14
GE Medical Systems – Americas: 888-202-5528 P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A. GE Ultrasound - Europe: +49-212-28020 P.O. Box 1105 60 D-42665 Solingen GE Ultrasound - Asia: 65-291-8528 Singapore: 298 Tiong Bahru Road, #15-01/06 Tiong Bahru Plaza, Singapore 168730 Internet: http://www.gemedicalsystems.com Rev3.0.1 July 9, 2009 LOGIQ P5 TruScan Imaging Technology Product Description User Interface Operator Keyboard The LOGIQ® P5 is a high performance, highly mobile and easy to use multipurpose color Doppler Ultrasound imaging system designed for Obstetrical, Gynecological, Cardiac, Musculoskeletal, Vascular, Urological, Small Parts, Superficial, Pediatric, Neonatal, Transcranial, Abdominal and other applications. Keyboard Width: 43cm Keyboard Height: 84cm Alphanumeric Keyboard Ergonomic Hard Key Operations Indicator Lights Identify Activated Keys Integrated Recording Keys for Remote Control of Up to 2 Peripheral Devices and DICOM Devices 8 TGC Pods, with Re-mapping Functionality at Any Depth System Architecture Monitor TruScan™ Architecture – GE’s innovative, software-intensive platform provides exceptional computational power, image-manipulation capabilities, real time 4D capabilities, workflow flexibility and product upgradeability. 15 inch TFT LCD Electrical Power XGA Format: - Display size: 1024 x 768 Voltage: 100-120Vac or 220-240Vac - Recording area: 800 x 600 Frequency: 50/60 Hz Adjustable Tilt/Swivel Power: Max. 750 VA with Built-in and On-Board Peripherals Digital Brightness/Contrast Adjustment TruAccess is the GE advanced, raw-data processing technology that has helped transform ultrasound imaging. By accessing raw data, TruScan achieves excellent image quality and ensures exceptional image management capabilities. Maximum Thermal Output: 2200 BTU/hr. System Overview Console Design Applications 2 Active Probe Ports Abdominal Integrated HDD (Capacity: 160GB) Obstetrical Integrated DVD-R/W Drive SmartScan utilizes new advances in operating algorithms and system operations to aid in improving image acquisition and patient throughput while helping to increase diagnostic confidence and exam consistency. Gynecological On-board Storage for Peripherals (Max 3 peripherals) Cardiac Musculoskeletal Wheels Vascular - Wheel diameter: 12.5cm Urological - Integrated locking mechanism that provides rolling lock and optional caster swivel lock Small Parts and Superficial ComfortScan, our advanced user- friendly designed software suite helps to maximize productivity. It simplifies every exam you perform. Pediatric and Neonatal Transcranial Probe Holders, Removable for Cleaning and Washing Scanning Methods Gel Holder, Removable for Cleaning and Washing Electronic Sector General Specification Electronic Convex Air Filters, Easily Removable Electronic Linear Dimensions and Weight Front Handle Mechanical Volume Sweep Height: 135/141 cm Rear Handle (Option) Transducer Types Width: 43cm Drawer (Option) Sector Phased Array Depth: 64cm Probe Cable Hanger (Option) Convex Array Weight: approx. 75 kg (165 lb.) Urology Probe Holder (Option) LOGIQ P5 DOC0518685 Page 1 of 14

LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

  • Upload
    vohanh

  • View
    237

  • Download
    1

Embed Size (px)

Citation preview

Page 1: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

GE Medical Systems – Americas: 888-202-5528 P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A. GE Ultrasound - Europe: +49-212-28020 P.O. Box 1105 60 D-42665 Solingen GE Ultrasound - Asia: 65-291-8528 Singapore: 298 Tiong Bahru Road, #15-01/06 Tiong Bahru Plaza, Singapore 168730

Internet: http://www.gemedicalsystems.com Rev3.0.1 July 9, 2009

LOGIQ P5 TruScan Imaging Technology

Product Description User Interface

Operator Keyboard The LOGIQ® P5 is a high performance, highly mobile and easy to use multipurpose color Doppler Ultrasound imaging system designed for Obstetrical, Gynecological, Cardiac, Musculoskeletal, Vascular, Urological, Small Parts, Superficial, Pediatric, Neonatal, Transcranial, Abdominal and other applications.

• Keyboard Width: 43cm Keyboard Height: 84cm

• Alphanumeric Keyboard • Ergonomic Hard Key Operations • Indicator Lights Identify Activated Keys • Integrated Recording Keys for Remote

Control of Up to 2 Peripheral Devices and DICOM Devices

• 8 TGC Pods, with Re-mapping Functionality at Any Depth

System Architecture

Monitor TruScan™ Architecture – GE’s innovative, software-intensive platform provides exceptional computational power, image-manipulation capabilities, real time 4D capabilities, workflow flexibility and product upgradeability.

• 15 inch TFT LCD Electrical Power • XGA Format:

- Display size: 1024 x 768 • Voltage: 100-120Vac or 220-240Vac - Recording area: 800 x 600 • Frequency: 50/60 Hz

• Adjustable Tilt/Swivel • Power: Max. 750 VA with Built-in and On-Board Peripherals • Digital Brightness/Contrast Adjustment TruAccess is the GE advanced, raw-data

processing technology that has helped transform ultrasound imaging. By accessing raw data, TruScan achieves excellent image quality and ensures exceptional image management capabilities.

• Maximum Thermal Output: 2200 BTU/hr. System Overview

Console Design Applications • 2 Active Probe Ports • Abdominal • Integrated HDD (Capacity: 160GB) • Obstetrical • Integrated DVD-R/W Drive

SmartScan utilizes new advances in operating algorithms and system operations to aid in improving image acquisition and patient throughput while helping to increase diagnostic confidence and exam consistency.

• Gynecological • On-board Storage for Peripherals (Max 3 peripherals) • Cardiac

• Musculoskeletal • Wheels • Vascular - Wheel diameter: 12.5cm • Urological - Integrated locking mechanism that

provides rolling lock and optional caster swivel lock

• Small Parts and Superficial ComfortScan, our advanced user-friendly designed software suite helps to maximize productivity. It simplifies every exam you perform.

• Pediatric and Neonatal • Transcranial • Probe Holders, Removable for Cleaning

and Washing Scanning Methods • Gel Holder, Removable for Cleaning

and Washing • Electronic Sector

General Specification • Electronic Convex • Air Filters, Easily Removable • Electronic Linear Dimensions and Weight • Front Handle • Mechanical Volume Sweep

• Height: 135/141 cm • Rear Handle (Option) Transducer Types • Width: 43cm • Drawer (Option) • Sector Phased Array • Depth: 64cm • Probe Cable Hanger (Option) • Convex Array • Weight: approx. 75 kg (165 lb.) • Urology Probe Holder (Option)

LOGIQ P5 DOC0518685 Page 1 of 14

Page 2: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

- B + CFM/M • Real-time Auto Doppler Calculations • Microconvex Array - Realtime Triplex Mode • TruAccess, Raw Data Processing • Linear Array

(B + CFM or PDI/PW) • MpegView • Doppler Pencil - B-Flow + PW (option)

• Key Macro • Bi-plane Microconvex Array • Dual B (B/B) • Network Storage • Real time 4D volume probes - Dual B + CFM or PDI • Auto Focus • Selectable Alternating Modes Operating Modes • Wide Field of View on convex - B/M • B-Mode - B/PW • Simultaneous Display on BE9CS probe

• M-Mode - B + CFM/M • Quick Save • Anatomical M Mode (option) - B + CFM (PDI)/PW • B-Steer • Color Flow Mode (CFM) - 3D – Mode • Virtual Convex with X-Beam - 3D – Color Mode (option) • Power Doppler Imaging (PDI) with

Directional Map • ED Report - B/CW (option)

- B + CFM (PDI)/CW (option) System Options • PW Doppler with High PRF • Multi Image Split Screen • DICOM 3.0 Connectivity • Crossbeam (option) - Live and/or frozen

• ECG • SRI (option) - B + B/CFM or PDI • Coded Contrast Imaging • B-Flow (option) - Independent CINE playback • CW Doppler • Steered CW Doppler (option) - Quad screen format

• Zoom: Write/Read/Pan • Anatomical M-Mode • Dedicated CW Doppler (option) • Colorized Image • LOGIQView • Coded Contrast Imaging (option) - Colorized B • Crossbeam • 3D/4D Volume Modes (option) - Colorized M • SRI System Standard Features - Colorized PW

• B-Flow - Colorized CW (option) • Hard Disk for image storage • Easy 3D (Baby Face, Vascular) • Time line Display • Without compression: • Advanced 3D, with 3D Landscape - Independent Dual B/PW

Display • Raw DICOM: HDD capacity 33.000

images • Real time 4D

Display Formats: • 3rd. probe port • DICOM: Top/ Bottom selectable format (Size:

1/2:1/2; 1/3:2/3; 2/3:1/3) • Flexible Monitor Arm

- Image Only: HDD capacity 71.000 images

• 3-Pedal Foot Switch, with Programmable Functionality Side/Side selectable format

(1/2:1/2; 1/3:2/3; 0:1) • CINE Memory Frames (60 sec) (192MB) >1000 depending on probe, imaging depth, line density and field of view

• Rear Handle Switchable after freeze • Drawer • Probe Cable Hanger Display Annotation • Real-time Triplex Mode at any Depth

and PRF • Urology Probe Holder • Patient Name: First, Last, & Middle

name each store 27 characters. Up to 62 total characters displayed.

