48
Operating Instructions for Compact Disinfector G 7783 CD MIELABOR Laboratory Glasswasher It is essential to read these Operating Instructions before Q installing or using the machine, to avoid the risk of accident, or damage to the machine. M.-Nr. 04 195 811

Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Operating Instructions for

Compact Disinfector G 7783 CD MIELABORLaboratory Glasswasher

It is essential to read theseOperating Instructions before Qinstalling or using the machine,to avoid the risk of accident,or damage to the machine. M.-Nr. 04 195 811

Page 2: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

ContentsGuide to the machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Warning and safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Warning and safety instructions (detachable) . . . . . . . . . . . . . . . . . Centre pagesDescription of the machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Setting the water softener . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Opening and closing the door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Applications and loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Rinsing agent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Setting the dosage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Adding neutralising agent and (liquid) detergent . . . . . . . . . . . . . . . . . . . . . . . 18Venting the liquid dispensing system / Maintenance . . . . . . . . . . . . . . . . . . . . 19Adding powder detergent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Programme selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Programme chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Switching the machine on and off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Selecting a programme . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Selecting an additional programme . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Programme sequence indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Interrupting a programme. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Programming special functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Reactivating the water softener . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Cleaning and care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Drying unit (TA) cleaning and care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

Changing the Coarse Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Exchanging the S-class particle filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

Correcting minor faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Plumbing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Caring for the environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Contents

2

Page 3: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Guide to the machine

1 "Check water inlet/outlet" check light

2 "Reactivate softener" indicator light

3 Add (liquid) detergentindicator light

4 "Add neutralising agent" indicator light

5 Programme sequence indicator

6 Display

7 Door release

8 ON-OFF Button (I-0)

9 Programme touch controls

•••

DOS

°Cmin•

DESIN

•••

••••

•••

•••

I-O

123 4 5 6 7 8 9

•••A

B

CD

E

F

T

T

1

2

10’- 80’

Guide to the machine

3

Page 4: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Guide to the machine

10 Dispenser for neutralising agent (DOS 10 - red) and liquid detergent (DOS 60 - blue)

11 Drying unit (TA)

12 Safety fuse for TA Drying Unit

13 Time counter for TA Drying Unit

14 Temperature selector for TA Drying Unit

15 Service panel

16 Dispenser for powder detergent

17 Dispenser for liquid rinsing agent with dosage selector

18 Level indicator

19 Filter combination

20 Connection socket for saltcontainer (water softener)

10 11 12 13 14 15 16 17 18 19 20

Guide to the machine

4

Page 5: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Warning and safety instructionsThis machine conforms to currentsafety requirements. Inappropriateuse can however lead to personalinjury and damage to property.Read the operating instructionscarefully before starting to use thismachine. This way you will avoidthe risk of accidents, and damageto the machine.Keep these instructions in a safeplace, and make them available tofuture users.

Correct usage

This machine is designed for com-mercial use and for specialised ap-

plications only, as described in theseOperating Instructions. Using it for pur-poses other than those for which it wasdesigned would be unauthorised andcould cause harm.The manufacturer cannot be held re-sponsible for damage caused by im-proper use.

Please pay attention to the followingnotes, to maintain safe procedures.

The machine should be commis-sioned and then maintained only

by the Miele Service Department, anauthorised Miele Service Dealer orother approved competent person.

Do not install the machine in anarea where a danger of explosion

may be present.

The electrical safety of this ma-chine can only be guaranteed if

connected to a correctly installed ear-thing system on site. It is most import-ant that this basic safety requirement ismet and complies with local and na-tional regulations. It should be checkedby an expert if necessary.The manufacturer cannot be held re-sponsible for damage caused by mis-sing or defective earthing, such aselectric shock.

A damaged machine is dan-gerous. Switch off at the mains im-

mediately and call an authorised Mieleservice dealer, or suitably competentengineer.

Repair work may only be carriedout by an authorised, suitably com-

petent person. Repairs by unqualifiedpersons could be dangerous.

Personnel operating the machineshould be trained regularly. Child-

ren and untrained personnel should notbe allowed access to the machine orits controls.

Warning and safety instructions

5

Page 6: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Warning and Safety instructionsTake care when handling liquidssuch as detergents, wetting

agents, or neutralising agents. Thesemay contain irritant or corrosive ingre-dients, acids or alkalis. On no accountuse any organic solvent, as there maybe the danger of an explosion occur-ring.Follow the instructions and relevantsafety procedures carefully. Wear pro-tective gloves and goggles.

Avoid inhalation of powdercleaners. They can cause burning

in the mouth and throat if swallowed, orinhibit breathing.

The water in the machine must notbe used as drinking water.

Do not sit or lean on the open door,or rest objects on it. This could

cause the machine to tip and be dam-aged.

Be careful when sorting items withsharp pointed ends and position-

ing them in the machine that you do nothurt yourself or create danger for otheroperators.

When using this machine in thetemperature ranges from 70 to

95°C be especially careful not to scaldor burn yourself. Baskets and insertsmust first cool down. Any water whichmay remain in containers will be veryhot and must be emptied into the washcabinet.After drying with the TA drying unit thenopen the door to allow the items and in-serts to cool down.

Do not touch the heating elementsduring or directly after the end of a

programme, you could burn yourself.

Never use an ordinary or a highpressure hose on or in the vicinity

of the machine.

Before any maintenance or repairwork is undertaken, the machine

must be disconnected from the mainssupply, by switching off at the mainsisolator, or withdrawing the mains fuse.

The following points should be ob-served to avoid damage to the ma-chine and the loads being cleaned.

The standard of disinfection in theDisinfection programmes must be

confirmed by the user as a routine mat-ter. The process should be regularlychecked using chemical indicators,and bacteriologically at intervals usingbio-indicators, with full documentation.Proceed according to local health auth-ority regulations.

Disinfection programmes must notbe interrupted, as this can have a

limiting effect on the disinfection result.However, should an interruption benecessary. the complete programmemust be repeated.

Only use cleaning agents formu-lated for special processes and ap-

proved by Miele for use with this ma-chine. Use of unsuitable cleaningagents could adversely affect the loador the machine.

Warning and safety instructions

6

Page 7: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Warning and Safety instructionsFor pre-treatment and / or cleaningonly use low-foaming detergents

and disinfecting agents which havebeen approved by Miele. Foam canhave an adverse effect on the disinfec-tion and cleaning process.

Where a chemical additive is re-commended on technical applica-

tion grounds (for example with a clean-ing agent), this does not imply that themanufacturer of the machine takes re-sponsibility for the effect of the chemi-cal on the material of the items beingcleaned. Please be aware that changesin formulation, storage conditions etc,which may not be publicised by thechemical producer, can have a nega-tive effect on the cleaning result.

