    RAPID OIL TEST GAUGE

  • 8/16/2019 Lubmarine Rapid Oil Test Gauge (UK)

    2/9

    Pages

    3

    6

    9

    1011

    12

    13

    14

    16

    Water content measurement

    B.N. determination

    Introduction to the Rapid Oil Test GaugeProduct overview

    Subassemblies description and locationSpecifications

    Unpacking and inspection

    Initial start-upBattery

    Preserving the Rapid Oil Test GaugeMaintenance

    Power on / Main Run Screen / Power OffAdjusting the internal clock

    Oil monitoring functionTurning On and Off

    Setting the maximum acceptable water contentSetting the minimum acceptable BN

    Test history functionViewing the test history

    Downloading the test history to a computerErasing the test history

    Replenishment of consumables

    SUMMARY

    2

    Water Content MeasurementWith Reagent WT1/NH & Rapid Oil Test Gauge

    Make sure that the apparatus and accessories are perfectly clean before starting the test.

    1

    Read the instructionsand MSDS contained

    in this manual.

    Wear protective gloves. Wear goggles.

    Before each test, mind your own safetyand the environment.

    Wash thoroughly and obtain medical advice if eye or skincontact with a reagent or solution has occurred.

    3

  • 8/16/2019 Lubmarine Rapid Oil Test Gauge (UK)

    3/9

    5

    11 12

    1716

    Water Content MeasurementWith Reagent WT1/NH & Rapid Oil Test Gauge

    4

    75 6

    108 9

    42 3

    Clean the reaction flask usinga solvent such as PetroleumDistillate p/n DP9206.Wipe the reaction flask andRapid Test Gauge with a clea-ning fabric or absorbing paper.Make sure the measuring portis free of any dirt.

    Do not dispose of the residuesin the drainage system.

    Press the [Go] key. When the instrument memoryis full, a “History Full” messageshows. Press [Go] to confirm test(result will not be stored) or[Abort] to cancel.

    Water Content MeasurementWith Reagent WT1/NH & Rapid Oil Test Gauge

    Press the [ ▼ ] key for moreinstructions.

    Drain a full sachet of reagentWT1/NH in the reaction flask.

    Using the 10 ml syringe, inject10 ml of solution WT/BN in thereaction flask.

    Do not shake.

    Homogenize the oil to be tested, then introduce 5 mlof oil to be tested into the reaction flaskusing the syringe provided.

    Do not shake.

    Hold the reaction flask ona stable surface then closeand tighten the Rapid Oil TestGauge.

    Press the [Start] key.

    Grease the thread and O’ring of the Rapid Oil Test Gauge.

    Press any of the 3 front panelkeys to turn the Rapid Oil TestGauge on.

    Then press the [Test] key.

    Press the [H2O] key then follow the test instructions displayedby the instrument.

    LUBMARINE

    TestConf Hist

    RAPID

    OIL TEST GAUGE

    H2O BN Exit

    RUN TEST

    à Start Abort

    1° Add full WT REAGENT

    2° Add 10ml of SOLUTION

    3° Add 5ml of OIL

    Run Water Test

    á Start Abort

    4° Close and tighten5° Press START

    Run Water Test

    Go Abort

    Shake Gauge welluntil count is down to 0

    Run Water Test

    AbortGo

    Run Water Test

    History Full

    14The Rapid Oil Test Gaugecounts down the time required to complete the test.

    Abort

    80

    Run Water Test

    13Shake vigorously to start the reaction and throughout the test.

    15When the test is complete, theinstrument beeps, shows andlogs the water content of theoil sample. If the oil monitoringfunction is activated, theinstrument shows if the testresult is acceptable according to the function settings.

    Exit

    0.45%H2O

    Water Content

    OK

    If the overpressure alarm soundsand flashes, stop the test immediately

    and open to release pressure.

    RELEASE PRESSURE

    Run Water Test

  • 8/16/2019 Lubmarine Rapid Oil Test Gauge (UK)

    4/9

    BN DeterminationWith Reagent BN1 & Rapid Oil Test Gauge

    Make sure that the apparatus and accessories are perfectly clean before starting the test.

