Manual Medidor de Dureza Th160_ba_e

Embed Size (px)

Citation preview

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    1/41

    Operating ManualTH-160

    IMP-Ingenieurgesellschaft mbHWesterbreite 7

    49084 OsnabrückTel 0049-!0"#4$-9778-%&0fa' 0049-!0"#4$-9778-$0(

    eMailinf)*im+-,eutschlan,,ehtt+..///im+-,eutschlan,,e

    mailto:eMail:[email protected]:eMail:[email protected]

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    2/41

     1 INTRODUCTION 31.1 TH160 advantage 31.2 Main applications & testing range 4

    1.2.1 Main applications 4  1.2.2 Testing range 4 

    1.3 Configuration 61.3.1 Standard configuration 6 1.3.2 Optional configuration 6 

    1.4 Working conditions 7

    2 S TRUCTURE &  TESTING PRINCIPLE 82.1 Structure 8

    2.1.1 The hardness tester appearance 8  2.1.2 Names of the parts 8  2.1.3 Names of the parts of D type Impact device 9 2.1.4 Different types of Impact devices 9 

    2.2 Testing principle 9

    3 SPECIFICATIONS  10

     4 PREPARATION &  TESTING  114.1 Preparation & inspection before testing 11

    4.1.1 Preparation for sample surface 11 4.1.2 System setting 11 4.1.3 Preset test parameters 11  

    4.2 Testing 12

    4.2.1 Fix printing paper 12  4.2.2 Start-up 12  4.2.3 Loading 12  4.2.4 Testing 13  4.2.5 Read measured value 13  4.2.6 Print out the result 13  4.2.7 Press “On/off” to switch off. 13  4.2.8 Test result processing 13  4.2.9 Testing result 13  

    5 SPECIAL PROMPT   14

    6 OPERATION EXPLAIN IN DETAILS  156.1 Switch on 156.2 Switch off 156.3 Testing 15

    6.3.1 Instruction of the main display interface 16 6.3.2 The main display interface testing operation 16  6.3.3 The main display interface key operation 17 6.3.4 Instruction of the secondary display interface 18 6.3.5 The secondary display interface testing operation 18 6.3.6 The secondary display interface key operation 18 

    6.4 Menu structure 196.5 Preset test parameters 20

    6.5.1 Preset impact direction 21  6.5.2 Preset average times 22  

    Manual TH-160 1

    CONTENTS

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    3/41

    6.5.3 Preset material 22 6.5.4 Preset hardness scale 24  

    6.5.5 Preset tolerance limit 24  6.5.6 Preset histogram display range 25 6.5.7 Preset hardness / σ b 25 

    6.6 Print function 266.6.1 Print current data 26  6.6.2 Print data by memory group no. 27 6.6.3 Print all data of memory 27  6.6.4 Print current data and histogram 28  6.6.5 Print data and histogram by memory group no. 29  6.6.6 Print all data and histogram of memory 29 

    6.7 Memory management 306.7.1 View from No. 1 group / view from the last group 30  6.7.2 View from selected group 31  6.7.3 Data transfer 31  6.7.4 Delete selected group 31 6.7.5 Delete all 32  6.7.6 Deletion confirmation 32 

    6.8 Browsing display screen 326.9 System set 33

    6.9.1 Preset LCD brightness 34  6.9.2 Preset time and date 34  

    6.9.3 Preset language 34  6.10 Software information 356.11 Software calibration 366.12 Charging 376.13 Background light 376.14 Auto power off 376.15 Battery replacing 376.16 Connection of data transmission cable 37

     7 T ROUBLE SHOOTING 38

    8 M AINTENANCE 398.1 Impact device maintenance 398.2 Conservation of printing record 398.3 Maintenance 39

    9 NOTICE OF TRANSPORTATION AND STORAGE  40

    2 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    4/41

     1.1 TH160 ADVANTAGE

    - Printer and main unit are designed together.- 320*200 matrixes LCD show beautiful font, figure, and abundant information.- Lithium ion rechargeable battery and charge control circuit are without memorial

    reaction, convenient to charging and longevity.- Thermal line printer works quiet and could print any number of copies of testing

    results. Trouble of exchange ribbon is avoided.- Menu operation with a few keys is simple and convenient.

    - With RS232 interface, alternative communication modes satisfy personalrequirement of different users.

    - Identification of ordinary impact direction automatically is designed (except G typeimpact device).

    - 6 kinds of different impact devices; no readjust is needed while exchanging,identification impact device automatically.

    - Large memory can remember 240~1000 groups (Relative to average times 32~1 )information includes single measured value, mean value, testing date, impactdirection, impact times, material and hardness system etc.

    - Preset limit value of hardness and alarm automatically. Be convenient for the

    requirement of the batch measuring.- Screen with backlight makes it convenient for test in darkness.- The main display interface increased the contents of date, time, memory

    information, battery information, analogue indicator, exceed limit value alarm, typeof impact device and on line help etc., more convenient and practical.

