Reprocessing of Single Use Devices (SUDs)

MEDEC Regulatory ConferenceMay 9, 2016

Page 2: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Objective:– Health Canada’s requirements around licensing of

Reprocessed Single Use Devices (RSUDS)– Content to be provided in a licence application for RSUDs

• Outline:– Description/definition of RSUDs– Regulatory approach to RSUDs– Licence applications for Class III/IV devices– Expectations and requirements for reprocessors

• Labelling• Device tracking• Cleaning validation• Device validation

Objective and Outline

Page 3: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Single-use medical devices (SUDs) are intended by their manufacturers to be used once during a single procedure and not to be cleaned and reused.

• Reprocessed SUDs are those that have previously been used on a patient and have been subjected to additional processing and manufacturing for the purpose of an additional single use on a patient.

• Devices that may be candidates for RSUDs:– Surgical instruments, drills, burrs, catheters (PTA, EP),

compression sleeves, hemodialysis accessories, etc.

Reprocessed Single Use Devices

Page 4: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessing a medical device involves cleaning, reconditioning, function testing, and disinfection or sterilisation.

• A hospital’s use of third-party commercial reprocessorsmay be either:– A closed-loop procurement model (the hospital receives only its own

devices back from the third-party reprocessor)– An open-loop model (the hospital does not get its own devices back

but rather buys them from a pool of reprocessed devices).

Reprocessing Single Use Devices

Page 5: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

Background on Regulatory Approach

• Health Canada published a notice outlining the regulatory approach to commercial reprocessing of SUDs

http://www.hc-sc.gc.ca/dhp-mps/md-im/activit/announce-annonce/md_notice_sud_uu_avis_im-eng.php

• All RSUDs are to be licensed by September 1, 2017

… will

application.

… will be held to the same requirements as manufacturers of new devices. This means they must meet requirements for licensing, quality system management, labelling, investigating and handling complaints, maintaining distribution records, conducting recalls, reporting incidents and informing Health Canada of any changes to the information in their licence application.

Page 6: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

Regulatory Approach

6

• Health Canada will issue medical device licences to commercial reprocessors who reprocess SUDs for healthcare facilities.– Reprocessed devices will keep the same risk classification as the

original equipment manufacturer (OEM) device.– Reprocessors must demonstrate that the reprocessed devices they

sell meet the safety and effectiveness requirements of the Medical Devices Regulations either directly, or by demonstrating equivalence to a currently licensed device.

• Reprocessors of licensed SUD’s will not be required to demonstrate that the OEM devices meet the safety and effectiveness requirements of the Medical Devices Regulations.

• Reprocessors of unlicensed SUD’s will be required to demonstrate, through the provision of objective evidence, that the OEM devices meet the safety and effectiveness requirements of the Medical Devices Regulations

Page 7: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

Regulatory Approach

7

• In addition to the standard medical device licence application package, the reprocessor should submit:

1. An overview specifically outlining whether the device in question is appropriate for reprocessing, including:a. Description of the deviceb. Detailed risk analysisc. Data analysis of reverse engineering d. Performance specifications (functional specs., cleanliness, SAL,

etc.) including a justification that the specifications are appropriate.

2. Complete and detailed description of the entire validated reprocessing process / procedures, including initial decontamination, cleaning, acceptance testing, refurbishing, packaging, labelling, sterilisation, and release.

Page 8: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

3. Validation of post-processing device performance, in accordance with item 1 above, up to the maximum number of cycles, including:a. Acceptance criteriab. Pre-clinical testingc. Biocompatibility, etc.

4. Packaging validation including description of sterile barrier system and stability studies.

5. Product labelling, including IFU’s and product marking/etching scheme.

6. A method of tracking the number of processing cycles that have been applied to a given device.

Regulatory Approach

8

Page 9: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Requirements for licence applications the same as any other manufacturer.

