24
IMPLANT SYSTEM SSO ® / STO ® / STW ® FIELD OF APPLICATION IMPLANT SYSTEM FOR ENOSSAL DENTAL IMPLANTATIONS

IMPLANT SYSTEM SSO/ STO / STW SYSTEM SSO® / STO® / STW ® FIELD OF APPLICATION IMPLANT SYSTEM FOR ENOSSAL DENTAL IMPLANTATIONS 2 1.SURGERY 1.1 Tap for lower jaw (only for use with

Embed Size (px)

Citation preview

IMPLANT SYSTEM SSO® / STO® / STW®

FIELD OF APPLICATION IMPLANT SYSTEM FOR ENOSSAL DENTAL IMPLANTATIONS

2

1.SURGERY

1.1 Tap for lower jaw (only for use with SSO®)

TAP 1 Size 3.3 (for implants Ø 3.3 mm) Size 4.1 (for implants Ø 4.1 mm) Size 4.8 (for implants Ø 4.8 mm)

Ratchet (Position “IN”) insert clockwise

Genuine outer packaging with implant

Open packaging using the flap. Remove the label and stick in into your patient’s record.

The opened packaging contains the implant in a sterile tube (primary packaging).

1.2 Implant packaging SSO®/STO®/OSS®

1.3 Remove the implant from the sterile tube.

1. Open the lid. The implant is connected to the lid with a break joint.

2. Remove the implant without touching the inner walls of the tube.

3. For implants of the Premium line the implant comes with an insertion tool already attached.

Fasten screw with HT 1.25 in UZS*

Insertion tool IT 1,2,3 STOMake sure that the octagon is in the correct position

SSO implant

Break joint

Lid with implant holder

1.4 Assembling of the placement aid Attach the insertion tool to the implant (SSO or STO) by holding the Lid, to which the implant is secured, with your other hand.

Insertion tool ITW SSO or ITWH SSOinsert firmly into the implant

SSO implant

Break joint

Lid with implant holder

Fasten screw with HAS Insertion tool ITV

SSO implant

Break joint

Lid with implant holder

*UZS = clockwise

Version A with IT 1,2,3 STO

Version B with ITW SSO / ITWH SSO

Version C with ITV

S-SYSTEM

3

1.SURGERY

1.8

Result

Close the implant with a cover screw CST STI (tool: HT 1.25)After the healing period: Remove cover screw SSO® implant

Jaw

1.6

Using the ratchet or contra-angle handpiece: screw the implant clockwise into the bone cavity.

The enossal part of the implant must be completely covered by the bone. The polished implant neck is located partially outside of the bone or at bone level. We recommend to screw min. 1 mm of the polished implant neck inside of the bone. OSS® implants are completely submerged into the bone.

Ratchet in position “IN“

Jaw

Contra-angle handpiece with insertion tool ITW(H) SSO Jaw

Contra-angle hand-piece Adapter

Jaw

1.7

Release of the insertion tool or the contra-angle handpiece from the implant:

Using HT 1.25 loosen the screw in the insertion tool while loading the insertion tool IT STO in insertion direction with the ratchet.

Pull insertion tool off the implant and separate from the contra-angle hand piece

Attach IT ITV Ratchet adapter with RAT2 ratchet to the ITV insertion tool. Using the HAS flat wrench firmly hold the lower hexagon of the ITV. Remove the ITV from the implant using the ratchet (“Out” position).

HT 1.25

Ratchet in position “OUT“

Jaw Jaw Jaw

IT ITV and ratchet in “OUT” position for loosening

1.5

After you have attached the insertion tool, firmly hold the lid in your hand and break the implant off the lid. Insert the implant manually until it fits tightly into the jaw.

Break joint

ts tightly into the jaw.

Break joint Break joint

ITV : use for securing

CST STI

ITW(H) SSO

4

2. IMPRESSION

2.1

Impression taking with perforated, individual impression tray The long pick up screw must be clearly visible, when the impres-sion try is inserted over the impression post. The impression material must be removed in the area of the screw access prior to hardening.

Hex-instrument HT 1.25

Fasten impression post OLT STO

SSO® / STO® implant

Jaw

2.5

After the impression is taken, the implant is closed with a healing screw (HS) and the impression is sent to the laboratory.

HT 1.25

Close healing screw clockwise

SSO® / STO® implant

Jaw

2.4

View of the impression post OLT STO in the impression (Pick up method)

Position of the impression post OLT STO

Impression material

2.2

Impression taking with an A silicone® such as Safeprint® or Duraprint®. The use of open or closed impression trys is possible.

Impression tray

Impressiaon material

Impression post OLT STO

SSO® / STO® implant

Jaw

2.3

Remove OLT STO from the implant. OLT STO remains in the impression.

Loosen screws with HT 1.25

Window in impression try

OLT STO

SSO® / STO® implant

Jaw

S-SYSTEM

5

3. PROCESSING IN THE LABORATORY

3.1

Screw IA STO or M-analogue IAO M against it. (You can find application examples and information for M-analogues in our products catalog).

Tighten the screw using HT 1.25

OLT STO

IA STO or IAO M

3.3

The laboratory analogue is now in the proper position and orienta-tion in the putty. Gipsum

3.2

The impression is casted . Then the transfer OLT STO is unscrewed from the laboratory analogue IA STO.

In IAO M block the access to the lock screw out. Laboratory analogue

Pour gipsum in

3.4

Positioning the solid secondary part TLA STO (straight) or angulated TLA 15 STO, TLA 20 STO or TLA 25 STO, at which the optimal position and adequate angulation must be determined.

Note The octagon must be completely entered into the analogue. Ensure correct position of the octagon.

