INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

by

Susan Jett Lawson RCP, RRT-NPS

RC Educational Consulting Services, Inc. 16781 Vlan Buren Blvd, Suite B, Riverside, CA 92504-5798

(800) 441-LUNG / (877) 367-NURS www.RCECS.com

Page 2: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

BEHAVIORAL OBJECTIVES UPON COMPLETION OF THE READING MATERIAL, THE PRACTITIONER WILL BE ABLE TO: 1. Describe the main characteristics of neonatal ventilation: Positive pressure Time cycling Continuous flow Positive pressure ventilation/volume ventilation differences Tidal volume as it relates to pressure limiting 2. Describe the four common modes of newborn mechanical ventilation: CPAP via endotracheal tube and nasal prongs IMV SIMV 3. Describe the other “newer” modes of infant ventilation. 4. Outline the goals, indications and hazards of CPAP/PEEP and mechanical ventilation. 5. Summarize the most important considerations during management of the infant requiring

mechanical ventilation. 6. State the suggested initial settings for mechanical ventilation of the neonate. 7. Describe the ability to properly monitor the ventilator/neonate interface to include setting/troubleshooting for alarms:

Inspiratory or patient pressure alarm Low PEEP/CPAP alarm Gas pressure failure alarm Ventilator inoperative alarms

8. List and classify the ventilators that combine infant, pediatric and adult ventilation. 9. List and classify the ventilators that are available for infant/pediatric use only.

10. List and describe the calculations that may be helpful in the practice of neonatal and pediatric Respiratory Care. 11. Demonstrate, in a clinical practice exercise, the ability to troubleshoot infant

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 2

Page 3: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

ventilator/patient interface malfunctions. COPYRIGHT © 2001 By RC Educational Consulting Services, Inc. TX 5-360-010 Authored by: Susan Jett Lawson, RCP, RRT-NPS 2001 Revised by: Susan Jett Lawson, RCP, RRT-NPS 2004 Revised by: Susan Jett Lawson, RCP, RRT-NPS 2008 ALL RIGHTS RESERVED This course is for reference and education only. Every effort is made to ensure that the clinical principles, procedures and practices are based on current knowledge and state of the art information from acknowledged authorities, texts and journals. This information is not intended as a substitution for a diagnosis or treatment given in consultation with a qualified health care professional.

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 3

Page 4: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

TABLE OF CONTENTS

PREFACE............................................................................................................................7 INTRODUCTION ...............................................................................................................7

MAIN CHARCTERISTICS OF NEONATAL VENTILATION........................................7 PRESSURE-LIMITED...................................................................................................7 TIME-CYCLED .............................................................................................................7 CONTINUOUS FLOW ..................................................................................................8 COMMON MODES OF NEONATAL MECHANICAL VENTILATION........................8 CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP)........................................8 INDICATIONS FOR CPAP...........................................................................................9 HAZARDS OF CPAP.....................................................................................................9 CONTRAINDICATIONS OF CPAP ...........................................................................10 INTERMITTENT MANDATORY VENILATION.....................................................10 SYNCHRONIZED INTERMITTENT MANDATORY VENTILATION ..................11 NEWER MODES AND TERMS NOW BEING USED IN THE VENTILATION OF NEONATES ....................................................................................11 VOLUME CONTROL (VC) ........................................................................................11 PRESSURE SUPPORT (PS) ........................................................................................11 PRESSURE REGULATED VOLUME CONTROL (PRVC)......................................12 PRESSURE SUPPORT VOLUME GUARANTEE (PSVG).......................................12 OVERVIEW OF NEONATAL MECHANICAL VENTILATION.................................12

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 4

Page 5: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

GOALS OF MECHANICAL VENTILATION.......................................................12 INDICATIONS FOR INITIATING MECHANICAL VENTILATION.................12 HAZARDS OF MECHANICAL VENTILATION.................................................13 MONITORING THE VENTILATOR - PATIENT INTERFACE AND TROUBLESHOOTING ALARMS SYSTEMS................................................................15 INSPIRATORY OR PATIENT PRESSURE ALARM................................................15

LOW PEEP/CPAP ALARM ........................................................................................15 GAS PRESSURE FAILURE ALARM ........................................................................15 VENTILATOR INOPERATIVE ALARMS ................................................................15 A DISCUSSION OF VENTILATOR MANAGEMENT..................................................15 USEFUL CALCULATIONS FOR THE CONVENTIONAL VENTILATOR MANAGEMENT OF THE NEONATE............................................................................16 VENTILATORS................................................................................................................18 EXCLUSIVELY NEONATAL/PEDIATRIC ..............................................................19 SECHRIST MILLENNIUM....................................................................................19 SECHRIST IV 100 B...............................................................................................19 VIP BIRD GOLD ....................................................................................................21 BEAR CUB 750.......................................................................................................21 DRAGER BABYLOG 800 & 8000 PLUS..............................................................22 eVENT INSPIRATION INFANT VENTILATION SYSTEM...............................23

COVIDIEN (TYCO PURITAN BENNETT) 840 ...................................................24 VIASYS AVEA.......................................................................................................24

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 5

Page 6: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

GALILEO GOLD AND CLASSIC.........................................................................25 GE ENGSTROM CARESTATION ........................................................................26 MAQUET SERVO i ................................................................................................26 NEWPORT e 360 ....................................................................................................26 NEWPORT e 1001 ..................................................................................................27 SUMMARY.......................................................................................................................28

TROUBLESHOOTING AT A GLANCE .........................................................................29

CLINICAL SCENARIO QUESTIONS.............................................................................30 CLINICAL SCENARIO QUESTIONS DISCUSSION ....................................................31 SUGGESTED READING AND REFERENCES .............................................................33

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 6

Page 7: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

PREFACE

his course, specifically designed for the Respiratory Therapist just beginning their practice in the Neonatal Intensive Care Unit (NICU), will provide a basic understanding of neonatal ventilation including the main characteristics and the common modes of

ventilation. An overview of current ventilators capable of infant modes is included. Also discussed will be monitoring of the ventilator/neonate interface and setting and troubleshooting for alarms. High frequency ventilation is a separate continuing education course, “High Frequency Ventilation of the Neonate”.

