Operating manualincl. Declaration of conformity
BG 805 096 BE (9907) 1
Vacuum measurement and control unitfor Compact Gauges
2 Contents BG 805 096 BE (9907) DualGauge.om
The DualGauge™ (TPG 252 A) is used together withPfeiffer Vacuum Compact Gauges for total pressuremeasurement. The products must be operated asdescribed in the respective manuals.
In all communications with Pfeiffer Vacuum Instruments,please specify the information given on the productnameplate. For convenient reference copy thatinformation into the nameplate replica below.
Typ:No:F-No:
V Hz VA
Pfeiffer Vacuum, D-35614 Asslar
This document applies to products with part numbersPT G28 270 (230 V∼)PT G28 272 (120 V∼)PT G28 273 (100 V∼)
The part number can be taken from the product name-plate.
This document is based on firmware number
BG 509 727 -C
If your unit does not function as described in this docu-ment, please check that it is equipped with the abovefirmware version (or higher) (→ Appendix C).
We reserve the right to make technical changes withoutprior notice.
DualGauge™ Balzers AGFullRange™ Balzers AG
Intended use
Product identification
Validity
Trademarks
BG 805 096 BE (9907) DualGauge.om Contents 3
Table of contents
Intended use 2Product identification 2Validity 2Trademarks 2
1 Safety 51.1 General safety instructions 51.2 Symbols used 51.3 Liability and warranty 6
2 Design 72.1 Front panel 72.2 Compatible gauges 82.3 Display formats 92.4 Rear panel 11
3 Installation 123.1 Installing the equipment 123.2 Connections 133.2.1 Mains power connector 133.2.2 <sensor 1>, <sensor 2> connector 143.2.3 <control> connector 153.2.4 <s1>, <s2> connector 15
4 Operation 174.1 Initial start up 174.2 Power on 174.3 Power off 244.4 Selecting a gauge 244.5 Displaying the connected gauges 244.6 Turning a gauge on / off 254.7 Displaying the threshold values 264.8 Adjusting the threshold values 284.8.1 Logarithmic gauges 294.8.2 Linear gauges 304.9 Adjusting other parameters 324.9.1 General procedure 324.9.2 Setting the parameters 334.9.3 Selecting the pressure unit 454.9.4 Enabling the entry lock function 454.9.5 Setting the transmission rate 464.9.6 Selecting the display change mode 474.9.7 Adjusting the measured value filter 474.9.8 Adjusting the calibration factor 484.9.8.1 Logarithmic gauges 494.9.8.2 Linear gauges 49
4 Contents BG 805 096 BE (9907) DualGauge.om
4.9.9 Adjusting the measurement range of a lineargauge 50
4.9.10 Offset function (zero) of a linear gauge 504.9.10.1 Displaying and adjusting the offset value
524.9.10.2 Enabling the offset correction function of a
linear gauge 524.9.11 Adjusting the gauge control or 534.9.12 Adjusting the gauge underrange control 54
5 Technical data 56
6 Maintenance and care 65
7 Status indications 66
8 Error messages 71
9 Accessories 74
10 Spare parts 75
11 Disposal 76
Appendix 77A: Adapting the equipment voltage, exchanging the
fuses 77B: Dangerous voltages at the <s1>, <s2> connector79C: Test programs 80D: Literature 83E: Conversion of pressure units 86
Declaration of conformity 87
For cross-references within this document, the symbol(→ 2 XY) is used, for cross-references to other docu-ments, the symbol (→ & [Z]).
BG 805 096 BE (9907) DualGauge.om Safety 5
1 Safety
• Take the necessary precautions when doing installa-tion, maintenance or repair work (→ 2 12, 65, 77,79).
• According to EN 61010, voltages above 30 V∼ or60V= present a hazard of contact. Make sure to takethe corresponding safety measures (→ 2 12, 13, 79).
Communicate the safety instructions to all other users.
Skilled personnel
All work described in this document may only becarried out by persons who have suitable technicaltraining and the necessary experience or who havebeen instructed by the end-user of the product.
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment andenvironmental damage.
Note
Information on correct handling or use. Disregard canlead to malfunctions or minor equipment damage.
Press this key briefly.
Press this key additionally.
Press these keys simultaneously.
The result is O.K.
1.1 General safetyinstructions
1.2 Symbols used
6 Safety BG 805 096 BE (9907) DualGauge.om
The result is not as expected.
Waiting time, reaction time, duration oftest
Tip, recommendation
Lamp / display is lit
Lamp / display flashes
Lamp / display is dark
<set point> Lettering
«sen1» LED display
« » 7-segment display
»on« / »off« State of a switching function (e.g. relay)
Pfeiffer Vacuum assumes no liability and the warrantybecomes null and void if the end-user or third partiesmake any kind of modifications to the product whichexceed the work mentioned in the productdocumentation.
Since the individual electronic components are delicate,appropriate measures must be taken to protect themfrom static electricity. Modules must be stored in anti-static bags or containers.
Damage resulting from disregard of the above warningmay lead to a revocation of the warranty.
1.3 Liability andwarranty
BG 805 096 BE (9907) DualGauge.om Design 7
2 Design
Key labeling:
Measurement modeSetpoint mode
1 Display Meaningsen 1 sen 2 Sensor 1, 2 → 2 24
cal 1 cal 2 Calibration factorsensor 1, 2
→ 2 48
on 1 on 2 Sensor 1, 2 on → 2 25
sp 1 sp 2 Setpointsensor 1, 2
→ 2 26
hi lo Upper / lowerthreshold
→ 2 26
mbar/bar Pressure units → 2 45
TorrPa
err Error → 2 71
2 Control keys
3 Pressing two keyssimultaneously
→ 2 29, 30
2.1 Front panel
Fig. 2-1 Front panel
1
2
3
8 Design BG 805 096 BE (9907) DualGauge.om
Comprehensive list of gauge types → 2 56
PiraniGauge Compact Pirani Gauge
(Pirani gauge)
IKRCold CathodeGauge Compact Cold Cathode Gauge
(cold cathode gauge)
PKRFullRangeTM CCGauge
Compact FullRange™ CCGauge(Pirani/cold cathode gauge)
IMR
Process IonGauge Compact Process Ion Gauge 1)
(Pirani/high pressure gauge)
CapacitanceGauge CMR
ACRCompact Capacitance Gauge 1)
(capacitive sensor)
PiezoGauge
APRCompact Piezo Gauge(piezoresistive sensor)
Sensor X Any of the above gauges 1)
or no sensor at all
1) Restrictions → 2 15.
2.2 Compatible gauges
BG 805 096 BE (9907) DualGauge.om Design 9
50 bar
1000 mbar
1 mbar
Floating point display
e.g. or 1 Torr
Exponential display
e.g.
10-11 mbar
50 bar
1000 mbar
Floating point display
e.g.
10-4 mbar
10-11 mbar
2.3 Display formats
Logarithmic gauges
IKRCold CathodeGauge
PKRFullRangeTM CCGauge
IMR
Process IonGauge
Linear gauges
PiezoGauge
APR
CapacitanceGauge CMR
ACR
10 Design BG 805 096 BE (9907) DualGauge.om
Note
The unit switches automatically betweenexponential and floating point display.When Pascal is selected as pressure unit,only exponential display is available(→ 2 45).
When the IMR gauge is used, the pressure isindicated only with 1 digit accuracy in Piranimode.
BG 805 096 BE (9907) DualGauge.om Design 11
Fig. 2-2 Rear panel
VHQVRU
56 VHQVRU
FRQWURO
IXVHLQVLGH
V
V
1 Mains switch → 2 13
2 Mains power connector → 2 17
3 Connectors for gauges → 2 14
4 Connectors for external control signaland analog outputs
→ 2 14
5 Relay output → 2 14
6 Serial interface → 2 46
2.4 Rear panel 3
6
1
2
3
5
4
12 Installation BG 805 096 BE (9907) DualGauge.om
3 Installation
DANGER
Caution: possibly damaged productPutting a damaged product into operation canbe extremely dangerous.Make sure the product is not damaged. If youassume that the product is damaged makesure it is not put into operation again.
