Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version #: Author: Janice Thompson Page of 9 Approval Date: 1 Core Equipment ID: 15922 Description: LI-COR Biosciences, Odyssey Infrared Imaging System Room: B446 (Molecular & Biochemical Core) Champion: 1.0 Purpose Standardize the process for control, maintenance, ownership and use of the LI‐COR Biosciences Odyssey Infrared Imaging System located in B‐446 Life Sciences. 1.1 LI-COR Biosciences, Odyssey Infrared Imaging System Capabilities 1 The Odyssey® Infrared Imaging System sweeps away the old paradigms of chemiluminescence and visible fluorescence, and introduces a new standard for Western blot analysis – direct infrared fluorescence detection. a. Infrared detection gives you the quantitative analysis and wide linear dynamic range that chemiluminescence can’t. b. Strong and weak bands on the same blot are accurately detected without the uncertainty and inconvenience of multiple exposures, and without spending time in the darkroom. c. The Odyssey System gives you clear, sharp, reproducible bands without fuzziness or “blowout”. d. Bands hidden by overexposure with chemiluminescence become clear when imaged with Odyssey. LI-COR Biosciences, Odyssey Additional Capabilities 2 The Odyssey® System provides a flexible, multifunctional platform to accommodate a variety of applications, so you get results faster. a. Western Detection including two color blots and in‐gel westerns 1 LI‐COR Biosciences – Biotechnology Product Line. LI‐COR Biosciences. April 1, 2009. http://www.licor.com/bio/odyssey/OdysseyTour_1.jsp 2 LI‐COR Biosciences – Odyssey: Clear Image Data. LI‐COR Biosciences. April 1, 2009. http://www.licor.com/bio/odyssey/documents/Odyssey.pdf

LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Embed Size (px)

Citation preview

Page 1: LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Pharmacology & Toxicology Core Facilities    SOP #:15922.B446.001      Version #:  

Author: Janice Thompson  Page      of  9  Approval Date:      

Core Equipment ID:  15922  

Description: LI­COR Biosciences, Odyssey Infrared Imaging System 

Room: B446 (Molecular & Biochemical Core) 

Champion:  

 

1.0 Purpose Standardize the process for control, maintenance, ownership and use of the LI‐COR Biosciences Odyssey  Infrared  Imaging  System  located  in B‐446 Life Sciences.   1.1 LI­COR  Biosciences,  Odyssey  Infrared  Imaging  System 

Capabilities1 The  Odyssey®  Infrared  Imaging  System  sweeps  away  the  old paradigms  of  chemiluminescence  and  visible  fluorescence,  and introduces a new standard for Western blot analysis – direct infrared fluorescence detection.   a. Infrared detection gives you  the quantitative analysis and wide 

linear dynamic range that chemiluminescence can’t.   b. Strong and weak bands on the same blot are accurately detected 

without  the  uncertainty  and  inconvenience  of  multiple exposures, and without spending time in the darkroom.  

 c. The Odyssey System gives you clear,  sharp,  reproducible bands 

without fuzziness or “blowout”.   d. Bands hidden by overexposure with chemiluminescence become 

clear when imaged with Odyssey.  

LI­COR Biosciences, Odyssey Additional Capabilities2 The Odyssey® System provides a flexible, multifunctional platform to accommodate a variety of applications, so you get results faster.  a. Western  Detection  –  including  two  color  blots  and  in‐gel 

westerns  

                                                        1 LI‐COR Biosciences – Biotechnology Product Line. LI‐COR Biosciences. April 1, 2009. http://www.licor.com/bio/odyssey/OdysseyTour_1.jsp 2 LI‐COR Biosciences – Odyssey: Clear Image Data. LI‐COR Biosciences. April 1, 2009. http://www.licor.com/bio/odyssey/documents/Odyssey.pdf 

Page 2: LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Pharmacology & Toxicology Core Facilities    SOP #:15922.B446.001      Version #:  

