The Analysis of Steels and Alloys using Laser Ablation Coupled with the Thermo Scientific iCAP 6500 ICP-OES Andrew Clavering, ICP Applications Specialist, Thermo Fisher Scientific, Cambridge, UK. Benefits in Brief Unique, enhanced coupling between the Thermo Scientific iCAP 6500 ICP-OES, the Thermo Scientific iTEVA Software and the New Wave Research UP 266 Macro Laser enables unattended analysis of multiple solid samples Minimal sample preparation for direct trace element analysis of metals and other solid materials Laser ablation enables the pinpoint analysis of sections of material, small samples or large areas of sample using user defined patterns of ablation Introduction The analysis of steel and its alloys is a common application in many laboratories around the world. A variety of techniques may be used for elemental analysis of metals and these may include sample dissolution techniques and solid analysis techniques, for example, Spark Optical Emission Spectroscopy (Spark OE) and X-Ray Fluorescence (XRF). For sample dissolution techniques, the samples are dissolved in a combination of acids by either microwave digestion or using a hot plate and then analyzed by Atomic Absorption Spectrometry (AAS), Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES) or Inductively Coupled Plasma Mass Spectrometry (ICP-MS) or a combination of one or more of these techniques. The analysis of solid steel or metals alloys typically employs a combination of XRF and Spark OES techniques. In addition, techniques such as Laser/Spark ablation can also be employed in conjunction with ICP to produce metal particles for direct introduction into the instrument. The technique of Laser Ablation (LA) combined with ICP-OES (LA-ICP-OES or LA-ICP) is becoming increasingly popular because it enables the analyst to analyze metals, refractory elements, insoluble alloys, glass and almost any solid material (including very small samples) without having to dissolve or destroy the sample. Using these combinations of instrumentation also enables the analyst to quickly switch from solid sampling to solution sampling allowing the analysis of both liquids and solids on the same instrument. When using a Laser Ablation system, the surface of the material is ablated to produce fine particles which are then passed into the torch of the ICP instrument using an argon or helium carrier gas. These particulates are then processed in the plasma similarly to aerosols (as derived from the nebulization of liquid samples). Detection limits are typically in the order of μg/g in the solid when using this approach and the total time for analysis is greatly reduced because the dissolution process is removed from the process. Instrumentation The Thermo Scientific iCAP 6500 ICP-OES, with the dedicated radial plasma, is used here for the analysis of a set of certified metal standards with a New Wave UP 266 Macro laser. The iCAP 6500 ICP-OES is a compact, robust instrument which contains all the core technologies of the Thermo Scientific iCAP 6000 Series ICP-OES such as an efficient, robust solid state generator; high resolution echelle spectrometer with excellent detection capability and a CID86 detector with improved dynamic range, non- blooming design and optimum signal-to-noise measurements. Additionally, the iCAP 6500 ICP-OES also has the capability for enhanced accessory interfacing to external accessories including triggering and control. The Thermo Scientific iTEVA Software, combined with the capabilities of the UP 266 Macro Laser, enables the analyst to use the laser as an autosampler which means unattended analysis can be performed for as many samples as will fit in to the laser’s ablation chamber. The iCAP 6500 ICP-OES is connected to the New Wave UP266 Macro Laser Ablation system via a bespoke communications link (see Figure 1) and is also connected to the main instrument PC. The communication link allows the unique triggering and parameter control of the UP266 Laser Ablation system via the iTEVA Software in combination with the New Wave MEO Software. The iCAP 6500 ICP-OES also incorporates enhanced speed of data acquisition which enables the time for the ablation process to be reduced if required, saving valuable time, gas and sample. Key Words • Laser ablation • Steel • Trace elemental analysis • ICP Application Note: 43063

Note: 43063 Ablation Coupled with the Thermo Scientific ....pdfiCAP 6500 ICP-OES, the Thermo Scientific iTEVA Software and the New Wave Research UP 266 Macro Laser enables unatten

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Note: 43063 Ablation Coupled with the Thermo Scientific ....pdfiCAP 6500 ICP-OES, the Thermo Scientific iTEVA Software and the New Wave Research UP 266 Macro Laser enables unatten

The Analysis of Steels and Alloys using LaserAblation Coupled with the Thermo ScientificiCAP 6500 ICP-OESAndrew Clavering, ICP Applications Specialist, Thermo Fisher Scientific, Cambridge, UK.

