Method Validation 2370

Embed Size (px)

Citation preview

  • 8/8/2019 Method Validation 2370

    1/37

    1

    Method validation and

    verification

    W. W. Wong

    Senior Accreditation OfficerHKAS

    20 November 2009

  • 8/8/2019 Method Validation 2370

    2/37

    2

    Method Validation

    What is it?

    When is it required?

    Why is it necessary?

    What are required? How much is adequate?

    How should it be done?

  • 8/8/2019 Method Validation 2370

    3/37

    3

    Method Validation what are required

    Validation of methods (cl. 5.4.5.3 )

    The range and accuracy of the valuesobtainable from validated methods (e.g.

    Uncertainty of results

    Detection limit Selectivity Linearity Repeatability Reproducibility Robustness/cross-sensitivity

  • 8/8/2019 Method Validation 2370

    4/37

    4

    HOKLAS SC No. 37 Food TestCategory Chemical Testing

    Being commented by AAB WP on Food(today is the last day for comment)

    Will be published very soon Contains criteria specific to food testing With a view to assist labs

    In particular, method validation and samplepreparation

  • 8/8/2019 Method Validation 2370

    5/37

    5

    HOKLAS SC No. 37 Section5.3 Validation of methods

    Screening methods:

  • 8/8/2019 Method Validation 2370

    6/37

    6

    Confirmation of identity

    Mass spectrometric detection

    +50%+50%10% to 20%+25%+15%>20% to 50%

    +20%+10%>50%

    CI-GC-MS, GC-

    MSn

    , LC-MS,LC-MSn

    EI-GC-MS

    (relative)

    Relative intensity

    (% of base peak)

  • 8/8/2019 Method Validation 2370

    7/37

    7

    Limit of DetectionxLD = 3sO +xBl

    WherexLD is the limit of detection

    sO is the SD of the outlier free result of a matrix blank sample.

    should be based on at least 10 independent completedeterminations of analytes concentration in a typical matrix

    blank or low-level material, with no censoring of zero or

    negative results.

    xBl is the mean concentration of the matrix blank

  • 8/8/2019 Method Validation 2370

    8/37

    8

    Limit of Quantification

    The limit of quantification represents a

    concentration of the determinand that can

    reasonably be determined with an acceptable

    level of accuracy. Usually it is arbitrarily taken

    as a fixed multiple of the detection limit.

  • 8/8/2019 Method Validation 2370

    9/37

    9

    Limit of Quantification

    Estimated by:

    xLQ = 3xLD

    The factor k=3 corresponds to a relative result

    uncertainty of approximately 33%.

  • 8/8/2019 Method Validation 2370

    10/37

    10

    Limit of QuantificationAOAC

    smallest amount of analyte in a test sample that can bequantitatively determined with suitable precision and

    accuracy under previously established method conditions.

    It is often taken as the blank value plus 10 times the std

    dev.

  • 8/8/2019 Method Validation 2370

    11/37

    11

    Reporting limits

    Set at a level at which quantitativeresults may be obtained with a

    specified degree of confidence

    Limits of detection and reportingshall be verified.

  • 8/8/2019 Method Validation 2370

    12/37

    12

    Reporting limits

    ISO/TS 13530:2009

    RL is a specific concentration at orabove the limit of quantification that isreported to the client with a certain

    degree of confidence. It is often definedon a project-specific basis. If the RL isset below the limit of quantification by

    the client, method modification isrequired.

  • 8/8/2019 Method Validation 2370

    13/37

    13

    Reporting limits

    ISO/TS 13530:2009

    For verification of limit of detectionand limit of quantification, spiked

    blank matrix samples at theseconcentration levels and blank

    matrix samples shall be analysed inthe same manner as real samples.

  • 8/8/2019 Method Validation 2370

    14/37

    14

    Reporting limits

    ISO/TS 13530:2009

    If the uncertainty of results for thesamples spiked at the limit of

    quantification level is smaller thanor equal to the relative precision

    corresponding to the factork, thelimit of quantification is verified.

  • 8/8/2019 Method Validation 2370

    15/37

    15

    Reporting limits

    ISO/TS 13530:2009

    x = 1xLQ k

    where

    x = t .s/n

    kis the factor for calculating the limit of

    quantification as a multiple of detection,here: k=3

  • 8/8/2019 Method Validation 2370

    16/37

    16

    Reporting limits

    In essence, the laboratory needs to provide

    supporting data that, at the reporting limit,the measurement uncertainty (including biasand precision) meets the acceptance criteria.

    The direct method is to find the recovery andRSD at the reporting limit and compare themwith the acceptance criteria.

