Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Engineered by HBM FiberSensing
User Manual
MuST VII Dynamic Reading Unit
English
MuST VII Dynamic Reading Unit
Equipment User Manual
DynamicMONITOR User Manual
User manual version: 2.0
07.2015
Reading Unit version: 1.0
DynamicMONITOR Software version: 1.2
Subject to modifications.
All product descriptions are for general information only.
They are not to be understood as a guarantee of quality or durability.
Manual-2.0
Smartec 3
1 Technical details .......................................................................................... 5
1.1 General Information........................................................................................ 5
1.2 System Components ...................................................................................... 5
1.3 MuST VII Dynamic Technical Data .................................................................. 6
1.3.1 Performance .................................................................................................. 6
1.3.2 Technical Characteristics ................................................................................ 6
2 Regulatory and Certification Considerations ............................................... 7
2.1 Environment Considerations ........................................................................... 7
2.1.1 Disposal of your Old Appliance........................................................................ 7
2.2 Laser Safety .................................................................................................. 7
2.2.1 Symbols ........................................................................................................ 8
2.2.2 Class 1 Laser ................................................................................................. 8
2.2.3 General Precautions ....................................................................................... 9
3 Operation ................................................................................................... 10
3.1 Connectors .................................................................................................. 10
3.1.1 Type of connectors ....................................................................................... 11
3.2 Setting Up.................................................................................................... 11
3.2.1 Power supply ............................................................................................... 11
3.2.2 Optical connectors ....................................................................................... 12
3.2.3 Bottom plate ................................................................................................ 13
3.3 Switching On ............................................................................................... 13
3.3.1 Internal PC .................................................................................................. 13
3.3.2 Acquisition Module ....................................................................................... 13
3.4 Switching OFF ............................................................................................. 14
3.4.1 Internal PC .................................................................................................. 14
3.4.2 Acquisition Module ....................................................................................... 14
3.5 Power and Status LEDs ................................................................................ 14
3.5.1 Power LED .................................................................................................. 14
3.5.2 Status LED .................................................................................................. 14
3.6 Interface ...................................................................................................... 15
3.7 Control ........................................................................................................ 15
3.8 Cleaning Procedure...................................................................................... 15
3.8.1 Dry cleaning technique ................................................................................. 18
3.8.2 Wet cleaning technique ................................................................................ 20
3.9 Fuse Replacement ....................................................................................... 22
4 DynamicMONITOR Software ...................................................................... 24
4 Smartec Manual-2.0
4.1 General Bar ................................................................................................. 24
4.1.1 Acquisition ................................................................................................... 26
4.1.2 Exit Application ............................................................................................ 30
4.2 Graphical Area ............................................................................................. 30
4.2.1 Graphical View ............................................................................................. 30
4.2.2 Numeric View............................................................................................... 31
4.2.3 Configuration ............................................................................................... 32
4.2.4 Spectral View ............................................................................................... 39
4.2.5 FFT View ..................................................................................................... 41
4.2.6 SCPI Interface ............................................................................................. 42
Regulatory and Certification Considerations
Manual-2.0
Smartec 5
1 Technical details
1.1 General Information
The MuST VII Dynamic is a continuous swept laser scanning
measurement unit for interrogating SOFO and fiber Bragg grating
(FBG) sensors. It includes a NIST traceable wavelength reference that
provides continuous calibration to ensure system accuracy over long
term operation. The high dynamic range and high output power allows
high resolution to be attained even for long fiber leads and lossy
connections.
Multiple FBG sensors can be connected in series in each optical fiber.
This, in combination with the four or eight optical channels with parallel
acquisition, make the MuST VII Dynamic particularly suited for large
scale sensing networks, acquiring a large number of sensors
simultaneously , providing 100S/s acquisition rates with 5 pm
resolution.
