Operating Instruction MA235 1

Embed Size (px)

Citation preview

  • 8/16/2019 Operating Instruction MA235 1

    1/20

    51709598 a11/96   1

    Instruction Manual 

    This manual contains complete instructions for setting up and using

    the MA235 ion analyzer. Applications information is also available.

    The information contained in this manual was correct at the time of going to print. However, we continue to improve products and reserve

    the right to change specifications, equipment and maintenanceprocedures at any time.

    This manual is copyrighted, and all rights are reserved. No part of 

    this manual may be reproduced by any means or in any form withoutprior consent in writing.

    The power supply unit is classed as IEC Class II equipment (equipment

    providing an adequate degree of protection against electric shocks, inwhich additional safety precautions, for example, double or reinforced

    insulation, are included). The MA235 is intended for use by personsknowledgeable in safe laboratory practices. If the MA235 is not usedin accordance with these instructions for use, the protection provided

    by the equipment may be impaired.

    The MA235 is suitable for direct current only.

    This equipment generates, uses and can radiate radio frequencyenergy and if not installed and used in accordance with the instruc-tion manual, may cause interference with radio communications. Ithas been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC Rules,

    which are designed to provide reasonable protection against suchinterference when operated in a commercial environment. Operationof this equipment in a residential area is likely to cause interferencein which case the user at his own expense will be required to takewhatever measures may be required to correct the interference.

    There are no user replaceable parts in the MA235 or power supplyunit. Do not remove the covers.

    FS 981/SM 1746

  • 8/16/2019 Operating Instruction MA235 1

    2/20

    2 51709598 a11/96

    Contents 

    1 Set Up  

    1.1 Unpacking and Installation ................................................ 5 1.2 Input and Output Connections ........................................... 6 1.3 Controls ............................................................................. 7 

    2 Operation  

    2.1 pH Measurements ............................................................. 8 2.1.1 Calibrating a pH Electrode ................................................ 8 

    2.2 mV Measurements .......................................................... 10 2.2.1 Relative mV Measurements ............................................ 10 

    2.3 Concentration Measurements ......................................... 112.3.1 Calibrating an Ion Selective Electrode ............................ 11

    2.4 Known Addition/Subtraction Measurements ................... 13 

    2.5 Sample Addition/Subtraction Measurements .................. 14 

    2.6 Temperature Measurements ........................................... 15 

    2.7 Manual and Auto Endpoint, and Endpoint Time ............. 15 2.8 Sample ID ........................................................................ 16 

    2.9 Program Menu ................................................................. 17 

    2.10 Calibration Programming................................................. 19 2.10.1 Calibration Point Setting .................................................. 19 2.10.2 Last Calibration Recall .................................................... 20 2.10.3 mV Offset Setting ............................................................ 212.10.4 Isopotential Point Setting (pH) ........................................ 21

    2.11 Display Menu ................................................................... 22 2.11.1 Measurement Channel Selection .................................... 22 2.11.2 Units for Measurement Selection .................................... 22 2.11.3 Display Resolution for pH Selection ................................ 23 

    2.12 System 1 Menu ................................................................ 26 2.12.1 Time and Date Setting ..................................................... 24 2.12.2 Language Selection ......................................................... 24 2.12.3 Baud Rate Selection ........................................................ 25 

    2.12.4 Key Press Beeper Setting ............................................... 25 

    51709598 b6/97

  • 8/16/2019 Operating Instruction MA235 1

    3/20

    51709598 a11/96   3

    Contents (cont) 

    2 Operation (cont)  

    2.13 System 2 Menu ................................................................ 26 2.13.1 Temperature Mode Selection .......................................... 26 2.13.2 Manual Temperature Setting ........................................... 26 

    2.13.3 Endpoint Time Setting ..................................................... 27 2.13.4 Meter Self Test ................................................................ 27 

    2.14 Limits Menu ..................................................................... 29 

    2.14.1 Alarm Option Selection .................................................... 29 2.14.2 High/Low Limits for pH .................................................... 30 

