Upload
ernesto-arzabala
View
6
Download
1
Embed Size (px)
DESCRIPTION
Thermistor Calibration
Citation preview
Laboratorio de Instrumentacin
Mecatrnica
ITESM Chihuahua
Thermistor Calibration
Laboratorio de Instrumentacin Mecatrnica
2
Calibration equipment
3
The thermistor and a reference thermometer are placed in
an insulated mug. Different mixtures of hot and cold water
are used to set the water temperature inside the mug.
DMM Leads
(Ohmeter mode)
(or Thermocouple lead
of the DMM,
temperature mode )
Calibration equipment
Thermistor probe ( 10k
thermistor) Thermocouple probe
4
Measuring resistance
5
You will need to use the bench multimeter to measure the
resistance of the thermistor. Make sure that the probes are
plugged into the high and low jacks labeled for voltage and
resistance, not current. By convention, black probes are
used for the ground and red probes for measured voltage.
Although this polarity does not matter for measuring the
thermistor resistance, it is good to get in the habit of
connecting the probes up correctly.
Measuring resistance
6
You will need to select the resistance measurement (the
button labeled ) and automatic continuous measurement.
If the MAN for manual or TRIG for trigger lights are on
on the display, try fiddling with the buttons (shift-trigger
may help). Better would be to read the DMM Users Guide.
Depending on what sort of probe is connected to the
multimeter, you may be able to connect them directly to
the thermistor wires, or you may need to use alligator clip
leads to connect them.
Measuring resistance
7
For making precise measurements, it helps to subtract off
the measurement of the wires connecting up to the device
being measured, by shorting them together and recording
that measurement.
There is another technique that does an even better job of
compensating for wiring resistance: using 4-wire
measurement and Kelvin clips. For an explanation, see
http://www.allaboutcircuits.com/vol_1/chpt_8/9.html
Measuring Resistance
8
Although the DDM is capable of 4-wire measurement, we
do not have the Kelvin clip probes for simultaneously
measuring current and voltage, so this technique is not
available to us. In any case, the resistances were looking at
on the thermistors are large enough compared to the wire
resistance that we dont really need the extra accuracy
obtainable with 4-wire measurement.
Measuring Resistance
9
You will want to measure the temperature and resistance of
the thermistor at the same time (since the uninsulated
water baths will equilibrate to room temperature fairly
quickly). Having one person hold and read the temperature,
while the other records the temperature and resistance
makes the recording easier.
You need to both record the measurements in a lab
notebook and type them directly into a computer file
(Excel spreadsheet).
Calibration procedure
10
Once the equipment is assembled and ready to use, perform the following steps to obtain the calibration data.
1. Fill the coffee mug with water at a desired temperature. Mix hot and cold water as necessary.
2. Insert the reference thermometer and the thermistor probe into the water in the mug.
3. Wait for the thermistor probe signal and the reading of the reference temperature to stabilize.
4. Record the temperature and the thermistor resistance. (Notebook and Excel)
5. Return to step 1until you record the thermistor resistance for the range of 0 to 100 celsius, going up and down.
If your group has two mugs, you can be refilling one mug while you wait for the reference thermometer and thermistor probe to come into equilibrium with the water in the other mug.
Calibration procedure
11
Temperature (oC) Resistance
(temperature going up)
Resistance
(temperature going
down)
0
10
20
30
40
100
Thermistor T-R Curve
12
NTC (Negative Temperature Coefficient) thermistors
have the advantage of a very high sensitivity to
temperature changes, but the disadvantage of an
aggressively nonlinear characteristic.
The following curve shows the resistance of a typical NT
C thermistor device over a temperature range from 0 to 5
0C.
NTC thermistor T-R curve
13
Steinhart-Hart Equation
14
Thermistors T R curve can be described in equations. The most commonly used form is the Steinhart-Hart
Equation shown below:
1
= + + ()3
Where Rt the thermistor resistance at temperature T (Kelvin);
A, B, C the thermistors constants. Manufacturers may provide typical
values of the A, B, and C coefficients, or you can calibrate these values
for better accuracy.
NTC thermistor T-R curve
15
From the previous curve, resistance value changes from a
ratio of 3.0 to a ratio of 0.5 within 0~50C.
The change is most rapid at low temperatures, giving
great resolution for determining the corresponding tem
perature values there.
At higher temperature, resistance changes relatively less
with temperature and the measurement resolution will
be relatively poor.
Calibration procedure
16
It is not hard to calibrate your own response curves,
if you have an accurate temperature measurement
standard. For your calibration , three points need to
be selected, two close to the ends of the operating
range and one near the center.
Insert the three pair of resistance values and
temperatures to the equation to form three equations.
Solve the equations for the unknowns (A,B,C).
http://www.thinksrs.com/downloads/programs/Therm%20Calc/NTCCalibrator/NTCcalculator.htm
Calibration procedure
17
Compare your A,B, and C coefficients with the
manufacturers coefficients.
Compare the experimental T-R curve of the thermistor
agianst the T-R curve of the manufacturer.
Compare the experimental T-R curve of the thermistor
against the T-R curve with your coefficients.
Repeat everything for the next circuit
18
Additional Notes
19
http://www.digikey.com.mx/Web%20Export/Supplier%20Content/api-technologies-1171/pdf/api-ntc-engineering-notes.pdf?redirected=1
http://www.omega.com/Temperature/pdf/44000_THERMIS_ELEMENTS.pdf
20