6 Membrane Selection Guide Reference Chart of Membrane Filter Nylon CA(Cellulose acetate) Hydrophilic property Ideal for aqueous(Non-acidic) or organic sample High protein binding, Not used when protein recovery is important, Commonly used to for general lab filtration of HPLC Naturally Hydrophilic Low protein binding suitable for aqueous samples Nitrate free suitable for ground water filtration Excellent choice for maximum protein recovery, bind less protein than PVDF Ideal for tissue culture media and sensitive biological sample PTFE(Polytetrafluoroethylene) PES(Polyethersulfone) Hydrophobic PTFE Strong hydrophobic Strong chemical stability and inertia Strong acid and alkali solvent filtration Chemical resistance to all solvent, Ideal for filtering gas Hydrophilic PTFE Inorganic solvent with strong chemical causticity filtration Mixed solvent filtration in HPLC(Acetonitrile/Water) Alkali solvent filtration Lowest protein binding Low binding of protein and nucleic acid, faster than PVDF, excellent flow rates Suitable filtration for tissue culture media Broad chemical resistance , aggressive solvents PVDF(Polyvinylidene difluoride) MCE(Mixed cellulose ester) Good heat endurance and chemical stability Strong hydrophobic, Non Aggressive aqueous, mild organic solutions, Solvent mixture Extremely low protein binding HPLC sample, general biological filtration Filtration of aqueous sample, effectively bind trace protein Little resistance Oil, Alcohol, and bacteria filtration and then inspect, Particle Analysis Microbiology PP(Polypropylene) GMF Naturally hydrophilic membrane, Wide range of chemical compatibility to organic solvents Low no-specific absorbing membrane for maximum protein recovery in critical analysis Suited for biological sample filtration Designed specially to filter high particulate solutions. With four layers, the first filter is a composite membrane of 10μm glass fiber and 1.0μm PP, the second prefilter is 0.7μm GF/F membrane media, the last one is filtration media as speci- fied. Step-1 Determine whether required prefiltration Syringe Filter with PP Prefilter Step-2 Choose the suitable Membrane Filtration Medium. by Characteristics of Samples Aqueous Non-Polar Protein MCE/CA/NY PTFE/PVDF/NY PES/PVDF Step-3 Choose the suitable diameter. by Volume of Samples <10ml <100ml <250ml 13mm 25mm 30mm Step-4 Choose the suitable Pore Size. by size of Target compounds >3um <3um 0.45um 0.22um To Place an Order or Receive Technical Assistance +98-21-22024327-8 Ext 612 Arash Abed Mobile: 09128911277 Email: [email protected]

Membrane selection guide

Embed Size (px)

Citation preview

Page 1: Membrane selection guide

6

Membrane Selection Guide

Reference Chart of Membrane Filter

Nylon CA(Cellulose acetate)• Hydrophilic property• Ideal for aqueous(Non-acidic) or organic sample• High protein binding, Not used when protein recovery is

important, Commonly used to for general lab filtration of HPLC

• Naturally Hydrophilic• Low protein binding suitable for aqueous samples• Nitrate free suitable for ground water filtration• Excellent choice for maximum protein recovery, bind less

protein than PVDF• Ideal for tissue culture media and sensitive biological

sample

PTFE(Polytetrafluoroethylene) PES(Polyethersulfone)Hydrophobic PTFE• Strong hydrophobic• Strong chemical stability and inertia• Strong acid and alkali solvent filtration• Chemical resistance to all solvent, Ideal for filtering gas

Hydrophilic PTFE• Inorganic solvent with strong chemical causticity filtration• Mixed solvent filtration in HPLC(Acetonitrile/Water)• Alkali solvent filtration

• Lowest protein binding• Low binding of protein and nucleic acid, faster than

PVDF, excellent flow rates• Suitable filtration for tissue culture media• Broad chemical resistance , aggressive solvents

PVDF(Polyvinylidene difluoride) MCE(Mixed cellulose ester)• Good heat endurance and chemical stability• Strong hydrophobic, Non Aggressive aqueous, mild

organic solutions, Solvent mixture• Extremely low protein binding• HPLC sample, general biological filtration