This Manual applies to the following equipment:
PN S12.2030-1CH
MuST VII Dynamic Reading Unit • 1 OC • Standard Temperature Operation Range
PN S12.2030-4CH
MuST VII Dynamic Reading Unit • 4 OC • Standard Temperature Operation Range
1.2 System Components
The MuST VII Dynamic set includes:
► Measurement Unit
► Power cord
► Adapter protection caps
Regulatory and Certification Considerations
6 Smartec Manual-2.0
1.3 MuST VII Dynamic Technical Data
1.3.1 Performance
MuST /FBG Sensor Measurement
Measurement Resolution 5 pm
Linearity/Accuracy ±10 pm
Measurement Range 100 nm (1500 to 1600 nm)
Calibration NIST traceable wavelength reference
Measurement time 0.1 s
Available channel count 1 or 4 channels total
1.3.2 Technical Characteristics
AC power supply 230 V 50 Hz / 110 V 60 Hz Auto detect
External connections Ethernet connection, 1 or 4 optical ports, power supply, display
and screen (optional)
Dimensions 500 mm x 500 mm x 210 mm
Weight ~25 Kg
Operating temperature 10 ºC to +40 ºC
-40 ºC to + 40 ºC with heating option*
Humidity 90% non-condensed
Protection Index IP 66†
Dimensions 163mm x 103 mm x 53 mm
Weight 0.43 Kg
Material Aluminum
Operating temperature -25 to 70 ºC
Humidity < 90% at 40 ºC
* optional
† if the base of the Reading Unit is properly drilled and hermetic gland nuts are used.
Regulatory and Certification Considerations
Manual-2.0
Smartec 7
2 Regulatory and Certification Considerations
2.1 Environment Considerations
2.1.1 Disposal of your Old Appliance
When the attached symbol combination - crossed-out wheeled bin and
solid bar symbol – is attached to a product it means the product is
covered by the European Directive 2002/96/EC and is applicable in
the European Union and other countries with separate collection
systems.
All electrical and electronic products should be disposed of separately
from the municipal waste stream or household via designated
collection facilities appointed by the government or the local
authorities. The correct disposal of your old appliance will help prevent
potential negative consequences for the environment and human
health. For more detailed information about disposal of your old
appliance, please contact your city office, waste disposal service or
distributor that purchased the product.
HBM FiberSensing is a manufacturer registered in the ANREEE -
"Associação Nacional para o Registo de Equipamentos Eléctricos e
Electrónicos" under number PT001434. FiberSensing celebrated a
"Utente" type contract with Amb3E - "Associação Portuguesa de
Gestão de Resíduos de Equipamentos Eléctricos e Electrónicos",
which ensures the transfer of Electrical and Electronic appliance waste
management, i.e. placing Electronic and Electrical appliances in the
Portuguese market, from the manufacturer (HBM FiberSensing) to
Amb3E.
2.2 Laser Safety
The MuST VII Dynamic product contains a laser in its core. A laser is a
light source that can be dangerous to people exposed to it. Even low
power lasers can be hazardous to a person's eyesight. The coherence
and low divergence of laser light means that it can be focused by the
eye into an extremely small spot on the retina, resulting in localized
burning and permanent damage.
Regulatory and Certification Considerations
8 Smartec Manual-2.0
The lasers are classified by wavelength and maximum output power
into the several safety classes: Class 1, Class 1M, Class 2, Class 2M,
Class 3R and Class 4.
2.2.1 Symbols
Warning symbol
Class 1 Laser symbol
Fig. 2.1
2.2.2 Class 1 Laser
The MuST VII Dynamic is a class 1 laser product:
«Any laser or laser system containing a laser that cannot emit laser
radiation at levels that are known to cause eye or skin injury during
normal operation. »
It is safe under all conditions of normal use. No safety requirements
are needed to use Class 1 laser devices. This product contains a laser
within an enclosure that prevents exposure to the radiation and that
cannot be opened without shutting down the laser.
Regulatory and Certification Considerations
Manual-2.0
Smartec 9
2.2.3 General Precautions
Everyone who uses a laser’s equipment should be aware of the risks.
The laser radiation is not visible to the human eye but it can seriously
damage user’s eyesight.
The laser is enabled when the interrogator is turned on.
Users should never put their eyes at the level of the horizontal plane of
the optical adapters of the interrogator or uncovered optical
connectors.
Adequate eye protection should always be required if there is a
significant risk for eye injury.
Do not attempt to open or repair a malfunction interrogator. It must be
returned to Smartec for repair and calibration.