T INTRODUCTION

basic understanding of neonatal ventilation is required in any newborn nursery. Goals of ventilation for these newborns are to provide adequate ventilation and oxygenation (alveolar gas exchange) with minimal compromise to the circulatory system and minimal

damage to the pulmonary tissues. A MAIN CHARACTERISTICS OF NEONATAL VENTILATION Most ventilators designed for infants are pressure-limited, time-cycled and have continuous flow. Within the past few years, other modes have increased in popularity. Let us first define the most common terms: Pressure-limited

he operator limits the maximum pressure exerted against the patient’s airway during the breath. This is called the peak inspiratory pressure or PIP. The level of PIP is one of the determinants of the delivered tidal volume and is changed as needed to alter

ventilation. Due to the influence of PIP on mean airway pressure (MAP), PIP also can affect oxygenation. MAP is the average pressure exerted on the airway and lungs from the beginning of inspiration until the beginning of the next inspiration. Unlike volume-cycled ventilators, pressure-limited ventilators deliver a variable tidal volume depending on the patient’s lung compliance. As lung and thoracic compliance worsens, the pressure limit must be increased to maintain the same tidal volume. Conversely, as compliance improves, the pressure limit must be decreased to avoid excessive ventilation and barotrauma. Tidal volume (VT) as it pertains to mechanical ventilation, is the amount of gas entering the patient’s lungs during the inspiratory phase of ventilation.

T

Time-cycled

nspiratory time (IT) is set to cycle off the inspiratory breath. This time determines how long the gas is in contact with the alveoli for gas exchange. IT and frequency or rate together determines the I: E ratio. There is a delicate balance of the inspiratory to

expiratory time needed to maintain good oxygenation and prevent air trapping. Frequency is the I

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 7

Page 8: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

number of inspirations that will be delivered in one minute. During mechanical ventilation of the infant, generally a 1:2 ratio is desired. Inspiratory times set range from 0.25-0.45 seconds. At this writing, there seems to be no definitive studies as to appropriate inspiratory times for infants, but the trend is to use the shorter times. Longer inspiratory times increase the MAP, which in turn increases the chance of barotrauma. But…increased MAP also increases oxygenation. Continuous Flow

ince the inspiratory efforts and volumes of newborns are so small, it is technically difficult to achieve sensitive, accurate patient sensing mechanisms. A continuous flow (sometimes called bias flow) of gas past the patient’s airway ensures that when the patient needs

inspiratory flow it is already there without having the patient first activate a demand flow system. Flow rates of blended gas are required to achieve the PIP as these ventilators have pneumatically circuits that operate on compressed gas at 40-70 psig. Simply put, without flow, these ventilators will not operate. Flow is usually set at 6-12 LPM depending on neonatologist preference and weight of infant.

S There are calculations for I:E ratio, tidal volume and other parameters that may be indicated to appropriately monitor and manage the patient. COMMON MODES OF NEONATAL MECHANICAL VENTILATION Continuous Positive Airway Pressure (CPAP)

PAP applies continuous positive distending pressure to the airways. CPAP is used during spontaneous breathing to prevent the need for mechanical ventilation. CPAP does not have the same risk of barotrauma as does mechanical ventilation. Also, if nasal prongs

are used instead of an endotracheal tube, the chance of infection decreases. Less invasive equals less danger to the patient. Endotracheal tubes often create a ridge in the soft palate of the neonate when the need for mechanical ventilation is lengthy.

C The main physiological goals of CPAP are to:

• increase the functional residual capacity (FRC)

• increase compliance

• decrease total airway resistance

• decrease respiratory rate It is the actual increase in FRC that allows the other three goals to occur. CPAP physically holds the alveoli and airways open during exhalation thereby increasing the FRC. An increase in FRC

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 8

Page 9: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

leads to improved lung compliance, decreased work of breathing, increased PaO2, and decreased PaCO2. Indications for CPAP:

• Conditions that decrease FRC:

• Pneumonia • Atelectasis • Pulmonary edema • Thoracotomy • Meconium aspiration • Increased mucus • Respiratory Distress Syndrome (RDS) • RDS Type II/Transient Tachypnea of the Newborn (TTN) • Left-to-right shunting

• Airway Collapse:

• Tracheobronchial malacia • Apnea of prematurity (AOP) • Obstructive sleep apnea (OSA)

• Traditional Usage:

• Weaning from mechanical ventilation

• Abnormal Physical Assessment:

• Increased respiratory rate (30-40%) • Retractions • Grunting • Nasal flaring • Cyanosis

• Abnormal Arterial Blood Gas Values:

• PaO2 < 50 Torr at an FIO2 of .60 (with adequate ventilation)

Hazards of CPAP:

• Decreased pulmonary blood flow secondary to compression of pulmonary vessels*

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 9

Page 10: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

• Reduced cardiac output due to decreased venous return to the heart*

• Renal

• decrease in glomerular filtration rate • decrease in sodium excretion • decrease in urine output

• Increase in intracranial pressure (ICP)

• Pneumothorax

• Nasal obstruction (with nasal prongs)

• Gastric distension (orogastric (OG) tube recommended)

• Necrosis or erosion of the nasal septum and nasal deformities from the use of nasal

prongs *Generally associated with high pressures. Contraindications of CPAP:

• Untreated air leaks

• Cardiovascular instability

• Severely apneic patient

• Patients who cannot maintain an adequate spontaneous tidal volume

• Patients with alveolar instability

• Patients with persistent pulmonary hypertension (PPHN)

• Increased intracranial pressure that can cause intraventricular hemorrhage (IVH)

Intermittent Mandatory Ventilation (IMV)

