55
Multiwave Imaging and Elastography Mathias Fink [email protected] Institut Langevin, Ecole Supérieure de Physique et Chimie de La Ville de Paris, ESPCI ParisTech, Paris, France

Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

  • Upload
    lytram

  • View
    234

  • Download
    1

Embed Size (px)

Citation preview

Page 1: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Multiwave Imaging and Elastography

Mathias Fink

[email protected]

Institut Langevin, Ecole Supérieure de Physique et Chimie de La Ville de Paris,

ESPCI ParisTech, Paris, France

Page 2: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Fibrotic

Lesion

Carcinoma

Grade II

Viscous cyst

A difficult problem for radiologists : breast cancer detection

Begnin Begnin Malign

Ultrasound Images of different breast lesions

Good sensitivity but bad specificity

Page 3: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

How to improve specificity ?

Multi-modality : superposition of two images, one morphology and one metabolic

Page 4: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Two examples : morphology /metabolic activity

PET/CT Scan PET/MRI

Page 5: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Multi-Modality Imaging

Superposition of 2 images each obtained with a single wave

One single wave is sensitive only to a given Contrast :

Spatial Resolution depends on wave physics laws

and on sensor technology.

Ultrasound to bulk compressibility , Optical wave to dielectric permitivity and optical absorption, Sonic Shear wave to shear modulus, viscosity LF Electromagnetic wave to electrical impedance, conductivity X ray to density Gamma ray to radio tracer distribution ……..

Page 6: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Spatial resolution with one-wave imaging system

Spatial resolution d ? z observation depth l wavelength l* transport mean free path

In strongly heterogeneous medium ( multiple scattering) , waves lost their coherence on a distance called the transport mean free path l*

l*

Transport mean free path l*

Distance needed for a wavefront to loose the memory of its initial direction. In all tissues l*> l

Optical wave in tissues l* ~ 500 µm Ultrasound in tissues , l * > 1 m LF Electromagnetic waves, l * >> 1 m

Page 7: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Spatial resolution with one-wave imaging system

1. z<l

2. l<z<l*

3. l<l*<z

Near field Imaging

Coherent wave propagation

Diffusive regime

- Elec. Impedance Tomography (EIT) - EEG, MEG - Near Field Optics

- Ultrasound - Opt.Coh.Tomography - C.T., X-Ray diffraction

Deep optical tomography

Spatial resolution d ? 3 different regimes

d ~ z

d ~ l

d ~ z

Page 8: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Multi-Wave Imaging: A physicist approach

Wave n°1

Wave n°2

Poor spatial resolution Very interesting contrast

Excellent spatial resolution Poor contrast

Interactions with

For example : Ultrasound

Page 9: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

How to play Multiwave Imaging ? Three potential Interactions between different waves

• The interaction of the first wave with tissues can generate a second kind of wave

• The first wave IS TAGGED locally by a second kind of wave

• A first wave travelling much faster than the second one can be used to produce a movie of the slow wave propagation

PhotoAcoustic Imaging ThermoAcoustic Imaging

AcoustoOptical Imaging Electrical Impedance Imaging with Ultrasound

Transient Elastography Shear Wave imaging (Supersonic mode)

A unique case that allows the observation of the near field of the slow

wave inside the body ,

Page 10: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

I – A wave generates a second type of wave : Photo-Acoustics

Page 11: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Photo-Acoustics

L.. V Wang, Nature Biotech 24, 848, 2006, S. Emilianov et al, Physics Today, 2009

Page 12: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

II - A wave is tagged by another wave : Tagging Photons with ultrasound

Optical Speckle

Speckle is modulated

fl

fus

Ultrasound induces 1. Displacement of the scattering centers 2. Modulation of the refractive index G. Maret, LH. Wang, C. Boccara, S Leveque, F. Ramaz M. Groos

Optical

coherent

illumination

Page 13: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Imaging an absorbing inclusion

