Analyzer
JAS SimDis AnalyzerSimulated Distillation
JAS SimDis AnalyzerSimulated Distillation
JAS SimDis Analyzer
Simulated Distillation (SimDis) is a gas chromatographic method to determine the true boiling point distribution curves of petroleum raw materials.
JAS offers a turnkey solution for simulated distillation up to C120. The key components of the JAS SimDis Analyzer are the JAS UNIS HT SimDis inlet and the unique JAS SimDis Software.
The JAS SimDis Analyzer is designed to meet ASTM methods D2887, D3710 (with packed column inlet and detector adaptors), D5399, D6352 and D7096, D7169, D7213, D7398 and D7500.
Features & Benefits
• Available with Agilent 7890 GCs and as retrofits for 6890 GCs• Dedicated JAS UNIS HT SimDis inlet → max. temperature of 0 °C → low thermal mass for fast heating and cooling
• User-friendly and easy to understand JAS SimDis Software
JAS UNIS HT SimDis Inlet
With its unique temperature programmable heating cartridge, its low thermal mass and a heating rate up to 720 °C/min the
demanding high-temperature analyses.
With an upper operating temperature of 45minimizing any carry-over and memory effects from the inlet.
• SL and PTV mode• Low thermal mass • Long heating cartridge with heated
column nut to avoid cold spots• Patented SimDis liner• SimDis up to C120• External controller
JAS SimDis Software
The JAS SimDis Software includes all relevant functions for comprehensive analyses and offers a user-friendly and easy to understand graphical user interface.
• Calibration• Blank run substraction• Reference sample• Chromatogram with IBP and FBP• Boiling point distribution curve• Boiling point % result• Standard cut point• Custom cut point• Built-in wt% to vol% conversion
The JAS SimDis Analyzer includes
• Agilent 7890 GC with FID• JAS UNIS HT SimDis inlet• JAS SimDis Software• Application
Overview of ASTM Methods
D2887 Petroleum products and fractions having a final boiling point of 538 °C (1000 °F) or lower and a boiling range greater than 55.5 °C (100 °F).
D3710 Petroleum products and fractions with a final boiling point of 260 °C (500 °F) or lower (Packed column)
D5399 Hydrocarbon solventshaving a minimum initial boiling point of 37 °C (99 °F), a maximum final boiling point of 285 °C (545 °F), and a boiling range of 5 to 150 °C (9 to 270 °F)
D6352 Petroleum distillatefractions having an initial boiling point greater than 174 °C (345 °F) and a final boiling point of less than 700 °C (1292 °F, C90)
D7096 Gasoline and liquid gasoline blending components with a final boiling point of 280 °C (536 °F) or lower (Capillary column)
D7169 Crude oils and atmospheric and vacuum residues through 720 ºC (1328 °F, C100)
D7213 Petroleum distillates having an initial boiling point greater than 100 °C (212 °F) and a final boiling point less than 615 °C (1139 °F; C60)
D7398 Blends of diesel and biodiesel (B1 through B100) having an initial boiling point greater than 100 °C (212 °F) and a final boiling point less than 615 °C (1139 °F)
D7500 Distillates and lubricating base oils having initial boiling point above 100 ºC (212 °F) and final boiling point below 735 ºC (1355 °F, C110)
JAS SimDis AnalyzerSimulated Distillation
UNIS HT SimDis
JAS SimDis Software
Carl-Zeiss-Str. 49
47445 Moers
Germany
Phone: +49 2841 9871 100
Fax: +49 2841 9871 222
e-Mail: [email protected]
Internet: www.jas.de
joint analytical systems GmbH
Copyright © 20
joint analytical systems GmbH
All rights reserved
About joint analytical systems
Since 1995 JAS has been a Premier Solution Partner and
Value Added Reseller of Agilent Technologies. We are
an innovative-driven organization that offers customized
solutions for GC, GCxGC, µGC, GC-AED, GC-MS, GC-QQQ,
LC, LC-MS, LC-QQQ and Q-TOF LC-MS applications.
JAS serves key industries such as
• Chemical• Environmental• Food and Flavor• Forensic• Clinical• Petrochemical• HPI
JAS Products for GC
• Atomic Emission Detector• CryoTrap• Customized Valving System• EzPrep - Preparative Fraction Collector• Olfactometer• UNIS Inlets