18
HYDROGEN INDUCED DISBONDING: FROM LABORATORY TESTS TO ACTUAL FIELD CONDITIONS L. Coudreuse, S. Pillot, P. Bourges Centre de Recherche des Matériaux du Creusot (CRMC) INDUSTEEL Creusot B.P. 19, 71202 Le Creusot France A. Gingell Chateauneuf plant INDUSTEEL France B.P. 68, 42803 Rive-de-Gier France ABSTRACT Disbonding tests are often required during fabrication of hydrotreating vessels. However it is difficult to make the link between the testing conditions and the actual service conditions in the reactor wall. From numerical simulations of hydrogen profiles through disbonding test specimens and reactor walls it is possible to assess both the severity of disbonding tests and the actual risk of disbonding in service. It is shown that in-service temperature and hydrogen pressure are important parameters that control hydrogen distribution throughout the wall of a reactor, however, the influence of the wall thickness will also be underlined: the greaterer the thickness of the vessel, the greaterer the risk of disbonding. On the basis of a data bank of disbonding test results performed for fabricators world-wide, a proposal is made to adapt the severity of the test to the severity of the service conditions (pressure, temperature, wall thickness). Keywords: Disbonding, Hydrogen Induced Disbonding, Hydrotreatment, 2.25Cr-1Mo, 1.25Cr-1Mo, V modified grades.

Hydrogen Induced Disbonding - From Laboratory Tests to Actual Field Conditions - Nace 2005 - Coudreuse Et Al

Embed Size (px)

DESCRIPTION

HID

Citation preview

  • HYDROGEN INDUCED DISBONDING: FROM LABORATORY TESTS TO ACTUAL FIELD CONDITIONS

    L. Coudreuse, S. Pillot, P. Bourges Centre de Recherche des Matriaux du Creusot (CRMC)

    INDUSTEEL Creusot B.P. 19, 71202 Le Creusot

    France

    A. Gingell Chateauneuf plant

    INDUSTEEL France B.P. 68, 42803 Rive-de-Gier

    France

    ABSTRACT

    Disbonding tests are often required during fabrication of hydrotreating vessels. However it is

    difficult to make the link between the testing conditions and the actual service conditions in the reactor wall. From numerical simulations of hydrogen profiles through disbonding test specimens and reactor walls it is possible to assess both the severity of disbonding tests and the actual risk of disbonding in service. It is shown that in-service temperature and hydrogen pressure are important parameters that control hydrogen distribution throughout the wall of a reactor, however, the influence of the wall thickness will also be underlined: the greaterer the thickness of the vessel, the greaterer the risk of disbonding. On the basis of a data bank of disbonding test results performed for fabricators world-wide, a proposal is made to adapt the severity of the test to the severity of the service conditions (pressure, temperature, wall thickness). Keywords: Disbonding, Hydrogen Induced Disbonding, Hydrotreatment, 2.25Cr-1Mo, 1.25Cr-1Mo, V modified grades.

  • INTRODUCTION

    Hydrogen Induced Disbonding, called Disbonding in the following, can occur after cooling down of hydrotreating reactors which operate at elevated temperature under high hydrogen pressure. Disbonding, is characterized by a crack propagating at the interface between the base material, and the austenitic stainless steel weld overlay or cladding (In the paper, overlay will be use to describe both weld overlay and cladding). It is the consequence of both hydrogen over-saturation at the interface after cooling down and the presence of a sensitive microstructure at this interface [1]. Typical hydrogen profiles through the wall of a reactor in service, steady state condition, and after cooling down, are shown on Figure 1. During operation, there is a linear gradient of hydrogen content in the stainless steel overlay and in the base material. There is more hydrogen in the stainless steel overlay than in the base material because of the higher hydrogen solubility in the austenitic stainless steels. The increase of hydrogen content at the interface is the consequence of the variation of hydrogen diffusivity and solubility laws when the temperature decreases. As will be explained later, there is diffusion of hydrogen from the parent material to the weld overlay, but since the diffusion coefficient of hydrogen is low in the overlay, there is a strong increase of hydrogen at the interface.