T

he definition of Intermittent Mandatory Ventilation (IMV) is a mode of ventilatory support that allows spontaneous breathing in between mandatory breaths from the ventilator. In between the preset frequency or respiratory rate of the ventilator that

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 10

Page 11: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

delivers a specific pressure, the infant is allowed to breathe warmed, humidified continuous flow gas. The problem with this mode of ventilation is that the timing of the mandatory breaths and the spontaneous breaths are not synchronized. A mandatory breath may be given anywhere in the patient’s respiratory cycle. This asynchrony of breathing may cause the infant to “fight” the ventilator, not to mention the physiologic complications of increased intrathoracic pressures and high airway pressures. This is the reason synchronized intermittent mandatory ventilation (SIMV) was designed. Synchronized Intermittent Mandatory Ventilation (SIMV)

ynchronized Intermittent Mandatory Ventilation is also a mode of ventilation in which mandatory breaths are given at a preset rate and spontaneous respirations are allowed in-between just as in IMV. The difference is that the patient’s spontaneous respirations are

never interrupted. A timing mechanism in the ventilator senses the appropriate time to give the mechanical breath without interfering with the spontaneous breath. This mode of ventilation is beneficial to the spontaneously breathing infant who does not require sedation or paralyzation. Many infant ventilators are not flow or pressure sensitive enough to sense patient triggering at the patient airway, therefore abdominal sensors may be used to sense diaphragmatic and chest wall movement.

S

Remember………there is no need to place an infant who is paralyzed on an SIMV mode. NEWER MODES AND TERMS NOW BEING USED IN THE VENTILATION OF NEONATES VOLUME CONTROL (VC) Historically, volume control has been used for adult and pediatric ventilation and pressure control has been the method of choice for ventilating infants. Recently, though, adults have been receiving more pressure control and infants more volume control. In part, the reason for the emergence of volume control in the infant is our technological advances in ventilators. A stable minute ventilation with a known tidal volume can be achieved, but with any pure volume control, the pressure will vary to maintain the tidal volume set by the clinician. This increases the likelihood of baro and volutrauma. Some ventilators that provide volume control do not have continuous flow; the infant has to “trigger” their flow, possibly increasing their work of breathing. Leaks around endotracheal tubes may be a more significant problem, causing alarming when set tidal volume is not achieved. As you may be aware, infants may “outgrow” their tubes before a new one is placed! PRESSURE SUPPORT (PS) Pressure support is patient-triggered (pressure or flow), pressure-limited and flow-cycled. It provides a constant level of positive pressure during a spontaneous inspiration. Tidal volume is variable. The PS is measured above baseline (ZEEP or PEEP).

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 11

Page 12: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

PRESSURE REGULATED VOLUME CONTROL (PRVC) Tidal volume and rate are preset. The minimum inspiratory pressure necessary to provide the ordered tidal volume will be delivered. PRESSURE SUPPORT VOLUME GUARANTEE (PSVG) Pressure support with a volume guarantee (pressure limited with a set tidal volume) is not a difficult mode to understand. There are two sets of values-set (tidal volume, inspiratory time, inspiratory pressure limit, rate and PEEP) and measured (spontaneous). The set values are used as an “apnea” back up, otherwise, the infant determines their own inspiratory pressure to meet the set tidal volume. As compliance improves, the PIP will decrease. With this mode, endotracheal tube leaks are compensated for and so is lost volume due to tubing compliance. This is a weaning mode. OVERVIEW OF NEONATAL MECHANICAL VENTILATION Goals of Mechanical Ventilation

• Normalization and maintenance of blood gases and acid-base balance (ventilation and oxygenation)

• Prevention of iatrogenic complications:

• Barotrauma and volutrauma

Careful regulation of rate and pressures • Infection

Use sterile technique during suctioning • Sedation

Pharmacologic sedation and analgesia as needed to reduce anxiety and pain.

• Support of the Patient’s Respiratory Needs: Indications for Initiating Mechanical Ventilation

• Increased PaCO2 with a pH less than 7.20-7.25, PaO2 less than 50mmHg despite the use of CPAP and FIO2s .60 or greater than with grunting, flaring, retractions, cyanosis and agitation.

• Neurologic conditions that compromise the drive to breathe:

• Apnea of prematurity (AOP) • Intracranial hemorrhage (IVH or ICH)

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 12

Page 13: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

• Drug depression

• Congenital neuromuscular disorders

• RDS

• Meconium aspiration syndrome (MAS)

• Pneumonia

• Bronchopulmonary dysplasia (BPD)

• Bronchiolitis

• Diaphragmatic hernia

• Sepsis

• Decreased lung volume as seen on chest x-ray

• Persistent Pulmonary Hypertension (PPHN)

• Post-resuscitation

• Congenital Heart Disease

• Shock

• Postoperative with impaired ventilatory function Hazards of Mechanical Ventilation

• Oxygen

• Oxygen toxicity • Hyaline membranes (Infant Respiratory Distress Syndrome) • BPD (Bronchopulmonary dysplasia) • Retinopathy of prematurity (ROP)

• PEEP and CPAP

• Excessive pressures:

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 13

Page 14: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

Hypoventilation from excessive FRC Decreased cardiac output Barotrauma

• Peak Inspiratory Pressure

• Barotrauma: Air leaks such as: Pneumothorax Pneumomediastinum Pneumopericardium Pulmonary interstitial emphysema (PIE)

• Bronchopulmonary dysplasia

• Hyperinflation:

• Hyperventilation • Respiratory alkalosis • Hemodynamic depression

• Respiratory rate

• Respiratory alkalosis • Air leaks • Decreased ventilation-to-perfusion ratios • Increased intrapleural pressure • Decreased pulmonary perfusion • Diminished cardiac output

• General hazards

• Infection

• Hypoxic-ischemic injuries

• Intracranial hemorrhage • Gastric distension • Complications of endotracheal intubation