    2.14.3 High/Low Limits for Temperature .................................... 30 2.14.4 High/Low Limits for pH Electrode Slope.......................... 31

    2.15 Datalog Menu .................................................................. 32 2.15.1 Datalog Selection ............................................................ 32 2.15.2 Interval Timer Setting ...................................................... 32 2.15.3 Recalling Data ................................................................. 33 2.15.4 Clearing Data ................................................................... 34 

    2.16 GLP Menu........................................................................ 35 

    2.16.1 GLP Option Selection ...................................................... 35 2.16.2 Cal Alarm Setting............................................................. 35 2.16.3 Entering ID Number ......................................................... 36 

    2.17 Method Menu ................................................................... 37 2.17.1 Method Saving ................................................................. 37 2.17.2 Method Loading ............................................................... 38 2.17.3 Resetting the MA235 ....................................................... 38 

    2.18 ISE Menu ......................................................................... 39 

    2.18.1 Electrode ID Selection ..................................................... 39 2.18.2 Incremental Technique Selection .................................... 39 2.18.3 Known Values for Incremental Techniques..................... 40 2.18.4 Sample Volumes for Incremental Techniques ................ 41

    2.19 Chart Recorder ................................................................ 42 

    2.20 Interfacing via the RS232 Output .................................... 43 

    2.21 Using an Electrode Selector Box .................................... 46 

  • 8/16/2019 Operating Instruction MA235 1

    4/20

    4 51709598 a11/96

    Contents (cont) 

    3 Support Information  

    3.1 Basic Theory .................................................................... 47 

    3.2 Operating Hints ................................................................ 48 

    3.3 Maintenance .................................................................... 50 3.3.1 MA235 Maintenance........................................................ 50 3.3.2 Electrode Maintenance .................................................... 50 

    3.4 Problem Solving .............................................................. 51

    3.5 Specifications .................................................................. 54 

    Consumables and Accessories ...................................................... 56 

  • 8/16/2019 Operating Instruction MA235 1

    5/20

    51709598 a11/96   5

    1.1 Installation 

    1. Unpack the meter, power supply, electrode, electrode arm,

    buffer sachets, Guide Book and calibration certificate. Store

    the calibration certificate in a safe place.

    2. Check the power supply unit is the correct voltage for your

    local supply. If it is not, contact your distributor.

    3. Attach the electrode arm to the left or righthand side of the

    meter:

    a. Turn the meter upside down and seat the dome of the

    electrode of the electrode arm base into the left or

    righthand meter recess.

    b. Insert and tighten the fixing screw.

    c. Turn the meter and base the right way up and locate the

    electrode arm onto the base fixing post. Press down to

    snap into place.

    1

  • 8/16/2019 Operating Instruction MA235 1

    6/20

    6 51709598 a11/96

    1.2 Input and Output Connections 

    1. Disconnect the shorting clips from the pH sockets and retainby clipping over the socket. If you are using two sensing

    electrodes connect them to Channel A and B. If you are using

    one electrode, connect it to Channel A.

    2. Connect the reference electrode(s), if using, to the Refsocket(s).

    3. If you are using an ATC probe connect it to theATC

     socket.

    If you are using a PT1000 probe connect it to the PT sockets.

    4. Connect the power supply unit to the DC socket.

    5. When the MA235 is powered up it carries out a display self test.

    1power supplyunit socket

    reference electrode sockets,Channel A & B

    ATC (automatic temperaturecompensation) probe socket

    combination or pH electrodesockets, Channel A & B

    PT1000 (automatictemperature compensation)probe sockets

    data output socket,see Section 2.20for interfacing details

    recorder outputsocket, seeSection 2.19for details

    51709598 b6/97

  • 8/16/2019 Operating Instruction MA235 1

    7/20

    51709598 a11/96   7

    1.3 Controls 

    Selects pH, mV, relative mV, concentration, knownaddition or sample addition mode.

     Accesses software menu from current mode. Exitssoftwaremenu one step at a time.