Author: Janice Thompson  Page      of  9  Approval Date:      

b. Cell‐Based  Assays  –  including  in‐cell  westerns  and  on‐cell westerns 

 c. Small Animal Imaging 

 d. Protein Detection – including 1‐D/2‐D Coomassie Arrays 

 e. Nucleic Acids – including EMSA/Gel Shift and DNA Staining 

 f. Microwell Assays – including ELISA/FLISA, transcription factors 

and arrays  

g. Tissue and Tissue Sections    

2.0 Reason for Issue Maintain a document that describes the Standard Operating Procedures that allows  for  the  standard  safe  and  maximal  use  of  the  LI‐COR  Biosystems Odyssey within the Pharmacology and Toxicology Core Facilities.   

3.0 Process Description Allows  Core  Facility  Users  within  the  Pharmacology  and  Toxicology Department to properly and effectively use the LI‐COR Biosciences Odyssey. The process description details and creates a standard for the use of the LI‐COR Biosciences Odyssey. The controlled standard must maintain and adhere to proper and approved research and regulatory qualitative conditions.   3.1 SOP:  15922.B446.001  for  LI‐COR  Biosciences  Odyssey,  authored  by 

Janice Thompson, created on April 1, 2009, issued on (insert date).   

3.2 SOP: 15922.B446.001 applies to any user and/or trainer of the LI‐COR Biosciences Odyssey.  

3.3 Responsibilities:  All  Users  are  responsible  for  obtaining  the proper approval and training before the use of the LI‐COR Biosciences Odyssey.  All  users  are  responsible  for  the  proper  use,  according  to defined protocol, when using the LI‐COR Biosciences Odyssey.  a. New Users need a new LI‐COR Biosciences Odyssey user name and 

password  created  for  equipment  access,  before  initial  use.  New accounts are authorized and created by the Equipment Champion and/or the Core Facility Manager. A new account may be created after training and equipment approval has occurred.    

Page 3: LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Pharmacology & Toxicology Core Facilities    SOP #:15922.B446.001      Version #:  

Author: Janice Thompson  Page      of  9  Approval Date:      

b. All Users must schedule equipment use in the Equipment Schedule Logbook prior to use. The Equipment Schedule Logbook is located on  the  desk  next  to  the  LI‐COR  Odyssey.  Please  note  that  the Nanodrop Technologies Nanodrop ND‐1000 Spectrophotometer is run on the same computer as the LI‐COR Biosciences Odyssey and Users must check the Equipment Schedule Logbook for this piece of equipment when considering the use of the LI‐COR Biosciences Odyssey. 

 c. All users must  record all  equipment use  in  the Equipment Usage 

Logbook  post‐use  on  the  same  day  as  the  recorded  use.  The Logbook  is  located  on  the  desk  next  to  the  LI‐COR  Biosciences Odyssey. Within the Logbook on the current log sheet, Users must record the following: Date, PI, Name, Lab Location, Phone Number, Channel(s) Used, Number of Scans, Comments.  

 d. Immediately  after  use,  the  LI‐COR  Biosciences  Odyssey  must  be 

cleaned with distilled water and a Kimwipe to ensure residue does not remain on the scanning surface. 

 3.4 Equipment Safety Issues 

 a. Laser3  –  The  Odyssey  Infrared  Imaging  System  contains  two 

lasers; one emitting at 785 nm, and one at 685 nm. Each has a peak power rating of 30 milliwatts. The 685 nm laser emits visible laser radiation – direct exposure to either beam may cause eye damage. Laser  radiation  is  emitted  through  apertures  at  the  top  of  the Odyssey microscope assembly. Because  the microscope assembly moves  in both planes relative  to  the glass scanning surface,  laser radiation could be  focused at  any position on  this  surface.  Safety interlocks automatically turn the lasers off when the Odyssey lid is opened. 

 b. Safety Interlocks3 – The Odyssey Infrared Imager has two safety 

interlocks that prevent access to the laser when the instrument lid is  opened  during  operation.  Do  not  attempt  to  defeat  these interlocks.  The  interlocks  are  located  on  the  front  of  the instrument bezel as shown in the Operator’s Manual, Version 2.1. Audible tones sound when the lid is opened and closed. 