Axial pattern

L = 2.3 cm µs’=µs(1-g)= 6 cm-1

US: 2.3 MHz, 4 cycles, 1 ms

« Photorefractive acousto-optic imaging in thick scattering media at 790 nm with a Sn2P2S6:Te crystal »

S.Farahi, G Montemezzani, A. Grabar, JP. Huignard, F. Ramaz Optics Letters (2010)

Axial resolution = 2.6 mm

Experimental results in vitro laser @790 nm + US bursts

Agar Agar + Intralipid-10% + 1 inclusion 3 mm x 10 mm

More difficult to implement in-vivo because the tissue speckle

has a coherence time smaller than 1 ms

Page 14: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

III - A wave produces a movie of another wave : Transient Elastography :

How to image elastic properties of tissues with millimetric resolution ?

Page 15: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

K Bulk Modulus (Compression) almost constant, of the order of 109 Pa, Fluctuations 5%

Quasi incompressible medium

µ Shear Modulus, Strongly heterogeneous, varying between 10 2 and 10 7 Pa

(A. Sarvazian)

K >> m

K

m

What kind of mechanical waves can propagate in soft tissues ?

Two types of waves related to the two mechanical coefficients

K and µ used to define the elasticity of a solid material

Young modulus

E 3 µ

Page 16: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Human Body Seismology : Mechanical waves in soft tissues

Compressional Waves propagate at

Shear waves propagates at

P

Kc

( 1500 m.s-1)

( 1-10 m.s-1)

Two kind of waves propagating at totally different speeds !!

At Sonic frequency, Shear waves can propagate < 1000 Hz (High Shear Viscosity), at 200 Hz, large wavelength = 2cm

Ultrasonic radiation force

At Ultrasonic frequency, only Compressional waves can propagate, at 5MHz, wavelength = 0.3mm.

sc

m

Page 17: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Transient Elastography : a Multiwave approach

• Generation of transient low frequency shear wave (10 Hz to

1000 Hz) with some microns amplitude

3E

scm

5.000 images.s-1 !

On observe alors la propagation de ces ondes lentes, soit avec un unique transducteur ultrasonore, soit avec un réseau de transducteurs capable de délivrer des images à une cadence très élevée (5 000 images par seconde) On mesure en chaque point de l’image la vitesse des ondes de cisaillement et on en déduit une carte d’élasticité à partir de

la relation :

• One follows tissue motion induced by shear waves 5.000 times/s. Local measurement of the shear velocity and E ou m are deduced by relation :

Page 18: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

In a first step (1994) we observed transient shear waves with a single ultrasonic transducer . Then in 2000 a company ECHOSENS was created (45 persons) to develop the Fibroscan to get a global measurement of liver elasticity

1994 2001

1D Transient Elastography

M.Fink, S. Catheline, L. Sandrin

Page 19: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

From 1D Transient Elastography to SuperSonic Shear Wave Imaging

Research Work in Laboratoire Ondes et Acoustique (now Institut Langevin)

Fibroscan Aixplorer

Page 20: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

2D Transient Elastography needs an ultrafast ultrasound imaging system

Page 21: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

D

F

How to make an ultrafast ultrasound scanner ? : Time reversal

RAM

Time reversal and numerical propagation

traitement

128 shots for one image, 50 frames /second

1 shot for one image, 5000 frames/s

Page 22: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Ultrasound Transducer

Imaged Area

x

z

Focal zone

Force

),(),( 2

2trp

ctrF

Ultrasonic Radiation Force non-linear and dissipative effects

Typical ultrasonic bursts of 100 µs to create low frequency pushes (10 micrometers displacement)

A. Sarvazian, J. Greenleaf, C. Nithingale, G. Trahey, M. Fink, M Tanter

Transient Elastography and Ultrasonic Radiation Force

Page 23: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Dc

F

shear

USl

2 4 6 8 10 12 18 20 -0.1

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Point Spot experiment

14 16 22 ms

F

Dcf

US

shearshear

l

4

1 mm (FWHM)

2 m/s

fshear= 500 Hz

100 µs excitation

Shear Wave Bandwidth generated by the Ultrasonic Radiation force ?