    An API survey, conducted several years ago, on heavy wall reactors has revealed that out off 100

    reactors investigated, twenty-eight (28) exhibited some form of disbonding [2]. The operating conditions of most of these reactors were in the following range: Temperature (455C- 466C/850F- 870F) and hydrogen partial pressure (138- 245 bar/ 2000- 3450psi). Since then the question of disbonding resistance of stainless steel weld overlays has become a question of great concern. More and more specifications for hydrotreating vessels are asking for disbonding tests. In 1996, an ASTM standard was published to describe the way to run a test [3]. More recently API publication 934 [4], which describes materials and fabrication requirements for heavy wall pressure vessels for high temperature and high hydrogen pressure service, introduced a requirement for disbonding tests. However, both the ASTM standard and the API publication are not clear on the testing conditions to be used. An industry joint research program conducted in Europe led to the conclusion that in many cases, use of actual service conditions for testing, (hydrogen pressure and temperature), results in excessively severe test conditions [5].

    The objective of this paper is to discuss the influence of several parameters, (hydrogen pressure, temperature, cooling rate, thicknesses), on the increase of hydrogen content at the interface and to demonstrate why the use of actual service conditions for disbonding tests may result in too severe conditions. On the basis of numerical calculation and test results, a matrix of disbonding test conditions is proposed based on the severity of in-service conditions.

    DESCRIPTION OF HYDROGEN ENRICHMENT AT THE INTERFACE Why is there hydrogen enrichment at the interface?

    Figure 1 describes the hydrogen profile in a reactor wall at the interface, for the steady state condition and after cooling down. Knowing the hydrogen diffusivity and solubility laws for both the

  • base material and the weld overlay, it is possible to determine the hydrogen profile in steady state condition.

    C0, Hydrogen concentration in the overlay at the internal surface, can be calculated from the Sievert laws. C0 = pH21/2*S (1) p2: Partial hydrogen Pressure S : Hydrogen solubility in the austenitic stainless steel

    C01 and C02 are the hydrogen concentration at the interface in the overlay and in the base material respectively. At the equilibrium, the hydrogen activity is the same in the two materials, then: C01/ C02 = S/S (2) S : Hydrogen solubility in the base material

    It can also be established that the ratio C0/C01 can be calculated from the following formula [6]: C0/C01 =1+ (S/S)*(D/D) *tov/tbm (3) tov : Overlay, or cladding, thickness tbm: base material thickness

    In order to calculate, the hydrogen concentration at the interface after cooling down, it is necessary to consider that there is at any time, t, an equilibrium between the hydrogen activity at the interface in the base material and in the weld overlay: Ct1/ Ct2 = S()/S() (4) S() and S() are the values of the hydrogen solubility respectively in the overlay and in the base material at the temperature T (in K) corresponding to time t.

    The evolution of hydrogen solubility with temperature can be expressed by an Arrhenius law: S =So*exp (-E/RT) (5)

    Figure 2 shows the relationship between hydrogen solubility and 1000/T for an austenitic stainless steel and a 2.25Cr-1Mo low alloy steel. The following laws, well accepted for the two materials have been used [6]: S = 33*exp(-3333/T) (6) S = 8.93*exp(-650/T) (7)

    The solubility is expressed in ppm bar1/2

    At 450C (1000/T = 1,383), the ratio between the hydrogen solubility in the austenitic stainless steel and the base material is about 11. There is 11 times more hydrogen in the overlay than in the base material at the interface. When the temperature decreases to room temperature (1000/T = 3,356), the ratio increases to about 2200. The increase of hydrogen content in the overlay, close to the interface, can be explained by this increase of the ratio S/S. Since the hydrogen solubility remains high in the overlay there is diffusion from the base material to the overlay. The maximum hydrogen content