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 14

Page 15: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

MONITORING THE VENTILATOR - PATIENT INTERFACE AND TROUBLESHOOTING ALARM SYSTEMS Inspiratory or Patient Pressure Alarm This alarm alerts the practitioner to leaks or a patient disconnect. The pressure alarm should be adjusted from between ½ to 2 cmH2O less than the peak airway pressure. Low PEEP/CPAP Alarm This alarm is similar to the low patient pressure alarm except that it alerts the practitioner to loss of PEEP/CPAP pressure. The alarm should be adjusted to 1-2 cmH2O less than the PEEP/CPAP level. Gas Pressure Failure Alarm Many ventilators have alarms that alert the practitioner to the loss of oxygen or air pressure. In the event of a gas supply line loss; cylinders may be temporarily used until the supply line problems are corrected. The clinician may wish to check the oxygen or compressed air line pressures if this alarm is actuated. Ventilator Inoperative Alarms Some ventilators have a ventilator inoperative alarm. These alarms are not operator-adjustable, but alert the practitioner to serious internal problems that will render the ventilator incapable of supporting the patient. These internal problems include electrical power failures, microprocessor failures and internal mechanical failures. In order to obtain more sophisticated alarm capabilities and patient monitoring, additional monitors may be employed. This would be particularly advantageous when using older generation ventilators. A DISCUSSION OF VENTILATOR MANAGEMENT

understand that when first dealing with these “little ones”, the clinician would find it advantageous to have a sort of recipe for ventilator set-up. As you are well aware, there is no simple cookbook method to assist the clinician in this endeavor. I will attempt, though, to

put some common practices on paper.

I The first step in management of your infant, is to set realistic goals for oxygenation and ventilation. The goals depend on several factors, mostly the infant’s disease state. For example, the infant on the ventilator due to meconium aspiration syndrome (MAS)/persistent pulmonary hypertension of the newborn (PPHN) will most likely have the FIO2 kept close to 1.0 in an effort

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 15

Page 16: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

to dilate pulmonary vasculature. On the contrary, the premature infant with respiratory distress syndrome (RDS) will likely be placed on an FIO2 as low as possible to keep their PaO2 over 50-60 Torr to avoid retinopathy of prematurity (ROP). It may be difficult to aim for a normal PaCO2 in the infant who is developing air leak syndromes such as pulmonary interstitial emphysema (PIE). Many clinicians think it is best to tolerate a more acidotic pH and a higher PaCO2 in lieu of the potential of causing a pneumothorax. Remember to consider your source of the blood gas during interpretation. Initial settings for PIP during pressure ventilation may be derived from blood gases drawn while hand-bagging the infant. Adjustments can be made according to the results of the gases and considering the frequency and pressure and FIO2 at which the infant was being ventilated manually. When placing the infant on volume ventilation, the initial settings are usually chosen based on minute ventilation requirements. Volume is calculated at 4-8 ml/kg. Rates are generally 20-30 although ultimately determined by ventilation measurements. Flow rates are generally set between 8 and 12 LPM. Eight for the infants weighing 500 grams and twelve for those appropriate for gestational age (AGA) term infants-somewhere in-between for the others. PEEP ranges from 3 to 5 cmH2O during conventional ventilation based on the FIO2 requirements of the infant. Also, during pulmonary hemorrhage, PEEP may be increased to tamponade the bleed. The parameters set during ventilation of the infant depend on the mode of ventilation being used by the clinician. Literature suggests initial settings of: PIP 15-20 cmH2O PEEP 3-5 cm H2O Frequency 40 Flow rate 6-8 LPM I-time of 0.5 I:E 1:1.5 or 1:2 USEFUL CALCULATIONS FOR THE CONVENTIONAL VENTILATOR MANAGEMENT OF THE NEONATE

Delivered Flow Rate Flow rate = set minute volume I-time % Expiratory Time Example: What expiratory time is needed to achieve a rate of 30 with an I-time of 0.4 seconds?

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 16

Page 17: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

Total breath time = seconds per min/desired breath rate Total breath time = 60/30 Total breath time = 2.0 seconds Expiratory time = total breath time (2.0 seconds) – Inspiratory time (0.4 seconds) Expiratory time = 2.0 – 0.4 = 1.6

Calculation of FIO2 When O2 flow and airflow are known, the FIO2 may be calculated. FIO2 = O2 flow + (0.21 X air flow) Total flow Example: What is the FIO2 with the following flow rates? O2 = 6 Air = 4 FIO2 = 6 + (0.21 X 4) 10 FIO2 = 6 + 0.84 10 FIO2 = 6.84 10 FIO2 = .684 Or O2 % of 68 Tidal Volume During Pressure-limited Ventilation VT = I-time X flow rate Example: Inspiratory time = 0.40 seconds Flow rate = 8 LPM (convert to ml)

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 17

Page 18: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

8 LPM = 8,000 ml/min. or 133.3 ml/sec. (8,000 / 60) VT = 0.4 X 133.3 VT = 53.32 ml. Calculation of I:E / Determination of I-time Example: Ventilator rate = 40 breaths/minute Desired I:E = 1: 2 Respiratory cycle time = 60 seconds Set ventilator rate Respiratory cycle time = 60 seconds 40 Respiratory cycle time = 1.5 seconds Add the 2 numbers of the desired I:E (1 + 2 = 3) Divide into the respiratory cycle time (1.5 ÷ 3) I time = 0.5 seconds VENTILATORS Newborn or infant ventilators like adult mechanical ventilators, are for support of the respiratory system using positive-pressure ventilation. VENTILATORS AVAILABLE IN THE U.S. FOR INFANTS Infant Infant/Neonatal Mode Drager Babylog 800 GE Engstrom Carestation Sechrist Millennium Maquet Servo i eVent Medical Ltd. Inspriation® Infant Ventilation System

Newport NMI e360

VIASYS Bear Cub 750 Hamilton-G5 VIASYS VIP Gold VIASYS AVEA eVent Medical Ltd. Inspiration™ Ventilator System

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 18

Page 19: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

EXCLUSIVELY NEONATAL/PEDIATRIC

Millennium

Compliments of Sechrist Industries

The Sechrist Millennium is for use with infant and pediatric patients weighing less than 50 kg. It uses a triggering flow sensor termed SmartSync™

IV 100B

Compliments of Sechrist Industries

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 19

Page 20: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

Basic Mechanisms

• Pneumatically powered

• Electronically and fluidic valve controlled

• The gas delivered to the patient does not pass through the ventilator. The 50psi source gases are connected directly to the oxygen/air blender located on the side of the ventilator. The blender reduces the pressure of the gases, mixes them and then directs the flow to an attached flowmeter, calibrated from 2 to 32 LPM. The gas flow enters the patient circuit from the flowmeter.