• 1-Pedal Foot Switch • Automatic Optimization

• Remote Control Switch - Auto Tissue Optimization: ATO

• Patient ID: 31 characters. Up to 27 total characters displayed.

• PDF report conversion - Auto Spectrum Optimization: ASO

- Auto Color Optimization: ACO • Age, Sex and Birth Date (optional)

• Coded harmonic Imaging Media & Peripheral Options • Hospital Name: 23 characters. • Coded Excitation • Integrated Mounting Kits and Remote

Controls Provided for • Date: 3 Types selectable

MM/DD/YY, DD/MM/YY,

• Virtual Convex • Patient Information Database - B/W Digital thermal printer • Image Archive on Hard Drive and DVD • YY/MM/DD - Color Digital thermal printer • Easy Backup for Media • Time: 2 types selectable - DVD Video Recorder

• 24 hours, 12 hours • Report Designer Display Modes • Gestational Age from

LMP/EDD/GA/BBT • Vascular Calcs • Live and Stored Display Format: Full

size and split screen - both w/ thumbnails. For Still and CINE

• Cardiac Calcs • Probe Name • OB Calcs • Gray Map names • Review Image Format: 4x4, and

"thumbnails". For Still and CINE • Fetal Trending

• Probe Orientation • Multi Gestational Calcs

• Depth Scale Marker • Simultaneous Capability • Hip Dysplasia Calcs • Lateral Scale Marker - B/PW • Gynecological Calcs • Focal Zone Markers - B/CFM or PDI

• Urological Calcs • Image Depth - B/M

LOGIQ P5 Page 2 of 14

• Renal Calcs

Page 3: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

• Languages: English, French, German, Spanish, Italian, Portuguese, Russian, Greek, Swedish, Danish, Dutch, Finnish, Norwegian, Japanese, Chinese

• Conversion to JPEG, AVI (SaveAs) and WMV (MPEGvue) file formats

• Zoom Depth • B-Mode

• Internal Hard Drive Image Storage: 50 GB

- Gain - Dynamic Range

• OB Report Format: 5 Types, Tokyo Univ., Osaka Univ., USA, Europe, and ASUM

• External USB 2.0 Hard Drive support for Import, Export, DICOM Read, SaveAs and MPEGVue

- Imaging Frequency - Edge Enhance - Frame Averaging

• EFBW: 10 types, Japan, USA and Europe (Tokyo Univ., Osaka Univ., Tokyo Shinozuka, JSUM, German, Shephard, Merz, Hadlock/Shephard, Williams, Brenner)

• USB 1.1 / 2.0 Memory Stick support for SaveAs and MPEGVue

- Gray Map - ATO On/Off

• Network Storage support for Import, Export, DICOM Read, SaveAs, MPEGVue

• M-Mode - Gain

- Dynamic Range • 348 Pre-defined Annotations and

User Programmable Libraries/Annotations

Connectivity - Time Scale • Doppler Mode • Ethernet network connection

- Gain • RS-232 serial data output (need a converter cable) • Body Patterns: 140 human types

plus 14 animal types - Angle - Sample Volume Depth and Width • DICOM 3.0 (option)

• Customized Comment Home Position

- Wall Filter - Verify - Velocity and/or Frequency Scale - Print

- Spectrum Inversion - Store Complete User Manual available on board through Help (F1)

- Time Scale - Modality Worklist - PRF - Storage Commitment - Doppler Frequency - Modality Performed Procedure

Step (MPPS) User Manual and Service Manual are included on CD with each system. A printed Manual is available upon request.

• Color Flow Mode - Line Density - Media Exchange - Frame Averaging - Off network / mobile storage

queue - Packet Size - Color Scale - Query/Retrieve; supported on

Centricity and other compatible vendors

CINE Memory/Image Memory - Color Velocity Range and Baseline - Color Threshold Marker • CINE Memory: 192MB - Color Gain • Dual Image CINE Display - Structured Reporting; compatible

with LOGIQworks - PDI • Quad Image CINE Display - Color Scale Inversion • CINE Gauge and CINE Image

Number Display - Public SR Template

- Color Doppler Frequency - Media Store of SR • TGC Curve • CINE Review Loop - iLinq capability • Acoustic Frame Rate • CINE Review Speed: 20 steps (10, 20,

30, 40, 50, 60, 70, 80, 90, 100, 150, 200, 300, 400, 500, 600, 700, 800, 900, 1000%)

• Cine Frame Number Scanning Parameters • VCR Counter • VCR Status • Digital Beamformer • VCR Playback Counter • Selectable CINE Sequence for CINE

Review • 2560 processing channels

• Body Pattern • Frame Rate: > 700 depending on • Application Name • Measurements, Calculations and

Annotations on CINE Playback preset

• Measurement Results • Displayed Imaging Depth: 0 – 30 cm • Operator Message • Scrolling Timeline Memory • Minimum Depth of Field: 0 –2cm • Displayed Acoustic Output • Cine Capture Function (probe dependent)

- TIS: Thermal Index Soft Tissue • Maximum Depth of Field: 0 – 30 cm - TIC: Thermal Index Cranial (Bone) Image Storage (probe dependent) - TIB: Thermal Index Bone • Transmission Focus • On-board database of patient

information from past exams - MI: Mechanical Index 1 – 8 Focus Points selectable (probe

and application dependent) • % of Power output • Storage Format: DICOM/Raw Data • Biopsy Guide Line and/or Zone • Continuous Dynamic Receive Focus/Aperture

• DICOM Still Image Storage Size: • Heart Rate - Gray Image: ~300K to ~1.3 MB • Multi-Frequency/Wideband Technology

- Color Image: ~900K to 1.9 MB General System Parameters • Multiframe • Frequency Range 1 to 15MHz • Display Format: Full size, 4x4, and

"thumbnails" System Setup • 256 Shades of Gray • Up to 197 db processing dynamic

range • Live image and stored image side-

by-side display • Pre-programmable Categories • User Programmable Preset

Capability • 16,777,216 Hues of Color • CD-R storage: 650, 700 MB • DVD storage: -R 4,7GB • Factory Default Preset Data

LOGIQ P5 Page 3 of 14

Page 4: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

Anatomical M-Mode • Adjustable Field of View (FOV) up to 170 degree depending on probe

• Wall Filter:4 steps depend on probe/application M-mode cursor adjustable at any

plane • Image Reverse: Right/Left • Scanning Size (FOV or Angle): probe dependent • Image Rotation: 4 steps Can be activated from a CINE loop

from a live or stored image • CF/PDI Vertical Size (mm) of ROI: default pre-settable

Rotation: 0°, 90°, 180°, 270° Available with Color Flow Mode

• CF/PDI Center Depth (mm) of ROI: default pre-settable

B-Mode M & A capability • B/M Acoustic Output: 0 – 100%,

2% step • CF/PDI Frequency: Up to 3 steps, depend on probe

PW/CW-Mode • Maximum and Minimum Velocity

Scales • Image Reverse: On/Off

• Color Map: 20 types depend on application

• B Color: 10 types - Max: 10 m/sec • Thermal Index: TIC, TIS, TIB • Transparent: 5 steps - Min: 5 cm/sec • Softener: 4 steps • Color Threshold: 0 – 100 %, 5 % step • Gray Scale Map: 4 types • Focus Number: 8 steps • Auto Line Density: On/Off pre-

settable • Dynamic Range: 24 – 60, 4 dB step • Focus Width: 3 types • Base Line: 5 – 95 %, 11 steps • Range Focus: On/Off • PW/CF Ratio: 1, 2, 4 • SV Gate: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 12,

14, 16 mm • Compression: 0.5 – 1.5, 0.1 step • Accumulation: 8 steps • Line Density: 4 steps • Angle Correct: +/- 90°, 1° step • Line Density Zoom: 4 steps Power Doppler Imaging • Spectral Color: 6 types • Suppression: 6 steps • PDI Map: 13 types • PW Sweep Speed: 8 steps • Frame Average: 8 steps • CF/PDI Flash Suppression: 2 steps • Invert: On/Off • Edge Enhance: 6 steps • CF/PDI Focus Depth: default pre-

settable for 0 – 100 % of ROI in depth, 10 % step

• M/PW Display Format: V-1/3B, V-1/2B, V-2/3B, H-1/2B, H-1/4B, TL only • Scanning Size (FOV or Angle):

probe dependent, see probe specifications

• Duplex: On/Off(PW only) • CF/PDI Acoustic Output: 0 – 100%,

10% step • PW Acoustic Output: 0 – 100 %,

10 % step • Gray Scale Map: 23 types • CF/PDI Angle Steer: 0, +/- 20° • Clear Map: 23 types • Spectral Averaging: 4 steps • Packet Size: 5 – 16, dependent on

probe/application • Tint Map: 10 types • Time Resolution: 4 steps

• PW/CF Ratio: 1, 2, 4 • Gain: 0 – 98 dB, 2 dB step • Spatial Filter: 6 steps • Rejection: 15 steps • Dynamic Range: 30 – 120 dB, 3 dB

step • Frame Average: 7 steps • Gain: 0 – 32 dB, 1 dB step • PRF: 280 Hz – 19600 Hz • Wall Filter: 5 – 1500 Hz, 22 steps,

depend on probe/application • Depth: 1 – 30 cm, 1 cm step,

depend on probe. • Power Threshold: 0 – 100 %, 5 % step • PW Angle Steer: 0, +/- 5, 10, 15, 20° • Rejection: 6 steps