It can be assumed that the equipment is not safe tooperate when:
• it has sustained any visible damages
• it no longer functions
• it has been stored for a longer period underunfavorable conditions
• it has been subjected to severe transport stress
The product has been designed for installation in a 19"rack frame built according to the DIN 41 494 standard(→ 2 74).
DANGER
Caution: protection class of the rackIf the product is installed in a rack, it is likelyto lower the protection class of the rack(protection against foreign bodies and water)e.g. the EN 60204-1 regulations for switchcabinets.Take appropriate measures for the rack tomeet the specifications of the protectionclass.
Take the appropriate safety measures (→ 2 14, 79)when voltages above 30 V∼ or 60 V= are applied at theconnector <s1> or <s2> .
3.1 Installing theequipment
... in a rack
BG 805 096 BE (9907) DualGauge.om Installation 13
Note
The temperature inside the rack must notexceed 50 °C.
If you intend to use the Dual-Gauge™ as desktop unit,attach the two enclosed rubberfeed to and slide the enclosedrubber bar on its front panelfrom below.
WARNING
Before connecting the DualGauge™ to themains power supply, verify that the powerrating (indicated on the product nameplate)agrees with the local mains power specifica-tions. If they do not agree convert the voltageof the unit according to the instructions inAppendix A.
... as desktop unit
3.2 Connections
3.2.1 Mains powerconnector
14 Installation BG 805 096 BE (9907) DualGauge.om
DANGER
The power connector may only be pluggedinto a socket with protective ground. Use only3-pole mains power cables with the protectiveground connected correctly. This protectionmust not be nullified by using an extensioncable without ground conductor.To assure continuous grounding protection,connect the mains power cable before con-necting any other cables. In the same way,disconnect all other cables before unpluggingthe mains power cable.
If the unit is installed in a rack, the mains voltage mustbe supplied and turned on via a central distributor.
Here, you can connect either:
• a Pirani gauge
• a cold cathode gauge
• a Compact FullRange™ CC Gauge
• a Pirani/high pressure gauge
• a linear gauge
3.2.2 <sensor 1>,<sensor 2>connector
BG 805 096 BE (9907) DualGauge.om Installation 15
DANGER
Since the IMR 260 and CMR 27x have ahigher power consumption than other gauges,only one of these types can be connectedand operated in combination with any othergauge type. 1)
Always turn the unit off before connecting or discon-necting a gauge.
Use this connector to:
• turn the Pirani/high pressure gauge on/off e.g. via anexternal signal
• further evaluate the analog measurement signals
• evaluate any error signals
Use the enclosed connector (Pin assignments → 2 61).
Use this relay output if you want to use the switchingfunctions for external control.
1) Two IMR 260 may be connected. The DualGauge™
ensures that only one IMR 260 is operated at a time.
3.2.3 <control>connector
3.2.4 <s1>, <s2>connector
16 Installation BG 805 096 BE (9907) DualGauge.om
DANGER
Take the appropriate safety measures(→ 2 79) and consider the technical data(→ 2 56, 61). When the DualGauge™ is usedas desktop unit a maximum voltage of 30 V∼or 60 V= may be connected (protective lowvoltage).
Skilled personnel
Use the enclosed connector and screw it tothe connector of the unit. The connector mayonly be plugged in or unplugged when thepower line is idle.
BG 805 096 BE (9907) DualGauge.om Operation 17
4 Operation
WARNING
Before turning the unit on, make sure that theoperating voltage of the equipment agreeswith the local mains power. Also verify thatthe unit is correctly grounded, that all thegauges and cables are correctly installed(→ 2 12) and take the technical data(→ 2 56) into consideration.
DANGER
Any break in the protective grounding insideor outside the unit, and any loosening of thegrounding connection can make the unit dan-gerous to operate. Intentional interruption ofthe grounding line is not admissible.
The mains switch is located on the rear panel. Turn onstand-alone units from that power switch, rack mountedunits from the central power distributor.
After power on the unit:
• runs through a self-diagnostics routine
• identifies the gauges connected
• activates the parameters effective when it was lastturned off
• switches to measurement mode (on sensor 1)
• adjusts the parameters (if any other gauges wereconnected in the meantime)
• displays a pressure indication, a status indication, oran error message
4.1 Initial start up
4.2 Power on
18 Operation BG 805 096 BE (9907) DualGauge.om
Note
Status indications « » and « » as wellas error messages « » are outputaccording to the pressure unit and calibra-tion factor settings. The indications in thisdocument refer to the pressure unit mbarand a calibration factor of 1.00.
BG 805 096 BE (9907) DualGauge.om Operation 19
Sensor 1PiraniGauge
Sensor X
The unit displays, for instance, « » (1.3×10-2)
or « » (floating point format).The measured va-lue is within the measurement range of the gauge.
The unit displays « ». The pressure in thevacuum system is lower than the lower limit of themeasurement range of the connected Pirani gaugeor the gauge is no longer aligned (→ & 1, 2, 3, 4).
The unit displays « ». The pressure in the vac-uum system is higher than the upper limit of themeasurement range of the connected Pirani gaugeor the gauge is no longer aligned (→ & 1, 2, 3, 4).
The unit displays « ». No gauge is con-nected (at the <sensor 1> connector).
The unit displays « ». In addition, the «err»lamp flashes. The Pirani gauge has an error(→ 2 72).
The unit displays « ». In addition, the «err»lamp flashes. You have turned the unit back on tooquickly after turning it off (→ 2 24).
Turning the unit on
20 Operation BG 805 096 BE (9907) DualGauge.om
IKRSensor 1Cold CathodeGauge
Sensor X
The unit displays « ». A connected coldcathode gauge is not turned on (→ 2 25).
The unit displays, for instance, « »(2.8×10-6). The measured value is within themeasurement range of the gauge.
The unit displays « ». The pressure in thevacuum system is lower than the lower limit of themeasurement range of the connected coldcathode gauge or the gas discharge has notignited yet (→ & 5, 6, 7, 8).
The unit displays « ». The pressure in thevacuum system is lower than the lower limit of themeasurement range of the connected coldcathode gauge or the gas discharge has notignited yet (→ & 9).
The unit displays « ». The pressure in thevacuum system is higher than the upper limit ofthe measurement range of the connected coldcathode gauge.
The unit displays « ». No gauge is con-nected (at the <sensor 1> connector).
The unit displays « ». In addition, the «err»lamp flashes. The cold cathode gauge has anerror (→ 2 72).
The unit displays « ». In addition, the «err»lamp flashes. The cold cathode gauge has anerror (→ 2 72).
The unit displays « ». In addition, the «err»lamp flashes. You have turned the unit back on tooquickly after turning it off (→ 2 24).
Turning the unit on
BG 805 096 BE (9907) DualGauge.om Operation 21
PKRSensor 1
FullRangeTM CCGauge
Sensor X
The unit displays, for instance, « »(2.8×10-6) or « » (floating point format). Themeasured value is within the measurement rangeof the gauge.
The unit displays « ». The pressure in thevacuum system is lower than the lower limit of themeasurement range of the connected CompactFullRange™ CC Gauge (→ & 10, 11, 12, 13).
The unit displays « » and the green lampon the gauge is lit. The measured value is lowerthan 5×10-4 mbar and the gas discharge of theconnected Compact FullRange™ CC Gauge hasnot ignited yet.
The unit displays « ». The pressure in the vac-uum system is higher than the upper limit of themeasurement range of the connected CompactFullRange™ CC Gauge.
The unit displays « ». No gauge is con-nected (at the <sensor 1> connector).
The unit displays « », « » or « ».In addition, the «err» lamp flashes. The CompactFullRange™ CC Gauge has an error (→ 2 72).
