Upload
cristina-velasco
View
214
Download
0
Embed Size (px)
Citation preview
Original article
Influence of pump compliance (peristaltic vs. infusion) on urodynamic
measurement during cystometry in conscious rats
Cristina Velasco*, Luciano Guarneri, Amedeo Leonardi, Rodolfo Testa
Pharmaceutical R & D Division-Recordati S.p.A., Via M. Civitali I-20148, Milano, Italy
Received 12 January 2001; accepted 9 June 2001
Abstract
Cystometry, employing natural or pump-induced bladder filling, is the most widely used method for studying bladder reflexes and
micturition in conscious rats. However, discrepancies in basal values of urodynamic parameters are often reported, especially for micturition
pressure. The aim of this study was to establish whether the type of pump used (peristaltic or infusion) might yield different urodynamic
parameters. Differences between natural filling (evaluated in water-loaded animals and considered ‘‘physiological micturition’’) and pump-
evoked cystometrograms, as well as the compliance of these systems, and the effects of pharmacologically diverse drugs (prazosin,
oxybutynin, and naproxen) acting on the bladder voiding were evaluated. Micturition pressure recorded from pump-evoked
cystometrograms showed differences from natural micturition that were related to the total compliance of the system (pump + tube) and
not only to the nature of the pump used. Drug-induced changes of micturition pressure during natural micturition resembled those recorded
during bladder infusion with a peristaltic pump more than those with an infusion pump. Other basal values and drug-induced changes of
bladder capacity were the same during natural and pump-evoked micturition. The present findings indicate that cystometrographic
parameters obtained during pump-evoked micturition with a system at high compliance (peristaltic pump) are equivalent to those observed
during physiological micturition. D 2002 Elsevier Science Inc. All rights reserved.
Keywords: Cystometry in conscious rats; Micturition pressure; Oxybutynin, Prazosin; Naproxen
1. Introduction
Since its first description by Mosso and Pellacani (1892),
cystometry has been employed in humans to measure the
properties of the bladder wall and to evaluate changes in
urodynamic parameters associated with urinary inconti-
nence (Abrams et al., 1988).
Cystometry is also used routinely in vivo to evaluate
bladder volume capacity and micturition pressure (the max-
imal bladder pressure during micturition) in several animal
species e.g., guinea-pig (Peterson et al., 1989), rabbit (Sjog-
ren, 1976), dog (Nishizawa et al., 1989), pig (Blok et al.,
1985), mini-pig (Peterson et al., 1990), cat (Klevmark, 1974),
monkey (Shoukry & Ghoniem, 1992), as well as awake
(Yaksh et al., 1986) and anesthetized (Maggi et al., 1986a,
1986b, 1986c) rats.
In view of the fact that anesthetics may interfere with the
micturition reflex in rats (Yaksh et al., 1986; Matsuura &
Downie, 2000), continuous cystometry in conscious animals
is the preferable model. Urodynamic parameters during
spontaneous voiding are similar to those of pump-evoked
micturition in conscious rats (Yaksh et al., 1986).
Nevertheless, continuous cystometry in conscious rats
has revealed wide between-study discrepancies in basal
values of urodynamic parameters. In particular, values of
micturition pressure ranging from 26 to 122 cm H2O (19 to
90 mmHg) have been reported using different types of
pumps (Durant & Yaksh, 1988; Conte et al., 1988; Guarneri
et al., 1991; Yamamoto et al., 1995; Ishizuka et al., 1997;
Persson et al., 1998; Chai et al., 1999). Cystometric meth-
odology in each of these studies was almost the same except
for slight differences in flow rate and the type of pump used
to fill the bladder (infusion versus peristaltic).
Since it has been demonstrated (Yaksh et al., 1986;
Conte et al., 1988; Guarneri et al., 1991) that different
infusion rates (e.g., between 0.05 and 0.25 ml/min) do not
modify basal micturition parameters, we hypothesized that
the choice of the pump used for filling the bladder during
continuous cystometry (peristaltic versus infusion pump of
1056-8719/02/$ – see front matter D 2002 Elsevier Science Inc. All rights reserved.
PII: S1056 -8719 (01 )00151 -4
* Corresponding author. Fax: +39-2-48-70-9017.