  • depends also on a kinetic aspect, which can be described by the diffusion coefficient. As shown in figure 3, the diffusion coefficient in the austenitic material is very low, when compared with the low alloy steel base material. At 450C, the difference is about 2 orders of magnitude; at room temperature it is about 6 orders of magnitude. The following diffusion coefficient laws have been used in Figure 3 [6]: D = 2.4*10-3 exp(-2132/T) (8) D = 7.1*10-4 exp(-4555/T) (9)

    Diffusion coefficient is expressed in cm2/s

    The low hydrogen diffusion coefficient in the austenitic material explains the sharp aspect of the hydrogen peak in the overlay. Difference between the hydrogen distribution in the wall of a reactor and in a test specimen

    In order to study disbonding resistance of overlay, specimens are exposed in an autoclave under hydrogen pressure at elevated temperature. After exposure and cooling at a selected cooling rate, the amount of disbonded area is evaluated through UT examination.

    Different specimen sizes have been used for disbonding tests. Figure 4 shows examples of

    specimens used. Figure 4a represents a parrallelepipedic specimen, used by the authors of this paper, while figure 4b represents the specimen described in the ASTM standard G146. The main difference between the two specimens is the presence of a side austenitic weld overlay in the case of the ASTM specimen. The reason for this side overlay is to avoid hydrogen diffusion from the cylindrical surface of the specimen. It has been shown with numerical calculations, that if the cooling rate is high enough (Greater than 150C/H), the hydrogen content at the interface in the middle of a parrallelipedic specimen is the same than for the ASTM specimen. However for both types of specimen, the hydrogen exposure consists in the exposition of the whole specimen under hydrogen pressure in an autoclave. When the time necessary to achieve a saturation of hydrogen in the specimen is achieved (steady state condition), the specimen is cooled at a specified cooling rate. So in the steady state condition in the specimen a homogeneous hydrogen distribution at the interface is obtained (Figure 5). C0 = pH21/2*S (10) C0 = pH21/2*S (11)

    Then if the exposure condition in the autoclave corresponds to the actual service conditions, it will result in higher hydrogen content at the interface, at the steady state. The initial conditions are not the same at the interface before cooling down.

    In order to determine hydrogen profile through the wall of a reactor or in a test specimen, a finite element program has been used. Hydrogen profiles have been calculated for a wide range of conditions for the temperature, the hydrogen pressure, the cooling rate, the base material thickness, and the overlay thickness. A statistical analysis of the obtained results has permitted to establish relationships between the different parameters and the maximum hydrogen content at the interface, after cooling down, for both a reactor wall and a laboratory test specimen [5].

  • PARAMETERS ACTING ON MAXIMUM HYDROGEN CONTENT AT THE INTERFACE

    The occurrence of disbonding at the interface between an overlay and the base material, depends on two kinds of parameters: Parameters acting on the metallurgical quality of the interface, and parameters acting on the amount of hydrogen at the interface. The quality of the interface, its sensitivity to disbonding, depend on the base material composition and more particularly, on its carbon content[7], the welding conditions when apply, the overlay, and the PWHT[8]. It is not the purpose of the paper to describe the influence of these parameters. Then for a given quality of the interface, the occurrence of disbonding will depend on the hydrogen content at the interface. Disbonding will occurs when a critical hydrogen concentration will be reached.

    According to formulas (1),(2),(3) and (4), the hydrogen content at the interface after cooling down depends on the following parameters. Hydrogen pressure (pH2) Temperature (T)

    Hydrogen diffusion coefficient and hydrogen solubility laws for the overlay and the base material (D(T); D(T); S(T) and S(T)) Base material thickness (tbm) Overlay thickness (tov) Cooling rate (Cr): An average cooling rate value has been considered Time

    In the following, the influence of these parameters on maximum hydrogen content at the

    interface, CHmax, is discussed. Influence of time:

    During and after cooling, there is an increase of hydrogen content at the interface. After a fast increase of hydrogen content during cooling, there is a slower increase when the reactor remains at room temperature. Maximum hydrogen content, CHmax, is obtained after several hundred hours. As wall thickness increases, more time is required to reach maximum hydrogen content [5]. For disbonding test specimens, numerical simulations show that the maximum hydrogen content is achieved within about 130h for a 40mm thick specimen. This is in accordance with the experimental observation which show that the amount of disbonding measured by UT increases with time after cooling down.