• All functions of ventilation (rate, inspiratory time, PIP and PEEP) are handled by the

expiratory valve that is controlled by the ventilator settings. A small portion of the inlet oxygen is diverted internally to the ventilator to power the exhalation valve.

• The Sechrist utilizes two types of fluidic valves, a backpressure switch and an

OR/NOR gate. A complex discussion of fluidic gates and OR/NOR valves is not necessary in this module.

Unique Features

• Sine-wave or square wave flow patterns

• Sine wave by turning a small Allen screw located just above the expiratory block. As the screw is tightened, the normal pathway of flow to the expiratory diaphragm is cut off, forcing the flow of gas from the fluidic valve to pass through a restricted orifice before reaching the expiratory diaphragm slowing the flow and creating a rounded flow.

• Square-wave is preset unless the Allen screw is adjusted for a sine wave.

• Manual breath button is inline with the flow from the backpressure switch. If button is pressed for a manual breath, the flow is occluded and the inspiratory phase is started. Inspiration lasts as long as the button is held in so carefully monitor the I: E ratio when using this feature.

• Digital display of:

• I: E ratio • Rate • Inspiratory time • Expiratory time

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 20

Page 21: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

• Expiratory timer is adjusted by reading the rate display when determining rate.

VIP BIRD GOLD Compliments of: Viasys Critical Care

V.I.P. Bird Gold is microprocessor controlled and pneumatically powered. This ventilator has the ability to ventilate neonate to 30 kg. patients. It has a comprehensive graphics monitor with waveforms and loops. The V.I.P. performs mechanical calculations such as compliance and resistance. Modes offered are assist/control and SIMV/CPAP in both volume-cycled and time-cycled settings.

BEAR CUB 750 Compliments of: Viasys Critical Care

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 21

Page 22: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

The Bear Cub 750 ventilator has the capability of A/C, SIMV, IMV and CPAP modes. It also has integrated synchronized ventilation and tidal volume monitoring. It provides proximal airway flow triggering. The heated-wire flow sensor provides breath-to-breath measurement of spontaneous and mechanical inspiratory and expiratory tidal volume and percentage of tube leak. A graphics monitor is available to view measurements, waveforms and loops.

Drager Babylog 800 Compliments of: Drager

Drager Babylog 8000 Plus Compliments of: Drager

DRAGER BABYLOG 800 & 8000 Plus

Compliments of: Drager This ventilator is microprocessor controlled, pneumatically powered, time-cycled, pressure-limited, constant flow generator designed for infants up to 10 kilograms (22 pounds). It provides A/C, SIMV, PSV, VG, VIVE and CPAP modes of ventilation. It is capable of both volume and pressure ventilation. A flow sensor at the patient wye measures tidal volume and senses airflow

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 22

Page 23: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

initiated by the infant allowing triggering of the ventilator cycle. The sensor is able to compensate for small endotracheal tube leaks. The 8000 Plus model (not available in the United States) includes high frequency ventilation and is designed for patients up to 20 kg.

eVent MEDICAL has both an Inspiration™ Infant Ventilation System and an adult, pediatric and infant system; Inspiration™ Ventilator System. Compliments of: eVent Medical

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 23

Page 24: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

AGE RANGE FROM INFANT THROUGH ADULT

COVIDIEN (TYCO PURITAN BENNETT) 840 Compliments of: Tyco Puritan Bennett

This ventilator features dual microprocessor electronics and high-performance pneumatics. The 840 is designed to offer adult and pediatric modes of ventilation with an optional NeoMode. NeoMode software allows the 840 to ventilate patients as small as 500 grams without requiring a flow sensor at the patient’s airway.

AVEA Compliments of: Viasys Critical Care

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 24

Page 25: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

It is a fourth generation ventilator, servo-controlled, software-driven ventilator. It has a dynamic range of breathing gas delivery that provides for neonatal through adult patients. The AVEA may be configured as a conventional ventilator or non-invasive positive pressure ventilator. Capabilities include precise delivery of heliox, artificial airway compensation, leak compensation, circuit compliance compensation, and independent lung ventilation. Neonatal ventilation modes available are volume and pressure assist/control, volume and pressure SIMV, time-cycled, pressure limited assist/control and SIMV and CPAP/PSV.

G5 Compliments of: Hamilton Medical

The Galileo Gold and Classic can ventilate all patients, from neonates to adults. The newest offering from Hamilton is the G5 ventilator. Hamilton offers closed-loop controlled mode AStV (adaptive tidal volume support - not available in the neonate modes), DuoPAP (BiPAP), DuoPAP+ (APRV), tube resistance compensation, the P/V Tool (to find best PEEP) and noninvasive ventilation. The monitoring package includes 26 parameters, lung mechanics with waveform, loop and trend displays on a color screen. It has a built-in nebulizer and internal battery back up.

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 25

Page 26: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

GE Engstrom Carestation Compliments of: GE

GE Healthcare, Madison, Wis, provides the Engström Carestation , a critical care respiratory carestation that offers integration throughout the care process. From the ICU to the step-down unit, users have the ability to integrate ventilation with monitoring modules capable of measuring

®

advanced parameters. The Carestation offers ETCO2, energy expenditure monitoring, metabolics, gas exchange calculations, volume guarantee, (SBT) spontaneous breathing trial, SpiroDynamics, FRC, PEEP, Lung INview.