Journal of Pharmacological and Toxicological Methods 45 (2001) 215–221
varying volume) is the main source of baseline urody-
namic variation.
To compare the two different bladder filling systems,
experiments have been performed to evaluate: a) the diffe-
rences between natural and pump-evoked cystometrograms,
b) the compliance of the systems, c) the effects of drugs
acting on the bladder voiding cycle via different mechanisms.
The anticholinergic drug oxybutynin, the a1-adrenoceptor
antagonist prazosin, and the COX inhibitor naproxen
were selected.
2. Materials and methods
2.1. Animals
Male Sprague-Dawley rats [Crl: CD� (SD) BR] of 300–350 g b.w. from Charles River Italia were used. Animals
were housed with free access to food and water and
maintained on forced light-dark cycle at 22–24�C. The
animals were handled according to internationally accepted
principles for care of laboratory animals (E.E.C. Council
Directive 86/609, O.J. No L358, 18/12/86).
2.2. Surgical procedure
Rats were anesthetised with 3 ml/kg Equitensin (com-
position for 125 ml of aqueous solution: 1.2 g of pentobar-
bital, 5.3 g of chloral hydrate, 2.7 g of MgSO4, 49.5 ml of
propylene glycol, 12.5 ml of ethanol) and placed in a
supine position. An approximately 10 mm incision was
made in the shaved and cleaned abdominal wall. The
urinary bladder was gently freed from adhering tissues and
a polyethylene tube (Portex, ID 0.58 mm, OD 0.96 mm),
heated to create a collar, was passed through a small incision
at the apex of the bladder dome. A suture was tightened
around the collar of the catheter and, to prevent urine
leakage, the external tip of the catheter was closed by heat.
For intravenous (iv) injection, another polyethylene catheter
(Portex, ID 0.58 mm, OD 0.96 mm) filled with physiological
heparinized saline (40 UI/ml) was inserted into the jugular
vein. A trocar was passed through the abdominal muscles at
1 cm from the upper end of the incision and tunneled
subcutaneously to emerge through the skin in the back of
the neck. The catheters were then passed through the trocar,
where a 1 cm-long piece of Silastic tubing (ID 0.30 mm;
OD 0.65 mm) was inserted around the external part of the
bladder catheter to prevent its slipping back under the skin of
the animal. The rats were allowed to recover from the
anesthesia, and the experiments were carried out 1 day after
catheter implantation.
2.3. Evaluation of urodynamic parameters
On the day of the experiment, the rats were placed in
modified Bollman’s cages that were large enough to
permit them to adopt a normal crouched posture but
narrow enough to prevent turning around. These cages
also enabled measurements of micturition volumes by
means of a fluid collector connected to a force isometric
transducer (Basile 7003 DY1).
The following schemes of experiments were performed
(see Fig. 1 for details):
1. After a stabilization period of 20 min, the free tip
of the bladder catheter was connected by a T-shaped tube
to a transducer (Statham P 23 XL or Marb type P82).
The physiological micturition was evaluated on water-
loaded rats (15 ml/rat by oral gavage) by recording three
to five cystometrograms.
2. On the same rats, cystometrograms were also obtained,
opening the other side of the T-shaped tube connected with
switched-off pumps.
3. In another group of water-loaded rats, after record-
ing the physiological micturition, pump-evoked cystomet-
rograms were obtained by connecting the T-shaped tube
alternatively to two different pumps to perform a con-
tinuous infusion into the bladder, at a constant rate of
0.1 ml/min.
Two different kinds of pumps were utilized: a peristaltic
pump (Gilson minipuls II), equipped with polyvinylchlo-
ride tube 0.38 mm ID, or an infusion pump (Harvard
model 22), equipped with plastic syringes of different
volumes (2.5–10–20–50 ml).
Pumps, pressure transducers, and bladder catheters were
connected to each other by polyethylene tube (Portex, ID
0.58 mm, OD 0.96 mm) of the same length filled with
saline, as exemplified in Fig. 1.
The urodynamic parameters evaluated from cystome-
trograms recorded on the polygraph (ALFOS WK-480R
with BM 614/2 preamplifier from Biomedia Mangoni)
were: micturition pressure (MP), micturition time (MT),
and micturition volume (MV). Bladder volume capacity
(BVC) was evaluated with pump-evoked cystometro-
grams. MP (in mmHg) is defined as the maximal
intravesical pressure induced by contraction of the
detrusor during micturition. MV (in milliliters) is defined
as the urinated volume expelled in a single micturition.