TT 1.25

TLA 15 STO

or

3.5

Ensure proper position of the abutment when transferring into the mouth.Tightening torque of the screw during fastening on the implant with: 25 Ncm.

TLA 15 STO

IA STO IAO M

or

Bone

IA STO IAO M

or

Note in contraindications:Pure titanium implants are generally less flexible and resilient than implants made of a titanium alloy. We recommend against using pure titanium implants with a diameter of 3.3mm for single tooth replacements nor for bridges in the premolar or molar area of both jaws. The same applies to STO 3.7. STO 3.7 and SS0 3.3 are in general at best treatment Alternativees in stress reduced areas and should be used as supporting implants.

STO 3.7 mm and SSO 3.3 mm implants may not be used in the molar region and not as single tooth implants with off axis load. STO 3.7 mm and SSO 3.3 mm implants are used as supporting implants, for example to increase the number of abutments in immediate load situation. Never use STO 3.7 mm and SSO 3.3 mm implants when cantilevers are involved.

Note the following contra-indications:STO 3.7, SSO 3.3: Do not use in areas where off-axis load is present. Do not use in areas with strong chewing forces. Do not use in flexion areas of the jaws.

Gipsum

6

d

a

b

c

IMPLANTS

Octagonal system

ALLFIT® SSO® - IMPLANTS

a) esthetic head: 1.8 mmb) enossal length: from 7 mmc) enossal Ø: 3.3/4.1/4.8 mmd) head-Ø: 4.8 mm

Perfectly manufactured implant.Double blasted on the enossal section. SSO: Strengthened titanium alloy (Ti6AI4V ELI) according to ASTM F 136/ISO 5832- 3or SSO RT: Pure titanium medical grade 4 according to ASTM F67 / ISO 5832-2. Successful implant with octagon socket and 8°inner cone. For rotationally-secured supra-structures.

Internal octagon

d

c

b

a

SSO® - implant cylindrical Connection type: Inner cone + internal octagon / head-Ø: 4.8 mm

Description enossal Ø Length REF Price cat.

SSO 3.3 9 3.3 mm 9 mm 420000 F

SSO 3.3 11 3.3 mm 11 mm 420001 F

SSO 3.3 13 3.3 mm 13 mm 420002 F

SSO 3.3 15 3.3 mm 15 mm 420003 F

Incl. surgical screw

SSO® RT - implant cylindrical Connection type: Inner cone + internal octagon / head-Ø: 4.8 mm

Description enossal Ø Length REF Price cat.

SSO RT 4.1 7 4.1 mm 7 mm 420004 F

SSO RT 4.1 9 4.1 mm 9 mm 420005 F

SSO RT 4.1 11 4.1 mm 11 mm 420006 F

SSO RT 4.1 13 4.1 mm 13 mm 420007 F

SSO RT 4.1 15 4.1 mm 15 mm 420008 F

SSO RT 4.1 17 4.1 mm 17 mm 420009 F

SSO RT 4.8 7 4.8 mm 7 mm 420010 F

SSO RT 4.8 9 4.8 mm 9 mm 420011 F

SSO RT 4.8 11 4.8 mm 11 mm 420012 F

SSO RT 4.8 13 4.8 mm 13 mm 420013 F

Incl. surgical screw

Octagonal system

ALLFIT® STO® - IMPLANTS

Internal octagon

STO® - implant conical Connection type: Inner cone + internal octagon / head-Ø: 4.8mm

Description enossal Ø Length REF Price cat.

STO 3.7 9 3.7 mm 9 mm 420020 F

STO 3.7 11 3.7 mm 11 mm 420021 F

STO 3.7 13 3.7 mm 13 mm 420022 F

STO 3.7 15 3.7 mm 15 mm 420023 F

STO 4.1 7 4.1 mm 7 mm 420024 F

STO 4.1 9 4.1 mm 9 mm 420025 F

STO 4.1 11 4.1 mm 11 mm 420026 F

STO 4.1 13 4.1 mm 13 mm 420027 F

STO 4.1 15 4.1 mm 15 mm 420028 F

STO 4.8 7 4.8 mm 7 mm 420029 F

STO 4.8 9 4.8 mm 9 mm 420030 F

STO 4.8 11 4.8 mm 11 mm 420031 F

STO 4.8 13 4.8 mm 13 mm 420032 F

STO 4.8 15 4.8 mm 15 mm 420033 F

a) esthetic head: 1.8 mmb) enossal length: ab 7.0 mmc) enossal Ø: 3.7 / 4.1 / 4.8 mmd) head-Ø: 4.8 mm

Perfectly manufactured precision implant - double radiated on the enossal section. Strengthened titanium alloy (Ti6AI4V ELI) according to ASTM F 136/ISO 5832-3 or Ti6 AI7 Nb Implant with octagon socket. Compression screw thread for stable enossal anchorage e.g.

S-SYSTEM

7

FOR SSO®, STO®

SURGICAL SCREW

GINGIVAFORMER

TEMPBASE® ABUTMENTS FOR TEMPORARY TREATMENT

SURGICAL ACCESSORIES

Description Code REF Price cat.

Cover screw CST STI 420300 A

Description Code REF Price cat.

Gingivaformer

for 2 mm mucous membrane height, cylindrical HS2 STI 420301 B

for 4 mm mucous membrane height, cylindrical HS4 STI 420302 B

for 5 mm mucous membrane height, anatomical HS5 STI 420303 B

Description Code REF Price cat.