The unit displays « ». In addition, the «err»lamp flashes. You have turned the unit back on tooquickly after turning it off (→ 2 24).
Turning the unit on
22 Operation BG 805 096 BE (9907) DualGauge.om
IMR
Sensor 1Process IonGauge
Sensor X
The unit displays « ». A connected Pirani/highpressure gauge is not turned on (→ 2 25).
The unit displays, for instance, « » (1×10-2) or
« » ( floating point format). The measured valueis within the Pirani measurement range of thegauge.
The unit displays, for instance, « »(2.8×10-6). The measured value is within themeasurement range of the gauge (hot ionizationmode).
The unit displays « ». The pressure in thevacuum system is lower than the lower limit of themeasurement range of the gauge in Pirani adjust-ment mode (→ & 14).
The unit displays « ».The pressure in thevacuum system is lower than the lower limit of themeasurement range of the connected Pirani/highpressure gauge (→ & 14).
The unit displays « ». The pressure in the vac-uum system is higher than the upper limit of themeasurement range of the connected gauge(Pirani mode).
The unit displays « ». No gauge is con-nected (at the <sensor 1> connector).
The unit displays « ». In addition, the «err»lamp flashes. The gauge has an error (→ 2 72).
The unit displays « ». In addition, the «err»lamp flashes. You have turned the unit back on tooquickly after turning it off (→ 2 24).
Turning the unit on
BG 805 096 BE (9907) DualGauge.om Operation 23
APR
Sensor 1PiezoGauge
or
Sensor 1CapacitanceGauge CMR
ACR
Sensor X
The unit displays, for instance, « » (floatingpoint format). The measured value is within themeasurement range of the gauge.
The unit displays « ». The pressure in the vac-uum system is lower than the lower limit of themeasurement range of the connected gauge or notwithin the offset range (→ 2 50).
The unit displays « ». The pressure in the vac-uum system is higher than the upper limit of themeasurement range of the connected linear gaugeor not within the offset range (→ 2 50).
The unit displays « ». No gauge is con-nected (at the <sensor 1> connector).
The unit displays « ». In addition, the «err» lampflashes. The linear gauge has an error (→ 2 72).
The unit displays « ». In addition, the «err»lamp flashes. You have turned the unit back on tooquickly after turning it off (→ 2 24).
Turning the unit on
24 Operation BG 805 096 BE (9907) DualGauge.om
Turn the unit off with the power switch (or centrally viathe power distributor, for rack mounted units).
turning the unitoff
Wait at least10 seconds beforeturning the unitback on.
10 seconds
turning the unitback on
Select the gauge the pressure of which you want todisplay by pressing the <sensor> key <sensor> 1)
sensor
on/offsen 1 sen 2
When the unit is in the measurement mode, you candisplay the types of gauges connected by brieflypressing all three keys together.
set point
sen 1 sen 2 sen 2sen 1
The types of gauges connected will be displayed forapprox. 1 second each in the sequence shown above.When the display has run through, the unit returns tomeasurement mode.
1) The <sensor> and < > keys are identical
4.3 Power off
Turning the unitback on
4.4 Selecting a gauge
4.5 Displaying theconnected gauges
BG 805 096 BE (9907) DualGauge.om Operation 25
« » Pirani gauge
« » Cold cathode gauge with a lower meas-urement range limit of 10-9 mbar
« » Cold cathode gauge with a lower meas-urement range limit of 10-11 mbar
« » FullRange™ CC Gauge with a lower meas-urement range limit of 10-9 mbar
« » Pirani/high pressure gauge with a lowermeasurement range limit of 10-6 mbar
« » Linear gauge
« » No gauge connected
« » Gauge not identified
This function is available for the following gauges:
Pirani gauge
æ Cold cathode gauge
Compact FullRange™ CC Gauge
æ Pirani/high pressure gauge
Linear gauge
Any attempt to turn off another gauge results in the dis-play of the type of the gauge connected (→ 2 25).
WARNING
Turning a gauge on / of may affect the statusof the relay.
Display codes andtheir meaning
4.6 Turning a gaugeon / off
26 Operation BG 805 096 BE (9907) DualGauge.om
IKRCold CathodeGauge
or IMR
Process IonGauge
When the unit is in measurement mode, each connectedcold cathode and Pirani/high pressure gauge can beturned on / off manually if the following two conditionsare met: the entry lock function is disabled (→ 2 45) and
the gauge control is set to « » (→ 2 53).
sensor
on/off
First select the gauge by pressing the<sensor> key.
sensor
on/off
Turn the gauge on / off by pressing the<on/off> key. 1) When the gauge is on,the «on 1» / or «on 2» lamp is lit.
Note
To prevent unnecessary contamination of thecold cathode gauges, turn them on at pres-sures <10
-2 mbar only.
The unit has one switching function with two adjustablethreshold values per gauge. The switching function isused for controlling the behavior of the relay output.
In the measurement mode, you can display the thresholdvalues of the switching functions for the individualgauges.
sensor
on/off
To select the gauge of which you wishto display the switching function val-ues, press the <sensor> key.
1) The <on/off> and < > keys are identical
Turning a gaugeon / off
4.7 Displaying thethreshold values
BG 805 096 BE (9907) DualGauge.om Operation 27
set point
Then press and hold down the<set point> key; the unit is now in theswitching function mode.
Each time the key is pressed, the display changes fromthe upper to the lower threshold or vice versa. The indi-cators «sp...» and «hi» or «lo») are flashing to showwhich threshold value is being displayed.
sp hi sp lo
When the <set point> key is released, the display returnsto the measurement mode.
«sp 1» «lo» lower threshold of the switching functionfor gauge 1
«sp 1» «hi» upper threshold of the switching functionfor gauge 1
«sp 2» «lo» lower threshold of the switching functionfor gauge 2
«sp 2» «hi» upper threshold of the switching functionfor gauge 2
Note
Take account of the display formats (→ 2 9).
sp1 hi
sp1 lo
p
t
measured value
upper threshold
lower threshold
Contact position ofthe relay (→ 2 61)
Switching function 1»off» »on« »off«
Display codes and theirmeaning
28 Operation BG 805 096 BE (9907) DualGauge.om
The Setpoint low (sp1 lo) defines the pressure at whichSwitching function 1 is activated when the pressure isdropping.The Setpoint high (sp1 hi) defines the pressure at whichSwitching function 1 is deactivated when the pressure isrising.
For Switching function 2 the above definitions apply byanalogy.
Note
The threshold values can only be adjustedwhen the entry lock is disabled. If the lockfunction is enabled « » is displayed whenattempting to adjust a threshold value(→ 2 45).
WARNING
If other gauges were connected previously,the threshold values may have been adjustedautomatically.Any threshold value entered that is outsidethe measurement range will be changed sothat it corresponds to the lower (upper) meas-urement range limit.If both threshold values entered are outsidethe measurement range, they will be modifiedcorrespondingly, and the minimum hysteresiswill be taken into account (→ sections 4.8.1and 4.8.2).
Definitions (sp lo),(sp hi)
4.8 Adjusting thethreshold values
BG 805 096 BE (9907) DualGauge.om Operation 29
PiraniGauge
IKRCold CathodeGauge
PKRFullRangeTM CCGauge
IMR
Process IonGauge
If you want to modify a threshold value, you must firstdisplay it (→ 2 26).
set point
additionally press:
• < > to increasethe thresholdvalue
sp para
or
• < > to decreasethe thresholdvalue
Briefly tapping the < > or < > key changes the mantissavalue by 0.1 for each tap.
If you keep pressing the key, the mantissa value scrollsautomatically in steps of 1.0.
Depending on the calibration factor entered (→ 2 48) ,the adjustment range for a particular threshold value cango above or below the measurement range for thegauge by a factor 10.
When the two keys are released, the changed thresholdis automatically stored in nonvolatile memory.