    Influence of hydrogen diffusivity and solubility:

    The hydrogen profiles depend on the hydrogen solubility and hydrogen diffusion coefficient in the base material and in the overlay [5]. Lower increase of hydrogen at the interface is obtained when the differences in hydrogen solubility and hydrogen diffusivity between the two materials decrease. Then for a given overlay, the choice of a base material with a low diffusion coefficient and a high hydrogen solubility, such as V modified Cr-Mo grades [9], will result in a lower CHmax values, than the one obtained for a standard Cr-Mo grade working in the same condition. However the hydrogen solubility and the hydrogen diffusion coefficient are not known for each material used in the fabrication of hydrotreating reactors, and even for a given material these values can depend on slight variations in chemical analysis or on the heat treatment. So for the following, we have considered the solubility and diffusivity laws given by relationships (6),(7),(8),and (9). Then, the CHmax value calculated cannot be

  • considered as the exact hydrogen content at the interface for each material combination. It can be considered as a value indicating the severity of service conditions.

    From the numerical calculations performed, it has been possible to establish a formula [5].

    CHmax (reactor) = f( T, pH2, Cr, tbm and tov) (12) Influence of temperature, hydrogen pressure and cooling rate:

    Figures 6 to 8 give the influence of pressure and cooling rate on CHmax, calculated from relation (12), for different operating temperature for a 250mm thick reactor, with a 6mm thick overlay. It is not surprising to see that an increase of temperature or hydrogen pressure results in higher hydrogen content at the interface. There is not a big influence of cooling rate on CHmax, this confirms the fact that degassing occurring during cooling down is not efficient to reduce the amount of hydrogen in the reactor.

    The influence of wall thickness is shown on Figures 9 and 10. Figure 9 gives the evolution of CHmax for a service temperature of 450C for hydrogen pressure between 50 bar and 200 bar and for a wall thickness between 50 and 350mm. The figure clearly shows that the wall thickness has a strong influence on CHmax. This can be explained by the following reasons: . When the thickness decreases there is more hydrogen diffusion to the atmosphere from the external side of the reactor. There is less hydrogen available to go to the interface. . According to formula (3), and (2), the hydrogen contents at the interface, C01 and C02, increase when the base material thickness increases. There is more hydrogen content at the interface before cooling down in a thick reactor than in a thin one, for the same operating conditions.

    Figure 10 gives the influence of weld overlay thickness. It shows that an increase of the weld overlay thickness reduces CHmax. This is well explained by formula (3): C01 decreases when overlay thickness increases.

    These results confirm that disbonding is a more critical issue for thick wall reactors. Thick wall reactors are also those operating in the more severe service conditions (elevated temperature and high hydrogen pressure).

    SIGNIFICANCE OF DISBONDING TESTS

    There are several possible objectives for a disbonding test. It can be a quality control test, with fixed testing conditions, assessing that the weld overlay or the cladding, give a sufficient quality level by comparison with previous test results. A second objective can be to have a test representative of service conditions to ensure that the overlay will be exempt of problems in service. In the first case, the test may not necessarily be representative of service conditions; it can be more severe. However, it is necessary to have sufficient feedback on in-service behaviour to determine what the acceptance criteria must be for a severe test. The first approach can only be used for comparison purposes as it is not possible to determine acceptable disbonding test results for a given service condition. In the second case, the result of the test must be no disbonding.

  • The second approach is probably more adapted to the qualification of overlay procedures. However it is necessary to have a disbonding test which can represent the actual situation of a reactor wall. It means that CHmax in the disbonding test specimen must be equal or close to the CHmax value obtained in the reactor wall.