Benefits in Brief• Unique, enhanced coupling between the Thermo ScientificiCAP 6500 ICP-OES, the Thermo Scientific iTEVASoftware and the New Wave Research UP 266 MacroLaser enables unattended analysis of multiple solidsamples

• Minimal sample preparation for direct trace elementanalysis of metals and other solid materials

• Laser ablation enables the pinpoint analysis of sections ofmaterial, small samples or large areas of sample using userdefined patterns of ablation

IntroductionThe analysis of steel and its alloys is a common applicationin many laboratories around the world. A variety oftechniques may be used for elemental analysis of metals andthese may include sample dissolution techniques and solidanalysis techniques, for example, Spark Optical EmissionSpectroscopy (Spark OE) and X-Ray Fluorescence (XRF).For sample dissolution techniques, the samples are dissolvedin a combination of acids by either microwave digestion orusing a hot plate and then analyzed by Atomic AbsorptionSpectrometry (AAS), Inductively Coupled Plasma OpticalEmission Spectrometry (ICP-OES) or Inductively CoupledPlasma Mass Spectrometry (ICP-MS) or a combination ofone or more of these techniques.

The analysis of solid steel or metals alloys typicallyemploys a combination of XRF and Spark OES techniques.In addition, techniques such as Laser/Spark ablation canalso be employed in conjunction with ICP to produce metalparticles for direct introduction into the instrument.

The technique of Laser Ablation (LA) combined withICP-OES (LA-ICP-OES or LA-ICP) is becoming increasinglypopular because it enables the analyst to analyze metals,refractory elements, insoluble alloys, glass and almost anysolid material (including very small samples) without havingto dissolve or destroy the sample. Using these combinationsof instrumentation also enables the analyst to quickly switchfrom solid sampling to solution sampling allowing theanalysis of both liquids and solids on the same instrument.

When using a Laser Ablation system, the surface of thematerial is ablated to produce fine particles which are thenpassed into the torch of the ICP instrument using an argonor helium carrier gas. These particulates are then processedin the plasma similarly to aerosols (as derived from thenebulization of liquid samples). Detection limits aretypically in the order of µg/g in the solid when using thisapproach and the total time for analysis is greatly reducedbecause the dissolution process is removed from the process.

InstrumentationThe Thermo Scientific iCAP 6500 ICP-OES, with thededicated radial plasma, is used here for the analysis of a setof certified metal standards with a New Wave UP 266Macro laser. The iCAP 6500 ICP-OES is a compact, robustinstrument which contains all the core technologies of theThermo Scientific iCAP 6000 Series ICP-OES such as anefficient, robust solid state generator; high resolution echellespectrometer with excellent detection capability and aCID86 detector with improved dynamic range, non-blooming design and optimum signal-to-noisemeasurements. Additionally, the iCAP 6500 ICP-OES alsohas the capability for enhanced accessory interfacing toexternal accessories including triggering and control. TheThermo Scientific iTEVA Software, combined with thecapabilities of the UP 266 Macro Laser, enables the analystto use the laser as an autosampler which means unattendedanalysis can be performed for as many samples as will fit into the laser’s ablation chamber.

The iCAP 6500 ICP-OES is connected to the New WaveUP266 Macro Laser Ablation system via a bespokecommunications link (see Figure 1) and is also connected tothe main instrument PC. The communication link allowsthe unique triggering and parameter control of the UP266Laser Ablation system via the iTEVA Software incombination with the New Wave MEO Software. TheiCAP 6500 ICP-OES also incorporates enhanced speed ofdata acquisition which enables the time for the ablationprocess to be reduced if required, saving valuable time, gasand sample.

Key Words

• Laser ablation

• Steel

• Trace elementalanalysis

• ICP

ApplicationNote: 43063

Page 2: Note: 43063 Ablation Coupled with the Thermo Scientific ....pdfiCAP 6500 ICP-OES, the Thermo Scientific iTEVA Software and the New Wave Research UP 266 Macro Laser enables unatten

Figure 1: Gas and Communication Connections for the iCAP 6500 ICP-OESand Laser Unit.

The ablation gas line may be connected directly intothe base of the torch of the iCAP 6500 ICP-OES. In thisconfiguration, the spray chamber and nebulizer of thestandard sample introduction system is bypassed and theablated sample material is passed directly into the plasmausing an argon carrier gas. Helium carrier gas, used withmuch the same flow rate as argon, can also be used andcan enhance the laser ablation and sample transportefficiency to the plasma to provide further improvedsensitivity as required. Further details on the connectionsbetween iCAP 6000 Series ICP-OES, PC and laser unitsare provided in the User Guide provided with the LaserAblation unit1.