4.8.1 Logarithmicgauges
Increasing a displayedthreshold value
Decreasing a displayedthreshold value
Exponential display(threshold value <1 mbar)
30 Operation BG 805 096 BE (9907) DualGauge.om
Note
The minimum hysteresis between the lowerand upper threshold value must be a mini-mum of 10% of the lower threshold value.This prevents unstable states. If you shoulddefine the upper threshold value lower thanthe lower threshold value, the minimum hys-teresis is entered automatically.
Briefly tapping the < > or < > key changes the secondvisible digit from the left by 1 for each tap.
If you keep pressing the key, the first visible digit scrolls.
Depending on the calibration factor entered (→ 2 48),the adjustment range for a particular threshold value cango above or below the measurement range for thegauge by a factor 10.
When the two keys are released, the changed thresholdis automatically stored in nonvolatile memory.
Note
The minimum hysteresis between the lowerand upper threshold value must be a mini-mum of 10% of the lower threshold value.This prevents unstable states. If you definethe upper threshold value lower then thelower threshold value, the minimumhysteresis is automatically applied.
CapacitanceGauge CMR
ACR
PiezoGauge
APR
If you want to change a threshold value, you must firstdisplay it (→ 2 26).
Floating point display(threshold value ≥1 mbar)
4.8.2 Linear gauges
BG 805 096 BE (9907) DualGauge.om Operation 31
set point
additionally press:
• < > to increasethe thresholdvalue
sp para
or
• < > to decreasethe thresholdvalue
Briefly tapping the < > key or < > key changes thesecond digit from the right by 1 for each tap.
If you keep pressing the key, the third digit from the rightscrolls.
Depending on the calibration factor entered (→ 2 48),the adjustment range for a particular threshold value cango above or below the measurement range for thegauge by a factor 2.
When the two keys are released, the changed thresholdis automatically stored in nonvolatile memory.
Note
The minimum hysteresis between the lowerand upper threshold value must be a mini-mum of 1% of the upper threshold value (Fullscale range). This prevents unstable states. Ifyou should define the upper threshold valuelower then the lower threshold value, theminimum hysteresis is entered automatically.
Increasing a displayedthreshold value
Decreasing a displayedthreshold value
32 Operation BG 805 096 BE (9907) DualGauge.om
2 s
set point
Press the < > and < >keys for 2 seconds tochange frommeasurement to pa-rameter mode. Thisenables you –depending on theconnected gauges –to adjust the followingparameters:• pressure unit (→ 2 45)• entry lock function (→ 2 45)• transmission rate of the
RS232C interface (→ 2 46)• display-change mode (→ 2 47)• measured value filter (→ 2 47)• calibration factor (→ 2 48)• measurement range of the linear
gauge (→ 2 50)• offset correction (→ 2 52)• offset value (→ 2 52)• gauge control (→ 2 53)• gauge underrange control
(→ 2 54)
set point
Press the <set point> as often asnecessary.When you keep pressing the key,the name (Unit, LOC, etc.) for thecurrent parameter is displayed.First the parameter valid for bothgauges (Unit, LOC, bAUd) and thenthe individual parameters forgauges 1 and 2 are displayed.
4.9 Adjusting otherparameters
4.9.1 General procedure
Selecting theparameter mode
Selecting theparameters to bemodified
sp para
BG 805 096 BE (9907) DualGauge.om Operation 33
set point
or Select the possible set-tings by pressing the< > or < > key as oftenas necessary.
30 s
set point
or
2 s
Return to the measure-ment mode by:
• not pressing any keyfor 30 seconds or
• pressing the < > and< > keys together for2 seconds.
The modified parametersettings are automati-cally stored in non-volatile memory.
On the following pages you will find parameter settingsfor configurations which are often used.
Selecting thesettings
Returning to themeasurement mode
4.9.2 Setting theparameters
34 Operation BG 805 096 BE (9907) DualGauge.om
combinedwith
PiraniGauge
Sensor 2
PiraniGauge
Sensor 1
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Parameter settings
BG 805 096 BE (9907) DualGauge.om Operation 35
combinedwith
Sensor 1
IKR
Cold CathodeGauge
Sensor 2
IKR
Cold CathodeGauge
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Gauge control 1 (→ 2 53)
Gauge underrange control 1 (→ 2 54)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Gauge control 2 (→ 2 53)
Gauge underrange control (→ 2 54)
Parameter settings
36 Operation BG 805 096 BE (9907) DualGauge.om
combinedwith
Sensor 1
PKR
FullRange™ CCGauge
Sensor 2
PKR
FullRange™ CCGauge
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Parameter settings
BG 805 096 BE (9907) DualGauge.om Operation 37
combinedwith
IMR
Process IonGauge
Sensor 1
IMR
Process IonGauge
Sensor 2
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Gauge control 1 (→ 2 53)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Gauge control 2 (→ 2 53)
Parameter settings
38 Operation BG 805 096 BE (9907) DualGauge.om
combinedwith
CapacitanceGauge
CMR
ACR
Sensor 1
APR
PiezoGauge
Sensor 2
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Measurement range gauge 1 (→ 2 50)
Offset correction gauge 1 (→ 2 52)
Offset value gauge 1 (→ 2 52)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Measurement range gauge 2 (→ 2 50)
Offset correction gauge 2 (→ 2 52)
Offset value gauge 2 (→ 2 52)
Parameter settings
BG 805 096 BE (9907) DualGauge.om Operation 39
combinedwith
PiraniGauge
Sensor 1 Sensor 2
IKR
Cold CathodeGauge
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Gauge control 2 (→ 2 53)
Gauge underrange control 2 (→ 2 54)
Parameter settings
40 Operation BG 805 096 BE (9907) DualGauge.om
combinedwith
PiraniGauge
Sensor 1 Sensor 2
PKR
FullRange™ CCGauge
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Parameter setting
BG 805 096 BE (9907) DualGauge.om Operation 41
Sensor 2
PKR
FullRange™ CCGauge
combinedwith
Sensor 1
IKR
Cold CathodeGauge
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Gauge control 1 (→ 2 53)
Gauge underrange control 1 (→ 2 54)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Parameter settings
42 Operation BG 805 096 BE (9907) DualGauge.om
CapacitanceGauge
CMR
ACR
Sensor 1
APR
PiezoGauge
Sensor 1
combinedwith
or
PiraniGauge
Sensor 2
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Measurement range gauge 1 (→ 2 50)
Offset correction gauge 1 (→ 2 52)
Offset value gauge 1 (→ 2 52)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Parameter settings
BG 805 096 BE (9907) DualGauge.om Operation 43
Sensor 2
IKR
Cold CathodeGauge
CapacitanceGauge
CMR
ACR
Sensor 1
APR
PiezoGauge
Sensor 1
combinedwith
or
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Measurement range gauge 1 (→ 2 50)
Offset correction gauge 1 (→ 2 52)
Offset value gauge 1 (→ 2 52)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Gauge control 2 (→ 2 53)
Gauge underrange control 2 (→ 2 54)
Parameter settings
44 Operation BG 805 096 BE (9907) DualGauge.om
CapacitanceGauge
CMR
ACR
Sensor 1
APR
PiezoGauge
Sensor 1
combinedwith
or
Sensor 2
PKR
FullRange™ CCGauge
Pressure unit (→ 2 45)
Entry lock function (→ 2 45)
Transmission rate setting (→ 2 46)
Filter time constant gauge 1 (→ 2 47)
Calibration factor gauge 1 (→ 2 48)
Measurement range gauge 1 (→ 2 50)
Offset correction gauge 1 (→ 2 52)
Offset value gauge 1 (→ 2 52)
Filter time constant gauge 2 (→ 2 47)
Calibration factor gauge 2 (→ 2 48)
Parameter settings
BG 805 096 BE (9907) DualGauge.om Operation 45
Changing the pressure unit affects:æ the displayæ the switching functions (threshold display)o the analog outputs of the unitso RS232C interface (display always in mbar)
mbar / bar
Torr
Pascal
The current pressure unit setting is indicated by the lamp«mbar», «Torr» or «Pa».
