e.m.f. temperature relationsare more nearly linear. They are open to the objection thatthey must be renewed frequently, but upon the other hand they

Page 36: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 37: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 38: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 39: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

15

are cheaper. They are not suitable for the most precisework at the highest temperatures, (See Curves),

2, Apparatus.

The standard thermo-couple recommended by thePhysikalische Reichanstalt of Berlin is made of a wire ofplatinum and a wire of platinum + 10 percent rhodium, 0.6mm, (About No. 23 B. & S. gage) in diameter, 1.5 meters (59inches) long, fused together at one end. The resistanceof the couple should be 1.6 ohms. The metals used must bepure and the alloy uniform throughout. The uniformity ofthe alloy is as important as the purity of its constituents.Points of variation along the wire will act as junctionsgiving rise to parasitic currents. Such a couple willgenerate an E. M. P. of approximately 17 milli-volts at 1600°G, (See Curve)

.

For the most precise work it is essential thatthe junctions of the instrument leads and the wires of thecouple be kept at a constant temperature. This is usuallyaccomplished by insulating then in glass tubes immersed inice water, (See drawing of thermo-electric circuit)

.

The hot junction and at least those portions ofthe wires of the couple likely to be exposed to hot gasesshould be protected from contamination. It is also veryessential that the wires be insulated, not only from theground, but from each other in order to prevent short cir-cuits. This is easily accomplished by threading themthrough a small porcelain, or fire clay, tube provided withtwo holes and by enclosing this tube in a larger porcelaintube, externally glazed and closed at one end. Some times,when the couple is to be used at relatively low temperatures,the outer porcelain tube is enclosed in an iron pipe for me-chanical protection. Tubes of fused quarts are now cominginto use for the protection of couples. They are not sofragile as porcelain tubes and may be used at very high temp-eratures. They also possess the valuable property of with-standing sudden changes in temperature. It is said thatthey may be plunged into water when red hot without injury.Porcelain tubes must be heated gradually and allo7/ed to coolslowly. If this precaution is not taken, it is almost cer-tain that they will be broken. While the inner tube may bein short sections, the sheath must be intact to prevent gasesgetting to the wires. Pire clay tubes are some times usedfor the outer envelope in special cases, , Platinum tubescan be used, but their cost is almost prohibitive.

Nearly all of the manufacturers of pyrometricapparatus make what they call "fire rods", or "elements" for

Page 40: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 41: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

THERMO-ELECTRIC

Page 42: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 43: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

insertion into the region where the temperature is to be meas-ured. These are generally arranged with a handle and some-times a guard to shield the hand is provided. They corres-pond to the thermometers used in resistance pyrometry. Thethermo-couples contained in them may be either platinum-rho-dium, or base metal elements, depending upon the temperaturesto be measured. Flexible . leads extend from the rod to theinstrument.

The electromotive forces generated by thermo-couples are measured by potentiometer, or galvanometric methodThe former is well adapted for strictly laboratory purposes,but is not suitable for use in installations requiring aninstrument that can be moved readily. The method is usedonly for the most precise work.

The galvanometric method involves the use of someform of D'Arsonval milli-voltmeter of which there are manyupon the market at the present time. The principal consider-ation involved in the choice of such an instrument, asidefrom mechanical features, is that its resistance should behigh in comparison with that of the couple and the remainderof the circuit. This precaution is necessary in order torender negligible the effects of changes in the resistanceof the external circuit. In general, a resistance of 400to 500 ohms is sufficient. Many instruments for this pur-pose are calibrated to read temperatures directly when usedin connection with a given couple. Some instruments arealso calibrated to read millivolts and may be used in con-nection with any couple if its temperature e.m.f, curve isat hand.

Pivot instruments suitable for use with thenoble metal couples are now manufactured. Base metal coupleswhich yield higher electromotive forces than platinum couples,give excellent results with pivot instruments. Most of thehigher grade galvanometers for thermo-electric work are ofthe suspension type, however. They consequently requirecareful handling and leveling to avoid introducing errors.The current produced in a thermo-couple circuit is exceed-ingly small. In the case of the standard LeChatelier coupleused in connection with a circuit of 500 ohms total resist-ance, it is only ,000034 ampere. Compared with .01 ampererequired to produce full scale deflection upon an ordinaryvoltmeter, for example, it is easily seen that there is notvery much power available for overcoming friction.