    - Display interfaces have on line help make it convenient to handle.- screen has the display of charging process. With charging indicating lamp, if it is in

    the condition of charging is clear at a glance.- TH160 has function of software calibration.- Data-View software for TH160 satisfies the higher requirement of quality guarantee

    and management.

    Manual TH-160 3

     1. INTRODUCTION

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    5/41

     1.2 M AIN APPLICATIONS &  TESTING RANGE

     1.2.1 M AIN APPLICATIONS

    - Installed machinery and permanently assembled parts.- Die cavity of molds- Heavy work piece- Failure analysis of pressure vessel, steam generator and other equipment.- Testing surface of a small hollow space.- Bearing and other parts.- Material identification in the warehouse of metallic materials.- Rapid testing on multi-point measurement for Large-scale work piece.

     1.2.2 T ESTING RANGE

    Testing range see Table 1 and Table 2.

    Table 1

    4 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    6/41

    Table 2No.Material HLD Strength σb(MPa)

    1 Mild steel 350~522 374~7802 High-Carbon steel 500~710 737~16703 Cr steel 500~730 707~18294 Cr-V steel 500~750 704~19805 Cr-Ni steel 500~750 763~20076 Cr-Mo steel 500~738 721~18757 Cr-Ni-Mo steel 540~738 844~19338 Cr-Mn-Si steel 500~750 755~19939 Super strength steel 630~800 1180~265210 Stainless steel 500~710 703~1676

    Manual TH-160 5

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    7/41

     1.3 CONFIGURATION

     1.3.1 S TANDARD CONFIGURATION

    Main unit 1D type impact device with cable 1Small support ring 1Cleaning brush 1Standard test block 1Charger 1Thermal printing paper 1Communication cable 1

     1.3.2 OPTIONAL CONFIGURATION

    Types of impact device and the shaped support ring see Table 3 and Table 4.Data-View software for TH160 is also optional.

    6 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    8/41

     1.4 WORKING CONDITIONS

    Environmental temperature: 0°C~40°C;Relative humidity: 90%;The surrounding environment should be no vibration, strong magnetic field, corrosivemedium and heavy dust.

    Manual TH-160 7

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    9/41

    2.1 S TRUCTURE

    2.1.1 T HE HARDNESS TESTER APPEARANCE

    1. Main body2. Impact device cable3. Impact device

    2.1.2 N AMES OF THE PARTS

    1. Socket for impact device 14. Enter key2. Reset 15. Print compartment cover3. Charging indicate lamp 16. Buzzer4. Socket for charge 17. Socket for communication cable5. Menu in key 18. Paper compartment6. Left/save key 19. Handle of printer7. Back light key 20. Screw

    8. LCD screen 21. Screw9. On/off key 22. Name plate10. Upper/ paper feed key 23. Screw11. Menu esc key 24. Battery compartment and cover12. Right/delete key 25. Battery compartment and cover13. Down/ print key

    8 Manual TH-160

    2. S TRUCTURE FEATURE &  TESTING PRINCIPLE

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    10/41

    2.1.3 N AMES OF THE PARTS OF D  TYPE IMPACT DEVICE

    1 Release button2 Loading sheath3 Guide tube4 Coil part5 Cable

    6 Impact object7 Support ring

    2.1.4 DIFFERENT TYPES OF IMPACT DEVICES

    2.2 T ESTING PRINCIPLE

    The basic principle is: The impact object of certain weight pounds at the testing surfaceunder certain test force. Measure the impacting velocity and the rebounding velocity of the impact object respectively when it is 1mm above the testing surface. The calculationformula is as followed:

    HL=1000×VB/ VA

    Where,HL -- Leeb hardness valueVB -- Rebounding velocity of the impact objectVA -- Impacting velocity of the impact object

    Manual TH-160 9

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    11/41

    - Accuracy and repeatability of displayed value see Table 5.

    Table 5

    No. Type of Hardness value of Error of displayed Repeatability of impact device Leeb standard value displayed value

    hardness block 1 D 760±30HLD ±6 HLD 6 HLD

    530±40HLD ±10 HLD 10 HLD

    2 DC 760±30HLDC ±6 HLDC 6 HLD530±40HLDC ±10 HLDC 10 HLD

    3 DL 878±30HLDL ±12 HLDL 12 HLDL736±40HLDL ±12 HLDL 12 HLDL

    4 D+15 766±30HLD+15 ±12 HLD+15 12 HLD+15544±40HLD+15 ±12 HLD+15 12 HLD+15

    5 G 590±40HLG ±12 HLG 12 HLG500±40HLG ±12 HLG 12 HLG

    6 E 725±30HLE ±12 HLE 12 HLE508±40HLE ±12 HLE 12 HLE

    7 C 822±30HLC ±12 HLC 12 HLC590±40HLC ±12 HLC 12 HLC

    - Measuring range: HLD(170~960)HLD- Measuring direction: 360°- Hardness test method: HL, HB, HRB, HRC, HV, HS- Display: 320×200 dot matrix LCD- Data memory: 240~1000 groups( Relative to impact times32~1)- Limits range: Same as testing range.- Width of thermal printing paper: 57.5±0.5mm