    1

    Read the instructionsand MSDS contained

    in this manual.

    Wear protective gloves. Wear goggles.

    Before each test, mind your own safetyand the environment.

    Wash thoroughly and obtain medical advice if eye or skincontact with a reagent or solution has occurred.

    6 7

    Grease the thread and O’ring of the Rapid Oil Test Gauge.

    Press any of the 3 front panelkeys to turn the Rapid Oil TestGauge on.

    Then press the [Test] key.

    Press the [BN] key.

    42 3

    LUBMARINE

    TestConf Hist

    RAPID

    OIL TEST GAUGE

    H2O BN Exit

    RUN TEST

    BN=30

    Select BN Range

    Run BN Test

    à Start Abort

    1° Add 10ml of OIL

    2° Add 10ml of SOLUTION

    3° Add 10ml of REAGENT

    Run BN Test

    BN DeterminationWith Reagent BN1 & Rapid Oil Test Gauge

    Select the BN rangeaccording to the BN of the oilwhen it is fresh.Select [BN =< 30] if the oil has aBN of 30 or less when freshSelect [BN > 30] if the oil has aBN higher than 30 when fresh.

    Follow the test instructionsdisplayed by the instrument.

    Press the [ ▼ ] key for moreinstructions.

    Homogenize the oil to be tested, then introduce 10 mlof oil to be tested into the reaction flaskusing a syringe provided.

    Using a clean 10 ml syringe,inject 10 mlof solution WT/BNin the reaction flask.

    Do not shake.

    Using a clean 10 ml syringe,inject 10 mlof reagent BN1 in the reaction flask.

    Do not shake.

    Hold the reaction flask ona stable surface then closeand tighten the Rapid Oil TestGauge.

    75 6

    108 9

  • 8/16/2019 Lubmarine Rapid Oil Test Gauge (UK)

    5/9

    LUBMARINE

    TestConf Hist

    RAPID

    OIL TEST GAUGE

    9

    RAPID OIL TEST GAUGEProduct Overview

    1. Product overview

    The Total Lubmarine Rapid Oil Test Gauge is a high precision battery operated digital lubricant testing gauge intended to measure and monitor the water content of lubricants up to 1.5% H2O and/or the remaining alkaline reserve (BN) of lubricants up to 80 BN.The gauge can be controlled locally by the operator using a front panel display and special function keys.A USB communication port can be used to download test results to a computer.

    2. Subassemblies description and location

    Keyboard

    USB port

    Reaction flask

    Battery

    Battery power indicator

    Identification plate

    Current function name

    Main information display

    Soft function keys

    Keys

    Vent groves

    O’ring

    Measuring port

    Main volume graduations

    Secondary volume graduations

    Overpressure fuse

    8

    á Start Abort

    4° Chose and tighten5° Press START

    Run BN Test

    Go Abort

    Shake Gauge welluntil count is down to 0

    Run BN Test

    Exit

    28 BN

    BN Value

    OK

    BN DeterminationWith Reagent BN1 & Rapid Oil Test Gauge

    Press the [Start] key. Press the [Go] key.

    When the test is complete, the instrument beeps, showsand logs the BN value of the oil sample.If the oil monitoring functionis activated, the instrumentshows if the test result isacceptable according to thefunction settings.

    When the instrument memoryis full, a “History Full” messageshows. Press [Go] to confirm test(result will not be stored) or[Abort] to cancel.

    Clean the reaction flask usinga solvent such as PetroleumDistillate p/n DP9206.Wipe the reaction flask andRapid Test Gauge with a clea-ning fabric or absorbing paper.Make sure the measuring portis free of any dirt.

    Do not dispose of the residuesin the drainage system.

    11 12

    16 18

    13

    17

    Shake vigorously to start the reaction and throughout the test.14 If the overpressure alarm sounds and

    flashes, stop the test immediately andopen to release pressure.