Several recording milli-voltmeters for use inconnection with thermo-couples are now made# In most cases,these instruments are indicating milli-voltmeters with a penattached to the needle which swings over a strip of papermoved by clockwork. The principal consideration involvedin the design of such an instrument is the elimination of the

Page 44: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 45: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

A Base Metal Couple Pyrometer.

A Thermo-Electric Kecorder.

Page 46: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 47: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

friction between the pen and the paper^ This is often ac-complished by allowing the pen to swing free normally and providing clock controlled mechanism for lifting the paperagainst the pen, or for pushing the latter down at regular intervals. The record thus consists of a series of dots.Sometimes instead of attaching the pen to needle, the latteris pushed down against an inked thread which makes a dot onthe record tape. At least two manufacturers, use a circularrecord chart, the recording part of the instrument being verymuch like that of the well known recording gages and electrical instruments made by the Bristol Company, Provision ismade for intermittent contact of the pen with the paper.

V. OPTICAL AND RADIATION PYROMETERS.

Optical pyrometers may be defined as convenientaids for measuring temperatures with the eye. Optical andradiation methods are based upon the fact that the energy ra-diated by a hot body is proportional to its absolute tempera-ture. The optical instruments usually depend upon the pho-tometric principle which may be somewhat crudely stated tobe the comparison of a light of known temperature with thelight emitted by the body whose temperature we would measure.Electricity is employed inithese instruments principally asa convenient means of lighting this standard lamp. Two op-tical pyrometers are so constructed that the heated filamentof a lamp is viewed in the same field as the hot body. V/henthe radiation from the body and from the filament are ofequal intensity, the filament becomes invisible. The mostimportant factor in the accuracy of this instrument is theconstancy of the standard lamp.

The radiation pyrometer utilizes the total ra-diation incident upon it from the hot body. It consists ofa convex lens, or a concave mirror, with a delicate thermo-couple, or resistance thermometer, at the focus. When athermo-couple is used, it is connected to a milli-voltmeterwhich is so calibrated that it indicates the temperature ofthe hot body. By use of specially sensitive couples it ispossible to m.easure temperatures as low as 600°C (llOOop.)The instrument is assembled in a tube open at the end turnedtoward the source of heat.

Optical and radiation pyrometers are used inmuch the same manner as a telescope- that is they are sight-ed upon the object, the temperature of which is to be measur-ed. In this manner of use lie their advantages and theirdisadvantages. They are obviously not subject to the de-teriorating effects of heat and there is theoretically no up-per limit to the temperature range over which they may beused. Their disadvantages are that they are not suitable

Page 48: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 49: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

Radiation Pyrometry

FEKY RADIATION PYROMETER, IN WEATHER-HOOD SIGHTED INTO FIRECLAY TUBE.

About 1-10 full size.

Page 50: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 51: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

for measuring temperatures of small spaces and provisionmust be made to sight them upon the region, the temperatureof which is to be measured. They are however, the onlymeans available for the measurement of temperatures above therange of the striotly electric pyrometers.

The optical pyrometers cannot be employed, ex-cept for temperatures high enough to cause luminosity. Theradiation types may be used for lower temperatures sincethey depend upon total radiation. The optical instrumentsare not easily made self registering, but recording radiationpyrometers are manufactured by at least two companies. TheCambridge Scientific Company, makes a thread recorder andthe Brown Instrument Co, of Philadelphia, employs the sameinstrument that it uses in connection with its themo-elec-tric pyrometers.

VI, INDUSTRIAL APPLICATIONS.

1,- Resistance Method,

Resistance pyrometry is well adapted for use inmany different kinds of industrial processes. The lists ofpurchasers of resistance pyrometers given in the catalogs ofthe two principal makers include chemical works, gas plants,glass works, breweries, steel works, annealing, hardeningand smelting furnaces and many others. The method is ofcourse limited in its application to those processes involv-ing temperatures of 1200^C, , or lower. Both indicating andrecording instruments are employed. The former are of mostvalue as a guide for the workman in the performance of hisduties. The latter enable the superintendent to know thetemperature conditions that existed at any time and furnishan impartial, unbiased cheek upon the workman. Both kindsof instruments are often used in conjunction. It is quitecommon practice to employ switch boards to which a number ofthermometers may be connected for use with a common instru-ment.