    - Diameter of printing paper roll: 40mm ( Thermal paper )- Printing speed: 0.27~17.36mm/sec (Grey scale=1)- Printing resolution: 8dot/mm- Working voltage: 3.7V- Charging duration: 2.5~4 hour- Charging power source:6V/500mA- Continuance working period: about 100 hour ( Without printing )- Communication interface:RS232- Dimensions: 230×90×46.5 mm.- Weight: about 0.42kg ( main unit only ).

    10 Manual TH-160

    3. SPECIFICATIONS

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    12/41

     4.1 PREPARATION & INSPECTION BEFORE TESTING

     4.1.1 PREPARATION FOR SAMPLE SURFACE

    Preparation for sample surface should conform to the relative requirement inTable 3.- In the preparation process for sample surface, the hardness effect of being

    heated or cold processing on the surface of sample should be avoided.- Too big roughness of the measured surface could cause measure error. So, the

    measured surface must be metallic luster, smoothing and polish, without oilstain.

    - Curved surface: The best testing surface of sample is flat. When the curvatureradius R of the surface to be tested is smaller than 30mm while use D, DC,D+15,C, E and DL type of impact device or smaller than 50mm while use G typeof impact device, the small support ring or the shaped support rings should bechosen.

    - Support of test sampleSupport is not necessary for heavy sample.Medium-weight parts must be set on the smoothing and stable plane. Thesample must set absolutely equability and without any wobble.

    - The sample should have enough thickness, minimum thickness of sampleshould conform to Table 3.

    - For the sample with hardened layer on surface, the depth of hardened layershould conform to Table 3.

    - couplingLight-weight sample must be firmly coupled with a heavy base plate. Bothcoupled surfaces must be flat and smooth. And there should be no redundant

    coupling agent exists. The impact direction must be vertical to the coupledsurface.

    - When the sample is a big plate, long rod or bending piece, it can be deformedand become unstable, even though its weight and thickness is big enough, andaccordingly, the test value may not be accurate. So should be reinforced orsupported at its back.

    - Magnetism of the sample itself should be avoided.

     4.1.2 S YSTEM SETTING

    See 6.9.

     4.1.3 PRESET TEST PARAMETERS

    See 6.5.

    Manual TH-160 11

     4 PREPARATION &  TESTING

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    13/41

     4.2 T ESTING

    - If necessary, calibrate the TH160 with the standard test block. The accuracy andrepeatability displayed should be within the regulation of Table 5.

     4.2.1 F IX PRINTING PAPER

    - Hold lightly on the prismatic area at two ends of paper compartment cover bypollex and middle finger, turn the paper compartment cover upward, press printhandle fore-upwards, make paper entry at open condition.

    - Put printing paper into the paper compartment, draw printing paper out and putit into the paper entry at the end of paper compartment.

    - Press “Upper/paper feed” key to make printing paper feed in the cutting exit,then press down the printer handle. Hold lightly on the prismatic raised area attwo ends of paper compartment cover by pollex and middle finger and pressdown, until heard a click, the two ends of paper compartment cover fall in thelock entirely.

    - If the printer handle does not return, TH160 will indicate: It can not print now.

     4.2.2 S TART -UP

    - Insert the plug of impact device into the socket on the tester.- Press “on/off” key, now power is connected. The tester is into testing condition.

     4.2.3 L OADING

    - Pushing loading-sheath downwards until contact is felt, then allow it to slowlyreturn to the starting position to lock the impact body.For DC type of impact device, place loading stick next to test area. Plunge faceof impact device(support ring side) over stick and press down till it reaches thestop position.

    - Press impact device on the surface of sample firmly, the impact direction shouldbe vertical to testing surface.

    12 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    14/41

     4.2.4 T ESTING

    - Press release button at top of impact device to test. The sample and the impactdevice as well as the operator are all required being stable now, and the actiondirection should pass the axis of the impact device.

    - Usually, test 5 times on each measure area of sample. The data dispersionshould not be more than mean value±15HL.The distance between any two indentations or from the center of anyindentation to the edge of tested sample should conform to the regulation ofTable 6.

