50 years’ experience at your service• On-site calibration of all current measuring principles
• System alignment from sensor to complete loop check
• Linearization of complex container geometries
Calibration of level instruments
Instruments International
Endress+Hauser Instruments International AGKägenstrasse 24153 ReinachSwitzerlandTel. +41 61 715 81 00Fax +41 61 715 25 00http://[email protected]
Other locations
Please visit: www.endress.com/worldwide
IN00103H/29/EN/01.06PF/INDD CS4
Long-term stability and guaranteed precision are essential requirements for level measurement devices. Regular calibration is particularly important for quality and safety - critical measuring points.
Calibration of level instruments
On-site calibration and system alignment Various measuring principles can be used to measure the level of liquids and bulk goods, depending on your requirements. Endress+Hauser offers 8 physical principles, each tailored to your application. The calibration and alignment options are as varied as the measuring principles. Parameterisation of the level measurement instrument is often undertaken during the commissioning phase. In the majority of cases this is sufficient, and is particularly simple for pre-aligned sensors. If level measurement instruments are to be recalibrated following an operational phase, this is often done by gauging the capacity of a container in liters, so there is no need to remove the device from the process. We can provide you with specific advice on this.
We also offer factory calibration at our in-house reference calibration centre,
Place of calibration On-site In the laboratory (at E+H)
Calibration range
Pressure using handpump, volumetric measurement with water or original
medium upon request, following technical clarification
Absolute measuring accuracy of 0.5 mm
Equipment concerned
• Pressure (hydrostatic) instruments • Radar instruments• Ultrasonic instruments• Capacitative instruments • Others (on request)
Radar instruments
Calibration method
By generation of a reference pressure or by tank contents measurement (emptying of tank via a standard
flowmeter)
By comparison to a laser or a target
Number of calibration points 1, 3, 5 points or other, as agreed 5 points
10 points for Micropilots (accuracy 1 mm)
Resources used• Pressure: handpump• Level: mobile bench with specific
standard flowmeter Laser interferometer
Metrological compliance
To national and international standards
To national and international standards
Documents issued
• Calibration certificate• On request, copy of the proof
of conformity to national/international standards
Calibration certificate
specifically for radar measuring devices subject to mandatory calibration. Here, devices with
an absolute measuring accuracy of 0.5 mm can be calibrated and adjusted.