Upload
others
View
11
Download
0
Embed Size (px)
Citation preview
Laryngeal mask airway
(LMA-ProSeal) malfunction causing acute airway obstruction Binu Puthur SIMON and Syed Harun HABIBULLAH
Department of Anaesthesiology, RIPAS Hospital, Brunei Darussalam
ABSTRACT
The Laryngeal Mask Airway ProSealTM (LMA-ProSealTM; Laryngeal Mask Company Limited) is a reusable
supraglottic airway device developed to enhance supraglottic airway protection and extend the benefits
of the classic LMA (Laryngeal Mask Airway) to greater number of patients. Added features include an
additional drain tube to channel fluid away from the airway and a tighter seal against the glottic open-
ing with no increase in mucosal pressure. Clinicians have extended the use of the LMA-ProSealTM inside
and outside the operating theatre including use for difficult airway management and airway rescue.
However, even these new devices have their limitations. We report an unforeseen acute airway ob-
struction caused by LMA-ProSealTM malfunction during ophthalmic surgery. The cuff of the device was
deformed with herniation to one side upon insufflation of the balloon.
Keywords: Laryngeal mask airway, malfunction, complications, LMA-ProSeal
INTRODUCTION
Laryngeal masks are used broadly for elective
and emergency airway management and are
an essential part of the American and Europe-
an difficult airway management algorithm. 1-3
Due to their wide use, noticeable complica-
tions and side effects have been reported
over the last years. The rare reports of airway
obstruction directly triggered by the laryngeal
mask are swelling of the pharyngeal soft tis-
sues caused by the leakage of irrigation fluid,
4 herniation of the laryngeal mask airway
cuff, 5-7 foreign bodies (Ascaris lumbricoides),
Case Report
Correspondence author: Binu Puthur SIMON,
Department Of Anaesthesiology RIPAS Hospital, Bandar Seri Begawan BA 1710
Brunei Darussalam E mail: [email protected]
CASE REPORT
A 48-year-old man with a body mass index of
38.9 kg/m2 was scheduled for Trans Pars Pla-
na Vitrectomy and intra-ocular lens insertion
of the left eye under general anaesthesia. His
8 and intermittent obstruction related to a va-
gal nerve stimulator. 9 We report the case
where an unanticipated airway problem arose
due to a defect in a reusable Laryngeal Mask
Airway ProSealTM (LMA-ProSealTM) when it was
used for an ophthalmic procedure. Upon care-
ful investigation post procedure, the inflatable
part of the device had weakened and bal-
looned out causing deformation of the mask
resulting in loss of proper fit and seal.
Brunei Int Med J. 2012; 8 (4): 205-209
past medical history was relevant for obs–
tructive sleep apnoea, hypertension and dia-
betes mellitus.
The patient was pre-oxygenated with
100% oxygen for three minutes and general
anaesthesia was induced with 200mgs of
propofol and 75mcgs of fentanyl. Anaesthesia
was maintained with oxygen, nitrous oxide
and isoflurane. Mask ventilation was fairly
easy and intravenous atracurium (40mg) was
given for relaxation. A size 4 cuff LMA-
ProSealTM laryngeal mask airway was suc-
cessfully placed after checking of the cuff with
30ml air insufflation. The cuff was inflated
with 20ml of air. The patient was ventilated
with pressure controlled ventilation (PCV)
mode with following parameters; inspiratory
pressure of 15cm H2O, respiratory rate of 12
per minute and an inspiratory: expiratory
(I:E) ratio of 1:1. An I:E ratio of 1:1 was pre-
ferred to achieve a higher tidal volume at a
lower pre-set inspiratory pressure (normal
ratio is 1:2). The patient could generate
around 550mls of tidal volume with the above
settings. As a precautionary measure, a size
12 nasogastric tube was introduced through
the drain tube of the laryngeal mask airway
(LMA). His end tidal CO2 was 35-40mm of Hg
(Normal range 35 to 45mm of Hg).
Thirty minutes into the surgery a
gradual decrease in tidal volume (350ml) and
a rise in the end tidal CO2 were noted. An ad-
ditional dose of atracurium (40mg) and fenta-
nyl (75mcgs) were given on the assumption
that the effect of the initial dose was wearing
off, but there was no improvement. Then the
ventilator mode was changed to volume con-
trol with the following setting; tidal volume of
500ml and respiratory rate of 12/minute.
