Upload
george-lopez
View
213
Download
1
Embed Size (px)
Citation preview
Abstract— Tightly controlling blood glucose in selected
patients has been shown to significantly decrease complications
such as infection, neurologic injury, and respiratory failure.
Intravenous microdialysis has been used to continuously
monitor glucose levels yet variable recovery may occur due to
probe fouling, osmosis, and ultrafiltration. A portable
microdialysis system is presented that: (1) uses retrodialysis
with fluorescein to monitor probe fouling, (2) measures
downstream flow rate to gauge fluid losses from
ultrafiltration/osmosis, and (3) employs a wireless
microcontroller to sequence the above events and
record/transmit the data. In vitro glucose microdialysis is
performed with a CMA Microdialysis IView catheter and
Medtronic CGMS® iPro™ glucose sensor. Providing
self-diagnostics can eventually lead to a medical embedded
system with calibration, safety alarms, and wireless
communication of blood glucose readings to a drug delivery
system for closed loop control.
I. INTRODUCTION
ICRODIALYSIS is a versatile technique for sampling
molecules (analytes) from tissues or blood in living
animals and man. Microdialysis is performed by circulating a
small volume of buffer through a porous polymeric catheter
(probe) that is about 500 microns in diameter and 1 – 4 cm in
length. Pores in the probe allow the diffusion of molecules
from blood or the tissue of interest into the circulating buffer.
The buffer exiting the probe, also known as the dialysate, is
then collected for later analysis. Microdialysis has been used
to continuously measure glucose levels in the subcutaneous
tissue of diabetics [1]-[3].
Recently, commercially manufactured microdialysis
probes have become available for continuous blood glucose
monitoring (CMA Microdialysis, Chelmsford, MA, U.S.A.).
Researchers have demonstrated that tight blood glucose
control in critically ill patients has multiple benefits [4]. Such
control is achieved by continuous intravenous insulin
infusion. It is not without certain risks; subgroups of patients
Manuscript received February 15, 2009. This work was supported in part by
the the Commonwealth of Pennsylvania under A Pennsylvania
Infrastructure And Technology Alliance (PITA) Grant.
A. J. Rosenbloom is with the Institute For Complex Engineered Systems,
Carnegie Mellon University, Pittsburgh, PA. 15213 USA; Phone:
412-268-8638; fax: 412-268-6571; e-mail: [email protected].
H.R. Gandhi is with the Electrical and Computer Engineering Dept.,
Carnegie Mellon University, Pittsburgh, PA. 15213 USA; Phone:
412-268-5212; e-mail: [email protected].
G.L. Subrebost is with the Institute For Complex Engineered Systems,
Carnegie Mellon University, Pittsburgh, PA. 15213, Carnegie Mellon
University, Pittsburgh, PA 15213 USA; Phone: 412-268-5212; e-mail:
are actually harmed by tight control, perhaps by the ill effects
of hypoglycemia [5]. Blood glucose must be followed
closely in order to keep levels within 80 to 140 mg/dL.
Measurements are normally taken every 1 – 2 hours to adjust
the insulin dose properly. Continuous blood glucose
monitoring via microdialysis would eliminate repeated
testing and allow adjustment of glucose level more smoothly,
quickly and with increased safety. However, there are factors
that potentially create variability with estimates of blood
glucose obtained by microdialysis. Fouling of the probe, i.e.
clogging of its pores by blood proteins or clotting, can
decrease the recovery of glucose. Furthermore, hydrostatic
or osmotic pressure differences between the microdialysis
buffer and the blood or tissue being sampled will cause a net
flow of fluid across the probe membrane (i.e. ultrafiltration
or osmosis of water). Ultrafiltration into the probe will
augment analyte recovery while ultrafiltration out of the
probe will reduce recovery. The importance of ultrafiltration
increases with membranes having larger pore sizes [6].
Microdialysis modeling has been used to provide more
insight to the complex trans-membrane dynamics [7].