    Selects automatic endpoint, timed endpoint or manualendpoint facility.

    Turns the display off, and places the meter in standbymode. Wakes the meter up.

    Starts a calibration sequence in pH, concentration andrelative mV modes. Starts sequence in known additionand sample addition modes.

    Starts sample measurement in all modes. Press againto manually endpoint. Starts and stops timed datalogging. Returns to measurement from softwaremenu.

    Enters sample ID mode.

    Selects options in software menu, and enters data orselection.

    Used to move around software menu.

    Data entry keys, to set calibration buffers, date/time,high/low limits etc. Can also be used to move aroundsoftware menu quickly.

    1

    7 8 9

    4 5 6

    1 2 3

    0 . -

       >>   >>   >          >          >          >          >          >

     > > > > >>  >  >  >  >  

    Enter

    ID

     A 

    Prog

    Mode

    Read

    OnOff

    Cal

    51709598 b6/97

  • 8/16/2019 Operating Instruction MA235 1

    8/20

    8 51709598 a11/96

    2.1 pH Measurements 

     We recommend regular calibration before sample measurement

    (see Section 2.1.1).

    To measure the pH of a sample:

    Place electrode in sample and press or start the

    measurement.

    The display shows the pH of the sample. If you are using a

    temperature probe the sample temperature is shown, or you can

    enter the temperature manually (see Section 2.13.2).

    If you are using auto endpoint or timed endpoint the display

    freezes automatically (see Section 2.7). To manually endpoint

    and freeze the display press

    To start a new measurement press

    2.1.1 Calibrating a pH Electrode 

    NOTES If you are using a temperature probe the buffertemperature is shown, or you can enter the temperature

    of the buffer manually (see Section 2.13.2).

    If you are using auto endpoint or timed endpoint the

    display freezes automatically (see Section 2.7). To

    manually endpoint and freeze the display press

    To calibrate an electrode:

    Press If you are using both measurement channels

    (A and B), select the one you want to calibrate

     using and thenpress>  

              >

    CalpH 

    Read

    Enter

    ReadpH 

    Read

    Read

    Read

    Cal

    2

  • 8/16/2019 Operating Instruction MA235 1

    9/20

    51709598 a11/96   9

    2.1.1 Calibrating a pH Electrode (cont) 

    The calibration buffers are shown on the display.

    If you want to use other buffers, you will need to enter the values

    (see Section 2.10.1).

    1 point calibration

    Place the electrode in the first buffer. At the measurement

    endpoint a tick (4 ) appears on the display.

    If you are using a standard buffer the display updates to the

    temperature corrected value (see Section 3.5).

    The electrode offset is displayed.

    To return to sample measurement press

    2 point calibration

    To continue with a 2 point calibration:

    Place the electrode in the second buffer and press

     When the display freezes the electrode slope value will bedisplayed.

    To return to sample measurement press

    3, 4 and 5 point calibration

    Repeat procedure using the third, fourth and fifth calibration

    buffer.

    NOTE The buffers can be selected for calibration in any order.

    Use and to select the buffer you want to use

    for the calibration and press

    2

    Read

    Cal

    Read

    Cal

              >  >  

  • 8/16/2019 Operating Instruction MA235 1

    10/20

    10 51709598 a11/96

    2

    2.2 mV Measurements 

    To measure the absolute mV of a sample:

    Place electrode in sample and press to start the measurement.

    The display shows the absolute mV of the sample.

    If you are using auto endpoint or timed endpoint the display freezes

    automatically (see Section 2.7). To manually endpoint and freeze the

    display press

    To start a new measurement press

    2.2.1 Relative mV Measurements 

    Place electrode in the reference sample and press

    The display shows the mV of the reference sample.

    If you are using auto endpoint or timed endpoint the display

    freezes automatically (see Section 2.7). To manually endpoint and

    freeze the display press

    The display updates to 0.0, or an entered value (see Section 2.10.3).

    To measure the relative mV of a sample:

    Place electrode in sample and press to start the measurement.