    Table 6

    Type of Impact Device Distance of center of Distance of center of thethe two indentations indentation to sample edgeno less than (mm) no less than (mm)

    D,DC 3 5DL 3 5D+15 3 5G 4 8E 3 5C 2 4

    - For special material, if accurate conversion from the Leeb hardness into other

    hardness method is wanted, contrastive test must be done to get conversionrelations. Use inspected qualified Leeb hardness tester and corresponding otherhardness tester to test at same sample respectively. For each hardness value,measure 5 Leeb hardness values. using Leeb hardness arithmetic average valueand corresponding hardness method average value as correlative value to makeindividual hardness contrastive curve. Contrastive curve at least should include threeunits of correlative data.

     4.2.5 READ MEASURED VALUE

     4.2.6 PRINT OUT THE RESULT 

    See 6.3.3 and 6.6.

     4.2.7 PRESS “ON/OFF ”  TO SWITCH OFF .

     4.2.8 T EST RESULT PROCESSING

    Take the arithmetic average value of 5 efficient testing points as a Leeb hardnesstesting data.

     4.2.9 T ESTING RESULT 

    - In front of Leeb hardness symbol HL, expresses the hardness value; behind HL

    expresses type of impact device. For example: 700HLD expresses using D typeof impact device measured Leeb hardness value is 700.

    - For other hardness method converted from Leeb hardness, put the symbol ofcorresponding hardness method in front of Leeb hardness symbol. For example:400HVHLD expresses using D type of impact device measured Leeb hardnessvalue converted Vickers hardness value is 400.

    Manual TH-160 13

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    15/41

    - Exchange the impact device in the power off condition. Otherwise it can not identifyimpact device automatically, and the circuit board of tester might be damaged.

    - In normal condition, it could not print present testing value while not reachsetting Average . If want to print, press “Menu esc”key at the main display interfaceto end the testing in advance, then print.

    - When the test is end in advance as mentioned above, all the functions of Autosave, Auto print and Auto transfer etc., of System set menu will not work.

    - When the Impact direction was set as AUTO, if the Impact direction is or,the user need to confirm it manually.

    - G type impact device has not the function of identifying impact directionautomatically. So its Impact direction can not be set as AUTO ( the cursor couldnot be moved at this symbol in the menu). If the user used other impact device andset Impact direction as AUTO , then exchanges to G type impact device, theImpact direction will be changed to? automatically.

    - Only D type impact device has the function of strength measure. So, when other

    type of impact device is used,Set Hardness orσ

    b can not be select ( The cursorcould not be moved at Set Hardness or σ

    b). If the user used D type impact deviceand set to σb, then exchanged to other type impact device, Set Hardness or σbwill be set as Hardness automatically.

    - When the user sets σb to measure, he can not set hardness method ( the cursorcould skip over Hardness Scale).

    - Not all hardness methods can be converted to each other for all materials. Thesetting of hardness scale will recover to HL automatically after material setting ischanged. So select material first then select hardness method when setting

    measure condition.

    - Reset switch is used to cut off the power thoroughly. All settings will be recoveredto ex-work settings after it is turned off. It needs readjustment using standardhardness block when the TH160 is used again. Data in memory will be lost too. Theswitch is sealed by a round piece of paper in 'On' position. Break it first when theswitch must be used.

    14 Manual TH-160

    5. SPEC IAL PROMPT  

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    16/41

    6.1 S WITCH ON

    Press “On/off” to switch on, the screen shows:

    Then, it enters the main display interface.

    Note:The tester will switch on automatically, while connecting charging power in the condition of switch off.

    6.2 S WITCH OFF 

    In any conditions, Press key “On/off” , it will be switch off.

    Note:In the charging state, the tester will switch on again automatically on the instant when switch off, to monitoring charging condition.

    6.3 T ESTING

    This tester can measure in two display interfaces named the main display Interface andthe aecondary display Interface.The tester displays the main display Interface automatically after switch on as follows:

    Adopting large character to display measured value and rich information, with fastoperation of memory and printing, this interface is used in most conditions.

    Manual TH-160 15

    6. OPERAT INO EXPLA IN IN DETAI LS

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    17/41

    6.3.1 INSTRUCTION OF THE MAIN DISPLAY INTERFACE

    Date: Current date: month/day/year.Time: Current time: hour/minute/second.Memory information: Percent of memory used in the whole memory

    capacity.Battery information: Display the percentage of rest capacity or display

    charging degree while the tester is charging.Over tolerance alarm: When the present measured value or average

    value is over the tolerance limit, it will appear,otherwise it will not.

    Lower than tolerance alarm: When the present measured value or average

    value is lower than tolerance limit, it will appear,otherwise not.Analogue indicator: Use the height of the black post to display the

    position of present measured value or averagevalue in the tolerance.

    Impact device type: Showing the impact device type in use.Impact direction: Showing the present impact direction.Average value indicator: After reaching the preset impact times, It appears

    to show the mean value.Hardness scale: Hardness method of the present measured value.Measured value: Display present single time measured value

    (without average value indicator), or display thepresent average value (with average valueindicator). Display ↑ means over conversion ormeasured range, ↓ means lower thanconversion or measured range.