The variability in analyte recovery caused by clogging,
osmosis, and ultrafiltration can become important when
potentially harmful therapy will be based on glucose
readings. Furthermore, intravenous microdialysis systems
can fail for many reasons such as the probe pulling out of the
vein, broken connections or fluidic lines, clotting in the vein,
etc. In order for intravenous microdialysis to move toward
being a robust clinically useful modality, there must be
alarms to warn of system failure and self-checks to confirm
adequate function. These capabilities would ideally be
integral to the microdialysis setup. In this paper we
demonstrate a microdialysis system that performs continuous
glucose measurements using a commercially available
glucose electrochemical sensor. We explore methods to
make the system self-monitoring.
The two monitoring methods are: (1) detection of flow rate
after the microdialysis probe – this will confirm pump
function and system integrity, and detect excess fluid being
taken from or added to the system by ultrafiltration or
osmosis, (2) modified retrodialysis to assess probe patency
and membrane diffusion capacity. Retrodialysis is normally
performed by adding a marker molecule very similar to that
being sampled to the dialysis buffer in order to calibrate the
recovery of analyte [8]. The diffusion rate of the marker out
of the probe is semi-quantitative for determining the
diffusion rate of the similar analyte into the probe. This and
other methods of calibration [9] are particularly useful in
tissue microdialysis where the kinetics of transport of
analytes into the microdialysis probe are very complex.
2009 IEEE/ICME International Conference on Complex Medical Engineering
Glucose Microdialysis with Continuous On-Board Probe
Performance Monitoring
Alan John Rosenbloom, Heer Robin Gandhi, George Lopez Subrebost
M
978-1-4244-3316-2/09/$25.00 ©2009 IEEE
These methods are less applicable with bloodstream
microdialysis of small molecules such as glucose. Due to a
continuous blood flow that mimics sampling from a stirred
solution, the diffusion rate of small molecules is very rapid,
allowing good equilibration with the microdialysis buffer.
Thus, the principal barrier to analyte recovery is fouling
(clogging) or clot formation on the probe membrane. We
confirm probe patency by continuously quantitating
fluorescein, a fluorescent molecule. Numerous researchers
have demonstrated miniature fluorescence detection systems
with various dyes [10-12]. Fluorescein is approved for
intravenous use in humans and has the potential for use in
clinical settings. We explore continuous flow rate
determination after the probe to quantitate ultrafiltration,
confirm fluidic transport within the system and to detect
system failures such as kinked or broken lines or
connections. By using ‘timing marks’, created by
photobleaching fluorescein present in the microdialysis
buffer and tracking this bleached ‘plug’ further downstream,
time-of-flight calculations are used to determine volumetric
flow rate [13]. Alternate methods such as ultraminiature
thermodilution sensors can also be used.
II. METHODS AND MATERIALS
A. Optofluidic Setup
A CMA 107 syringe pump (CMA Microdialysis,
Chelmsford, MA, U.S.A.) is used to provide various volume
flow rates from 0.1 to 10 µL (see Figure 1). In vitro glucose
sampling was performed with an IView CMA 64
microdialysis catheter (also manufactured by CMA
Microdialysis). This catheter is intended for continuous
intravenous monitoring of hospitalized patients during
surgery, intensive care or in the general wards [14]. With a
membrane cut-off of approximately 20,000 Daltons, the
catheter can also be used to measure free fractions of drugs in
blood during pharmacokinetic and pharmacodynamic
studies. The porous membrane material is made of
polyarylethersulphone (PAES) with an outer diameter of 0.6
mm and length of 10 mm. The outlet tubing of the catheter
(inner diameter 120 µm), normally connected to microvials
for fraction collection of the dialysate, was instead
lengthened to accommodate the monitoring schemes
mentioned above by attaching Tygon tubing (inner diameter
508 µm).
Downstream from the microdialysis catheter, a
photobleaching and fluorescence detection region has been
assembled to perform fouling and ultrafiltration
measurements. Photobleaching of the fluorescein in the
dialysate is achieved by locating the dialysate tubing within a
beam of focused light from a light emitting diode (LED),
model Luxeon V, LXHL-LB5C (Philips Lumileds, San Jose,
CA, U.S.A.). This LED has a Lambertian radiation pattern,
Fig. 1. Continuous glucose microdialysis setup that uses a CMA Microdialysis IView CMA 64 microdialysis catheter and Medtronic CGMS®
iPro™ glucose sensor. An embedded systems microcontroller is used to control the photobleaching as well as monitor the fluorescence of the
dialysate.