• Guidance Document: Information to Be Provided by Manufacturers for the Reprocessing and Sterilization of Reusable Medical Devices

http://www.hc-sc.gc.ca/dhp-mps/md-im/applic-demande/guide-ld/md_gd_reprocessing_im_ld_retraitement-eng.php

• Guidance Document: Guidance on supporting evidence to be provided for new and amended licence applications for Class III and Class IV medical devices, not including In Vitro Diagnostic Devices (IVDDs)

http://www.hc-sc.gc.ca/dhp-mps/md-im/applic-demande/guide-ld/md_gd_data_im_ld_donnees_ciii_civ-eng.php

Regulatory Approach

Page 10: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

Device Licence Applications - Class III and Class IV 1. Background Information

– Device Description– Design Philosophy– Indications

2. Labelling

3. Marketing History– Canadian and International– Incident Reports

4. Manufacturing and Quality Control

– Materials– Biological materials– Quality Plan

– Manufacturing Process– Process Validation Studies– Sterilization Validation

• Bioburden, pyrogen, residuals– Packaging Validation

5. Safety and Effectiveness Studies– List of standards– Preclinical studies

• Bench testing, shelf life, biocompatibility, animal studies

– Clinical studies– Literature Studies &

Bibliography

6. Risk Assessment

Page 11: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessors will be required to include their name and address on the labelling of the SUDs they reprocess and will therefore be considered the ‘manufacturer’ as defined in the MDR.

Expectations for Reprocessors

11

Page 12: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Section 9 of the MDR requires that the manufacturer ensures that its devices meet the safety and effectiveness requirements (S&E) [Sec.9(1)] and to keep objective evidence to establish that its devices meet the S&E requirements [9(2)].

• Reprocessors of all risk classes of RSUD’s will therefore be required to maintain objective evidence to demonstrate that their RSUD’s meet the S&E requirements of the MDR.

Expectations for Reprocessors

12

Page 13: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Clause 4.2.1 of ISO 13485:2003 requires that a manufacturer establish and maintain, for each type or model of device, a file either containing or identifying documents defining product specifications and quality management system requirements. These documents shall define the complete manufacturing process.

Expectations for Reprocessors

13

Page 14: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessors will need to develop a complete, documented set of acceptance criteria and functional requirements for each type and model SUD to be reprocessed.

• This should include a documented and justified rationale for the suitability of each type and model of SUD for reprocessing.

Expectations for Reprocessors

14

Page 15: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• This also entails a detailed risk analysis covering:– the device – the maximum number of reprocessing cycles– acceptance criteria for incoming used SUD’s– the decontamination and cleaning process– any refurbishment activities– functional performance criteria and testing activities– packaging and labelling activities– the sterilization process– tracking the number of reprocessing cycles– identifying the device in the field as reprocessed– tracking of design changes made by the OEM

Expectations for Reprocessors

15

Page 16: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessors are expected to have documented acceptance criteria and functional / performance specifications.

• Preference is given to using the OEM’s specifications.

• Reverse-engineering of specifications is possible, but this requires justification through, for example, pre-clinical validation.

• Monitoring for OEM changes to the device.

Expectations for Reprocessors

16

Page 17: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• RSUD’s shall be labelled as being ‘reprocessed’ and shall bear the name and mark of the reprocessor as well as a device identifier specific to the reprocessed version of the SUD.

• The OEM’s information can remain on the device and label for post-market and compatibility issues, but the reprocessor’s information shall be prominent.

Labelling Requirements

17

Page 18: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessed SUD’s shall be permanently marked (indelible ink, laser etching, etc.) to indicate the number of use and reprocessing cycles the devices have undergone.

• RSUD’s are to be labelled to not be re-used and to be returned to the reprocessor.

Labelling Requirements

18

Page 19: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• The labelling / IFU for RSUD’s should include details for the initial collection, cleaning, and storage of used devices prior to their return to the reprocessor.

• Unless deviations are specifically addressed by the reprocessor’s objective evidence, the labelling of RSUD’s should include the same indications and contraindications, and instructions for use as the OEM device.

Labelling Requirements

19

Page 20: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Collection and Tracking

• Cleaning

• Function Testing

• Packaging

• Sterilisation

Manufacturing (Reprocessing) Process

Page 21: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Applications for licensing should include descriptions of the collection system and device tracking system that are used. The descriptions should relate to the safety of the device and validations performed.