TempBase® Abutment

TempBase® for SSO®, STO®, 2 mm mu-cous membrane height, incl. screw TPB 2 SSO 420304 D

TempBase® for SSO®, STO®, 4 mm mu-cous membrane height, incl. screw TPB 4 SSO 420305 D

Temporary prosthesis base for SSO implants, with anti-rotation protection, PEEK, individually abradable, including screw. For cemented crowns and bridges. Screw in with HT 1.25.

8

Bone

Bone

DRILL SEQUENCES

SSO® Ø 3,3 mm

SSO® Ø 4,1 mm

SSO® Ø 4,8 mm

STO® Ø 3,7 mm

STO® Ø 4,1 mm

STO® Ø 4,8 mm

RECOMMENDED DRILL SEQUENCES FOR SSO® IMPLANTS

RECOMMENDED DRILL SEQUENCES FOR STO® IMPLANTS

Marking bur DS 2 DSL 2

DS 2.2 DSL 2.2

DS 2.8 DSL 2.8

DRA 3.3 CS Countersink head

Marking bur DS 2 DSL 2

DS 2.2 DSL 2.2

DS 2.8 DSL 2.8

DS 3.5 (DSL 3.5)

DRA 4.1 CS Countersink head

Marking bur DS 2 DSL 2

DS 2.2 DSL 2.2

DS 2.8 DSL 2.8

DS 3.5 DSL 3.5

DS 4.0 DSL 4.0

(DSL 4.2) DRA 4.1 CS Countersink head

Marking bur DS 2 DSL 2

DRK 1 (for 9, 11, 13 mml)

DRK 2 (for 15, 17, 19 mml)

Marking bur DS 2 DSL 2

DRK 3 (for 7, 9, 11, 13 mml)

DRK 2 (for 15, 17, 19 mml)

DRK 6 (for 21, 23, 25 mml)

Marking bur DS 2 DSL 2

DRK 4 (for 7, 9, 11 mml)

DRK 5 (for 13, 15 mml)

S-SYSTEM

9

ABUTMENTS FOR CEMENTED PROSTHETICS

Description Code REF Price cat.

Straight, Anti-rotation, screw-in abutment withpre-assembled screw for cemented crowns or bridges Fasten with TT 1.25.

TLA STO 420306 D

Description Code REF Price cat.

15°angled, anti-rotation, screw-in abutment with pre-assembled screw for cemented crowns or bridges. Fasten with TT 1.25. The angulation of 15° runs axial over the edge of the octagon.

TLA 15 STO 420307 E

The angulation of 15°runs axial over the area of the octagon (color: metallic yellow). TLA 15F STO 420308 E

Description Code REF Price cat.

20°angled, anti-rotation, screw-in abutment with pre-assembled screw for cemented crowns or bridges. Fasten with TT 1.25. The angulation of 20°runs axial over the edge of the octagon.

TLA 20 STO 420309 E

Description Code REF Price cat.

25°angled, anti-rotation, screw-in abutment with pre-assembled screw for cemented crowns or bridges. Fasten with TT 1.25. The angulation of 25°runs axial over the edge of the octagon.

TLA 25 STO 420311 E

The angulation of 25°runs axial over the area of the octagon (color: metallic yellow). TLA 25F STO 420312 E

Description Code REF Price cat.Screw SF OLT, short, 14.5 mml for OLT STO 5 pieces/pack SF OLT 420909 B

FOR SSO®,STO®

With octagon (anti-rotation)

There are 16 possibilities to position the abutment.

Angulation over the area (metallic yellow abutment)

Angulation over the edge (metallic yellow abutment)

Impression and laboratory accessories

Code OLT STO IA STO IAO M PS PS 15 PS 20 PS 25

REF 420350 420348 420352 420354 420355 420356 420357

Impression post with pick-up- screw

Lab analogue (inner octa)

Lab analogue with 5 sleeves, straight (inner octagon)

Synthetic base cauterizableTLA STO, TLA STI 7.4 mm high

TLA 15 STO

4 mm high

TLA 20 STO

4 mm high

TLA 25 STO

4 mm high

Price cat. B B C A A A A

or

Alternative:

10

ABUTMENTS FOR CEMENTED PROSTHETICS

FOR SSO®, STO®

With conus (without anti-rotation)

Description Code REF Price cat.

One-piece abutment with solid thread, for cemented crowns or bridges, height above implant 6.5 mm, 8°incline. Can be shortened or trimmed. Flat on one side. Direct impression or transfer to lab analogue IA STI. Fasten with HT 1.25

TCA STI 420313 D

Impression and laboratory accessories

Code TS STI TS 15 STI /TS 25 STI TSX STI IA STI PS PS 15

REF 420345 420402 / 420401 420411 420347 420354 420355

Impression post 9 mm high

Transfer post angled 15°/ 25°

Impression post 13 mm high

Lab analogue, with cone

Plastic-Basis ausbrennbar

TLA STO TLA 15 STO 7.4 mm high 4 mm high

Fasten with HT 1.25

Price cat. B B B B A A

Single-piece abutment with a fixed thread; flat on one side for increased rotation stability

6°angle, tighten with IT TCA tool

Hight above implant Color Code REF Price cat.

4 mm yellow SA4 STI 420318 C

5.5 mm grau SA5 STI 420319 C

7 mm blue SA7 STI 420320 C

Impression and laboratory accessories

Code TZ SA 4, 5, 7 AA SA 4, 5, 7 PA SA PA SR

REF 420404, 420405, 420406 420324, 420325, 420326 420327 420328

Color-coded transfer post for SA-abutments

Use with REF 460230

Abutment analogue, used with TZ SA 4, 5, 7

Castable abutment, 10 mm high, reducible anti rotation

Castable abutment, 10 mm high, reducible internally round

Price cat. A B A A

S-SYSTEM

11

ABUTMENTS AND TOOLS FOR CEMENTED PROSTHETICS

Description REF Price cat.