Fig. 2. (Top) Medtronic CGMS® iPro™ data recorder and sensor probe
used to continuously record dialysate glucose values. (Bottom)
Amperometric readings from the Medtronic CGMS® iPro™ glucose sensor
placed inside a tube with inner diameter of 794 µm and connected to a 10
mL syringe. Large dips between reading plateaus are due to manual
flushings from the syringe.
outlet luminous flux of 48 lumens, and has a dominant
wavelength of 470 nm, which is necessary for exciting
fluorescein. A narrow beam lens (Model FLP, Fraen Corp.,
Reading, MA, U.S.A.) with an aspheric profile, and total
beam divergence of 12 degrees, is mounted over the LED to
provide a collimated light source. Another narrow beams
lens is placed in front of the collimated light (forming an oval
enclosure) to focus the beam to the dialysate tubing for
photobleaching. Due to significant heating, the LED is
mounted onto an aluminum heat sink to provide adequate
temperature regulation.
A miniature fluorometer setup (8 cm x 4 cm x 2.5 cm) was
made from a machined plastic block (acrylonitrile butadiene
styrene) to measure fluorescence of the dialysate as well as
the photobleaching front. A collimated light source is
provided by another Luxeon V LED and FLP series narrow
beam lens. To limit the output light from corrupting the
fluorescence detection, the collimated light is filtered by an
exciter (ET470/40x, Chroma Tech. Corp., Rockingham, VT,
U.S.A). An aspheric lens (Model 352330-A, ThorLabs Inc.,
Newton, NJ, U.S.A.) with a diameter of 6.35 mm, focal
length of 3.1 mm, and numerical aperture of 0.68 is used to
focus the light to a 0.5 mm spot on the dialysate tubing.
Fluorescence emission is detected orthogonally from the
light source path. A second aspheric lens with the same
specifications is used to magnify the emitted light output.
This output is filtered with an emitter (ET525/50m, Chroma
Tech. Corp.) and collected by a silicon photodiode (BPW21,
Vishay Semiconductor, Heilbronn, Germany). The output
signal from the photodiode is conditioned and amplified by
an embedded systems microcontroller that will be further
described.
B. Glucose Sensing and Reagents
100 µM fluorescein in 1X calcium magnesium free –
phosphate buffered saline (CMF-PBS) was used as the
source buffer for the syringe pump. Fluorescein (MW 332
g/mol) was obtained from Sigma Aldrich (St. Louis, MO,
U.S.A.). To test the response of the photodiode, eight
different fluorescein concentrations were prepared from 5 to
85 µM.
The microdialysis probe was placed in a 15 mL test tube
that contained glucose at a concentration of 180 mg/dL in a
1X CMF-PBS buffer. Anhydrous dextrose (glucose, MW
180 g/mol) was obtained from Fisher Scientific (Pittsburgh,
PA, U.S.A.).
An electrochemical glucose sensor, CGMS® iPro™
(Medtronic Inc., Northridge, CA) was used to continuously
record dialysate glucose values (see Figure 2a). The outer
diameter and total length of the sensor tip is approximately
650 µm and 2 cm, respectively. The sensor is typically used
by diabetic patients to monitor subcutaneous glucose levels
and provides a reading every 5 minutes. A calibration was
performed to determine the linearity of the glucose readings
(see Figure 2b). The glucose sensor is placed downstream
from the microdialysis probe inside a larger diameter Tygon
tubing (inner diameter of 794 µm) in order to accommodate
the larger size of the sensor tip. Glucose readings are stored
on-board using a recorder that attaches to the backside of the
probe and can be subsequently downloaded to a computer
using a wireless connection.