MT (in second) is defined as the time elapsed between
the threshold pressure point and the return to basal
pressure after voiding. BVC (in milliliters) is defined
as the volume of saline infused into the bladder between
two micturitions.
2.4. Evaluation of the compliance of the systems
The total compliance (catheter + pump) of the system
was evaluated after clamping off flow on the animal
catheter and connecting alternatively the two (switched-
on) different pumps to the pressure transducer via the
T-shaped tube by using the pump catheter. Compliance
was determined for pressure increase between 0 and
100 mmHg.
C. Velasco et al. / Journal of Pharmacological and Toxicological Methods 45 (2001) 215–221216
2.5. Evaluation of mechanistically different drugs on the
bladder voiding cycle
To verify whether the system used in awake animals
might interfere with the effects of drugs, groups of rats were
treated iv with 0.3 mg/kg of prazosin or oxybutynin, as well
as with 3.0 mg/kg of naproxen. Matched vehicle-treated rats
were also evaluated. Cystometrograms were obtained by
infusing saline into the bladder at the constant rate of
0.1 ml/min by an infusion pump equipped with 50ml syringe
or a peristaltic pump using 0.38 mm (ID) tubing. Basal
BVC and MP values were evaluated as mean of two
complete and reproducible cystometrograms. Basal MV
was the mean of volumes collected in the same cystometro-
grams. At this point, drugs or vehicles were administered
intravenously by bolus injection, with concomitant continu-
ous infusion of the bladder. Changes in urodynamic param-
eters were evaluated as average of the second and third
cystometrogram after treatment. The same protocol was
applied to groups of rats in which natural fill was studied.
2.6. Drugs and chemicals
Oxybutynin HCl and prazosin HCl were purchased from
Sigma-Aldrich (Milano, Italy). Naproxen was from Record-
ati S.p.A., Milano. All the other substances were from
commercial sources.
2.7. Statistical analysis
Statistical significance of the differences in urodynamic
parameter values in conscious rats before and after drug
treatments was evaluated by Student’s t test for paired data.
Two-way ANOVA and Dunnett’s test was used to evaluate
the differences of cystometrographic parameters recorded
during physiological micturition and during pump-evoked
cystometry. To improve the degree of confidence, 1% level
of probability was always retained to distinguish two
different means.
3. Results
3.1. Comparison between natural and pump-evoked
cystometrograms
Urodynamic parameters (MP, MV, & MT) were eval-
uated in a group of rats after an oral water load to facilitate
natural micturition without bladder infusion. Cystometro-
grams were initially obtained without any connection to the
Fig. 1. Scheme showing equipment utilized to perform cystometry in unanesthetized rats. Saline was alternately delivered into the bladder through chronically
implanted catheter by two different pumps: a peristaltic pump or an infusion pump equipped with syringes of different volumes. Animals were relatively
restrained and located over a stainless steel funnel that emptied into a strain gauge-mounted cup. Bladder pressure was monitored by a transducer connected
with the rats and with one of the pumps via T-shaped connection tube. Outputs of transducer and strain gauge were simultaneously monitored on a polygraph.
The following schemes of experiments were performed: (a) physiological micturition was evaluated on water-loaded rats choking the catheter connecting
pumps. On the same rats, natural cystometrograms were also obtained connecting the T-shaped tube with switched-off pumps. Pump-evoked cystometrograms
were obtained connecting the T-shaped tube alternatively with the different pumps; (b) total compliance (catheter + pump) of the system was evaluated after
choke of flow on the animal catheter and connecting alternatively the two different pumps to the pressure transducer via the T-shaped tube by using the pump
catheter. Compliance was determined for pressure increase between 0 and 100 mmHg; (c) the effects of the different drugs tested were evaluated on pump-
evoked micturition filling the bladder alternatively with the peristaltic pump or the syringe pump connected with the T-shaped tube by the pump catheter.