When using cleaning agents andspecialised products it is essential

that the manufacturer’s instructions arefollowed. Only use the product for theapplication described by the manufac-turer, to avoid any material damage orthe occurrence of strong chemical reac-tions such as oxyhydrogen.

The machine is designed only foroperation with water and additive

cleaning agents. Organic solventsmust not be used in the machine, asthere is the danger under certain cir-cumstances of explosion. Although thisis not the case with all organic sol-vents, other problems could arise withtheir use, for example damage to rub-ber and synthetic materials.

In critical applications where verystringent requirements have to be

met, it is strongly recommended that allthe relevant factors for the process,such as cleaning agent, quality ofwater are discussed with the Miele Ap-plication Technology specialists.

If the cleaning result is subject to theseparticularly stringent requirements (e.g.chemical analysis, specialised pro-cesses), a regular quality control testshould be carried out by the user to en-sure that required standards of cleanli-ness are being achieved.

Please follow the advice on installa-tion in these instructions and the

separate Installation Instructions.

Empty any containers or utensilsbefore arranging them in the ma-

chine.

Do not allow any remains of acidsor solvents, or chloride solutions to

get into the wash cabinet. Similarlyavoid any materials with a corrosive ef-fect.The presence in compounds of any sol-vents should be minimal, (especiallythose in hazard class A1).

Ensure that solutions or steam con-taining acid salts do not come into

contact with the steel outer casing ofthe machine, to avoid any corrosiondamage.

Warning and safety instructions

7

Page 8: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Warning and Safety instructionsAccessories

Use special inserts in accordancewith the instructions provided.

The only accessories to be usedwith the special commercial ma-

chine must be those supplied or ap-proved by MIELE for the specific appli-cation. Consult your dealer or the MieleProfessional department for advice.

Disposal of an old machine

When disposing of an old machinefirst make it unusable. Disconnect,

cut off the cable and any plug from thecable. For environmental and safetyreasons ensure the machine is com-pletely drained of any residue waterand cleaning agent. (Observe safetyregulations and wear safety gogglesand gloves). Make the door lock inoper-ative, so that children cannot acciden-tally shut themselves in. The simplestway to do this is to remove the lockingpin (2 philips screws above the door).Make appropriate arrangements for thesafe disposal of the machine. For tanksystem machines ensure that any wateris emptied out of the tank.

In countries where there are areaswhich may be subject to infesta-

tion by cockroaches or other vermin,pay particular attention to keeping themachine and its surroundings in aclean condition at all times. Any dam-age which may be caused by cock-roaches or other vermin will not becovered by the machine guarantee.

The manufacturer cannot be held re-sponsible for any damage causedthrough non-observance of theseWarning and Safety instructions.

Warning and safety instructions

8

Page 9: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Description of the machineThe G 7783 CD is a front-loading auto-matic machine for the efficient washing- neutralising - rinsing - final rinsing oflaboratory glassware. It can be oper-ated on normal water supply. There are"AD" (aqua destillata) programmes withfinal rinses using purified water (dis-tilled, R.O., or de-ionised).

On completion of the automatic pro-gramme sequence, laboratory glass-ware is cleaned to the standard re-quired for analysis.

The electronic control unit offers achoice of temperatures for the cleaningand final rinsing phases. A combinationoften preferred is cleaning - T1 85° Cand final rinse T2 70 °C.

With the SPECIAL "B" programme themachine can thermally disinfect at90°C with relevant holding times.

"Instrument disinfection in automaticmachines" is to be found under subsec-tion 3.2.4. of paragraph 10c ofBSeuchG, according to the GermanFederal Health Authority, Berlin (BGA).The disinfection parameters of between90°C and 93°C held for 10 minutesapply to the effective areas A + B. Re-duction factor 7 (by reduction factor~107).

The effective areas are defined as fol-lows:

A = Suitable for destruction of vegeta-tive bacteria, including myco-bacteriaand fungus and fungal spores.

B = Suitable for inactivation of virusesIncluding HBV and HIV).

The SPECIAL "B" programme operatesat 90°C with 10 minutes holding time.

This machine is fitted with a TA dryingunit and water softener as standard.

Available at extra cost:

– Machine lid

Description of the machine

9

Page 10: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Setting the water softenerTo avoid the build-up of calcium de-posits on glassware and in the machinethe water needs to be softened, wherethe supply hardness exceeds 107ppm, (6°d).

To ensure a steady supply of soft waterthe water softener unit must always be:

1. correctly set

2. reactivated with "dishwasher" salt assoon as the k indicator light comes on.

The water softener must be set to corre-spond with the water hardness the firsttime the machine is used.

Your local water authority will adviseyou on the water hardness in your area.

Setting the WaterSoftenerThe hardness range is divided into 18"time" units ad a zero position. The ma-chine is set at the factory, correspond-ing to 339 ppm, (19°d).

If your water supply is harder or softerthan this, then:

Use the touch pads on the controlpanel to change the time setting (seeTable).

Timeunit

°d mmol/l

° f °e ppmCaCO3

1817161514131211109876543210

67891011121315171922242832384871

1.071.251.421.601.781.962.142.312.673.033.383.924.274.985.706.768.54

12.64

10.712.514.316.017.919.621.423.226.830.433.939.342.850.057.167.885.7126.7

7.58.810,011.312.513.815.016.318.821.323.827.530.035.040.047.660.188.9

1071251431601791962142322683043393934285005716788571267

Press T1 and c touch pads simul-taneously and, whilst depressed,switch the machine on, using themain switch "I-0".

"P..." will appear in the display field (toindicate current programming status).

Press touch pad k. The "P" will dis-appear and setting "0" will appear inthe display field. This setting "0" cannow be changed if necessary. To dothis:

Press touch pad cas many timesas is necessary for the desired valueto appear in the display field.

Press touch padh. "SP" appears.

Press h once more. The setting willnow be stored. The display fieldclears.

The machine is now ready for oper-ation.

Setting the water softener

10

Page 11: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Setting the water softenerIf the water hardness in your area ispermanently below 107 ppm (6°d):

Select setting "0". The control lightk will not come on and the watersoftener does not need to be re-generated.

Reactivating the water softener , seepage 29.

Setting the water softener

11

Page 12: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Opening and closing the doorTo open the door a

Press the release catch as far as itwill go and, at the same time, graspthe door grip and open the door.

The door should only be opened dur-ing operation for emergencies: e.g. ifarticles are knocking together (see in-terrupting a programme).

To close the doorLift the door upwards and push untilit clicks into position. Do not pressthe release catch.

Opening and closing the door

12

Page 13: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Applications and loadingThis laboratory glasswasher normallyrequires a top and a bottom basket.