• Filtration of aqueous sample, effectively bind trace protein

• Little resistance• Oil, Alcohol, and bacteria filtration and then inspect,

Particle Analysis Microbiology

PP(Polypropylene) GMF• Naturally hydrophilic membrane, Wide range of chemical

compatibility to organic solvents• Low no-specific absorbing membrane for maximum

protein recovery in critical analysis• Suited for biological sample filtration

Designed specially to filter high particulate solutions. With four layers, the first filter is a composite membrane of 10μm glass fiber and 1.0μm PP, the second prefilter is 0.7μm GF/F membrane media, the last one is filtration media as speci-fied.

Step-1 Determine whether required prefiltration Syringe Filter with PP PrefilterStep-2 Choose the suitable Membrane Filtration Medium. by Characteristics of Samples

Aqueous Non-Polar ProteinMCE/CA/NY PTFE/PVDF/NY PES/PVDF

Step-3 Choose the suitable diameter. by Volume of Samples<10ml <100ml <250ml13mm 25mm 30mm

Step-4 Choose the suitable Pore Size. by size of Target compounds>3um <3um

0.45um 0.22um

To Place an Order or ReceiveTechnical Assistance+98-21-22024327-8 Ext 612Arash AbedMobile: 09128911277Email: [email protected]

Page 2: Membrane selection guide

FUTECS your chromatography specialist

7www.futecs.com

Chemical Compatibility Chart for Syringe Filters

NY PTFE PVDF PP REG PES CAACIDSAcetic, Glacial L C C C C N NAcetic, 25% C C C C C C CFormic, 25% N C C C C - LHydrochloric, 25% N C C C N C NNitric, 25% N C C C N - NPhosphoric, 25% N C C C L - CSulfuric, 25% N C C C N - NTrichloroacetic, 10% N C - - C - C

ALCOHOLSAmyl Alcohol, Butanol C C L C C L CBenzyl Alcohol C C C C C N LEthanol, 70% L C C C C C CEthanol, 98% C C C C C C CEthylene Glycol C C C C C C CGlycerol C C C C C C CIsopropal, n-Propanol C C C C C C CMethanol, 98% C C C C C C CPropylene Glycol C C C C C C L

ALKALIESAmmonium Hydroxide, 25% C C C C L C CSodium Hydroxide 3 Normal C C C C L - N

HYDROCARBONSHexane C C C N C L CSodium C C C N C C CToluene, Benzene C C C N C S C

HALOGENATED HYDROCABONSCarbon Tetrachloride C C C L C C LChloroform C C C L C N NMethylene Chloride L C C L C N NMonochlorobenzene, Freon C C C C C - CTrichloroethylene C C C C C N C

KEYTONESAcetone, Cyclohexanone C C C C C N NMethy Ethyl Ketone C C L L C N LMethyl Isobutyl Ketone - C N L C - -

ESTERSAmyl Acetate C C L L C - LButyl Acetate C C L L C - LEthyl acetate, Methyl acetate C C L L C N NPropyl Acetate C C L L C - L

OXIDES - ETHERSDimethylsulfoxide (DMSO) - E N C C N NDioxane and Tetrahydrofuran C E N C C - NEthyl Ether C E C L C C C

SOLVENTS WITH NITROGEN (AMIDES)Acetonitrile C E L C C L NAniline - - N - C - NDiethylacetamide C - N - C - NDimethyl Formamide L N N C L - NPyridine C L S N C L N

MISCELLANEOUSFormaldehyde Solution, 30% C C C C L C LHydrogen Peroxide, 30% N C C C C C CPhenol, Aqueous, 10% N C L C N N N

*N =Not CompatibleC =CompatibleL =Limited Compatibility. May be okay for small volumes and/or short term contact.