SIMON and HABIBULAH. Brunei Int Med J. 2012; 8 (4): 206
After this adjustment, the peak airway pres-
sure went up to 29-30cm of H2O and the
mean airway pressure was around 17cm of
H2O. The end tidal CO2 was maintained at 48
to 50mm of Hg. We managed to continue with
this strategy for another 40 minutes.
Approximately 10 to 15 minutes be-
fore completion of surgery, breathing at-
tempts by the patient were noted and we at-
tempted to manage the patient on spontane-
ous ventilation mode with manual assistance.
It was noticed that at this point there was
hardly any end tidal CO2 trace and the reser-
voir bag did not show any movement. The
patient seemed to be experiencing acute air-
way obstruction during spontaneous attempts
of breathing. The patient was paralysed with a
10mg dose of atracurium. Volume control
mode was reinstated with the same settings
as before. The patient was then satisfactorily
ventilated until the surgical procedure was
completed.
An intravenous dose of neostigmine
and atropine were then administered to re-
verse the residual effect of the muscle relax-
ants. When breathing attempts commenced,
the patient appeared to be further exhibit air-
way obstruction. At this point the LMA-
ProSealTM was removed and a Guedel airway
(size 4) was inserted. The patient was venti-
lated with a face mask until a good respirato-
ry effort was observed. Further recovery was
uneventful. The LMA-ProSealTM was later ex-
amined and it was noted that the cuff was
deformed with herniation to one side (Figure
1). A faulty LMA with herniation after insuffla-
tions, together with formation of a fold on the
cuff was suspected as the cause of the airway
problem we had encountered in this case.
DISCUSSION
Cases of herniation have been reported with
the Classic LMA™ airway. There was a case
where the plastic layers between the inflated
cuffs of a disposable LMATM had separated and
resulted in a herniation 10 and, another case
where the classic reusable LMA had herniated,
probably as a result of material fatigue follow-
ing repeated sterilisation. 7
In this case, we suspected herniation
(deformation) of the LMA occurred intra-
operatively causing leakage around the LMA-
ProSeal TM. A fold formed on the cuff facing
the laryngeal inlet (Figure 1 b). The defor-
mation led to loss of proper seal causing the
subsequent events.
Though we were successful in venti-
lating the patient using the volume control
mode, the peak airway pressures were high
(30cm of H2O) despite administering a repeat
dose of muscle relaxant. Initially we were
able to get the same tidal volume at a setting
of 15 cm H2O with pressure control mode.
With intermittent positive pressure
ventilation (IPPV) and adjustment to volume
control we succeeded in ventilating the pa-
tient adequately. Since a constant flow of gas
is delivered in the volume control mode, more
tidal volume was achieved with the same air-
way pressure, unlike the decelerating flow
noticed in pressure control mode. Moreover
an airway obstruction at laryngeal inlet could
generate an auto positive end expiratory
pressure (PEEP). This intrinsic PEEP together
with long inspiratory time places the patient
on a less compliant (over distended) part of
the volume pressure curve.
When the patient started breathing
spontaneously the obstruction became more
severe causing inadequate ventilation. The
reasons for almost complete airway obstruc-
tion once spontaneous breathing commenced
could be due to the deformed cuff displacing
the LMA inlet away from the glottis. This ef-
fect possibly got worse with the negative
pressure caused by spontaneous inspiratory
SIMON and HABIBULAH. Brunei Int Med J. 2012; 8 (4): 207
a b
Figures 1: a) Inflated ProSeal TM showing
herniation of the right side (arrow) and b)
View from the other side showing the
herniation resulting in deformation and
loss of seal contour.
efforts. In this particular case, LMA-ProSealTM
was preferred over the endotracheal tube to
obtain smooth recovery and avoid the stress
of extubation occurring after posterior cham-
ber eye surgery. 11 LMA-ProSealTM was used
as we could take extra precautionary
measures against aspiration and gastric dis-
tension (like insertion of an orogastric tube).
Additionally, LMA ProSeal has proven to have
better seal at higher ventilating pressures. 12
There have been case reports of her-
niation of the classic reusable type LMA, pos-
sibly as a result of overuse, over inflation or
repeated sterilisation. 13 Repeated uses of
LMA™ airways beyond 40 uses increases the
probability of device malfunctions. 15 In this
particular case we were unable to determine
how often the LMA-ProSealTM had been sub-
jected to repeated use. As a standard practice
we utilise a cuff volume of 20-30 ml for LMA-
ProSealTM size 3 in accordance with manufac-
turer recommendations and, our sterilisation
complies with manufacturer instructions.
Since the head was draped and cov-
ered for our case, we did not have access to
the airway such that, in the worst case sce-
nario of total airway obstruction we would
have been left with no choice but to stop the
surgery and replace the faulty LMA with an
oral endotracheal tube.