Depending on the type and shape ofthe glassware, inserts can be suppliedand there is a range of other basketsfor special applications. The diversity ofthe inserts is so great that it is notpossible to illustrate them all or give de-tailed instructions on their use here. Asan example there are inserts for directinjection spraying of volumetric flasks,butyrometers, pipettes etc.

General glassware used in labora-tories, e.g. beakers, conical flasks, testtubes, etc., can be arranged in full-size, half or quarter inserts in top or bot-tom baskets. Your Miele laboratorydealer will advise on specific require-ments.

The following instructions relate only tobasic preparation and loading of glass-ware.

Loading the machine

Removing excess soils

Empty all glassware before loadinginto the machine. Take any hygienemeasures necessary against infec-tion.

Ensure that no acid or solventresidues, especially hydrochloricacid or chlorides get into the washcabinet.

Remove all agar residues from petridishes.

Remove blood clots and residuesfrom test tubes, etc.

Remove all stoppers, corks, labels,sealing wax residues, etc.

Notes

Load the glassware so that water willcome into contact with all surfaces.This ensures that they will beproperly cleaned.

Do not place items to be cleaned in-side other pieces where they may beconcealed.

Hollow vessels such as beakers,measuring cylinders, flasks, etc.should be inverted and placed in thecorrect inserts. A cover net can beused to reduce the risk of movementduring the wash process.

Petri dishes and similar should beplaced in the correct insert with thesoiled side facing the centre.

Pipettes should be placed with themouthpiece facing upwards.

Deep based items should be placedat enough of an angle for water torun off them freely.

Tall, narrow, hollow pieces should beplaced in the centre of the basket toensure good coverage of water.

Do not cover the funnel in the middleof the top basket. Mobile injectorunits or baskets with an adaptermust engage correctly, (see below).

Applications and loading

13

Page 14: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Applications and loadingThe spray arms must not be blockedby items which are too tall for thebasket or which hang down. Rotatethe arms manually to test.

Important

Make sure that the spring adapter forwater connection engages correctlywhen a basket, injector or mobile unit isinserted into the machine. It must be 4 -5 mm higher than the water connectioninlet in the roof of the cabinet.If it is not, adjust the adapter accord-ingly.

Loosen the lock ring.

Push up the adapter (4-5 mm higherthan the water connection inlet) andtighten the lock ring.

Select baskets and inserts which areappropriate for the application.

Loading example:

E 380 "Intermiel" Injector mobile unit (with TA connec-tion)For narrow necked glassware, com-plete with 15 nozzles 4x1 mm/160 mmhigh, 18 nozzles 6x1 mm/220 mm highwith 33 "Mielava" clips.

O 175 Injector mobile unit (with (TA) Drying unit connection)(top basket) for direct injection of nar-row necked glassware, 34 nozzles 4x1mm/ 160 mm high with "Mielava" clips.

Please note:

powder detergent can only be placedin the inner side of the door when usingthe O 175 unit .

Applications and loading

14

Page 15: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Applications and loading the machine

O 188 top basket framecan take various inserts.

U 174 bottom basket framecan take various inserts.

E 109Half insert, stainless steel, with 21 glassbeakers up to 250 cm3, round and conical flasks etc.

E 106Half insert, stainless steel, with 28spring hooks at two heights for glass-ware, such as narrow necked bottles,measuring cylinders, medicine bottlesetc.

Applications and loading

15

Page 16: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Applications and loading the machineAdjusting the top basketAdjustable top baskets can be used inthis machine, They can be adjusted upor down by 2 cm from the middle posi-tion set on delivery.

The top basket will accommodate:

Vessels and glassware up to 19 cmin height.

The bottom basket will accommodate:

Vessels and glasses up to 25 cm inheight.

Top basketposition

Top basketheight (cm)

Bottombasket

height (cm) Top 17 27

Middle 19 25Bottom 21 23

To adjust the top basket

Pull out the top basket until a resist-ance is felt, lift from the runners andremove.

Unscrew the roller supports on bothsides of the basket with a 7 mmspanner, reposition as required andre-tighten.

Applications and loading

16

Page 17: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Rinsing agentA liquid storage container incorporatedin the door has a capacity of approxi-mately 200 ml. The required amount ofappropriate liquid rinsing agent (e.g.Mielclear, in those countries whereavailable), is dispensed automaticallyfrom this container.

Adding Rinsing agent

Undo the screw cap.

Pour rinsing agent into the storagecontainer until the level indicator isdark (see arrow in illustration). Use afunnel if necessary.

Firmly tighten the screw cap. Cleanup any spillage.

Refilling with rinsing agent

The level indicator shows how muchrinsing agent is left in the container.When there is no dark colouring visiblein the level indicator, the containerneeds to be refilled.

Setting the dosage

The dosage adjuster in the openinghas settings from 1-6 (1-6 ml). It is setto "3" (3 ml) at the factory.

Use a higher setting if spots are leftafter drying.

If smearing appears, use a lower set-ting.

Rinsing agent

17

Page 18: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Adding neutralising agent and (liquid) detergentAdding neutralising agentTo reduce the pH to neutral after alkalinecleaning, special neutralising acids canbe used in programme stage 4.

Important "Special functions" must firstbe used to set the programme to dis-pense neutralising agent in stage 4, in-stead of rinsing agent.

When the w indicator light on the con-trol panel lights up:

Fill the container DOS 10 (red mark-ings) with neutralising agent, orchange containers, and then:

Place the siphon tube firmly in theopening of the free-standing con-tainer (pay attention to colour cod-ing).

Re-fill the container in good time, itshould never be allowed to empty com-pletely.

Adding detergentsDetergent can be added in either liquidor powder form (for powder detergent,see relevant section, p. 18)

When the indicator light "DOS" on thecontrol panel lights up:

Fill the container DOS 60 (blue mark-ings) with liquid detergent, orchange containers, and then:

Place the siphon tube firmly in theopening of the free-standing con-tainer (pay attention to colour cod-ing).

Re-fill the container in good time, itshould never be allowed to empty com-pletely.

Note: If adding an extra dose of deter-gent in the pre-wash of programmes Eand F (see "Programme chart"), onlypowder detergent can be used.

Adding neutralising agent and (liquid) detergent

18

Page 19: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Venting the liquid dispensing system / MaintenanceVenting the liquid dispensingsystemBefore using the machine for the firsttime, or if the container had not beenfilled in time and allowed to empty, thedispensing system for liquid agents(apart from rinsing agent) must bevented.

Proceed as follows:

Press the I-0 (ON-Off) button.

Switch "A":Dispensing system for neutralisingagent.

Switch "B":Dispensing system for detergent (ifusing liquid detergent).

Use a pointed object to press inswitch "A" and/or "B" at the left handside end of the control panel forthree and a half minutes.