    Abort

    120

    Run BN Test

    The Rapid Oil Test Gaugecounts down the time required to complete the test.15

    AbortGo

    Run BN Test

    History Full

    RELEASE PRESSURE

    Run BN Test

  • 8/16/2019 Lubmarine Rapid Oil Test Gauge (UK)

    6/9

    RAPID OIL TEST GAUGEProduct Overview

    1 off 9 volts 6LR61/PP3 battery2½ years or 700 tests minimum0 to 50 °C (32 to 122 °F)- 20 to 65 °C (-4 to 149 °F)0.25 kg (0.55 lb)175 x 100 x 72 mm (6.89 x 3.94 x 2.84 in)USB Mini-B128 x 64 dots dot matrix LCDPMMA, POM, NBR, AISI 316 stainless steel(i) Water in oil content(ii) Oil alkaline reserve (BN) (Option)90 tests, including measured data, test evaluation, date and time

    Lubmarine reagent WT1/NHLubmarine solution WT/BN5 ml0 to 1.5 %H2O0.05 %H2O3.3 %H2O6.6 %H2O9.9 %H2O

    Lubmarine reagent BN1Lubmarine solution WT/BN10 ml0 to 80 BN1 BN

    (*: Data valid for 5 ml oil sample only)(**: Data valid for 10 ml oil sample only)

    Power requirement :Battery life :

    Operating temperature :Storage temperature :

    Weight :Dimensions :

    Communication port :Display :

    Materials in contact with tested oil :Test application :

    Test memory :

    Water TestWater test reagent :Water test solvent :

    Required oil sample :Measuring range* :Display resolution :

    Maximum allowable water content* :Maximum water content without damage on reaction flask* :

    Maximum water content without damage on measuring sensor* :

    BN Test (optional)BN test reagent :BN test solvent :

    Required oil sample :Measuring range** :

    Display resolution :

    3. Specifications

    Declaration of Conformity

    We GESERCO SARL 16 RUE FRANCOIS ARAGO 33700 MERIGNACOn behalf of TOTAL LUBRIFIANTSDeclare thatProduct LUBMARINE RAPID OIL TEST GAUGE

    Part Numbers WT1150, BN2150

    Complies with the clauses of the Directive of the Council 2004/108/EC Electromagnetic Compatibility.

    Standards : EN55022 EN61000-4-3 EN61000-4-11 EN61000-4-6 EN61000-4.4 EN61000-4-5 EN61000-4-2Tests conditions and results are given in the CERE-ENSAM test report No. QC14 dated 04.01.2012.

    Mérignac, October 11th, 2012

    Luc DARGENTChief Executive

    A battery power indicator is located in the upper left corner of thedisplay screen.When the battery power indicatorflashes, change the battery assoon as possible.

    Installing and connecting the battery:1 - Remove the battery cover from the back of the rapid oil test gauge.2 - Take the battery out.3 - Line up the positive and negative terminals on the new battery with the correspondingconnectors of the connecting cap and snap the cap on the battery terminals.4 - Insert the battery with connected cap into the battery compartment of the instrumentand replace the cover.

    Note:When first powered on after the battery has been replaced, the instrument clock must be adjusted.

    LUBMARINE

    TestConf Hist

    RAPID

    OIL TEST GAUGE

    Check that all items are present and have no visible damage.Delivery includes:

    Description

    Tech’Care/Std kit: Rapid Oil Test Gauge for water test onlyor Tech’Care/Pro.2 kit: Rapid Oil Test Gauge for water and BN tests

    Reaction flaskO’ring (installed in the test gauge)

    9 volt 6LR61/PP3 battery (in battery compartment but not connected)

    For water test

    Pack of sachets of reagent WT1/NH (50 sachets)Flask of solution WT/BN (500 ml)

    5 ml syringe (1)10 ml syringe (1)

    For BN test (optional)

    Flask of reagent BN1 (500 ml)Flask of solution WT/BN (500 ml)

    10 ml syringe (3)

    Part number

    WT1150BN2150WT1152DP6150DP6151

    WT1203WT1205DP6106DP6107

    BN2200WT1205DP6107

    The 9 volts 6LR6I/PP3 battery which powers your Rapid Oil Test Gauge should provide two years operation or more.

