Luminosity Monitor Status Luminosity Monitor Status MICE Collaboration Meeting 27 5 July 2010 Paul Soler, David Forrest Danielle MacLennan

Luminosity Monitor Status

Embed Size (px)

DESCRIPTION

Luminosity Monitor Status. MICE Collaboration Meeting 27 5 July 2010 Paul Soler, David Forrest Danielle MacLennan. Purpose of Luminosity Monitors. - PowerPoint PPT Presentation

Citation preview

Page 1: Luminosity Monitor Status

Luminosity Monitor StatusLuminosity Monitor StatusLuminosity Monitor StatusLuminosity Monitor Status

MICE Collaboration Meeting 27 5 July 2010

Paul Soler, David ForrestDanielle MacLennan

Page 2: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

2

Purpose of Luminosity MonitorsPurpose of Luminosity Monitors Luminosity monitor to determine particle rate close to target

and extract protons on target as function of depth – independent of beam loss monitors.

The luminosity monitor was installed and commissioned in the ISIS vault in January and February 2010

Further test runs to validate the gate width were carried out in April 2010

The luminosity monitor (LM) is now an integral part of MICE and has been taking data regularly since February

The LM scaler information is now standard and can be used to normalise all future analyses with a measure that is proportional to the protons on target (POT) and independent of the beamloss (which is written on a separate data stream)

Meanwhile, simulations to understand and normalise LM to POT are ongoing.

Page 3: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

3

Final design of luminosity monitor:

Luminosity Monitor DesignLuminosity Monitor Design

Beam

Cuts off: protons ~500 MeV/c pions ~150 MeV/c

(6 mm thick)

(6 mm thick)

Page 4: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

4

Final design of luminosity monitor:

Luminosity Monitor DesignLuminosity Monitor Design

Beam

Cuts off: protons ~500 MeV/c pions ~150 MeV/c

Page 5: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

5

Use four Hamamatsu H5783P PMTs Readout consists of three signals within

the 3.23 ms MICE experimental gate:— Coincidence 12 (LMC-12) – scaler 08— Coincidence 34 (LMC-34) – scaler 09— Coincidence 1234 (LMC-1234) – scaler 10

PhotomultipliersPhotomultipliers

50 mm

Power provided by two low voltage power supplies:

Discriminator set at 500 mV: very low noise!

LMC-12

LMC-34

LMC-1234

PMT-1: 10.5 V

PMT-2: 10.7 V

PMT-3: 10.7 V

PMT-4: 10.5 V

Page 6: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

6

Commissioning Luminosity MonitorsCommissioning Luminosity Monitors Commissioning mainly 7 February LM signals inside 3.23 ms experimental trigger gate

ISIS bunches 100 ns long and 325 ns separation, so net gate width is 3.23x(100/325) = 1 ms

In February, set width of LM signals to 40 ns, but concern that gate width too big and may cause saturation at high rate

In April 20-21 performed runs as a function of gate width to check dependence with gate width (BL limited to <1.4 V)

Page 7: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

7

Data taken Feb 2010Data taken Feb 2010

Luminosity Monitor Commissioning RunsCoincidence of PMTs 1,2

y = 1955x - 268.3

0

1000

2000

3000

4000

5000

6000

0 0.5 1 1.5 2 2.5 3

Beamloss S7 Integral method (Vms)

Sca

lar

for

Co

inci

den

ce o

f P

MT

1, 2

Page 8: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

8

Data taken Feb 2010Data taken Feb 2010

Luminosity Monitor Commissioning RunsCoincidence of PMTs 3, 4

y = 2086.4x - 402.35

0

1000

2000

3000

4000

5000

6000

0 0.5 1 1.5 2 2.5 3

Beamloss Sec7 Integral Method

Sc

ala

r o

f C

oin

cid

en

ce

of

PM

Ts

3 a

nd

4

Page 9: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

9

Luminosity Monitor Commissioning RunsCoincidence of PMT 1,2,3,4

y = 888.59x - 281.25

0

500

1000

1500

2000

2500

0 0.5 1 1.5 2 2.5 3

Beamloss Sec7 Integral Method

Sc

ala

r o

f C

oin

cid

en

ce

of

PM

T 1

2 3

4

Data taken Feb 2010Data taken Feb 2010

Page 10: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

10

Data taken Feb 2010Data taken Feb 2010 Summary of results:

• LMC-12: 1955 particles per V.ms / 4 cm2

• LMC-34: 2086 particles per V.ms / 9 cm2

• LMC-1234: 889 particles per V.ms / 4 cm2

Assume beamloss calibration of 3.5x10-14 V.s/pot at 9 ms, therefore: 1 V ms =2.9x1010 pot• LMC-12: 1.71x10-8 particles/(pot . cm2)• LMC-34: 0.81x10-8 particles/(pot . cm2)• LMC-1234: 0.78x10-8 particles/(pot . cm2)

Coincidence 34 and 1234 have same rate per cm2

Page 11: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

11

Luminosity Monitor 12 - 40 ns gate

y = 2056.8x - 249.97

0

500

1000

1500

2000

2500

3000

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 12

Luminosity Monitor 12 - 30 ns gate

y = 2110.4x - 270.77

0

500

1000

1500

2000

2500

3000

3500

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 12

Data taken April 2010 – LM12Data taken April 2010 – LM12Luminosity Monitor 12 - 10 ns gate

y = 2069.6x - 267.26

0

500

1000

1500

2000

2500

3000

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 12

Luminosity Monitor 12 - 20 ns gate

y = 2132.5x - 292.73

0

500

1000

1500

2000

2500

3000

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 12

10 ns 20 ns

40 ns30 ns

2070 part/V ms

2057 part/V ms2110 part/V ms

2133 part/V ms

Page 12: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

12

Data taken April 2010 - LM 34Data taken April 2010 - LM 34

Luminosity Monitor 34 - 40 ns gate

y = 2292.3x - 415.81

0

500

1000

1500

2000

2500

3000

3500

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 34

Luminosity Monitor 34 - 30 ns gate

y = 2374x - 466.15

0

500

1000

1500

2000

2500

3000

3500

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 34

Luminosity Monitor 34 - 10 ns gate

y = 2304.6x - 371

0

500

1000

1500

2000

2500

3000

3500

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 34

Luminosity Monitor 34 - 20 ns gate

y = 2424x - 486.19

0

500

1000

1500

2000

2500

3000

3500

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 34

10 ns 20 ns

40 ns30 ns

2304 part/V ms

2292 part/V ms2374 part/V ms

2424 part/V ms

Page 13: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

13

Luminosity Monitor 1234 - 40 ns gate

y = 1073.1x - 297.12

0

200

400

600

800

1000

1200

1400

1600

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 123

4

Luminosity Monitor 1234 - 30 ns gate

y = 1026.8x - 285.44

0

200

400

600

800

1000

1200

1400

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 123

4

Luminosity Monitor 1234 - 20 ns gate

y = 929.62x - 246.01

0

200

400

600

800

1000

1200

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 123

4

Luminosity Monitor 1234 - 10 ns gate

y = 717.61x - 136.6

0

100

200

300

400

500

600

700

800

900

1000

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Beamloss S7 Integral method (Vms)

Lum

inos

ity M

onito

r 123

4

Data taken April 2010 - LM Data taken April 2010 - LM 12341234

10 ns20 ns

40 ns30 ns

718 part/V ms

1073 part/V ms1026 part/V ms

930 part/V ms

Page 14: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

14

Data taken April 2010Data taken April 2010 Summary of results:

LMC-12 and LMC-34 seem to be independent of gate width, while LMC-1234 seems to increase with gate width

Conclusion: set gate at 10 ns to minimise pile-up

Rate vs gate width (ns)

LMC-12 (part/ V ms)

LMC-34 (part/ V ms)

LMC-1234 (part/ V ms)

10 ns 2070 2304 718

20 ns 2133 2424 930

30 ns 2110 2374 1026

40 ns 2057 2292 1073

Average 2092 2349 937

Feb 2010 1955 2086 889

Page 15: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

15

We have run simulations using G4Beamline (D. MacLennan) Set up cylindrical target (R=3mm,r=2.3mm), and two

detectors 100x100cm2, separated by 15 cm plastic at 10 m and 25o angle. Include 6 mm thick steel from target enclosure

Comparison to simulationsComparison to simulations

Page 16: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

16

Only select particles within acceptance of detectors (100x100cm2 at 10 m) and kill all other particles

Test that we don’t kill valid particles by changing kill volumes

Comparison to simulationsComparison to simulations

Proton Beam

Target

(yellow volumes)

Page 17: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

17

Only select particles within acceptance of detectors (100x100cm2 at 10 m) and kill all other particles