Resistance pyrometers are used a great deal formeasuring moderate and low temperatures. For instance theyare often used to indicate such temperatures as that of airsoils and water, of grain in elevntors and cars, of coldstorage rooms and of food in the process of manufacture.

Page 52: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 53: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

22

2. Thomo-Electric Method,

The thermo-electric method is suitable for usein practically any process that the resistance method may beused, and in addition to those involving temperatures as highas 1600^0, Thermo-couples are used extensively in the burn-ing of clay products, to some extent in foundry work, steelmaking, experimental work, glass making, tempering and an-nealing, galvanizing and many other processes involving hightemperatures. The number of users Is constantly increasingas the value of accurate control of temperature conditionsbecomes better understood. The users of thermo-electricpyrometers out number those employing the resistance type.

It is becoming common practice when a number ofcouples are used in a plant, in different ovens for example,to install a switchboard and a common instrument which maybe connected to any couple attached to the board. Record-ing instruments which may be under lock and key are oftenemployed as a means of supervision. The installation of per-manent wiring in a substantial, workmanlike manner is comingto be recognized as a profitable investment from an engineer-ing and economic standpoint. Some of the manufacturers ofpyrometers prefer to install their own apparatus and one con-cern at least, insists upon so doing. A new field of en-gineering is slowly being evolved. In twenty-five years the py-rometer will be considered as essential a part of a plant asthe furnace.

3. Optical and Radiation Method.

The optical and radiation pyrometers are especial-ly adapted for use in connection with furnaces involving temp-eratures too high for the electric pyrometers, or conditionsunder which the latter can not be installed. Their intelli-gent use under varying conditions calls for greater technicalknowledge than either of the other methods. On the otherhand, they are not subject to the deteriorating effects ofheat and if used continually for the same purpose are probablyto be preferred to the electrical instruments. Their in-herent accuracy is greater at the higher temperatures owingto the fact that radiant energy increases as the fourth powerof the absolute temperature of its source.

Page 54: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 55: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

23

VII. CALIBRATION OF PYROMETERS.

Owing to the deteriorating effects of heat, itis essential that pyrometers be calibrated at intervals.This is true, particularly if the metal of the element hasbeen subjected to hot gases. Base metal couples should bechecked frequently to detect changes in constants. Platinumresistance thermometers made by either of the two companiesmentioned can be depended upon for considerable periods oftime, unless subjected to abuse. They should be checked oc-casionally, however for changes in constants, especially ifused for precise work.

Electric pyrom.eters are calibrated by subject-ing them to known temperatures. This may be done in eitherof two ways. One method is to use a number of fixed temper-atures, such as the melting, or freezing points, of certainsubstances. The other method is to compare readings of thepyrometer with those of a standard instrument subjected tothe same conditions.

The first method is the more exact of the twoand is used for precise work« The freezing points of certainmetals have been determined many times with great exactnessin terms of the scale of the gas thermometer, the standardwith which all pyrometers are ultimately compared. Thefreezing points of tin, cadmium, lead, zinc, antimony, silver,copper, gold, and platinum may be used. In the metals men-tioned, the freezing point is marked by a period of station-ary temperature of several minutes duration in some cases.This affords ample time for a number of readings. Readingstaken with several metals afford sufficient data for a cali-bration. The number of points required depends upon theformula used. In the case of a resistance thermometer, threepoints are necessary, if the Callendar formula is used. Twopoints a considerable distance apart, as for example, thezinc and copper freezing points are sufficient to establishthe scale of the Pt- (Pt +• Rh) couple, if the formula givenabove is used.

Freezing point determinations should be madewith pure materials. It is not necessary to use chemicallypure metals in all cases however, except for the most refinedwork. Commercially pure copper gives results within a degreeof those given by chemically pure metal. Pig and pipe leadhave freezing points I-1/20 to 2°C. below that of pure lead.Commercial tin has about the same freezing point as the puremetal. Antimony upon the contrary is a metal that must bepure to give correct results. Antimony also exhibits thephenomenon known as, "under cooling". That is, the temper-ature goes down for a time and then suddenly rises 30^0, or

Page 56: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 57: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

more, at the moment of freezing, after which it again de-creases.

The electric resistance furnace is usually em-ployed in the freezing point method, the metal charge beingplaced in a graphite crucible. The electric furnace is ad-mirably adapted for calibration purposes because it is easi-ly controlled and because there are no contaminating gasespresent

.

