Multiple Stage Tandem Mass Spectrometryfor identification of plastic contaminations in blood plasma
using nano electrospray ionization
Tobias Kind and Kwang-Hyeon LiuFiehnLab UC Davis 2009
http://fiehnlab.ucdavis.edu/staff/kind/Metabolomics/LipidAnalysis
File under: Anecdotal mass spectrometryCC-by Free Reuse License granted; Tobias Kind FiehnLab 2009
Instrumentation – robotized chip based nano electrospray ionizationwith linear iontrap
Advion NanoMate robot not shown
+
nanoESI chip LTQ Linear IonTrap
Plasma_Pos_LowMS_60MSMS_CE45_01 #35-73 RT: 0.07-0.15 AV: 39 NL: 2.53E4T: ITMS + p ESI Full ms [200.00-1200.00]
640 650 660 670 680 690 700 710m/z
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
Rel
ativ
e A
bund
ance
650.55
703.64
678.64
654.18680.27
676.55 705.64690.55686.27664.55655.27648.55 701.64 706.55666.55660.45 707.55644.45
Full scan nanoESI positive mode with data dependent multiple stage tandem MS
We will fragment this ion with the awe-inspiring power ofMS/MS/MS/MS/MS/MS/MS/MS or MS8
Slowly repeat with me: MS1..MS2..MS3..MS4..MS5..MS6..MS7..MS8...
What could it be?
Plasma_Pos_LowMS_60MSMS_CE45_01 #184 RT: 1.15 AV: 1 NL: 2.29E3T: ITMS + c ESI d Full ms2 [email protected] [170.00-675.00]
200 250 300 350 400 450 500 550 600 650m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rel
ativ
e A
bund
ance
607.32
551.24
495.26
359.24 604.28 631.35540.52325.27 437.36212.44 283.20
MS^2precursor m/z 664.00 (ESI+)
-56.8
-112.76
-168.74
Plasma_Pos_LowMS_60MSMS_CE45_01 #190 RT: 1.27 AV: 1 NL: 1.16E2T: ITMS + c ESI d Full ms8 [email protected] [email protected] [email protected] [email protected] 439.26 ...
100 150 200 250 300m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100R
elat
ive
Abu
ndan
ce233.03
214.97
153.09 251.07
291.17171.07309.12
174.99
MS^8 (ESI+)precursor m/z 327.00
Plasma_Pos_LowMS_60MSMS_CE45_01 #184-190 RT: 1.15-1.27 AV: 7 NL: 2.99E3T: Composite spectrum MS8 664.00,607.32,551.33,495.34,439.26,383.17,327.06
100 200 300 400 500 600m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rel
ativ
e A
bund
ance
551.29
607.32495.31
327.08 439.26383.17
233.03359.24 631.35540.52153.09 251.07 480.52419.50
56 Da
56 Da
56 Da
56 Da
56 Da
MS^8 composite tandem mass spectrumof precursor m/z 664.00 (ESI+)
Sample: KH Liu (FiehnLab)
Plasma_Pos_LowMS_60MSMS_CE45_01.rawFriday, April 03, 2009
MS2 precursor 664.00|__MS3 precursor 607.32| |__MS4 precursor 551.33| | |__MS5 precursor 495.34| | | |__MS6 precursor 439.26| | | | |__MS7 precursor 383.17| | | | | |__MS8 precursor 327.06| | | | | | |__Composite spectrum MS8 664.00,607.32,551.33,495.34,439.26,383.17,327.06| | | | | | |__Single spectrum MS8 664.00,607.32,551.33,495.34,439.26,383.17,327.06 (190)| | | | | |__Composite spectrum MS7 664.00,607.32,551.33,495.34,439.26,383.17| | | | | |__Single spectrum MS7 664.00,607.32,551.33,495.34,439.26,383.17 (189)| | | | |__Composite spectrum MS6 664.00,607.32,551.33,495.34,439.26| | | | |__Single spectrum MS6 664.00,607.32,551.33,495.34,439.26 (188)| | | |__Composite spectrum MS5 664.00,607.32,551.33,495.34| | | |__Single spectrum MS5 664.00,607.32,551.33,495.34 (187)| | |__Composite spectrum MS4 664.00,607.32,551.33| | |__Single spectrum MS4 664.00,607.32,551.33 (186)| |__Composite spectrum MS3 664.00,607.32| |__Single spectrum MS3 664.00,607.32 (185)|__Single spectrum MS2 664.00 (184)
List of ions from precursor m/z 664.00 (ESI+)
List of product ions:m/z 664 , 607 ,551 ,495 ,439 ,383 ,327accurate masses see above
Solutions – well Plasma – well Lipids use LipidMaps tools!