• Tracking system– Each device is marked with a unique identifier (number) – The number of reprocessing cycles the device has undergone

is tracked with each device’s identifier– Tracking will facilitate product identification in case of a recall– Tracking will facilitate adverse event reporting

Device Tracking

Page 22: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessed SUD’s shall be traceable to the OEM device’s control number.

• A recall to the OEM device should be evaluated for applicability to the reprocessed version.

• Reprocessors are required to have a process to ensure that RSUD’s are of a known and valid version / design / configuration / revision.

• Changes to the OEM device may require revalidations.

Device Tracking

22

Page 23: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessors are required to perform a thorough validation of cleaning activities / processes.

• ISO 14937:2009 can be used as a model for the validation of the cleaning process, including:– Cleaning agent characterisation– Process and equipment characterisation– Product definition– Process validation, including IQ, OQ, PQ– Routine monitoring and control– Product release and acceptance– Assessment of changes

Process Validation - Cleaning

23

Page 24: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Cleaning validation should be performed on worst-case scenarios using devices contaminated with artificial test soils.

• A measured endpoint for the cleaning process is required and should be in accordance with the risk analysis; visually clean is not an acceptable endpoint, nor is a demonstrated reduction in contamination.

• The validation should address residual contamination by cleaning agents.

Cleaning Validation

24

Page 25: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• The cleaning validation should extend up to the maximum number of re-use cycles that the device may be reprocessed.

• The cleaning validation can include any refurbishing steps or activities, if applicable.

Cleaning Validation

25

Page 26: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessors are required to validate packaging and shelf life, providing the following information in licence applications:– Packaging description– Packaging and device shelf life– Sterile barrier testing– Device function testing

• Validation testing should be performed after environmental conditioning, aging, and transportation simulation.

Packaging and Shelf Life Validations

26

Page 27: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Sterilisation processes shall be validated in accordance with applicable standards. This validation should account for re-sterilisation up to the maximum number of re-use cycles.

• Particular attention should be paid to potentiation to residuals due to repeated sterilisation cycles.

• Pyrogen testing shall be performed on blood contact / surgically invasive devices, as applicable.

Sterilisation Validation

27

Page 28: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessors shall have objective evidence of device validation (ref. MDR 9(2), 10-20) post-processing following a maximum number of re-use cycles.

• This should be in accordance with documented device and performance specifications and the reprocessor’s risk analysis.

• Assessed after maximum number of reprocessing cycles and cycles should be fully simulated (clinical use, cleaning, sterilisation, etc)

Device Validation

28

Page 29: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Post-processing device validation can take many forms, but should address all relevant aspects, including:

– Functional testing against specifications (bench testing)– Pre-clinical testing (animal studies)– Biocompatibility testing– Materials Characterisation– etc.

Device Validation

29

Page 30: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Other information considered during scientific review:– Number of intended reprocessing cycles – all validations

at maximum– Monitoring and tracking OEM device changes– Revalidating process when necessary– Determining when a licence amendments is required– Adverse event reporting– Human factors – Prions –control over any infectious agent that would not

be removed by cleaning and sterilisation methods (i.e. TSEs)

Additional Considerations for RSUDs

Page 31: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Reprocessed devices need to meet the requirements of the Regulations and obtain a medical device licence.

• Emphasis will be placed on the following during review of licence applications:– Cleaning validations– Device validations following all reprocessing and manufacturing

steps (simulated clinical use, cleaning, function testing, packaging, sterilisation, shipping, shelf life, etc)

– Device tracking– Labelling

Summary

Page 32: MEDEC Regulatory Conference May 9, 2016 · 2018-04-02 · MEDEC Regulatory Conference May 9, 2016 ... Canada of any changes to the information in their licence . Regulatory Approach

• Notice to Stakeholders - Health Canada's Regulatory Approach to Commercial Reprocessing of Medical Devices Originally Labelled for Single Use – March 12, 2015

• Medical Device Regulations (SOR/98-282)• Guidance for Industry and FDA Staff – Medical Device User Fee and

Modernization Act of 2002, Validation Data in Premarket Notification Submissions (510(k)s) for Reprocessed Single-Use Medical Devices –September 25, 2006

• AAMI TIR30:2011 – A Compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices

• ISO 14937:2009 - Sterilization of health care products -- General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices

References