    As for reactors, on the basis of numerical calculations, performed for different specimen

    thickness and for different exposure and cooling conditions, a formula has been established between the maximum hydrogen content at the interface, CHmax, and these parameters.

    CHmax (specimen) = g(T, pH2, Cr, tbm and tov) (13)

    Again the relationship has been established for the solubility and diffusivity laws (6),(7),(8),and (9). It means, as, for reactor walls, that the value is not necessary the actual value at the interface. It can be considered as a value giving the severity of the testing conditions.

    In order to check if there is a relation between the disbonding test and the hydrogen content at the interface, a series of tests have been conducted for a wide range of conditions (temperature, hydrogen pressure, cooling rate, and specimen thickness), for the same overlay condition. Figure 11 represents the relation between the percent of disbonded area on the surface of the specimen (D%), determined by UT, and the calculated CHmax. The figure clearly shows that below a critical hydrogen content at the interface, about 375ppm, there is no disbonding, while the amount of disbonded area increases with CHmax values higher than 375ppm. This confirms that disbonding depends on the hydrogen content at the interface.

    In table 1, the CHmax values calculated for reactors (operating conditions are T=450C, pH2=150bar, and the cooling rate is 50C/h), for different wall thicknesses are given together with the value obtained for a specimen. For the laboratory test, the exposure conditions have been taken equal to the reactor service condition (450C; 150bar H2). CHmax values in laboratory test specimens are higher than the values obtained for reactor walls. This indicates that for the cases considered, the disbonding test conditions are not representative of actual service conditions.

    Table 2 gives for each wall thickness, a disbonding test condition, which would result in the same CHmax value. The temperature is kept constant, but it is shown that it is necessary to decrease the hydrogen pressure. The cooling rate has also been kept constant and equal to 150C/h. In fact for disbonding tests, the cooling rate must be high enough to limit hydrogen degassing from the sides of the specimen; 150C/h could be considered as the minimum cooling rate value to be used. The lower the thickness of the reactor wall, the lower the hydrogen pressure must be.

    In the following, a disbonding test guideline, based on the actual severity of in service conditions is proposed.

    DISBONDING TEST GUIDELINE

    Figure 12 gives the amount of disbonding measured by UT on test specimens, versus the CHmax values calculated for the exposure conditions used during those tests. The figure gives test results performed in our laboratory, for many fabricators world-wide. The results concern different weld overlay procedures, welding conditions, PWHT While most of the results concern 2.25Cr-1Mo base

  • materials, results for 1.25Cr-0.5Mo and V modified grades (2.25Cr-1Mo-0.25V and 3Cr-1Mo- 0.25V) are also given. Typical testing conditions are also reported on the CHmax scale. Testing conditions 1, 2, 3 and 4 gives CHmax values of respectively 333, 385, 464 and 496ppm.

    For 2.25Cr-1Mo, the discrepancy in the result corresponds to the wide variety of specimens tested: different welding procedure, heat input, PWHT, base materialHowever it appears that the higher amounts of disbonded areas correspond to the most severe test conditions. Below 350ppm, disbonding was never observed on 2.25Cr-1Mo. For CHmax > 500ppm most of the specimens are disbonded. 1.25Cr-0.5Mo appears more sensitive to disbonding : Disbonding is observed for a CHmax value of 330ppm. At the opposite for all the testing conditions, no disbonding has been found with V modified 2.25Cr 1Mo or 3Cr-1Mo grades, up to CHmax value of 690ppm. It is necessary to repeat that CHmax has been calculated with hydrogen diffusivity and solubility laws typical of standard 2.25Cr-1Mo, and that the scale has to be considered as a severity scale, describing the testing conditions. Due to the increased hydrogen solubility and lower hydrogen diffusion coefficient for V modified grades, the actual CHmax value at the interface is well below the values indicated. It is also possible that the increased sensitivity with 1.25cr-0.5Mo is linked with lower hydrogen solubility in this grade. The choice of testing conditions can be made by comparing the severity scale of the testing conditions with the severity scale for the service conditions (which take into account temperature, hydrogen pressure, cooling rate and thickness of both base material and weld overlay).