• Gain: 0 – 40 dB, 0.5 dB step • PRF: 640 – 29800 Hz with PW, 50000 Hz with CW

• Frequency: Up to 5 steps, depend on probe • Wall Filter: 7 steps depend on

probe/application • Sample Volume Depth: 29 steps default pre-settable

• CF/PDI Frequency: Up to 3 steps,

depend on probe • Auto Line Density: On/Off pre-

settable • CW-Mode (option) is Available on the Following Probes • Auto Line Density: On/Off pre-

settable • Steered Linear: +/- 15° - 3S • Transparent: 5 steps - 5S Color Flow Mode • Invert: On/Off - 7S

• Base Line: 0 – 100 %, 10 % step • Accumulation: 8 steps - 3Sp • Invert: On/Off - 5Sp • CF/PDI Focus Depth: default pre-

settable for 0 – 100 % of ROI in depth, 10 % step

M-Mode - P2D - P6D • Sweep Speed: 8 steps

• M Color: 10 types • CF/PDI Flash Suppression: 2 steps Coded Harmonic Imaging • M/PW Display Format: V-1/3B, V-1/2B, V-2/3B, H-1/2B, H-1/4B, TL only • CF/PDI Acoustic Output: 0 – 100%,

10% step • Available on the Following Probes • B/M Acoustic Output: 0 – 100 %, 2 %

step - 3.5C • CF/PDI Angle Steer: 0, +/- 20° - 3.5CS • Packet Size: 5 – 16, dependent on

probe/application • Rejection: 6 steps - 3CRF • Dynamic Range: 30 – 120 dB, 3 dB

step - 4C • Line Density: 5 steps - 5CS • Line Density Zoom: 5 steps • Edge Enhance: 6 steps - E8C • Frame Average: 7 steps • Gray Scale Map: 23 types - E8CS • PRF: 280 Hz – 19600 Hz • Clear Map: 23 types - 8C • Spatial Filter: 6 steps • M Gain: 0 – 98 dB, 2 dB step - ERB • Gain: 0 – 40 dB, 0.5 dB step

LOGIQ P5 Page 4 of 14

Page 5: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

- BE9CS - ERB - Linear • Accumulation: 8 steps • SRI-HD - 3S - 4C

- 5S - 5CS The LOGIQ P5 is designed for compatibility with commercially available ultrasound contrast agents. Because the availability of these agents is subject to government regulation and approval, product features intended for use with these agents may not be commercially marketed nor made available before the contrast agent is cleared for use. Contrast related product features are enabled only on systems for delivery to an authorized country or region of use. GE Medical Systems makes no claims concerning the safety or effectiveness of contrast agents.

- 7S - 4D3C-L - 3Sp - 3CRF

- 5Sp - 8L Automatic Optimization - 9L • Optimize B-Mode, B-Flow image to

improve contrast resolution. - 11L - i12L • Selectable amount of contrast

resolution improvement (low, medium, high)

- i739 - T739 - 4D3C-L • Auto-TGC in B-Mode and Color –

adjusts overall and axial gain - 4DE7C - 4D8C • Optimize Spectral Waveform –

adjusts baseline, invert, PRF (on live image), and angle correction

• Softener: 4 steps • Line Density: 4 steps • Line Density Zoom: 4 steps • Algorithm works on focal zone/

number and depth changes • Suppression: 6 steps LOGIQView (option) • Edge Enhance: 6 steps • Available on stored or live image • Available on all probes • Gray Scale Map: 23 types • Available in B-Mode, B-Flow, PW

Doppler, Color • Extended Field of View Imaging

• Clear Map: 23 types • For use in B-Mode • Tint Map: 10 types • LOGIQView Status • Gain: 0 – 98 dB, 2 dB step • Auto detection of scan direction B-Flow (option) • Dynamic Range: 30 – 120 dB, 3 dB

step • Pre or post-process zoom up to 10X • Available on 8L, 9L, 10L, 11L, i739

and T739 Probes • Rotation • Rejection: 6 steps • Auto best fit on monitor • Background: On/Off • Auto Line Density: On/Off pre-

settable • Measurements in B-Mode • Sensitivity/PRI: 14 steps • Up to 60 cm scan length • Line Density: 4 steps • Frequency: Up to 5 steps, depend on

probe • Edge Enhance: 6 steps Easy 3D (option) • Frame Average: 8 steps • Gray Scale Map: 23 types Colorize Coded Excitation • Clear Map: 23 types Threshold (Opacification) • Available on the following probes

the Following Probes • Tint Map: 10 types Render • Dynamic Range: 30 – 120 dB, 3 dB

step Texture - 11L - 12L Gray Surface • Rejection: 6 steps - 8C Scalpel • Gain: 0 – 98 dB, 2 dB step - E8C • Auto Line Density: On/Off pre-

settable Auto Movie

- E8CS Undo - BE9CS • Accumulation: 8 steps Reset - ERB Coded Contrast Imaging (option) Advanced 3D (option) Virtual Convex • Coded Harmonic Angio: Available on 3.5C, 3.5CS, 4C and 5CSprobes

• Acquisition of Color data • Provides a Convex Field of View for Convex, Linear and Sector probes • Automatic rendering

- Tissue Background Selection: 4 steps

• 3D Landscape technology • Available on the Following Probes • 3D Movie - 8L • TruAgent Detection: Available on

3.5C, 3.5CS, 4C and 5CS probes • Main Mode - 9L - 10L - 2 frequencies on 3.5C and 3.5CS, 1

on 4C and 5CS probes CrossBeam (option) - 11L • Provides Spatial Compounding - 12L • Coded Phase Inversion: • Available on the linear and convex

probes - i12L 2 types, Available on 3.5C and 3.5CS

probes - T739 - 3CRF - I739 3 type, Available on 4C, 5CS probes - 4C - 3S - Tissue Background Selection: 4

steps - 5CS - 5S - 8C - 7S • Time Trigger Scan: 0.3, & 0.5 – 10

sec, 0.5 sec step - E8C - 3Sp - E8CS - 5Sp • Time Intensity Curve Analysis

LOGIQ P5 Page 5 of 14

- 8L

Page 6: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

- Gain - 9L - Post Gain - TGC - 11L - Compression - Acoustic Output - I12L - Rejection - Transmission Focus Position - I739 - Spectral Averaging - Transmission Focus Number - T739 - Colorized D - Transmission Focus Width - BE9CS - Display Format - Imaging Frequency - 4D3C-L - Sweep Speed - Edge Enhancement - 4DE7C • Color Flow Mode - Line Density Control - 4D8C - ACO (Auto Color Optimization) - Live Anatomical M-mode - ERB - Base Line Shift

- Color Map • PW/CW-Mode • Provides 3, 5 or 7 angles for

Compounding on Linear probes, 3 or 5 angles on Convex probes

- Post Gain - Gain - Frame Averaging - Dynamic Range - CFM Display Threshold - Acoustic Output • Compatible with Side by Side

Display - Angle Correct (PW mode)

- Doppler Frequency - Quick Angle Correct (PW mode) - Velocity Scale (PRF) • Compatible with: Color mode,

Timeline mode, SRI-HD, Coded Harmonic Imaging, Virtual Convex

- Auto Angle Correct (PW mode) - PW/CF Ratio - Spectral Invert for Color and

Doppler - Wall Filter - Time Resolution • • 3D reconstruction from a stored

CINE loop SRI-HD (option) - Sample Volume Gate for PW-Mode • Anatomical M-Mode • High Definition Speckle Reduction

Imaging • Accumulation Length • Cine Capture Depth • Provides 6 levels of speckle

reduction • Color Flow Mode Physiological Input Panel (Option) - CFM Gain • Side by Side Display

- CFM Velocity Scale (PRF) • Physiological Input • Compatible with ALL linear, convex and sector transducers - Acoustic Output - ECG, 1 channel

- CFM Frequency • Dual R-Trigger • Compatible with ALL scanning mode - Wall Filter • Pre-settable ECG R Delay Time

- Packet Size • Pre-settable ECG Position • Compatible with Side by Side Display - CFM Spatial Filter • Adjustable ECG Gain Control