Regulatory and Certification Considerations
10 Smartec Manual-2.0
3 Operation
3.1 Connectors
Front View
Bottom View
Fig. 3.1
Technical details
Manual-2.0
Smartec 11
The connectors and buttons on Fig. 3.1 are:
1. Power and Status LEDs
2. Optical Output Connectors
3. ON/Off Button
4. USB Connectors
5. LAN Connector
6. VGA Connector
7. Power Connector
3.1.1 Type of connectors
The Reading unit has 1 or 4 E2000/APC Optical Output Connectors to
which can be plugged only E2000/APC connector. The E2000/APC
connectors and Mating are identified by the green color, in the Fig. 3.2
are shown an E2000/APC connector and an E2000/APC Mating
adapter:
Fig. 3.2
3.2 Setting Up
3.2.1 Power supply
To power supply the MuST VII Dynamic Reading Unit connects the
supplied power cable to 100 - 240 V power line to the measurement
Regulatory and Certification Considerations
12 Smartec Manual-2.0
unit Power Connector (7 on Fig. 3.1). Then switch the power switch to
its “I” position. The measurement unit powering off can be performed
by disconnecting the power supply or by switching the power switch to
its “O” position.
The measurement unit has internally: one acquisition module and a
PC. Whenever the power supply is interrupted (either by a power
shortage or a mechanical switch off of the power switch or the power
plug) the internal PC will start up and the acquisition module will
resume to its configuration at the time of power shortage when device
is powered on again.
Information
Check the status and power led (1 on Fig. 3.1) to ensure that the unit
is ready to measure.
Fig. 3.3
3.2.2 Optical connectors
The MuST VII Dynamic Measurement Unit can be purchased either
with 1 or 4 E2000/APC optical connectors. Number 2 on Fig. 3.1
exemplifies a 1 connector removable base.
Technical details
Manual-2.0
Smartec 13
Attention should be paid to the cleaning of the optical connectors. A
dirty connector can compromise the measurement and will degrade
the measurement unit performance. It is advisable to frequently clean
the connectors using appropriate tools (see Cleaning Procedure on
page 15).
3.2.3 Bottom plate
The bottom plate of the measurement unit is prepared for receiving
gland nuts ensuring the closet’s water tightness. Two PG21 and 24
PG11 can be used. It is supplied with covers on the holes and without
glands.
Fig. 3.4
3.3 Switching On
3.3.1 Internal PC
The internal PC will start every time the power is turned from off to on.
3.3.2 Acquisition Module
Pressing the “ON/OFF” button (3 on Fig. 3.1) will start the acquisition
module of the unit.
Regulatory and Certification Considerations
14 Smartec Manual-2.0
3.4 Switching OFF
3.4.1 Internal PC
The internal PC can be shut down either by software – using the Start
Menu in Windows – or when power is turned off.
3.4.2 Acquisition Module
To avoid accidental shutting-down of the acquisition module of the
measurement unit, it is necessary to press the “ON/OFF” button (2 on
Fig. 3.1) during 4 seconds to power-off the module.
Information
The “ON/OFF” button (2 on Fig. 3.1) will not shut down the PC.
3.5 Power and Status LEDs
Power and status LEDs are related to the acquisition module of the
measurement unit.
3.5.1 Power LED
Whenever the acquisition module of the measurement unit is running,
the power LED will be on. If the acquisition module of the
measurement unit is off, the power LED will be off.
LED Acquisition module
Green steady Running
OFF Not running
3.5.2 Status LED
The Status LED (1 on Fig. 3.1) will start blinking at 2Hz. After
approximately 30 seconds it will start blinking at 1Hz. This means that
the acquisition module of the measurement unit is already on and
responsive, but the optoelectronic module is still warming up. After
approximately one and a half minutes (90 s) it should stay on
Technical details
Manual-2.0
Smartec 15
permanently. This means that the measurement unit is able to
measure.
Important
If the acquisition module of the unit does not start correctly, the status
led will blink faster. If this happens please contact Smartec technical
support.
LED Acquisition module
Green blinking 2Hz Powering up
Green blinking 1Hz Warming up
Green steady Running
Green blinking 0.5Hz Running in error mode
OFF Not running
3.6 Interface
To operate the measurement unit a Monitor and a mouse and
keyboard should be connected to the VGA connector (6 on Fig. 3.1)
and to the USB ports (4 on Fig. 3.1).