 c. Chemical Safety3    ‐ LI‐COR recommends that all biochemicals be 

handled carefully, and that safe laboratory procedures be followed 

                                                        3 Odyssey Infrared Imaging System Operator’s Manual, Version 2.1. LI‐COR Biosciences. 

Page 4: LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Pharmacology & Toxicology Core Facilities    SOP #:15922.B446.001      Version #:  

Author: Janice Thompson  Page      of  9  Approval Date:      

at all times. Be aware of the hazards associated with any chemical before you begin work. 

 d. Radiation Safety4  ‐ The Odyssey Imager should not be used with 

any radioactive materials.   

3.5 Quality Measures   

a. Each Use: Each time the LI‐COR Bioscience Odyssey is used, wipe all  chemical  spills  and  residues  from  both  the  case  and  the scanning  surface.  This  prevents  the  surface  coating  from  being damaged. Use only warm water and a damp cloth to clean the LI‐COR Biosciences Odyssey. 

 b. Every Six Months: Inspect cables and power cords to ensure their 

integrity.   

3.6 Procedure for Western Blot Scanning; LI­COR Biosciences   

Setup 

a. Double  click on Odyssey V3.0  software desktop  icon  to  start use. This software runs the LI‐COR Biosciences Odyssey system. 

b. When  initial  screen  has  loaded,  click  on  “File”  and  select  “New” from the dropdown menu. 

c. In the prompt box that opens, select “Browse” from the right side of the box. Within the window that opens, select the folder to save the  file  in.  Ensure  that  it  is within  the  folder  for  your  laboratory only.  

d. In the “File Name” area, type in the name of the file. 

e. Select “Scan” from the bottom of the window. 

f. When the “Please note the project location” message appears, click “OK”. 

g. Type in your User Name and Password. (These will be assigned to you when training has been completed.) Click “OK”. 

h. Open  the  lid  of  the  LI‐COR  Biosciences  Odyssey.  A  soft,  short beeping  sound  will  be  heard.  This  indicates  the  laser  has  been 

                                                        4 Odyssey Infrared Imaging System Operator’s Manual, Version 2.1. LI‐COR Biosciences. 

Page 5: LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Pharmacology & Toxicology Core Facilities    SOP #:15922.B446.001      Version #:  

Author: Janice Thompson  Page      of  9  Approval Date:      

turned off and that it is safe for the User to work on the scanning surface. 

i. Place  the  membrane  face  down  on  the  lower  left  corner  of  the scanning  surface,  noting  the  coordinates  from  the  scanning surface. Roll the membrane with the brayer to remove air bubbles.  

j. Close  the  lid  of  the  LI‐COR  Biosciences  Odyssey.  A  soft,  short beeping  sound will  be  heard.  This  indicates  the  safety  interlocks will  allow  the  laser  to  function when  the  parameters  have  been set. 

k. Identify  “Group”  from  the  dropdown  menu.  This  should  be  the same as your User Name. 

l. Identify the Preset. For western blots, choose “Membrane”. 

m. The “Scan Parameters” remain at the factory default settings. 

n. Identify the wavelength to be scanned. This will coincide with the label  attached  to  the  secondary  antibody.  If  using both  channels, check boxes for both the 700 and 800 nm wavelengths. 

o. The “Intensity” remains at the factory default settings. 

p. Identify the scan area by typing in the coordinates OR resizing the highlighted red box on the grid. To resize the red box, hover over a line with the cursor until  it changes from a four‐sided arrow to a two‐sided arrow. 

q. Select  “Start  Scan”.  The LI‐COR Biosciences Odyssey will  connect with  the  computer  and  the  scanning  process  will  begin.  A  noise will be heard from the Odyssey while the laser is moved back and forth over  the area  to be scanned. As  lines are  transferred  to  the computer, an image will appear in the highlighted area on the grid. 

r. When scan  is  complete, press  “Save”. Type  in  the  scan name and analysis  name,  using  features  identifiable  for  future  reference. Press “OK”. 

s. Close  the  Scanner  Console  screen  by  pressing  the  red  “X”  in  the upper  right  corner or press  the  “Close” button  in  the  lower  right corner of the screen. 

