Upload
julie-dyke
View
221
Download
0
Embed Size (px)
Citation preview
IAEAInternational Atomic Energy
Agency
Radiation Protection in Paediatric Radiation Protection in Paediatric RadiologyRadiology
Radiation Protection of Radiation Protection of Children in Screen Film Children in Screen Film
RadiographyRadiographyL03L03
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
2
Educational objectives
At the end of the programme, the participantsshould:
•Become familiar with specific radiation protection issues in paediatric radiography
• Identify the features of radiographic imaging equipment used in paediatric radiology
•List important operational considerations in paediatric radiography
•Discuss important considerations in paediatric radiography using mobile X-ray units
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
3
Answer True or False
1. Added filtration will reduce the dose to the patient.
2. Short exposure time is a disadvantage.
3. Proper collimation reduce dose.4. Shielding of radiosensitive organs is
recommended in paediatric radiography.
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
4
Contents
• Justification in radiography• Practical optimisation in paediatric
radiography•Equipment related•Radiographic technique related
• Important consideration for mobile radiography
• Image quality and patient dose
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
5
Introduction
• Children have higher radiation sensitivity than adults due to a longer life expectancy
• For children under age of 10, the probability for fatal cancer is 2-3 times higher than for whole population
• The higher radio-sensitivity of the patients should be taken into account
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
6
Introduction
• Radiologists and radiographers should be specifically trained for paediatrics
• A paediatric radiological procedure should be individually planned and projections should be limited to what is absolutely necessary for a diagnosis
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
7
General recommendations
Key areas in radiation protection in paediatricradiology:
• Justification•Optimisation•Evaluation of patient dose and image quality
“Do you really need a glossy picture to make that diagnosis”
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
8
Justification in radiography
• Justification is required for all radiographic studies• Ask referring practitioner, patient, and/or family
about previous procedures• Use referral guidelines where appropriate and
available• Use alternative approaches, such as ultrasound,
MRI where appropriate• Consent, implied or explicit, is required for
justification• Include justification in clinical audit
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
9
Justification in radiography
• Referral guidelines for radiological examinations:
• EUROPEAN COMMISSION, Referral Guidelines for Imaging, Luxembourg, Radiation Protection 118, Office for Official Publications of the European Communities, Luxembourg (2001) and Update (2008)
• THE ROYAL COLLEGE OF RADIOLOGISTS, Making the Best use of Clinical Radiology Services (MBUR), 6th edition, RCR, London (2007)
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
10
Examples of radiography examinations not routinely
indicated
• Skull radiograph in a child with epilepsy• Skull radiograph in a child with headaches• Sinus radiograph in a child, under 5 years,
suspected of having sinusitis• Cervical spine radiograph in a child with
torticollis without trauma• Radiographs of the opposite side for
comparison in limb injury
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
11
Optimisation in radiography
• Justified studies must be optimised• Various actions taken contribute to
systematic dose savings (from a factor of two to ten, with the result that their combined effect can dramatically reduce dose)
• Sustain good practice through a quality assurance and constancy checking program
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
12
Optimisation in radiography
Selection of equipment:
• Influence on patient dose and image quality
•But, good radiographic technique is the main factor in improving quality without increasing dose
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
13
Practical optimisation measures in radiography (I)
• Have a standard type and number of projections for specific indications
• Views in addition to standard should only be performed on a case-by-case basis
• Use manual technique selection pending equipment developments on small patients or body parts
• Where practical use a long (or the recommended) Focus-to-Film Distance
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
14
Practical optimisation measures in radiography (II)
• Carefully collimate the X-ray beam to area of interest, excluding other regions, especially gonads, breast, thyroid and eyes
• Use appropriate gonad, thyroid, and breast shielding
• Fast film-screen combinations are acceptable for the majority of indications
• Antiscatter grid is often unnecessary in children – do not use grid for abdominal examination in patients under age of 3, for skull radiography for patients under age of 1 and any fluoroscopy examination unless high detail is required (Cook, V. Imaging, (13) 2001:229–238)