Page 3: Membrane selection guide

8

Sample Syringe Filter

Syringe filters are simply quality filters, well packaged, and offered at a fair and competitive price. The Classic range is available in all of the major membranes including Nylon, CA, PP, Glass Fiber, PTFE, PES, MCE and PVDF which are supplied in 13mm, 25mm and 17mm, 30mm formats in virgin plypropylene housings

Page 4: Membrane selection guide

9

FUTECS your chromatography specialist

www.futecs.com

I. Sample Preparation

Superpure -Syringe Filter

• Larger filtration area(Diameter) : 17mm, 30mm• Increase sample volume throughput• High reducing pressure• Designed to speed up membrane flow rate• Nylon, Hydrophobic-PTFE, Hydrophiliic-PTFE, PVDF, PES, MCE• Pore Size; 0.22um, 0.45um• Color Housing Designed

Chrompure -Syringe Filter

• High quality membrane• Gear rim ring for better handling• Improved membrane flow rate• Membrane Materials; Nylon, PTFE, PVDF, CA, PES, MCE, GF• Pore Size; 0.22um, 0.45um• Size(Diameter); 13mm, 25mm• Color Housing Design

Simplepure -Syringe Filter

• Specially designed to maximize sample recovery• Minimum sample hold-up• High quality membrane• Improved membrane flow rate• Materials; Nylon, Hydrophobic-PTFE, Hydrophilic-PTFE,

PVDF, CA, PES, MCE• Pore Size; 0.22um, 0.45um, Available 0.8 - 5um• Size(Diameter); 13mm, 25mm• Housing; Natural white, Virgin medical polypropylene housing

Page 5: Membrane selection guide

10

Non-Sterile Syringe Filter

I. Sample Preparation

MS LLC

Nylon Syringe Filter (Polyamide)

• Hydrophilic property• No need to moist beforehand• Uniform aperture• Strong tenacity and adsorbability• Compatible with aqueous and alcoholic solutions and solvents; suitable for HPLC

Parameter 13mm 25mm 30mmMaterial Nylon Nylon NylonHousing material PP PP PPFiltration area(cm2) 0.65 3.90 4.60Pore size(um) 0.22, 0.45 0.22, 0.45 0.22, 0.45Holdup volume(ul) <10 <30 <55Sample volume(ml) <12 <100 <140Max. Operation Temp 100 100 100Max. pressure(psi) 75 95 110pH 3-12 3-12 3-12

Ordering informations

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFNY013022NC Nylon 0.22 13 No 100SFNY013045NC Nylon 0.45 13 No 100SFNY013022N Nylon 0.22 13 No 100SFNY013045N Nylon 0.45 13 No 100SFNY025022NC Nylon 0.22 25 No 100SFNY025045NC Nylon 0.45 25 No 100SFNY025022N Nylon 0.22 25 No 100SFNY025045N Nylon 0.45 25 No 100SFNY017022N Nylon 0.22 17 No 100SFNY017045N Nylon 0.45 17 No 100SFNY030022N Nylon 0.22 30 No 100SFNY030045N Nylon 0.45 30 No 100

Page 6: Membrane selection guide

11

FUTECS your chromatography specialist

www.futecs.com

I. Sample Preparation

PVDF Syringe Filter(Polyvinylidene Fluoride)

• Good heat-endurance chemical stability• Strong hydrophobiclity filters• Widely used in gas filtration, vapor filtration, high-temperature• Food industry• Medicine filtration

Parameter 13mm 25mm 30mmMaterial PVDF PVDF PVDFHousing material PP PP PPFiltration area(cm2) 0.65 3.90 4.60Pore size(um) 0.22, 0.45 0.22, 0.45 0.22, 0.45Holdup volume(ul) <10 <30 <55Sample volume(ml) <12 <100 <140Max. Operation Temp 100 100 100Max. pressure(psi) 50 95 110pH 1-14 1-14 1-14

Ordering informations

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFPVDF013022NC PVDF 0.22 13 No 100SFPVDF013045NC PVDF 0.45 13 No 100SFPVDF013022N PVDF 0.22 13 No 100SFPVDF013045N PVDF 0.45 13 No 100SFPVDF025022NC PVDF 0.22 25 No 100SFPVDF025045NC PVDF 0.45 25 No 100SFPVDF025022N PVDF 0.22 25 No 100SFPVDF025045N PVDF 0.45 25 No 100SFPVDF017022N PVDF 0.22 17 No 100SFPVDF017045N PVDF 0.45 17 No 100SFPVDF030022N PVDF 0.22 30 No 100SFPVDF030045N PVDF 0.45 30 No 100