Each time the system is vented any liq-uid which runs into the wash cabinetmust be washed out as follows:

Select programme "A" (Cold).

After this has run the dispensing sys-tem is filled and ready for use.

Checking the dispensingsystemTo maintain the performance of the dis-pensing system, it should be servicedat regular intervals.

This work should only be carried out bysuitably qualified people or by a MIELEService Engineer.

Every 12-18 months

– Replace the dispensing hoses ineither the dispensing systems or theDOS modules.

AB

Venting the liquid dispensing system / Maintenance

19

Page 20: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Adding powder detergentOnly use specific detergents forspecial cleaning applications. Donot use domestic dishwasher deter-gents.

If not using liquid, fill dispenser in thedoor with powder detergent beforeeach programme (See "Programme ’B’and..." for exceptions).

Press the opening catch on the con-tainer flap forwards (see arrow). Theflap will spring open. (The flap willbe open at the end of a programme).

Fill the dispenser with the appropri-ate amount.

Dosage: approximately 3 g/l, corre-sponding to approximately 30 g perload. Follow manufacturer’s instructions.

Close the dispenser lid.

Adding powder detergent

20

Page 21: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Adding detergent Programme selection

Programme "B", and when using O-175 or a pipette plate

Add powder detergent directly to theinside of the door or dispense liquiddetergent via DOS 60.

An additional dose of powder detergentcan be added to the inside of the doorwhere required in applications E and F(see Programme chart).

When selecting a cleaning agentplease bear in mind the ecological ef-fects, in particular the following con-siderations:

How much alkalinity is needed to dothe required cleaning?

Is bleach required for disinfection orstain removal?

Are tensides at all necessary for dis-persal or emulsification?

The thermal disinfection programmerequires the use of a suitable mildlyalkaline or active chlorine-freecleaner.

Under certain circumstances, i.e. spe-cial soiling, different combinations ofdetergents and other additives may benecessary. Consult your Miele labora-tory dealer.

Take care when handling liquidssuch as detergents, disinfectingand neutralising agents. These maycontain acids, alkalis or disinfec-tants. Follow the relevant safetyprocedures and the instructions onthe packaging carefully. Wear protective gloves andgoggles if necessary.

Programme selectionChoose your programme according tothe type of glassware and the degreeof soiling.

The different programmes and their ap-plications are described overleaf in the"Programme chart" .

21

Page 22: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Programme chart

Programme When to use Detergent(when not using liquid

detergent)

follow manufacturer’s

instructions

A = COLD For the quick removal of chemicals whichcause problems for washing technology or candamage the machine.

B = SPECIAL Thermal disinfection to BGA standard. Cleaing of particularly stubborn soil.

30 grammes on to the righthand side door interior

C = MINI For lightly soiled items without AD rinsing. 30 grammes in the powderdispenser

D = STANDARD A wash programme with two AD rinses, wherethe items are intended for analytical use.

30 grammes in the powderdispenser

E = UNIVERSAL Programme with two AD rinses for items,intended for analytical use, which are moreheavily soiled with for example oil and greaseresidues. (Connect the cold water inlet to hotwater if residues to be dispersed are oilbased).

30 grammes in the powderdispenser

(and also 30 grammes onthe right hand side doorinterior if needed)

F = INTENSIVE-PLUS

For particularly encrusted residues (agar etc.). 30 grammes in the powderdispenser

(and also 30 grammes onthe right hand side doorinterior if needed)

Programme chart

22

Page 23: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Warning and safety instructions

This machine conforms to currentsafety requirements. Inappropriateuse can however lead to personalinjury and damage to property.Read the operating instructionscarefully before starting to use thismachine. This way you will avoidthe risk of accidents, and damageto the machine.Keep these instructions in a safeplace, and make them available tofuture users.

Correct usage

This machine is designed for com-mercial use and for specialised ap-

plications only, as described in theseOperating Instructions. Using it for pur-poses other than those for which it wasdesigned would be unauthorised andcould cause harm.The manufacturer cannot be held re-sponsible for damage caused by im-proper use.

Please pay attention to the followingnotes, to maintain safe procedures.

The machine should be commis-sioned and then maintained only by

the Miele Service Department, an auth-orised Miele Service Dealer or other ap-proved competent person.

Do not install the machine in anarea where a danger of explosion

may be present.

The electrical safety of this ma-chine can only be guaranteed if

connected to a correctly installed ear-thing system on site. It is most import-ant that this basic safety requirement ismet and complies with local and na-tional regulations. It should be checkedby an expert if necessary.The manufacturer cannot be held re-sponsible for damage caused by mis-sing or defective earthing, such aselectric shock.

A damaged machine is dan-gerous. Switch off at the mains im-

mediately and call an authorised Mieleservice dealer, or suitably competentengineer.

Repair work may only be carriedout by an authorised, suitably com-

petent person. Repairs by unqualifiedpersons could be dangerous.

Personnel operating the machineshould be trained regularly. Child-

ren and untrained personnel should notbe allowed access to the machine orits controls.

G 7783 CD

Instructions for the operator:Remove these pages and ensure that they are fixed in a position clearly visibleto the system operator in direct proximity to the installation site. Instruct and encourage operating personnel to heed the warning and safety in-structions.

Warning and safety instructions

1

Page 24: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Warning and Safety instructionsTake care when handling liquidssuch as detergents, wetting

agents, or neutralising agents. Thesemay contain irritant or corrosive ingre-dients, acids or alkalis. On no accountuse any organic solvent, as there maybe the danger of an explosion occur-ring.Follow the instructions and relevantsafety procedures carefully. Wear pro-tective gloves and goggles.

Avoid inhalation of powdercleaners. They can cause burning

in the mouth and throat if swallowed, orinhibit breathing.

The water in the machine must notbe used as drinking water.

Do not sit or lean on the open door,or rest objects on it. This could

cause the machine to tip and be dam-aged.

Be careful when sorting items withsharp pointed ends and position-

ing them in the machine that you do nothurt yourself or create danger for otheroperators.

When using this machine in thetemperature ranges from 70 to

95°C be especially careful not to scaldor burn yourself. Baskets and insertsmust first cool down. Any water whichmay remain in containers will be veryhot and must be emptied into the washcabinet.After drying with the TA drying unit thenopen the door to allow the items and in-serts to cool down.

Do not touch the heating elementsduring or directly after the end of a

programme, you could burn yourself.

Never use an ordinary or a highpressure hose on or in the vicinity

of the machine.

Before any maintenance or repairwork is undertaken, the machine

must be disconnected from the mainssupply, by switching off at the mainsisolator, or withdrawing the mains fuse.

The following points should be ob-served to avoid damage to the ma-chine and the loads being cleaned.