.

Maquet Servo i Compliments of: Maquet

Maquet Inc, Bridgewater, NJ, provides the Servo i ventilator platform for all patient types, including neonatal and pediatric, as well as invasive and noninvasive ventilation, and support of inter-hospital transport. It offers a comprehensive array of modalities, such as volume support, PRVC, BiVent, Automode®, and Flow-adaptive VC. The Open Lung Tool® provides physicians with parameters to support alveolar recruitment procedures. The Servo i also provides information about end inspiratory pressure (EIP) and PEEP, tidal volume, as well as dynamic

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 26

Page 27: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

compliance. Mainstream CO2 monitoring and two nebulizers are available.

Newport e360 Compliments of: Newport NMI

The Newport e360 is Newport Medical Instruments newest ventilator. The Newport e360 ventilator builds on the design and features of the e500 ventilator. Simple to use, the e360 provides comprehensive mode selections, with graphics and extensive monitoring built into a single compact package. The e360 can easily transition from invasive to noninvasive ventilation for adult, pediatric, or infant patients. It offers weaning tools. TCPL modes are not available at this writing. NEWPORT E 1001 The Newport E 1001 is a compact universal transport ventilator for infant, pediatrics and adults. It offers assist/control, SIMV and spontaneous modes. An air/oxygen blender and four hour battery may be added for transport.

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 27

Page 28: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

SUMMARY The goals of ventilation for infants are focused on providing adequate ventilation and oxygenation with minimal compromise to the circulatory and other systems and minimal damage to the pulmonary tissues. Most infants are ventilated with pressure-limited, time-cycled, continuous flow ventilators. The most common modes of neonatal mechanical ventilation are CPAP, pressure-IMV and SIMV. Alternative modes are becoming more popular. These include, high frequency, volume control, pressure support, pressure regulated volume control (PRVC) and pressure support volume guarantee (PSVG). The three main goals of mechanical ventilation are normalization and maintenance of blood gases, prevention of iatrogenic complications and support of the patient’s respiratory needs. The list of indications for initiating mechanical ventilation. Hazards of mechanical ventilation generally stem from oxygen administration, excessive pressures, bronchopulmonary dysplasia, hyperinflation, hyperventilation, hypoxic-ischemic injuries and infection. Essential alarms during mechanical ventilation include, but are by no means limited to peak inspiratory pressure, gas pressure failure and ventilator inoperative. Other useful alarms and indicators are low pressure, low expired tidal volume, low expired minute volume, low PEEP, high respiratory rate, low respiratory rate; the list is almost endless depending on the clinician’s perspective and the mode of ventilation being employed. Literature suggests initial ventilator settings for the neonate, but there are numerous variables to consider. The clinician may wish to use a few simple calculations to assist them in the management of their neonatal patient. The most useful (depending on the ventilator used) are delivered flow rate, expiratory time, FIO2, tidal volume and I:E. Ventilators that are specific to neonatal and pediatric ventilation are the Infant Star, Sechrist, Babybird, V.I.P. Bird, Bear BP 200, Bear Cub, Sensormedics 3-10, Newport Breeze, Drager Babylog and Bio-Med MVP-10. Ventilators designed to ventilate infants through adults include the Siemens 300A, Puritan-Bennett 840, Viasys Avea, Hamilton Galileo and Newport E 1001. With all these ventilators, the clinician’s ability to troubleshoot the patient-ventilator interface is essential.

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 28

Page 29: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

TROUBLESHOOTING AT A GLANCE PROBLEM POSSIBLE SOLUTION (S) Whistling or hissing sound from ventilator Gas leak Check psi connections Blender alarm or gas pressure failure PSI of air and/or oxygen reduced to less than 45 psi Check source gas psi. Call Engineering. Ventilator inoperative alarm Serious internal ventilator Malfunction Hand-bag patient Inspiratory or patient pressure alarm Leak or patient disconnect Reconnect patient-correct leak Low PEEP/CPAP alarm Loss of PEEP/CPAP Correct leak Sudden drop in CPAP level to zero Disconnection at patient or in breathing circuit. Check all

connections and reassemble break. CPAP level drops more than 2cmH2O during inspiration Inadequate flow Increase flow During CPAP, patient demonstrates an increased use of accessory muscles or work of breathing (WOB) Inadequate flow Increase flow Inadvertently high CPAP/pt. having difficulty exhaling Too high of flow Decrease flow Increased CPAP reading Plugged nasal prong(s) Suction patient and clean prongs Drop in oxygenation, increase in Respiratory rate, and/or retractions during CPAP Airway obstruction Suction patient and clean prongs

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 29

Page 30: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

CLINICAL SCENARIO QUESTIONS 1. Baby Boy Smith, 24 week, appropriate for gestational age (AGA) infant, is on the Star Ventilator in the demand IMV mode set at 30 breaths per minute, flow of 7 liters, I-time of .40, PIP of 15cmH2O and PEEP of 3cmH2O. The high peak pressure alarm is activated several times. What is your appropriate response? 2. Infant Girl Jones is being weaning from mechanical ventilation. She has been placed on the CPAP mode at a PEEP of 5 cmH20 via nasal CPAP prongs. Although expected to do well on this modality, Baby Girl Jones is showing signs of irritation and cannot fully exhale. The manometer intermittently reads much higher than the PEEP you originally set. What would you do? 3. You just received the message from engineering that the piped in compressed air will have to be turned-off to repair the compressor. There is no back-up source. How would this affect your infants on ventilators and what course of action should you take? 4. You are in a 10 bed NICU with 2 infants on ventilators. One of the ventilators begins alarming. The ECG on the infant shows a cardiac rate of 50 beats per minute. What would be your course of action?