The selection of pressure units depends on the gaugesused. The DualGauge™ allows for a particular pressureunit only, if the entire measurement range can be dis-played in that pressure unit.
Measurement range mbar /bar
Torr Pascal
10-11 mbar ... 1000 mbar æ æ æ
Full Scale value(defined measurement range)
mbar /bar
Torr Pascal
1 mbar æ æ æ
10 mbar æ æ æ
100 mbar æ æ
1000 mbar æ æ
2 mbar æ æ
5 mbar æ æ
10 mbar æ
50 mbar æ
Conversion table → 2 86
The entry lock function has been included to prevent in-advertent entries.
4.9.3 Selecting thepressure unit
Display range fordifferent pressureunits
Logarithmic gauges
Linear gauges
4.9.4 Enabling the entrylock function
46 Operation BG 805 096 BE (9907) DualGauge.om
Entry lock function disabled:
All operating steps are possible.
Entry lock function enabled:
The following operating steps are notpossible:
• Adjusting the threshold values
• Turning a gauge on / off
When the entry lock function is enabled, « » is dis-played when the threshold value is changed or a gaugeis turned on / off.
The DualGauge™ can be controlled via the (RS232C)serial interface. It also offers the possibility of transfer-ring measurement data and changing parameters.
The transmission rate can be setbetween 300 and 19200 baud.
Communication protocolDualGauge™ → & 18.
4.9.5 Setting thetransmission rate
BG 805 096 BE (9907) DualGauge.om Operation 47
The automatic display-change mode («sen 1» ⇒«sen 2» or «sen 2» ⇒ «sen 1») is available for thecombination Compact Pirani gauge and linear gauge(1000 mbar Full Scale) only.
Automatic display-change when themeasured value of the linear gaugerises above or drops below 10 mbar
Manual display-change
Combination Pirani - linear gauge (1000 mbar)(automatic display change):
5×10-4 mbar 10 mbar 1000 mbar
Pirani gauge Linear gauge
The measured value filter allows for a better evaluationof unstable or disturbed measurement signals. It affects:æ the displayæ the switching functionso the analog outputs of the unitæ the RS232C interface
Normal filter time(320 ms)
Slow filter time(2.6 s)
Fast filter time(35 ms)
4.9.6 Selecting thedisplay changemode
4.9.7 Adjusting themeasured valuefilter
48 Operation BG 805 096 BE (9907) DualGauge.om
t
p Standard setting with a good relation-ship between the reaction time andthe sensibility of display and switchingfunctions with regard to changes inthe measured value.
t
p Select this setting if you wish that thedisplay an the switching functions donot react to small variations of themeasured value. The unit will, how-ever, react slower to changes in themeasured value.
t
p Select this setting if you wish that thedisplay and the switching functionsreact fast to changes in the measuredvalue. The unit will, however, be moresensitive to disturbances with regardto the measured values.
The filter for gauge 2 (« ») is adjusted in the sameway.
The calibration factor allows the measured value to becalibrated for gases other than N2. It affects:æ the displayæ the switching functions (threshold display) 1)
o the analog outputs of the unitæ the RS232C interface 1)
WARNING
The status indications « » and « » aswell as the error messages « » are outputaccording to the pressure unit and calibrationfactor settings. The indications in this docu-ment refer to the pressure unit mbar and acalibration factor of 1.00.
1) For Compact Process Ion Gauges in the hot cathode
measurement range only
4.9.8 Adjusting thecalibration factor
BG 805 096 BE (9907) DualGauge.om Operation 49
max. calibrationfactor (×9.99)
min. calibrationfactor (×0.10)
Briefly tapping the < > or < > key changes the value ofthe mantissa by 0.01 for each tap.
If you keep pressing the key, the value scrolls automati-cally in steps of 0.1.
If the calibration factor does not equal 1.00, the «cal1»and / or the «cal2» lamp is lit.
The calibration factor for gauge 2 (« ») is ad-justed in the same way.
max. calibrationfactor (×2.000)
min. calibrationfactor (×0.500)
Briefly tapping the < > or < > key changes the value ofthe mantissa by 0.001 for each tap.
If you keep pressing the key, the value scrolls automati-cally in steps of 0.01.
If the calibration factor does not equal 1.000 the «cal1»and / or «cal2» lamp is lit.
The calibration factor for gauge 2 (« ») is adjustedin the same way.
4.9.8.1 Logarithmicgauges
4.9.8.2 Linear gauges
50 Operation BG 805 096 BE (9907) DualGauge.om
In order for the correct measured value of a linear gaugeto be displayed it is necessary to save the full scalevalue of the measurement range of that gauge as para-meter. The measurement range is adjusted by indicatingthe upper limit (full scale) value.
«mbar»
«mbar»
«mbar»
«mbar»
«bar»
«bar»
«bar»
«bar»
The offset function allows for zeroing to the value beingmeasured within -5 ... +110 % of the set full scale value.
The offset function affects:æ the displayæ the switching functions (threshold display)o the analog outputs of the unitæ the RS232C interface
This function can be used for two different purposes:
There are two methods for adjusting the zero of a lineargauge. Note, however, that the actual pressure must belower than the lower limit of the measurement range ofthe gauge:
− Set the zero by adjusting the „ZERO“ potentiometerof the gauge → & 15, 16, 17
− With the offset function of the measurement andcontrol unit set the current pressure reading to zero
4.9.9 Adjusting themeasurementrange of a lineargauge
4.9.10 Offset function(zero) of a lineargauge
Zero adjustment
BG 805 096 BE (9907) DualGauge.om Operation 51
The advantage of the second method is that no directaccess to the potentiometer of the gauge is required.
The pressure reading of the measurement and controlunit can be set to zero at any pressure within the meas-urement range of a linear gauge. All subsequent read-ings will then be relative to that offset value and maytherefore be positive or negative. This method allows formonitoring of pressure variations during a process.
Zeroing at anypressure
52 Operation BG 805 096 BE (9907) DualGauge.om
Action Result
Select « » menu(→ 2 32)
«cal1»(Last offset value saved)
set point
2 s
or
set point
2 s
«cal1»The value being measured(without offset) is acceptedas new offset value.
The new offset value issaved by returning to themeasurement mode(→ 2 32).
The offset value of gauge 2 is adjusted in the same way.
The offset correctionfunction is enabled
The offset correctionfunction is disabled
With the offset correction function enabled, the savedoffset value is subtracted from the value being meas-ured.
4.9.10.1 Displaying andadjusting theoffset value
4.9.10.2 Enabling theoffset correctionfunction of a lineargauge
BG 805 096 BE (9907) DualGauge.om Operation 53
Measuredvalue
Offset valuesaved
Displaywhen
=
Displaywhen
=
« » « » « » « »
« » « » « » « »
« » « » « » « »
WARNING
For readjusting the zero point of the gauge,set to .
WARNING
Turning a gauge on / of may affect the statusof the relay.
Controlling gaugeæ Pirani gaugeo Cold cathode gaugeæ Compact FullRange™ CC Gaugeo Pirani/high pressure gaugeæ Linear gauge
Controlled gaugeo Pirani gaugeæ Cold cathode gaugeo Compact FullRange™ CC Gaugeæ Pirani/high pressure gaugeo Linear gauge
IKRCold CathodeGauge
orIMR
Process IonGauge
Example
4.9.11 Adjusting thegauge control or
Controllinggauge Ø
54 Operation BG 805 096 BE (9907) DualGauge.om
automatic control byfixed threshold values(→ 2 58)
external control via the<control> connector(→ 2 15, 56)
manual control with the<on/off> key
WARNING
Turning a gauge on / of may affect the statusof the relay.