It is necessary to prevent oxidation of the chargein some cases by covering the surface of the molten metalwith powdered graphite. Copper so protected gives a freez-ing point of 1083OG, If exposed to the air, the freezingpoint is 1065°C, a very constant temperature, however.

In the second method mentioned, the pyrometer tobe calibrated Is inserted in a furnace beside a standard py-rometer and the readings of the two instruments compared.The horizontal type of furnace is usually employed. Thismethod is the one used by the Bureau of Standards, exceptfor the most refined work.

The method is one that can be used to advantageby industrial concerns. It of course necessitates keepinga standard fire element that should be used for no other pur-pose. Such a thermometer, or couple, may well be a certi-fied instrument, the fees of the Bureau of Standards beingmoderate in comparison with the cost of pyrometric apparatus.Ten dollars and transportation charges will meet the expenseof any ordinary calibration of an electric pyrometer.

In works using many thermo-couples under tryingconditions, it may result in considerable saving to employcheap base metal couples and to check them at frequent inter-vals against a standard platinum couple. The base metalcouples give higher electromotive forces and can be used withmore rugged instruments than platinum couples. They alsocan be made heavy so as to resist mechanical injury and donot require such elaborate protection from the effects ofhot gases as do platinum couples. Of course, plants whoseprocesses involve temperatures greater than base metalcouples are capable of withstanding must necessarily employplatinum couples exclusively.

Resistance furnaces for calibrating purposes maybe purchased from a number of concerns. A furnace suitablefor many purposes and giving good results up to temperaturesas high as lOOOOP, can be constructed at a slight cost frommaterials easily picked up in almost any shop. It is some-what short lived, but this objection is offset by the lowcost of replacing it. The writer has constructed a number

Page 58: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 59: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

25

of such furnaces. The materials required are a clay, orporcelain, battery jar, several yards of iron wire of aboutNo. 20 gage, some mineral wool, asbestos, or magnesia pipecovering, and a few fire bricks. One layer of wire iswound around the jar beginning about one half inch from thebottom and ending about the same distance from the top. Aspace equal to several times the diameter of the wire is leftbetween each convolution. The jar is then placed in a re-ceptacle made of brick laid in fire clay, or inside a draintile, and packed with mineral wool. A section of magnesiapipe covering may be used instead of the mineral wool. Pack-ing is placed around the closed end of the jar and a remov-able plug of non-inflammable material is inserted in the openendo Pyrometers are introduced through holes in this plug.The ends of the wire are left long enough to bring out forconnection to the circuit, A theatre dinner in series withthe winding furnishes an admirable means of regulating thecurrent flow, A water rheostat may be used if desired.

The writer is at present constructing a resist-ance furnace of slightly modified design. A porcelain tube3 inches in diameter and l-l/z inches long is wound withNichrome ribbon .007 inch thick and l/8 inch wide. Aboutl/s inch space is left between each convolution. The insu-lation is to be at least 3 inches of mineral wool packedaround the tube in a drain tile mounted in a horizontal posi-tion upon suitable iron legs. The ends of the porcelaintube are closed, except for a hole 3/4 inch in diameter forthe insertion of pyrometers. While this furnace has notbeen tested, it is almost a duplicate of one that has givensatisfactory service for tempering tools. It is expectedthat a temperature of 1200°?., or higher, can be attained.

The accurate calibration of electric pyrometersis absolutely essential. The effect of heat is deteriorateany known material suitable for pyrometric use. Since thereis no means by which these destructive effects can be elimi-nated, the only recourse is to subject the instrument inquestion to test as often as experience shows to be necessaryto maintain the standard of accuracy required. Platinumcouples can be restored in most cases by glowing them atyellow heat for several hours in an electric circuit. Incase this does not restore the material to its original con-dition, it still has a market value for other uses. 3asemetal couples, being cheap, can be thrown away if contaminated.

The testing of radiation pyrometers involves theapplication of principles of optical methods lying out sidethe field of electrical measurements.

Page 60: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 61: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

VI. COWPARISON OP METHODS

No one method of pyrometry is suitable for usein every situation. The proper method cannot be chosenwithout careful consideration of all of the conditions underwhich the apparatus is to be used. The questions of temper-ature range, precision of measurement desired, protectionrequired for tho elements, intelligence necessary for oper-ation, necessity for records, ease of installation, firstcost and cost of up keep must be considered in the choiceof a pyrometer for a given plant. These are discussedbriefly in the following paragraphs which are later summarizedin tabular form.