First run with QGSP hadronic model

Comparison hadronic modelsComparison hadronic models

Shielded detectorsUnshielded detectors

QGSPQGSP

Page 18: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

18

Only select particles within acceptance of detectors (100x100cm2 at 10 m) and kill all other particles

Now run with QGSP_BERT (QGSP+Bertini cascade model) for comparison – shows proton peak at ~1200 MeV/c

Comparison hadronic modelsComparison hadronic models

Shielded detectorsUnshielded detectors

QGSP_BERTQGSP_BERT

For total particle yields: assume neutron efficiency from GEANT4 simulations ~2.2%

Page 19: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

19

Compare number particles (proton+pion+neutronx2.2%) crossing detectors for hadronic models (up to a factor 2):

Comparison hadronic modelsComparison hadronic models

Hadronic model

Number particles in unshielded detector

Number particles in shielded detector

Particles/ (pot cm2)

Unshielded

Particles/ (pot cm2)

Shielded

LHEP (107

pot)183 74 (1.83±0.14)x10-9 (7.40±0.86)x10-10

LHEP_BERT (107 pot)

342 257 (3.42±0.19)x10-9 (2.57±0.16)x10-9

QGSC (107

pot)192 67 (1.92±0.14)x10-9 (6.70±0.82)x10-10

QGSP (107 pot)

157 57 (1.57±0.13)x10-9 (5.70±0.76)x10-10

QGSP_BERT (109 pot)

31145 22911 (3.12±0.02)x10-9 (2.29±0.02)x10-9

QGSP_BIC (109 pot)

27380 21693 (2.74±0.02)x10-9 (2.17±0.02)x10-9

Page 20: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

20

Compare number protons crossing unshielded detector (104 cm2) for the new target (cylinder with outer radius 3 mm and inner radius 2.3 mm) compared to the old target (10 mm x 1 mm) geometry

Comparison target geometryComparison target geometry

Volume material in each target is very similar (assume depth inside beam=10mm):

• Old target: 10x1x10 mm3

• New target: (3.02-2.32)x10=116.7 mm3

New Old

Page 21: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

21

Compare number protons crossing unshielded detector (104 cm2) for two target geometries (using QGSP_BIC)

Comparison target geometryComparison target geometry

Target geometry

Number particles in unshielded detector

Protons on target (pot)

Area detector (cm2)

Protons/ (pot cm2)

Unshielded detector

New 27380 109 104 (2.74±0.02)x10-9

Old 3639 108 1600 (2.27±0.04)x10-8

There is a factor of 8.31 difference in normalisation, Old target has 10 mm thickness New target has variable thickness due to geometry of cylinder (effective

average thickness 1.945 mm=116.7/60) Another simulation was run to determine fraction of particles interacting in

each target:27.8

0051.0

0422.0

New

Old (therefore net number of pot in

new target is 8.27 times smaller!)

Page 22: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

22

Assume all protons on target traversing 10 mm target are lost in ISIS (9 MeV Eloss), so for new target simply multiply by 8.27 (ratio of old/new)

Comparison to simulationsComparison to simulations

Hadronic model

Particles/ (pot cm2)

Unshielded

Particles/ (pot cm2)

Shielded

Ratio Shielded/ Unshielded

LHEP 1.51x10-8 6.12x10-9 0.41

LHEP_BERT 2.83x10-8 2.12x10-8 0.75

QGSC 1.59x10-8 5.54x10-9 0.35

QGSP 1.30x10-8 4.71x10-9 0.36

QGSP_BERT 2.58x10-8 1.89x10-8 0.73

QGSP_BIC 2.27x10-8 1.79x10-8 0.79

Data 1.71x10-8 8.10x10-9 0.47

No model describes data accurately (not even ratio is well described) However, to do normalisation more accurately would need to determine

number of protons lost in ISIS when target is traversed by beam

Page 23: Luminosity Monitor Status

LM Status, MICE CM27, 5 July 2010

23

Luminosity Monitors have been installed in ISIS vault and are now working regularly for MICE analyses

LM data scales very well with beam loss data Up to ~1.4 V.ms beamloss, LM rate independent gate width

10 ns-40 ns (have chosen 10 ns as final width) Comparison of yields for different hadronic models shows big

differences in yields (about a factor of 2) Normalisation of simulations for cylindrical target are about a

factor of 8 smaller than the data, due to the fact that not all protons interact in target.

Need to understand how protons are lost to the beam better to do a proper comparison of LM data to simulations

ConclusionConclusion