MEASUREMENT OF OIL AND GAS Accuracy: difference from the actual measurement and the meter reading, accuracy is stated in following terms: Repeatability :meter’s ability to reproduce same measurement for a set of constant conditions of flow rate, temperature, viscosity, density, pressure. Repeatability of a custody transfer meter should be within +/-.025% in 3 runs or +/- 0.05% of each other in 5 consecutive prove runs. Linearity :ability to maintain a meter factor through-out the stated turndown. Depending on meter size and application this is typically +/- 0.15% or +/-0.25% for custody transfer use.

Measurement of Oil and Gas

  • Upload
    mts1234

  • View
    36

  • Download
    0

Embed Size (px)

DESCRIPTION

This presentation presents various methods to measure oil and gas in petroleum industry.It deals with basics of various types of meters,working principles, various parameters for accuracy etc

Citation preview

Page 1: Measurement of Oil and Gas

MEASUREMENT OF OIL AND GAS

Accuracy: difference from the actual measurement and the meter reading, accuracy is stated in following terms:

Repeatability :meter’s ability to reproduce same measurement for a set of constant conditions of flow rate, temperature, viscosity, density, pressure. Repeatability of a custody transfer meter should be within +/-.025% in 3 runs or +/- 0.05% of each other in 5 consecutive prove runs.

Linearity :ability to maintain a meter factor through-out the stated turndown. Depending on meter size and application this is typically +/-0.15% or +/-0.25% for custody transfer use.

Page 2: Measurement of Oil and Gas

Resolution. is a measure of the smallest increment of total flow that can be individually recognized by the meter.

Turndown. Turndown is the meter’s flow range capability. The flow range of the meter is the ratio of maximum flow to minimum flow over which the specified accuracy or linearity is maintained. a meter with a minimum flow rate of 100 bbl/hr and a maximum flow of 1,000 bbl/hr is said to have a 10:1 turndown.

Page 3: Measurement of Oil and Gas

SELECTION CONSIDERATIONS

• FLOW RATE• PRESSURE• TEMPERATURE• VISCOSITY• ACCURACY

Page 4: Measurement of Oil and Gas

TYPICAL POSITIVE DISPLACEMENT METER

Page 5: Measurement of Oil and Gas
Page 6: Measurement of Oil and Gas
Page 7: Measurement of Oil and Gas
Page 8: Measurement of Oil and Gas

MEASURING ELEMENT

low pressure drop low mechanical friction non jamming rotation high driving torque

Page 9: Measurement of Oil and Gas

POSITIVE DISPLACEMENT METERS

FLOW OUTPUT mechanical counters electrical ACCURACYFactors affecting accuracy:1.Displacement,Temperature,Wear,Viscosity,coating2.Slippage Hydraulically induced : flow, viscosity Mechanically induced : friction, wear

Page 10: Measurement of Oil and Gas

PROS AND CONS

• Low cost• Good for viscous liquids

• High maintenance cost• High pressure drop• Not suitable for low flow rates• Clean fluid with no solids

Page 11: Measurement of Oil and Gas

INDIRECT METERS INFERENCE METERS

• TURBINE METERS• CORIOLIS METERS PRINCIPLE : INFERS FLOW BY MEASURING SOME DYNAMIC

PROPERTY OF THE FLOW STREAM

Page 12: Measurement of Oil and Gas

TURBINE METERS

• Principle: Angular velocity of rotor is proportional to linear velocity of liquid

• DESIGN CONSIDERATIOS: Housing• Measuring Element : Rotor with multiple blades on a spindle

mounted on a free running bearing assembly FLOW OUTPUT: Voltage pulse

Page 13: Measurement of Oil and Gas

Accuracy based on two assumptions: (1) the flow area remains constant and; (2) the rotor velocity accurately represents the stream velocity

“K” factor (pulses / unit volume), Erosion, Corrosion, Deposits, Boundary Layer

Thickness, Cavitation , Obstructions Rotor velocity Rotor Blade Angle, Rotor

Stability ,Bearing Friction, Velocity Profile,Swirl, Fluid Density

Page 14: Measurement of Oil and Gas

TYPICAL TURBINE METER

Page 15: Measurement of Oil and Gas

Typical turbine meter

Page 16: Measurement of Oil and Gas
Page 17: Measurement of Oil and Gas

OSCILLATION

Page 18: Measurement of Oil and Gas

TWIST

Page 19: Measurement of Oil and Gas

FLOW RATE AND VISCOSITY FACTORCHOICE OF FLOWMETER FOR LIQUIDS

Page 20: Measurement of Oil and Gas
Page 21: Measurement of Oil and Gas

Meter Proving. Meter proving is the physical testing of the performance of a liquid meter in a liquid service. The main purpose of the test is to assure accuracy.

Meter factor = prover known volume meter reading

Page 22: Measurement of Oil and Gas

While proving a meter, the process-fluid conditions must be as stable, includes temperature, pressure, flow rate, and density. Before starting a meter proving, let the liquid flow through the meter and prover long enough so that the conditions stabilize. Check for leaks or fluid bypassing around the prover or meter. The only way to obtain a reliable meter factor is to have all the liquid that is measured by the meter also measured by the prover.

Page 23: Measurement of Oil and Gas

LACT Units. LACT units are designed for unattended custody transfer of crude oil from a seller to a buyer. Flow rate, operating pressure, gravity, and temperature of the oil determine the LACT design. Minimum pressure drop through the piping and components is desirable.

Page 24: Measurement of Oil and Gas

TYPICAL LACT SYSTEM

Page 25: Measurement of Oil and Gas
Page 26: Measurement of Oil and Gas

TURBINE METER

Page 27: Measurement of Oil and Gas

CORIOLIS METER

Page 28: Measurement of Oil and Gas

ORIFICE METER

• Principle: differential pressure proportional to squire of

flow rate Standards: AGA 3 ISO 5167

Page 29: Measurement of Oil and Gas

BASIC EQUATION: qh =C’ (hw pf )1/2

where: qh - quantity rate of flow at base

conditions, cfh C’- orifice flow constant, hw differential pressure in inches of water at

600F, pf absolute static pressure, psia

Page 30: Measurement of Oil and Gas

C’=(Fb)(Fr)(Y)(Fpb)(Ftb)(Ftf)(Fg)(Fpv)(Fm)(Fl)(Fα)

Where: (Fb) basic orifice factor, cfh(Fr) Reynolds number factor (Y) expansion factor (Fpb) pressure base factor (Ftb) temperature base factor (Ftf) flowing temperature factor (Fg) specific gravity factor(Fpv) supercompressibility factor(Fm) manometer factor for mercury meter(Fl)gauge location factor,(Fα) orifice thermal expansion factor

Page 31: Measurement of Oil and Gas
Page 32: Measurement of Oil and Gas

Typical orifice meter installation details

Page 33: Measurement of Oil and Gas

GAS TURBINE METERS

• Principle: flow rate is proportional to velocity of gas Standards: ISO 9951 OIML R32

Page 34: Measurement of Oil and Gas

Turbine meter installation

Page 35: Measurement of Oil and Gas

ULTRASONIC METERS

Principle: Difference in transit time of high frequency

sound waves travelling between a pair of fixed sound transducers with the flow and against the flow determines the flow

Page 36: Measurement of Oil and Gas

TYPICAL ULTRASONIC METER

Page 37: Measurement of Oil and Gas
Page 38: Measurement of Oil and Gas
Page 39: Measurement of Oil and Gas

Ultrasonic meter installation