1,- Temperature Range,

The temperature range of the resistance methodis limited to 1200°C. (2200OF.) if platinum is employed.The range of the thermo-electric pyrometer is greater, butcouples of different materials must be used in different partsof the range. Gopper-constantan, iron-constantan and plat-inum- gold couples are available for use from the lowesttemperatures to the lower limit of the Pt - Rh, or Ft - Ir,couple and the Ir - Ru couple from the upper limits of thelatter to the highest temperatures that can be measured bythe thermo-couple pyrometer. As pointed out previously,the last named couple is not a commercial instrument. Therange of the radiation pyrometer is from 600^0. (IIOOOF.) up.Theoretically, it has no upper limit,

2,- Precision of Measurement.

The precision of measurement is greatest whenthe resistance method is employed. In many industries theprecision of the laboratory is not required. The precisionwith which a workman is likely to make a given measurement isnot the same as the precision of which the instrument is ca-pable. The workman is prone to judge the accuracy of hismeasurement by the length of the portion of the scale covered,rather than by the temperature to which the space corresponds.For this reason instruments with long scales are especiallyrecommended. In this respect the resistance type of instru-ment, such as the Wiipple indicator with its long temperaturescale, meets the above requirement admirably.

Page 62: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 63: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

27

3,- Protection of Element.

The resistance method is at a distinct dis-advantage considered from the standpoint of protection re-quired for the element. The radiation pyrometer having nopart subjected to high temperatures requires no protectionfrom heat. Platinum couples need the same protection as theresistance thermometer, but are more easily re-calibrated Ifcontaminated. The base metal couples may be used withoutprotection.

4,- Operation.

The operation of the \Vhipple indicator requiressome manipulation upon the part of the observer before a

reading can be taken. It is not possible for him to know,except by going to the indicator whether the temperature isvarying and then he can not determine the amount, except bybalancing the bridge. If the temperature is fluctuatingvery much it is impossible to measure it with any degree ofcertainty. With the Leeds and Northrup instrument shown,it is possible to detect variations after balancing thebridge and the amount and direction of the change is also ap-parent. In other respects the operation of the apparatusmade by the two companies is about the same.

The thermo-electric pyrometers give visual in-dications of their readings and require no preliminary oper-ations when a reading is to be taken. The method of usinga radiation pyrometer depends upon the installation, but theactual readings are made as in the thermo-electric pyrometer.

5.- Recording Instruments.

The three methods are about upon a par in thematter of recording temperatures. The thermo-electric re-corders are simpler, but probably no more reliable.

6.- Permanent Installations.

The principal consideration in the matter of in-stallation is the wiring, which should be well done in eithercase. The resistance pyrometers require three to four leads,but their resistance is immaterial provided it is the samefor each conductor. Thermo-electric installations requireonly two leads for each couple, or in some cases one load and

Page 64: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 65: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

28

a common return, but these must be of low resistance which inno case should be more than one to two ohms. Radiation py-rometers use the same kind of wiring as thermo-couples.

7.- Cost of Apparatus.

The cost of apparatus ranges from |50,00, orless, to as high as $450 according to quotations and catalogs.Some base metal couples and suitable indicating instrumentscan be purchased for the lower figure. The larger figureis for a recording, radiation pyrometer, A certified Pt -

Rh couple from Heraeus with a Siemens and Halske galvanometerand protecting tubes can be purchased for about $150.00, Afive foot Pt - Rh couple costs about $60.00. Shorter lengthscost proportionately less. Base metal couples cost $6.00 to$10.00 if purchased of makers. Some kinds can be made fora few cents. Pivot galvanometers, or milli-voltmeters , suit-able for use with base metal couples can be purchased for lessthan $50.00, A good imported suspension type of galvanometercan be gotten for $75,00 (Siemens and Halske). Porcelainand quartz protecting tubes cost from $3.00 to $15.00.

The tipple indicator costs in the neighborhoodof $200,00 with a thermometer, A Leeds and Northrup indi-cator and thermometer can be bought for about $150,00,

Radiation pyrometers cost from $150.00 to $300,depending upon the make— the latter price being for an im-ported instrument.