The measurement unit operates on a Windows 7 Professional
environment.
Please refer to the software manual for further details.
3.7 Control
The MuST VII Dynamic Measurement Unit measurement unit can be
fully controlled using the DynamicMONITOR software. Please refer to
the DynamciMONITOR software manual for further details.
3.8 Cleaning Procedure
Proper performance of a fiber optic connection is strongly dependent
on the cleanliness of the mated ferrules. After repeated matings,
Regulatory and Certification Considerations
16 Smartec Manual-2.0
however, or when degraded performance is observed, it may become
necessary to clean the individual ferrules and mating sleeve. In this
section are outlined the proper cleaning and inspection procedures to
help ensure optimal connector performance.
When a Measurement Unit is repeatedly being plugged in and out with
optical connectors, it is very important that the connectors are cleaned
prior to any connection. If not, dust and moister can be deposited in
the measurement unit adaptor and this will compromise
measurements. On Fig. 3.5 a picture of a magnified connector is
presented. The dark gray circle corresponds to the fiber cladding and
the small light gray circle is the core of the fiber. One picture for a
clean connector and one picture for a dusty connector are presented.
Clean connector
Dirty connector
Fig. 3.5
The most common effect of dirt on the connections is that there is a
large amount of broad band light that is being reflected at the
connection, at both directions, meaning that the dynamic range for
measuring becomes smaller.
Technical details
Manual-2.0
Smartec 17
On the market there are several tool aimed to clean the connectors,
some of them are:
► Dust and shutter caps
► Isopropyl alcohol
► Cotton swabs
► Soft tissues
► Pipe cleaner
► Compressed air
Information
Before reporting the general procedure for cleaning connectors it is
important to highlight some reminders that shall always be kept into
account:
► Always turn off any laser sources before you inspect fiber
connectors
► Always inspect the connectors or adapters before you clean.
► Always inspect and clean the connectors before you make a
connection.
► Always use the connector housing to plug or unplug a fiber.
► Always keep a protective cap on unplugged fiber connectors.
► Always store unused protective caps in a sealable container in
order to prevent the possibility of the transfer of dust to the
fiber. Locate the containers near the connectors for easy
access.
Some warnings as well:
► Never use alcohol or wet cleaning without a way to ensure
that it does not leave residue on the end face. It can cause
damage to the equipment.
► Never look into a fiber while the system lasers are on.
► Never use unfiltered handheld magnifiers or focusing
optics to inspect fiber connectors.
► Never connect a fiber to a fiberscope while the system
lasers are on.
► Never touch the end face of the fiber connectors.
► Never twist or pull forcefully on the fiber cable.
Regulatory and Certification Considerations
18 Smartec Manual-2.0
► Never touch the clean area of a tissue, swab, or cleaning
fabric.
► Never touch any portion of a tissue or swab where alcohol
was applied.
► Never touch the dispensing tip of an alcohol bottle.
► Never use alcohol around an open flame or spark; alcohol
is very flammable.
The here below section describes the connector cleaning process.
► Inspect the fiber connector, component, or bulkhead with a
fiberscope with a proper E2000 adapter.
► If the connector is dirty, clean it with a dry cleaning technique.
► Inspect the connector.
► If the connector is still dirty, repeat the dry cleaning technique.
► Inspect the connector.
► If the connector is still dirty, clean it with a wet cleaning technique
followed immediately with a dry clean in order to ensure no residue
is left on the end-face.
Important
Wet cleaning is not recommended for bulkheads and receptacles.
Damage to equipment can occur.
► Inspect the connector again.
► If the contaminate still cannot be removed, repeat the cleaning
procedure until the endface is clean.
Here below dry and wet cleaning technique are described:
3.8.1 Dry cleaning technique
Necessary tool
Lint-free wipes, preferably clean room quality.
Technical details
Manual-2.0
Smartec 19
Fig. 3.6
Starting Pre-Caution
Read the reminders and warnings before you begin this process.
► Make sure that the lasers are turned off before you begin the
inspection.
► Remove the protective cap using the E2000 service adapter.
► Fold the wipe into a square about 4 to 8 layers thick, see Figure 8.
