Sample to Insight
Integrated Solutions For Sexual Assault Analysis
Sample to Insight
Differential Extraction
Separation of non-sperm DNA and sperm DNA
Differential wash workflow
Identify stains and take sample
that contains
Epithelial/non-sperm cells
Sperm cells
Digestion of non-sperm cells
Washing and separation
of non-sperm cells
Digestion of sperm cells
Non sperm/Epithelial DNA Sperm Fraction
Sperm DNA
Sample to Insight
Pyrosequencing for Sperm identification in forensic casework
Sample to Insight
Epigenetics for tissue identification
4
Epigenetic marker analysis CpG methylation degree
Each tissue has its own marker pattern
Requires sensitive analysis of multiple, consecutive CpGs
Sample to Insight
Methylation analysis for tissue identification
5
Analysis of tissue specific DNA methylation using Pyrosequencing
Methylation analysis
o CpG methylation works as epigentic switch to control tissue specific gene regulationo Each tissue has its own DNA methylation pattern o Loci were identified that can be used to distinguish between the following sample types:
o Bloodo Salivao Semeno Epithelial tissue
Sample to Insight
Methylation analysis for tissue identification
6
Analysis of tissue specific DNA methylation using Pyrosequencing
The determination of tissue-specific DNA methylation patterns in forensic biofluids using bisulfite modification and pyrosequencing. Madi, T., Balamurugan, K., Bombardi, R., Duncan, G, and McCord, B. Electrophoresis 2012, 33, 1736–1745
Sample to Insight
Find out more…
7
Application Note: San Francisco Police Department, Florida International University
Sample to Insight
QIAGEN’s workflow for sexual assault analysis
8
Main goals
o Automate differential separation of sexual assault samples to accelerate your analysiso Simplify isolation of challenging samples for best possible yield and quality of DNAo Obtain detailed and accurate information on the male DNA in your sampleso Perform rapid and reliable downstream analysis with the Investigator assays
Sample to Insight
Differential wash, lysis and DNA purification
9
Sample to Insight
Automated “manual” processing
10
The QIAcube differential wash workflow
Add wash buffer
Pellet sperm
Repeat 3x
Add sperm lysis buffer
Sperm fraction
Non-sperm fraction
Load mixture, Epi lysed
Pellet sperm
Transfer non-spermfraction to 2 ml tubes
1
2
3
4
5
6
6 samples <1hr12 samples ~1.5hr
Sample to Insight
QIAcube workflow for differential lysis
11
Up to 48 differential wash samples per day, 2 operators
EZ1 Adv XL QIAcube for automateddifferentiall wash
• Fully automated differential fraction separation and sperm pellet lysis • Up to 12 samples/run• The QIAcube can be used up front of other automation workflows• Designed to run on the QIAcube, the protocol is directly compatible with QIAGEN’s EZ1,
QIAsymphony or QIAamp DNA Investigator Kit chemistry for downstream DNA purification.
Sample to Insight
Find out more…
12
Customer poster: Alabama Department of Forensic Sciences
Sample to Insight
Assay setup
13
Sample to Insight
Assay setup
14
Quantiplex Pro RGQ reaction setup
Template Files
QProtocols
Sample to Insight
DNA Quantification
15
Sample to Insight
Rotor-Gene Q system
o Outstanding thermal and optical performance
o Fast cycling (40 cycles in 40 minutes)
o Up to 6 fold multiplexing
o High resolution melting and fast PCR
o Tube & disc formats (1-100 samples)
o Minimal maintenance and calibration
o Versatile software package
Sample to Insight
Investigator Quantiplex Pro RGQ Kit
17
Key features
• New male target
◦ Superior sensitivity with new male target sequence and small target length
◦ Detection threshold up to 1:2,500,000 (male:female)
• New Degradation Index System
◦ Unique design for small and large autosomal targets
◦ Unique design for small Y and large Y targets
◦ Assess degradation status specifically of male DNA in female:male mixtures
◦ Minimize the risk that your Y-STR fails due to degraded or partially degraded male DNA
• Internal Control for inhibitor monitoring
◦ Correlates very well with the STR results
Sample to Insight
Investigator Quantiplex Pro RGQ Kit – Performance Data
18
Sensitivity
• Dynamic range from 200 ng/µl down to 0.5 pg/µl
• Accurate quantification of low amounts of DNA
Sensitive detection of traces, robust at high template input
Amplification PlotStandard Curve
Sample to Insight
Investigator Quantiplex Pro RGQ Kit – Performance Data
19
Mixture analysis
• 0.2 pg/µl male DNA was spiked to up to 200 ng/µl female DNA• Male and female DNA was robustly and accurately quantified
400.000:1
Reliable quantification of male DNA in the presence of excess female DNA
0.5pg/µl male DNA + 100ng/µl female DNA (1:200.000)0.25pg/µl male DNA + 100ng/µl female DNA (1:400.000)0.2pg/µl male DNA + 200ng/µl female DNA (1:1.000.000)
Sample to Insight
Investigator Quantiplex Pro RGQ Kit – Performance Data
20
Degraded samples
• Male genomic DNA was sheared with a Covaris™ S220 Focused-ultrasonicator to an average fragment size of 1500 bp, 500 bp, 400 bp , 300 bp, 200 bp and 150 bp
• 4.6 ng used as template for quant
Investigator 24plex QS Kit with 150 bp sampleInvestigator Quantiplex Pro RGQ Kit
Reliable prediction of sample integrity and accurate DNA degradation detection
Sample to Insight
Investigator Quantiplex Pro RGQ – detects degraded male DNA
Comparison data
The QuantiplexPro RGQ system for DNA degradation allows for detection of degraded male DNA even in a high background of female DNA – competitor kits fail to detect degraded male DNA in female DNA background.