Current mass accuracy for unit mass resolution LTQ: 78 ppm (external calibration) or 0.046 Da @ 594.46 Da for reference purposesLipidMaps Tools used with M+H
CE(18:2(Ep)) 18 2 663.5711 0.4289 C45H75O3+CE(18:2(Ke)) 18 2 663.5711 0.4289 C45H75O3+CE(18:3(OH)) 18 3 663.5711 0.4289 C45H75O3+CE(18:1(Ep))-cyclo 18 1 663.5711 0.4289 C45H75O3+CE(18:1(Ke))-cyclo 18 1 663.5711 0.4289 C45H75O3+CE(18:2(OH))-cyclo 18 2 663.5711 0.4289 C45H75O3+CE(16:2(Ke,Ep2)) 16 2 663.4983 0.5017 C43H67O5+CE(16:2(Ke2,Ep)) 16 2 663.4983 0.5017 C43H67O5+CE(16:3(Ep,OO)) 16 3 663.4983 0.5017 C43H67O5+CE(16:3(Ke,OO)) 16 3 663.4983 0.5017 C43H67O5+CE(16:3(OH,Ep2)) 16 3 663.4983 0.5017 C43H67O5+CE(16:3(OH,Ke,Ep)) 16 3 663.4983 0.5017 C43H67O5+CE(16:3(OH,Ke2)) 16 3 663.4983 0.5017 C43H67O5+CE(16:4(OOH,Ep)) 16 4 663.4983 0.5017 C43H67O5+CE(16:4(OOH,Ke)) 16 4 663.4983 0.5017 C43H67O5+CE(16:4(OH,OO)) 16 4 663.4983 0.5017 C43H67O5+CE(16:4(OH2,Ep)) 16 4 663.4983 0.5017 C43H67O5+CE(16:4(OH2,Ke)) 16 4 663.4983 0.5017 C43H67O5+CE(16:5(OH,OOH)) 16 5 663.4983 0.5017 C43H67O5+CE(16:5(OH3)) 16 5 663.4983 0.5017 C43H67O5+CE(16:1(Ke,Ep2))-cyclo 16 1 663.4983 0.5017 C43H67O5+CE(16:1(Ke2,Ep))-cyclo 16 1 663.4983 0.5017 C43H67O5+CE(16:2(Ep,OO))-cyclo 16 2 663.4983 0.5017 C43H67O5+CE(16:2(Ke,OO))-cyclo 16 2 663.4983 0.5017 C43H67O5+CE(16:2(OH,Ep2))-cyclo 16 2 663.4983 0.5017 C43H67O5+CE(16:2(OH,Ke,Ep))-cyclo 16 2 663.4983 0.5017 C43H67O5+CE(16:2(OH,Ke2))-cyclo 16 2 663.4983 0.5017 C43H67O5+CE(16:3(OOH,Ep))-cyclo 16 3 663.4983 0.5017 C43H67O5+CE(16:3(OOH,Ke))-cyclo 16 3 663.4983 0.5017 C43H67O5+CE(16:3(OH,OO))-cyclo 16 3 663.4983 0.5017 C43H67O5+CE(16:3(OH2,Ep))-cyclo 16 3 663.4983 0.5017 C43H67O5+CE(16:3(OH2,Ke))-cyclo 16 3 663.4983 0.5017 C43H67O5+CE(16:4(OH,OOH))-cyclo 16 4 663.4983 0.5017 C43H67O5+CE(16:4(OH3))-cyclo 16 4 663.4983 0.5017 C43H67O5+
Solutions – well Plasma – well Lipids!