- CFM Line Density • Automatic Heart Rate Display - CFM Regression Filter Real Time 4D (option) - Accumulation Measurements / Calculations • Acquisition Modes:

- Realtime 4D Bmode Post-Processing w/ TruAccess (Raw Data)

- Static 3D Bmode General B-Mode • Visualization Modes: • Depth & Distance - 3D Rendering (diverse surface and

intensity projection modes) • SRI-HD – Selectable level of Speckle

Reduction Imaging • Circumference (Ellipse / Trace) • Area (Ellipse / Trace) - Sectional Planes (3 Section planes

perpendicular to each other) • Max Read Zoom to 16x • Volume (Ellipsoid) • B/M-Mode • % Stenosis (Area or Diameter) • Render Mode: - ATO (Auto Tissue Optimization) • Angle between two lines - Surface texture, Surface Smooth,

max-, min- and X-ray (average intensity projection), Gradient, Inversion, Glass Body, Mix Mode of two render Modes

- Image Reverse - Image Rotation General M-Mode - Gray Map • M-Depth - Colorized B and M • Distance - Gain • Time • Curved 3point Render start - Dynamic Range • Slope • 3D Movie - TGC • Heart Rate • Scalpel: 3D Cut tool - Compression

• Display Format: - Rejection General Doppler - Quad: A-/B-/C-Plane/3D - Frame Averaging Measurements/Calculations - Dual: A-Plane/3D - B Softener • Velocity - Single: 3D or A- or B- or C-Plane - Suppression • Time - Sweep Speed for M-Mode • A/B Ratio (Velocities / Frequency

Ratio - Anatomical M Mode Pre-Processing • PW/CW-Mode

• PS (Peak Systole) - ASO (Auto Spectral Optimization) • Write Zoom up to 8x • ED (End Diastole) - Base Line Shift • B/M-Mode

LOGIQ P5 Page 6 of 14

• PS/ED (PS/ED Ratio) - Gray Map

Page 7: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

• ED/PS (ED/PS Ratio) • SYS DECA (Systolic Distal External Carotid Artery)

- AC, HC • AT (Acceleration Time) • Calculations and Ratios • ACCEL (Acceleration) • DIAS DECA (Diastolic Distal External

Carotid Artery) - FL/BPD

• TAMAX (Time Averaged Maximum Velocity

- FL/AC • SYS PECA (Systolic Proximal External

Carotid Artery) - FL/HC

• Volume Flow (TAMEAN and Vessel Area)

- HC/AC • DIAS PECA (Diastolic Proximal

External Carotid Artery) - CI (Cephalic Index)

• Heart Rate - AFI (Amniotic Fluid Index) • PI (Pulsatility Index) • VERT (Systolic Vertebral Velocity) • Measurements / Calculations by:

Jeanty, Merz, Tokyo University, Mercer, Hansmann, Erickson, Hill, Shephard, Hadlock, Hohler, Campbell

• RI (Resistivity Index) • SUBCLAV (Systolic Subclavian Velocity)

• Summary Reports Real-time Doppler Auto • Mean IMT Measurement Tools • Fetal Graphical Trending Measurements / Calculations • Growth Percentiles • PS (Peak Systole)

• Multi-Gestational Calculations (4) • ED (End Diastole) Cardiac • Fetal Qualitative Description (Anatomical survey)

• MD (Minimum Diastole) Measurements/Calculations • PI (Pulsatility Index)

• Fetal Environmental Description (Biophysical profile)

• RI (Resistivity Index) B-Mode Measurements • AT (Acceleration Time) • Aorta • Programmable OB Tables • ACC (Acceleration)

Aortic Root Diameter (Ao Root Diam) • Over 20 selectable OB Calcs • PS/ED (PS/ED Ratio) • Expanded Worksheets Aortic Arch Diameter (Ao Arch Diam) • ED/PS (ED/PS Ratio) Ascending Aortic Diameter (Ao Asc) • HR (Heart Rate) GYN Measurements/Calculations • TAMAX (Time Averaged Maximum

Velocity) Descending Aortic Diameter (Ao Desc Diam) • Right Ovary Length, Width, Height

• Left Ovary Length, Width, Height • PVAL (Peak Velocity Value) Aorta Annulus Diameter (Ao Annulus Diam)

• Uterus Length, Width, Height • Volume Flow (TAMEAN and Vessel Area) • Ovarian Volume

Aorta Isthmus (Ao Isthmus) • ENDO (Endometrial thickness) Aorta *** (Ao st junct) • Ovarian RI OB Measurements/Calculations

• Uterine RI • Aortic Valve • Gestational Age by: • Follicular measurements Aortic Valve Cusp Separation (AV Cusp) - GS (Gestational Sac) • Summary Reports Aortic Valve Area Planimetry (AVA Pl

ani met ry) - CRL (Crown Rump Length)

- FL (Femur Length) Vascular Measurements/Calculations

- BPD (Biparietal Diameter) *** ( Trans AVA) - AC (Abdominal Circumference) • Left Atrium

• SYS DCCA (Systolic Distal Common Carotid Artery)

- HC (Head Circumference) Left Atrium Diameter (LA Diam) - APTD x TTD (Anterior/Posterior Trunk Diameter by Transverse Trunk Diameter)

LA Length (LA Major) • DIAS DCCA (Diastolic Distal Common Carotid Artery) LA W idth (LA Minor)

Left Atrium Diameter to AoRoot Diameter Ratio (LA/Ao Ratio)

• SYS MCCA (Systolic Mid Common Carotid Artery)

- LV (Length of Vertebra) - FTA (Fetal Trunk Cross-sectional

Area) • DIAS MCCA (Diastolic Mid Common Carotid Artery)

Left Atrium Area (LAA(d), LAA(s)) - HL (Humerus Length) Left Atrium Volume, Single Plane,

Method of Disk (LAEDV A2C, LAESV A2C ) (LAEDV A4C, LAESV A4C)

• SYS PCCA (Systolic Proximal Common Carotid Artery)

- BD (Binocular Distance) - FT (Foot Length)

• DIAS PCCA (Diastolic Proximal Common Carotid Artery)

- OFD (Occipital Frontal Diameter) • Left Ventricle - TAD (Transverse Abdominal

Diameter) Left Ventricle Mass (LVPWd, LVPWs) • SYS DICA (Systolic Distal Internal Carotid Artery) - TCD (Transverse Cerebellum

Diameter) Left Ventricle Volume, Teichholz/Cubic (LVIDd, LVI Ds) • DIAS DICA (Systolic Distal Internal

Carotid Artery) - THD (Thorax Transverse Diameter) Left Ventricle Internal Diameter (LVIDd, LVI Ds) • SYS MICA (Systolic Mid Internal

Carotid Artery) - TIB (Tibia Length) - ULNA (Ulna Length)

Left Ventricle Length (LVLd, LVLs) • DIAS MICA (Diastolic Mid Internal Carotid Artery)

• Estimated Fetal Weight (EFW) by: Left Ventricle Outflow Tract Diameter (LVOT Diam)

- AC, BPD • SYS PICA (Systolic Proximal Internal

Carotid Artery) - AC, BPD, FL - AC, BPD, FL, HC Left Ventricle Posterior Wall Thickness

(LVPWd, LVPWs) • DIAS PICA (Diastolic Proximal Internal Carotid Artery)

- AC, FL - AC, FL, HC Left Ventricle Length (LV Major)

LOGIQ P5 Page 7 of 14

Page 8: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

Right Atrium Volume, Systolic, Single Plane, Method of Disk (RAAs)

Left Ventricle Volume, Teichholz/Cubic (LVIDd, LVI Ds)

Left Ventricle Width (LV Minor) Left Ventricle Outflow Tract Area (LVOT)

• Right Ventricle Left Ventricle Internal Diameter (LVIDd, LVI Ds)

Left Ventricle Area, Two Chamber/Four Chamber/Short Axis (LVA (d), LVA (s)) Right Ventricle Outflow Tract Area

(RVOT Planimetry) Left Ventricle Posterior Wall Thickness (LVPWd, LVPWs)

Left Ventricle Endocardial Area, Width (LVA (d), LVA(s)) Left Pulmonary Artery Area (LPA Area)

Right Pulmonary Artery Area (RPA Area) Left Ventricle Ejection Time (LVET) Left Ventricle Epicardial Area, Length (LVAepi (d), LVAepi (s)) Right Ventricle Internal Diameter

(RVIDd, RVIDs) Left Ventricle Pre-Ejection Period (LVPEP) Left Ventricle Mass Index (LVPWd,

LVPWs) Right Ventricle Diameter, Length (RVD Ma)

Interventricular Septum (IVS) Left Ventricle Internal Diameter (LVI D) Ejection Fraction, Teichholz/Cube

(LVIDd, LVIDs) Right Ventricle Diameter, Width (RVD Mi)

Left Ventricle Posterior Wall Thickness (LVPW) Left Ventricle Posterior Wall Fractional

Shortening (LVPWd, LVPWs) • Mitral Valve Right Ventricle Wall Thickness (RVAWd, RVAWs) E-Point-to-Septum Separation (EPSS) Left Ventricle Stroke Index,