Sample and Standard PreparationThe certified reference standards are chosen from a seriesmanufactured by the National Institute for Science andTechnology (NIST) to show typical concentrations ofimpurities and minor elements in steel. The standards aresolid low-alloy steel discs obtained from MBH AnalyticalLtd (UK) and are specifically designed to be standards forsolid sampling analysis of steels (see Table 1), whether bythe two ablation techniques (laser and spark) or sparkoptical emission spectrometry analysis.

Element NIST 1761 NIST 1763 NIST 1765 NIST 1767

Al 0.055 0.043 - 0.004

Co - 0.095 0.0012 0.005

Cr 0.22 0.5 0.051 0.0015

Cu 0.3 0.043 0.0013 0.0014

Mn 0.678 1.58 0.144 0.022

Mo 0.103 0.5 0.005 0.02

Ni 1.99 0.51 0.154 0.002

Si 0.18 0.63 0.026

Ti 0.18 0.31 0.0055 0.006

V 0.053 0.3 0.004 0.033

Table 1: NIST Calibration Standards used for the Analysis– All Concentrations in %

Samples or standards are prepared (i.e. if the surface hasbeen previously ablated or altered in some way) byskimming the surface using a lathe or by rubbing withabrasive material to smooth and clean the sample surface.The surface should ideally be mounted perpendicular to thelaser beam within the ablation chamber to get the maximumbenefit of the ablation energies as well as reproducibility.

Method DevelopmentMethod development with this hyphenated techniquerequires the analyst to optimize the laser parameters for thesample type being analyzed in addition to the plasma sourcefor the ICP. In this case, the laser program utilized a pre-ablation step to ensure that any surface contamination wasremoved before the analytical ablation is performed. Thelaser parameters are user defined and are required to beoptimized for both the sample type and the requiredsensitivity.

The laser parameters that are user defined were set asfollows:

Parameter Setting

Pre-ablation rate 100µm/sec

Laser spot size (diameter) 750µm

Laser output 90%

Laser firing rate 10Hz

Laser step rate 30µm/sec

Carrier gas flow (Ar) 0.5L/min

Table 2: New Wave Macro UP266 Laser Parameters

A “raster” ablation pattern is used here as this isconsidered to be the most appropriate for bulk analysisapplications since it can cover a large area of the materialbeing analyzed. However, multiple pattern types may beused on more than one sample and the sample patterns willbe ablated in the order that they are inserted into the workqueue. Figure 2 shows an example of a combination ofpatterns in readiness for analysis. These patterns could alsobe used on multiple samples with the change of samplescontrolled by the unique triggering options between theiTEVA Software (see Figure 3) and the New Wave MEOSoftware used to control the New Wave UP266 MacroLaser. These triggering options allow the analysis of multiplesamples, each with its own gas flush, pre-ablation andablation steps controlled by iTEVA Software.

Figure 2: New Wave MEO Laser Software showing some types of ablationpatterns and options – from top left, clockwise: spot, line and rasterablations

Page 3: Note: 43063 Ablation Coupled with the Thermo Scientific ....pdfiCAP 6500 ICP-OES, the Thermo Scientific iTEVA Software and the New Wave Research UP 266 Macro Laser enables unatten

Figure 3: iTEVA Software - some of the unique triggering options

The iTEVA Software has an “Optimize Source” featurewhich can be used to optimize the plasma conditions tomaximize the analyte signal intensities. A homogenous solidsample is ablated using a large pattern and the “OptimizeSource” feature is then activated during the sample ablationto enable optimization using the analyte signal intensities asrequired. The plasma parameters established for theapplication are shown in Table 3 for reference:

Parameter Setting

Pump Tubing NA

Pump Rate NA

Nebulizer NA

Nebulizer gas flow NA

Spray Chamber NA

Auxiliary gas flow 1.0 L/min

Coolant gas flow 12 L/min

Center tube 2 mm

RF Power 1150 W

Radial Viewing Height 12mm

High/Low Integration time 20s UV/20s Vis

Table 3. Sample Introduction and Plasma Parameters* NA= Not Applicable

After optimization, the Fullframe data acquisitionmode was used to identify all elements in the samples andlook for possible interferences. A Fullframe is a display ofthe full spectrum in a two dimensional array showing allthe elements present in the sample. It is an excellentmethod development tool which can be stored indefinitelyand can also be used for real time and retrospective semi-quantitative analysis as required.