Patented gingiva retractor for SSO®, STO®, 4 pieces / pack (use instead of retraction cord). Material: PE (Polypropylen) 460230 A

With cone (without anti-rotation)

TOOL FOR IMPRESSION

FOR SSO®, STO®

TEMPBASE® PROTECTION CAP

Description Code REF Price cat.

Plastic, white, cementable and augmentable

for SA4, TCA 4 5 mm high CSTK 4 420334 B

for SA5, TCA 5 6.5 mm high CSTK 5 420335 Bfor SA7, TCA 7 8 mm high CSTK 7 420336 B

SECONDARY PART SETS

TLA Set

Code TLA STO Gingiva retractor TZ TLA AA TLA PS

REF 420306 460230 420390 420354

Price categorie for SET F

SA Set

Code SA R STI Gingiva retractor TZ SA 7 AA SA PA SA

REF 420330 460230 420406 420321 420327

Price categorie for SET D

Single-piece abutment with a fixed thread and vertical cement escape grooves

6°angle, tighten with IT TCA tool

Impression and laboratory accessories

Code AA TCA 4 AA TCA 5 AA TCA 7 MB TCA PA TCAREF 420407 420408 420409 420410 420332

Abutment analogue in combination with TCA 4

Abutment analogue in combination with TCA 5

Abutment analogue in combination with TCA 7

CoCrMo metal base 10 mm high, reducible, for direct ceramic veneering, suitable for soldering and laser-welding

Castable abutment 10 mm high, reducible, white

Price cat. B B B C A

Hight above implant Color Code REF Price cat.

4 mm yellow TCA4 STI 420415 C

5.5 mm grau TCA5 STI 420416 C7 mm blue TCA7 STI 420417 C

12

ABUTMENTS FOR SCREW-IN PROSTHETICS

Description Code REF Price cat.Abutment with double-threaded screw and octagon for crowns, bridges and bases. Anti-rotation anchorage of the restoration. Height above implant 1.5 mm. Fasten with HT 1.25

OSA STO 420337 E

FOR SSO®, STO®

With octagon (anti-rotation)g

OCTAGON

external

internal

Description Code REF Price cat.Single-piece and anti-rotational octagonal screw-in abutment; for crowns, bridges and bar connectors. Height above implant: 1.5 mm Fasten with HT 1.77 / HTX 1.77

OSA STI 420338 D

With cone (without anti-rotation)

Description Code REF Price cat.

Protective cap, plastic white, screwable, reducible, 5 pieces/pack, for OSA STI or OSA STO. Height above implant: 3.0 mm

CSTO 420344 B

CONE

External octagon

3 mm

Impression and laboratory accessories

Code TS STI IA STIREF 420345 420347

Impression post, fasten with HT 1.259 mm high

Lab analogue with cone

Impression taking directly on the implant, no transfer of the edge. For bridge treatment only

Price cat. B B

Impression and laboratory accessories

Code TST STI SF 365 SF 415 OA STI PAOA PAOR SF 350 SF 365

REF 420339 420938 420928 420340 420342 420343 420930 420938

Transfer post

Screws for anchorage of TST STI

Lab analogue for OSA

white castable abutments 10 mm high5 pieces/pack

Screws

Price cat. D B B B B B

or or

Impression and laboratory accessories

Code TST STI SF 365 SF 415 OA STI PAOA PAOR SF 350 SF 365

REF 420339 420938 420928 420340 420342 420343 420930 420938

Transfer post

Screws for anchorage of TST STI

Lab analogue for OSA

white castable abutments 10 mm high5 pieces/pack

Screws

Price cat. D B B B B B

or or

anti rotation

anti rotation

internally round

internally round

S-SYSTEM

13

ABUTMENTS FOR REMOVABLE PROSTHETICS

Impression and laboratory accessories

CodeREF

TST418147

SFL420428

BTT418100

PSTR (grau) 418124

PSTA 418123

SF418151

Transfer post Long screw TCT-analogue Castable abutment, 12 mm high. Internally round. 5 pieces/pack

Castable abutment, 12 mm high. Internally edged. 5 pieces/pack

Screw for fastening

Preiskat. B A B B B B

With this procedure the position of the TCT-Hex is being transfered

Abutment for removable prosthetics, fasten with HT 1.77.

LOCALICER®

ACCESSORIES FOR LOCALICER®

Description Code REF Price cat.

Instrument for mounting and de-mounting Localicer caps LOC Tool 462335 C

Analogue + impression set AA LOC 462337 C

Set with 4 caps + housing NCS 462338 D

Set of two, yellow, with increased friction strength R-Cap 462336 B

Pull-off force: *black 125 g, red 350 g, blue 500 g, pink 1000 g *Black cap is for temporary solutions up to one month.

ABUTMENTS

Description Hight REF Price cat.

LOC S 2.5 2.5 mm 420386 D

LOC S 4 4.0 mm 420385 D

We recommend to use minimally 6 implants per jaw and to use a single denture as splinting.

Description Code REF Price cat.

TCT SSO 0.5

for screw-in prostheticsfasten with HT 1.77

TCT SSO 0.5 420420 D

TCT SSO 1.5 TCT SSO 1.5 420421 D

TCT SSO 2.5 TCT SSO 2.5 420422 D

TCT SSO 4 TCT SSO 4 420423 D

TITAN BASE (FOR CAD CAM)

Description Code REF Price cat.

