Effective Date: 20 JAN 2015 Page 1 of 3 CCF: 15-0033 Ver 10
AU680 Contamination Avoidance Parameters OUS^
NO. PRECEDING TEST NAME
FOLLOWING TEST NAME
REAGENT PROBE CLEANER
WASH COUNT
EFFECTIVE
OF WATER
CLEANING
SAME USE
MIXER CUVETTE
1. TG 118 LIP 030 Clin Lipase² 3 No Yes No
2. HDL-C 087 LIP 030 Clin Lipase² 5 No Yes No
3. LDL-C 083 LIP 030 Clin Lipase² 5 No Yes No
4. CRP 047 CRP 099 Water 1 Yes Yes No
5. All Gent 420b CLN 20%* 5 No Yes Yes
6. Crea 078 C3 059 Water 1 Yes Yes No
7. Crea 078 C4 060 Water 1 Yes Yes No
8. Crea 078 Malb 067 CLN 100%* 3 No Yes No
9. Crea 078 Prealb 075 Water 1 Yes Yes No
10. Crea 078 ∞ UAlb 224 CLN 100%* 3 No Yes No
11. TP 032 Mg 089 Water 0 No Yes No
12. AST 009 LDH 026, 027, 028 Water 1 Yes No No
13. Chol 016 IP 022 Water 1 Yes No No
14. DBili Colour 011 Urea 034 Water 1 Yes No No
15. DBili Blank 010 Urea 034 Water 1 Yes No No
16. IP 022 UIBC 205 Water 0 No No Yes
17. Crea 078 Fe 086 Water 0 No No Yes
18. TG 118 Fe 086 Water 0 No No Yes
19. Digox 404 Malb 067 CLN 100%* 1 No No No
20. Digox 404 ∞ UAlb 224 CLN 100%* 1 No No No
21. Malb 067 Digitox 403 CLN 100%* 1 No No No
22. ∞ UAlb 224
Digitox 403 CLN 100%* 1 No No No
23. ◊ HbA1c 218 All CLN 100%* 2 No No No
24. ◊ HbA1c 218 ◊ HbA1c 218 CLN 100%* 2 No No Yes
25. ◊ THb 219 THC 322 Water 1 Yes No Yes
26. ◊ THb 219 UCSFP 070 CLN 100%* 2 No No No
27. ◊ HbA1c 218 AMP 323 CLN 100%* 3 No No No
28. ◊ THb 219 AMP 323 CLN 100%* 3 No No No
Effective Date: 20 JAN 2015 Page 2 of 3 CCF: 15-0033 Ver 10
* CLN OSR0001/ODR2000, replenish cleaner on a daily basis or as required. b Recommend to run Gentamycin tests in batch mode. ² ODR20067 Cleaning Solution (for contamination avoidance). Replenish weekly or as required. ◊ Refers to reagent part number B00389 ∞Refers to reagent part numbers B38858/B46435 ^ OUS: For Europe, Emerging Markets, Asia
Effective Date: 20 JAN 2015 Page 3 of 3 CCF: 15-0033 Ver 10
AU680 Sample Carryover Contamination Avoidance Parameters OUS^
Contamination Parameters → Carry-Over Prevention (Test)
TEST NAME
PRE-DISPENSE WASH COUNT POST- DISPENSE WASH COUNT
DETERGENT -1 DETERGENT -2 WATER DETERGENT -1 DETERGENT -2 WATER
Urine/CSF Albumin
B38858/B46435 1 0 1 0 0 0
AU680 Sample Carryover Contamination Avoidance Parameters OUS^
Contamination Parameters →Carry-Over Prevention (Type Changes)
NO. COMBINATION
WASH COUNT
DETERGENT -1 DETERGENT -2 WATER
4. Serum to Urine 1□ 0 1□
Detergent 1 = 2% Wash Solution (OSR0001/ODR2000) □ Recommended sample carryover parameters for Urine/CSF Albumin B38858/B46435.Test or type changes parameters can be used for Urine/CSF Albumin B38858/B46435 depending on lab usage. ^ OUS: For Europe, Emerging Markets, Asia