After the selection of several wavelengths for eachelement, some weak iron lines were selected as internalstandard lines. Using the iron as an internal standard inthis way compensates for any fluctuation in the transportor ablation efficiencies. Most analyses showed % RSDs tobe <5%. The iCAP 6500 ICP-OES was calibrated usingthe four NIST standards shown in Table 1. Twoadditional NIST steel standards (NIST 1762 and NIST1764) were then analyzed as unknowns against thesecalibrations.

Results All calibration curves used showed correlation coefficientsbetter than 0.995. The analytical results of the LA-ICPanalysis of the NIST standards are shown below in Table3. The analyzed values are compared against the acceptedvalues as determined by various methods in a variety oflaboratories guided by NIST.

Element NIST 1762 % NIST 1762 % NIST 1764 % NIST 1764 % Method DL Analysed Accepted Analysed Accepted (%)

Al 308.215 nm 0.057 0.069 0.008 0.009 0.003

Co 228.616 nm 0.063 0.062 0.0089

Cr 267.716 nm 0.83 0.92 1.55 1.48 0.0012

Cu 327.396 nm 0.11 0.12 0.48 0.51 0.0058

Mn 279.482 nm 1.89 2.00 1.15 1.21 0.008

Mo 202.030 nm 0.37 0.35 0.192 0.200 0.008

Ni 231.604 nm 1.19 1.15 0.223 0.202 0.004

Si 212.412 nm 0.29 0.35 0.056 0.057 0.002

Ti 337.280 nm 0.102 0.095 0.022 0.028 0.0007

V 292.402 nm 0.173 0.200 0.116 0.106 0.007

Table 4: Comparison of analyzed and accepted values in NISTreference materials

The method detection limits shown were estimatedusing the built-in iTEVA Software facility that utilizestailored signal to background calculations to estimatemethod detection limits. The facility is instantly availablefor all methods which have been calibrated and producesreliable method detection limits that approximate themore traditional 3σ approach.

Figure 4. Comparison of found and expected values of the NIST standards

Page 4: Note: 43063 Ablation Coupled with the Thermo Scientific ....pdfiCAP 6500 ICP-OES, the Thermo Scientific iTEVA Software and the New Wave Research UP 266 Macro Laser enables unatten

The majority of the comparison results show goodcorrelation between found and expected values. Almost allof the results fall well within a ±20% range and mostvalues were within a ±10% range of the expected values(see Figure 5). The exception is one of the titanium resultswhich is one of the lower concentration elementsanalyzed.

Figure 5: % Recovery of NIST standards

ConclusionThe analysis of steels by laser ablation is rapid, powerfuland flexible when using the Thermo Scientific iCAP 6500ICP-OES. The unique triggering options between theThermo Scientific iTEVA Software and the New WaveMEO Laser Software make the combination of theplatforms especially powerful with regard to theunattended analysis of multiple samples. The user cananalyze as many samples as the laser compartment willaccommodate and the autosampler options in the iTEVASoftware application can be used to control the LaserAblation accessory effectively.

References1 New Wave Research Laser documentation: can be requested at:

http://www.esi.com/Products/NewWaveResearch/LaserAblation/UPMACRO.aspx

Part of Thermo Fisher Scientific

Thermo Electron ManufacturingLtd (Cambridge) is ISO Certified.

AN43063_E 08/10C

www.thermoscientific.com©2010 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

In addition to these

offices, Thermo Fisher

Scientific maintains

a network of represen -

tative organizations

throughout the world.

Africa-Other+27 11 570 1840Australia+61 3 9757 4300Austria+43 1 333 50 34 0Belgium+32 53 73 42 41Canada+1 800 530 8447China+86 10 8419 3588Denmark+45 70 23 62 60 Europe-Other+43 1 333 50 34 0Finland /Norway /Sweden+46 8 556 468 00France+33 1 60 92 48 00Germany+49 6103 408 1014India+91 22 6742 9434Italy+39 02 950 591Japan +81 45 453 9100Latin America+1 561 688 8700Middle East+43 1 333 50 34 0Netherlands+31 76 579 55 55New Zealand+64 9 980 6700South Africa+27 11 570 1840Spain+34 914 845 965Switzerland+41 61 716 77 00UK+44 1442 233555USA+1 800 532 4752