    6. Battery

    Upon delivery the battery of the Rapid Oil Test Gauge shall be connected and the internal clock shall be adjusted.No other adjustment is required; the instrument is ready to be used.

    5. Initial start-up

    4. Unpacking and inspection

    Rapid Oil Test GaugeOperating the Rapid Oil Test Gauge

    1110

  • 8/16/2019 Lubmarine Rapid Oil Test Gauge (UK)

    7/9

    Rapid Oil Test GaugeOperating the Rapid Oil Test Gauge

    12

    It is recommended to grease from time to time the internal O’ring using silicon freegrease p/n DP6108.

    n order to ensure the accuracy of subsequent tests and to preserve the instrument over time, the Rapid Oil Test Gauge shall always be keptlean and dry :

    The battery compartment cap and USBport cap shall be kept in place at all times in order to protect the internalelectronics of the instrument.

    Any trace of oil, solutions or reagentshall be removed from the gauge andeaction flask after each use. Specialttention shall be paid to the measuringort which must be kept perfectly free,nd to the internal O’ring.

    Use dedicated cleaning fabric orbsorbing paper for cleaning purpose.roper cleaning solvents include Whitepirit or Petroleum Distillates p/n DP9206.

    Never use toluene or other benzeneerivatives.

    7. Preserving the Rapid Oil Test Gauge

    he Lubmarine Rapid Oil Test Gauge was designe d for maintenance free operation.No maintenance is required apart from:

    • Keeping the instrument clean with its O’ring greased as described in §7.• Replacing the battery when necessary.• If required by your quality assurance system: recalibrating the instrument at regular intervals.

    8. Maintenance

    Rapid Oil Test GaugeOperating the Rapid Oil Test Gauge

    13

    5 - The year is displayed.Press the [ ▲ ] key to increment the year.Press the [ ▼ ] key to decrement the year.Press [Next] when done.

    Press any of the 3 front panel keys to turn the Rapid Oil Test Gauge on.

    The instrument initializes and displays itsserial number and embedded softwareversion for a few seconds.

    The instrument then displays the MainRun Screen which offers three functions:[Conf], [Test] and [Hist]

    Press the key corresponding to thefunction you want to run:Select [Conf] to adjust the internal clockand the oil monitoring function of yourinstrumentSelect [Test] to run a water test or a BN test (optional)Select [Hist] to view the test history

    Refer to the appropriate section of thismanual for details on each function.

    The instrument turns off automaticallyif not used for 60 seconds in order topreserve battery life.

    Note: When first powered on after the battery has been replaced, theinstrument accesses automatically the clock adjustment function.

    The Rapid Oil Test Gauge has an internal clock which is used to time stamp test results when they are stored in the instrument memory.To adjust the internal clock:

    6 - Adjust the month and press [Next]when done.7 - Adjust the day and press [Next] whendone.8 - Adjust the hour and press [Next]when done.9 - Adjust the minutes and press [Save] to store the data and return to the mainrun screen.

    1 - Press any of the 3 front panel keys to turn the Rapid Oil Test Gauge on.2 - Press the [Conf] key.3 - Press the [Other] key.4 - Press the [Time] key.

    2 Adjusting the internal clock

    1 Power on / Main run screen / Power off

    The Lubmarine Rapid Oil Test Gauge has an oil monitoring function which can warn the operator in case of: - Excessive water in oil content - Insufficient oil alkaline reserve (low BN)*

    If the function is On, the water in oil content or the BN value* is evaluated at the end of each test and the instrument shows if the test result isacceptable or not. The information is also stored in the instrument memory with the test result.

    The default settings of the function are: - Function status: Off - Water content alarm set to 1.5 %H2O - BN alarm* set to 1 BN (* The BN monitoring function is available only for Lubmarine Rapid Oil Test Gauges having the BN Test option).

    3 Oil Monitoring Function

    Note:When first powered on after the battery has been replaced, the instrument accesses automatically the clock adjustment function.