    The display shows the relative mV of the sample. If you are using

    auto endpoint or timed endpoint the display freezes automatically

    (see Section 2.7). To manually endpoint and freeze the display

    press

    If you entered a mV offset (see Section 2.10.3) the sample result

    will be the difference between the measured and the set value.

    To start a new measurement press

    Read

    Read

    Read

    Cal

    Read

    CalRel mV Mode Read

    Read

    Read

    mV   ReadMode

    Read

  • 8/16/2019 Operating Instruction MA235 1

    11/20

    51709598 a11/96   11

    22.3.1 Calibrating an Ion Selective Electrode NOTES If you are using a temperature probe the temperature of 

    the standard is shown, or you can enter the temperature

    of the standard manually (see Section 2.13.2).

    If you are using auto endpoint or timed endpoint the

    display freezes automatically (see Section 2.7). To

    manually endpoint and freeze the display press

    To calibrate an electrode:

    Press If you are using both measurement channels

    (A and B), select the one you want to calibrate

    using and thenpress

    2.3 Concentration Measurements 

     We recommend regular calibration before sample measurement(see Section 2.3.1).

    To measure the concentration of a sample:

    Place electrode in sample and press to start the

    measurement.

    The display shows the concentration of the sample (you can

    change the units of measurement - see Section 2.11.2). If you areusing a temperature probe the sample temperature is shown or

    you can enter the temperature manually (see Section 2.13.2).

    If you are using auto endpoint or timed endpoint the display

    freezes automatically (see Section 2.7). To manually endpoint

    and freeze the display press

    To start a new measurement press

    concentration 

    Cal

    CalMode

    Read

    Read

    Read

              >>  

      Enter

    concentration Mode   Read

    Read

  • 8/16/2019 Operating Instruction MA235 1

    12/20

    12 51709598 a11/96

    2

    2.3.1 Calibrating an Ion Selective Electrode (cont) 

    Five calibration standards are shown on the display. If you want

    to use other standards, you will need to enter the values (see

    Section 2.10.1).

    1 point calibration

    Place the electrode in the first standard. At the measurement

    endpoint a tick (4 ) appears.

    The electrode offset is displayed.

    To return to sample measurement press

    2 point calibration

    To continue with a 2 point calibration:

    Place the electrode in the second standard and press

     When the display freezes the electrode slope value will be

    displayed.

    To return to sample measurement press

    3, 4 and 5 point calibration

    Repeat the procedure using the third, fourth and fifth calibration

    standard.

    NOTES The standard set for the first calibration standard is theisopotential point, calibrate in this standard first.

    The five standards can be selected for calibration in anyorder.

    Use and to select the standard you want

    to use for the calibration and

    press

    Read

    Read

    Cal

              >  >  

    Cal

  • 8/16/2019 Operating Instruction MA235 1

    13/20

    51709598 a11/96   13

    2.4 Known Addition/Subtraction Measurements 

    Enter electrode type and electrode slope, volumes and standardconcentration before measurement (see Section 2.18).

    To measure a sample by known addition/subtraction:

    Place electrode in sample and press to start the

    measurement.

    The display shows the first part of the measurement (you can

    change the units of measurement - see Section 2.11.2). If you areusing a temperature probe the sample temperature is shown or

    you can enter the temperature manually (see Section 2.13.2).

    If you are using auto endpoint or timed endpoint the display

    freezes automatically (see Section 2.7). To manually endpoint

    and freeze the display press

     Add the standard to the sample and press

    The sample result is shown. If you are using auto endpoint or

    timed endpoint the display freezes automatically (see Section 2.7).

    To manually endpoint and freeze the display press

    NOTE known values (Section 2.18.3) - enter the volume andconcentration of the standard you are using, and the

    slope of the electrode.

    sample volumes (Section 2.18.4) - enter the volume of your sample as sample volume 1.