Page 7: Membrane selection guide

12

Non-Sterile Syringe Filter

I. Sample Preparation

MS LLC

Hydrophobic PTFE Syringe Filter(Polyfluortetraethylene)

• Broad chemical compatibility• Strong chemical stability and inertia• Strong hydrophobicity• Organic solvent with strong chemical causticity filtration• Strong acid solvent filtration• Alkali solvent filtration

Parameter 13mm 25mm 30mmMaterial PTFE PTFE PTFEHousing material PP PP PPFiltration area(cm2) 0.65 3.90 4.60Pore size(um) 0.22, 0.45 0.22, 0.45 0.22, 0.45Holdup volume(ul) <10 <30 <55Sample volume(ml) <12 <100 <140Max. Operation Temp 130 130 130Max. pressure(psi) 130 130 130pH 1-14 1-14 1-14

Ordering informations

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

Hydrophobic PTFE SFPTFE013022NBC PTFE 0.22 13 No 100SFPTFE013045NBC PTFE 0.45 13 No 100SFPTFE013022NB PTFE 0.22 13 No 100SFPTFE013045NB PTFE 0.45 13 No 100SFPTFE025022NBC PTFE 0.22 25 No 100SFPTFE025045NBC PTFE 0.45 25 No 100SFPTFE025022NB PTFE 0.22 25 No 100SFPTFE025045NB PTFE 0.45 25 No 100SFPTFE017022NB PTFE 0.22 17 No 100SFPTFE017045NB PTFE 0.45 17 No 100SFPTFE030022NB PTFE 0.22 30 No 100SFPTFE030045NB PTFE 0.45 30 No 100

Page 8: Membrane selection guide

13

FUTECS your chromatography specialist

www.futecs.com

I. Sample Preparation

Hydrophilic PTFE Syringe Filter

Company Product DescriptionMS Hydrophilic PTFE

syringe filterA versatile filter for aqueous and aggres-sive organic solvent-based solutions and especially ideal for HPLC operation.

Pall GHP syringe filter Hydrophilic PP membrane media. Highly recommended for filtering HPLC samples and mobile phases.

Millipore Millex-GV Hydrophilic PVDF membrane media. Preparation of protein-containing solutions prior to chromatographyMillex-HV

Ordering informations

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

Hydrophilic PTFE SFPTFE013022NLC PTFE 0.22 13 No 100SFPTFE013045NLC PTFE 0.45 13 No 100SFPTFE013022NL PTFE 0.22 13 No 100SFPTFE013045NL PTFE 0.45 13 No 100SFPTFE025022NLC PTFE 0.22 25 No 100SFPTFE025045NLC PTFE 0.45 25 No 100SFPTFE025022NL PTFE 0.22 25 No 100SFPTFE025045NL PTFE 0.45 25 No 100SFPTFE017022NL PTFE 0.22 17 No 100SFPTFE017045NL PTFE 0.45 17 No 100SFPTFE030022NL PTFE 0.22 30 No 100SFPTFE030045NL PTFE 0.45 30 No 100

Syringe filters are using special hydrophilic PTFE membrane as the filter medium. This material has a maximal chemical and pH resistance, and optically clear when wet with water. Hydrophilic PTFE syringe filter is a versatile filter for aqueous and aggressive organic solvent-based solutions, and especially ideal for HPLC operation. Pure polypropylene housings and color-coding housings include smooth rim or gear rim are all available

Features• Low protein binding fine particle removal from aqueous and organic solutions.• Accurate analysis HPLC certified for low levels of UV-absorbing extractables.• High flow rates with minimal aqueous extractables.• Low surface friction coefficient: 0.1.• Wide range of working temperature:180-260℃

Application

Cross Reference

• Highly recommended for filtering HPLC samples and mobile phases.• Organic solvent with strong chemical causticity filtration

Page 9: Membrane selection guide

14

Non-Sterile Syringe Filter

I. Sample Preparation

MS LLC

PES Syringe Filter(Polyether Sulfone)