Teichholz/Cube (LVIDd, LVIDs, and Body Surface Area)

Right Ventricle Outflow Tract Diameter (RVOT Diam)

Mitral Valve Leaflet Separation (D-E Excursion)

Left Pulmonary Artery (LPA) Mitral Valve Anterior Leaflet Excursion (D-E Excursion)

Left Ventricle Fractional Shortening (LVIDd, LVIDs) Main Pulmonary Artery (MPA)

Right Pulmonary Artery (RPA) Mitral Valve D-E Slope (D-E Slope) Left Ventricle Stroke Volume, Teichholz/Cubic (LVIDd, LVIDs) • System Mitral Valve E-F Slope (E-F Slope)

• Pulmonic Valve Interventricular Septum Thickness (IVSd, IVSs)

Left Ventricle Stroke Index, Single Plane, Two Chamber, Method of Disk (LVI Dd, LVIDs, LVSd, LVSs)

QRS complex to end of envelope (Q-to-PV close) Inferior Vena Cava

• Right Ventricle Pulmonary Artery Diameter (MPA) Left Ventricle Stroke Index, Single Plane, Four Chamber, Method of Disk (LVI Dd, LVIDs, LVSd, LVSs)

Systemic Vein Diameter (Systemic Diam)

Right Ventricle Internal Diameter (RVIDd, RVIDs)

Patent Ductus Arterosis Diameter (PDA Diam)

Right Ventricle Wall Thickness (RVAWd, RVAWs)

Left Ventricle Stroke Index, Bi-Plane, Bullet, Method of Disk (LVAd, LVAs)

Pericard Effusion (PEs) Right Ventricle Outflow Tract Diameter (RVOT Diam)

Interventricular Septum (IVS) Patent Foramen Ovale Diameter (PFO Diam)

Left Ventricle Internal Diameter (LVI D) Right Ventricle Ejection Time (RVET) Left Ventricle Posterior Wall Thickness

(LVPW) Ventricular Septal Defect Diameter (VSD Diam)

Right Ventricle Pre-Ejection Period (RVPEP) • Mitral Valve

Interventricular Septum (IVS) Fractional Shortening (IVSd, IVSs)

Velocity Circumferential Fiber Shortening (Vcf)

Mitral Valve Annulus Diameter (MV Ann Diam)

• Tricuspid Valve • System E-Point-to-Septum Separation (EPSS) Tricuspid Valve Area (TV Panimetry) Interventricular Septum Thickness

(IVSd, IVSs) Mitral Valve Area by Pressure Half Time (MVA By PHT) Tricuspid Valve Annulus Diameter (TV

Annulus Diam) Pericard Effusion (PE(d)) Mitral Valve Area Planimetry (MVA Planimetry) Interventricular Septum (IVS) Fractional

Shortening (IVSd, IVSs) M-Mode Measurements • Pulmonic Valve • Tricuspid Valve • Aorta Pulmonic Valve Area (PV Planimetry)

QRS complex to end of envelope (Q-to-TV close)

Pulmonic Valve Annulus Diameter (PV Annulus Diam)

Aortic Root Diameter (Ao Root Diam) • Aortic Valve

Pulmonic Diameter (Pulmonic Diam) Aortic Valve Diameter (AV Diam) Doppler Mode Measurements • Right Atrium Aortic Valve Cusp Separation (AV Cusp) • Aortic Valve Right Atrium Diameter, Length (RAD

Ma) Aortic Valve Ejection Time (LVET)

• Left Atrium Aortic Insufficiency Mean Pressure Gradient (AR Trace) Right Atrium Diameter, Width (RAD Mi) Left Atrium Diameter to AoRoot

Diameter Ratio (LA/Ao Ratio) Right Atrium Area (RAA) Aortic Insufficiency Peak Pressure Gradient (AR Vmax) Right Atrium Volume, Single Plane,

Method of Disk (RAAd) Left Atrium Diameter (LA Diam)

• Left Ventricle Aortic Insufficiency End Diastole

LOGIQ P5 Page 8 of 14

Page 9: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

Mitral Valve Acceleration Time/Deceleration Time Ratio (MVAcc/Dec Time)

Pressure Gradient (AR Trace) Aortic Insufficiency Mean Velocity (AR Trace) Aortic Insufficiency Mean Square Root Velocity (AR Trace) Aortic Insufficiency Velocity Time Integral (AR Trace) Aortic Valve Mean Velocity (AV Trace) Aortic Valve Mean Square Root Velocity (AV Trace) Aortic Valve Velocity Time Integral (AV Trace) Aortic Valve Mean Pressure Gradient (AV Trace) Aortic Valve Peak Pressure Gradient (AR Vmax) Aortic Insufficiency Peak Velocity (AR Vmax) Aortic Insufficiency End-Diastolic Velocity (AR Trace) Aortic Valve Peak Velocity (AV Vmax) Aortic Valve Peak Velocity at Point E (AV Vmax) Aorta Proximal Coarctation (Coarc Pre-Duct) Aorta Distal Coarctation (Coarc Post-Duct) Aortic Valve Insufficiency Pressure Half Time (AR PHT) Aortic Valve Flow Acceleration (AV Trace) Aortic Valve Pressure Half Time (AV Trace) Aortic Valve Acceleration Time (AV Acc Ti me) Aortic Valve Deceleration TIme (AV Trace) Aortic Valve Ejection Time (AVET) Aortic Valve Acceleration to Ejection Time Ratio (AV Acc Time, AVET) Aortic Valve Area according to PHT

• Left Ventricle Left Ventricle Outflow Tract Peak Pressure Gradient (VLOT Vmax) Left Ventricle Outflow Tract Peak Velocity (LVOT Vmax) Left Ventricle Outflow Tract Mean Pressure Gradient (LVOT Trace) Left Ventricle Outflow Tract Mean Velocity (LVOT Trace) Left Ventricle Outflow Tract Mean Square Root Velocity (LVOT Trace) Left Ventricle Outflow Tract Velocity

Time Integral (LVOT Trace) Left Ventricle Ejection Time (LVET) Cardiac Output by Aortic Flow (AVA Pl ani met ry, AV Trace) Stroke Volume Index by Mitral Flow

(MVA Planimetry, MVTrace) Stroke Volume Index by Aortic Flow (AVA Planimetry, AV Trace) Mitral Valve Area from Continuity

Equation (MVAPlanimetry, LVOT Vmax, MV Vmax)

• Mitral Valve Mitral Valve Regurgitant Flow Acceleration (MR Trace) • Pulmonic Valve

Pulmonic Insufficiency Peak Pressure Gradient (PR Vmax)

Mitral Valve Regurgitant Mean Velocity (MR Trace)

Pulmonic Insufficiency End-Diastolic Pressure Gradient (PRTrace)

Mitral Regurgitant Mean Square Root Velocity (MR Trace)

Pulmonic Valve Peak Pressure Gradient (PV Vmax)

Mitral Regurgitant Mean Pressure Gradient (MR Trace)

Pulmonic End-Diastolic Pressure Gradient (PR Trace)

Mitral Regurgitant Velocity Time Integral (MR Trace)

Pulmonic Insufficiency Peak Velocity (PR Vmax)

Mitral Valve Mean Velocity (MR Trace) Mitral Valve Mean Square Root Velocity (MR Trace) Pulmonic Insufficiency End-Diastolic

Velocity (Prend Vmax) Mitral Valve Velocity Time Integral (MR Trace) Pulmonic Valve Peak Velocity (PV

Vmax) Mitral Valve Mean Pressure Gradient (MR Trace) Pulmonic End-Diastolic Velocity (PV

Trace) Mitral Regurgitant Peak Pressure Gradient (MR Vmax) Pulmonary Artery Diastolic Pressure

(PV Trace) Mitral Valve Peak Pressure Gradient (MR Vmax) Pulmonic Insufficiency Mean Pressure

Gradient (PR Trace) Mitral Regurgitant Peak Velocity (MR Vmax) Pulmonic Valve Mean Pressure

Gradient (PV Trace) Mitral Valve Peak Velocity (MR Vmax) Pulmonic Insufficiency Mean Velocity (PR Trace)

Mitral Valve Velocity Peak A (MV A Velocity)

Pulmonic Insufficiency Mean Square Root Velocity(PR Trace)

Mitral Valve Velocity Peak E (MV E Velocity)

Pulmonic Insufficiency Velocity Time Integral (PR Trace)

Mitral Valve Area according to PHT (MV PHT)

Pulmonic Valve Mean Velocity (PV Trace)

Mitral Valve Flow Deceleration (MV Trace)

Pulmonic Valve Mean Square Root Velocity (PV Trace)

Mitral Valve Pressure Half Time (PV PHT)

Pulmonic Valve Velocity Time Integral (PV Trace)

Mitral Valve Flow Acceleration (MV Trace)

Pulmonic Insufficiency Pressure Half Time (PR PHT)

Mitral Valve E-Peak to A-Peak Ratio (A-C and D-E) (MV E/ARatio)