    Material: The present preset material.Impact times' count: Times that have been impacted.Impact times: The preset impact times.On line help: Instruction of main operation that could proceed

    presently.

    6.3.2 T HE MAIN DISPLAY INTERFACE TESTING OPERATION

    The user can measure under this Interface. It displays the current measured valueafter each measurement; Analogue indicator displays according to measuredvalue; 1 is added to the impact times' count. If the measured value exceeds thetolerance limits, Over tolerance alarm or Lower than tolerance alarm willappears, and the buzzer will alert a long howl. After reaching the preset impacttimes, the buzzer will alert a long howl, and after 2 seconds, the buzzer will alert ashort howl, and display the average value.

    16 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    18/41

    6.3.3 T HE MAIN DISPLAY IINTERFACE KEY OPERATION

    - Press “save”key to save present number value only valid in condition ofdisplaying the average value, and only conserve it for one time.

    - Press “delete” key to delete the latest single measured value, after conform it asfollowing interface:

    Press “delete” key, move the cursor to Yes and press “enter” key to confirm todelete the latest single measured value.

    - Press “delete” key, move the cursor to No and press to cancel the delete

    operation.- Press “esc” key could cancel the delete operation also, no matter where the

    cursor is.- Press “paper feed” key to feed the paper for the thermal printer.- Press “print” to print present unit value. It is only valid after displaying the

    average value. The user can print multi-times.- Press “esc” key could end the testing even if it does not reach the preset impact

    times, and display the average value.- Press “back light” key could switch on or off the background light of LCD- Press “enter” key could enter the secondary display interface.

    - Press “menu” key could enter the main menu.

    At the main display interface, press “enter” key to enter the secondary displayinterface, as following:

    This interface displays all single measured value, average value, maximum,minimum value and histogram at any moment after each measurement. It is usedin the case of immediate detailed analysis for the value.

    Manual TH-160 17

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    19/41

    6.3.4 INSTRUCTION OF THE SECONDARY DISPLAY INTERFACE

    Histogram: The two transverse lines are the preset tolerance limits,the whole display range could be 2~5 times of thetolerance limit.

    Average value: The average value of all measured values of presentgroup.

    Maximum: The maximum value in all measured values of presentgroup.

    Minimum: The minimum value in all measured values of presentgroup.

    Single measured value: every single measured value of present group.

    Arrangement order is from left to right.Impact times: Preset and finished impact times. The first number is theimpact times which has been finished, the secondnumber is the preset impact times.

    Cursor position: The cursor is on one of the single measured values.On line help: Instruction of present main operation available.

    6.3.5 T HE SECONDARY DISPLAY INTERFACE TESTING OPERATION

    The user can measure under this interface. It displays the current measured valueafter each measure. The measured value of this time is increased into histogramalso. The tester recalculates and displays the average value, maximum andminimum value and impact times +1. If the measured value exceeds the tolerancelimits, the buzzer will make a long beep. After reaching the preset impact times, thebuzzer will make a long beep too.

    6.3.6 T HE SECONDARY DISPLAY INTERFACE KEY OPERATION

    - Press “left”, “right”, “up” or “down” key to move the cursor.- Press “enter” to delete the single measured value at the cursor.

    Confirm to delete as followed :

    Press “left” or “right” key to move the cursor to Yes , and press “enter” to confirm it.Press “enter” could confirm deleting the last single measured value.Press “left” or “right” key move the cursor to No , press “enter” to cancel deleting

    operation.Press “esc” key can cancel deleting operation too, no matter where the cursor is.

    - Press “esc” key to return to the main display interface.- Press “back light” to open or close the background light of LCD.- Press “menu” to enter the main menu.

    18 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    20/41

    6.4 MENU STRUCTURE

    Both preset tester parameters and the additional functions can be performed by themenu operation. At the main display interface or the secondary display interface, press“menu” to enter the main menu.

    Manual TH-160 19

    Impact DirectionAverageMaterialHardness ScaleTolerance LimitHistogram Display RangeSet Hardness or σb:Hard

    Print current dataPrint data by Memory No.Print All Data of MemoryPrint Current Data & MapPrint Data & Map by Mem No.

    Print All Data & Map Memory

    View from No.1View from EndView from select No.TransferDelete by No.Delete All

    Auto Save: Off Auto Print: Off Auto Delete: Off Auto Transfer: Off Key Sound: OnWarning Sound: OnPrint Intensity: 1LCD BrightnessTime Date Set

    Language

    The Main Display Interface

    The Secondary Display Interface

    Set Test ParameterPrint FunctionMemory ManagerSystem SetAbout Software

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    21/41

    6.5 PRESET TEST PARAMETERS

    At the main display interface or the secondary display interface, press “menu”key toenter the main menu.

    Press key “enter” key to enter Set Test Parameter menu.

    Press “up” or “down” to move the cursor to the line you need to set, and press “enter” toconfirm it.

