ABX Pentra 80 rangeLMNE Lamp Adjustment Tool

Tn

SIL

Page 2: LMNE Lamp Adjustment Toolits-information.horiba-abx.com/_basedoc/hematology/abx... · 2016-09-05 · ABX Pentra 80 range LMNE Lamp Adjustment Tool Tn SIL. LMNE Lamp Adjustment Tool

LMNE Lamp Adjustment Tool SIL: RAN738A

Objective Adjusting the LMNE lamp height using the adjustment tool

Intervention

informationIntervention priority For information - As needed - Mandatory on next visit - Immediate and mandatory

Tracking number SIL-RAN738A Intervention estimated duration: xhxx

User process impact YES NO Potential patient result impact YES NO

User training needed YES NO User documentation update YES NO

Applied in factory on instrument serial #

since instrument serial #

Note history Note index Reason for change

A to B

Software

version

Note index Previous New Mini. requirements

A to B

B to C

C to D

New parts Parts designation Previous New Reason for change

Lamp Adjust Tool P60/P80

XDB302AS Creation

New

adjustments

Adjustments Previous New Reason for change

New

documents

Documentation Previous New Reason for change

Additionnal

requirements

Designation Part number Link

2/10

Page 3: LMNE Lamp Adjustment Toolits-information.horiba-abx.com/_basedoc/hematology/abx... · 2016-09-05 · ABX Pentra 80 range LMNE Lamp Adjustment Tool Tn SIL. LMNE Lamp Adjustment Tool

LMNE Lamp Adjustment Tool SIL: RAN738A

1. Lamp Adjustment ToolA lamp adjustment tool is now available for ABX Pentra 80 range instruments. It allows you to adjust the height of theLMNE lamp more precisely.

It can also be used on ABX Pentra 60 range instruments.

Part detailskit or Assy Designation

Component p/n

Component designation Qtyavaila-ble for order

XDB302ASLamp Adjust Tool P60/P80

yes

KAA004A CHC M3 X 10 screws 2 no

KAA049A CHC M2 X 10 screws 2 no

FAA072A O’ring Intern. Water Tank 1 no

FCA002A Focusing Screen 1 no

FCA026A Optic Bench Mirror 1 no

GBL0635 Gun Support 1 no

GBL0636 Optical Gun 1 no

MAJ004A 20 Wedges Kit 0.04 to 1mm 1 no

RAN738ATechnical Note Lamp Adjustment Tool P80

1 no

3/10

Page 4: LMNE Lamp Adjustment Toolits-information.horiba-abx.com/_basedoc/hematology/abx... · 2016-09-05 · ABX Pentra 80 range LMNE Lamp Adjustment Tool Tn SIL. LMNE Lamp Adjustment Tool

LMNE Lamp Adjustment Tool SIL: RAN738A

2. Procedure for ABX Pentra 80 and ABX Pentra XL80Refer to RAS341 of the Pentra 80 Range Technical Manual (P/n: RAA022E) for the entire lamp replacement procedure.

Before starting the adjustment, make sure there are no air bubbles in the flowcell. If there are, go to "Service > Super User Menu > Hydraulic System > Clean Cycles" and press "Rinse Cytometer" to clean theflowcell.

2.1. Lamp Height Check

1 - Loosen the 2 screws of the LMNE amplifier board and lift the small plate.

2 - Disconnect the coaxial cable and remove the LMNE amplifier board.

3 - Remove the 2 screws.

4 - Screw the first part of the adjustment tool in place using the appropriate fixation screws attached to the tool.

5 - Set the second part of the adjustment tool on the first part.

4/10

Page 5: LMNE Lamp Adjustment Toolits-information.horiba-abx.com/_basedoc/hematology/abx... · 2016-09-05 · ABX Pentra 80 range LMNE Lamp Adjustment Tool Tn SIL. LMNE Lamp Adjustment Tool

LMNE Lamp Adjustment Tool SIL: RAN738A

2.2. Lamp Height Adjustment

If the image is too low, decrease the distance between the bench and the lamp holder.

If the image is too high, increase the distance between the bench and the lamp holder.

To do so:

6 - Check the position of the lamp using the adjustment tool.The width of the top part (a) and the width of the bottom part (b) of the light must be the same.

7 - If the adjustment is not correct, perform the Lamp Height Adjustment procedure.

1 - Check the distance between the bench and the lamp holder.

2 - Use the measured distance and your observation to estimate the thickness of the gauges required.

3 - Loosen the small screw on the side of the lamp assembly using a 2 mm hexagonal key.

