Medical Market Laboratory EMEA
Penetration tests in according to the ISO 11140
Josephus P.C.M van Doornmalen Gomez Hoyos
Croatia, March 27, 2009
2
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Outline
� Purpose penetration tests
� Original Bowie and Dick test
� Alternatives for original Bowie and Dick test
� Norms
� Conclusions
3
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Sterilization conditions
� Sterilization conditions:
� 100 % saturated steam
� At the aimed temperature, e.g., 134 °C
� For the right amount of time, e.g., 134 °C for 3 minutes
� On all surfaces
� Textile fibers
� Inside of tubes
4
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
• The task of the BDT is to check pre-vacuum sterilisers
independently from their own gauges etc. for:
• Air removal
• Steam penetration
• Routine (Daily) test
for (steam) sterilizer
AirAir
SteamSteam
??
The task of the BDT
5
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
The History of the Bowie and Dick Test
• First mentioned in 1961:
"Recent Developments in the Sterilization of Surgical
Materials";
Report of a symposium organised by the Pharmaceutical
Society of Great Britain
• Description of the Test in 1963:
"THE BOWIE AND DICK
AUTOCLAVE TAPE TEST"
THE LANCET, March 16, 1963
6
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
The original Bowie and Dick Test Pack
• 29 "Huckaback-Towels"
• Each 36 x 24 inch in size (91 x 61 cm)
• Washed prior to the test
• Unfolded and aired prior to a test for at least one hour
• Folded 3 times to give 8 layers
• Stack 11 inch high (28 cm)
7
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
The original Bowie and Dick Test Pack
• 2 lengths of 12 inch (30 cm)
autoclave tape, "type 1222"(Minnesota Mining and Manufacturing Company)
• In the form of a
St. Andrew’s cross on
a sheet of unglazed paper
• In the middle of the stack
8
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
THE BOWIE AND DICK AUTOCLAVE TAPE TEST, The Lancet, (1963), S.586
The original "Huckaback Towel Pack" as set up byDr. J. H. Bowie and Mr. J. Dick
9
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
250 ml
residual air
Reasons for BDT failure
• Bad vacuum
• Leak
• Non-condensable
gases
10
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Time
Pressure
Phase 1:
Conditioning
Phase 2:
Sterilization
Phase 3:
Drying and
pressure equilibration
Process: Vacuum assisted sterilisers
11
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
CEN Towel Pack
Bowie and Dick: CEN Towel Pack
• EN 285 (26.1)
• Plain, bleached cotton
sheets
• Size: 90 x 120 cm
• Threads: warp 30 +/- 6
weft 27 +/- 5
• Stack size: 22 x 30 x 25 cm
• Wrapped in similar fabric
• Weight: 7 kg +/- 10%
12
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
EN 285:2006+A1:2008 (E)
� Bowie and Dick package
and test defined
� Dimensions
� Material
� Way of folding
� Conditioning
13
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Bowie and Dick package simplified
30 cm
22 cm
25 cm
14
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Steam sterilization process
Time
Pressure
Start of sterilization phase
Steam sterilization condition
Suction
Steam injection
15
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Air (gas mix) removal
Time
Pressure
16
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Steam injection
Time
Pressure
17
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Steam penetration and condensation
Time
Condensation front
Gases sucked out, textile stays wet
Good penetration
18
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Insufficient deep vacuum
� Condense front will not disappear
19
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
None condensable gases in steam
� Water vapour condenses on surfaces colder than the
vapour
� Gases that will not condense in similar conditions are called
None Condensable Gases
� Oxigen (O2), Nitrogen (N2), Carbon dioxide (CO2)
� Possibilties of introduced in the steam:
� Feed water
� Too smal dimension of steam boiler
� Leak in pipe work, vessel
20
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Effect NCG’s
O2, N2, CO2, H2O
Steam Injection
(Steam with NCG’s)
Condensation of water
Steam injection, pressure increasing
21
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
steam
steam
steam
steam
NCG’s do not change their volume
during condensation.
