21/5/2007 Thomas S. Bauer - Nikhef 1

Input Mode Cleaner

Update

LIGO-G070343-00-Z

Page 2: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 2

IMC

Reminder:

New end mirror ordered (14 cm vs.. 8 cm);

needs new Payload -- Roma 1;

and new installation tooling -- Nikhef ;

In addition:

losses in IMC close to 50 %, needs action at level of input bench: DiHedron.

Since beg. 2007, Nikhef responsible for IMC.

Page 3: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 3

IMC End Mirror

End mirror Payload: already started by Roma -1

Installation tooling: at present designed by

Frans Mul (Nikhef).

next few slides are by Frans.

Page 4: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 4

VIRGO MODE CLEANER END-MIRROR INSTALLATION [email protected]

First step, mirror on support

CONCEPT

Page 5: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 5

Vacuum clamping

Mirror support

MirrorRecoil mass

Recoil support

VIRGO MODE CLEANER END-MIRROR INSTALLATION [email protected]

Gear boxLifting device

Installation frame

Vacuum pump2 mbar vacuum4.8 l/minDC 12 V 2 Amp

Page 6: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 6

VIRGO MODE CLEANER END-MIRROR INSTALLATION [email protected]

Front viewRear view

Translation

Side view

TranslationTranslation

Tra

nsla

tion

Tra

nsla

tion

Rotation

Tilt & keep

Movements

Page 7: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 7

CENTRE OF GRAVITY

Total weight 71 kg

VIRGO MODE CLEANER END-MIRROR INSTALLATION [email protected]

Page 8: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 8

CENTRE OF GRAVITY

Total weight 71 kg

VIRGO MODE CLEANER END-MIRROR INSTALLATION [email protected]

VARTADeep Cycle battery 12V/ 44Ahhxbxd: 225 x 176 x 409 38 kg€ 192

Page 9: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 9

VIRGO MODE CLEANER END-MIRROR INSTALLATION [email protected]

Page 10: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 10

VIRGO MODE CLEANER END-MIRROR INSTALLATION [email protected]

Page 11: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 11

VIRGO MODE CLEANER END-MIRROR INSTALLATION [email protected]

THE END

114kg

Page 12: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 12

TimescaleMirrorRM design

productionMarionetta design

productionFrame design

productionTests and optim.Installation

M J J A S O N D J F M 2007 2008

Page 13: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 13

Timescale ITF ShutdownOld Mirror dismountMarionetta dismount

machiningre-mounting

SA pre-tuningNew MC installationSA post-tuning

TotalTower closed:

Few days1 -2 days

1 week ? more ?1-2 days1 week Few days2 weeks

min 5 … 6 weeksend April 08

Page 14: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 14

End mirror Action :Finalize tooling design:

few measures need to be fixed, no work;

Approve and produce it.

Decide about Marionetta: Options: - modifying the existing one, will take

precious time when tower is open;

- produce a new one (can be done at Nikhef) will save time (and surprises)

Page 15: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 15

DiHedron…

Problems:

a) Present DiHedron "jumps";

b) IMC transmission about ~ 55..60 %- due to too small aperture?- or rather to poor quality*) mirrors?

*) - could be "dirty"

Page 16: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 16

Present DiHedron

Injection Bench

Page 17: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 17

Present DiHedron

Page 18: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 18

Options DiHedronClamp existing DiH on Bench, or suspend it:

solves only jumping, not losses

Clean mirrors:not easy - chance for success… ?

Replace DiH by 2 suspended mirrors (GEO600)quite involved, crowded IB, commissioning?

new DiH with larger apertures:solves only part of problems

Other ???

Page 19: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 19

Present DiHedron

Mirr

or

Mirror Beam

Channel for alignment

Page 20: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 20

inverted DiHedron (?)

Mirr

or

Mirror

Beam

Page 21: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 21

FEA inverted DiHedron L = 55

D = 54

S =20

R1 =10

R2 =17.5

R3 =40

T =30

H1 =90

H2 =70

H3 = 5

L

H1

H2

Page 22: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 22

Oscillation in X-Y plane Oscillation in X-Z plane 3D-Torsional mode

4250 Hz 7500 Hz 4600 Hz

Page 23: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 23

QuickTime™ and aMicrosoft Video 1 decompressorare needed to see this picture.

Page 24: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 24

Advantages vs. Problems2 "minitowers" for independent mirrors complicated:

- very crowded area,

- commissioning needs quite considerable time

Inverted DiHedron:

need to get good idea on whether it is possible;

prelim. FEA indicates: lower frequency

Seems to be a nice idea, but is it a viable option??

Page 25: Input Mode Cleaner

21/5/2007 Thomas S. Bauer - Nikhef 25

SummaryEnd mirror:

Design tooling, Gearbox, started

need to finalize and approve it

start production

Marionetta: produce a new one? or modify existing one?

DiHedron:

available information collected, no decision yet.

Independent mirror suspension looks quite complicated.New Dihedron with larger aperture ? inverted?