CE(18:2(Ep)) 18 2 663.5711 0.4289 C45H75O3+CE(18:2(Ke)) 18 2 663.5711 0.4289 C45H75O3+CE(18:3(OH)) 18 3 663.5711 0.4289 C45H75O3+CE(18:1(Ep))-cyclo 18 1 663.5711 0.4289 C45H75O3+CE(18:1(Ke))-cyclo 18 1 663.5711 0.4289 C45H75O3+CE(18:2(OH))-cyclo 18 2 663.5711 0.4289 C45H75O3+
Sphingosine CerP 18:1 644.5013 0.5013 C36H71NO6P+Sphingosine HexCer 12:0 644.5096 0.5096 C36H70NO8+Sphinganine Cer 24:4 644.5976 0.5976 C42H78NO3+
Problem: Our mass accuracy is 0.046 Da but the solutions are 0.5 Da difference
Without Google Scholar we would be lost to search single m/z valuesor numbers. Still its a horrible way to do mass spectral library search.Has anybody a Google API example?
Solutions – well Plasma – well Google It!
On the lipid composition of human meibum and tears: Comparative analysis of nonpolar lipidsIgor A. Butovich http://www.iovs.org/cgi/content/abstract/iovs.08-1889v1
Researcher working with tears = The Dacriologist (Book by Djerassi)Dakryology, the neologism for the science of tears10.1007/BF00137944also CRYING: The Natural and Cultural History of Tears By Tom Lutz Norton
"However, rushing to this conclusion would have been a mistake because subsequent fragmentation of ion m/z 663 in sequential MSn experiments gave us the following chain of product ions differing by 56 amu: 663 (MS1) -> 607 (MS2) -> 551 (MS3) -> 495 (MS4) -> 439 (MS5) -> 383 (MS6) -> 327 (MS7) (not shown). Note that the first four ions (663, 607, 551, 495) were also easily detectable in a simple MS1 experiment (Fig. 11A) . These transformations were indicative of a sequential loss of several up to six) t-butyl groups and were formerly described for oxidized form of Irgafos 168 (C42H63O4P, Fig. 11A , inset) a polymer additive present in polyethylene and polypropylene and its derivatives."
Solutions – well Plasma – well Plasticizers!
Analysis of Additives in Polyethylene with DesorptionChemical Ionization/Tandem Mass Spectrometry
S. W. CHEN and G. R. HER*Appl. Spectrosc. 47, 844-851 (1993) http://www.opticsinfobase.org/abstract.cfm?URI=as-47-6-844
Also Oxidized Naugard 524 - http://dx.doi.org/10.1016/0003-2670(95)00183-Z
Picture Source: S. W. CHEN and G. R. HER Appl. Spectrosc. 47, 844-851 (1993)
Solution – probably Irganox / Irgafos contamination
Picture Source: Analysis of Additives in Polyethylene with DesorptionChemical Ionization/Tandem Mass SpectrometryS. W. CHEN and G. R. HER*Appl. Spectrosc. 47, 844-851 (1993) http://www.opticsinfobase.org/abstract.cfm?URI=as-47-6-844
Plasma_Pos_LowMS_60MSMS_CE45_01 #184-190 RT: 1.15-1.27 AV: 7 NL: 2.99E3T: Composite spectrum MS8 664.00,607.32,551.33,495.34,439.26,383.17,327.06
100 200 300 400 500 600m/z
0
10
20
30
40
50
60
70
80
90
100
Rel
ativ
e A
bund
ance
551
607
495
327
439
383
Our MS^8"Lipid"
Problem solved, "lipid" is oxidized plasticizer.
Is the problem really solved?
CID 14572930
InChIKey: AZSKHRTUXHLAHS-UHFFFAOYSA-N
No, because for the formula = C42H63O4Pthe exact mass = 662.446397016.
The [M+H]+ would be 663.4542 Da and not 664.0
In this case an ultrazoom investigation orinfusion into FT-ICR-MS with resolving power50,000 to 100,000 would probably help.
MS/MS on isotopic peak, ion suppression of isobars, overlap with multiple ceramides or cholesterol and oxycholesterol species possible
Final question – How can we avoid rediscovery (reinventing the wheel)?
• sharing experiments (free)• sharing mass spectra (free or sell)• putting publications into OA for easier searching
How can we prevent plasticizer contaminations?
• running solvent blanks with each experiment• running methods blanks with each experiment• using only (baked) glassware
Other solutions?
• library search (but compound is not in NIST)• remember the awesome power of MS/MS/MS/MS/MS/MS/MS/MS• modern iontraps are currently not used to the full extend Data Dependent • exclusion libraries for unwanted "contaminations"
See: XLS supplement from Keller, B.O.; Sui, J.; Young, A.B.; Whittal, R.M. Interferences and contaminants encountered in modern mass spectrometry.