The standard of disinfection in theDisinfection programmes must be

confirmed by the user as a routine mat-ter. The process should be regularlychecked using chemical indicators,and bacteriologically at intervals usingbio-indicators, with full documentation.Proceed according to local health auth-ority regulations.

Disinfection programmes must notbe interrupted, as this can have a

limiting effect on the disinfection result.However, should an interruption benecessary. the complete programmemust be repeated.

Only use cleaning agents formu-lated for special processes and ap-

proved by Miele for use with this ma-chine. Use of unsuitable cleaningagents could adversely affect the loador the machine.

Warning and safety instructions

2

Page 25: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Warning and Safety instructionsFor pre-treatment and / or cleaningonly use low-foaming detergents

and disinfecting agents which havebeen approved by Miele. Foam canhave an adverse effect on the disinfec-tion and cleaning process.

Where a chemical additive is rec-ommended on technical applica-

tion grounds (for example with a clean-ing agent), this does not imply that themanufacturer of the machine takes re-sponsibility for the effect of the chemi-cal on the material of the items beingcleaned. Please be aware that changesin formulation, storage conditions etc,which may not be publicised by thechemical producer, can have a nega-tive effect on the cleaning result.

When using cleaning agents andspecialised products it is essential

that the manufacturer’s instructions arefollowed. Only use the product for theapplication described by the manufac-turer, to avoid any material damage orthe occurrence of strong chemical reac-tions such as oxyhydrogen.

The machine is designed only foroperation with water and additive

cleaning agents. Organic solventsmust not be used in the machine, asthere is the danger under certain cir-cumstances of explosion. Although thisis not the case with all organic sol-vents, other problems could arise withtheir use, for example damage to rub-ber and synthetic materials.

In critical applications where verystringent requirements have to be

met, it is strongly recommended that allthe relevant factors for the process,such as cleaning agent, quality ofwater are discussed with the Miele Ap-plication Technology specialists.

If the cleaning result is subject tothese particularly stringent require-

ments (e.g. chemical analysis, spe-cialised processes), a regular qualitycontrol test should be carried out bythe user to ensure that required stand-ards of cleanliness are being achieved.

Please follow the advice on installa-tion in these instructions and the

separate Installation Instructions.

Empty any containers or utensilsbefore arranging them in the ma-

chine.

Do not allow any remains of acidsor solvents, or chloride solutions to

get into the wash cabinet. Similarlyavoid any materials with a corrosive ef-fect.The presence in compounds of any sol-vents should be minimal, (especiallythose in hazard class A1).

Ensure that solutions or steam con-taining acid salts do not come into

contact with the steel outer casing ofthe machine, to avoid any corrosiondamage.

Warning and safety instructions

3

Page 26: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Warning and Safety instructionsAccessories

Use special inserts in accordancewith the instructions provided.

The only accessories to be usedwith the special commercial ma-

chine must be those supplied or ap-proved by MIELE for the specific appli-cation. Consult your dealer or the MieleProfessional department for advice.

Disposal of an old machine

When disposing of an old machinefirst make it unusable. Disconnect,

cut off the cable and any plug from thecable. For environmental and safetyreasons ensure the machine is com-pletely drained of any residue waterand cleaning agent. (Observe safetyregulations and wear safety gogglesand gloves). Make the door lock inoper-ative, so that children cannot acciden-tally shut themselves in. The simplestway to do this is to remove the lockingpin (2 philips screws above the door).Make appropriate arrangements for thesafe disposal of the machine. For tanksystem machines ensure that any wateris emptied out of the tank.

In countries where there are areaswhich may be subject to infesta-

tion by cockroaches or other vermin,pay particular attention to keeping themachine and its surroundings in aclean condition at all times. Any dam-age which may be caused by cock-roaches or other vermin will not becovered by the machine guarantee.

The manufacturer cannot be held re-sponsible for any damage causedthrough non-observance of theseWarning and Safety instructions.

Warning and safety instructions

4

Page 27: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Programme cycle Miscellaneous

1.Prelimi-

nary pre-wash

2.Prewash

II

3.Main

wash &possiblythermaldisin-

fection

4.Interimrinse Iand

possiblyneutral-isation*)

5.Interimrinse

II

6.Interimrinse

III

7.Finalrinse

8.Dryingphase(extra prog)

*) see"Programming special functions"

X

X93°C

**)

X X X70°CAD

(X) **) Increase intemp. holding timesee "Specialfunctions".

X X60°C

X X X60°C

(X)

X85°C

X X XAD

X70°CAD

(X)

X X85°C

X X XAD

X70°CAD

(X)

X X X85°C

X XAD

XAD

X 70°CAD

(X)

AD (Aqua destillata) = purified water, "H2O", (distilled, R.O., or de-ionised).X = sections included in standard programmes

23

Page 28: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Switching the machine on and off

1. Switching onPress the I-0 button.

2. Selecting a programmeAfter switching on the machine the indi-cator lights next to the selectors willlight up to show the programmes avail-able.

Press the programme selector forthe required programme. The washtemperature of the programme se-lected will show up in the displayfield.

Select additional programme T1 / T2 "TEMPERATURE" and/orc "DRYING" (see "Selecting an additional pro-gramme").

Press the h "Start" touch pad.. Thedisplay panel will now show the run-ning time in ascending minutes forthe selected programme.

During the heat-up stage in "Wash-ing/disinfecting" and "Final rinse" theend temperature will be displayed inthe panel.

All other programmes are automaticallyblocked and only the lights for the pro-gramme selected stay on.

Programme Changes

Provided the "Start" touch pad has notyet been pressed, a wrongly selectedprogramme may be changed as fol-lows:

Press the appropriate touch pad toselect the required programme.

Press the h "Start" touch pad.

Otherwise "Interrupt programme"(see page after next).

3. Selecting an additional pro-grammeThe following additional programmesare available:

T1 TEMPERATURE FOR WASHINGT2 TEMPERATURE FOR FINAL RINS-ING

Immediately after the selection of awashing programme the temperaturefor washing and for final rinsing can bealtered through the additional pro-gramme "Temperature". Press on theappropriate touch pad T1 or T2 for asmany times as necessary to changethe temperature step by step (40°C /50°C / 60°C / 70°C / 85°C / 93°C). Thelast temperature chosen will be storedfor the following wash/rinse cycles, andwill be shown in the display field.

•••

DOS

DESIN

•••

••••

••••

••I-O

°Cmin

•••A

B

CD

E

F

T

T

1

2

10’- 80’

Switching the machine on and off

24

Page 29: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Switching the machine on and offc DRYINGThe additional programme "DRYING"can be selected for 10 minutes stepsup to a total of 80 minutes immediatelyafter the selection of a cleaning pro-gramme except for (except A "COLD").