This function is available for the following gauge:o Pirani gaugeæ Cold cathode gaugeo Compact FullRange™ CC Gaugeo Pirani/high pressure gaugeo Linear gauge
IKRCold CathodeGauge
There is a number of possible causes of an underrange(e.g. « »):æ the pressure in the vacuum system is lower than the
lower limit of the measurement rangeæ the measurement element has not ignited (yet)æ the discharge has failedæ a defect has occurred
An underrange state is interpretedas an admissible measured value.« » is displayed. The switch-ing function remains »on«.An underrange state is interpretedas an inadmissible measured value.
4.9.12 Adjusting thegauge underrangecontrol
BG 805 096 BE (9907) DualGauge.om Operation 55
« » is displayed. The switch-ing function changes to »off«.
Note
In the « » setting, the evaluation of theswitching function is suppressed for approx.10 seconds when the cold cathode gauge isturned on and each time after an underrangehas occurred. During this time, the switchingfunction remains »off«.
If chances are that the pressure in the vacuumchamber is under the lower limit of the measure-ment range of the gauge, it is advisable to select« ».
The underrange control of cold cathode gauge 2(« ») is adjusted in the same way.
56 Technical data BG 805 096 BE (9907) DualGauge.om
5 Technical data
Admissible temperaturestorageoperation
–40 °C … +65 °C+ 5 °C … +50 °C
Relative humidity max. 80 % at temperaturesup to +31 °C, decreasing to50 % at +40 °C
Use indoors onlyaltitude up to 2000 m
Number of gauges 2
Note
Since the IMR 260 and CMR 27x have ahigher power consumption than other gauges,only one of these types can be connectedand operated in combination with any othergauge type. 1)
TypesCompact Pirani Gauges TPR 250, TPR 260,
TPR 261, TPR 265Compact Cold CathodeGauges
IKR 250, IKR 251,IKR 260, IKR 261, IKR 270
Compact FullRange™ CCGauges
PKR 250, PKR 251,PKR 260, PKR 261
Compact Process IonGauges
IMR 260
Compact CapacitanceGauges
CMR 261, CMR 262,CMR 263, CMR 264,CMR 271, CMR 272,CMR 273, CMR 274,ACR 261, ACR 262,ACR 263, ACR 274
1) Two IMR 260 gauges can be connected. The elec-
tronics ensure that only one of them is operated at atime.
Ambient conditions
Compatible gauges
BG 805 096 BE (9907) DualGauge.om Technical data 57
Compact Piezo Gauges APR 250, APR 260,APR 261, APR 265,APR 266, APR 267
Measurement range,measurement error,reproducibility
depending on the gaugeused (→ 2 1 … 17)
Measurement error ofmeasurement unitamplification error
offset error
≤ 0.2 % of the current inputvoltage≤ 20 mV
Measurement rate 100 per sDisplay rate 4 per sFilter time constants
normalfastslow
320 ms (fg = 0.5 Hz) 35 ms (fg = 4.5 Hz) 2.6 ms (fg = 0.06 Hz)
Number of switchingfunctions
1 per gauge
Type of contact isolated changeover con-tact (contact open ⇒switching function »off« ormeasurement unit turnedoff)
Maximum load 240 V∼; 5 A; 1200 VA200 V=; 5 A; 40 W
DANGER
If the DualGauge™ is used as desktop unit, amaximum of 30 V∼ or 60 V= may be applied(protective low voltage).
Uptimemechanicalelectrical
5×107 cycles1×105 cycles (at maximumload)
Reaction time ≤15 ms, if the thresholdvalue is close to the meas-ured value (at higher dif-ferences, take the filtertime constant into con-sideration)
Measurement
Switching functions
58 Technical data BG 805 096 BE (9907) DualGauge.om
Error signal (error contact)Type of contact potential free normally
open (contact open ⇒error or measurement unitturned off
Maximum load 50 V∼; 1.5 A50 V=; 1.5 A; 40 W
Uptimemechanicalelectrical
5×107 cycles1×105 cycles (at maximumload)
Voltage 20.5 … 30 V unstabilizedripple max. 1 Vpp
Current ≤ 200 mA per gauge pro-tected with PTC element,self-resetting at power-off
manually with <on/off> keyautomatically
cold cathode turned oncold cathode turned off
with Pirani gauge, Full-Range™ CC Gauge, lineargauge (1 mbar Full Scale)at p ≤ 6.0×10-3 mbarat p ≥ 8.0×10-3 mbar
externally at <control> connector; in-ternal pull-up 3.3 kΩ to 5 Vcontrol via- TTL signal- open collector- relay contacthigh: ≥ 2.0 V or input open⇒ gauge offlow: ≤ 0.8 V⇒ gauge on
Error signal
Gauge supply
Turning the coldcathode gauge on / off
BG 805 096 BE (9907) DualGauge.om Technical data 59
manually with <on/off> keyautomatically
hot cathode onhot cathode off
with Pirani gauge, Full-Range™ CC Gauge, lineargauge (10 mbar, 100 mbarFull Scale)p ≤ 1.0×10-1 mbarp ≥ 1.5×10-1 mbar
externally at <control> connector; in-ternal pull-up 3.3 kΩ to 5 Vcontrol via- TTL signal- open collector- relay contacthigh: ≥2.0 V or input open⇒ gauge offlow: ≤0.8 V⇒ gauge on
Voltage range 0 V … +10 V
Internal resistance Ri 660 ΩRelationship betweenmeasrement signal andpessure
(→ 2 1 … 17)
Safety
protection classcontamination degreeEMC
EN61010-1 : 1993(IEC 1010)IIIemission EN 50081-1immunity EN 50082-2
Turning the Pirani/highpressure gauge on / off
Analog output signal
Standards applied
60 Technical data BG 805 096 BE (9907) DualGauge.om
Voltage 230 V∼, +10 %, -14 %208 V∼, +10 %, -14 %120 V∼, +10 %, -14 %100 V∼, ±10 %
Frequency 50 … 60 Hz
Power consumption 17 VAOvervoltage category IIFuses F1, F2 incorporated in the unit
230 V∼: 0.1 AT, 250 V;ø5×20 mm
208 V∼: 0.1 AT, 250 V;ø5×20 mm
120 V∼: 0.2 AT, 250 V;ø5×20 mm
100 V∼: 0.2 AT, 250 V;ø5×20 mm
Minimum waiting timebetween turning the unitoff and back on
10 seconds12
8.5
70.8
231
As desktop unit or for rack installation
Weight 1.4 kg
Power specifications
Waiting time
Dimensions
Use
Weight
BG 805 096 BE (9907) DualGauge.om Technical data 61
Skilled personnel
Cables may only be made and adapted bypersons with suitable training in electro-technics.
DANGER
Take account of the Technical data to preventany hazard, equipment damage or overload.
VHQVRU
56 VHQVRU
FRQWURO
IXVHLQVLGH
V
V
Pin assignments
Rear panel
Mains power connector
62 Technical data BG 805 096 BE (9907) DualGauge.om
Amphenol C91B, 6-pole, female
Pin assignmentPin 1: IdentificationPin 2: GNDPin 3: Measurement signal +Pin 4: Measurement signal –Pin 5: Screen 1)
Pin 6: Vcc
45
1 2
36
View to DualGauge™
Amphenol C91B, 6-pole, female
Pin assignment:see <sensor 1>
45
1 2
36
View to DualGauge™
Amphenol C91B, 7-pole, female
Pin assignmentPin 1: Analog output gauge 2Pin 2: Analog output gauge 1Pin 3: Error (Error)Pin 4: Gauge 1 on/offPin 5: GND / screen 1)
Pin 6: Gauge 2 on/offPin 7: Error (Error)
4
5
1
2
3
6
7
n.c.
View to DualGauge™
Phoenix Combicon, 6-pole, male
View to DualGauge™
Note
The contacts are shown in idle position.Switching function = »off«.
1) Use screened cables only (electromagneticcompatibility).