  • 8/8/2019 Method Validation 2370

    17/37

    17

    Recommended LOD and LOQ

    ML= specified maximum and/or

    minimum level

    For ML 0.1 mg/kg, LODML X 1/5

    For ML 0.1 mg/kg, LODML X 2/5

    LOQ

    For ML 0.1 mg/kg, LODML X 1/10

    For ML

    0.1 mg/kg, LOD

    ML X 1/5

    LOD

  • 8/8/2019 Method Validation 2370

    18/37

    18

    Recommended Recovery

    40-1201g/kg

    60-11510g/kg

    80-110100g/kg80-1101mg/kg

    80-11010mg/kg

    90-107100mg/kgRecovery (R)95-105

    0.1% (1mg/g)

    97-103 1% (1g/100g)

    98-102 10% (10g/100g)

    98-102100%(100g/100g)

    Recovery (%)Unit

  • 8/8/2019 Method Validation 2370

    19/37

    19

    Precision

    AOAC Official Methods of Analysis (2005)

    Appendix E: Laboratory Quality Assurance

    Common values of within lab precision

    2010ppb

    101ppm

    50.01%

    21%

    1100%

    RSDr

    Concentration

  • 8/8/2019 Method Validation 2370

    20/37

    20

    Estimation of standard deviation by meanrange

    s =R/d2

    d2is 1.128 for duplicate

    R is the mean range of the duplicates

  • 8/8/2019 Method Validation 2370

    21/37

    21

    Range of food matrices

    If a method is to be accredited undergeneral foodstuffs, satisfactoryvalidation data shall be obtained for atleast five different food matrices(protein, carbohydrate, oil, dietary fibre

    and water) with at least three food typesrepresentative of each food matrix.

  • 8/8/2019 Method Validation 2370

    22/37

    22

    0%

    %FAT

    %PROTEIN

    %CARBOHYDRATES0%

    0%

    100%

    100%

    100%

    J. AOAC, 83 (2), 413 (2000)The Referee, AOAC Intl, July 1993

  • 8/8/2019 Method Validation 2370

    23/37

    23

  • 8/8/2019 Method Validation 2370

    24/37

  • 8/8/2019 Method Validation 2370

    25/37

    25

    Range of food matrices

    The range of matrices shall also in line

    with those listed in relevant regulations.Due consideration shall also be taken

    for the food matrices with potentialinterferences, e.g. high chloride effect

    on the ICP-MS determination.

  • 8/8/2019 Method Validation 2370

    26/37

    26

    Range of food matrices

    General foodstuffs all foods Test procedure shall document the food matrices used in

    the validation studies

    Labs need to assess and determine the applicability oftheir methods to the food samples received.

  • 8/8/2019 Method Validation 2370

    27/37

    27

    Method verification

    Released early 2008

    http://www.aoac.org/alacc_guide_2008.pdf

  • 8/8/2019 Method Validation 2370

    28/37

    28

    How should it be done?

    The following are some examples of useful references

    Harmonised guidelines for single-laboratoryvalidation of method of analysis (IUPAC Technical

    Report), 2002

    How to meet ISO 17025 Requirements for MethodVerification, AOAC, 2007

    ISO/TS 21748:2004 Guidance for the use of

    repeatability, reproducibility and trueness estimatesin measurement uncertainty estimation

  • 8/8/2019 Method Validation 2370

    29/37

    29

    How should it be done?

    CRMs

    How many replicate results are required for propercomparison of the certified reference value and the

    actual analysis result?

    Is it necessary for the mean of test result for a CRM tolie within the uncertainty range of the certified value?

    Is a result outside the uncertainty range acceptable ?

  • 8/8/2019 Method Validation 2370

    30/37

    30

    How should it be done?

    Useful references- ISO Guide 33: 2000

    Uses of certified reference materials

  • 8/8/2019 Method Validation 2370

    31/37

  • 8/8/2019 Method Validation 2370

    32/37

    32

    How should it be done?

    Comparison of results with another laboratory(ISO 5725-6:1994 clause 7.2.4.3)

    - the means of the two laboratories are compared

    if ,

    then statistically no significant difference

    where = repeatability standard deviation

    = reproducibility standard deviation

    222

    22

    21r

    Ryy

    2

    r

    2

    R

  • 8/8/2019 Method Validation 2370

    33/37

    33

    How should it be done - verification

    Worked examples of method verification areavailable from AOAC

    Determination of Total, Saturated, andMonosaturated Fats In Foodstuffs by Hydrolytic Extraction and Gas ChromatographicQuantitation: Collaborative Study

    Determination of Low-Level Glucose andFructose in Raw and Refined Crystalline Sugarby High-Performance Anion ExchangeChromatography

  • 8/8/2019 Method Validation 2370

    34/37

    34

    Common deficiencies

    Number of food matrices used not adequate

    Number of food types for each matrix notadequate

    Concentration levels used not adequate

    Reporting limits estimated but not verified Food samples chosen not representative of the

    food matrices

  • 8/8/2019 Method Validation 2370

    35/37

    35

    Common deficiencies

    Confirmation of identity technique not availableor adequate Validation/verification data analyses not done

    correctly

    Measurement uncertainty not available/notestimated correctly

    Equipment and/or test procedures notconforming to the test std requirements

  • 8/8/2019 Method Validation 2370

    36/37

    36

    Common deficiencies

    Lack of or inappropriate sampling procedure QC plan acceptance limits not appropriate

  • 8/8/2019 Method Validation 2370

    37/37

    37

    Thank you