Selecting the drying time

Press c as many times as necess-ary to change the drying time stepby step until the desired time is dis-played (shown in the display field).For example:Press once - 10 minsPress twice - 20 mins and so on.

The drying cycle will automaticallybegin 10 minutes after the final rinse.The drying time display will then countdown as time goes on.

Selecting the Drying Temperature

The drying temperature is set using thetemperature selector for the (TA) DryingUnit.

To do this:

Take off the service panel (see "Elec-trical Connection").

Check the setting of the TA tempera-ture selector. the temperature can befreely set between 55°C and 95°Cand is electronically controlled.

– approx. 95°C (max. temperature) fornormal items (set ex-works),

– approx. 70°C for items with parts made of rubber,latex or plastics, with a specified tem-perature tolerance,

– approx. 55°Cfor items which are particularly heat-sensitive.

Please use the key supplied to switchover.

Note:

If on completion of a programme sub-sequent drying is necessary:

Select the programme q DRAIN(running time only approx. 1 minute)with the additional programme cDRYING (see above).

Switching the machine on and off

25

Page 30: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Switching on and offk REACTIVATION(see "Reactivating water softener").

q DRAIN,

if for example the programme has beeninterrupted (n.b. observe preventivemeasures to avoid infection):– Press the I-0 button.– Press the q programme touchpad.– Press the "Start"-button h.

Programme sequence indicatorWhen a washing programme has beenselected, the indicator lights in the dis-play will show the programme se-quence.

I Pre-wash

l Wash/possibly disinfecting

m Interim rinse

H Final rinse

c Drying (additional programme)

DESIN Disinfection (Only with the "B" SPECIAL pro-gramme maintaining parameters90-93 °C - 10’, 90-93 °C - 25’,85 °C - 10’, or 85 °C - 25’)

When part of a programme is com-plete, the corresponding indicator lightgoes out.

The programme is completed when noindicator lights are left on (expect for"DESIN" for the "B" SPECIAL pro-gramme) and when the "Start" h.lightgoes out. The display panel will showthe total running time of the programme.

Switching offPress the I-0 switch and let it springout.

The machine can now be opened andunloaded.

Interrupting a programmeOnce a programme is running it shouldonly be interrupted in extreme cases,e.g.: when articles in the machine rattleor bump against each other and haveto be rearranged.

The machine must then be switchedoff, the water drained and the pro-gramme restarted:

Switch the machine on and off againwith the I-0 switch.

Select the "DRAIN" programme (Ad-ditional programmes)

On completion, open the door andreplace the items securely and cor-rectly. Take any hygiene measuresnecessary against infection andwear protective gloves wherenecessary.

Replenish the powder detergent dis-penser in the door if necessary.

Close the door.

Reselect the programme.

Switching the machine on and off

26

Page 31: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Programming special functionsThe following special functions can beprogrammed where appropriate:

1. Setting water inlet to level plustime control instead of timed inlet (60 secs).

If the inlet water pressure is lower than2.5 bar, too little water will flow into thewash cabinet. By selecting level plustime control flow, more water will flowinto the wash cabinet, if there is an inletwater pressure between 1 and 2.5 bar.

The setting is adjusted separately forcold, hot and AD-water.

Switch off the machine.

1a. Cold water connection

Press T1 and c at the same time,hold them in and also swtich on themachine with "I-0".The current programme status "P..."appears in the display.

Press F. "10" or "11" will appear inthe display."10" = Cold water connection time

control (60 secs)"11" = Cold water connection level

plus time control

Press c to change from "10" to "11"or vice versa.

Press h. "SP" appears.

Press h again. The change hasbeen recorded and the display fieldclears.

1b. Hot water connection

Switch off the machine.

Press T1 and c at the same time,hold them in and also switch on themachine with "I-0".The current programme status "P..."appears in the display.

Press E. "20" or "21" will appear inthe display.

"20" = Hot water connection time control (60 secs).

"21" = Hot water connection level plustime control

Press c to change from "20" to "21"or vice versa.

Press h. "SP" appears.

Press h again. The change hasbeen recorded and the display fieldclears.

1c. AD water connection

Switch off the machine.

Press T1 and c at the same time,hold them in and also switch on themachine with "I-0". The current pro-gramme status "P..." appears in thedisplay.

Press D. "30" or "31" will appear inthe display."30" = AD water inflow time control

(60 secs)"31" = AD water inflow level plus time

control.

Press c to change from "30" to "31"or vice versa.

Switching the machine on and off

27

Page 32: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Programming special functionsPress h. "SP" appears in the display.

Press h again. The change hasbeen recorded and the display fieldclears.

2. Increasing the temperature hold-ing from 10 to 25 minutes in the"Wash" part of the programme, (onlyin the "B" SPECIAL programme).

Switch off the machine.

Press T1 and c at the same time,hold them in and also switch on themachine with "I-0". The current pro-gramme status "P..." appears in thedisplay.

Press C. "40" or "41" will appear inthe display."40" = Temp. holding time 10 mins."41" = Temp. holding time 25 mins.

Press c to change from "40" to "41"or vice versa..

Press h. "SP" appears in the display.

Press h again. The change hasbeen recorded and the display fieldclears.

Note:The temperature T1 must not exceed85 °C, (see "Selecting additional pro-grammes").

3. Adding neutralising agent in the "In-terim rinse 1" part of the programme.

The machine is set at the factory forrinsing agent to be dispensed in the"Final rinse" part of the programme. Ifneutralising agent is to be dispensed in-stead make the following adjustment:

Switch off the machine.

Press T1 and c at the same time,hold them in and also switch on themachine with "I-0". The current pro-gramme status "P..." appears in thedisplay.

Press B. "50" or "51" will appear inthe display."50" = Rinsing agent dosage"51" = Neutralising agent dosage.

Press c to change from "50" to "51"or vice versa..

Press h. "SP" appears in the display.

Press h again. The change hasbeen recorded and the display fieldclears.

Programming special functions

28

Page 33: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Reactivating the water softenerWhen the k indicator light on the lefthand side of the control panel comeson during a programme the built-inwater softener is depleted and cannotsupply any more softened water. Itmust therefore be reactivated with dish-washer salt as soon as the pro-gramme has finished.

If this cannot be done immediately, andfurther batches have been washed,then the reactivation process must becarried out twice in succession.

Only use special course grained dish-washer salt (granule size approx. 1-4mm), for reactivation, as other saltsmay contain insoluble impurities or ad-ditives which can impair the functioningof the softener.

The salt container holds approx. 2 kgsalt.

Important

If the only reactivation salt available isfine grain, please consult your Mieleservice representative.

Reactivation salt with granules largerthan 4 mm cannot be used.

