Endoscope reprocessing –from the washer to the patient

Wayne Spencer

Page 2: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Reprocessing Cycle

Pre-Cleaning

Transport

Manual Cleaning + Leak Test

Auto Cleaning

Disinfection

Inspection

Packaging

Transport

Storage

Use

Page 3: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Inspection• It is commonly accepted that reprocessed devices should be inspected to

show that they clean and safe for reuse.

• But in practice does this mean only the outside of the scope?

• Is it practical to insect the inside of the scope on a routine basis?

• Or is sampling the scope – e.g. testing the output the answer?

Ofstead, Cori et al (2018). Residual moisture and waterborne pathogens inside

flexible endoscopes: Evidence from a multisite study of endoscope drying

effectiveness. American Journal of Infection Control. 10.1016/j.ajic.2018.03.002.

Page 4: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Packaging• Endoscopy packaging can be divided up into two groups:

– Packaging for sterilization and maintaining a sterile state– Packaging when used as an alternative to post disinfection

storage

• Sterilization packaging - well defined standards (EN 868 Series and EN ISO 11607 Series)– In most cases the method used needs to be compatible with low

temperature systems (most commonly not cellulose based for H2O2 methods)

• Post disinfection systems - no standards!– Is it portable storage or packaging?– Lack of long-term published evidence in this area– Check the new Standards and Recommended Practices

document

Page 5: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Portable storage systems

Page 6: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Standards and Recommended Practices: Portable storage systems

• Should not reduce standards, but maintain patient safety.

• Manufacturers of such systems need to provide all relevant type-test data to verify and validate process effectiveness and to identify key variables that need maintaining to ensure continued operation in line with type testing.

• Where endoscopes need to be processed in a controlled environment storage cabinet first, what is the actual validated drying time required for each type of endoscope?

• Is the storage period based upon validated data and will it require retesting if a new type of endoscope is added to the process.

• Periodic maintenance and validation is required to verify performance.

• The manufacturer should provide a table and schedule of tests and identify maintenance tasks.

• Storage times of endoscopes in such systems must be verified and checked at least annually.

• Consideration should be given to keeping the validated storage time to a minimum acceptable period which meets the needs of the user.

Page 7: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Standards and Recommended

Practices -validation:

Page 8: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Is it a storage cabinet or a drying cabinet?

Page 9: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Is it a storage cabinet or a drying cabinet?

• EN 16442: – Storage cabinet:

• equipment controlled by an automatic control system that maintains the microbiological quality of processed thermolabile endoscope

• Can be single or double ended (pass through)

– Drying:• additional feature of a storage cabinet carried out in

the sequence as regulated by the automatic controller to remove moisture

• if the storage cabinet does not have a drying phase, it has to be specified that the endoscope (outside surfaces and internal channels) has to be dried before storage

Page 10: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Standards and Recommended

Practices -Validation:

Page 11: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

“Many outbreaks of HCAI after endoscopy and cross contamination due to Ps. aeruginosa, S. marcescens, M. tuberculosis and M. chelonae have been associated with drying of flexible endoscopes without ethanol flushing or lack of a drying procedure”

Kovaleva et al (2013) Clinical Microbiology reviews 26 231-254

“Endoscope reprocessing guidelines are inconsistent with one another or give no exact recommendations about drying and storage of flexible endoscopes. There is no conclusive evidence on the length of time endoscopes can be safely stored before requiring re-disinfection and before they pose a contamination risk.”

Kovaleva et al (2017) Journal of Hospital infection 97 319-328

Page 12: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

“Endoscope drying is as important to the prevention of nosocomial

infection as cleaning and high level disinfection.

Whereas wet or inadequately dried endoscopes pose an increased risk of

contamination and have been associated with transmission of water borne

micro-organisms and nosocomial infection, thoroughly dried (and properly

cleaned and disinfected) endoscopes have not been linked to nosocomial

infection”.

Page 13: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

UK Guidelines (2017)(BSG)

• The current guidance is that endoscopes need to be reprocessed before reuse if they have been stored for >3 hours because of bacterial growth in damp lumens.

• However, if scopes are stored in cabinets with a constant flow of clean air through all lumens to keep them dry, the time before reprocessing can be extended to 72 hours to 31 days ( period must be validated).

• The use of an alcohol flush to aid drying is not recommended in the UK due to the fixative properties of alcohol.

• When transporting endoscopes to and from areas outside the endoscopy unit, they must be transferred in in a rigid container with a solid lid.

Page 14: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

French Guidelines

• Disinfect endoscopes stored for more than 12 hours.

