Upload
kentsang822
View
238
Download
0
Embed Size (px)
Citation preview
8/8/2019 Method Validation 2370
1/37
1
Method validation and
verification
W. W. Wong
Senior Accreditation OfficerHKAS
20 November 2009
8/8/2019 Method Validation 2370
2/37
2
Method Validation
What is it?
When is it required?
Why is it necessary?
What are required? How much is adequate?
How should it be done?
8/8/2019 Method Validation 2370
3/37
3
Method Validation what are required
Validation of methods (cl. 5.4.5.3 )
The range and accuracy of the valuesobtainable from validated methods (e.g.
Uncertainty of results
Detection limit Selectivity Linearity Repeatability Reproducibility Robustness/cross-sensitivity
8/8/2019 Method Validation 2370
4/37
4
HOKLAS SC No. 37 Food TestCategory Chemical Testing
Being commented by AAB WP on Food(today is the last day for comment)
Will be published very soon Contains criteria specific to food testing With a view to assist labs
In particular, method validation and samplepreparation
8/8/2019 Method Validation 2370
5/37
5
HOKLAS SC No. 37 Section5.3 Validation of methods
Screening methods:
8/8/2019 Method Validation 2370
6/37
6
Confirmation of identity
Mass spectrometric detection
+50%+50%10% to 20%+25%+15%>20% to 50%
+20%+10%>50%
CI-GC-MS, GC-
MSn
, LC-MS,LC-MSn
EI-GC-MS
(relative)
Relative intensity
(% of base peak)
8/8/2019 Method Validation 2370
7/37
7
Limit of DetectionxLD = 3sO +xBl
WherexLD is the limit of detection
sO is the SD of the outlier free result of a matrix blank sample.
should be based on at least 10 independent completedeterminations of analytes concentration in a typical matrix
blank or low-level material, with no censoring of zero or
negative results.
xBl is the mean concentration of the matrix blank
8/8/2019 Method Validation 2370
8/37
8
Limit of Quantification
The limit of quantification represents a
concentration of the determinand that can
reasonably be determined with an acceptable
level of accuracy. Usually it is arbitrarily taken
as a fixed multiple of the detection limit.
8/8/2019 Method Validation 2370
9/37
9
Limit of Quantification
Estimated by:
xLQ = 3xLD
The factor k=3 corresponds to a relative result
uncertainty of approximately 33%.
8/8/2019 Method Validation 2370
10/37
10
Limit of QuantificationAOAC
smallest amount of analyte in a test sample that can bequantitatively determined with suitable precision and
accuracy under previously established method conditions.
It is often taken as the blank value plus 10 times the std
dev.
8/8/2019 Method Validation 2370
11/37
11
Reporting limits
Set at a level at which quantitativeresults may be obtained with a
specified degree of confidence
Limits of detection and reportingshall be verified.
8/8/2019 Method Validation 2370
12/37
12
Reporting limits
ISO/TS 13530:2009
RL is a specific concentration at orabove the limit of quantification that isreported to the client with a certain
degree of confidence. It is often definedon a project-specific basis. If the RL isset below the limit of quantification by
the client, method modification isrequired.
8/8/2019 Method Validation 2370
13/37
13
Reporting limits
ISO/TS 13530:2009
For verification of limit of detectionand limit of quantification, spiked
blank matrix samples at theseconcentration levels and blank
matrix samples shall be analysed inthe same manner as real samples.
8/8/2019 Method Validation 2370
14/37
14
Reporting limits
ISO/TS 13530:2009
If the uncertainty of results for thesamples spiked at the limit of
quantification level is smaller thanor equal to the relative precision
corresponding to the factork, thelimit of quantification is verified.
8/8/2019 Method Validation 2370
15/37
15
Reporting limits
ISO/TS 13530:2009
x = 1xLQ k
where
x = t .s/n
kis the factor for calculating the limit of
quantification as a multiple of detection,here: k=3
8/8/2019 Method Validation 2370
16/37
16
Reporting limits
In essence, the laboratory needs to provide
supporting data that, at the reporting limit,the measurement uncertainty (including biasand precision) meets the acceptance criteria.
The direct method is to find the recovery andRSD at the reporting limit and compare themwith the acceptance criteria.