C. Velasco et al. / Journal of Pharmacological and Toxicological Methods 45 (2001) 215–221 217
pumps (see Fig. 1), and the recorded values are summarized in
the Table 1 as physiological micturition (Table 1). On the
same animals, other cystometrograms were later recorded
after the connection of the switched-off pumps. In the case of
the infusion pump, different volume syringeswere also tested.
No statistically significant differences in MV and MT
were observed between physiological micturition (with no
pumps connected) to that obtained when the different pumps
(switched off) were connected into the system. In contrast,
MP was significantly reduced after connection with infusion
pump equipped with 50 ml and 20 ml syringes. MP values
recorded connecting the peristaltic pump were the same as
those recorded during physiological micturition.
In another group of orally water-loaded rats, physio-
logical micturition (no pumps connected, then followed by
bladder infusion using two different pumps) yielded no
significant differences in MV and MT (Table 1). MP values
recorded with the infusion pump equipped with 50 and 20 ml
volume syringes were not significantly different from those
during physiological micturition. In contrast, significantly
higher MP values were observed with syringes of 10 and
20 ml and also with the use of a peristaltic pump (Table 1).
3.2. Measurement of total (catheter + pump) compliance
The peristaltic pump fitted with 0.38 mm (ID) tube or
the infusion pump (fitted with different volume syringes)
were connected alternatively to the pressure transducer via
the T-shaped tube by using the same catheter and setting the
flow to 0.1 ml/min (1.67 ml/s). Initially the T-tube allowed
free outflow through the animal catheter (Fig. 1). A
pressure ramp was obtained by clamping off flow on the
animal catheter, and compliance was determined for pres-
sure increases between 0 and 100 mmHg. Under these
experimental conditions, the pressure increased as a linear
function of time according to the following equation (Ask
et al., 1977; Ask & Hok, 1990): P = F�t / C, where F is
the infusion flow, C is the compliance of the system and t
is the time. The slope of the pressure versus time plot i.e.,
the pressure rise rate represents therefore the ratio F/C.
The compliance values obtained with the infusion pump
varied from 1.7089 ml/mmHg with the 50 ml syringe to
0.0830 ml/mmHg with the 2.5 ml syringe. These values
correspond to pressure rise rates of about 1 mmHg/s by
using the 50 ml syringe to 20 mmHg/s with the 2.5 ml
syringe. The compliance of the system using the peristaltic
pump was 0.0123 ml/mmHg, corresponding to pressure rise
rate of about 136 mmHg/s. These data indicate that the
compliance of the system utilizing the infusion pump
equipped with a 50 ml volume syringe is typically about
100-fold lower than that utilizing the peristaltic pump
equipped with 0.38 mm (ID) tube.
On the other hand, when the peristaltic pump was
equipped with a 2.28 mm (ID) tube the compliance
evaluated as described above was 0.1412 ml/mmHg, cor-
responding to a pressure rise rate of about 11.8 mmHg/s.
This is similar to the value obtained with the infusion
pump equipped with 10 ml syringe (compliance = 0.1315 ml/mmHg and pressure rise = 12.7 mmHg/s).
As shown in Fig. 2, the different compliance values
obtained might explain the results of Table 1. In both the
experimental situations (switched-off or switched-on
pumps) the differences in MP values recorded during
physiological micturition and after the connection with the
pumps were linearly related to compliance values.
In a further experiment, cystometrographic recordings
performed on rats with bladder infused with the peristaltic
pump equipped with a tube of 2.28 mm (ID) gave a mean MP
value of 57.2 mmHg. This value was very similar to that
obtained during infusion with 10 ml syringe, having similar
compliance, confirming that the different MP values
recorded during pump-evoked micturition are related to the
compliance of the system and not only to the pump utilized.
3.3. Comparison of the effects of different drugs on BVC
and MP values obtained during physiological or
pump-evoked micturition
As far as the BVC is concerned, basal values recorded in
the different groups of animals were in the range from 0.4 to
0.8 ml, and no difference in these values was observed with
the different pumps or during physiological micturition
(Table 2). MV values were always similar to the BVC
values (data not shown).