5/10

Page 6: LMNE Lamp Adjustment Toolits-information.horiba-abx.com/_basedoc/hematology/abx... · 2016-09-05 · ABX Pentra 80 range LMNE Lamp Adjustment Tool Tn SIL. LMNE Lamp Adjustment Tool

LMNE Lamp Adjustment Tool SIL: RAN738A

2.3. Lamp orientation adjustment

4 - Insert the appropriate gauges (supplied with the adjustment tool) between the lamp holder and the optical bench on each side to adjust the height of the lamp.Add gauges until the the light is centered in the circle.

5 - Make sure you have the same height on each side of the lamp holder plate. For example, use a 0.9 mm gauge on one side, and a 0.8 mm gauge + a 0.1 mm gauge on the other side.

6 - Perform the Lamp orientation adjustment procedure when you have found the correct gauges.

1 - Push the lamp holder down and turn the lamp until the light is as homogeneous as possible.

2 - Once you have found the correct position of the lamp, push the lamp holder down and tighten the small screw on the side of the lamp assembly.

3 - Remove the gauges and check the adjustment again.

4 - Re-assemble the LMNE amplifier board.

5 - Adjust the lamp voltage to 6.7 V +/- 0.8 with the R414 potentiometer on the LMNE amplifier board.

Not OK

Filaments are visible An elliptic form is visible on the right

OK

6/10

Page 7: LMNE Lamp Adjustment Toolits-information.horiba-abx.com/_basedoc/hematology/abx... · 2016-09-05 · ABX Pentra 80 range LMNE Lamp Adjustment Tool Tn SIL. LMNE Lamp Adjustment Tool

LMNE Lamp Adjustment Tool SIL: RAN738A

3. Procedure for Pentra XLRRefer to RAS735 of the Pentra 80 Range Technical Manual (P/n: RAA022E) for the entire lamp replacement procedure.

Before starting the adjustment, make sure there are no air bubbles in the flowcell. If there are, go to "Service > Super User Menu > Hydraulic System > Clean Cycles" and press "Rinse Cytometer" to clean theflowcell.

3.1. Lamp Height Check

3.2. Lamp Height Adjustment

If the image is too low, decrease the distance between the bench and the lamp holder.

1 - Disconnect and remove the APD board.

2 - Screw the first part of the adjustment tool in place using the appropriate fixation screws attached to the tool.

3 - Set the second part of the adjustment tool on the first part.

4 - Check the position of the lamp using the adjustment tool.The width of the top part (a) and the width of the bottom part (b) of the light must be the same.

5 - If the adjustment is not correct, perform the Lamp Height Adjustment procedure.

7/10

Page 8: LMNE Lamp Adjustment Toolits-information.horiba-abx.com/_basedoc/hematology/abx... · 2016-09-05 · ABX Pentra 80 range LMNE Lamp Adjustment Tool Tn SIL. LMNE Lamp Adjustment Tool

LMNE Lamp Adjustment Tool SIL: RAN738A

If the image is too high, increase the distance between the bench and the lamp holder.

To do so:

1 - Check the distance between the bench and the lamp holder.

2 - Use the measured distance and your observation to estimate the thickness of the gauges required.

3 - Loosen the small screw on the side of the lamp assembly using a 2 mm hexagonal key.

4 - Insert the appropriate gauges (supplied with the adjustment tool) between the lamp holder and the optical bench on each side to adjust the height of the lamp.Add gauges until the the light is centered in the circle.

5 - Make sure you have the same height on each side of the lamp holder plate. For example, use a 0.9 mm gauge on one side, and a 0.8 mm gauge + a 0.1 mm gauge on the other side.

6 - Perform the Lamp orientation adjustment procedure when you have found the correct gauges.

8/10

Page 9: LMNE Lamp Adjustment Toolits-information.horiba-abx.com/_basedoc/hematology/abx... · 2016-09-05 · ABX Pentra 80 range LMNE Lamp Adjustment Tool Tn SIL. LMNE Lamp Adjustment Tool

LMNE Lamp Adjustment Tool SIL: RAN738A

3.3. Lamp orientation adjustment

1 - Push the lamp holder down and turn the lamp until the light is as homogeneous as possible.

2 - Once you have found the correct position of the lamp, push the lamp holder down and tighten the small screw on the side of the lamp assembly.

3 - Remove the gauges and check the adjustment again.

4 - Make sure the shutter flap can move freely up and down and that it is not in contact with the shutter.

5 - Put the APD board back in place and connect the cables back.

6 - Adjust the lamp voltage to 6.7 V +/- 0.8 on the APD board.

7 - Make sure there are no air bubbles in the flowcell.

Not OK

Filaments are visible An elliptic form is visible on the right

OK

9/10