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
15
Practical optimisation measures in radiography (III)
• Use PA projections, where practical, for chest and spine radiographs
• Make sure the correct filtration is used to reduce entry dose
• Use as high a kVp as is consistent with examination requirements
• Consider additional filtration at higher kVp• Balance the use of a small focal spot size
and short exposure times
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
16
Practical optimisation measures in radiography (IV)
• Use of quality assessment, quality assurance and audit programs for all aspects of the department’s work, including film processing and justification
• Introduce and use a system that allows patient dose be assessed regularly
• Monitor reject rate and the causes (overexposure, underexposure, positioning, motion, and collimation problems)
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
17
Equipment, practice, dose and image quality
1. Generators
For paediatric examinations, the generator should be:• a high frequency• multi-pulse
(converter) • of sufficient power• nearly rectangular
waveform with minimal voltage ripple
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
18
Equipment, practice, dose and image quality
2. Exposure time
• When children are uncooperative they may need immobilization
• They have faster heart and respiratory rates
• Short exposure times improve quality without increasing dose
• Only possible with powerful generators and accurate exposure time switches
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
19
Equipment, practice, dose and image quality
3. Focal Spot• Small focal spot
• Improves image quality• May in some machines increase
exposure time and motion artefacts
• Choice depends on exposure parameters: time, kVp and FFD (Focus-to-Film Distance)
• Recommendation: focal spot should be 0.6 -1.3mm
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
20
Equipment, practice, dose and image quality
4. Additional filtration
• Additional filtration may lead to dose reduction
• 0.1 mm of Cu in addition to 2.5 mm of Al* • reduce ESAK by 20%• barely noticeable reduction
in image quality• Some modern systems can
automatically insert either 0.1mm or 0.2 mm Cu depending on the examination
*Cook, V., Imaging, (13) 2001:229–238
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
21
Dose reduction with added filtration
From: Mooney and Thomas : Dose reduction in a paediatric X-ray department following optimization of radiographic technique, BJR (77) 1998:852-860
Added filtration
0 mm Al 3 mm Al
Examination Mean ESD (Gy) Reduction
Abdomen AP 10 months(62 kVp)
200 30 %
Chest AP 10 months(55 kVp)
64 40 %
Pelvis AP 4 months(50 kVp)
94 51 %
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
22
Equipment, practice, dose and image quality
5. Exposure factors
Increased kVp (reduced mAs):• Greater penetration and less
absorption • Reduced patient dose for a
constant film density
Neonatal chest:• Minimum 60kVp: less contrast but
better assessment of lung parenchyma
• Lower kVp if looking for bone detail
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
23
Equipment, practice, dose and image quality
6. Antiscatter grid
• Often unnecessary in children because smaller irradiated volume (and mass) results in less scattered radiation.
• Limited improvement in image quality but increased dose of ~50% with the use of antiscatter grids
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
24
Antiscatter grids
• Antiscatter grid should be removable in paediatric equipment
• Remove antiscatter grid for:• abdominal imaging in young
children especially <3 years old
• skull imaging <1 year old • in most fluoroscopic imagingCook, V., Imaging, (13) 2001:229–238
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
25
Antiscatter grids
If used for children,Antiscatter grids should have*:
• Grid ratio (r) > 8:1 • Line numbers: >100 cm-1
• Low attenuation intersperse material, such as carbon fibre
Alternative: air gap technique(reduces the effect of scatterwithout dose increase, but theimage is magnified)
*Cook, V., Imaging, (13) 2001:229–238
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
26
Equipment, practice, dose and image quality
7. Automatic Exposure Control (AEC)• Generally not appropriate for small children• Sensors (size and geometry) are normally
designed for adult patients• AEC use may be associated with the use of the
grid (where the grid is not removable), which is frequently unnecessary
• AEC should have specific technical requirements for paediatrics
• If not appropriate or available, carefully applied exposure charts are preferred
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
27
Automatic Exposure Control
• Specially designed paediatric AEC:
•Small mobile detector for use behind a lead-free cassette
•Position can be selected with respect to the most important region of interest
•This must be done extremely carefully, as even minor patient movement may be disastrous
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
28
Equipment, practice, dose and image quality
8. Focus-to-film distance (FFD)