Never fill the salt reservoir with dish-washer cleaning agent this can lead to– damage to the water softener and– blockage of the salt container filter insert, resulting in a build up of pressure in the salt reservoir.The irritant alkali solution could be ahazard and personal injury could becaused when the salt container isremoved.Make sure that you have picked upthe right packet of dishwasher saltbefore filling the salt container.

Filling the salt container

Unscrew the filter insert from the saltcontainer and remove.

Fill the salt container with granulardishwasher salt and screw the filterinsert back in place.

Reactivating the water softener

29

Page 34: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Position the salt container asfollows:

Remove the bottom basket from themachine.

Unscrew the plastic cap in the baseof the wash cabinet.

Place the salt container on the reacti-vation socket and screw firmly intoplace.

Switch on the machineOpen the stopcocks if previously closed.Close the door.Press I-0.Press the k touch pad.Press the "Start" h touch pad.

The reactivation programme proceedsautomatically and finishes when the indi-cator light k at the left hand side of thecontrol panel and the "Start" indicatorlight h have gone out.

Then:Switch off the machine.Unscrew the salt container carefully,so that any water pressure that hasbuilt up can be released. Do not useforce. If you cannot uncrew the saltcontainer by hand, contact the MieleService department.Screw the plastic lid back on the socket.Reposition bottom basket.Rinse the salt container and filter in-sert thoroughly. Turn off the stopcocks if the machineis not to be used again for some time.

ImportantThe water pressure (flow pressure at thetake off point) must be at least 2.5 bar. Ifit less than this see "Programming spe-cial functions".

If the flow pressure is below 2.5 bar, orif it fluctuates, the water softener can-not be reactivated in the normal way.After reactivation, salt remains may stillbe found in the salt container.

To use up the salt remains and to washout the water softener, the "reactivatesoftener" programme must be switchedon again.

Reactivating the water softener

30

Page 35: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Cleaning and careCleaning the filters in thewater cabinetThe filter combination in the base of thecabinet should be inspected regularlyand cleaned if necessary.

Caution!Watch out for glass splinters whichcould cause injury.

Cleaning the coarse filter

Press the 2 lugs together, removeand clean the filter.

Put the clean filter back in positionand press until it clicks in place.

Cleaning the fine, flat and micro-finefilters

Remove the coarse filter

Remove the fine filter (if fitted) frombetween the flat and micro-fine filters.

To unscrew the micro-fine filter, takehold of the two lugs, turn twice in ananti-clockwise direction.

Remove, together with the flat filter.Clean the filters.The filter combination is then re-placed in the reverse order to whichit was removed.

The machine must not be used withoutall the filters in place.

Cleaning and care

31

Page 36: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Cleaning and careCleaning the filters in the waterinletFor the protection of the inlet valve(s),filters are incorporated in the screwconnection of the water inlet hose. Thefilters must be cleaned when dirty,otherwise insufficient water flows intothe wash cabinet.

Turn off stopcocks.

Unscrew the water inlet hoses.

Clean large area filter (1) and fine filter(2) or replace if necessary.

After reconnection, turn on the stop-cock to test for leaks.

There is a further filter in front of thewater inlet valve which must only becleaned or replaced by a suitably quali-fied and competent person.

2

1

Cleaning and care

32

Page 37: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Drying unit (TA) cleaning and careChanging the Coarse FilterThe coarse filter should be changedwhen it is soiled or after approx. 100operating hours, (see time counter onthe TA unit).

Grip the inlet grille to the right andthe left and pull up out of the re-tainers and then downwards and off.

Take out the perforated plate.

Change the coarse filter.- The smooth side of the filter mustface the rear.

Position the perforated plate at thebase and press into place at the top.- The edge surround should face for-ward.

When putting back the inlet grille, itshould audibly click into place.

Exchanging the S-class particlefilterIf is advisable for this filter to be re-placed after approx. 500 operatinghours (see the time counter on the TAcontrol unit).

This filter should only be changed byan authorised Service Dealer or theMIELE Service department.

The effective operating of the machinecan only be achieved if original "Miele"S - class particle filters are used.

Drying unit (TA) cleaning and care

33

Page 38: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Correcting minor faultsWith the aid of the following explana-tory notes, minor faults can usually becorrected without contacting the MieleService Department.Any work on the electrical componentsof the machine should only be carriedout by a suitably qualified and compe-tent person.

Faults/Possible Causes

The machine does not start

– Check that the door is properly shut.

– Check the plug connection or isolator.– Check the fuses.

A few minutes after switching on, theindicator light p q flashes.

– The stopcocks are closed.– The water inlet hose filters are dirty.

Switch off the machine.Open the stopcocks or clean the fil-ters (see "Cleaning and care").Switch on the machine again and re-select the programme.

The wash cycle finishes too earlyand the p q indicator light flashes

– The drain hose is kinked.– The door was opened and closed

during the programme.– There was an interruption to the

power supply.

Correct the faultPump away the water and start theprogramme again (see procedureunder "Interrupting a programme").

The water in the wash cabinet is notheated; the programme sequencelasts too long.

This machine has a resettable heaterlimiter which will switch off the heatersin the case of over-heating. This couldbe caused for example, if large articlesobscure the heating elements or if thefilters in the wash cabinet are blocked.

Remove cause of fault.

Take off service panel (see under"Electrical connection").

Press in the re-set button (blue) onthe heater limiter.

If this switch trips repeatedly, your Ser-vice Dealer or the Miele Service Depart-ment must be contacted.

In the event of any faults which cannotbe remedied, please contact your Ser-vice Dealer or the Miele Service Depart-ment.

Correcting minor faults

34

Page 39: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

InstallationPlease refer to the Installation diag-ram supplied with the machine.

The machine must be installed cor-rectly and levelled.

Any unevenness in the floor level canbe compensated by adjusting thescrew feet on the right and left handsides at the front of the machine.

The service panel must first be lifted offand the cover plate must be un-screwed (see under "Electrical Connec-tion").

When the screw feet are screwed in,the disinfector can be moved back-wards or forwards on its rollers.

The machine can be installed in the fol-lowing ways:

Free-standing *)

Slot-in or at the end of a row: *)

The machine can be installed adjacentto other equipment or furniture or in arecess at least 90 cm wide and 70 cmdeep.*) A machine lid is essential with thistype of installation (available as a spe-cial accessory).

Building uner:

The machine should be positionedunder a continuous worktop. The nichedimensions must be at least 90 cmwide, 70 cm deep and 82 cm high.

Important for machines with DHKsteam condensor:

To avoid steam damage to the worktopthe self-adhesive protective sheet sup-plied (25 x 58 cm) should be stuckunder the worktop in the area of thecondensor.

Position and secure the machine

To guarantee stability the machinemust first be aligned and then screwedto the worktop.

