Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution Techniques Charlene Cuellar Electrical Engineering, Contra Costa Community College Faculty Advisor: Dr. Michael Liebling Mentor: Nikhil Chacko Electrical and Computer Engineering, UCSB Funding Source: Hellman Fellowship granted to Dr. Michael Liebling

Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution Techniques

Charlene Cuellar

Electrical Engineering,

Contra Costa Community College

Faculty Advisor: Dr. Michael Liebling

Mentor: Nikhil Chacko

Electrical and Computer Engineering, UCSB

Funding Source: Hellman Fellowship granted to Dr. Michael Liebling

Page 2: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

Fluorescence is exploited inFluorescence Microscopy

2

Fluorescence Imaging

Simplified Model of Fluorescence Imaging

Page 3: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

We acquire 3D datasets by Optical Sectioning

Microscopy

Out-of-focus adjacent planes contaminate image.

Bad axial (z) resolution.

Drawbacks

3

Z

A point source thatexists in only one

focal plane along zspreads to other focal

planes:Point Spread Function (PSF)

Page 4: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

4

We can alleviate the blur by:

(a) Illuminating the whole sampleand using a physical barrier toblock rogue light rays.(Confocal Microscopy)

(b) Restricting the excitation tothe plane of interest alone.(Light-sheet Microscopy)

(c) Image processing (Deconvolution)

Page 5: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

5

Deconvolution methods assume a shift-invariant model

A shift-invariant system is completely characterized by its response to a point source:

Point source Point-spread function (PSF)

Any signal can be represented as a linear combination of many points.

Any signal response can be represented as a linear

combination of many PSFs.

Fourier Transforms simplify convolutions to multiplications

Given f(x) and h(x), find g(x): convolution problemGiven g(x) and h(x), find f(x): deconvolution problem

Page 6: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

6

1. Model the available microscope (Leica DMI 6000B)Determine the PSF characteristic to the microscope in thelab.

2. Deconvolve data from Single-View observationUse PSF to deblur 3-D data acquired by the microscope.

3. Deconvolve data from Multi-View observationAcquire data from multiple angles and perform deblurring

using a multi-channel deconvolution algorithm.

Project Steps

Page 7: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

7

fluorescent beads blurred observationLeica DMI6000b

1. PSF determination

Fluorescent beads having a diameter less than the spatial resolution of the device approximate point sources.

The blurred observation of any single bead hints to the PSF of the microscope.

Page 8: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

8

Measured PSF Averaged PSF

Born & Wolf PSF Gibson & Lanni PSF Richards and Wolf PSF

3D PSF Intensity Graphs (x-z plane)Comparison of measured PSF to theoretical models

Theoretical models generated according to parameters in experimental setup:

A single measured PSF exhibits significantamount ofnoise.

Average over multiplesimilar data

Averagingreducesthe standard deviation of the noise component, improving signal quality.

Page 9: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

9

Page 10: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

z =0 μm

10

3D PSF Intensity Graphs (x-y plane)Comparison of measured PSF to theoretical models

Measured PSF (averaged) in x-y plane at z = 0 µm

Richards-Wolf (RW) PSF Modelin x-y plane at z = 0 µm

Born-Wolf (BW) PSF Modelin x-y plane at z = 0 µm

Gibson-Lanni (GL) PSF Modelin x-y plane at z = 0 µm

Page 11: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

11

2. Deconvolve data from Single-View observation

Zebrafish Tg(f1ia:EGFP), fluorescent along the vetebral column, was used in the experimental setup for imaging.

Bright-field Microscopy Fluorescence Microscopy

Page 12: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

12

xy-plane xz-plane

yz-plane

Page 13: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

13

Deconvolution Results – xy-plane

Original Data

Algorithm: Richardson LucyPSF: Richards & Wolf

Algorithm: ThresholdedLandweberPSF: Richards & Wolf

Page 14: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

14

Deconvolution Results- xz-plane

Original Data

Algorithm: Richardson LucyPSF: Richards & Wolf

Algorithm: ThresholdedLandweberPSF: Richards & Wolf

Page 15: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

15

3. Deconvolve data from Multi-View observation

We place the specimen within a tube that is connected to a stepper motor. The stepper motor is controlled by an Arduinoprogramming board, which is interfaced to Matlab.

Page 16: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

16

Zebrafish Multi-View - xy-plane

Page 17: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

17

Zebrafish Multi-View - xz-plane

The blur along different directions makes it difficult to spatially register two data sets, making any registration algorithms based on spatial landmarks difficult to use

Page 18: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

1818

3. Deblur data from Multi-View observation (SimulationResults)

Blurred “Observation”

Single-View Deconvolution Results

Landweber Regularized Inverse Filtering

Richardson-Lucy

ThresholdedLandweber

Original Object

x

z

Page 19: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

19

Page 20: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

20

Page 21: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

21

Original Data

Single -View:ThresholdedLandweber

Multi-View:ThresholdedLandweber

Page 22: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

22

Conclusions and Future Prospective

(a) The theoretical PSF models generated according to the instrument and experimental parameters were close approximations to the actual measured data.

(b) Deconvolved results using the theoretical PSF models were observed to be superior (less noisier) than that with the measured PSF.

(c) The faithful registration of actual 3D datasets blurred along different angles still remains a problem that is unsolved, though the simulation results for the multi-angle deconvolution look promising.

Page 23: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

23

Thank You

•INSET - Internships in NanosystemsScience, Engineering and Technology

•Systems Bioimaging LabDr. Michael LieblingNikhil ChackoKevin ChanMichael Lee

Page 24: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

24

Page 25: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

25

x

z

x

z

x

z

x

z

Page 26: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

26

Zebrafish – Z-Stack

Page 27: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

27

Zebrafish Single-View Deconvolution ResultsMeasured VS Theoretical PSF

Page 28: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

28

Method Signal Estimate Comments

Inverse Filtering• Linear filtering operation

•Amplifies noise when H(ω) 0

Landweber• Iterative

• Doesn’t amplify noise

Regularized Inverse Filtering•Tries to smoothen image in addition

to inverse filtering

Wiener Filtering• Linear filtering operation

• Used in noisy cases

Richardson-Lucy

•Assumes that input is Poisson

distributed (appropriate for photon

noise in data)

• Developed from Bayes’ Theorem

Thresholded Landweber (TL)•Assumes wavelet coefficients of the

data to be estimated are sparse.

Multi-Channel Thresholded

Landweber

• Extension of TL to multi-channel

framework.

fe~

argmin || g~

H~ f~

||22

2

2~~~~

||||minarg fHgfe

fe argmin || g~

H~ f~

||22 || f ||2

2

fe argminE |F() Fe() |2

dffPfgP

fPfgPgfP

fe

)()|(

)()|()|(

maxarg

fe~

argmin || g~

H~ f ||2

2 ||W~ f~

||11

1

0

1

1~~

2

2~~~

||||||||minargM

i

iie fWfHgf

Page 29: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

29

PSF- Fourier Transform Representation

Page 30: Improving Resolution in Wide-field Fluorescence …inset-csep.cnsi.ucsb.edu/sites/inset-csep.cnsi.ucsb.edu/...Improving Resolution in Wide-field Fluorescence Microscopy Using Deconvolution

30

f

f

Deblur Multi-Angle(Channel

) Image