    Figure 13, is identical to Figure 12, but the CHmax values calculated for different reactor wall thickness have been reported. The operating conditions are: 450C and 150 bar H2. A cooling rate of 50C/h has been chosen for the calculation. This cooling rate can be considered as an extreme cooling rate for a reactor, so that to have a conservative approach. The CHmax values obtained for these conditions are given in Table 1. According to figure 13, for the overlaying conditions used on 2.25Cr-1Mo, no disbonding is expected for reactors C, D and E. For reactors B and A, some of the overlay conditions tested would result in disbonding. If in order to qualify an overlaying procedure, test condition 3 is selected (exposure conditions corresponding to in-service conditions), there is a high probability to have disbonding, but this does not mean that the quality of the overlay is not acceptable for the service conditions. For this reason, it is suggested to select the disbonding test condition, as a function of actual service conditions.

    The proposed approach, is illustrated on Figure 14, for a temperature of 450C, but such

    diagrams can be drawn for other temperatures. On the diagram, a mapping of CHmax values is given as a function of hydrogen pressure and thickness. The CHmax values are given for the different combinations of thickness and pressure. In order to have a conservative approach, the calculations have been made with a cooling rate of 50C/h, and an overlay thickness of 4mm. Iso CHmax curves have been drawn for 450, 400, 300, and 250ppm respectively. Then domains have been defined, up to 200 bar H2 and up to 350mm thick reactors. Domain A includes combination of thickness and pressure giving 480>CHmax > 400ppm, domains B, C, D, and E correspond respectively to 450/400, 450/350, 350/300 and 300/250ppm and

  • probably possible to consider that combination of operating conditions which results in less than 300ppm for CHmax will not result in disbonding, and that tests are not necessary for reactor operating in domains E and D. For 1.25Cr-0.5Mo probably domain F will not need disbonding test qualification. For V modified grade whatever the domain no disbonding is expected, however it can be suggested to perform disbonding tests for domain A.

    CONCLUSIONS

    Hydrogen induced disbonding, at the interface between an austenitic stainless steel weld overlay or cladding and a Cr Mo base material, is the consequence of strong hydrogen enrichment at the interface during and after cooling. The influence of in-service conditions, hydrogen pressure and temperature can be calculated and then used to determine equivalent test conditions. More severe service conditions result in higher hydrogen contents. While the influence of the cooling rate of the vessel does not have a strong influence on the hydrogen content, it has been shown that the greater the thickness of the vessel, the greater the amount of hydrogen at the interface. The use of thick overlays, can be a way to reduce the hydrogen content. Calculations of hydrogen profiles in laboratory test specimens show that in most of the cases, use of actual service conditions, hydrogen pressure and temperature, results in too severe testing conditions. This is particularly true for thin reactors. Analysis of many laboratory test results confirms that in many cases testing conditions used are too severe and may result in disbonding. An approach has been proposed to define a disbonding test guideline. Disbonding test conditions have been proposed. They depend on the severity of the operating condition of the reactor which take into account the temperature, the hydrogen pressure and the wall thickness. On the basis of many test results, it is suggested that disbonding test qualification may not be necessary for thin reactors or for less severe in-service conditions, depending on the base material selected.

  • REFERENCES

    1. H.Okada, K.Naito, J.Watanabe, K.Onishi and R.Chiba: Current Solutions to Hydrogen Problems in Steels, Ed. C.G. Interrante and G.M. Pressouyre, ASM 1982, p349

    2. M.S. Cayard, R.D. Kane and C.E. Stevens: Paper N518, CORROSION/94, NACE

    international, March 1994.