<sensor 1> connector
<sensor 2> connector
<control> connector
Relay connector
s1Gauge 1
s2Gauge 2
BG 805 096 BE (9907) DualGauge.om Technical data 63
RS232C, D-Sub, 9-pole, male
Pin assignment:Pin 1: not assignedPin 2: RXDPin 3: TXDPin 4: DTRPin 5: SGNDPin 6: not assignedPin 7: RTSPin 8: CTSPin 9: not assignedHousing: screen 1)
95
61
View to DualGauge™
1) Use screened cables only (electromagneticcompatibility).
Serial connector
64 Technical data BG 805 096 BE (9907) DualGauge.om
External control of the cold cathode or Pirani/highpressure gauge.
WARNING
Pin 5 of the DualGauge™ is directly con-nected to protective ground. Therefore, pro-tective measures must be taken to protect theunit from earth currents.
a) b)
5
6
4
5
6
4
c)
5
6 TTL
4 TTL
5V 5V
Evaluation of the analog output voltage
5V
+
–
2
V+
–
1
Ri ≥ 100 kΩ
Evaluation of the Error signal
5V
3
7
Details with regard to the<control> connector
BG 805 096 BE (9907) DualGauge.om Maintenance and care 65
6 Maintenance and care
The DualGauge™ Vacuum measurement and controlunit requires no maintenance. For maintenance of thegauges please refer to the corresponding documents1 … 17.
WARNING
Turn the unit off and remove all cables (themains cable at last), before doing any of thework described below.
For cleaning the outside of the unit, a slightly humid clothwill usually do. Do not use under any circumstances anaggressive or scouring cleaning agent. Before puttingthe unit into operation again, allow it to dry thoroughly.
Skilled personnel
In a very dusty environment, the dust insidethe unit should be periodically removed(Opening the unit → Appendix A). Carefullyremove the dust with dry compressed air.
Cleaning the outside
Cleaning the inside
66 Status indications BG 805 096 BE (9907) DualGauge.om
7 Status indications
err
«err» lamp dark(Error contact = »closed«)
« »Specification of the statusmessage on the display.
Note
Status messages are for information only.They do not release any error messages. Fordetails regarding gauge specific indications→ & 1 … 17.
WARNING
The status indications « » and « » aswell as the error messages « » are outputaccording to the pressure unit and calibrationfactor setting. The indications in this docu-ment refer to the pressure unit mbar and acalibration factor of 1.00.
Display Possible cause Correction
« » isdisplayedwhen the<on/off> keyis pressed.
In automatic gaugecontrol mode, youhave attempted toswitch the coldcathode gauge orthe Pirani/highpressure gauge onor off.
Gauge control(→ 2 53).
« » isdisplayedwhen the<on/off> keyis pressed.
In external gaugecontrol mode, youhave attempted toswitch the coldcathode gauge orthe Pirani/highpressure gauge onor off.
Gauge control(→ 2 53).
Visualization of statusmessages
Status indications
BG 805 096 BE (9907) DualGauge.om Status indications 67
Display Possible cause Correction
« » isdisplayedwhen the<on/off> keyis pressed.
You have attempt-ed to switch thecold cathode gaugeor the Pirani/highpressure gauge onor off with the entrylock functionenabled.
Entry lock function(→ 2 45).
« » isdisplayedwhen the<set point>key ispressed to-gether withthe < > or< > key.
You have attempt-ed to change thethreshold value withthe entry lock func-tion enabled.
Entry lock function(→ 2 45).
« » Gauge identificationerror.
Check the corre-sponding connec-tion cable.Use an appropriategauge (→ 2 56).
« » No gaugeconnected
Switch the unit off,connect a gauge,and switch the uniton again. (The unitcan also be oper-ated with only onegauge connected).
« » isdisplayedwhen the<on/off> keyis pressed.
You have attempt-ed to turn off theFullRange™ CCGauge.
The FullRange™CC Gauge cannotbe turned off.
« » isdisplayedwhen the<on/off> keyis pressed.
You have attempt-ed to turn off alinear gauge.
Linear gaugescannot be turnedoff.
« » isdisplayedwhen the<on/off> keyis pressed.
You have attempt-ed to turn off thePirani gauge.
The Pirani gaugecannot be turnedoff.
« » Linear gauge:The measured –
68 Status indications BG 805 096 BE (9907) DualGauge.om
Display Possible cause Correctionvalue is under thelower limit of themeasurementrange of the con-nected gauge.
(Standard casewhen used for highvacuum measure-ment)
Incorrect HVadjustment
Adjustment→ & 15, 16, 17
« » Pirani/highpressure gauge:Measurementresult below lowerlimit of Piranimeasurementrange.
Quit Pirani adjust-ment mode of thegauge → & 14.
« » Pirani:Measurementresult below meas-urement range ofconnected gauge.
–(Standard casewhen used for highvacuum measure-ment)
Pirani:Incorrect HVadjustment
Adjustment→ & 1, 2, 3, 4
FullRange™ CCGauge:The gas dischargehas not yet ignited.
Wait until the gasdischarge ignites→ & 10, 11, 12,13.
« » Pirani/high pres-sure gauge:Measurement resultbelow lower limit ofmeasurementrange.
–
« » FullRange™ CCGauge:Measurement resultbelow lower limit ofmeasurementrange.
–(Standard casewhen used for ultrahigh vacuummeasurement).
Cold cathodegauge:Measurement resultbelow lower limit of
(Adjusting thegauge underrange
BG 805 096 BE (9907) DualGauge.om Status indications 69
Display Possible cause Correctionmeasurementrange.
control → 2 54 ff)
The gas dischargemay not have ig-nited yet.
Wait until the gasdischarge ignites→ & 5, 6, 7, 8
Gauge severelycontaminated(isolating layers).
Cleaning the gauge→ & 5, 6, 7, 8
« » Cold cathodegauge:Measurementresult below lowerlimit of measure-ment range.
-(Standard casewhen used for ultrahigh vacuummeasurement)(Adjusting thegauge underrangecontrol → 2 54 ff)
The gas dischargemay not haveignited yet.
Wait until the gasdischarge ignites→ & 9.
Gauge severelycontaminated(isolating layers).
Cleaning the gauge→ & 9.
« » Linear gauge:Measurement resultabove the upperlimit of themeasurementrange.
-
Pirani / Pirani/highpressure gauge /FullRange™ CCGauge:Measurement resultabove upper limit ofPiranimeasurementrange.
-
Incorrect ATMadjustment
(Adjustment→ & 1, 2, 3, 4, 10,11, 12, 13, 14)
70 Status indications BG 805 096 BE (9907) DualGauge.om
Display Possible cause Correction
« » Cold cathodegauge:Measurementrange above upperlimit of measure-ment range.
Turn the cold ca-thode gauge off (riskof contamination).Automatic controlwith a Pirani gauge.
BG 805 096 BE (9907) DualGauge.om Error messages 71
8 Error messages
err
«err» lamp flashing(Error contact = »open«)
« »Specification of the errormessage on the display.
WARNING
The status indications « » and « » aswell as the error messages « » are outputaccording to the pressure unit and calibrationfactor setting. The indications in this docu-ment refer to the pressure unit mbar and acalibration factor of 1.00.
set point
After you have acknowledged the errormessage:
• the unit returns to the measurementmode, if the error has been cor-rected
• the error is displayed again if it hasnot been corrected
• the next error is displayed, if thereare any further errors.
Pfeiffer-Vacuum
If you cannot correct the error yourself,please contact your nearestPfeiffer Vacuum Service Center.
Display Possible cause Acknowledge-ment / Correction
No status indica-tion when theunit is turned on.
Power cable inter-rupted.
Check the powercable.
Fuse defective. Replace the fuse(→ 2 77).
The voltage ratingof the unit does notagree with thelocal line voltage.
Adapt the equip-ment voltage(→ 2 77).
Visualization of errormessages
Acknowledging of errormessages
72 Error messages BG 805 096 BE (9907) DualGauge.om
Display Possible cause Acknowledge-ment / Correction
« » Pirani:No power supply,measurementelement defective
Check supplyand cable.Service gauge orreplace it→ & 1, 2, 3, 4.