    Note: 1. When switch Hardness or σ 

    b to σ b , The hardness scale could not be selected,The cursor will skip over Hardness Scale while moves the cursor.

    2. Only D type of impact device has the function of σ b measure, so, the cursor could not move toHardness or σ b while use other impact device.

    20 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    22/41

    6.5.1 PRESET IMPACT DIRECTION

    Press ‘left” or “right” key to move the cursor to the impact direction that you willpreset, and press “enter” key to confirm it. Press “esc” key to cancel it.

    Note:1. The tester could adjust the value according to identified impact direction while preset impact direction

    as automatic.2. G type impact device has not the function of identify impact direction automatically.

    3. But if the direction is at or  , the user need to confirm it manually.Press “left” or “right” key to move the cursor to the actual impact direction, and press key to confirm it.

    Manual TH-160 21

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    23/41

    6.5.2 PRESET AVERAGE TIMES

    The average times can be set within the range of 1~32.Press “left” or “right” key to move the cursor.Press “up” to increase the number or press “down” to reduce the number at thecursor.Press “enter” to confirm the modifying.Press “menu esc” key to cancel the modifying.

    6.5.3 PRESET M ATERIAL 

    - In Set Hardness or σb, if preset to hardness, it will display the following material:

    Press “up” or “down” key to move the cursor to the material want to preset, andpress “enter” key to confirm it.Press “esc” to cancel the preset.

    22 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    24/41

    Note1. Preset of hardness scale will be recovered as HL automatically after material preset is changed.2. Please select material first, and then select hardness scale.

    3. The symbol  at left-down side of the menu means that the menu is not end, press “down” key tocontinue downward.The symbol at left - up side of menu means that the menu is not end, press to find upward.

    - In Set Hardness or σb, if preset to σb, it will display the following material:

    Press “up” or “down” key to move the cursor to the material want preset,Press “enter” key to confirm it.Press “menu” to cancel preset.

    Note1. The symbol at left-down side of menu means that the menu is not end, press “down” to continue.

    The symbol at left -upside of menu means that the menu is not end, press “up” to find upward.

    Manual TH-160 23

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    25/41

    6.5.4 PRESET HARDNESS SCALE

    Press “left” or “right” key to move the cursor to the hardness scale need to set.Press “enter” to confirm or press “menu esc” key to cancel the setting.

    Note:1. Here only displays the hardness scale existed for the present selected impact device and material.2. Please select material first, then select hardness scale.3. Preset of hardness scale will recover to HL automatically after preset of material is changed.

    6.5.5 PRESET TOLERANCE LIMIT 

    Press “left” or “right” key to move the cursor.Press “up” or “down” key to remark the value at the cursor.Press “enter” key to confirm the setting.Press “menu esc” to cancel the setting.

    Note:1. If the set value exceeds the measuring range, the tester will remind you to set again.

    2. If the bottom limit setting is bigger than the upper limit, they will exchange automatically.

    24 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    26/41

    6.5.6 PRESET HISTOGRAM DISPLAY RANGE

    The ratio of display range of histogram value to tolerance limit range could be set in2-5 times.Display range to maximum at 5 times, minimum at 2 times.Press “up” or “down” key to choose the multiple.Press “enter” key to confirm the setting.Press “menu keys” to cancel the setting.

    6.5.7 PRESET HARDNESS OR σB

    Press “enter” to choose Hardness or σb.The cursor switch between Hardness and σb.

    Manual TH-160 25

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    27/41

    6.6 PRINT FUNCTION

    In the main display interface or the secondary display interface, press “menu” key intothe main menu.

    Press ”up” or “down” key to move the cursor to?Print Function?.Press “enter” key to enter Print Function menu.

    Press ”up” or “down” key to move the cursor to the printing function wanted,then press “enter” to printing.

    6.6.1 PRINT CURRENT DATA

    Note:1. During the printing process, keep pressing “menu esc” key can halt the print after printing present line.2. After a group of test, press “down” to printing present number value of any group.

    Print Current Data can print: tester name, number,

    operator, time, date, impact device type, impact direction,average times, material, single measured value andaverage value.

    If it is not switched off, and the testing condition is notchanged, only single measured value and average valuewill be printed when print again.

    Note: Number and operator need to be filled manually.

    26 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    28/41

    6.6.2 PRINT DATA BY MEMORY GROUP NO.

    In Print Data by Memory No. , select the groups range first. The display shows thepreservative group range in the memory.Press “left” or “right” to move the cursor.Press “up” or “down” to modify the value at the cursor.Press “enter” to start printing.Press “menu esc” keys to cancel the printing order.

    Note:

    1. If the preset group number exceeds actual range, print available group.2. To print 1~5 group, you can preset from 1 to 5, or from 5 to 1.3. The bigger the number of the group is, the closer the group to current group number.