Analysis 

a. Click  on  “View”  and  select  “Alter  Image  Display”  from  the dropdown menu. 

Page 6: LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Pharmacology & Toxicology Core Facilities    SOP #:15922.B446.001      Version #:  

Author: Janice Thompson  Page      of  9  Approval Date:      

b. Adjust  “Brightness”  and  “Contrast”  levels  for  the  appropriate channels by sliding  the blue square on each bar until  the desired levels have been achieved. 

c. To further enhance the image, the “Linear Manual” can be changed by sliding the blue square on this bar. Further adjustments to the “Brightness”  and  Contrast”  levels  can  be  made  after  the  “Linear Manual” number has been changed. 

d. “Black  and White”  can  only  be  used  as  an  option  when  a  single channel is selected.  

e. When finished, press “OK”. 

f. If  everything  is  acceptable,  click  on  “File”  and  select  “Save”  from the dropdown menu.  

g. The  image  can be printed  (click on  “File”  and  select  “Print”  from the  dropdown  menu,  then  follow  the  prompts)  or  saved  to  a jumpdrive and taken with the User. 

Clean Up 

a. When  complete,  close  the  software by  clicking  the  red  “X”  in  the upper  right  corner  of  the  window  or  clicking  on  “File”  and selecting “Exit” from the dropdown menu. 

b. Open  the  lid  of  the  LI‐COR  Biosciences  Odyssey;  a  short,  soft beeping  will  be  heard  to  indicate  the  laser  has  been  turned  off. Remove the membrane.  

c. Clean  the  scanning  surface  of  the  Odyssey  using  a  Kimwipe dampened with  distilled water.  Dry  thoroughly with  a  Kimwipe, then  close  the  lid. A  short,  soft  beeping will  be heard  to  indicate the laser is ready for the next task. 

d. Sign out in the Equipment Usage Log as indicated in Section 3.3c. 

3.7 Core Materials A  brayer  is  provided  on  the  bench  with  the  LI‐COR  Biosciences Odyssey. Currently, there is also a bottle of distilled water and a box of Kimwipes located near the brayer.   

3.8 User Materials Secondary  antibodies  compatible  with  the  LI‐COR  Biosciences Odyssey  for use at  the 700 and 800 nm wavelengths may be  located and  purchased  from  the  below website.  These  antibodies will  allow for  the  proper  and  optimal  use  of  the  LI‐COR  Biosciences  Odyssey. 

Page 7: LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Pharmacology & Toxicology Core Facilities    SOP #:15922.B446.001      Version #:  

Author: Janice Thompson  Page      of  9  Approval Date:      

Users are individually responsible for the compatibility, maintenance and purchase of these materials.  Related Products http://www.licor.com/bio/reagents/irdye_secondary_antibodies.jsp http://www.licor.com/bio/reagents/irdyes.jsp  

3.9 Use Records  

a. Records  of  Use  –  All  LI‐COR  Biosciences  Odyssey  use  must  be recorded. Refer to 3.3c.   

b. Error Messages / System Issues – All error messages and system issues  must  be  relayed  to  the  Equipment  Champion  and  the Pharmacology  &  Toxicology  Core  Facility  Manager.  This information must  be  relayed on  the  same day  as  equipment use. Error messages/system issues must be recorded. Refer to 3.3c. 

 3.10 Resource Index  

LI‐COR Biosciences Odyssey literature and resources for the following items may be found at the link below.  a. Operator’s Manual (multiple versions) b. Application Protocols Manual c. User Guide (multiple versions) d. Etc.  http://biosupport.licor.com/index.jsp?spec=Manuals,Manuals&more=Manuals&sort=1 

 4.0 Competences, Authorization and Training 

New  Users  must  receive  proper  authorization  from  either  the  Equipment Champion and/or Pharmacology & Toxicology Core Facility Manager before equipment  use.  A  new  User  may  contact  the  Equipment  Champion  or Pharmacology  &  Toxicology  Core  Facility  Manager  to  schedule  training. Training  includes  SOP  and  instrument  familiarization  and  any  additional required  or  specialized  training.  Once  training  is  complete,  authorization maybe issued and a User Name and Password will be established. All Users are individually responsible for current SOP familiarization.  