    LUBMARINE

    TestConf Hist

    RAPID

    OIL TEST GAUGE

    Time Exit

    Config

    á à Next

    Year : 2012

    Set Date

    á à Save

    Minute : 30

    Set Time

  • 8/16/2019 Lubmarine Rapid Oil Test Gauge (UK)

    8/9

    14

    The instrument shows the currentminimum acceptable alkaline reserveof the oil.4 - Press the [BN] key

    - Press any of the 3 front panel keys tourn the Rapid Oil Test Gauge on.- Press the [Conf] key.- Press the [Alarm] key.

    The instrument shows the currentmaximum acceptable level of water in the oil.4 - Press the [H2O] key.

    5 - Press the [▲ ] key to increase the value.Press the [ ▼ ] key to decrease the value.6 - Press [OK] when done

    5 - Press the [ ▲ ] key to increase the value.Press the [ ▼ ] key to decrease the value.6 - Press [OK] when done

    - Press any of the 3 front panel keys tourn the Rapid Oil Test Gauge on.- Press the [Conf] key.- Press the [Alarm] key.

    3.3 Setting the minimum acceptable level of alkaline reserve (BN) of the oil*

    3.2 Setting the maximum acceptable level of water in the oil

    (* The BN monitoring function is available only for LubmarineRapid Oil Test Gauges having the BN Test option).

    5 - Confirm the desired status by pressing the [Yes] keyor the [No] key

    - Press any of the 3 front panel keys tourn the Rapid Oil Test Gauge on.- Press the [Conf] key.- Press the [Alarm] key.

    3.1 Turning the oil monitoring function On or Off

    The instrument shows the current statusof the function.

    4 - Press the [On/Off] key.

    Rapid Oil Test GaugeOperating the Rapid Oil Test Gauge

    Alarm Other Exit

    Config

    H2O BN On/Off

    Alarm is : On

    H2O : 0.5 % H2O

    BN : 60 BN

    Config Alarm

    Yes No

    Confirm alarm : off

    Config Alarm

    Alarm Other Exit

    Config

    H2O BN On/Off

    Alarm is : On

    H2O : 0.5 % H2O

    BN : 60 BN

    Config Alarm

    á à OK

    0.5 % H2O

    Config Alarm

    Alarm Other Exit

    Config

    H2O BN On/Off

    Alarm is : On

    H2O : 0.5 % H2O

    BN : 60 BN

    Config Alarm

    á à OK

    50 BN

    Config Alarm

    15

    4 - The Rapid Oil Test Gauge shows as aremovable USB drive named OILTSTR.The OILTSTR drive contains a file namedDATALOG.CSV which is the test results file.Copy and paste the DATALOG.CSV file onyour hard drive to save it.

    1 - Press any of the 3 front panel keys to turn the Rapid Oil Test Gauge on.2 - Press the [Hist] key.

    4 - Press [Exit] when done.

    4 - The Rapid Oil Test Gauge shows as a removable USB drivenamed OILTSTR.5 -The OILTSTR drive contai ns a file named DATALOG.CSV whichis the test results file.6 - Delete the DATALOG.CSV file.

    1 - Remove the USB port cover of the Rapid Oil Test Gauge andconnect a Mini-USB cable to the port.2 - Connect the USB cable to the USB port of the computer.3 - Open the file explorer of the computer.

    4.3 Erasing the test history

    5 - Use Microsoft Excel™ or equivalent to view and process the test h istory datacontained in the DATALOG.CSV file.

    1 - Remove the USB port cover of theRapid Oil Test Gauge and connect a Mini-USB cable to the port.2 - Connect the USB cable to the USBport of the computer.3 - Open the file explorer of the computer.

    4.2 Downloading the test history to a computer

    4.1 Viewing test history

    The Lubmarine Rapid Oil Test Gauge has a test history function whichautomatically stores the result of each test.Stored data include : Test tag number Date Time Measured value If the oil monitoring function was turned on at time

    of test: Test result evaluation (“OK” or “Fail”)

    The test history can be viewed on the Rapid Oil Test Gauge localdisplay using the [Hist] function or downloaded to a computer via theUSB communication port for subsequent processing.

    The test history function can store up to 90 test results. When theinstrument memory is full, a “History full” message shows and theinstrument will not store any additional data unless the memory iscleared.