8/8/2019 Method Validation 2370
17/37
17
Recommended LOD and LOQ
ML= specified maximum and/or
minimum level
For ML 0.1 mg/kg, LODML X 1/5
For ML 0.1 mg/kg, LODML X 2/5
LOQ
For ML 0.1 mg/kg, LODML X 1/10
For ML
0.1 mg/kg, LOD
ML X 1/5
LOD
8/8/2019 Method Validation 2370
18/37
18
Recommended Recovery
40-1201g/kg
60-11510g/kg
80-110100g/kg80-1101mg/kg
80-11010mg/kg
90-107100mg/kgRecovery (R)95-105
0.1% (1mg/g)
97-103 1% (1g/100g)
98-102 10% (10g/100g)
98-102100%(100g/100g)
Recovery (%)Unit
8/8/2019 Method Validation 2370
19/37
19
Precision
AOAC Official Methods of Analysis (2005)
Appendix E: Laboratory Quality Assurance
Common values of within lab precision
2010ppb
101ppm
50.01%
21%
1100%
RSDr
Concentration
8/8/2019 Method Validation 2370
20/37
20
Estimation of standard deviation by meanrange
s =R/d2
d2is 1.128 for duplicate
R is the mean range of the duplicates
8/8/2019 Method Validation 2370
21/37
21
Range of food matrices
If a method is to be accredited undergeneral foodstuffs, satisfactoryvalidation data shall be obtained for atleast five different food matrices(protein, carbohydrate, oil, dietary fibre
and water) with at least three food typesrepresentative of each food matrix.
8/8/2019 Method Validation 2370
22/37
22
0%
%FAT
%PROTEIN
%CARBOHYDRATES0%
0%
100%
100%
100%
J. AOAC, 83 (2), 413 (2000)The Referee, AOAC Intl, July 1993
8/8/2019 Method Validation 2370
23/37
23
8/8/2019 Method Validation 2370
24/37
8/8/2019 Method Validation 2370
25/37
25
Range of food matrices
The range of matrices shall also in line
with those listed in relevant regulations.Due consideration shall also be taken
for the food matrices with potentialinterferences, e.g. high chloride effect
on the ICP-MS determination.
8/8/2019 Method Validation 2370
26/37
26
Range of food matrices
General foodstuffs all foods Test procedure shall document the food matrices used in
the validation studies
Labs need to assess and determine the applicability oftheir methods to the food samples received.
8/8/2019 Method Validation 2370
27/37
27
Method verification
Released early 2008
http://www.aoac.org/alacc_guide_2008.pdf
8/8/2019 Method Validation 2370
28/37
28
How should it be done?
The following are some examples of useful references
Harmonised guidelines for single-laboratoryvalidation of method of analysis (IUPAC Technical
Report), 2002
How to meet ISO 17025 Requirements for MethodVerification, AOAC, 2007
ISO/TS 21748:2004 Guidance for the use of
repeatability, reproducibility and trueness estimatesin measurement uncertainty estimation
8/8/2019 Method Validation 2370
29/37
29
How should it be done?
CRMs
How many replicate results are required for propercomparison of the certified reference value and the
actual analysis result?
Is it necessary for the mean of test result for a CRM tolie within the uncertainty range of the certified value?
Is a result outside the uncertainty range acceptable ?
8/8/2019 Method Validation 2370
30/37
30
How should it be done?
Useful references- ISO Guide 33: 2000
Uses of certified reference materials
8/8/2019 Method Validation 2370
31/37
8/8/2019 Method Validation 2370
32/37
32
How should it be done?
Comparison of results with another laboratory(ISO 5725-6:1994 clause 7.2.4.3)
- the means of the two laboratories are compared
if ,
then statistically no significant difference
where = repeatability standard deviation
= reproducibility standard deviation
222
22
21r
Ryy
2
r
2
R
8/8/2019 Method Validation 2370
33/37
33
How should it be done - verification
Worked examples of method verification areavailable from AOAC
Determination of Total, Saturated, andMonosaturated Fats In Foodstuffs by Hydrolytic Extraction and Gas ChromatographicQuantitation: Collaborative Study
Determination of Low-Level Glucose andFructose in Raw and Refined Crystalline Sugarby High-Performance Anion ExchangeChromatography
8/8/2019 Method Validation 2370
34/37
34
Common deficiencies
Number of food matrices used not adequate
Number of food types for each matrix notadequate
Concentration levels used not adequate
Reporting limits estimated but not verified Food samples chosen not representative of the
food matrices
8/8/2019 Method Validation 2370
35/37
35
Common deficiencies
Confirmation of identity technique not availableor adequate Validation/verification data analyses not done
correctly
Measurement uncertainty not available/notestimated correctly
Equipment and/or test procedures notconforming to the test std requirements
8/8/2019 Method Validation 2370
36/37
36
Common deficiencies
Lack of or inappropriate sampling procedure QC plan acceptance limits not appropriate
8/8/2019 Method Validation 2370
37/37
37
Thank you