Pulmonic Valve Flow Acceleration (PV Acc Time)

Mitral Valve Acceleration Time (MV Acc Time)

Pulmonic Valve Acceleration Time (PV Acc Time)

Mitral Valve Deceleration Time (MV Dec Time)

Pulmonic Valve Ejection Time (PVET) Mitral Valve Ejection Time ((MV Trace) Pulmonic Valve Pre-Ejection Period (PVPEP)

Mitral Valve A-Wave Duration (MV A Dur)

QRS complex to end of envelope (Q-to-Mitral Valve Time to Peak (MV Trace)

LOGIQ P5 Page 9 of 14

Page 10: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

PV close) Pulmonic Valve Acceleration to Ejection TIme Ratio (PV Acc Time, PVET) Pulmonic Valve Pre-Ejection to Ejection Time Ratio (PVPEP, PVET)

• Right Ventricle Right Ventricle Outflow Tract Peak Pressure Gradient (RVOT Vmax) Right Ventricle Systolic Pressure (RVOT Vmax) Right Ventricle Outflow Tract Peak Velocity (RVOT Vmax) Right Ventricle Diastolic Pressure (RVOT Trace) Right Ventricle Outflow Tract Velocity Time Integral (RVOTTrace) Right Ventricle Ejection Time (RV Trace) Stroke Volume by Pulmonic Flow (RVOTPlanimetry, RVOTTrace) Right Ventricle Stroke Volume Index by Pulmonic Flow (RVOT Planimetry, RVOT Trace)

• System Pulmonary Artery Peak Velocity (PV Vmax) Pulmonary Vein Velocity Peak A (reverse) (P Vein A) Pulmonary Vein Peak Velocity (P Vein D, P Vein S) Systemic Vein Peak Velocity (PDA Diastolic, PDA Systolic) Ventricular Septal Defect Peak Velocity (VSD Vmax) Atrial Septal Defect (ASD Diastolic, ASD Systolic) Pulmonary Artery Velocity Time Integral (PV Trace) Systemic Vein Velocity Time Integral (PDA Trace) Pulmonary Vein A-Wave Duration (P Vein A Dur) IsoVolumetric Relaxation Time (IVRT) IsoVolumetric Contraction Time (IVCT) Pulmonary Vein S/D Ratio (P Vein D, P Vein S) Ventricular Septal Defect Peak Pressure Gradient (VSD Vmax) Pulmonic-to-Systemic Flow Ratio (Qp/Qs)

• Tricuspid Valve Tricuspid Regurgitant Peak Pressure Gradient (TR Vmax) Tricuspid Valve Peak Pressure Gradient

(TV Vmax) Tricuspid Regurgitant Peak Velocity (TR Vmax) Tricuspid Valve Peak Velocity (TV Vmax) Tricuspid Valve Velocity Peak A (TV A Velocity) Tricuspid Valve Velocity Peak E (TV E Velocity) Tricuspid Regurgitant Mean Pressure Gradient (TR Trace) Tricuspid Valve Mean Pressure Gradient (TV Trace) Tricuspid Regurgitant Mean Velocity (TR Trace) Tricuspid Regurgitant Mean Square Root Velocity (TR Trace) Tricuspid Regurgitant Velocity Time Integral (TR Trace) Tricuspid Valve Mean Velocity (TV Trace) Tricuspid Valve Mean Square Root Velocity (TV Trace) Tricuspid Valve Velocity Time Integral (TV Trace) Tricuspid Valve Time to Peak (TV Acc/Dec Time) Tricuspid Valve Ejection Time (TV Acc/Dec Time) Tricuspid Valve A-Wave Duration (TV A Dur) QRS complex to end of envelope (Q-to-TV close) Tricuspid Valve Pressure Half Time (TV PHT) Stroke Volume by Tricuspid Flow (TV Planimetry, TV Trace) Tricuspid Valve E-Peak to A-Peak Ratio (TV E/A Velocity)

Color Flow Mode Measurements • Aortic Valve

- Proximal Isovelocity Surface Area: Regurgitant Orifice Area (AR Radius)

- Proximal Isovelocity Surface Area: Radius of Aliased Point (AR Radius)

- Proximal Isovelocity Surface Area: Regurgitant Flow (AR Trace)

- Proximal Isovelocity Surface Area: Regurgitant Volume Flow (AR Trace)

Proximal Isovelocity Surface Area: Aliased Velocity (AR Vmax)

• Mitral Valve Proximal Isovelocity Surface Area:

Regurgitant Orifice Area (MR Radius) Proximal Isovelocity Surface Area: Radius of Aliased Point (MR Radius) Proximal Isovelocity Surface Area: Regurgitant Flow (MR Trace) Proximal Isovelocity Surface Area: Regurgitant Volume Flow (MR Trace) Proximal Isovelocity Surface Area: Aliased Velocity (MR Vmax)

Combination Mode Measurements • Aortic Valve

- Aortic Valve Area (Ao Root Diam, LVOT Vmax, AV Vmax)

- Aortic Valve Area by Continuity Equation by Peak Velocity (Ao Root Diam, LVOT Vmax, AV Vmax)

Stroke Volume by Aortic Flow (AVA Pl ani met ry, AV Trace) Cardiac Output by Aortic Flow (AVA Planimetry, AV Trace, HR) Aortic Valve Area by Continuity Equation VTI (Ao Root Diam, LVOT Vmax, AV Trace)

• Left Ventricle Cardiac Output, Teichholz/Cubic (LVIDd, LVI Ds, HR) Cardiac Output Two Chamber, Single Plane, Area-Length/ Method of Disk(Simpson) (LVAd, LVAs, HR) Cardiac Output Four Chamber, Single Plane, Area-Length/ Method of Disk(Simpson) (LVAd, LVAs, HR) Ejection Fraction Two Chamber, Single Plane, Area-Length/ Method of Disk(Simpson) (LVAd, LVAs) Ejection Fraction Four Chamber, Single Plane, Area-Length/ Method of Disk(Simpson) (LVAd, LVAs) Left Ventricle Stroke Volume, Single Plane, Two Chamber/Four Chamber, Area-Length (LVAd, LVAs) Left Ventricle Stroke Volume, Single Plane, Two Chamber/Four Chamber, Method of Disk(Simpson) (LVIDd, LVIDs, LVAd, LVAs) Left Ventricle Volume, Two Chamber/Four Chamber, Area-Length (LVAd, LVAs) Ejection Fraction, Bi-Plane, Method of Disk (LVAd, LVAs, 2CH, 4CH) Left Ventricle Stroke Volume, Bi-Plane, Method of Disk (LVAd, LVAs, 2CH, 4CH)

LOGIQ P5 Page 10 of 14

Page 11: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

- Physical Foot Print : 38 x 10 mm - Doppler Frequency: 2.5, 3.3 MHz Left Ventricle Volume, Bi-Plane, Method of Disk (LVAd, LVAs, 2CH, 4CH) - B-mode Imaging Frequency : 6.0, - Biopsy Guide Available: Single Angle,

Reusable - 8.0, 10.0 MHz Left Ventricle Stroke Index, Single Plane, Two Chamber/Four Chamber, Area-Length (LVSd, LVSs, and BSA)

- Harmonic Frequency : 8.0, 10.0, - 12.0 MHz • 3.5C Wide Band Convex Probe - Doppler Frequency : 5.0, 6.7 MHz

- Applications: Abdomen, OB Gyn, Urology, Vascular

Left Ventricle Volume, Single Plane, Two Chamber/Four Chamber, Method of Disk (LVAd, LVAs)

- Steered Angle : 0, +/- 5, 10, 15, 20° - Biopsy Guide Available : Multi Angle, Reusable - Probe Band Width: 2.0 – 5.0 MHz - Number of Element: 128 Left Ventricle Volume, Apical View,

Long Axis, Method of Disk (LVAd, LVAs) • 9L Wide Band Linear Probe - Convex Radius: 38 mmR - Applications: Vascular, Small Parts - FOV (Max): 68° Stroke Volume by Aortic Flow (AVA

Planimetry, AV Trace) - Probe Band Width : 3 – 10 MHz

- Physical Foot Print: 48 x 11 mm - Number of Element : 192 - B-mode Imaging Frequency: 2.0, 3.0,

4.0, 5.0 MHz • Mitral Valve - FOV (Max) : 44 mm

Stroke Volume by Mitral Flow (MVA Planimetry, MV Trace)

- Physical Foot Print : 44 x 6 mm - B-mode Imaging Frequency : 5.0, - Harmonic Frequency: 4, 5, 5.2, 5.5

MHz - 7.0, 9.0 MHz Cardiac Output by Mitral Flow (MVA Planimetry, MV Trace, HR) - Harmonic Frequency : 8.0, 10.0 - Doppler Frequency: 2.5, 3.3 MHz

- MHz - Biopsy Guide Available: Single Angle, Reusable

• Pulmonic Valve - Doppler Frequency : 4.0, 5.0 MHz Stroke Volume by Pulmonic Flow (PV Planimetry, PV Trace)