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 30

Page 31: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

5. Upon assessment of Baby Girl Flowers, you note that unlike your previous observations, she has increased work of breathing demonstrated by grunting, retractions and use of accessory muscles of respiration. You placed her on a CPAP of 6 via nasal prongs earlier today and now the digital reading indicates a drop of greater than 2 cmH20 during inspiration. What is the appropriate remedy? 6. You have just assisted the RN in turning and positioning Baby Boy Bunting who is intubated with a 3.5 ETT and is being ventilated via a Sechrist ventilator. You have reconnected the adapter to the ETT but the inspiratory or patient pressure alarm continues to sound. What should you do? 7. The gas source alarm is sounding on the ventilator of Baby Dickson. What does this indicate and what would be your course of action? CLINICAL SCENARIO QUESTIONS DISCUSSION 1. The increased peak inspiratory pressure (PIP) alarm was activated. On an Infant Star, an

AO1 message for the high pressure notifies the clinician that the PIP was detected 5 cm H2O above the PIP set point. Assess the infant, including observation and breath sounds. This may indicate an obstructed tube. This level of increased pressure usually indicates activity, the need for suctioning, or if rhythmical, that the infant has the hiccoughs.

2. Assess the infant, including observation and breath sounds. The cause of this problem is

generally too high a flow set. The suggested starting point for setting the continuous flow is from 8-12 LPM. 8 for the smaller infants and 12 for the term infants. 10 is for those in-between. For example, if you do not set a high enough flow on the Sechrist, you may not be able to obtain the indicated pressure. (Answer: decrease flow.)

3. Most infant ventilators are pneumatically controlled, thereby making us dependent on a gas

source. The infants will be on an FIO2 of 1.0 when the compressed gas source is turned off. Your facility may have compressed air cylinders or air compressors to temporarily replace the piped-in air gas source.

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 31

Page 32: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

4. Assess the infant. Reconnect the ventilator and possibly increase rate, PIP and FIO2 until the infant is stable, then wean. Some facilities prefer that the infant is hand-bagged. If the ventilator shows a ventilator inoperative alarm, there is a serious internal ventilator malfunction and the patient must be removed from the ventilator and hand-bagged until a functional replacement is obtained.

5. If, during CPAP, the infant demonstrates increased work of breathing and/or the CPAP level drops to greater than 2 cmH20 during the infant’s inspiration, the flow is generally inadequate. Correction: increase flow. 6. Look for another disconnection! Then proceed from there. The ventilator worked fine before the re-positioning so the ventilator is probably still fine. The most likely problem at this point is another disconnection from manipulating the tubing during the repositioning. 7. The gas source alarm indicates that the PSI of the air or oxygen sources has decreased to less than 45 PSI (exact number may be dependent on the ventilator). Check for loose connections and that your 50-PSI tubing is in place. If readily available, check the PSI readings near the oxygen and air mains for your area. Call engineering. But…most of all ensure the safety of your infant. Hand-bag until the situation is resolved.

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 32

Page 33: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

SUGGESTED READING AND REFERENCES American Association for Respiratory Care Clinical Practice Guideline “Neonatal”

Time-triggered, pressure-limited, time-cycled mechanical ventilation.” Resp Care 39(8): 808-16 (1994).

American Association for Respiratory Care Clinical Practice Guideline “Application of continuous positive airway pressure to neonates via nasal prongs or nasopharyngeal tube.” Resp Care 39(8): 817-23 (1994). American Association for Respiratory Care Clinical Practice Guideline “Patient-Ventilator System Checks” www.rcjournal.com/cpgs/mvsccpg.html Argavest, Inc. 6813Hobson Valley Drive, Suite 105 Woolridge, IL 60517 Bellettato, M. Assisted Ventilation of the Newborn, emedicine.com Last Updated June, 2006. Branson, R., Hess, D., Chatburn, R. (1999). Respiratory Care Equipment. (2nd Ed). Philadelphia: Lippincott, Williams and Wilkins. Burton, G.G., et al. (1997) Respiratory Care: A Guide to Clinical Practice. (4th Ed). Philadelphia: Lippincott-Raven. Koff, P., Eitzman, D., Neu, J. (1993). Neonatal and Pediatric Respiratory Care. St. Louis: Mosby-Year Book, Inc. General Biomedical Service, Inc. 1900 25th St. Kenner, LA 70062 (800) 558-9449 Hamilton Medical, AG, Via Nova, CH-7403 Rhazuns, Switzerland. MacIntyre, N., Branson, R. (2001) Mechanical Ventilation. W.B. Saunders Co. Scanlan, C.L., Wilkins, R., Stoller, J. (1999). Egan’s Fundamentals of Respiratory Care. (7TH Ed). St. Louis: Mosby. Sills, J. (1995). Respiratory Care Registry Guide. St. Louis: Mosby. University of Washington, Children’s Hospital and Regional Medical Center, NICU-WEB. Mechanical Ventilators, revised 2002. Viasys Healthcare Critical Care, 1100 Bird Center Drive, Palm Springs, CA 92262-8066 www.viasyshealthcare.com

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 33

Page 34: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

Whitaker, K. (1997). Comprehensive Perinatal and Pediatric Respiratory Care. (2nd Ed). Albany: Delmar, ITP. White, G. (1998). Basic Clinical Lab Competencies for Respiratory Care. (3rd Ed). Albany: Delmar, ITP.

