DEMINERALIZER MPW-LAB 16

M A D E I N

P O L A N D

Page 2: M A D E I O L A N D - img.unimed.cz · type according to: pn.en. iso 3696:1999 life time dem-mlab.16 prod demineralizer mpw-lab 16, 230v 50hz, efficency 16l/h, class of purity ii,

DEM-MLAB demineralizers use tap water and fit every modern laboratory’s requirements in areas of supply-ing with purified water for instrumental and analytical purposes. Conductivity of obtained water does not ex-ceed 0,055 µS/cm, so it is applicable for PN-EN ISO 3696:1999, ASTM, CLSI, FP standards.

Water may also be used for AAS, ICP/MS, IC, HPLC, GC instrumental analyses. In addition, DEM-MLAB demin-eralizers are equipped with a microprocessor control and measurement system.

SeriesDEM-MLAB 16

HYDROLAB.PL

CATALOGUE NUMBER MPW-LAB PRODUCT / ACCESORIES DESCRYPTION TYPE ACCORDING TO: PN.EN.ISO 3696:1999 LIFE TIME

DEM-MLAB.16 PROD DEMINERALIZER MPW-LAB 16, 230V 50Hz, efficency 16L/h, Class of purity II, III II i III

DEM-PREFL.5UM ACC MECHANICAL PRE-FILTER 5 µm 6 months

DEM-M1.SCS ACC MODULE M1 SENDIMENT-CARBON- SOFTENING 6 months

DEM-M3.IE ACC MODULE M3 ION EXCHANGE 5000 ml 1x5000 ml

DEM-MLAB.16UV PROD DEMINERALIZER MPW-LAB 16UV, 230V 50Hz, efficency 16L/h, Class of purity I, II, III I i II i III

DEM-PREFL.5UM ACC MECHANICAL PRE-FILTER 5 µm 6 months

DEM-M1.SCS ACC MODULE M1 SENDIMENT-CARBON- SOFTENING 6 months

DEM-M3.IE ACC MODULE M3 ION EXCHANGE 5000 ml 1x5000 ml

DEM-UV.254 ACC UV LAMP 254 nm 8500 h

DEM-CAP.02UM ACC MICROFILTRATION CAPSULE 0,2 µm 12 months

0,05 uS/cmSQF

CWC

CNC EZ

uS/cm

0,45/0,2 um

conductivity <0,05 uS/cmresistivity > 18 MOhm*cm

- I type (ISO 3696, ASTM, CLSI)

concentrate(drain)

tank

tank

III type(ISO 3696, ASTM, CLSI)

(option)

area of device

pump

manometer

membrane module

module M1

feed water

pressure min. P=3 barT=4-40 C

C max = 1200 mg/dm3

sediment

5 µm

control panel

module M3

UV lamp 254 nm

AUTOMATICS A

R=96-99%

Page 3: M A D E I O L A N D - img.unimed.cz · type according to: pn.en. iso 3696:1999 life time dem-mlab.16 prod demineralizer mpw-lab 16, 230v 50hz, efficency 16l/h, class of purity ii,

HYDROLAB.PL

General information:Supply: tap waterEfficiency: min. 16dm3/hPurified water intake speed: 1-2 dm3/minConductivity < 0,056 µS/cmBacteria < 1 cfu/ml*Particles > 0,2μm < 1/ml**device equipped with a UV lamp and a 0,2μm microfiltration capsule

Purity class:Water purified fits the requirements of the PN-EN ISO 3696:1999, ASTM, CLSI standard for first*, second and third purity class, microbiological and physicochemical parameters match the FP requirements for purified production water.

Usage:Obtained water may be used for instrumental analyses AAS, ICP/MS, IC, HPLC, GC, bacteria cultures, biochemical analyses.

Technical specification: » The device works under tap water pressure. » Fed by tap water. » Water purification levels:

– sediment pre-filter 5µm, – integrated module (sediment-carbon-softening), – reverse osmosis, – initial demineralization on a mixed ionex resin, – final demineralization on a mixed ionex resin, – UV lamp 254 nm, – microfiltration cascade capsule 0,45/0,2μm. – Efficiency: min. 16 dm3/h. – Demineralized water conductivity < 0,056 μS/cm. – Unattended and automated. – Equipped with a pump 24V increasing feed water pressure. – Mobile water intake point - second purity class (PN-EN ISO

3696:1999, ASTM, CLSI) - nozzle reach min. 2 m. – Additional water intake point - first class (PN-EN ISO 3696:1999

- with a 0,2μm microfiltration capsule). – Optional installation of an additional water intake point for

general-purpose water - third class (PN-EN ISO 3696:1999; glass washing, laboratory washer supply, autoclaves etc.).

– Mobile, adjustable inox arm holding water collection points – available adjustment ranges: up/down, front/back, left/right.

– Equipped with a 10 dm3 pressure tank. – Optional replacement with a bigger tank (40dm3, 80dm3). – Automated system shutdown when the tank is full. – Optional connection to a washer, an autoclave etc. – Maintenance procedures may be performed by the user (easy

replacement of disposables). – Maximum working pressure: 1 MPa – Fed by cold water: 5-40°C. – Energy consumption 50W. – Power supply: 230V/50Hz. – Can be installed by the user. – Acid-proof stainless steel (inox) housing.

Functions monitoring the device:The device is equipped with a microprocessor control and measurement system that includes:LCD display screen 2x16 charactersconductometer measuring conductivity and temperature of purified water (in µS/cm or MOhm),reading values compensated and uncompensated thermally,timer displaying current date and time,alarm informing about necessity to replace mechanical and carbon filters,alarm informing about necessity to replace ionex resins,alarm informing about necessity to replace microfiltration capsule,alarm informing about necessity to replace UV lamp,maintenance deadlines preview,menu in English, Russian, Spanish or German,built-in RS 232 connector for personal computers,individual adjustment of maintenance frequency and alarm levels,Manometer measuring feed water pressure.

Functions protecting the device:Pomp shutdown when:

– feed water pressure is too low (lack of feed water) – low pressure sensor,

– the tank is full – high pressure sensor.

Dimensions (width x depth x height): 230x480x560 mmTank 10l: height: 390 mm, diameter: 250 mm

Feed water parameters: » Conductivity < 1200 μS/cm » Pressure > 0,35 MPa » Temperature: 5-40ºC » Hardness < 250 mg CaCO3/dm3

» Fe < 0,2 mg/dm3

Required connections: » cold water connection ½” or ¾” » 230V socket » drain

300

200

Page 4: M A D E I O L A N D - img.unimed.cz · type according to: pn.en. iso 3696:1999 life time dem-mlab.16 prod demineralizer mpw-lab 16, 230v 50hz, efficency 16l/h, class of purity ii,

DEM-PREFL.5UM DEM-M1.SCS DEM-M3.IE DEM-CAP.02UM DEM-UV.254

AccessoriesDEM-MLAB 16

SERWIS/SERVICEtel. +48 22 610 81 07 +48 22 610 50 14fax +48 22 610 55 [email protected]

WORLDWIDEtel. +48 22 673 04 08 +48 22 879 70 97fax +48 22 610 81 [email protected], www.mpw.pl