Titan base for octa system / SSO, Ti6Al4V, ant-irotation Hight 3.9 mm above implant shoulder. Incl. screw SF N62 MB4 SSO 463120 B

Titan base for octa system / SSO, Ti6Al4V, ant-irotation Hight 8 mm above implant shoulder. Incl. screw SF N62 MB8 SSO 463122 C

14

ABUTMENTS FOR REMOVABLE PROSTHETICS

Description Code REF Price cat.

Ball abutment, fits nylon cap NC / NC1 / NC2 . Head-Ø 2.5 mm, hight above implant 3 mm. Fasten with HT 1.25

TB STI 420361 D

Description Code REF Price cat.

Ball abutment, fits nylon cap NC / NC1 / NC2 Head-Ø 2.5 mm, hight above implant 3 mm. Fasten with Tool E (alternatively with HAS2).

TB3 STI 420363 D

FOR SSO®, STO®

RETENTIVE BALL ATTACHMENTS

Description Code REF Price cat.Ball abutment, fits gold cap GC.Head-Ø 2.3 mm, hight above omplant 3 mm. Fasten with Tool E (alternatively with HAS2).

TB2 STI 420362 D

Impression and laboratory accessories

Code IA STI TS STI IAB 2 GCREF 420347 420345 420367 420366

Lab analogue with cone

Transfer post. Fasten with HT 1.25

Ball analogue Gold cap

Price cat. B B B Price upon request

Impression and laboratory accessories

Extraction forcesCode IA STI TS STI IAB NC NC 1 NC 2 R-NC R-NC 1 R-NC 2 H

REF 420347 420345 420566 465028 465029 465030 465034 465033 465032 465031Lab analoguewith cone

Transfer post. Fasten with HT 1.25

Ball analogue ca. 1200 g white

ca. 800 g pink

ca. 500 g yellow

green, strong

pink, medium

orange, soft

Sleeve for all NC

Price cat. B B C A A A A A A B

Impression and laboratory accessories

Extraction forcesCode IA STI TS STI IAB NC NC 1 NC 2 R-NC R-NC 1 R-NC 2 H

REF 420347 420345 420566 465028 465029 465030 465034 465033 465032 465031Lab analoguewith cone

Transfer post. Fasten with HT 1.25

Ball analogue ca. 1200 g white

ca. 800 g pink

ca. 500 g yellow

green, strong

pink, medium

orange, soft

Sleeve for all NC

Price cat. B B C A A A A A A B

Nylon cap NC

Nylon cap NC

Nylon cap R-NC with increased friction strength

Nylon cap R-NC with increased friction strength

S-SYSTEM

15

IMPLANTS AND SURGICAL ACCESSORIES

These implants with a head diameter of 6.5 mm can be used, if adequate bone is available, for posterior single crowns with the width of one premolar in dentate and partially/fully edentulous ridges as for superstructures retained by bars and ball attachments.

Description Code REF Price cat.

Surgical screw (Cover screw) for STW implant, non-sterile, 1 piece CST STW 420370 B

Description Code REF Price cat.

Gingivaformer

for 2 mm mucous membrane height, cylindrical HS2 STW 420371 B

for 4 mm mucous membrane height, cylindrical HS4 STW 420372 B

ALLFIT® STW® - IMPLANTS

Octagonal system

FOR STW®

SURGICAL SCREW

GINGIVAFORMER

a) esthetic head: 1.8 mmb) enossal length: ab 7.0 mmc) enossal Ø: 4.8 mmd) head-Ø: 6.5 mm

Perfectly manufactured precision implant with double-sandblasted endosseous section. Highly fracture-resistant titanium alloy (Ti6AI4V ELI) according to ASTM F 136/ISO 5832-3. Implant with internal octagon and 8°internal cone. For anti-rotational superstructures.

Internal octagon

STW® - implant cylindrical

Connection type: Inner cone and Inner octa / head - Ø: 6.5 mm

Description enossal Ø Length REF Price cat.

STW 4.8 7 4.8 mm 7 mm 420015 F

STW 4.8 9 4.8 mm 9 mm 420016 F

STW 4.8 11 4.8 mm 11 mm 420017 F

STW 4.8 13 4.8 mm 13 mm 420018 F

STW 4.8 15 4.8 mm 15 mm 420019 F

a

d

b

c

TITAN BASE (FOR CAD CAM)

Beschreibung Code REF Preiskat.

Titan base for octa system / STW, Ti6Al4V, ant-irotation Hight 3.9 mm above implant shoulder. Incl. screw SF N62 MB4 STW 463121 B

16

ABUTMENTS FOR CEMENTED PROSTHETICS

Description Code REF Price cat.

15°angled anti-rotation screw-in abutment with pre-mounted screw, for cemented crowns or bridges. Fasten with TT 1.25. The 15°line angle runs axially across the edge of the octagon

TLA 15 STW 420374 E

Description Code REF Price cat.Screw SF OLT, short, 14.5 mml for OLT STW 5 pieces/pack SF OLT 420909 B

Replacement screw for TLA abutments of all S-implants SF TLA 420910 A

Flattened on one side for increased rotational stability Code REF Price cat.

Hight above implant 4 mm SA4 STW 420375 C

Hight above implant 6 mm SA6 STW 420376 C

Solid - Abutment

6° angle, fasten with HT 1.77

LOCK - ABUTMENT FOR STW®

With Octagon (anti-rotation)

Alternative:

Description Code REF Price cat.