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 34

Page 35: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

POST TEST DIRECTIONS: IF COURSE WAS MAILED TO YOU, CIRCLE THE MOST CORRECT ANSWERS ON THE ANSWER SHEET PROVIDED AND RETURN TO: RCECS, 16781 VAN BUREN BLVD, SUITE B, RIVERSIDE, CA 92504-5798 OR FAX TO: (951) 789-8861. IF YOU ELECTED ONLINE DELIVERY, COMPLETE THE TEST ONLINE – PLEASE DO NOT MAIL OR FAX BACK. 1. Hazards of CPAP include:

I. Pneumothorax II. Gastric distension III. Decreased cardiac output IV. Increase in ICP

a. I, II, III b. II, III, IV c. II & III only

d. All of the above 2. Goals of mechanical ventilation include:

I. Careful regulation of ventilating pressures II. Maintenance of acceptable arterial blood gases III. Prevention of iatrogenic complications IV. Support of the neonates respiratory needs

a. I, II, III b. II, III, IV c. II & III only d. All of the above

3. Demand flow on the Infant Star ventilator establishes a 4 L/min continuous flow to the ventilator circuit.

a. True b. False

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 35

Page 36: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

4. The ventilator inoperative alarms sounds on your patient. You should:

a. Check for leaks in the system b. Suction the infant c. Hand-bag the infant

d. Check PSI connections 5. The main physiological goals of CPAP are:

I. Increase compliance II. Decrease airway resistance III. Increase inspiratory rate IV. Increase FRC

a. I, II, III b. II, III, IV c. I, II, IV d. All of the above

6. Indications for mechanical ventilation of the neonate include:

I. Pneumomediastinum II. Meconium aspiration syndrome III. PIE IV. RDS

a. I, II, III b. II, III, IV c. I & III only d. II & IV only

7. Ventilator inoperative alarms are operator-adjustable.

a. True b. False 8. The Sechrist ventilator is controlled:

a. Electronically and pneumatically b. With a fluidic valve c. By a microprocessor d. Electronically and with a fluidic valve

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 36

Page 37: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

9. The maximum pressure exerted against the infant’s airway during each breath is defined as:

a. MAP b. PIP c. PCWP d. VT

10. Indications for the use of CPAP on the neonate include:

I. Non-severe apnea of prematurity II. An increase of respiratory rate by 30% III. Persistent pulmonary hypertension IV. Cardiovascular instability

a. I, II, III b. II, III, IV c. I & II only

d. All of the above 11. Gas pressure alarms alert the clinician to air and oxygen pressures in excess of 60 PSI.

a. True b. False

12. Hazards of mechanical ventilation in the neonate include:

I. Congenital diaphragmatic hernia II. Air leak syndromes III. IVH IV. BPD

a. II & IV b. I & III c. II, III, IV d. All of the above

13. The AVEA is capable of ventilating neonates, pediatrics and adult patients.

a. True b. False

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 37

Page 38: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

14. Most neonatal ventilators are classified as:

a. Electrically controlled, pressure-cycled, volume-limited b. Pneumatically controlled, pressure-limited, time-cycled c. Microprocessor controlled, time-limited, volume-controlled

d. None of the above 15. The inspiratory time determines how long the gas is in contact with the alveoli for gas exchange.

a. True b. False

16. Which of the following ventilators IS NOT capable of adult ventilation?

a. VIP Gold b. Carestation c. G5 d. 840

17. Which of the following ventilators are used primarily for transport of the neonate?

a. Sechrist Millennium b. Newport E1001 c. Maquet Servo i d. VIASYS VIP Gold

18. A mode in which the tidal volume and rate are present and the minimum pressure necessary to provide the tidal volume will be delivered is:

a. Volume ventilation b. Pressure support volume guarantee c. Pressure regulated volume control d. Pressure support

19. The suggested “initial” settings for neonatal ventilation DO NOT include:

a. PIP 25-30 b. PEEP 3-5 c. Frequency 40 d. I:E of 1:1.5 or 1:2

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 38

Page 39: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

20. The first step in management of your infant (other than initial resuscitation measures) is to:

a. Interpret the chest X-ray for hyper expansion b. Set realistic goals for oxygenation and ventilation c. Place the infant on a transcutaneous monitor d. Assess the infant for congenital anomalies

SL: Test Version C

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 39

Page 40: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

ANSWER SHEET NAME____________________________________ STATE LIC #_______________________ ADDRESS_________________________________ AARC# (if applic.)___________________ DIRECTIONS: (REFER TO THE TEXT IF NECESSARY – PASSING SCORE FOR CE CREDIT IS 70%). IF COURSE WAS MAILED TO YOU, CIRCLE THE MOST CORRECT ANSWERS AND RETURN TO: RCECS, 16781 VAN BUREN BLVD, SUITE B, RIVERSIDE, CA 92504-5798 OR FAX TO: (951) 789-8861. IF YOU ELECTED ONLINE DELIVERY, COMPLETE THE TEST ONLINE – PLEASE DO NOT MAIL OR FAX BACK. 1. a b c d 16. a b c d 2. a b c d 17. a b c d 3. a b 18. a b c d 4. a b c d 19. a b c d 5. a b c d 20. a b c d 6. a b c d 7. a b 8. a b c d 9. a b c d 10. a b c d 11. a b 12. a b c d 13. a b 14. a b c d 15. a b SL: Test Version C

This material is copyrighted by RC Educational Consulting Services, Inc. Unauthorized duplication is prohibited by law. 40

Page 41: INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

INTRODUCTION TO MECHANICAL VENTILATION OF THE NEONATE

EVALUATION FORM NAME:____________________________________________ DATE:______________ AARC # (if applic.)________________________ STATE LICENSE #:______________ RC Educational Consulting Services, Inc. wishes to provide our clients with the highest quality CE materials possible. Your honest feedback helps us to continually improve our courses and meet CE regulations in many states. Please complete this form and return/submit it with your answer sheet. Thank you. YES NO Were the objectives of the course met? Was the material clear and understandable? Was the material well-organized? Was the material relevant to your job? Did you learn something new? Was the material interesting? Were the illustrations, if any, helpful? Would you recommend this course to a friend? What was the most valuable portion of the material? ________________________________________________________________________ What was the least valuable portion of the material? ________________________________________________________________________ Suggestions for future courses: ______________________________________________ Comments: ______________________________________________________________ What is your specialty area?______________________________ Credentials?________ How did you hear about RCECS?____________________________________________