Table 1
Comparison of micturition parameters measured in conscious rats
Infusion pump
Physiological micturition 50 ml 20 ml 10 ml 2.5 ml Peristaltic pump
A (n = 6) MP (mmHg) 21 ± 1.0 12 ± 1.1* 16 ± 1.6* 19 ± 2.0 18 ± 1.1 22 ± 1.5
MV (ml) 0.44 ± 0.08 0.42 ± 0.07 0.48 ± 0.14 0.46 ± 0.10 0.46 ± 0.09 0.41 ± 0.07
MT (s) 14 ± 1.8 14 ± 2.3 15 ± 2.3 13 ± 1.7 13 ± 1.8 13 ± 3.1
B (n = 9) MP (mmHg) 24 ± 4.2 19 ± 1.1 27 ± 2.4 51 ± 3.4* 66 ± 4.3* 87 ± 7.5*
MV (ml) 0.45 ± 0.08 0.46 ± 0.05 0.42 ± 0.09 0.46 ± 0.09 0.45 ± 0.07 0.37 ± 0.04
MT (s) 12 ± 1.3 13 ± 1.4 12 ± 1.2 12 ± 0.9 11 ± 0.7 11 ± 0.5
Parameters were measured on cystometrograms obtained from the same animals: (a) without connection to infusion pumps (physiological micturition), (b) after
connection to the pumps but with the pumps switched off (A), or (c) during pump-evoked cystometrograms (B). Data represent the mean values ± S.E.
* P < .01 versus physiological micturition values (two-way ANOVA and Dunnett’s t test).
C. Velasco et al. / Journal of Pharmacological and Toxicological Methods 45 (2001) 215–221218
In the vehicle-treated rats, no significant changes of BVC
were observed, either during natural micturition or when the
bladder was infused with the peristaltic or the infusion
pump. Prazosin and oxybutynin induced non-statistically
significant changes of BVC that were roughly of the same
magnitude in all kind of cystometrograms. Naproxen
induced a BVC increase of about 40% in water-loaded
animals and during pump-evoked micturition.
As expected, basal MP values were markedly different
when the two different pumps were utilized. Cystometro-
grams recorded with the infusion pump equipped with 50 ml
syringe gave basal MP values in the range 20–30 mmHg as
observed in the water-loaded groups of rats, whereas with
the peristaltic pump these values ranged 65–100 mmHg.
When the infusion pump was utilized, only oxybutynin
treatment induced a significant (but no more than 14%)
decrease of MP. When the peristaltic pump was used,
prazosin treatment did not changed MP, naproxen induced
a nonsignificant decrease of the parameter (about 20%),
and the effect of oxybutynin was much higher, reaching
55% decrease of MP. The drug-induced changes of MP
recorded on cystometrograms of water-loaded rats were
close to those observed in the groups infused with the
peristaltic pump.
Table 2
Effect of iv administration of prazosin, oxybutynin, and naproxen on bladder volume capacity (BVC) and micturition pressure (MP) during physiological
micturition (water-loaded animals) and pump-evoked cystometry parameters in conscious rats
BVC or MV (ml) MP (mmHg)
Treatments (mg/kg iv) No. rats Before After % change Before After % change
Physiological micturition Vehicle 5 0.60 ± 0.12 0.66 ± 0.14 10.0 23 ± 2.3 22 ± 2.2 � 4.3
Prazosin (0.3) 5 0.76 ± 0.08 0.84 ± 0.12 10.5 23 ± 2.5 21 ± 1.3 � 8.7
Oxybutynin (0.3) 5 0.72 ± 0.13 0.70 ± 0.11 � 2.8 18 ± 3.1 11 ± 2.8* � 38.9
Naproxen (3.0) 7 0.66 ± 0.10 0.91 ± 0.18* 37.9 25 ± 3.3 25 ± 3.3 0
Infusion pump Vehicle 9 0.44 ± 0.03 0.43 ± 0.05 � 2.3 22 ± 0.8 23 ± 1.0 4.5
Prazosin (0.3) 9 0.70 ± 0.11 0.76 ± 0.21 8.6 25 ± 1.8 21 ± 1.8 � 16.0
Oxybutynin (0.3) 8 0.51 ± 0.08 0.40 ± 0.04 � 21.6 21 ± 1.9 18 ± 1.9* � 14.3
Naproxen (3.0) 9 0.51 ± 0.07 0.71 ± 0.10* 39.2 23 ± 1.4 22 ± 1.5 � 4.3
Peristaltic pump Vehicle 5 0.54 ± 0.12 0.55 ± 0.15 1.9 91 ± 12.5 85 ± 10.4 � 6.6
Prazosin (0.3) 8 0.71 ± 0.19 0.73 ± 0.16 2.8 79 ± 9.2 80 ± 9.4 1.3
Oxybutynin (0.3) 10 0.51 ± 0.07 0.48 ± 0.09 � 5.9 100 ± 10.3 45 ± 4.9* � 55.0
Naproxen (3.0) 6 0.54 ± 0.11 0.71 ± 0.11* 31.5 110 ± 11.1 89 ± 8.5 � 19.1
Bladder infusion was performed with the infusion pump equipped with a 50 ml syringe or with the peristaltic pump equipped with 0.38 mm (ID) tube. Data
represent the mean values ± S.E. of BVC and MP and the % changes versus basal values.