    3. ASTM G146-96: Evaluation of Disbonding of Bimetallic Stainless Alloy/Steel Plate for Use in High-Pressure, High Temperature Refinery Hydrogen service, Annual Book of ASTM Standard, Vol03.02

    4. API Recommended practice 934: Materials and Fabrication Requirements for 2.25Cr-1Mo

    and 3Cr-1Mo Steel Heavy Wall Pressure Vessels for High Temperature, High Pressure Hydrogen Service, First Edition, December 2000

    5. L. Coudreuse: Paper 01533, CORROSION/2001, NACE International, Houston, march 2001

    6. K.Smit and P.F.Ivens: Interaction of Steels with Hydrogen in Petroleum Industry Pressure

    Vessel Service, Ed M.Prager, The Material Properties Council inc., 1993, p205

    7. G.M. Pressouyre, J.M. Chaillet and G.Valette, Current Solutions to Hydrogen Problems in Steels, Ed. C.G. Interrante and G.M. Pressouyre, ASM 1982, p331

    8. A.Vignes, R.Palengat and P.Bocquet, Interaction of Steels with Hydrogen in Petroleum

    Industry Pressure Vessel Service, Ed M.Prager, The Material Properties Council inc., 1993, P139

    9. L. Coudreuse and P. Bocquet: Hydrogen Transport and Cracking in Metals, ed A.Turnbull,

    The Institute of Materials, 1995, p227

  • TABLE 1: Influence of wall thickness on CHmax, comparison with CHmax for a specimen tested with exposure conditions corresponding to in service conditions (T=450C ; pH2 =150bar)

    Reactor wall

    (Cooling rate 50C/h) Specimen

    (cooling rate 150C/h) Thickness

    (wall+overlay) (mm)

    100+6 150+6 200+6 250+6 300+6 40+6

    CHmax (ppm) 237 300 338 361 375 462

    TABLE 2: Exposure conditions giving same CHmax values than those obtained for different wall thickness

    Thickness (wall+overlay) (mm)

    100+6 150+6 200+6 250+6 300+6

    CHmax (ppm) 237 300 338 361 375 T (C) 450 450 450 450 450

    Cr (C/h) 150 150 150 150 150 Testing

    conditions pH2 (bar) 35 58 72 87 95

    TABLE 3: Testing conditions suggested for thedifferent domains of severity described on figure 14

    Reactor service conditions Testing specimen thickness (40 +5mm) Temp (C)

    H2max (ppm)

    Thickness (mm)

    Pressure (bar) Domain Temp (C) Pressure Cr (C/h)

    H2max spec

    >=450 >=250 >= 170 A 450 150 675 509 >=400 >=180 >= 140 B 450 150 150 475 >=350 >=130 >= 110 C 450 120 150 430 >=300 >=100 >= 80 D 450 90 150 375 >=250 >=80 >= 60 E 450 70 150 335

    450C

  • FIGURE 1: Typical hydrogen profiles through the wall of a reactor before and after cooling down

    1

    Overlay Base Material

    Profile before cooling down Profile after cooling down

    Ct

    Hyd

    roge

    n

    C 0

    C 0 1

    C 0 2

    C t 2

    1,00E-04

    1,00E-03

    1,00E-02

    1,00E-01

    1,00E+00

    1,00E+01

    1 1,5 2 2,5 3 3,51000/T

    S (p

    pm b

    ar1/

    2 )

    450C 25C

    Austenitic Stainless Steel

    2.25Cr-1Mo

    FIGURE 2: Relationship between hydrogen solubility and temperature (1000/T )

  • 1,00E-10

    1,00E-09

    1,00E-08

    1,00E-07

    1,00E-06

    1,00E-05

    1,00E-04

    1,00E-03

    1 1,5 2 2,5 3 3,51000/T

    450C 25C

    Austenitic Stainless Steel

    2.25Cr-1Mo

    D (c

    m2 /s

    )

    FIGURE 3 : Relationship between hydrogen diffusion coefficient and temperature (1000/T)

    a) b)

    FIGURE 4: Disbonding test specimens a) parrallelepipedic specimen 100*50*45

    b) ASTM G146 specimen diam 80*45

    Austenitic side w elded overlay

  • FIGURE 5: Hydrogen distribution in a disbonding test specimen

    1

    Overlay Base Material

    C 0 C01

    C 02

    C t 2

    Profile before cooling down

    Profile after cooling down

    Ct

    Hyd

    roge

    n co

    ncen

    trat

    ion

    T =425C

    100150200250300350400450500

    1 10 100 1000

    Cooling rate (C/h)