    Print contents include: tester name, number,operator, time, date, impact device type, impactdirection, average times, material, singlemeasured values and average value.

    If the date, impact device type, impact direction,averaging times, material or hardness scale of the

    next group are the same as those of the formerone, only the number of unit, single measuredvalues and average value are printed. Otherwise,all content as figure display are printed.

    6.6.3 PRINT ALL DATA OF MEMORY 

    Print All Data of Memory Print the values of all groups in the memory as sameformat.

    Manual TH-160 27

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    29/41

    6.6.4 PRINT CURRENT DATA & HISTOGRAM

    Print Current Data & Map: Besides testername; numbering; operator; time; date;impact device type; impact direction;averaging times; material; single measuredvalues and average value, the histogram of present group is also printed.

    The two cross-lines are tolerance limits in thehistogram.

    If not switched off, and the testing conditionhas not be changed, only single measuredvalues, average value and histogram areprinted when printing again.

    28 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    30/41

    6.6.5 PRINT DATA & HISTOGRAM BY MEMORY GROUP NO.

    Print Data & Map by Memory No. Need to selectgroup range for printing too. The operation is thesame as Print Data by Memory No. . Comparing toPrint Data by Memory No., histogram is added .

    6.6.6 PRINT ALL DATA & HISTOGRAM OF MEMORY 

    Print All Data & Map Memory Prints the number values and histogram in allgroups of the memory in the same format.

    Manual TH-160 29

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    31/41

    6.7 MEMORY MANAGEMENT 

    In the main display interface or the secondary display interface, press “menu” key toenter the main menu.

    Press “up” or “down” key to move the cursor to Memory Manager .Press “enter” key to enter Memory Manager menu.If there is no data in the memory, it displays: and return.

    6.7.1 V  IEW FROM NO. 1 GROUP / V IEW FROM THE LAST GROUP

    Press “up” or “down” key to move the cursor to the function wanted, press “enter”to confirm.

     View from No. 1:Start to display number value in the memory from the first group. View from End: Start to display number value in the memory from the last group.

    30 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    32/41

    6.7.2 V IEW FROM SELECTED GROUP

     View from Select No.: Selecting the beginning group.Press “left” or “down” to move the cursor.Press “up” or “down” to modify the number at the cursor.Press “enter” to display data in the memory from the selected beginning group.Press “menu esc” key to cancel the operation.

    6.7.3 D ATA TRANSFER

     Transfer : Export the number value in the memory as text format from interfaceRS232 .

    6.7.4 DELETE SELECTED GROUP

    Delete by No.: Select the range of deleting group.Press “left” or “right” key to move the cursor.Press “up” or “down” to remake number at the cursor.Press “enter” to delete the selected group.Press “menu esc” to cancel the operation.

    Note:

    1. If the preset group number exceeds the actual range, then deletes the actual group among them.2. To delete 1~5 group, you can preset from 1 to 5, or from 5 to 1.3. After operation finished, the order of the rest groups number will realign.

    Manual TH-160 31

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    33/41

    6.7.5 DELETE A ALL 

    Delete All: Delete all data in the memory.

    6.7.6 DELETE CONFIRMATION

    Press “left” or “right” key to move the cursor to YES , Press “enter” to confirm todelete.Press “left” or “right” key to move the cursor to NO , press “enter” to cancel thedelete operation.Press “menu esc” could cancel delete operation, no matter where the cursor is.

    6.8 BROWSING DISPLAY SCREEN

    Press “up” or “down” key to turn the page for browsing.Press “menu esc” exit browsing.

    32 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    34/41

    6.9 S YSTEM SET 

    In the main display interface or the secondary display interface, press “enter” key toenter the main menu.

    Press “up” or “down” to move the cursor to System Set .Press “enter” key to System Set menu.

    Press “up” or “down” to move the cursor to the project demand to set, press “enter” tomodify directly or into corresponding screen.Press “menu esc” return.

    For Auto Save, Auto Print , Auto Delete, Auto Transfer , Key Sound and WarningSound, use “enter” key to choose On or Off . When Auto Save is set On, it stores thedata of current group automatically after measuring and displaying average value.When Auto Print is set On, it prints the number value of current group automaticallyafter measuring and displaying average value. When Auto Delete is set On, accordingto 3s rule, it cancels gross error automatically after measuring the preset average timesor press “menu esc” key in advance. If there is data canceled, supplemental

    measurement is needed to reach the preset times.When Auto Transfer is set On, it exports the number value of present group in textformat from interface RS232 after measuring and displaying average value.When Key Sound is set On, the buzzer will make a short hoot while pressing the keyeach time. When Warning Sound is set On, when the measured value exceeds thetolerance limit, after reaching the preset average times or deleting data, the buzzer willmake a long hoot.