* P < .01 versus basal values (Student’s t test for paired data).
Fig. 2. (a) Correlation between the differences of MP values (4 mmHg) recorded during physiological micturition minus the values in the same animals
recorded after connection with switched-off pumps (data from Table 1A), and the compliance of the pump + catheter systems. Correlation coefficient = .9737;
P < .01. (b) Correlation between compliance of the pump + catheter systems and the difference between the observed MP values recorded after connection with
the switched-on pumps and the MP values recorded during ‘‘physiological micturition’’ (data from Table 1B). Correlation coefficient = .9763; P < .01.
C. Velasco et al. / Journal of Pharmacological and Toxicological Methods 45 (2001) 215–221 219
4. Discussion
Cystometry is a widely used in vivo method for quant-
itative studies on physiopharmacology of micturition in
several animal species (Peterson et al., 1989, 1990; Sjogren,
1976; Nishizawa et al., 1989; Blok et al., 1985; Klevmark,
1974; Shoukry & Ghonicm, 1992; Yaksh et al., 1986;
Maggi et al., 1986a, 1986b, 1986c), as well as to evaluate
changes in urodynamic parameters associated with urinary
incontinence in humans (Abrams et al., 1988).
During cystometry in animals, continuous filling of
bladder is obtained by an intravesical catheter connected
to a delivery system, which pumps saline into the bladder at
constant rate. Usually, a peristaltic or an infusion pump is
used, simultaneously filling the bladder of four to eight
animals with saline continuously delivered to the bladder
also during the micturition.
A literature review of urodynamic parameters, especially
regarding micturition pressure, revealed remarkable diffe-
rences (Durant & Yaksh, 1988; Conte et al., 1988; Guarneri
et al., 1991; Yamamoto et al., 1995; Ishizuka et al., 1997;
Persson et al., 1998; Chai et al., 1999). Even under similar
experimental conditions micturition pressures between
about 40 and 90 cm H2O have been reported (Persson
et al., 1998; Ishizuka et al., 1997; Igawa et al., 1993).
We have examined, therefore, in the same conscious
animals, urodynamic parameters, from cystometrograms
obtained during spontaneous voiding after an oral water
load, and pump-evoked micturition, using two different kind
of pumps.
No significant differences were seen in MV and MT
recorded during natural voiding cycle after connection with
switched-off pump or during pump-evoked micturition.
In contrast, MP recorded when the switched-off infusion
pump, equipped with 50 and 20 ml syringes, was utilized was
significantly lower than that recorded during physiological
micturition. The hydrostatic pressure exerted by the bladder
duringcontractionmaybeconsideredconstant.Whenthe50ml
syringe is connected, the pressure rise speed of the system is
about 1 mmHg/s. Taking into account that micturition lasted
13–14 s, nomore than 13–14mmHgcould be recorded by the
system.After connection of the 20ml syringe, the rise speed of
pressure is about 4 mmHg/s. This could allow the system to
reach the same pressure as during physiological micturition.
The lower values of MP reached with both of these syringes
might due to the larger area (syringe diameter) on which the
stiffness of the bladder is exerted during contraction.