Straight anti-rotation screw-in abutment with pre-mounted screw, for cemented crowns or bridges. Fasten with TT 1.25

TLA STW 420373 D

Impression and laboratory accessories

Code OLT STW IA STW IAO STW M PA STW PA 15 STW

REF 420351 420349 420353 420359 420360

Impression post with pick-up screw. Fasten with HT 1.25

Lab analogue Lab analogue with 5 sleeves

Plastic base, castable

for TLA STW for TLA 15 STW

Price cat. B B C A A

Impression and laboratory accessories

Code TZ SA6 STW AA SA6 STW PA STW

REF 420377 420378 420359

Transfer post for SA6 abutment

Abutment analogue brown, anodized, for TZ SA6 STW

Castable abutmenttet

Price cat. A A A

S-SYSTEM

17

ABUTMENTS FOR REMOVABLE PROSTHETICS

RETENTIVE BALL ATTACHMENTS

Description REF Price cat.

Ball abutment TB STW, fits nylon cap NC / NC 1 / NC 2 Head - Ø 2,5 mm, hight above implant 3 mm. Fasten with HT 1.25 420365 D

Description REF Price cat.

Ball abutment TB3 STW, fits nylon cap NC. Head - Ø 2,5 mm, hight above implant 3 mm. Fasten with HAS2 and Tool E (REF 462377) 420364 D

Impression and laboratory accessories

Extraction forces

Nylon cap NC Nylon cap R-NC with increased friction strength

Code 420346 420349 420566 420353 NC NC 1 NC 2 R-NC R-NC 1 R-NC 2 HREF TS STW IA STW IAB IAO STW M 465028 465029 465030 465034 465033 465032 465031

Transfer post. Fasten with HT 1.25

Lab analogue

Ball analogue

Lab analogue with 5 sleeves

ca. 1200 g white

ca. 800 g pink

ca. 500 g yellow

green, strong

pink, medium

orange, soft

Sleeve for all NC

Price cat. B B C C A A A A A A B

Implants enossal Ø recommended drill sequences

SSO® 3.3 mmDS 2 425001 DS 2.2 DS 2.8

425005DRA 3.3 CS 425046

DSL 2 425002

DSL 2.2 425004

DSL 2.8 425006

DRA 3.3 CS 425046

SSO® 4.1 mm

DS 2 425001

DS 2.2 DS 2.8 425005

DS 3.5 425009

DRA 4.1 CS 425047

DSL 2 425002

DSL 2.2 425004

DSL 2.8 425006

(DSL 3.5) (425010)

DRA 4.1 CS 425047

SSO® 4.8 mmDS 2 425001 DS 2.2 DS 2.8

425005DS 3.5 425009

DS 4.0 425011

(DS 4.2) (425013)

DRA 4.1 CS 425047

DSL 2 425002

DSL 2.2 425004

DSL 2.8 425006

DSL 3.5 425010

DSL 4.0 425012

(DSL 4.2) (425014)

DRA 4.1 CS 425047

Implants enossal Ø Length recommended drill sequences

STW® 4.1 mm all length

DS 2 425001 DS 2.2 DS 2.8

425005DS 3.5 425009

DS 4.0 425011

(DSL 4.2) (425014)

DRA 4.8 CSW 425048

DSL 2 425002

DSL 2.2 425004

DSL 2.8 425006

DSL 3.5 425010

DSL 4.0 425012

(DSL 4.2) (425014)

DRA 4.8 CSW 425048

CYLINDRICAL IMPLANTS

CONICALE IMPLANTS

Implants enossal Ø Length in mm recommended drill sequences

STO® 3.7 mm9 / 11 / 13

DS 2 / 425001 DS 2.2 DRK 1 / 425020 DRA 3.3 CS / 425046

DSL 2 / 425002 DSL 2.2 / 425004 DRK 1 / 425020 DRA 3.3 CS / 425046

15DS 2 / 425001 DS 2.2 DRK 2 / 425021 DRA 3.3 CS / 425046

DSL 2 / 425002 DSL 2.2 / 425004 DRK 2 / 425021 DRA 3.3 CS / 425046

STO®4.1 mm

7 / 9 / 11 / 13DS 2 / 425001 DS 2.2 DRK 3 / 425022 DRA 4.1 CS / 425047

DSL 2 / 425002 DSL 2.2 / 425004 DRK 3 / 425022 DRA 4.1 CS / 425047

15 / 17 / 19DS 2 / 425001 DS 2.2 DRK 2 / 425021 DRA 4.1 CS / 425047

DSL 2 / 425002 DSL 2.2 / 425004 DRK 2 / 425021 DRA 4.1 CS / 425047

STO® 21 / 23 / 25DS 2 / 425001 DS 2.2 DRK 6 / 425025 DRA 4.1 CS / 425047

DSL 2 / 425002 DSL 2.2 / 425004 DRK 6 / 425025 DRA 4.1 CS / 425047

STO® 4.8 mm7 / 9 / 11

DS 2 / 425001 DS 2.2 DRK 4 / 425023 DRA 4.1 CS / 425047

DSL 2 / 425002 DSL 2.2 / 425004 DRK 4 / 425023 DRA 4.1 CS / 425047

13 / 15DS 2 / 425001 DS 2.2 DRK 5 / 425024 DRA 4.1 CS / 425047

DSL 2 / 425002 DSL 2.2 / 425004 DRK 5 / 425024 DRA 4.1 CS / 425047

18

SURGICAL INSTRUMENTS AND TOOLS

Description Type Working length REF Price cat.

BCD 1 non-sterile 15 mm 900240 C

BCDX 1 non-sterile 15 mm 900243 C

DRILL SPEED (RPM)

PATHFINDER-DRILLS

FOR SSO®, STW®

To determine the drilling speed, you must know the motor speed, which is then divided by the gear reduction factor of the handpiece.