    Read

    Read

    Cal

    Cal   Readk. add Mode

    2Read

  • 8/16/2019 Operating Instruction MA235 1

    14/20

    14 51709598 a11/96

    2

    2.5 Sample Addition/Subtraction Measurements 

    Enter electrode type and electrode slope, volumes and standardconcentration before measurement (see Section 2.18).

    To measure a sample by sample addition/subtraction:

    Place electrode in standard and press to start the

    measurement.

    The display shows the first part of the measurement (you can

    change the units of measurement - see Section 2.11.2). If you

    are using a temperature probe the temperature of the standardis shown or you can enter the temperature manually (see

    Section 2.13.2).

    If you are using auto endpoint or timed endpoint the display

    freezes automatically (see Section 2.7). To manually endpoint

    and freeze the display press

     Add the sample to the standard and press

    The sample result is shown. If you are using auto endpoint or

    timed endpoint the display freezes automatically (see Section 2.7).

    To manually endpoint and freeze the display press

     You can enter volumes for 4 samples (see Section 2.18.4), so it is

    possible to measure 4 samples using the same standard. You can

    measure more than 4 samples (the MA235 assumes the samplevolume is that entered for sample volume 4) but for maximum

    accuracy we recommend that you only measure 4 samples for

    each standard.

    To start a new measurement add the next sample to the

    standard (and previous sample) and press

    The MA235 uses the known values, plus the calculated samplevalue to calculate a new ‘standard value’ and therefore the new

    sample measurement.

    Cal

    Cal   Reads. add Mode

    Read

    Read

    Read

    Read

  • 8/16/2019 Operating Instruction MA235 1

    15/20

    51709598 a11/96   15

    2

    2.5 Sample Addition/Subtraction (cont) 

    NOTE known values (Section 2.18.3) - enter the volume andconcentration of the standard you are using, and the

    slope of the electrode.

    sample volumes (Section 2.18.4) - enter the volume of your samples (up to 4).

    2.6 Temperature Measurements 

    To measure the temperature of samples and calibration buffers

    you must have a temperature probe connected (see Section 1.2).

    Select which temperature mode you want to use for each channel

    (see Section 2.13.1). If you are using a temperature probe place

    the probe in the sample with the electrode when you calibrate

    and measure in any mode. The temperature is displayed with the

    reading, and the display shows ATC or PT1000.

    If you do not have a temperature probe connected or if you

    selected manual in temp mode, the display shows MTC. The

    temperature displayed will either be the default value of 25°C, ora temperature entered manually, see Section 2.13.2.

    2.7 Manual and Auto Endpoint, and Endpoint Time 

    The auto read control can be used in all modes toselect the endpoint method - manual, automatic or timed.

    The automatic endpoint facility checks each measurement to see

    if the electrode output is stabilizing, and when the stability

    requirements have been met automatically endpoints the

    reading. For solutions with long equilibration times (e.g. low

    ionic strength samples) it is better to use manual or timed

    endpoint.

    The endpoint time facility freezes the display after a presetamount of time (see Section 2.13.3). The endpoint time facility is

    turned off by setting the endpoint time to 00:00:00.

     A 

  • 8/16/2019 Operating Instruction MA235 1

    16/20

    16 51709598 a11/96

    2.8 Sample ID 

    The sample number counter increments 1 every time a sample

    measurement is endpointed.

    To change or reset the sample ID number press

    Enter your required sample number using the keypad (enter 0 if 

    you want to reset the number) and press again.

    NOTE If you press while measuring, the sample numberis assigned to that sample.

    If you press when the measurement hasendpointed, the number will be

    assigned to the next sample.

    A

    2.7 Auto Endpoint and Endpoint Time (cont) 

     Auto endpoint and endpoint time should not be used when using

    timed interval data logging.

    To use auto or timed endpoint:

    press the auto endpoint indicator appears

    Press again to select timed endpoint

    During calibration or sample measurement the

    decimal place flashes until a stable endpoint is

    reached, or until the endpoint time has elapsed,

    then the display freezes.

    To cancel timed endpoint enter 0 as the endpoint time.