C. Electrical Setup
An embedded systems microcontroller (MSP430, Texas
Instruments, Dallas, TX, U.S.A.) is used to control the
sequence of photobleaching and fluorometer measurements
(see Figure 3). A custom printed circuit board, attached to the
microcontroller, is used to efficiently amplify the photodiode
signal and determine the fluorescence level in the dialysate.
Its secondary tasks are to sustain itself in terms of power
usage as well as retaining or transmitting the data that it
collects from the photodiode. To perform these tasks the
electrical setup is governed by a central microcontroller that
controls the functioning of the complete system. The system
is divided into three functioning modules.
1) Analog Amplification and Filtering
Fluorescence from the dye emits green light at a 521 nm
wavelength when blue light (480 nm wavelength) is
projected on to the dialysate tubing. The photodiode is
subjected to this emission radiation as well as other external
noise sources (60 Hz noise from light sources such as
fluorescent light tubes). To suppress this noise, the source
LED light is modulated to provide an identifier for
subsequent filtering and amplification. The fundamental
signal frequency chosen for this operation is 965 Hz and an
active band-pass analog filter (2nd
order, signal gain 40 dB) is
implemented using discrete off-the-shelf components. The
signal gain is programmable using a digital potentiometer so
that the central microcontroller can calibrate the gain of the
system based on the input signal level. This prevents the
signal from saturating if the optofluidic setup is altered. The
amplified and filtered analog signal is sent over to the
on-chip analog to digital converter (ADC) on the central
microcontroller.
2) Digital Calibration and Signal Level Detection
To measure the fluorescence level digitally the LED
generates a 965 Hz signal that is focused on the dialysate
tubing for approximately 500 ms. Within this time frame, the
ADC is programmed to sample the amplified signal for 2,560
samples. The average DC level and AC RMS level of the
signal are measured using digital computations on the
microcontroller. The analog gain can be calibrated using this
information and all the data henceforth is measured relative
to the calibrated signal. The AC RMS signal level is what
denotes the fluorescence level.
3) Storing and Transmitting Data
The electrical system is powered by a 3.7 V, 400 mA
lithium ion battery. Two dedicated ADCs are used to
measure the battery voltage and input charging voltage. The
microcontroller charges the battery from the external
charging voltage when the battery level is low using a voltage
supply of 4.5-5.5 V. Based on the power consumption of the
electrical setup, it should be able to sustain itself on battery
power for at least 5 days. The AC RMS values of the signal
are stored in the data memory of the microcontroller. A
CC2500 2.4GHz RF chip, embedded along with the
microcontroller, transmits the data stored to a receiver node
where it can be viewed and archived.
III. EXPERIMENTAL RESULTS
Serial dilutions of fluorescein (from 5 to 100 µM) were
introduced into the setup in order to test whether the
photodiode was accurately measuring fluorescence levels as
shown in Figure 4. Calibration of the photodiode output is
performed by filling the dialysate tubing with CMF-PBS
buffer solely and setting this to 0 µM fluorescein. The
photodiode output recorded at 100 µM fluorescein was then
used to determine a conversion factor between voltage and
concentration. Each fluorescein concentration was
introduced for 40-60 seconds and the data was recorded by
the microcontroller every second during an LED pulse. The
data shows very good linearity as the concentration was
increased. While the fluorescein was stagnant within the
setup, a noticeable decline in fluorescence was observed.
Although the power supply on the LED was low (60 mA),
some inadvertent photobleaching occurred even though the
LED is only powered for 500 ms.
Fig. 4. Plot showing photodiode response with varying fluorescein
concentration. Calibration was done by using CMF-PBS buffer to set the 0
mM floor, while the 100 µM fluorescein photodiode output was used to
arrive at a voltage-to-concentration conversion factor.
In order to calculate volumetric flow rate in the dialysate
tubing, a photobleaching section was placed approximately
35 mm upstream from the fluorometer measurement. The
syringe pump was filled with 100 µM fluorescein and set to 2
µL/min. A tubing length of approximately 15 mm was
photobleached with a high LED intensity (500 mA) for 1
minute. Based on the inner diameter of the dialysate tubing
(508 µm), it should take 3.5 minutes for the photobleached
Fig. 3. Embedded systems controller includes a Texas Instruments
MSP430 microcontroller and a custom printed circuit board for signal
amplification and filtering.
plug to reach the fluorometer. As shown in Figure 5, this plug
takes about 3.7 minutes for the fluorescence intensity to
begin a steep decline caused by the photobleaching. The
difference in timing is probably due to inaccurate estimation
in length between the upstream bleaching and downstream
fluorescence measurement. A second photobleaching event,
as shown in Figure 5, was performed with a lower LED
intensity (275 mA) for 1 minute as well.