► Inspect the connector with a fiberscope.
If the connector is dirty, clean it with a lint-free wipe.
Be careful not to contaminate the cleaning area of the wipe with
your hands or on a surface during folding.
► Lightly wipe the ferrule tip in the central portion of the wipe with a
figure 8 motion.
Regulatory and Certification Considerations
20 Smartec Manual-2.0
Do not scrub the fiber against the wipe. If you do it, it can cause
scratches and more contamination.
► Repeat the figure 8 wiping action on another clean section of the
wipe.
► Properly dispose of the wipe.
► Inspect the connector again with the fiberscope.
► Repeat this process as necessary.
3.8.2 Wet cleaning technique
Necessary tool
Lint-free wipes, preferably clean room quality, and 99% isopropyl
alcohol.
► Make sure that the lasers are turned off before you begin the
inspection.
► Remove the protective cap using the E2000 service adapter.
► Inspect the connector with a fiberscope.
► Fold the wipe into a square, about 4 to 8 layers thick.
► Moisten one section of the wipe with one drop of 99% alcohol. Be
sure that a portion of the wipe remains dry.
► Lightly wipe the ferrule tip in the alcohol moistened portion of the
wipe with a figure 8 motion. Immediately repeat the figure 8 wiping
action on the dry section of wipe to remove any residual alcohol.
Do not scrub the fiber against the wipe, doing so can cause
scratches.
► Properly dispose of the wipe.
Technical details
Manual-2.0
Smartec 21
Important
Never reuse a wipe.
► Inspect the connector again with a fiberscope.
► Repeat the process as necessary.
To clean an optical measurement unit adapter, use an appropriate
cotton swab (there are several fiber clean swabs in the market
frequently used for telecom) embedded in isopropyl alcohol.
► Ensure the reading unit is disconnected and switched off
► put some isopropyl alcohol on cleaning stick
► Push the metallic protection inside the connector with the help of
the cleaning stick. You with then see a white plastic ring
► insert the stick inside the ring and gently clean the inner part of the
connector rotating the swab always on the same direction.
Regulatory and Certification Considerations
22 Smartec Manual-2.0
How to clean an FC connector of a reading unit
How to clean an E2000 connector of a reading unit
Fig. 3.7
3.9 Fuse Replacement
Electrical shorts may cause fuse failure and when that happens it is
necessary to replace the fuse located on the Power Connector (Fig.
3.8). To replace the fuses release the top and bottom springs of the
fuse’s support and then remove the piece (Fig. 3.8).
Technical details
Manual-2.0
Smartec 23
Realeasing the top and bottom springs
Fuses on the piece
Fig. 3.8
Fuse characteristics:
► Rated voltage: 250 Volt AC
► Interrupting rate: 2 Ampere
► Number of fuses: 2
Regulatory and Certification Considerations
24 Smartec Manual-2.0
4 DynamicMONITOR Software
The DynamicMONITOR graphical interface is divided into two different
areas (Fig. 4.1):
1. General Bar
2. Graphical Area
Fig. 4.1
4.1 General Bar
The General Bar, that is always active, is where the main acquisition
actions can be performed.
On the left, the current date and time are shown (number 1 on Fig.
4.2).
To establish connection between the measurement unit and the
software press the connect button (number 2 on Fig. 4.2). If the
software finds the device, the connect button changes to “disconnect”
and the start button (number 3 on Fig. 4.2) becomes active as shown
on Fig. 4.3.
Technical details
Manual-2.0
Smartec 25
Fig. 4.2
Fig. 4.3
If a measurement unit is not found in the previously used IP Address,
a window will pop up (Fig. 4.4) and a new IP Address has to be
defined (number 1 on Fig. 4.4), tested (number 2 on Fig. 4.4) and
confirmed (number 3 on Fig. 4.4) by the user.
Fig. 4.4
Information
The FS2x00 100S/s series measurement units default IP is 10.0.0.100
which is the IP address used on the first connection with the
DynamicMONITOR software.
Regulatory and Certification Considerations
26 Smartec Manual-2.0
4.1.1 Acquisition
Start Acquisition
Press the start button to start acquisition. This will start the acquisition
and representation of the measured values for all optical channels.