    200 bar H2 150 bar H2 100bar H2 50 bar H2

    FIGURE 6: Influence of cooling rate and hydrogen pressure on CHmax

    (Thickness 250+6 ; T=425C)

  • T =450C

    100150200250300350400450500

    1 10 100 1000

    Cooling rate (C/h)

    200 bar H2 150 bar H2 100bar H2 50 bar H2

    FIGURE 7: Influence of cooling rate and hydrogen pressure on CHmax

    (Thickness 250+6 ; T=450C)

    T =475C

    100150200250300350400450500

    1 10 100 1000

    Cooling rate (C/h)

    200 bar H2 150 bar H2 100bar H2 50 bar H2

    FIGURE 8: Influence of cooling rate and hydrogen pressure on CHmax

    (Thickness 250+6 ; T=475C)

  • T =450C

    100150200250300350400450500

    0 50 100 150 200 250 300 350 400

    wall thickness (mm)

    200 bar H2 150 bar H2 100bar H2 50 bar H2

    FIGURE 9: Influence of reactor wall thickness on CHmax

    (T=450C; tov = 4mm)

    INFLUENCE OF OVERLAY THICKNESS (T=450C, pH2 = 150bar))

    100150200250300350400450500

    0 50 100 150 200 250 300 350 400

    wall thickness (mm)

    4mm 6mm 8mm

    FIGURE 10: Influence of base material and overlay thickness

    (T=450C ; pH2 = 150bar)

  • 0

    5

    10

    15

    20

    25

    200 300 400 500 600 700

    CHmax (ppm)

    FIGURE 11: Percent of disbonded (D%) area versus CHmax calculated for

    different testing conditions

    Weld overlay

    0

    10

    20

    30

    40

    50

    60

    70

    100 200 300 400 500 600 700

    CHmax (ppm)

    1,25Cr-0,5Mo 2,25Cr-1Mo 3Cr-1Mo 2,25Cr-1Mo-V 3Cr-1Mo-V

    4321

    Testing conditions1 : 450C - 80bar -100C/h2: 450C - 100bar -150C/h3: 450C - 150bar -150C/h4: 450C - 150bar -675C/h

    FIGURE 12 : Results of disbonding tests performed on weld overlay, for different testing conditions

  • Weld overlay

    0

    10

    20

    30

    40

    50

    60

    70

    100 200 300 400 500 600 700

    CHmax (ppm)

    1,25Cr-0,5Mo 2,25Cr-1Mo 3Cr-1Mo 2,25Cr-1Mo-V 3Cr-1Mo-V

    4321

    Testing conditions1 : 450C - 80bar -100C/h2: 450C - 100bar -150C/h3: 450C - 150bar -150C/h4: 450C - 150bar -675C/h

    Reactor (450C -150bar)Wall thicknessA: 300+6B: 250+6C: 200+6D: 150+6E: 100+6

    E CD B A

    FIGURE 13 : Comparison of severity of testing conditions with actual hydrogen content expected in reactor walls

    FIGURE 14 : Definition of severity domains as a function of wall thickness and hydrogen pressure for a temperature of 450C

    50

    100

    150

    200

    F 81

    116

    143

    169 481472 425 455377298

    427418 367 392325257

    337330 298 318264208

    235231 208 223185146

    450 400

    350

    300

    250E

    D

    C

    B A

    0 0 50 100 150 200 250 300 350

    Wall thickness (mm)

    MAIN MENUPREVIOUS MENU---------------------------------Search CD-ROMSearch ResultsPrint

    copy_04: number: 0557301: 1copy_a: 02: 203: 304: 405: 506: 607: 708: 809: 910: 1011: 1112: 1213: 1314: 1415: 1516: 1617: 1718: 18