Page 24: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

It is possible to measure displacements of 1 m (l/1000) between 2 consecutive shots

How to measure the displacements induced by shear waves ?

Tissues behave as random distributions of scatterers. The speckle is moving with shear wave One repeat ultrasonic shots at high rate (> 5000 shots/s) and create an ultrafast movie

shot 1 shot 2 shot 3 shot 4 shot 5

Moving window cross-correlation gives the axial displacements

),,( tzxuz

t

Page 25: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

6 m/s

2 m/s

Transducer

The Supersonic Push !!!!!!!

Conventional US time

0 s 1 s

Transducer

A 20 ms Experiment !!

Ultrafast US

Supersonic moving source

20 ms

Page 26: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

kPa

Multiwave Imaging of a hard inclusion

Movie duration 20 ms

Ref: Supersonic Shear Imaging: a new technique for soft tissue elasticity mapping. J. Bercoff, M. Tanter and M. Fink, IEEE Trans., April 2004

Page 27: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

2

2

2

2

2

2

z

u

x

u

y

u zzz

z

z ut

u

m

2

2

2

2

2

2

z

u

x

uu zz

z

A Simple Inversion Algorithm

1) The medium is considered as infinite, isotropic, purely elastic and locally homogeneous.

2) l>>m => the bulk wave propagates instantaneously, and then:

3) =>

uut

u

mml ).()(

2

2

- Motion Equation : an ideal model : isotropic solid without dissipation

- Assumptions:

No diffraction outside the image plane

Compressional shear

Page 28: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

2

2

2

2

2

2

),(),(

),(

),(

z

zxu

x

zxu

t

zxu

zxzz

z

m

Inverse Problem

•Local inversion algorithm

z

z ut

u

m

2

2

Page 29: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Hard Inclusion with a liquid zone

Page 30: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

The Evolution of Ultrafast Imaging Technology

Aixplorer ©

1996-2002 2004-2005 2006-2007

Time Reversal Prototype

SuperSonic Imagine was founded in September 2005 by Jacques Souquet, 120 employees, Aix en Provence and Seattle

Page 31: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Echographic System with Real-time and Quantitative Elastography

First SSI experiment : May 2002 45 Minutes processing

SSI Prototype 2006 some seconds processing

October 2007 0.2 seconds processing

Page 32: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Supersonic Shear Wave Imaging:

Spatial resolution

Elasticity contrast Axial Res (mm) Lateral Res (mm)

2 1 1.1

3 1.2 1.2

10 1.3 1.1

Axial and lateral resolution in a two layers medium : around 1 mm

Lateral resolution Axial resolution

Page 33: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Multiwave imaging and super-resolution Shear wavelength : typicaly 10 mm Spatial resolution on the shear modulus : 1 mm (lUS)

Ultrasonic Array

Movie of the shear wave near-field closed to each heterogeneities

Multi-Wave Imaging allows to get the Contrast of One Wave with the Resolution of the Second Wave

several hundreds of lUS

Several tens of lshear

Page 34: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Medical applications of Elasticity Multiwave Imaging

• Breast • Tyroid • Liver • Kidney • Muscle • Vascular • Cardiac • Eye • Prostate • Monitoring therapy (RF ablation, HIFU)

Page 35: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

35

Case #2 – Slide 2 of 2

Dr Balu Maestro, Nice France

Unknown

September 09th 2008

Same patient

Secondary lesion undetected at mammo but at US. BIRADS 4 at US. 3mm length.

This secondary lesion is an IDC Grade III & HR+ of 15mm.

Unknown

Emax > 150kPa in the center of this 3mm lesion.

Breast Imaging

Invasive ductal carcinoma

Page 36: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

36

Case #4 – Slide 1 of 1

Dr Balu Maestro, Nice France

56 years old

September 17th 2008

BIRADS 5 at US.