• Longer focus-to-film distances • Smaller skin dose
• Combined with a small object-to-film distance, results in less magnification (less geometric distortion) and improved quality
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
29
Equipment, practice, dose and image quality
9. Image receptors• Fast screen-film
combinations have advantages (reduction of dose) and limitations (reduced resolution)
• Low-absorbing materials in cassettes, tables, etc., are specially important in paediatric radiology (carbon fibre)
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
30
Film-screen systems
• Fast screen-film system:• shorter exposure times (requires a
good generator)• reduction in radiation dose and
prevention of artefacts• Recommendations:
• 200 speed: bone• 400 speed: general• >700 speed – constipation transit
abdominal radiographs, follow-up films, e.g. scoliosis and hips, swallowed foreign body,…
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
31
Equipment, practice, dose and image quality
10. Collimation
• The most important factor for improving image quality whilst also reducing dose
• The most common radiographic fault• Good collimation/coning is essential
to achieve better contrast and avoid exposing unnecessarily other body parts (dose reduction)
• Body parts outside the region of interest should not be in the X-ray field
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
32
Collimation
• Require a basic knowledge of paediatric pathology
•Lung fields extremely large in congestive heart failure & emphysematous pulmonary diseases
•Diaphragm, high in intestinal meteorism, chronic obstruction or digestive diseases
• Beam-limiting devices automatically adjusting the field size to the full size of the cassette are inappropriate for children
• Minimal deviation from the radiation and light beam may have large effects in relation to the usually small field of interest - check light beam diaphragm regularly
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
33
Collimation
• Alignment agreement among the collimators, radiation beam and the light beam must be regularly assessed
• Beyond the neonatal period, the tolerance for maximal field size should be less than 2 cm greater than the minimal
• In the neonatal period, the tolerance level should be reduced to 1.0 cm at each edge
• In paediatric patients, evidence of the field limits should be apparent by clear rims of unexposed film
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
34
Lateral skull radiograph (horizontal beam and
round cone)
Neonatal anteroposterior supine chest and
abdomen radiograph of newborn: all four cone
marks visible, with no extraneous body parts
included and lead masking of the gonads.
Cook, J.V., Imaging, 13 (2001), 229–238
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
35
Equipment, practice, dose and image quality
11. Shielding
• Standard equipment of lead-rubber shielding of the body in the immediate proximity of the diagnostic field
• Special shielding has to be added for certain examinations to protect against external scattered and extra-focal radiation
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
363636
Shielding
• For exposures of 60 - 80 kV, maximum gonadal dose reduction of about 30 to 40% can be obtained by shielding with 0.25 mm lead equivalent rubber immediately at the field edge
• However, this is only true when the protection is placed correctly at the field edge
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
373737
Shielding
• The gonads in "hot examinations", when they lie within or close to (nearer than 5 cm) the primary beam, should be protected whenever this is possible without impairing necessary diagnostic information
• It is best to make one's own lead contact shields for girls and lead capsules for boys
• Must be available in varied sizes
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
383838
Shielding
• With appropriate shielding the absorbed dose in the testes can be reduced by up to 95%
• In girls, shadow masks within the diaphragm of the collimator are as efficient as direct shields.
• When shielding of the female gonads is effective, the reduction of the absorbed dose in the ovaries can be about 50%
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
393939
Shielding
• The eyes should be shielded for X-ray examinations involving high absorbed doses in the eyes, e.g., for conventional tomography of the petrous bone, when patient cooperation permits
• The absorbed dose in the eyes can be reduced by 50% - 70%
• In any radiography of the skull the use of PA-projection rather than the AP-projection can reduce the absorbed dose in the eyes by 95%
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
40
Equipment, practice, dose and image quality
12. Patient Positioning and Immobilization• Patient positioning must be exact, whether or not
the patient co-operates. • In infants, toddlers and younger children
immobilization devices, properly applied, must ensure that: the patient does not move the beam can be centred correctly the film is obtained in the proper projection accurate collimation limits the field size exclusively to the
required area shielding of the remainder of the body is possible.