• Hydrophilic, low protein binding,• Low extractable• High throughput• Tissue culture media filtration• Tissue culture additive filtration• Low affinity for proteins and extractable with substantially faster flow rates than PVDF

Parameter 13mm 25mm 30mmMaterial PES PES PESHousing material PP PP PPFiltration area(cm2) 0.65 3.90 4.60Pore size(um) 0.22, 0.45 0.22, 0.45 0.22, 0.45Holdup volume(ul) <10 <30 <55Sample volume(ml) <12 <100 <140Max. Operation Temp 90 90 90Max. pressure(psi) 50 95 120pH 1-14 1-14 1-14

Ordering informations

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFPES013022NC PES 0.22 13 No 100SFPES013045NC PES 0.45 13 No 100SFPES013022N PES 0.22 13 No 100SFPES013045N PES 0.45 13 No 100SFPES025022NC PES 0.22 25 No 100SFPES025045NC PES 0.45 25 No 100SFPES025022N PES 0.22 25 No 100SFPES025045N PES 0.45 25 No 100SFPES017022N PES 0.22 17 No 100SFPES017045N PES 0.45 17 No 100SFPES030022N PES 0.22 30 No 100SFPES030045N PES 0.45 30 No 100

Page 10: Membrane selection guide

15

FUTECS your chromatography specialist

www.futecs.com

I. Sample Preparation

MCE Syringe Filter (Mixed Cellulose Ester)

• Uniform aperture,• No medium dropping• Naturally hydrophilic• Filtration of aqueous solutions; effectively binds trace proteins.

Parameter 13mm 25mm 30mmMaterial MCE MCE MCEHousing material PP PP PPFiltration area(cm2) 0.65 3.90 4.60Pore size(um) 0.22, 0.45 0.22, 0.45 0.22, 0.45Holdup volume(ul) <10 <30 <55Sample volume(ml) <12 <100 <140Max. Operation Temp 110 110 110Max. pressure(psi) 120 120 120pH 4-8 4-8 4-8

Ordering informations

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFMCE013022NC MCE 0.22 13 No 100SFMCE013045NC MCE 0.45 13 No 100SFMCE013022N MCE 0.22 13 No 100SFMCE013045N MCE 0.45 13 No 100SFMCE025022NC MCE 0.22 25 No 100SFMCE025045NC MCE 0.45 25 No 100SFMCE025022N MCE 0.22 25 No 100SFMCE025045N MCE 0.45 25 No 100SFMCE017022N MCE 0.22 17 No 100SFMCE017045N MCE 0.45 17 No 100SFMCE030022N MCE 0.22 30 No 100SFMCE030045N MCE 0.45 30 No 100

Page 11: Membrane selection guide

16

Non-Sterile Syringe Filter

I. Sample Preparation

MS LLC

CA Syringe Filter (Cellulose Acetate)

• Ideal for aqueous, biological or protein filtration• Naturally hydrophilic• Nitrate free: suitable for groundwater filtration• Cell retention and particle collection• Filtration for aqueous sample and some organic solvents

Parameter 13mm 25mm 30mmMaterial CA CA CAHousing material PP PP PPFiltration area(cm2) 0.65 3.90 4.60Pore size(um) 0.22, 0.45 0.22, 0.45 0.22, 0.45Holdup volume(ul) <10 <30 <55Sample volume(ml) <12 <100 <140Max. Operation Temp 50 50 50Max. pressure(psi) 120 120 120pH 3-7 3-7 3-7

Ordering informations

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFCA013022NC CA 0.22 13 No 100SFCA013045NC CA 0.45 13 No 100SFCA013022N CA 0.22 13 No 100SFCA013045N CA 0.45 13 No 100SFCA025022NC CA 0.22 25 No 100SFCA025045NC CA 0.45 25 No 100SFCA025022N CA 0.22 25 No 100SFCA025045N CA 0.45 25 No 100SFCA017022N CA 0.22 17 No 100SFCA017045N CA 0.45 17 No 100SFCA030022N CA 0.22 30 No 100SFCA030045N CA 0.45 30 No 100

Page 12: Membrane selection guide

17

FUTECS your chromatography specialist

www.futecs.com

I. Sample Preparation

PP Syring Filter (Polypropylene)

Glass Fiber Syringe Filter

• Naturally hydrophilic membrane• Wide range of chemical compatibility to organic solvents• Highly solvent resistant• Low no-specific absorbing membrane for maximum protein recovery in critical analysis

• Hydrophilic Material Membrane• Excellent compatibility with organic solvents and strong acids

(apart from hydrofluoric acid) and bases• High dirt-handling capacity.