Example: Speed : Gear reduction =Drilling speed 20.000 400 : 1 50 Upm

Motor speed (rpm)

Gear reduction 20.000 27.000 33.000 35.000 40.000

1:1 20.000 27.000 33.000 35.000 40.000

4:1 5.000 6.750 8.250 8.750 10.000

16:1 1.250 1.688 2.063 2.188 2.222

18:1 1.111 1.500 1.833 1.944 571

70:1 286 386 471 500 571

256:1 78 105 129 137 156

400:1 50 68 83 88 100

1000:1 20 27 33 35 40

I 15 mm I

Drill (Externally irrigated)

Surgical steel, laser marked. Maximum speed with external cooling 700 RPM. For lifetime extension and consistent cutting performance and easy removal of contaminated areas. Drills can be used without cooling at 50 RPM on green or blue contra-angle handpieces.

These drills are made of premium material using state-of-the-art precision milling machines. The benefit for you and your patients: Extremely good and durable cutting performance due to sophisticated geometry of the blades. Almost no heat development and excellent as a result of this geometry.

Twistdrill / Length: 36.5 mm Working length: 18 mm

Description Type Drill - Ø REF Price cat.

DS 2 non-sterile 2 mm 425001 D

DS 2.2 non-sterile 2.2 mm 425003 D

DS 2.8 non-sterile 2.8 mm 425005 D

DS 3.2 non-sterile 3.2 mm 425007 D

DS 3.5 non-sterile 3.5 mm 425009 D

DS 4.0 non-sterile 4.0 mm 425011 D

DS 4.2 non-sterile 4.2 mm 425013 D

Twistdrill / Length: 45.5 mm Working length: 27 mm

Description Type Drill - Ø REF Price cat.

DSL 2 non-sterile 2 mm 425002 D

DSL 2.2 non-sterile 2.2 mm 425004 D

DSL 2.8 non-sterile 2.8 mm 425006 D

DSL 3.2 non-sterile 3.2 mm 425008 D

DSL 3.5 non-sterile 3.5 mm 425010 D

DSL 4.0 non-sterile 4.0 mm 425012 D

DSL 4.2 non-sterile 4.2 mm 425014 D

18 mm

7 11 159 13

27 mm

7 11 15 19 239 13 17 21

I 15 mm I

27 mm

The drilling speed is indicated directly on the surgical unit.

S-SYSTEM

19

SURGICAL INSTRUMENTS AND TOOLS

Description System Color Length Implant-Ø REF Price cat.

DRK 1 STO® yellow 9, 11, 13 mm 3.3 / 3.7 mm 425020 D

DRK 2 STO® black 15, 17, 19 mm 3.3 / 3.7 / 4.1 mm 425021 D

DRK 3 STO® red 7, 9, 11, 13 mm 4.1 mm 425022 D

DRK 4 STO® blue 7, 9, 11 mm 4.8 mm 425023 D

DRK 5 STO® green 13, 15 mm 4.8 mm 425024 DDRK 6 STO® metallic 21, 23, 25 mm 3.7 / 4.1 mm 425025 D

Description System Type Implant-Ø REF Price cat.

DRA 3.3 CS SSO®, STO® non-sterile 3.3 mm 425046 D

DRA 4.1 CS SSO®, STO® non-sterile 4.1 mm 425047 DDRA 4.8 CSW STW® non-sterile 4.8 mm 425048 D

Description Finish Length Implant - Ø REF Price cat.

TAP1 3.3 9 - 17 mm 34 mm 3.3 mm 425055 G

TAP1 4.1 9 - 17 mm 34 mm 4.1 mm 425056 G

TAP1 4.8 9 - 17 mm 30 mm 4.8 mm 425057 G

TAP2 3.3 9 - 15 mm 23 mm 3.3 mm 425058 G

TAP2 4.1 9 - 15 mm 23 mm 4.1 mm 425059 G

HEATLESS®-DRILL «DRK» FOR IMPLANTS WITH CONICAL CORE

COUNTERSINK HEADS

TAP (FROM CRONIDUR) FOR SSO®, STW®

DRILLSTOPP - TRAY

* for S-System only

Form drills made from surgical steel (SS), length and color coded, resterilizable. Maximum motor speed with external cooling 700 rpm. Note the information in the instructions for use for hygiene and preparation requirements with medical products intended for multiple use in implantology. Pilot drilling using DS2 (P.33), DSL2 (P.33) or Pathfinder (P.34)

It has been scientifically proven that Dr. Ihde Dental heatless drills generate 55% less heat than traditional bone drills from other manufacturers. This makes it possible to use higher rotational speeds: between 3,000 and 5,000 rpm are recommended with good external cooling and intermittent drill technique.

tem onlyem only

Not suitable for dry heat sterilizers

Description REF

Drillstopp A 500881Drillstopp C 500883Drillstopp F 500886Drillstopp H 500888Drillstopp K 500891Drillstopp L 500892Drillstopp M* 500893Drillstopp N* 500894Drillstopp O* 500895

Description REF Price €Drill DS 2.8 425005Drill DS 3.2 425007Drill DS 3.5 425009Drill DS 4.0 425011Drill DS 4.2 425013Drill DSL 2.8 425006Drill DSL 3.2 425008Drill DSL 3.5 425010Drill DSL 4.0 425012Drill DSL 4.2 425014

Tray with content 60034-K 779,00

20

Please read our detailed instructions for cleaning and re-sterilisation of surgical instruments on www.implant.com/de/instructions (Gebrauchsanweisungen - Implants - Hygienemerkblatt)

Description Type Length Index REF Price cat.