Fig. 5. Plot showing photobleached ‘plug’ passing the fluorometer
measurement region. A 12 mm section of outlet tubing was photobleached
upstream from the fluorometer. A high (500 mA) and low (275 mA) LED
power setting created the left and right dips, respectively.
Lastly, the Medtronic glucose sensor is combined with the
modified retrodialysis method in order to evaluate the
accuracy of using fluorescein as a metric for glucose
trans-membrane diffusion. The microdialysis catheter was
placed into a 15 mL test tube that contained a 180 mg/dL
glucose solution. Unfortunately, a mixing stirrer was not
placed inside the test tube to provide adequate refreshing of
the glucose solution. Before taking glucose readings, it is
necessary to allow for a 2 hour initialization time after the
sensor is wetted. A 100 µM fluorescein solution was used as
the microdialysis buffer while the volumetric flow rate on the
syringe pump was varied to 4 different intervals (2, 5, and 10
µL/min). A low LED intensity was used for the fluorometer
measurement (60 mA) and data was recorded every second.
The photobleaching setup was not used in this experiment.
Figure 6 shows two curves representing the relative
recovery of glucose and relative loss of fluorescein as the
flow rate was varied. A small flow rate causes the fluorescein
to leach out of the probe, while also providing adequate time
for the glucose to diffuse into the probe, yet this requires a
longer waiting period to perform the glucose sensing and
monitoring methods due to the length of the dialysate tubing.
High flow rate causes the fluorescein to bypass the diffusion
process out of the probe almost completely. A 10 µL/min.
flow rate shows a relative recovery of 80% fluorescein and
30% glucose, while a 2 µL/min. flow rate shows a relative
recovery of 35% fluorescein and 60% glucose. Since the
fluorescein molecule is approximately twice the size of
glucose (332 g/mol vs. 180 g/mol), it should provide an early
warning if fouling begins to occur on the probe membrane.
Further in vitro recovery tests will be performed using whole
blood before testing moves to a clinical trial with intensive
care unit patients. In addition, ultrafiltration will be induced
by altering the height of the dialysate tubing end, which is
exposed to atmospheric pressure. This should demonstrate
the adequacy of continuously measuring the fluorescein as a
gauge for probe fouling as well as measuring volumetric flow
rate to determine ultrafiltration values.
Fig. 6. In vitro glucose relative recovery and fluorescein relative loss at
varying volumetric flow rate. The microdialysis probe was placed in a test
tube containing 180 mg/dL glucose while the microdialysis buffer
contained 100 µM fluorescein.
IV. FUTURE APPLICATIONS
An embedded system microcontroller can be used to
perform the previously mentioned monitoring functions as
well as control the microdialysis flow rate. It is well known
that the slower the flow rate of microdialysis buffer, the more
complete the equilibration of analyte concentration between
sample fluid and buffer in the probe. However, very slow
flow rates impose a large delay in transporting the fluid to the
analyzer section of the system. Also, in-line analyzers require
a minimum sample volume. With sophisticated control of
microdialysis flow rate, one can achieve a highly efficient
system that collects sample at a slow rate for a set time. A
higher flow rate can be used to transport the sample to the
in-line analyzer. Finally, a slow rate can again be used to
allow time for the sample volume to reach steady state with
the analyzer components. We plan to integrate a servo
actuator to control of the syringe pump flow rate. The use of
an embedded systems microcontroller along with
microdialysis has other benefits as well. In the future, the
microcontroller can contain alarms, provide continuous
wireless communication of glucose levels, and communicate
with drug delivery systems to achieve closed loop control of
blood glucose. This technology is applicable to other drug
therapies as well as insulin control of glucose level.