IDC Grade I, partially necrotic center proved by histology.

None

Emax > 200kPa on surrounding tisue.

E = 70kPa in the center.

Breast Imaging

Page 37: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

37

Case #5 – Slide 2 of 2

Dr Balu Maestro, Nice France

Unknown

July 17th 2008

BIRADS 3 at mammo & US.

Suspect margins.

Fibro-adenoma

Mother’s breat cancer.

Emean < 30kPa and totally homogeneous.

Breast Imaging

Page 38: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Diagnostic impact in breast :

benign benign Malignant

Page 39: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Shear Wave Imaging for Liver fibrosis Staging

F 3 versus F 4

Clinical Study on 118 patients with Hepatitis C

Fibroscan®

S.W.I.

Bavu E., Gennisson J.-L., Couade, M. Bercoff j., Mallet V., Fink M. Vallet-Pichard A., Nalpas B., Tanter M., Pol S. Non-invasive liver fibrosis staging using supersonic shear imaging: A clinical study on 113 HCV patients., under review, 2010.

Page 40: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave
Page 41: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Prostate – multiwave imaging

Suspicious lesion: much harder in SWE

Page 42: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Dynamics of Muscle Contraction

Gastrocnemius Contraction

Soleus Contraction

Shinohara S., Sabra K., Genisson J.-L., Fink M., Tanter M. "Real-time visualization of muscle stiffness distribution with ultrasound SWI during muscle contractions », Muscle and Nerve, June 2010

Coll. M. Shinohara, K. sabra, Georgia Tech. University, Usa

Page 43: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Shear Wave Dispersion

• Does the shear wave velocity depends of frequency ?

• Does always valid ?

• Origin 1 : viscosity

• Origin 2 : guides wave

sc

m

Page 44: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

SuperSonic Shear Imaging : Liver in vivo

2000 frames per second

50

mm

36 mm

• Linear Probe L7-4 4-7 MHZ 128 elts. • Mechanical Index Push 1.4 Imaging 0.7 • ISPTA Push+ Imaging 600 mW.cm-2 • Less energy deposit than Color Doppler !!!

26 Years old healthy volunteer

• Intercostal Exam

M. Tanter, G. Montaldo, T Deffieux, JL Gennisson, J Bercoff, M.Fink

Page 45: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Tissue Rheology with Shear wave dispersion

2.7 ms

-0.05

0

0.05

-0.05

-0.04

-0.03

-0.02

-0.01

0

0.01

0.02

0.03

0.04

0.05

4.6 ms

-0.05

0

0.05

-0.05

-0.04

-0.03

-0.02

-0.01

0

0.01

0.02

0.03

0.04

0.05

6.5 ms

-0.05

0

0.05

-0.05

-0.04

-0.03

-0.02

-0.01

0

0.01

0.02

0.03

0.04

0.05

Z

X

50-350 Hz

C B A

Page 46: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Can we assess viscoelastic properties of tissues using SSI ?

)()()( . xjxLtkxjtkrj eeee a

xc

x

)(

SuperSonic Wave Generation Plane Wave Approximation is valid !!!

0 100 200 300 400 500 600 7000

1

2

3

4

5

6

Frequency (Hz)

Shear

Wave S

peed

Shear Wave Phase Speed (m.s-1) Versus Frequency

0 100 200 300 400 500 600 7000

1

2

3

4

5

6

Frequency (Hz)

Shear

Wave S

peed

0 100 200 300 400 500 600 7000

1

2

3

4

5

6

Frequency (Hz)

Shear

Wave S

peed

Low Visc. Gel In vivo Liver

High Visc. Gel

Pushing Line Location

01

2

3

3210

Page 47: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Can we assess viscoelastic properties of tissues using SSI ?