IT1 STO long 20 mm HT 1.25 425060 K

IT2 STO short 12 mm HT 1.25 425061 K

IT3 STO medium 16 mm HT 1.25 425063 K

ITW SSO contra-angle 23 mm 463110 K

ITWH SSO contra-angle / hex* 23 mm 463111 K

* This instrument is with an additional 6-edge for force transmission to the contra-angle handpiece equipped. Fits to W&H contra-angle handpiece only.

INSERTION TOOLS FOR SSO®, STO®, STW®

SURGICAL INSTRUMENTS AND TOOLS

screwablescrewablebl

Description Type Length for REF Price cat.

ITV short 11 mm SSO, STO 500850 C

IT ITV ratchet adapter Adapter for ITV 500854 C

Adapter short / contra-angle 22 mm ITV 500850 500851 C

Adapter long / contra-angle 32 mm ITV 500850 500852 C

Adapter medium / contra-angle 27 mm ITV 500850 500853 C

HAS square 85 mm ITV 463108 H

INSERTION TOOLS FOR SSO®, STO®

Description Type Length Index for abutments REF Price cat.

IT TCA long 20 mm direct SA STI, TCA STI 425065 D

Tool E long 20 mm direct TB2 STI, TB3 STI,TB3 STW 462377 D

INSERTION TOOLS FOR ABUTMENTS

Description Type Length Ø Code REF Price cat.

Hex-instrument long 21 mm 1.25 mm HT 1.25 425100 C

Hex-instrument for contra-angle long 26.1 mm 1.25 mm HT 1.25 M 425112 B

Hex-instrument short 14 mm 1.25 mm HTS 1.25 425101 C

Hex-instrument extralong 45 mm 1.25 mm HTX 1.25 425102 C

Hex-instrument long 19 mm 1.77 mm HT 1.77 425103 C

Hex-instrument for contra-angle long 28.6 mm 1.77 mm HT 1.77 M 425113 B

Hex-instrument extralong 45 mm 1.77 mm HTX 1.77 425104 C

Torx-instrument for SSO and STO abutments 21 mm 1.25 mm TT 1.25 425105 C

Surgical steel, for use with a ratchet

INSTRUMENTS FOR SCREWS

S-SYSTEM

21

STARTER - TRAY

Please read our detailed instructions for cleaning and re-sterilisation of surgical instruments on www.implant.com/de/instructions (Gebrauchsanweisungen - Implants - Hygienemerkblatt)

Autoclavable up to 134 °C, not suitable for dry heat sterilizers This surgical kit contains all drills and tools for first works with the S-System. Material: autoclavable plastic

TOOLS AND TRAYS

TOOLS

Description Type Length Ø Code REF Price cat.

Guide jacketTitanium, for pilot drill,5 pieces / pack

10 mm 2.2 mm BFH 425401 A

X-ray measuring sphere

Surgical steel 5 pieces / pack 5.0 mm RM 425403 A

Drill extension Contra-angle, extends by 19 mm DX2 500704 D

Punch Contra-angle handpiece 4.9 mm PUW 1 425404 C

Punch Contra-angle handpiece 3.9 mm PUW 2 425405 C

Punch Manual 5.2 mm PU 425406 C

Standardized probe

Scale 1 mm, radio-graphical measuring for cylindrical implant types

22 mm PDG 425400 A

HAS Flat spanner for ITV 85 mm HAS 463108 H

HAS2 Flat spanner for TB3 STI 85 mm HAS2 463109 H

Ratchet For all Hex-instruments and insertion tools RAT 2 425051 K

Ratchet Angled, for all Hex-instruments and insertion tools RAT 3 425052 S

Torque wrench Torque wrench 10 - 70 Ncm for all insertion tools, Hex- and Torx instruments TW 2 425402 S

It is recommended to have the torque ratchets recalibrated by Dr. Ihde Dental once a year.

CONTENT

Description REF Price €IT3 STO 425063ITWH SSO 463111HT 1.25 425100TT 1.25 425105BCD 1 900240DS 2.2 425003DS 2.8 425005DS 3.5 425009DS 4.2 425013DRA 4.1 CS 425047PDG 425400PDG 425400TW2 425402

Starter-tray (with content)* 60042-K 788,60

* Empty tray upon request

We are certified DIN EN ISO 13485, and annex II of EEC Directive 93/42 EWG (2007). Due to technical reasons the product dimensions shown in this brochure might deviate from reality. SSOⓇ, STOⓇ and STWⓇ are registered trademarks. In case that implants would be reprocessed (cleaned, resterilized) infections could occur, because no validated procedures for reprocessing are available.

secure rotation protection by precision inner octagon

reliable insertion - ease of use

universal application for permanentand removable prosthodontics

satisfied patients in eachphase of the procedure

REF 6640EN_V006_03/2016

CONTACT

CH Dr. Ihde Dental AGDorfplatz 11 CH - 8737 Gommiswald / SG Tel. +41 (0)55 293 23 23 Fax +41 (0)55 293 23 00 [email protected]

D Dr. Ihde Dental GmbHErfurter Str. 19 D - 85386 Eching / Munich Tel. +49 (0)89 319 761-0 Fax +49 (0)89 319 761-33 [email protected]

www.implant.com (CH) www.ihde-dental.de (DE)

Symbols on the packs

Production No.

Sterilized by gamma radiation

Sterilization Ethylene oxide

Single-use

product

Instruction for use

Expiry date Store in a dry place

Temperature range from -5° C to 25° C

Store tightly keep closed

Do not use if packing is damaged

Do not resterilize

Manu-facturer

-5° C

25° C

Rx ONLYNon-sterile

Intended for use by dentists or surgeons only