    NOTES If 0 has been entered as the endpoint time, the manualendpoint indicator appears

    The method of endpointing can be saved as part of a

    method, see Section 2.17.1.

    2   IDID

    ID

    ID

    ID

     A 

  • 8/16/2019 Operating Instruction MA235 1

    17/20

    51709598 a11/96   17

            1 2 3 4   T   h  e   M

       A   2   3   5  u  s  e  s  a  m  e  n  u  s  y  s   t  e  m

       f  o  r   t   h  e  s  e   t  u  p   f  u  n

      c   t   i  o  n  s .   T   h  e  r  e  a  r  e  n   i  n  e

      m  e  n  u  s ,   t   h  e  s  e  a  r  e  :  c  a   l   i   b

      r  a   t   i  o  n ,

       d   i  s  p   l  a

      y ,  s  y  s   t  e  m    1 ,  s  y  s   t  e  m    2 ,

       l   i  m   i   t  s ,

       d  a   t  a   l  o  g ,

       G   L   P ,  m  e   t   h  o   d ,

       I   S   E .   T   h   i  s  c   h  a  r   t  s

       h  o  w  s   t   h  e  n   i  n  e  m  e  n  u  s .

            P      r      o      g

       1

       2

       3

       4

       5

       6

       7

       8

       9

       c   a   l   i

       b   r   a   t   i   o   n

       d   i   s   p   l   a  y

       s  y   s   t   e   m

        1

       s  y   s   t   e   m

        2

       l   i   m   i   t   s

       d   a   t   a   l   o   g

       G   L   P

       m   e   t   h   o   d

       I   S   E

      c  a   l  p  o   i  n   t  s

      c   h  a  n  n  e   l

       t   i  m

      e   /   d  a   t  e

       t  e  m  p  m  o   d  e

      a   l  a  r  m

      o  n   /  o   f   f

      o  n   /  o   f   f

      s  a  v  e

      e   l  e  c   t  r  o   d

      e   I   D

       l  a  s   t

      c  a   l

      c  o  n  c  u  n   i   t  s

       l  a  n

      g  u  a  g  e

      s  e   t   t  e  m  p

      s  e   t   l   i  m   i   t  s

       t   i  m  e  r

      c  a   l  a   l  a  r  m

       l  o  a   d

       i  n  c  r  e  m  e

      n   t  a   l

       t  e  c   h

      m   V

      o   f   f  s  e   t

      p   H

      r  e  s  o   l  u   t   i  o  n

       b  a  u   d

      e  n   d  p  o   i  n   t   t   i  m  e

       t  e  m  p

      r  e  c  a   l   l   d  a   t  a

       I   D

      n  o .

      r  e  s  e   t

       k  n  o  w  n  v

      a   l  u  e  s

       i  s  o

       b  e  e  p  e  r

      s  e   l   f   t  e  s   t

      s   l  o  p  e   l   i  m   i   t  s

      c   l  e  a  r   d  a   t  a

      s  a  m  p   l  e

      v  o   l  u  m  e  s

    >  

    >

              >

       >        E      n        t      e      r

       U  s  e   t   h  e  a  r  r  o

      w   k  e  y  s   t  o  m  o  v  e  a  r  o  u  n   d   t   h  e  m  e  n  u  s .

       T   h  e  n  u  m  e  r   i  c

       k  e  y  p  a   d  c  a  n   b  e  u  s  e   d  a  s  q  u   i  c   k  a  c  c  e  s  s   k  e  y  s .   F  o

      r  e  x  a  m  p   l  e ,   t  o  c   h  a  n  g  e   t   h  e   d   i  s  p   l  a  y

      r  e  s  o   l  u   t   i  o  n   t  o

       X .   X   X  p  r  e  s  s

       t   h  e  n

       (   t  o  a  c  c  e  s  s  m  e  n  u

       2   ) ,

       (   t  o  a  c  c  e  s  s   i   t  e  m

       3   ) ,

     

       (   t  o  s  e   l  e

      c   t  o  p   t   i  o  n   2   ) .