- Steered Angle : 0, +/- 5, 10, 15, 20° - Biopsy Guide Available : Multi • 4C Wide Band Convex Probe Cardiac Output by Pulmonic Flow (PV

Planimetry, PV Trace, HR) - Angle, Reusable

- Applications: Abdomen, OB Gyn, Urology, Vascular

• 10L Wide Band Linear Probe • Tricuspid Valve - Probe Band Width : 2 – 5 MHz

Cardiac Output by Tricuspid Flow (TV Planimetry, TV Trace, HR)

- Applications: Vascular, Small Parts, Neonatal, Pediatrics

- Number of Element: 128 - Convex Radius : 60 mmR

- Probe Band Width: 4 – 11 MHz - FOV (Max) : 58° Report Designer - Physical Foot Print : 60 x 18 mm - Number of Element: 192 • On-board reporting package

automates report writing - B-mode Imaging Frequency : 2.0, - FOV (Max): 39 mm - 3.0, 4.0, 5.0 MHz - Physical Foot Print: 39 x 10 mm

• Formats various exam results into a report suitable for printing to a windows printer or reviewing on a standard PC

- Harmonic Frequency : 4.0, 5.0, 5.2, - B-mode Imaging Frequency: 6.0, 8.0, 10.0 MHz

- 5.5 MHz - Doppler Frequency : 2.5, 3.3 MHz

- Doppler Frequency: 5.0, 6.7 MHz - Biopsy Guide Available : Multi Angle, Reusable - Steered Angle: +/- 10°, 20° • Exam results include patient info,

exam info, measurements, calculations, images, comments and diagnosis

- Biopsy Guide Available: Multi Angle, Reusable

• 5CS Convex Probe

• Standard templates provided - Applications: Abdomen, OB Gyn, Urology

Customizable templates • 11L Wide Band Linear Probe

- Probe Band Width : 2 – 6 MHz - Applications : Small Parts, - Number of Element : 128 Vascular, Neonatal, Pediatrics - Convex Radius : 60 mmR - Probe Band Width : 5 – 13 MHz Probes - FOV (Max) : 58° - Number of Element : 192 - Physical Foot Print : 60 x 18 mm - FOV (Max) : 38.4 mm - B-mode Imaging Frequency : 2.0, - Physical Foot Print : 38 x 10mm • 3.5CS Thin Wide Band Convex Probe - 3.0, 4.0, 5.0 MHz - B-mode Imaging Frequency : 7.0, - Applications: Abdomen, OB Gyn,

Urology, Vascular - Harmonic Frequency : 4.0, 5.0, 5.5, - 10.0, 12.0 MHz - 6.0 MHz - Harmonic Frequency : 10.0, 12.0, - Probe Band Width: 2.0 – 5.0 MHz - Doppler Frequency : 2.5, 3.3 MHz - 13.0 MHz

- Number of Element: 128 - Biopsy Guide Available : Multi Angle, Reusable

- Doppler Frequency : 5.0, 6.7 MHz - Convex Radius: 38 mmR - Steered Angle : 0, +/- 5, 10, 15, 20° - FOV (Max): 68° - Biopsy Guide Available : Multi

• 8L Wide Band Linear Probe - Angle, Reusable - Physical Foot Print: 48 x 11 mm - Applications : Vascular, Small - B-mode Imaging Frequency: 2.0, 3.0,

4.0, 5.0 MHz Parts • 12L Wide Band Linear Probe - Probe Band Width : 4 – 11MHz - Applications: Small Parts, Vascular,

Neonatal, Pediatrics - Harmonic Frequency: 4, 5, 5.2, 5.5

MHz - Number of Element : 128 - FOV (Max) : 38 mm

LOGIQ P5 Page 11 of 14

- Probe Band Width: 5 – 13.0 MHz

Page 12: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

- Biopsy Guide Available : None - Applications : Neonatal, Pediatrics - Number of Element: 192 - Probe Band Width : 4 11 MHz - FOV (Max): 39 mm

• 3Sp Wide Band Phased Array Sector - Number of Element : 128 - Physical Foot Print: 45 x 10 mm - Convex Radius : 11 mmR Probe

- B-mode Imaging Frequency: 7.0, 10.0, 12.0 MHz

- Applications : Cardiac, Trancranial, - FOV (Max) : 133° - Physical Foot Print : 26 x 10 mm Abdomen

- Harmonic Frequency: 10.0, 12.0 MHz - Probe Band Width : 1.5 – 5MHz - B-mode Imaging Frequency : 6.0, - Doppler Frequency: 5.0, 6.7 MHz - Number of Element : 64 - 8.0, 10.0 MHz

- FOV (Max) : 90° - Harmonic Frequency : 8.0, 10.0 - Steered Angle: +/- 10°, 20° - Physical Foot Print : 16 x 14 mm MHz Biopsy Guide Available: Single Angle,

Reusable - B-mode Imaging Frequency : 2.0, - Doppler Frequency : 4.0, 5.0 MHz - 3.0, 4.0, 5.0 MHz - Biopsy Guide Available : None - Harmonic Frequency : 3.0, 3.5, 4.0, • 3CRF Wide Band Microconvex Probe - 5.0, 5.5 MHz • 3S Wide Band Phased Array Sector - Applications: Abdomen, Urology,

Vascular - Doppler Frequency : 1.8, 2.0, 2.5, Probe - 3.3, 4.0 MHz - Applications : Cardiac, Trancranial, - Probe Band Width : 2.0 – 4 MHz - CW Doppler Frequency : 2.0 MHz Abdomen - Number of Element: 128

- Probe Band Width : 1.5 – 3.5 MHz Biopsy Guide Available: Multi Angle, Reusable

- Convex Radius : 20 mmR - Number of Element : 64 - FOV (Max) : 80° - FOV (Max) : 90° - Physical Foot Print : 28 x 16 mm

• 5Sp Wide Band Phased Array Sector - Physical Foot Print : 19 x 16 mm - B-mode Imaging Frequency : 2.0, - B-mode Imaging Frequency : 2.0, Probe - 3.0, 3.5, 4.0 MHz

- Applications : Cardiac, Trancranial, - 2.5, 3.0 MHz - Harmonic Frequency : 3.6, 3.8, 4.0, - Harmonic Frequency : 2.8, 3.0, 3.2, Abdomen - 4.2MHz - 3.6 MHz Probe Band Width : 2.0 – 8.0 MHz - Doppler Frequency : 2.5, 3.3 MHz

- Number of Element : 64 - Doppler Frequency : 1.7, 2.0, 2.2 - Biopsy Guide Available : Single Angle, Reusable - FOV (Max) : 90° - MHz

- Physical Foot Print : 10 x 10 mm - CW Doppler Frequency : 2.0 MHz - B-mode Imaging Frequency : 4.0, - Biopsy Guide Available : Multi • E8C Wide Band Microconvex Probe - 5.0, 6.7, 8.0 MHz - Angle, Reusable - Applications : OB Gyn, Urology,

Endocavity - Harmonic Frequency : 5.0, 6.0, 8.0, - 10.0 MHz • 5S Wide Band Phased Array Sector - Probe Band Width : 4 – 11 MHz - Doppler Frequency : 2.7, 3.0, 3.3 Probe - Number of Element : 128 - 4.0, 5.0 MHz - Applications : Cardiac, Trancranial, - Convex Radius : 11 mmR - CW Doppler Frequency : 2.5 MHz Abdomen - FOV (Max) : 133°

- Probe Band Width : 2 – 5 MHz Biopsy Guide Available: Multi Angle, Reusable

- Physical Foot Print : 26 x 10 mm - Number of Element : 96 - B-mode Imaging Frequency : 6.0, - FOV (Max) : 90° - 8.0, 10.0 MHz

• i12L Intraoperative Wide Band Linear Probe

- Physical Foot Print : 19 x 11 mm - Harmonic Frequency : 8.0, 10.0 - B-mode Imaging Frequency : 3.0 , - MHz

- Applications : Intraoperative, Small - 4.0, 5.0 MHz - Doppler Frequency : 4.0, 5.0 MHz - Parts, Vascular, Pediatrics - Harmonic Frequency : 4.0, 5.0 MHz - Biopsy Guide Available : Single

Angle, Disposable and Reusable - Probe Band Width : 4 – 12 MHz - Doppler Frequency : 2.5, 3.3 MHz - Number of Element : 96 - CW Doppler Frequency : 2.5 MHz - FOV (Max) : 25 mm Biopsy Guide Available: Multi Angle,

Reusable • E8CS Wide Band Microconvex Probe

- Physical Foot Print : 29 x 10 mm - Applications : OB Gyn, Urology, - B-mode Imaging Frequency : 6.0, Endocavity

• 7S Wide Band Phased Array Sector Probe

- 8.0, 10.0 MHz - Probe Band Width : 4 – 11 MHz - Harmonic Frequency : 10.0, - Number of Element : 128 - 12.0MHz - Applications: Cardiac, Pediatrics,