    4 Test history function

    Rapid Oil Test GaugeOperating the Rapid Oil Test Gauge

    LUBMARINE

    TestConf Hist

    RAPIDOIL TEST GAUGE

    3 - The instrument shows the result of thelast test performed. Use the [ ▲ ] and [▼ ]keys to browse test results.

    á à Exit

    History

    Test 4 / 406/09 /2012 -18:15

    0.00 %H2O OK

    á à Exit

    History

    Test 4 / 406/09/2012 -18:15

    0.00 %H2O OK

  • 8/16/2019 Lubmarine Rapid Oil Test Gauge (UK)

    9/9

    16

    Replenishment of consumables

    Test equipmentAny test equipment which requires replacement shall be sourced from GESERCO

    (Original Equipment Manufacturer authorized by TOTAL LUBRIFIANTS).

    Reagent and SolutionsAny reagent or solution shall be sourced from GESERCO

    (Original Equipment Manufacturer authorized by TOTAL LUBRIFIANTS).Reagents and solutions can be sourced individually or as a complete set for up to 50 tests.

    Product codesThe following is a summary of available spares & replenishments.

    Other spares and replenishments are available. Please contact GESERCO for additional details.For any request for replenishment, please contact :

    GESERCO SARLAttn : Sales department16 rue François Arago

    33700 MérignacFrance

    Tel : + 33 (0) 5 5634 9229Fax : + 33 (0) 5 5634 9544

    e-mail : [email protected]

    “Lubmarine recommends thatyou place your replenishment

    orders on the www.geserco-sarl.com website

    to ensure a rapid response.”

    Application

    RAPID OIL TEST GAUGE

    TECH’CARE/PRO.2

    TECH’CARE/STD

    WATER IN OIL TEST

    BN DETERMINATION

    SALT WATER TEST

    CARBON MATTERDETERMINATION

    SAMPLING

    MAINTENANCE

    Name

    Water in oil testerWater in oil + BN testerReaction flaskO'ringBatteryUSB cable

    Water test & BN testreplenishment pack

    Water testreplenishment pack

    Water test reagentWater test solvent

    BN test reagentBN test solvent

    Chloride test papers

    Chloride test solventPipettingTesting

    Carbon matter test papersCarbon matter sampling

    10 ml sampling 5 ml sampling 2 ml sampling 50 ml beaker100 ml beaker150 ml beaker250 ml beaker

    Cleaning solventCleanings budsO’ring & thread grease

    Content

    Complete testerComplete testerReaction flaskO'ring6LR51/PP3 batteryMini USB cable

    1 x reagent WT1/NH1 x reagent BN12 x solution WT/BN

    1 x reagent WT1/NH1 x solution WT/BN

    Reagent WT1/NHSolution WT/BN

    Reagent BN1Solution WT/BN

    Indicator papersfor chloridesSolution EPasteur pipettesTest tubes

    Spot test papersGlass rod

    10 ml syringes 5 ml syringes 2 ml syringes 50 ml beaker100 ml beaker150 ml beaker250 ml beaker

    Petroleum DistillateCotton budsGrease tube

    Packaging unit

    1 unit1 unit1 unit1 unit1 unit1 unit

    1 complete setfor 50 water tests and50 BN tests

    1 complete setfor 50 water tests

    1 Pack of 50 sachets1 x 500 ml flask

    1 x 500 ml flask1 x 500 ml flask

    1 Box of 30

    1 x 500 ml flaskSet of 5Set of 5

    Pad of 251 unit

    Set of 10Set of 10Set of 101 unit1 unit1 unit1 unit

    1 x 500 ml flask1 Pack of 1001 x 60g tube

    Code

    WT1207BN2106WT1152DP6150DP6151DP6152

    PP7102

    PS8102

    WT1203WT1205

    BN2200WT1205

    RC3200

    RC3201RC3101RC3100

    ET4200ET4100

    DP6107DP6106DP6105DP6101DP6102DP6103DP6104

    WT9206VT5101DP6108

    For technical support please contact : [email protected]