8.- Cost of Up Keep,

This is a variable quantity for either of thestrictly electrical methods. Leaving wiring out of con-sideration, the radiation instruments should cost almostnothing for up keep. The cost in the case of the resistanceand thermo-electric methods depends entirely upon use.Platinum thermo-couples can be restored by glowing and can berecalibrated. If past restoration the metal can be salvaged.Base metal couples are easily calibrated and if necessarycheaply replaced. If the use of resistance pyrometers issevere, new thermometers may be necessary, making up keepcharges high.

Page 66: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 67: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

29

9.- Tabular Comparison of Methods.

Feature Methods

Compared Resistance Thermo-elect ri c Radiat ion

Temp,Range

Absolute zero to1200°C.

Absolute zero to2IOOOC.

Prom 600° C.up.

Precisionof

Measurement

Within loc.-0°to 5000c. 30c.500^ to lOOO^C.with Whipple In-dicator.

Depends upon in-strument, coupleand leads. Underbest conditions20 at lOOOOC.using galvanom-eter.

100 at lOOOOC.Less at lowertemperaturesfor Fery.

Protectionof Element

Must be enclosedin porcelaintube to protectfrom contamina-tion by hotgases.

Platinum couplesmust be enclosedin porcelain.Base metalcouples may beexposed.

No part sub-jected tohigh temper-atures.

OperationBridge must bebalanced beforereading can bemade.

Readings made byinspection of aninstrument

,

Same asthermo-electric py-rometer.

RecordersComplicated,but on wholesatisfactory.

Simple and sat-isfactory inbest types.

Same as thermo-electric re-corders.

PermanentInstallation

Leads of 3 or4 conductors re-quired. Resis-tance immaterial.Can be used witha switchboard.Thermometers anydistance frominstrument

.

Only 2 wires re-quired, but re-sistance must below. Can beused with aswitchboard.Elements may beconsiderabledistance frominstrument

.

Must be locat-ed where itcan be Bight-ed on hotbody. Instru-ment may beat distance.2 heavy leadsrequired.

Page 68: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 69: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

30

Cost ofApparatus

Indicating in-strument andthermometer:^150 to $250Recorder, $200.to $400.

Platinum coupleand galvanom-eter $100 andup. Basemetal couplesand instru-ment $50 andup. Record-ers $125 andup.

Indicatinginstrumentfl50 to$300. Re-cording in-strument$250 to1450.

Cost ofUp Keep

Depends uponuse. Greatif thermometersmust be re-placed. Cal-ibration moreexpensive thanfor couples.

Couples ofPt can besalvaged.Base metalcouples cheapenough tothrow away.Calibrationfairly easy.

Practical-ly nothing.

IX. CONCLUSION.

The art of measuring high temperatures hasbeen developed greatly in the last few years and much ex-cellent pyrometric apparatus is now upon the market. Thegreatest development has been along a few lines, althoughprobably every physical, or chemical change, caused by heathas been investigated at some time with a view to using itas a means of measuring temperature. However great thedevelopment, it cannot be said that we are yet able to gagehigh temperatures repeatedly with the precision that we canmeasure v/eight, or electrical quantities, repeatedly.Single measurements can be made it is true, but not repeat-edly with such precision, owing to the fact that the effectof heat is to change the physical and chemical propertiesof the pyrometric material. Optical and radiation instru-ments are not open to this objection, but have other inherentlimitations. Optical methods depend to some extent uponthe eye and the radiation methods are limited in their ap-plication. Unless some new substance or principle is dis-covered, future improvements will be along the line of re-finements in present day methods. Since the limitationsof pyrometry as practiced to-day are inherent, the educa-tion of the users to recognize them and to govern them-selves accordingly is necessary if the arts are to be im-proved by their employment.

Page 70: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 71: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

31

X. BIBLIOGRAPHY.

The following list of references is by no meanscomplete, nor is it intended to be. It, however, includesa sufficient number of works to give an engineer all of theinformation needed upon the subject of electrical pyrometry,unless he intends to specialize. The works mentioned areeasily obtained. The catalogs sent out by the companieswhose n^mes are given in the accompanying list, contain agreat deal of valuable information about apparatus and manyhints concerning the use of pyrometers.