Installation

35

Page 40: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Installation

Open the door. Through each of thetwo holes in the front rim, left andright, screw the machine to the frontedge of the continuous worktop.

Do not use silicone sealant to sealthe gaps between the machine andany neighbouring units as this wouldhinder ventilation to the circulationpump.

Depending on the various requirementsfor building under, the following kit mayalso be required:

Cover plate (worktop protection)

The underside of the work surface isprotected from steam damage by aplate of stainless steel.

Installation

36

Page 41: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Electrical connectionAll electrical connection work must becarried out by a qualified electrician incompliance with local and nationalregulations.

Connection should be made via asuitable isolator, with an on-off switchwhich should be easily accessiblefor servicing.

For extra safety it is advisable to in-stall a residual current device (RCD)with a trip current of 30 mA (DINVDE 0664).

WARNINGTHIS APPLIANCE MUST BEEARTHED

For technical data see data plate orwiring diagram supplied.

The machine must only be operatedwith the voltage, frequency and fusingshown on the data plate.

The machine is not convertible .

The wiring diagram is to be found be-hind the cover plate on the left handside, attached to the floor of the ma-chine.

The data plate with test certification ison the rear of the machine, and on thecover plate (behind the service panel).

To remove service panel and cover plate

Holding the service panel at the topright and bottom left hand corners,pull up out of its retainers and lift outat the bottom.

Isolate the machine from the powersupply.

Unscrew the cover plate.

Pull off earth lead if necessary.

Electrical connection

37

Page 42: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Electrical connectionRemove the plastic protective cap.

To re-assemble:

Refit plastic protective cap, coverplate and service panel in thereverse order to which they wereremoved.

Make certain that the earth lead isreconnected.

Consult the Installation Instructions sup-plied.

Electrical connection

38

Page 43: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

PlumbingConnection to the water inlet

The machine must be connected tothe water supply in accordance withthe local and national regulations,(NWC in GB).

The machine is constructed to com-ply with DVGW regulations, in such away that an external non return valveis not required.

The water pressure (flow rate) mustbe between 2.5 and 10 bar. If thewater pressure is not within thisrange, the Customer Service Depart-ment will advise you of the measuresto take (see also "Programming spe-cial functions").

The machine is supplied as standardfor connection to cold (blue mark-ings) and hot (red markings) watersupply to max. 70°C. See next pagefor AD connection to purified water(H20 pure).The inlet hoses should be connectedto the respective cold and hot waterstopcocks.If no hot water supply is available,the hose for hot water (red marking)must also be connected to the coldwater supply. In this case both hosesmust be connected to the cold watersupply.In special cases, see Programme Ein the chart, connection of bothhoses to hot water is recommended.In order to prolong its life, the hosemarked 90°C should only be usedon a hot water connection supplyingwater no hotter than 70°C.

Stopcocks with 3/4 BSP male threadto be provided on site. They shouldbe easily accessible because thewater supply must be turned offwhen the machine is not in use.

The DN 10 inlet pressure hoses areapproximately 1.7 m long and termi-nate in a 3/4" female thread. On noaccount must the inlet filters beremoved.

For additional protection of the sole-noid valve against impurities in thewater supply, large surface area fil-ters are enclosed in the kit suppliedwith the machine (see illustration in"Cleaning and care, water inlet").The filter for AD purified water (H2Opure) is made from chrome-nickelsteel and has a matt upper surface.

Consult the installation instructions sup-plied.

The water in the machine is not suit-able for drinking.

Plumbing

39

Page 44: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

PlumbingAD connection to purified water,pressurised, (1.5 - 10 bar)

The machine is delivered from the fac-tory ready for connection to a press-urised system, 1.5 - 10 bar.

The AD pressure tested hose,marked H2O pure, with 3/4" threadmust be connected to the on-site ADstopcock for purified water.

Connection to purified water non-pressurised (0 - 1.5 bar)

For connection to a supply of 0 - 0.3bar, or 0.3 - 1.5 bar the machine mustbe converted, if not supplied convertedex-works. An “ADP” booster pump can only be fitted. This is only possiblewhen the machine has no DHK steamcondensor. As this work requires specialised knowledge, it should onlybe carried out by a Miele approved engineer.

The outlet of the purified water con-tainer (non-pressurised) must be atleast at the height of the upper edge ofthe machine. (see "Installation instruc-tions").

Connect the non-pressurised ADwater hose to the purified water con-tainer

If the AD connection to purifiedwater is not being used (or if itshould not be used) then the ma-chine must be converted by a Mieletrained or approved engineer.

Plumbing

40

Page 45: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

PlumbingDrainage

The drainage system is fitted with anon-return valve which prevents dirtywater from flowing back into the ma-chine via the drain hose.

The machine should preferably beconnected to a separate drainagesystem on site. If no separate drain-age system is available, seek profes-sional advice.

The machine is supplied with flexibledrain hoses with an internal diameterof 22 mm and approx. 1.5 m long.The drain hose must not be short-ened.Hose clamps for connection are sup-plied.

Longer drain hoses (up to 4 m) areavailable if required. Contact theMiele Service Department.

The drainage system must not ex-ceed 4 metres and the drain pumphas a maximum delivery head of 1metre.

Ventilation of the wash cabinetat rearFor machines without DHK Steam con-densor, the rear ventilation of the washcabinet should be connected to an on-site venting system.

Please refer to the Installation Instruc-tions provided.

Plumbing

41

Page 46: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Caring for the environmentDisposal of the packing materialThe transport and protective packinghas been selected from materialswhich are environmentally friendly fordisposal and can normally be recycled.

Rather than just throwing these ma-terials away, please ensure they are of-fered for recycling.

Disposal of your old machineOld machines contain materials whichcan be re-cycled. Please contact yourlocal authorities or scrap merchantabout potential recycling schemes, be-fore disposing of the machine.

When disposing of an old machinefirst make it unusable. Disconnect,cut off the cable and any plug fromthe cable. For environmental andsafety reasons ensure the machineis completely drained of anyresidue water and cleaning agent.(Observe safety regulations andwear safety goggles and gloves).Make the door lock inoperative, sothat children cannot accidentallyshut themselves in. The simplestway to do this is to remove the lock-ing pin (2 screws above the door).Make appropriate arrangements forthe safe disposal of the machine.For tank system machines ensurethat any water is emptied out of thetank.

Caring for the environment

42

Page 47: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Technical DataNoise emission in dB (A)Measurement of noise: 65

Technical Data

43

Page 48: Operating Instructions for Compact Disinfector G 7783 CD ...€¦ · 11 Drying unit (TA) 12 Safety fuse for TA Drying Unit 13 Time counter for TA Drying Unit 14 Temperature selector

Alteration rights reserved / 001 3397

This paper consists of cellulose which has been bleached without the use of chlorine.