During pump-evoked cystometrograms, MP showed
increasing values, moving from the 50 ml infusion pump to
infusion pumps with lower volumes (up to a 3-fold increase
with 2.5 ml volume syringe) and to the peristaltic pump (up to
a 4-fold increase). When the bladder pressure reaches a
threshold pressure, the urethral sphincter opens and, theoret-
ically, detrusor pressure could remain at the level of threshold
pressure during micturition. However, the continuous deliv-
ery of fluid to the bladder by the infusion pumps alters bladder
pressure. The degree of pressure detected by the transducer
could be again related to the surface of the system (syringes or
tube diameter) and speed in recording the pressure rise.
Therefore, the similarity between MP values recorded during
pump-evoked cystometrograms using the infusion pump
equipped with 50 and 20 ml volume syringes and those
recorded during physiological micturition (Table 1), as well
as the higher values recorded with low volume syringes and
peristaltic pump, were all influenced by system artefacts.
Finally the differences in the MP values recorded after
connection with the different pumps and the values during
physiological micturition were found to be a function of the
different compliance of the systems, and not dependent
upon the pump.
The effects of prazosin, oxybutynin, and naproxen on
BVC evaluated by performing cystometry with peristaltic
and infusion pump, as well as in water-loaded animals, were
almost the same. Prazosin and oxybutynin did not signifi-
cantly change BVC, and naproxen was found to increase
BVC (about 40% in all the experimental situations).
Using the infusion pump equipped with 50 ml volume
syringe, oxybutynin reduced slightly (about 15%) but
significantly bladder contractility, whereas naproxen was
inactive. On the contrary, when higher MP values are
reached by infusing the bladder with the peristaltic pump,
the effect of oxybutynin is highlighted and naproxen also
affected, although not significantly, the parameter, whereas
prazosin remained inactive.
It is noteworthy that although obtained in an experimental
condition in which the recorded MP values can be considered
an artefact, the results obtained with peristaltic pump are in
agreement with those obtained during physiological micturi-
tion in water-loaded animals. Water-loaded animals, how-
ever, cannot be considered a good screening model to
evaluate the effects of compounds on micturition. After water
load, a series of voiding cycles sufficiently close to allow the
evaluation of basal values and the effect of a treatment can be
obtained only considering the protocol utilized in the present
paper (IV treatment and evaluation of the effect considering
the second and third cystometrogram after treatment). Owing
to the fact that after less than 1 h diuresis become normal, a
time course of the effects after an iv administration or after
oral administration of a compound is not feasible.
In conclusion, the present findings indicate that only the
effects on cystometrographic parameters (expressed as per-
cent changes) obtained during pump-evoked micturition
with a high compliance system (peristaltic pump) can be
considered equivalent to those observed during physio-
logical micturition.
References
Abrams, P., Blaivas, J. G., Stanton, S. L., & Sndersen, J. T. (1988). The
standardisation of terminology of lower urinary tract function. Scand J.
Urol. Nephrol., 114 (Suppl.), 5–19.
C. Velasco et al. / Journal of Pharmacological and Toxicological Methods 45 (2001) 215–221220
Ask, P., & Hok, B. (1990). Pressure measurement techniques in urodynamic
investigations. Neurourol. Urodyn., 9, 1–15.
Ask, P., Oberg, A., & Tibbling, I. (1977). Static and dynamic characteristics
of fluid-filled esophageal manometry systems. Am. J. Physiol., 233,
E389–E396.
Blok, C., van Venrooij, G. E., & Coolsaet, B. L. (1985). Dynamics of the
ureterovesical junction: effectiveness of its peristalsis in high pressure
pig bladders. J. Urol., 134, 825–827.
Chai, T. C., Gemalmaz, H., Andersson, K. E., Tuttle, J. B., & Steers, W. D.
(1999). Persistently increased voiding frequency despite relief of blad-
der outlet obstruction. J. Urol., 161, 1689–1693.
Conte, B., D’Aranno, V., Santicioli, P., Giuliani, S., Mancinelli, A., Furio,
M., Maggi, C. A., & Meli, A. (1988). New method for recording cys-
tometrograms in conscious, freely moving rats. J. Pharmacol. Meth.,
19, 57–61.
Durant, P. A., & Yaksh, T. L. (1988). Drug effects on urinary bladder tone
during spinal morphine-induced inhibition of the micturion reflex in
unanesthetized rats. Anesthesiology, 68, 325–334.