Abdomen - Convex Radius : 9 mmR

- Doppler Frequency : 5.0, 6.7 MHz - FOV (Max) : 168° - Steered Angle : +/- 5, 10, 15, 20° - Probe Band Width : 3 – 8 MHz - Physical Foot Print : 15 x 10 mm - Biopsy Guide Available : None - Number of Element : 96 - B-mode Imaging Frequency : 6.0, - FOV (Max) : 90° - 8.0, 10.0 MHz

• t739 Intraoperative Wide Band - Physical Foot Print : 14 x 7 mm - Harmonic Frequency : 8.0, 10.0 MHz Linear Probe - B-mode Imaging Frequency : 5.0,

- Applications : Intraoperative, Small - 6.0, 7.0 MHz - Doppler Frequency : 4.0, 5.0 MHz Parts, Vascular, Pediatrics - Harmonic Frequency : 6.0, 7.0, 8.0, - Biopsy Guide Available : Single - Probe Band Width : 4 -10 MHz - 8.5 MHz - Angle, Disposable and Reusable - Number of Element: 192 - Doppler Frequency : 4.0, 5.0 MHz - FOV (Max) : 39 mm

LOGIQ P5 Page 12 of 14

- CW Doppler Frequency : 4.0 MHz • 8C Wide Band Microconvex Probe

Page 13: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

- Physical Foot Print : 39 x 10 mm - Harmonic Frequency : 8.0, 10.0, - Applications: Urology - B-mode Imaging Frequency : 6.0, - 11.0 MHz - Probe Band Width: - 8.0, 10.0 MHz - Doppler Frequency : 4.0, 5.0 MHz ERB Linear5 – 10 MHz - Harmonic Frequency : 8.0, 10.0, - Biopsy Guide Available : Single

Angle, Reusable (stainless steel) ERB Convex: 5 - 9MHz

- 12.0 MHz - Number of Element: 128 - Doppler Frequency : 5.0, 6.7 MHz - FOV (Max) : - Steered Angle : +/- 5, 10, 15, 20° • P2D Non-imaging Single CW ERB Linear: 51mm - Biopsy Guide Available : Multi ERB Convex: 123° Doppler Pencil Probe

- Applications : Cardiac - Angle, Reusable - Physical Foot Print: ERB Linear: 51 x 5 mm - Frequency: 2.0 MHz

• i739 Intraoperative Wide Band Linear Probe

ERB Convex: 17 x 5.5 mm - B-mode Imaging Frequency: 6.0, 8.0,

10.0 MHz • P6D Non-imaging Single CW

- Applications : Intraoperative, Small Doppler Pencil Probe - Parts, Vascular, Pediatrics - Harmonic Frequency: 8.0, 10.0 MHz - Applications : Cardiac, Vascular,

Pediatric - Probe Band Width : 4 – 10 MHz - Doppler Frequency: 5.0, 6.7 MHz - Number of Element: 192 - Biopsy Guide Available: Multi

Angle, Grid Reusable - Frequency: 5.0 MHz

- FOV (Max) : 39 mm - Physical Foot Print : 39 x 10 mm • UG7C Gastro Probe – Europe Only

- B-mode Imaging Frequency : 6.0, • 4D3C-L Convex Volume Probe - Applications Transesophageal, - 8.0, 10.0 MHz - Applications : Abdomen, OB Gyn Stomach - Harmonic Frequency : 8.0, 10.0, - Probe Band Width : 2 – 5 MHz - Probe Band Width : 4.5 – 10MHz - 12.0 MHz - Number of Element: 192 - Number of Element: 128 - Doppler Frequency : 5.0, 6.7 MHz - Convex Radius : 39.1 mmR - Convex Radius: 8.5 mmR - Steered Angle : +/- 5, 10, 15, 20° - Volume Sweep Radius: 19.8 mmR - FOV (Max) : 125°

- FOV: 87° Volume 85° x 80° Biopsy Guide Available : None - Physical Foot Print : 19 x 5 mm - Physical Foot Print : 60 x 13 mm - B-mode Imaging Frequency : 6.0,

• BE9C Wide Band Biplane Microconvex Probe

- B-mode Imaging Frequency : 3.0, - 8.0, 10.0 MHz - 4.0, 5.0 MHz - Harmonic Frequency : 8.0, 10.0

- Applications: Urology, Endocavity - Harmonic Frequency : 4.0, 4.5, 4.8, - MHz - Probe Band Width : 4 – 11MHz - .0 MHz - Doppler Frequency : 5.0, 6.7 MHz - Number of Element : 96 x 2 - Doppler Frequency : 2.5, 3.3 MHz - Biopsy Guide Available: None - Convex Radius : 9 mmR - Biopsy Guide Available : Single

Angle, Reusable (stainless steel, plastic)

* available only in Europe - FOV (Max) : 127° - Physical Foot Print : 20 x 5 mm

Inputs and Outputs - B-mode Imaging Frequency : 6.0, - 8.0, 10.0 MHz • 4D8C Micro Convex Volume Probe

• Video In - Applications : Neonatal, Pediatrics - Harmonic Frequency : 8.0, 10.0 - S-Video - MHz - Probe Band Width : 4 – 11 MHz

• Video Out - Doppler Frequency : 4.0, 5.0 MHz - Number of Element: 192 - S-Video - Biopsy Guide Available : Single

Angle Reusable (stainless steel), Disposable

- Convex Radius : 14.0 mmR - Volume Sweep Radius: 80 mm • Audio Stereo In - FOV: 120° Volume 37.4 mm x 29° • VGA out

- Physical Foot Print : 30 x 50 mm • Connectors • BE9CS Wide Band Biplane

Microconvex Probe - B-mode Imaging Frequency : 6.0, - Footswitch - 8.0, 10.0 MHz

- USB (6) - Applications: Urology, Endocavity - Harmonic Frequency : 8.0, 10.0, - Ethernet - Probe Band Width : 4 – 11MHz - 1.0 MHz - Power for Peripherals - Number of Element : 96 x 2 - Doppler Frequency : 4.0, 5.0 MHz

- Convex Radius : 9 mmR - Biopsy Guide Available : Single Angle Reusable(stainless steel, plastic)

- FOV (Max) : 127°

- Physical Foot Print : 20 x 10 mm - B-mode Imaging Frequency : 6.0, • 4DE7C Convex Volume Probe - 8.0, 10.0 MHz - Applications : OB Gyn, Urology - Harmonic Frequency : 8.0, 10.0 - Probe Band Width : 4 – 11 MHz

- MHz - Number of Element: 192 - Doppler Frequency : 4.0, 5.0 MHz - Convex Radius: 10.1 mmR - Biopsy Guide Available : Single

Angle Reusable (stainless steel), Disposable

- Volume Sweep Radius: 11.6 mmR - FOV: 133° Volume 146° x 90° - Physical Foot Print : 27 x 6 mm

- B-mode Imaging Frequency : 6.0, • ERB Wide Band Biplane Probe

LOGIQ P5 Page 13 of 14

- 8.0, 10.0 MHz

Page 14: LOGIQ P5 - Healthcare Products and Service · • Microconvex Array • Real-time Auto Doppler Calculations - B + CFM/M • Linear Array • TruAccess, Raw Data Processing - Realtime

Safety Conformance GE Healthcare. - America: 888-202-5528 P.O Box 414, Milwaukee , Wisconsin 53201 USA The LOGIQ P5 is:

GE Healthcare , GE Asia Technology Park • Listed to UL 60601-1 by a Nationally

Recognized Test Lab 3rd Floor, Building 1,1 Hua Tuo Road , Shanghai – 201203

• Certified to CAN/CSA C22.2 601.1 by an SCC accredited Test Lab

Tel : +86 21 38774000 GE Ultraschall Deutschland GmbH & Co KG • CE Marked to Council Directive

93/42/EEC on Medical Devices Beethovenstr. 239, D-42655 Solingen Tel.: (+49) 212-2802-0 • Conforms to the following standards

for safety: Fax: (+49) 212-2802-28 -IEC/EN 60601-1 Medical Electrical

equipment, General requirement for Safety

Internet : www.gehealthcare.com

-IEC/EN 60601-1-1 Safety requirements for Medical Electrical Systems

GE, GE Monogram and LOGIQ®, CrossXBeam , TruScan , LOGIQView and Inte are trademarks of General Electric Company.

-IEC/EN 60601-1-2 Electromagnetic compatibility GE Medical Systems Ultrasound & Primary

Care Diagnostics, LLC a General Electric company doing business as GE Healthcare -IEC/EN 60601-1-4 Programmable

medical systems -IEC/EN 60601-2-37 Particular

requirements for the safety of ultrasonic medical diagnostic and monitoring equipment

-IEC 61157 Declaration of acoustic output © 2008 General Electric Company - All rights reserved.

-ISO 10993 Biological evaluation of medical devices

·NEMA UD3 Acoustic output display (MI, TIS, TIB, TIC)

Not all features or specifications described in this document may be available in all probes and/or modes. General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product at any time without notice or obligation. Contact GE Representative for the most current information.

LOGIQ P5 Page 14 of 14