Important
Before start acquisition make sure that the network configuration on
«Configuration» tab is up to date. For more details refer to page 32.
Saving Data
To save data, select the file path to define the folder where data is to
be stored (see configuration section on page 32).
The naming of the file is automatic and corresponds to the timestamp
of the last measurement recorded.
Saved data file collects the data between the instant the save button is
pressed and the instant it is pressed back. Data files are divided in
accordance to a predefined maximum size. To define the maximum
size of the data files see «Configuration» section on page 32.
The first row of data files indicates the acquisition rate and the second
row indicates the columns meaning. The data file is organized in
columns as follows, see Fig. 4.5.
► 1st column – UTC Date: «DD-MM-YYYY»
► 2nd column – UTC Time: «HH:MM:SS.SSS»
► 3rd column – Sample: Sample number
► Following columns – Sensor Name (Formula): measured values
(formula computed) ordered by optical channel, then by
wavelength.
Technical details
Manual-2.0
Smartec 27
Fig. 4.5
General information on the software version can be called by pressing
the info button (number 5 on Fig. 4.2).
Exiting the application should be performed by pressing the exit button
(number 6 on Fig. 4.2).
Opening Data in Microsoft Excel
To import the data file to a Microsoft Office Excel Workbook proceed as follows:
Once Microsoft Excel is opened, press button “From Text” from menu Data>Get External Data (see Fig. 4.6) and select the file data file.
Fig. 4.6
Then it is necessary to complete text import wizard. All options must be set as shown on Fig. 4.7 to Fig. 4.11.
41 2 3
Regulatory and Certification Considerations
28 Smartec Manual-2.0
Fig. 4.7
Fig. 4.8
Technical details
Manual-2.0
Smartec 29
Fig. 4.9
Fig. 4.10
Fig. 4.11
Regulatory and Certification Considerations
30 Smartec Manual-2.0
Stop Acquisition
To stop acquisition press the start button again. It will turn from active
to the state showed in Fig. 4.3.
4.1.2 Exit Application
To exit DynamicMONITOR application, press the exit button (number
7 on Fig. 4.2).
4.2 Graphical Area
The graphical area is divided into six tabs:
► Graphical View
► Numerical View
► Configuration
► Spectral View
► FFT View
► SCPI Interface
4.2.1 Graphical View
On the Graphical View, graphical representation of the measured
values can be found over time (number 1 on Fig. 4.12).
Technical details
Manual-2.0
Smartec 31
Fig. 4.12
Using the drop-down arrow of the legend, as shown on number 2 on
Fig. 4.12, the user can select the sensors to represent and by
selecting the tick box enable or disable their representation.
Zoom In and Out
The user can perform zoom in by selecting an area on the displayed
graph and zoom out by pressing number 3 on Fig. 4.12. Alternatively,
scale numbers can be edited.
Clear Graph
All displayed data can be deleted by pressing clear button (number 4
on Fig. 4.12).
4.2.2 Numeric View
The Numeric View shows the acquired values (in accordance to the
configuration formula) for each optical channel, organized in columns
as in Fig. 4.13.
1
3 4
2
Regulatory and Certification Considerations
32 Smartec Manual-2.0
Fig. 4.13
4.2.3 Configuration
Acquisition configuration allows one to perform sensors configuration
for the acquisition of engineering values. To accomplish this step
correctly, sensors’ calibration sheets might be needed.
The DynamicMONITOR allows editing, saving and opening sensor
configurations. The configuration corresponds to the identification and
the definition of the calibration formula that converts measured
wavelengths into engineering values of the sensors.
The easiest way to start measuring and creating a configuration is to
perform an auto-scan (refer to section «AutoScan» on page 33).
Technical details
Manual-2.0
Smartec 33
Fig. 4.14
Clear configuration
If an old configuration is set, it is possible to clear it and start a new
configuration by pressing button New (number 8 on Fig. 4.14). Be
aware this operation will discard all previous unsaved configurations of
all channels.
AutoScan
When AutoScan is performed for the first time, the measurement unit
executes a measuring sweep and the DynamicMONITOR defines a
new configuration with all the found sensors. This configuration sets
the current measured WaveLength as Central WaveLength with a
default Range of 3 nm.