ACKNOWLEDGMENT
The authors would like to thank CMA Microdialysis for
donating the IView CMA 64 microdialysis catheter.
REFERENCES
[1] L. Heinemann, Glucose Monitoring Study Group, "Continuous
glucose monitoring by means of the microdialysis technique:
underlying fundamental aspects,” Diabetes Technology &
Therapeutics, vol. 5, no. 4, pp. 545-561, Nov. 2003.
[2] M. Varalli, et al., "A microdialysis technique for continuous
subcutaneous glucose monitoring in diabetic patients (part 2),"
Biosensors & Bioelectronics, vol. 18, no. 7, pp. 899-905, July 2003.
[3] F. Ricci, D. Moscone, G. Palleschi, “Ex vivo continuous glucose
monitoring with microdalysis technique: the example of GlucoDay,”
IEEE Sensors Journal, vol. 8, no. 1, pp. 63-70, Jan. 2008.
[4] G. Van den Berghe, P. Wouters, F. Weekers, C. Verwaest, F.
Bruyninckx, M. Schetz, D. Vlasselaers, P. Ferdinande, P. Lauwers, R.
Bouillon, "Intensive insulin therapy in the critically ill patients," New
England Journal of Medicine, vol. 345, no. 19, pp. 1359-1367, Nov.
2001.
[5] Van den Berghe, A. Wilmer, G. Hermans, W. Meersseman, P. J.
Wouters, I. Milants, W. E. Van, H. Bobbaers, R. Bouillon, "Intensive
insulin therapy in the medical ICU,", vol. 354, no. 5, pp. 449-461,
Feb. 2006.
[6] A. J. Rosenbloom, R. Ferris, D. M. Sipe, S. A. Riddler, N. C.
Connolly, K. Abe, and T. L. Whiteside, "In vitro and in vivo protein
sampling by combined microdialysis and ultrafiltration," Journal of
Immunological Methods, vol. 309, no. 1-2, pp. 55-68, 2006.
[7] P. Bungay, P. Morrison, R. Dedrick, “Steady-state theory for
quantitative microdialysis of solutes and water in vivo and in vitro,”
Life Sciences, vol. 46, no. 2, pp. 105-119, Jan. 1990.
[8] M. R. Bouw and M. Hammarlund-Udenaes, "Methodological aspects
of the use of a calibrator in in vivo microdialysis-further development
of the retrodialysis method," Pharmaceutical Research, vol. 15, no.
11, pp. 1673-1679, 1998.
[9] M. J. Cano-Cebrián, T. Zornosa, A. Polache, and L. Granero,
"Quantitative in vivo microdialysis in pharmacokinetic studies: some
reminders," Current Drug Metabolism, vol. 6, pp. 83-90, 2005.
[10] M. L. Chabinyc, D. T. Chiu, J. C. McDonald, A. D. Stroock, J. F.
Christian, A. M. Karger, and G. M. Whitesides, "An integrated
fluorescence detection system in poly(dimethylsiloxane) for
microfluidic applications," Analytical Chemistry, vol. 73, pp.
4491-4498, 2001.
[11] L. Novak, P. Neuzil, J. Pipper, Y. Zhang, and S. Lee, "An integrated
fluorescence detection system for lab-on-a-chip applications," Lab on
a Chip, vol. 7, pp. 27-29, 2007.
[12] E. Thrush, O. Levi, L. J. Cook, J. Deich, A. Kurtz, S. J. Smith, W. E.
Moerner, and J. S. Harris, "Monolithically integrated semiconductor
fluorescence sensor for microfluidic applications," Sensors and
Actuators B, vol. 105, pp. 393-399, 2005.
[13] E. Harel, C. Hilty, K. Koen, E. McDonnell, A. Pines, “Time-of-flight
flow imaging of two-component flow inside microfluidic chip,”
Physical Review Letters, vol. 98, no. 1, pp. 17601-(1-4), Jan. 2007.
[14] CMA Microdialysis, Product Note on IView CMA 64 Microdialysis
Catheter, http://www.microdialysis.se/