Guarneri, L., Cova, R., Angelico, P., Colli, E., & Testa, R. (1991). Effects
of different drugs on the cystometrogram in conscious rats. Pharm.
Research, 24, 175–187.
Igawa, Y., Mattiasson, A., & Andersson, K.-E. (1993). Functional impor-
tance of cholinergic and purinergic neurotransmission for micturition
contraction in the normal, unanaesthetized rat. Br. J. Pharmacol.,
109, 473–479.
Ishizuka, O., Pandita, R. K., Mattiasson, A., Steers, W. D., & Andersson,
K. E. (1997). Stimulation of bladder activity by volume, L-dopa
and capsaicin in normal conscious rats — effects of spinal
a1-adrenoceptor blockade. Naunyn. Schmied. Arch. Pharmacol.,
355, 793–797.
Klevmark, B. (1974). Motility of the urinary bladder in cats during
filling at physiological rates. I. Intravesical pressure patterns studied
by a new method of cystometry. Acta Physiol. Scand., 90, 565–577.
Maggi, C. A., Giuliani, S., Santicioli, P., & Meli, A. (1986a). Analysis of
factors involved in determining urinary bladder voiding cycle in ure-
thane.anesthetized rats. Am. J. Physiol., 251, R250–R251.
Maggi, C. A., Santicioli, P., & Meli, A. (1986b). The non-stop transvesical
cystometrogram in urethane-anesthetized rats: a simple procedure for
quantitative studies on the various phases of urinary bladder voiding
cycle. J. Pharmacol. Methods, 15, 157–167.
Maggi, C. A., Santicioli, P., & Meli, A. (1986c). Somatovesical and
vesicovesical excitatory reflexes in urethane-anesthetized rats. Brain
Res, 380, 83–93.
Matsuura, S., & Downie, J. W. (2000). Effect of anesthetics on reflex
micturition in the chronic cannula-implanted rat. Neurourol. Urodyn.,
19, 87–99.
Mosso, M. A., & Pellacani, P. (1892). Sur le fonctions de la vessie. Arch.
Ital. Biol., 1, 97-128, 291–325.
Nishizawa, O., Sugaya, K., Kohama, T., Satoh, K., Harada, T., & Tsuchida,
S. (1989). Effect of oxybutynin on reflex micturition in the decerebrate
dog as determined by urodynamic evaluation. Neurorol. Urodyn., 8,
513–519.
Peterson, J. S., Hanson, R. C., & Noronha-Blob, L. (1989). In vivo cysto-
metrogram studies in urethane-anesthetized and conscious guinea-pig.
J. Pharmacol. Meth., 21, 231–241.
Peterson, J. S., Patton, A. J., & Noronha-Blob, L. (1990). Mini-pig urinary
bladder function: comparisons of in vitro anticholinergic response and
in vivo cystometry with drugs indicated for urinary incontinence.
J. Auton. Pharmacol., 10, 65–73.
Persson, K., Pandita, R. K., Spitsbergen, J. M., Steers, W. D., Tuttle, J. B.,
& Andersson, K. E. (1998). Spinal and peripheral mechanisms con-
tributing to hyperactive voiding in spontaneously hypertensive rats.
Am. J. Physiol., 275, R1366–R1373.
Shoukry, M. S., & Ghoniem, G. M. (1992). Continuous monitoring of
micturition pattern in nonhuman primates: similarities to humans.
Neurourol. Urodyn., 11, 139–144.
Sjogren, C. (1976). The effects of some anticholinergic compounds on the
infused urinary bladder of unanaesthetized rabbits. Acta Pharmacol.
Toxicol., 13, 157–165.
Yaksh, T. L., Durant, P. A. C., & Brent, C. R. (1986). Micturition in rats: a
chronic model for study of bladder function and effect of anesthetics.
Amer. J. Physiol., 251, R1177–R1185.
Yamamoto, T., Koibuchi, Y., Miura, S., Sawada, T., Ozaki, R., Esumi, K., &
Ohysuka, M. (1995). Effects of vamicamide on urinary bladder func-
tions in conscious dog and rat models of urinary frequency. J. Urol.,
154, 2174–2178.
C. Velasco et al. / Journal of Pharmacological and Toxicological Methods 45 (2001) 215–221 221