    Manual TH-160 33

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    35/41

    For Print Intensity, use to select 1?2?3?4?5, totally 5 grade of grayness. The bigger thenumber is, the dark the printed literal or symbol will be. It applies to the condition of 

    lower environment temperature etc. But meanwhile, the print speed will be lower.

    6.9.1 PRESET LCD BRIGHTNESS

    The higher the brightness is, the darker the color is.The lower the brightness is, the shallower the color is.

    6.9.2 PRESET TIME AND DATE

    Press “left” or “down” key to move the cursor.Press “up” or “down” to modify the number at the cursor.Press “enter” to confirm.Press “menu esc” to exit.

    6.9.3 PRESET LANGUAGE

    34 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    36/41

    Press “up” or “down” key to move the cursor to the Language wanted.Press “enter” key to confirm it.

    Press “menu esc” to cancel the presetting.

    6.10 SOFTWARE INFORMATION

    In the main display interface or the secondary display interface, press “enter” key to themain menu.

    Press “up” or “down” key to move the cursor to About Software.Press “enter into About Software .

    This interface displays the information about the tester and firmware.

    Manual TH-160 35

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    37/41

    6.11 SOFTWARE CALIBRATION

    The tester and impact device should be adjusted using standard test block before usethis tester for the first time, or reuse after a long time, or use it after switched off thereset switch.While one main unit configures with multi-type of impact devices, each type needs toadjust for one time only. No readjust is needed after changing probe later.

    Press “enter” key and “on/off” key simultaneously. The TH160 is switched on and itenters the software calibration state.

    Preset the impact direction as .

    Measure 5 points on the standard hardness block.It will display average value after measurement.

    Press “up” or “down” key to input nominal value.Press “enter” key to confirm it.

    Press “menu esc” to cancel.Range of adjustment: ±15HL.

    36 Manual TH-160

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    38/41

    6.12 CHARGING

    - When battery capability run out, the battery display symbol will glint:

    Please charge as soon as possible for the tester.

    - Charging indicating lamp will be lighting while charging, indicates that chargepower is connected. If the tester is in power off condition, it will switch onautomatically. The battery symbol alternately shows:

    - When the charging is finished, it will glint:

    - The reset switch should be at On condition when charging.

    6.13 B ACKGROUND LIGHT 

    The LCD screen has background light. It is convenient to use in dark condition. Press“back light” key to open or close back light at any moment after switched on.

    6.14 AUTO POWER OFF 

    - The tester has the function of auto power off to save the power of battery.- If there is neither measurement nor any key operation within 5 minutes, the tester

    will shut off automatically. The screen will glint for 20 seconds before power off.Except “on/off” key, press any key to stop the twinkle of LCD screen and stop theoperation of power off at the moment.

    - While voltage of battery is too low, the tester will show < Battery Empty!>, thenpower off automatically.

    6.15 B ATTERY REPLACING

    - Screw off four screws on the battery cover, take off the cover of battery;- Take off the battery;- Insert the connection plug of new battery into the socket on the circuit board.

    Reset the battery cover.

    6.16 CONNECTION OF DATA TRANSMISSION CABLE

    Insert one connection plug of transmission cable (Optional parts) into the RS232 socketon the left side of main unit, and insert the other plug into the 9 pin socket in the backof computer box.

    Manual TH-160 37

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    39/41

    Problem Analysis Trouble shootingFail to switch on The reset switch is not open. Open the reset switch in one

    side of the TH160.Battery is exhausted Charging

    Recharge failure The reset switch is not open. Open the reset switch in oneside of the TH160.

    38 Manual TH-160

     7. T ROUBLE SHOOTING

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    40/41

    8.1 IMPACT DEVICE MAINTENANCE

    After the impact device is used for 1000--2000 times, use the cleaning brush provided toclean the guide tube and the impact body. To clean the guide tube, unscrew thesupport ring first, then take out the impact body, spinl the cleaning brush in counter-clock direction into the bottom of guide tube and take it out for 5 times, and then installthe impact body and support ring again.- Release the impact body after use.- Any lubricant is absolutely prohibited inside the impact device.

    8.2 CONSERVATION OF PRINTING RECORD

    Because the print paper is thermal, please avoid high temperature and direct lightirradiation during preservation. If record of long time is need, please copy forpreservation.

    8.3 M AINTENANCE

    - When using standard Rockwell hardness block to test, if all the error is bigger than 2HRC, maybe the test tip is invalid because of abrasion. Changing the test tip orimpact object should be considered.

    - When the hardness tester appears any other abnormal phenomena, please do notdismantle or adjust any fixedly assembled parts by yourself. You can contact withyour agent.

    Manual TH-160 39

    8. M AINTENANCE

  • 8/20/2019 Manual Medidor de Dureza Th160_ba_e

    41/41

    Please keep it away from vibration, strong magnetic field, corrosive medium,dumpiness and dust.

    9 NOTICE OF TRANSPORTATION AND STORAGE