Ordering informations

Ordering informations

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFPP013022N PP 0.22 13 No 100SFPP013045N PP 0.45 13 No 100SFPP025022N PP 0.22 25 No 100SFPP025045N PP 0.45 25 No 100

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFGF013070N Glass Fiber 0.7 13 No 100SFGF013100N Glass Fiber 0.1 13 No 100SFGF025070N Glass Fiber 0.7 25 No 100SFGF025100N Glass Fiber 0.1 25 No 100

Page 13: Membrane selection guide

18

Non-Sterile Syringe Filter

I. Sample Preparation

MS LLC

Syringe Filter with Prefilter

• Improve sample volume throughout with prefilter• High particulate load• Exceptionally low extractable level with no wetting agents utilized

Nylon with PP Prefilter

PVDF with PP Prefilter

PTFE with PP Prefilter

PES with PP Prefilter

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFNY013022NP NY/PP 0.22 13 No 100SFNY013045NP NY/PP 0.45 13 No 100SFNY025022NP NY/PP 0.22 25 No 100SFNY025045NP NY/PP 0.45 25 No 100

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFPVDF013022NP PVDF/PP 0.22 13 No 100SFPVDF013045NP PVDF/PP 0.45 13 No 100SFPVDF025022NP PVDF/PP 0.22 25 No 100SFPVDF025045NP PVDF/PP 0.45 25 No 100

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFPTFE013022NP PTFE/PP 0.22 13 No 100SFPTFE013045NP PTFE/PP 0.45 13 No 100SFPTFE025022NP PTFE/PP 0.22 25 No 100SFPTFE025045NP PTFE/PP 0.45 25 No 100

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFPES013022NP PES/PP 0.22 13 No 100SFPES013045NP PES/PP 0.45 13 No 100SFPES025022NP PES/PP 0.22 25 No 100SFPES025045NP PES/PP 0.45 25 No 100

Syringe filters with prefilter are purpose-built with features designed to bring the highest levels of performance and purity to your research. We incorporate a variety of membranes to offer separation and purification solutions for the majority of your laboratory needs. Syringe filter with prefilter prevents blockage and bursting of the actual membrane for the fine filtration.

Page 14: Membrane selection guide

19

FUTECS your chromatography specialist

www.futecs.com

I. Sample Preparation

Nylon with Glass Fiber Prefilter

PVDF with Glass Fiber Prefilter

PTFE with Glass Fiber Prefilter

PES Glass Fiber Prefilter

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFNY013022NG NY/GF 0.22 13 No 100SFNY013045NG NY/GF 0.45 13 No 100SFNY025022NG NY/GF 0.22 25 No 100SFNY025045NG NY/GF 0.45 25 No 100

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFPVDF013022NG PVDF/GF 0.22 13 No 100SFPVDF013045NG PVDF/GF 0.45 13 No 100SFPVDF025022NG PVDF/GF 0.22 25 No 100SFPVDF025045NG PVDF/GF 0.45 25 No 100

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFPTFE013022NG PTFE/GF 0.22 13 No 100SFPTFE013045NG PTFE/GF 0.45 13 No 100SFPTFE025022NG PTFE/GF 0.22 25 No 100SFPTFE025045NG PTFE/GF 0.45 25 No 100

Cat. No. Filter Medium

Pore Size(μm)

Diameter(mm)

Gama Sterile Qty/pack

SFPES013022NG PES/GF 0.22 13 No 100SFPES013045NG PES/GF 0.45 13 No 100SFPES025022NG PES/GF 0.22 25 No 100SFPES025045NG PES/GF 0.45 25 No 100

To Place an Order or ReceiveTechnical Assistance+98-21-22024327-8 Ext 612Arash AbedMobile: 09128911277Email: [email protected]