• Full decontamination procedure (as opposed to a shortened cycle) if more than one week has elapsed since last being processed.

Pineau et al (2008) Journal of Hospital Infection 68, 59-65

Page 15: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

• If endoscopes are not dried, they should be used within 4 hours.

• If dried within a cabinet, they can be stored for one month.

Dutch Guidelines (2017)

Professional Standard Handbook - Cleaning and Disinfection of Flexible Endoscopes

Version 4.1, 2017

Page 16: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Survival Of Bacteria On Drying

Page 17: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

• 10 cfu/100ml not Ps. aeruginosa or atypical mycobacteria

• Therefore, some bacteria are permissible and many endoscopy units carry out a risk assessment if the count is higher

Generally Accepted Requirements For Water Quality

With thanks to Christina Bradley

Page 18: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

If 9 cfu are present, this is a pass for EN 15883

• But if the multiplication time for the organism is 20 minutes then

– 18 cfu after 20 mins

– 36 cfu after 40 mins

– 72 cfu after 1 hour

– 2304 cfu after 3 hours

– 4.8 x 109 cfu after 24 hours

If The Channels Are Left Wet

With thanks to Christina Bradley

Page 19: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Endoscope At Room Temperature

Pre 30 mins 3 hours 6 hours 18 hours 24 hours 72 hours

Biopsy 4.54 3.95 3.25 4.18 5.98 6.57 7.41

Suction 4.74 3.64 3.55 3.89 5.92 6.63 7.54

Air 4.80 3.66 3.45 3.74 6.32 6.66 7.66

Water 4.83 3.34 2.70 3.79 5.58 6.63 7.26

Raiser 4.71 3.04 2.38 4.45 4.85 6.07 6.49

0.00

1.00

2.00

3.00

4.00

5.00

6.00

7.00

8.00

9.00

With thanks to Christina Bradley

Page 20: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Endoscope In Drying Cabinet

Pre 30 mins 3 hours 6 hours 18 hours 24 hours 72 hours

Biopsy 4.54 1.11 0.44 0.63 0.00 0.07 0.00

Suction 4.74 1.44 0.94 0.74 0.00 0.98 0.00

Air 4.80 1.95 1.91 1.96 0.00 0.24 0.00

Water 4.83 2.15 1.66 1.24 0.00 0.00 0.00

Raiser 4.71 3.19 2.90 3.78 0.00 0.00 0.00

0.00

1.00

2.00

3.00

4.00

5.00

6.00

With thanks to Christina Bradley

Page 21: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Can I Have My Cake and Eat it?

I used two hours before I put it in the drying cabinet – so can I have an hour

afterwards?

Page 22: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

If 9 cfu are present, this is a pass for EN 15883

• But if the multiplication time for the organism is 20 minutes then

– 18 cfu after 20 mins

– 36 cfu after 40 mins

– 72 cfu after 1 hour

– 576 cfu after 2 hours

• And what if I then don’t leave it in for the full drying period?

If The Channels Are Left Wet

Page 23: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Drying may not affect all types of microorganism in the same way:

– Gram positive bacteria – may survive drying anyway

– Gram negative bacteria – may thrive in the presence of even a little moisture

Drying is not a disinfection process!

If The Channels Are Not Dried Completely:

Page 24: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Even if the channels are dry:

Page 25: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

• A = Contaminated endoscope sampled at time 0 and 24 hours

• B = Contaminated endoscope sampled at time 0, 3, 6, 12, 18 and 24 hours

Testing Of Drying Cabinets

0

250

500

750

1000

0

2500

5000

7500

10000

AB

0 5 10 15 20 25

Time (hours).

CF

U r

em

ain

ing

.

Group

A

B

Death of organisms over time.

With thanks to Christina Bradley

Page 26: IDI Annual Conference 2018 Lecture: Endoscope …–Packaging for sterilization and maintaining a sterile state –Packaging when used as an alternative to post disinfection ... maintaining

Conclusions• Even basic visual inspection of endoscopes is essential to maintaining

a safe process• If sterilizing, use packaging that meets the EN and ISO standards but

make sure it is compatible with the process• Given the lack of standards, ask you post-disinfection storage system

supplier to provide you with evidence of its performance over the time period required and for the endoscope types you intend to process

• Use Storage/Drying cabinets according to the manufacturers instructions, taking note of the drying time

• Plan which endoscopes you wish to place in the cabinet in advance –the default position should be to dry if in doubt

• Cabinets and storage/drying systems should be validated regularly and………

Should endoscopes??