Cuff herniation of the LMA-Classic
may not be apparent if the cuff is inflated
with volume less than the recommended
maximal volume (e.g., 30 ml for size 4) and
may only become apparent when inflated with
a volume of air 50% greater (45 ml). 15 This
practice is now recommended by the manu-
facturer. We recommend this practice to be
followed in LMA–Pro SealTM prior to insertion.
In conclusion, when a ventilation
problem is encountered with a reusable
LMA™, particularly if this had experienced
repeated sterilisation, a herniated cuff should
be considered, even if initial testing was in-
conspicuous. 7 Removal of the malfunctioning
LMA™ and inspection of the cuff should be
considered to rule out this potentially deleteri-
ous technical problem.
SIMON and HABIBULAH. Brunei Int Med J. 2012; 8 (4): 208
AKNOWLEDGEMENT: We would like to acknow-
ledge our colleagues, Dr Salil G NAIR, Dr Anand H
KULKARNI and Dr Hj Zulaidi HJ ABDUL LATIF for
their help with the management of this patient.
REFERENCES
1: Cook TM, Lee G, Nolan JP. The ProSeal laryngeal
mask airway: a review of the literature. Can J
Anaesth. 2005; 52:739-60.
2: Practice guidelines for management of the
difficult airway: an updated report by the American
Society of Anesthesiologists Task Force on Manage-
ment of the Difficult Airway. Anesthesiology 2003;
98:1269–77.
3: Henderson JJ, Popat MT, Latto IP, Pearce AC.
Difficult Airway Society guidelines for management
of the unanticipated difficult intubation. Anaesthesia
2004; 59:675–94.
4: Yoshimura E, Yano T, Ichinose K, Ushijima K.
Airway obstruction involving a laryngeal mask air-
way during arthroscopic shoulder surgery. J Anesth
2005; 19:325–7.
5: Solaidhanasekaran S, Bharamgoudar M. Airway
obstruction secondary to herniation of the paediat-
ric laryngeal mask airway. Anaesthesia 2008;
63:785–6.
6: Christelis ND, Doolan GK. Complete airway ob-
struction caused by cuff herniation of an overused
laryngeal mask airway. Anaesth Intensive Care
2008; 36:274–5.
7: Wrobel M, Ziegeler S, Grundmann U. Airway
obstruction due to cuff herniation of a classic reusa-
ble laryngeal mask airway. Anesthesiology 2007;
106:1255–6.
8: Roy K, Kundra P, Ravishankar M. Unusual for-
eign body airway obstruction after laryngeal mask
airway insertion. Anesth Analg 2005; 101:294–5.
9: Bernards CM. An unusual cause of airway ob-
struction during general anesthesia with a laryngeal
mask airway. Anesthesiology 2004; 100:1017–8.
10: England AJ. Respiratory obstruction secondary
to laryngeal mask failure. Anaesthesia 2004;
59:1030-1.
11: Brimacombe J. The ProSeal laryngeal mask
airway. Anesthesiol Clinics North Am. 2002; 20:871
–91.
12: El-Ganzouri A, Avramov MN, Budac S, Moric M,
Tuman KJ. Pro Seal Laryngeal mask airway versus
endotracheal tube: ease of insertion, hemodynamic
responses and emergence characteristics. Anesthe-
siology 2003; 99: A571.
13: Keller C, Brimacombe J. Mucosal pressure and
oropharyngeal leak pressure with the Pro
Seal versus laryngeal mask airway in anaesthetized
paralysed patients. Brit J Anaesthesia 2000; 8:262
–6.
14: Overuse and failure to track number of uses of
LMATM airways. Available from: http://
www.lmana.com/files/section3-overuse-and-failure
-to-track-lma-424.pdf (Accessed 15th October
2011).
15: Asai T, Koga K, Morris S. Damage to the laryn-
geal mask by residual fluid in the cuff. Anaesthesia
1997; 52:977–81.
SIMON and HABIBULAH. Brunei Int Med J. 2012; 8 (4): 209
Brunei Darussalam Brunei Darussalam Brunei Darussalam Brunei Darussalam ———— Healthcare in PicturesHealthcare in PicturesHealthcare in PicturesHealthcare in Pictures
Hj Ahmed Yunos Bin Hassan (Dresser) doing his daily round with his team in the
surgical ward of the Brunei General Hospital.
(Picture courtesy of Dayangku Dr Siti Nur’Ashikin Bte Pengiran Tengah).