       T   h   i  s   i  s  u  s  e   f  u

       l  w   h  e  n  o  p  e  r  a   t   i  n  g   t   h  e   M

       A   2   3   5  r  e  m  o   t  e   l  y   f  r  o  m

      a   P   C ,  s  e  e   S  e  c   t   i  o  n   2 .   2   0 .

            2

            2

            3

            P      r      o      g

    2.9 Program Menu 

    2

  • 8/16/2019 Operating Instruction MA235 1

    18/20

    18 51709598 a11/96

    2.9 Program Menu (cont)  

    To access the menus:

    press the display shows three menu titles at a time (if you

    are using both measurement channels (A and B), youhave to select which channel you want to set up).

    Press and to move round the menus.You can move

    either way, and the menu wraps round

    at each end.

    Each time you press an arrow key

    the next menu title appears, e.g. press

    To access your chosen item press or

    To choose from options displayed use and then press

    Before you can enter or change a value the value must be flashing,

    for example: cal 1 = 7.000

    Press Change the flashing value using the keypad, then

    press again.

    Press to return directly to measurement.

    < display system 1 system 2 >

    channel

    conc units Use to move down through the items.

    pH resolution2

       >>

    >

    Prog

    < calibration display system 1 >

    >  

    < display system 1 system 2 >

    Enter

    Enter

    Enter

    Enter

    Enter

    Enter

    Enter

              >  >  

    >

    Read

    To access the menu items, press or when your

    required menu is highlighted,

    e.g. to access the display  menu items press

  • 8/16/2019 Operating Instruction MA235 1

    19/20

    51709598 a11/96   19

    2

    calibration

    cal points

    last cal

    mV offset

    iso

              >>  

    >

    Enter

    Enter

    Enter

    Enter

              >>  

    2.10 Calibration Programming 

    This menu allows you to set 5

    calibration points for pH or

    concentration; recall data for thelast 10 calibrations, set an offset

    value for relative mV mode and set

    the isopotential point for pH.

    Default refers to the factory

    setting which the MA235 uses if no

    information is entered.

    2.10.1 Calibration point setting 

    Press or when cal points  is highlighted.

    If you enter cal points from pH mode, the cal points for pHcalibration are set up. If you enter from concentration mode, the

    concentration calibration points are set up, and the standard set

    as the first calibration point is the isopotential point.

    Use and to select the calibration pointyou want to change and press

    The display flashes showing that the value can be changed.

    Enter the required value using the keypad and press

    # appears if you set a pH calibration value to a standard buffer,

    (the value must be the pH value at 25°

    C for the standard buffer).The MA235 stores temperature corrected values for the standard

    buffers (see Section 3.5).

    The next cal point is highlighted. If you want to change this next

    point, press and continue as before.

    Use and to select another calibration point to change,

    OR

  • 8/16/2019 Operating Instruction MA235 1

    20/20

    2

    2.10.1 Calibration point setting (cont)

    press or to leave cal points without any furtherchanges.

    Press to return directly to measurement.

    NOTE The concentration range of the MA235 is 1.00E –9 to9.99E+9. Enter values in the normal way (for example,

    100.0). If you enter a value within the range 0.0001 to

    19999 this is how it will be displayed. If you enter avalue outside this range (for example 20500) it will be

    converted to exponential format on the display (2.05E+4).

    2.10.2 Last calibration recall

    Press or when last cal  is highlighted.

    If you recall the data to the external printer it will print all the

    calibration data logged (up to 10 calibrations).

    If you recall the data to the display it is shown 4 calibrations at a

    time.

    Use and to scroll through the data; press to see

    further information.

    NOTES The MA235 can store data for the last 10 calibrations.

    So, if for example 15 calibrations have been carried out,the MA235 stores the data for calibrations 6 to 15.

    Calibration data cannot be cleared.

    Press to leave last cal.

    Press return directly to measurement.

    2.10 Calibration Menu (cont) 

    Prog   >

    Read

    Enter >

       >

    >          >  >  

    Read