1, High Temperature Measurements.By Chatelier and Boudouardo Translated by Burgess,This work is the most comprehensive treatise publish-ed on the subject of high temperature measurements.It contains an excellent bibliography. Published byWiley and Sons. Price |3.00,

2» Methods of Pyrometry,By C. W. Waidner. Reprint from the Proceedings ofthe Engineers' Society of 7/estern Pennsylvania, Avaluable paper by a reliable authority upon the subject.It covers practically the entire field in a brief,comprehensive manner. The pamphlet may be had by ap-plying to the Secretary of the society, Pittsburgh,

3, Optical Pyrometry,By C. W. Waidner and G. K, Burgess, Reprint fromBulletin No, 2 U. S. Bureau of Standards, IVa.shington,D. C, An authoritative discussion of the theory andapparatus of optical pyrometry. Some attention isdevoted to radiation pyrometers,

4, Platinum Resistance Thermometry at High Temperatures,By C. V/, 77aidner and G. K. Burgess. Published inBulletin No. 2. Vol, 6 of the U. S. Bureau of Stand-ards. A very comprehensive treatise upon the sub-ject. It contains much information regarding the cal-ibration of instruments by the melting point method.

5. Pyrometer Testing and Heat Measurements.Circular No. 7. issued by the U. S, Bureau of Stand-ards, T/ashington, D. C. This publication containsa great deal of valuable information regarding pyro-meters and their calibration. It is sent free ofcharge upon application.

6. University of Illinois.Bulletin No. 36, Eng. Exp. Station. By Clement andEgy,- Appendix contains definite information regard-ing the calibration of couples by the melting pointmethod.

Page 72: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 73: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

32

XI. MANUFACTURERS OF PYROMETRIG APPARATUS.

The manufacturers given are the concerns whoseadvertisements are seen most frequently in the technicaljournals of the day. The writer has used apparatus made by.at least half of the firms listed. Some of the concerns,however, are known to the author only through their litera-ture, or by their reputation in other lines.

1, Bristol Company, 7/aterbury, Conn,This company, which also manufactures the well knownline of Bristol recording gages and electrical instru-ments, makes the pyrometric apparatus formerly putout under the name of Wm. H. Bristol, New York, Thecompany makes thermo-electric apparatus includingthermo-couples, indicating and recording instruments.The indicating instrument is gotten up in the form ofthe ordinary voltmeter. The recording instrumentis designed for a circular record chart.

2, The Brown Instrument Company, Philadelphia, Pa.Makers of base metal and platinum- rhodium couples,indicating and recording instruments. The indicat-ing instruments are made in round and edgewise walltypes and in portable form. The recording instru-ment is of the type employing a circular recordsheet. The company also makes a radiation pyrome-ter.

3. Cambridge Scientific Company, Cambridge, England.Represented in this country by the Taylor InstrumentCompanies, Rochester, N. Y. This company manufac-turers platinum, resistance pyrometers, the WhippleIndicator, and the Callendar Recorder. The companyalso manufactures platinum- rhodium thermo-couples,indicating and recording instruments for use withthe same and the F6ry radiation pyrometer,

4. Charles Engelhard, New York,American sales agent for the Heraeus LeChatelierplat inum- rhodium thermo-couples probabl3r the bestupon the market. Also sells Siemens and Halskeinstruments for use in connection with temperaturemeasurements. The Siemens portable galvanometer isone of the best made for thermo-electric work,

6. The Frink Pyrometer Company, Lancaster, Ohio.This concern manufactures both base metal and plati-num couples together with indicating and recordinginstruments for use in connection with the same.The company has liad considerable experience with py-rometric installations in clay products works.

Page 74: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 75: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses

33

6« Hoskins Manufacturing Co., Detroit, Mich,Manufacturers of a general line of electric fur-naces and of base metal couples. Also manufactur-ers of portable and switch board types of indi-cating instruments. No recording instrument ismade by this company,

7, The Leeds and Northrup Co., Philadelphia,Makers of a general line of resistance pyrometricapparatus. Can furnish both indicating and recording instruments,

8. Wilson-Maeulen Co., New York*This company advertises base metal couples, plati-num couples, portable and stationary indicating in-struments and a recorder.

Page 76: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 77: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses
Page 78: High temperature measurement by electric means · 2015. 5. 27. · I.INTRODUCTION. Theaccuratemeasurementofhightemperatureshas cometobeaveryessentialfeatureofmanyindustrialpro- cesses