Each sample contains 0.1 ng/μl of degraded male DNA mixed with 100 ng/μl of non-degraded female DNA.
Sample to Insight
Investigator Quantiplex Pro RGQ Kit – Performance Data
Internal control
• Humic acid up to 40 ng/μl (final concentration in PCR) was added to reactions amplifying 1 ng of male DNA
◦ Corresponds to 250 ng/µl in STR if maximum template volume of 15 µl is applied
• Robust quantification results up to critical inhibitor concentrations
• The Quantiplex Pro IC correlates to the STRs and reflects the Quality Sensor of STR kits
0 ng/µL
150 ng/µL
200 ng/µL
250 ng/µL
Investigator 24plex QS
The Internal Control reports the presence of inhibitor with a CT shift, while quantification remains reliable up to a final humic acid concentration of 40 ng/μl
Quantiplex Pro RGQ
Sample to Insight
STR Analysis
23
Sample to Insight
Investigator STR QS Kits
Investigator Quality Sensor
Argus X-12 QS24plex QS Kit 24plex GO! Kit ESSplex SE QS• 23 markers of CODIS
expansion set• BT6 matrix• FRM 2.0
• 23 markers of CODIS expansion set
• BT6 matrix• FRM 2.0
• 17 markes of ESS set
• BT5 matrix• FRM 2.0
• 12 XChr markers, D21S11 and AM
• BT5 matrix• FRM 2.0
24
Sample to Insight
Investigator STR Kits
25
Kit features
• 24 plex QS/GO! kits : 23 markers recommended by CODIS Working Group
• Innovative Quality Sensors for more information and workflow optimization
• 6 color setup with new BT6 matrix
• New Reaction Mix FRM 2.0
◦ Even higher convenience and stability
Up to 16h stability of PCR mixed at RT
◦ Convenient manual and automated pipetting
◦ Even faster reaction speeds
60 minutes for casework, ca. 45 min for GO!
◦ Very high inhibitor tolerance
25
Sample to Insight
Investigator Quality Sensor
Internal Performance Control – more information for better results
• The kits include two Quality Sensors that will provide you with more information about a profile to improve your workflow and save unnecessary re-runs of a sample
• The two internal Quality Sensors are included in the Primer Mix and are amplified simultaneously with the STR markers
• The Quality Sensor gives one small (appr. 70 bp) and one large (435 bp) amplicon• Artificial sequence, with no similarity to known sequences, and tested against various human and non
human DNAs• You will be able to distinguish: ◦ Successful amplification◦ No DNA ◦ Failed PCR amplification◦ Inhibited DNA◦ Degraded DNA
The Quality Sensor provides additional information to improve quality of results and reduce costs
Investigator ESSplex SE QS Kit
Investigator 24plex QS and GO! Kits
Sample to Insight
Quality Sensor (QS) explained
Normal non-degraded / non-inhibited sample
No DNA
Failed PCR Amplification
QS confirms failed PCR. Re-run will most likely lead to better results.
QS confirms a successful PCR without DNA. Re-run will not lead to better results.
The Quality Sensor confirms that your sample is not degraded or inhibited
Sample to Insight
Quality Sensor (QS) explained
Inhibited DNA sample
Degraded DNA
QS indicates a degraded sample. Re-run will not improve profile.
QS indicates PCR inhibition. An inhibited profile may be rescued by re-running the PCR with a higher sample dilution.
Sample to Insight
Summary
29