Non-condensable gases
22
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
The temperature depression
Drain
Temperature:
135°C
Temperature:
130°C
Example
Temperature
depression
5 K
23
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
The Air Pocket
at 4,5 cm height
The air pocket’s form is not a perfect ball,
but egg shaped.
At 10 cm height
Animation
24
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
When to perform a penetration test
� EN 554
� 6.3.4 Process control and monitoring: "If the sterilization process includes air removal from the product, a steam penetration test shall be carried out at the commencement of each day the sterilizer is to be used."
� A.4.3 Routine monitoring and testing:"e) a steam penetration test (when specified).Note. A suitable penetration test for a saturated steam process is being prepared by CEN/TC 102.
� CEN/TC102 WG7 prepared EN 867-4
25
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
EN ISO 17665
� 12.1.6 Process control and monitoring: "If the sterilization process relies on the removal of air from the sterilizer chamber in order to achieve rapid and even penetration of steam into the sterilizer load, a steam penetration test shall be carried out each day before the sterilizer is used."
� A.6.2 Routine monitoring and testing:"An alternative test pack may be used providing the challenge to the sterilization process has been demonstrated to be equal to or greater than the challenge from the standard test pack described in A.3.”
� Statements to the same effect can be found in EN 554 which will be replaced by EN ISO 17665.
26
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
31300 Disposable Test
Pack
Alternative Products
• Alternative Bowie & Dick
Test products
• Increased reproducibility
• Constant test conditions
• Performance requirements
are covered by a
dedicated standard
(EN ISO 11140 - 4)
27
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
What the norms say
� Towel pack according to EN 285 with:
an Indicator test sheet in compliance with EN ISO 11140-3
or
alternative indicators according to EN ISO 11140-4
28
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Examples hollow instruments
Lumen
29
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Helix Examples
Lumen
Helix schematically
Tube Receptacle
30
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Time
Pressure
Length = L
Length = ½L Closed end
Penetration in lumen
31
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Relatively easy to penetrate
Relatively difficult to penetrate
Configuration of instruments
32
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Lumen as representative for lumen
� Material of lumen
� Multi layers of different materials
� Heat transfer
� Heat capacity
� Dimension (length tube, radius tube,
� Dimension of the tube:
� Length of the tube
� Radius of the tube
� Thickness of the tube
Lumen
33
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Wall thickness
Wall thickness
More heat capacity
(e.g., more condensation)Research ongoing
Heat transfer
34
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Dimensions of tube
Radius
rLenghth
Research ongoing!
35
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Swirl
� Well known and described phenomena
Axial dispersion and flow phenomena in helically ciled tubular reactors for flow analysis and
chromatography, R. Tijssen, Analytica Chimica Acta, 114, 1980, 71-89:
Swirl
36
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
EN ISO 11140-4 and EN 867-5
� The helix or hollow A is postulated in the EN867-5 and is to
be used in small sterilizers (type B and N processes
according to EN13060)
� Steam penetration test for large sterilizers (EN285) should
be a Bowie & Dick test or an alternative, as described in the
EN ISO 11140-4
� So far, no scientific proof has been found in literature (1).
(1) Publications in a scientific journal with an impact factor
37
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Proving for alternative tests
� Show that the alternative test is reacting as the original
Bowie and Dick test.
� Experimental data Bowie and Dick test
� Experimental data for alternative
� Comparison between data Bowie and Dick test and alternative
test
11111212 1313 1414
1515 16161717 1818 1919
23232525 2626
2727 2929 31313333
35353939 4040 4141
4242 434345454444 4646
4747
20202424
2121
28283434
2222
30303232
3636 37373838
38
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Hollow A and helix
� Is the hollow A of Helix a good representation of your
instruments with narrow lumen?
� Is a postulated test
� Is for small sterilizers type B and S
39
© 3M 2008. All Rights Reserved.
Medical Markets Laboratory (Europe)
Current situation
� ISO working group is started to write ISO 11140 part 6
� This standards is meant to address Process Challenge
devices
� Research concerning penetration in lumen is on going