100 150 200 250 300 350 400 4501

2

3

4

5

Frequency (Hz)

She

ar W

ave

Spe

ed

It can even be local ! A concept of real-time «Shear Wave Spectroscopy »

Dep

th in

mm

100 150 200 250 300 350 400 450 1

2

3

4

5

Frequency (Hz)

Shea

r W

ave S

pee

d(m

.s-1

)

100 150 200 250 300 350 400 450 1

2

3

4

5

Frequency (Hz)

Shea

r W

ave S

pee

d

Muscle

Liver

Liver

T. Deffieux, M. Tanter, G Montaldo, J. Bercoff. M. Fink

Page 48: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Shear Wave Spectroscopy : a broadband approach

Shear Wave Bandwidth

Static elastography

Dynamic elastography

SSI

50 150 250 350 450 550 0 650 Hz

10

20

30

5

10 15 20 25 30 35 40 45

(a)

Agar gel Viscous gel

Frequency in Hertz Lateral position in mm

100 200 300 400 500 600 0

2

4

6

8 (a) (b)

Agar gel

Viscous inclusionl

Agar gel

Viscous inclusion

Ph

ase

velo

city

in m

/s

Dep

th in

mm

Page 49: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Voigt Maxwell

0)(

2

2

2

U

iµx

U

)(2

)(α

)(

)(2C

222

2222

T

222

222

T

µ

µµ

µµ

µ

µ

µ

µ

µ

2

)11(

α

)11(

2C

22

2

2

T

22

2T

(ex: Polymères) (ex: Rubber)

Constitutive equation

.)(t

µ

µ1

µ2

µ1 µ2

.)(

1D Helmoltz equation

0)(2

2

2

U

µi

x

U

Isotropic, homogenous,

linear, viscoelastic What Viscoelastic model ?

Page 50: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Wave Dispersion on arterial phantoms (agar-agar, gelatin)

• Wave velocity is strongly reduced when shear wave is generated in a thin layer : guided propagation

- Frame rate ~ 5000 to 10000 Hz

- 8 MHz central

frequency

mc

Page 51: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Guided shear wave along a tube

• Guided shear wave is dispersive : phase velocity is a function of the frequency

?

Homogeneous medium Tube

Page 52: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

In vivo application

• In vivo experiment on healthy volonteers

• Dispersion curves behavior is similar to phantom

• Measured shear modulus : 𝝁~ 90 kPa

In vivo

Page 53: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Couade M, Pernot M., Messas E., Prada C., Emmerich J.,Criton A,Fink. M and Tanter M. Ultr. Med. Biol. , Aug. 2010

In vivo Acquisition on a 27 Years old Volunteer

Real Time Elasticity of the carotid during one single cardiac cycle

Generating a « pushing beam » at the surface of the arterial wall enables the precise estimation of local visco-elastic properties of arterial wall

Athérosclérosis, fibrodysplasia, myocardial fibrosis…

- Frame rate

- ~ 5000 to 10000 Hz

- 8 MHz central freq.

- 10 movies acquired

- per cardiac cycle

Collaboration Hospital G. Pompidou, Paris N=70 Healthy Volunteers Mean Intra-individual

Reproducibility (s.d)

Intra-individual Reproducibility (relative error)

local PWV (m/s) 5,45 0,68 12%

Shear wave velocity (m/s) @ 900 Hz 5,68 0,2 4%

Page 54: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

Arterial stiffness estimation

Propagation of shear wave (Lamb wave) in the arterial

wall

Exemple of a hard plaque

Page 55: Multiwave Imaging and Elastography - · PDF fileMultiwave Imaging and Elastography ... K and µ used to define the elasticity of a solid material ... Tissue Rheology with Shear wave

MultiWave Imaging

• Ultrasound based Shear Wave imaging

• MRI based Shear Wave imaging

• Shear Wave AcoustoOptics

• AcoustoOptics

• PhotoAcoustics

• Electric Impedance / Ultrasound Imaging

• Current Density / Ultrasound Imaging

• US Imaging of mechanical contraction

due to Action potentials

• MR Imaging of Ultrasonic Radiation force

• Transient Shear Waves / OCT

• …

One wave gives the contrast. The other wave gives the spatial resolution.