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
41
Patient Positioning and Immobilization
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
42
Patient Positioning and Immobilization
• Immobilization devices must be easy to use
• Their usefulness should be explained to the accompanying parent(s)
• Radiological staff members should only hold a patient under exceptional circumstances
• Even in quite young children the time allocation for an examination must include the time to explain the procedure not only to the parents but also to the child
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
43
Mobile radiography
• Mobile radiography is valuable on occasions when it is impossible for the patient to come to the radiology department
• It can result in • poorer quality images • unnecessary staff and patient exposures
• Where practicable, X-ray examinations should be carried out with fixed units in an imaging department
• Mobile units should only be used with those who cannot safely be moved to such a unit
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
44
Mobile radiography
• High output converter generators are recommended
• Capacitor discharge systems should be avoided (they have significant pre- and post-peak soft radiation)
• Appropriate collimation is essential to avoid exposing organs outside the diagnostic area of interest
• Other principles outlined above, should be followed with mobile radiography
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
45
Mobile radiography
• Scattered radiation must be managed to reduce dose to the patient, parents/guardians and to hospital personnel
• The advice of the medical physicist/radiation protection officer should be obtained on how best to do this.
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
46
Mobile radiography
Recommendations for Intensive Care Unit
(Duetting et. al. Pediat. Radiol. 29: 158-62 (1999)):• No additional protection for neighbouring
premature infants is necessary• The radiographer should wear a lead apron• Parents and personnel need not interrupt their
activities or leave the room during an X-ray examination
• When using a horizontal beam, the beam, must be directed away from other persons – use lead shield
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
47
Criteria related to images
• Incorrect positioning is the most frequent cause of inadequate image quality in paediatric radiographs
• Image criteria for the assessment of adequate positioning (symmetry and absence of tilting etc) are much more important in paediatric imaging than in adults
• A lower level of image quality than in adults may be acceptable for certain clinical indications
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
48
Criteria related to images
Guideline resources:
• European Guidelines on Quality Criteria for Diagnostic Radiographic Images in Paediatrics
• American College of Radiology
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
49
Quality Criteria List
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
5050
Chest-PA/AP projection
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
51
Chest radiography-PA/AP projection
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
52
Typical dose levels in paediatric radiography
Examination
ESAK (µGy)
Age
0 1 5 10 15
Abdomen AP 110 340 590 860 2010
Chest PA/AP 60 80 110 70 110
Pelvis AP 170 350 510 650 1300
Skull AP / 600 1250 / /
Skull LAT / 340 580 / /
NATIONAL RADIOLOGICAL PROTECTION BOARD, Doses to Patient from Medical X Ray Examinations in the UK: 2000 review, NRPB-W14, Chilton (2002).
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
53
ICRP-ISR “smart” message for paediatrics
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
54
http://rpop.iaea.org/RPoP/RPoP/Content/index.htm
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
555555
Summary
• Particular attention should be given to technical specifications of X-ray equipment
• Good radiographic technique is the main factor in improving quality without increasing dose for protocols used in X-ray paediatric radiology
• Justification of practice• Application of practical optimisation
measures in radiography
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
56
Answer True or False
1. Added filtration will reduce the dose to the patient.
2. Short exposure time is a disadvantage.
3. Proper collimation reduce dose.4. Shielding of radiosensitive organs is
recommended in paediatric radiography.
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
57
Answer True or False
1. True - Filtration absorbs low energy photons that are absorbed in patient’s skin and superficial organs and thus giving contributing to dose but not to image formation.
2. False - It prevents motion artefacts and unnecessary repetitions.
3. True - Collimation reduces exposed volume, and reduces scatter radiation that affects both image quality and dose.
4. True - It is especially important for radiosensitive organs as breast, gonads and eyes.
IAEA Radiation Protection in Paediatric Radiology L03.Radiation protection in screen-film radiography
58
References
• European Guidelines on Quality Criteria for Diagnostic Radiographic Images in Paediatrics, July 1996. EUR 16261. Available at: http://www.cordis.lu/fp5-euratom/src/lib_docs.htm
• Huda W, Assessment of the problem: paediatric doses in screen-film and digital radiography, Pediatr Radiol 34(Suppl 3) 2004:S173-S182
• Duetting,Foerste,Knoch,Darge and Troeger, Radiation exposure during chest X-ray examinations in a premature intensive care unit: phantom studies, Pediatr Radiol (29) 1999:158-162
• Mooney and Thomas : Dose reduction in a paediatric X-ray department following optimization of radiographic technique, BJR (77) 1998:852-860
• Cook, V., Radiation protection and quality assurance in paediatric radiology, Imaging, (13) 2001:229–238.