5.0 SOP Performance and Equipment Review The  effectiveness  of  the  SOP  15922.B446.001  will  be  monitored  by  the Pharmacology & Toxicology Core Facility Manager, Equipment Champion and all Users. Any procedural or qualitative deviations will be reflected within an updated  SOP.  Any  approved  User  should  aptly  report  any  procedural  or 

Page 8: LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Pharmacology & Toxicology Core Facilities    SOP #:15922.B446.001      Version #:  

Author: Janice Thompson  Page      of  9  Approval Date:      

qualitative  issues  and/or  errors  to  the  Pharmacology  &  Toxicology  Core Facility Manager or Equipment Champion. The Equipment Champion’s name and  contact  information  can  be  found  on  the  Core  Facility  Identification Sticker.  The  Core  Facility  Identification  Sticker  is  located  on  the  LI‐COR Biosciences Odyssey. The Core Facility Manager’s contact information may be found on the Core Facility Contact Posting within room B446. Updated SOPs will be published and approved Users will be notified. SOP: 15922.B446.001 review will occur every two years.   5.1 Update Date:     

Reason for SOP Change:     New Version #:  

5.2 Update Date Reason for SOP Change New Version #: 

 6.0 Definitions 

Any  standard  defined  term  specifically  used  in  the  Pharmacology  and Toxicology  Core  Facilities’  SOP.  These  are  terms  that  carry  a  unique  Core Facility  definition  that  is  pertinent  to  proper  SOP  understanding  and implementation. A list of terms and definitions is maintained to allow Users a quick  and  efficient  reference.  Listed  below  are  common Pharmacology  and Toxicology Core Facility terms. New common terms are to be approved and added, with definition, to the term reference guide.   a.  SOP  Standard  Operating  Procedure,  which  is  a  standard  guide  that officially standardizes the process of control, maintenance and ownership of the LI‐COR Biosciences Odyssey. The SOP number stand for (xxx  . xxx  . xxx) equipment number . room number . SOP version number.  b. Originator / Author  The  individual  representing  the Pharmacology and Toxicology Core Facilities that created SOP: 15922.B446.001.  c.  Stakeholder  Any  individual  that  uses  the  LI‐COR  Biosciences  Odyssey, including the Pharmacology and Toxicology Core Facilities Department.   d.  New  User  An  individual  who  has  not  completed  the  requirements  of section 4.0.  e. Approved User An  individual who uses or performs the task of which  is the  subject  of  the  SOP  and  has  fulfilled  section  4.0.  This  title may  only  be given  by  the  Equipment  Champion  and  /  or  the  Pharmacology  and Toxicology Core Facilities Manager.   

Page 9: LICOR Biosciences, Odyssey Infrared Imaging …core.phmtox.msu.edu/Scheduling/ItemDocs/20/SOP_LICOR_Odyssey.… · Pharmacology & Toxicology Core Facilities SOP #:15922.B446.001 Version

Pharmacology & Toxicology Core Facilities    SOP #:15922.B446.001      Version #:  

Author: Janice Thompson  Page      of  9  Approval Date:      

f.  Champion  An  individual  whose  direct  expertise  with  the  LI‐COR Biosciences  Odyssey  has  been  recognized  by  the  Pharmacology  and Toxicology  Core  Facilities  Committee.  The  Pharmacology  and  Toxicology Core Facility Committee may only award the champion title.   

7.0 Appendix The  below  signatures  and  dates  are  required  for  full  SOP  approval  and implementation.   This SOP was written by:        Janice Thompson _________________________________ 

 This SOP was reviewed by:       Dr. Stephanie Watts ______________________________  This SOP was authorized by:    Dr. J.R. Haywood __________________________________    Issue Date: _____________________________________