Confirm that gain and threshold values are correctly defined so
that no sensor is left out and that no noise is configured as a
sensor (see section «Measurement Configuration» on page 40).
Regulatory and Certification Considerations
34 Smartec Manual-2.0
Sensor Names are set by default as SensorXZZ, where X is the optical
channel and ZZ is the number of the sensor on that optical channel
ordered by wavelength.
Example:
Sensor315 corresponds to the 15th sensor of optical channel 3.
Add and Edit Configuration
In order to Add new sensors to an existing configuration, the user has
to repeat the AutoScan and choose the tick box “update only the new
or deleted sensors on the network” as shown on Fig. 4.15. If this
option is not selected all sensors configurations (including formulas)
will be lost and new AutoScan configurations are created.
Fig. 4.15
To configure a sensor individually select its cell and click on edit button
(number 11 on Fig. 4.14), hence a dialog box will pop up (Fig. 4.16).
Additional info about this dialogue box might be found at the next
table.
Technical details
Manual-2.0
Smartec 35
Fig. 4.16
Sensor Name Sensor’s identification field. It is not possible to have two sensors
with the same name.
Central Wavelength
(CWL)
Reference wavelength from which WaveLengthShift is calculated.
Range Safety bands defined to avoid crosstalk between sensors. Each
sensor has its own operation range. For instance, if we have a
CWL of 1522.659nm and a range spanning 3nm, only WaveLength
values between 1525.659 and 1519.659 are accepted.
WaveLength values out of the defined bands are returned as -998.
Inside each band only the highest peak is considered.
Formula Function defining the correlation between the WaveLengthShift (x)
and the engineering values.
x corresponds to the difference between the measured
WaveLength at each instant and the defined CentralWaveLength:
Here it is possible to set a formula manually, although it is also
possible to copy and paste from other locations. If the desired
output is WaveLengthShift, insert x on formula textbox.
Do Do not
-11.3 -11,3
-11.3*x -11.3x
-11.3*x^2+105.4*x+30 =-11.3*x^2+105.4*x+30
Regulatory and Certification Considerations
36 Smartec Manual-2.0
Test configuration
Test Configuration (number 3 on Fig. 4.14) allows the user to check
the current values from each sensor according to its configuration.
Information
Error code will be displayed as -998 when no sensor is found within
the defined range.
Save and Load configuration
Acquisition configuration can be saved in .txt extension file by pressing
the Save button (number 10 on Fig. 4.14) and choosing a save path as
shown on Fig. 4.17.
Fig. 4.17
In the configuration file the information is divided by optical channel
identified between brackets (number 1 on Fig. 4.18). For each sensor
there are 4 columns, namely: Sensor’s Name, Central WaveLength
(nm), Range (nm) and Formula separated by semicolons (as in
numbers 2, 3, 4 and 5 on Fig. 4.18).
Technical details
Manual-2.0
Smartec 37
Fig. 4.18
This file might be edited in Notepad or Excel (see “Opening Data in
Microsoft Excel” section on page 27) and saved as .txt extension file
after editing.
Configuration file can always be loaded by the user by pressing the
Load button (number 7 on Fig. 4.14) and pick the configuration file
(see Fig. 4.19).
Fig. 4.19
Regulatory and Certification Considerations
38 Smartec Manual-2.0
Changing Sample Rate
The sample rate can be altered using the dropdown box in number 5
from Fig. 4.14. Available rates are 1, 2, 5, 10, 20, 50 and 100 samples
per second.
Important
This sampling rate corresponds to averaged values of data acquired at
100S/s.
File Options
Measured data is recorded in the directory defined on the box number
7 on Fig. 4.14. To choose this directory the user can write on the box
or press the folder button next to the box, drive to the right directory
and then press “current folder” button.
Each saved file has a limited length defined by time. The maximum
time interval for each data file can be defined on number on Fig. 4.14.
When this interval is over, the data file created and named with the
timestamp of the last measurement recorded. The recording rate of
the saved data can be set on number 4 on Fig. 4.14.
Change IP Address
To change the IP Address of the measurement unit press the change
IP Address button (number 4 on Fig. 4.14) to pop-up the change IP
Address interface (see Fig. 4.20).
Fig. 4.20
Technical details
Manual-2.0
Smartec 39
Define the new IP Address (number 1 on Fig. 4.20). The Subnet Mask
(number 2 on Fig. 4.20) will be automatically defined according to the
IP Address. The Gateway is 0.0.0.0 by default. Then press the
Change IP button (number 1 on Fig. 4.20). After a few seconds a
validation message will appear (see Fig. 4.21).
When the IP is changed, a connection message appears as shown in
Fig. 4.21. At this point, the connection between the computer and the
equipment will be lost and the unit will reboot with the new IP Address.
This operation might take a few minutes.
The next time the Connect button (number 2 on Fig. 4.2) is pressed,
DynamicMONITOR will connect to the new IP.
Fig. 4.21
4.2.4 Spectral View
The spectral view tab is only accessible when data acquisition is
stopped.
This tab represents detected peaks with the defined gain and
threshold settings. Peaks are represented as pairs of wavelength and
power amplitude as a bar graph (number 1 on Fig. 4.22).
Regulatory and Certification Considerations
40 Smartec Manual-2.0
Fig. 4.22
The spectral view can only represent one channel at a time. Channel
selection can be performed on number 9 on Fig. 4.22.
Measurement Configuration
Measurement settings can influence sensor readings. Gain and
Threshold values should be defined accordingly to the FBG sensing
network status, so that peak amplitude is around 3000 and no FBG
side lobes are detected as peaks. These settings can be configured
independently for each optical channel.
Gain
Gain is directly related to the optical power used on the measurement
of the optical channel. The gain value should be raised to overcome
losses on the network. Gain is an integer value between 0 and 255.
The current gain value can be found using the get gain button (number
3 on Fig. 4.22). The gain value is updated every time the spectral view
tab is selected or a new channel is defined. To define a new gain
value, write it on the gain box (number 4 on Fig. 4.22) and press set
gain button (number 5 on Fig. 4.22).
Threshold
Threshold defines the line between usable signals and noise. It is an
integer value between 200 and 3200. Operations for getting and
defining threshold values can be performed on a similar way to the
gain, but using the threshold buttons (number 6, 7 and 8 on Fig. 4.22).
3 4 5 6 7 8
9
2
1
10
Technical details
Manual-2.0
Smartec 41
The threshold value is updated every time the spectral view tab or a
new channel are selected. This value is also represented graphically
as a horizontal line on the power bar graph.
4.2.5 FFT View
FFT View tab shows the fast Fourier transform graph of the acquired
measurements. This computation is based on LabView function
Spectral Measurements (FFT.VI), considering the measured values
exponential weighted (with no averaging), being the latest signals
more significant for the calculations.
Once the start button (number 3 on Fig. 4.2) is pressed all previous
measurements are erased from the FFT temporary memory.
FFT is shown for the active sensors on the sensors list, number 4 on
on Fig. 4.22 to enable/disable any sensor tick the sensor tick box.
Information
To choose sensors that are not present in the sensors list the user has
to edit the list on the Graphical View tab. For more details refer to
section «Graphical View» on page 30.
Fig. 4.23
4
1
2 3
Regulatory and Certification Considerations
42 Smartec Manual-2.0
4.2.6 SCPI Interface
This tab allows the user to send Standard Commands for
Programmable Instruments (SCPI) to the equipment. Please refer to
FS2x00 unit “User’s Manual” for a list of the commands available to
communicate with the measurement unit.
Commands are inserted on the command line (number 3 on Fig. 4.24)
and sent by pressing Enter or Send cmd button (number 2 on Fig.
4.24). The unit’s answers appear on the command board (number 1
on Fig. 4.24). Each time new answers are displayed, older ones are
pushed down.
Fig. 4.24
Each time a new command is sent, a shortcut at the scroll down box is
created, as shown at Fig. 4.25.
Fig. 4.25
1
2
3
Technical details
Manual-2.0
Smartec 43
When the user leaves SCPI Interface tab, unit status is changed
and Schedule acquisitions are disabled.
Any questions or comments regarding this guide please report to: [email protected]
M
uS
T V
II D
ynam
ic M
anual
v2.0
ww
w.h
bm
.co
m/f
s