FullRange™ CCGauge:No power supply. Check supply
and cable.
« » Pirani/highpressure gauge:No power supply,Pirani measure-ment elementdefective, fila-ment defective.
Check supplyand cable.Service gauge orreplace it→ & 14.
« » Cold cathodegauge:No power supply. Check supply
and cable.
« » Cold cathodegauge:No power supply. Check supply
and cable.
« » FullRange™ CCGauge:Pirani measure-ment elementdefective.
Service gauge orreplace it→ & 10, 11, 12,13.
Linear gauge:No power supply.
Check supplyand cable.
BG 805 096 BE (9907) DualGauge.om Error messages 73
Display Possible cause Acknowledge-ment / Correction
« »
« »
Line break in theconnection cableto the correspond-ing gauge or theconnection cableto the gauge hasbeen discon-nected duringoperation
Check the corre-sponding con-nection cable.
set point
If the error hasnot been cor-rected:
« » or« »
« » Severe electricdisturbance orerror of the ope-rating system(watchdog error).
set point 1)
You have turnedthe unit back ontoo quickly afterturning it off.
set point
« » EEPROM error(parametermemory)
set point
« » EPROM error(programmemory)
set point
« » RAM error(random accessmemory)
set point
« » Keyboard errorset point
« » Operating systemerror (Stackoverflow error)
set point
« » Operating systemerror (task Failerror)
set point
« » Operating systemerror (Idle error) set point
1) The error message is automatically acknowledged
when this option is selected (→ 2 82).
74 Accessories BG 805 096 BE (9907) DualGauge.om
9 Accessories
Orderingnumber
19" rack frame DIN 41 494 PT 544 083 -T
Cable for connection toCompact Gauge 3 m complete 6 m complete10 m completeOther lengths on request
PT 448 250 -TPT 448 251 -TPT 448 252 -T
Pos. 6Pos. 1
Pos. 7, 8
Orderingnumber
Measurement cable forCompact GaugePos. 1 Cable, 5 poles plus screen,
0.25 mm2 conductorCable, 5 poles plus screen,0.34 mm2 conductor
Pos. 6 Socket Hirschmann GO 6 WF,6-pole, angular, female
Pos. 7 Connector Amphenol C91B,6-pole, male
Pos. 8 Crimp contact(6 pieces required)
B 4590 198 BE
B 4590 198 CD
B 4707 283 MA
B 4722 126 CC
B 4722 841 CA
BG 805 096 BE (9907) DualGauge.om Spare parts 75
10 Spare parts
When ordering spare parts, always indicate:
• all information on the product nameplate
• description and ordering number according to thespare parts list
Ordering number
Fuses F1, F20.1 AT, 250 V, ø5×20 mm0.2 AT, 250 V, ø5×20 mm
B 4666 416B 4666 422
Connector Amphenol C91B,7-pole, male <control>
Connector Phoenix Combicon,6-pole, female <s1>,<s2>
Cable housing forPhoenix Combicon <s1>,<s2>
B 4722 107 CC
B 4717 900 WE
B 4717 900 3E
76 Disposal BG 805 096 BE (9907) DualGauge.om
11 Disposal
DANGER
Caution: mains voltageTurn the unit off and disconnect all cables(the mains power cable last), before dis-assembling the unit.
N
WARNING
Caution: substances detrimental to the envi-ronmentProducts, operating materials etc. may re-quire disposal in accordance with specialregulations.Dispose of such products in accordance withthe relevant local regulations.
After disassembling the product, separate its compo-nents and according to their materials and have themrecycled.
Separating thecomponents
BG 805 096 BE (9907) DualGauge.om Appendix 77
Appendix
DANGER
Disconnect all cables (the power cable last)before opening the unit.
• Screw driver No. 3
• Screwdriver for recessed-head screws No. 1
• Possibly two fuses (ratings → 2 60)
Remove the cover (eight screws altogether).
A: Adapting theequipment voltage,exchanging thefuses
Material / tools required
Procedure
78 Appendix BG 805 096 BE (9907) DualGauge.om
• Set the jumper to the desired voltage.
• Check the two fuses F1, F2 and exchange them ifnecessary (ratings → 2 60).
• Change the voltage indication on the nameplate onthe left side of the unit
230 V208 V120 V100 V
F1, F2
• Exchange the two fuses F1, F2 (ratings → 2 60)
• Place the cover.
• Fasten it with the eight screws.
DANGER
In order to ensure proper ground connec-tion, use the two lock washers when re-installing the back panel. Use the originalscrews only.
Adapting theequipment voltage
Exchanging the fuses
Placing the equipmentcover
BG 805 096 BE (9907) DualGauge.om Appendix 79
DANGER
According to EN 61010, voltages above30 V~ or 60 V= present a hazard of contact.
Skilled personnel
If such voltages are supplied to the <s1> or<s2> connector, the following instructionsmust be adhered to under any circumstances.
• The relevant rules and regulations for therespective application must be respected.In particular, conductors, isolation andfuses must be dimensioned correctly andmounted professionally.
• Depending on the voltage supplied, cablesproviding a double isolation must be used.
• The supplied connector is admitted for250 V~. The enclosed cable housing mustbe used; however, it does not providedouble isolation.
• Therefore, the current-carrying connectormust be protected against accidental con-tact. This is achieved by installing the unitin such a way that access to its back side(<s1>, <s2> connectors) is only possible bymeans of a tool or a key.
• The plug must be screwed onto the socketof the unit.
• The protective cover (e.g. door of theswitch cabinet) must be marked with acorresponding warning sign.
• Before performing any work at the <s1>,<s2> connectors, make sure it is idle.
• A voltage presenting a hazard of contactand a protective low voltage must not beapplied to the <s1>, <s2> connectors at thesame time.
• The current-carrying connector must not beaccessible when withdrawn.
B: Dangerousvoltages at the<s1>, <s2>connector
80 Appendix BG 805 096 BE (9907) DualGauge.om
Action Keys Result
while turningthe unit on until« » isdisplayed.
Unit in test mode(Test è)
2 s
(or turn the unitoff)
Unit returns tomeasurementmode.
set point
èØØØ…
è Display pro-gram number
(light writing)
Display test Start or stoptest Ø
RAM test Start test In progress
Passed
Error
EPROM test Start test In progress
Passed
Error
EEPROM test Start test PassedError
C: Test programs
Access to test mode
Quitting the test mode
Next test
BG 805 096 BE (9907) DualGauge.om Appendix 81
ò A/D convertertest
Start test Measuredvalue 1)
Start test Measuredvalue 1)
Start test Measuredvalue 1)
Start test Measuredvalue 1)
ó Relay or I/Otest The relays switch irrespec-
tive of the pressure. Re-move any cables that maybe connected to <s1> or<s2>.
Take the necessary safetymeasures when using thisfunction for checking thecurrent circuits connected.(See also Appendix B).
Start / stop test ...The 5 relayscyclically switchON / OFF twice.
Switching contact<sensor 1> = ON
Switching contact<sensor 1> = OFF
Switching contact<sensor 2> = ON
Switching contact<sensor 2> = OFF
Error contactclosed (no error)
Error contact open(error)
1) Displayed figure × 5 = voltage at the AD current
converter (mV)
82 Appendix BG 805 096 BE (9907) DualGauge.om
Supply contact<sensor 1> = ON
Supply contact<sensor 1> = OFF
Supply contact<sensor 2> = ON
Supply contact<sensor 2> = OFF
INPUT PE-CTRL 1 HIGH level(≥2.0 V) «sp1»
LOW level(≤0.8 V) «sp1»
INPUT PE-CTRL 1 HIGH level(≥2.0 V) «sp2»
LOW level(≤0.8 V) «sp2»
õ Watchdogerror control
In case of Error: err
Error contact =»open«Acknowledgement:
set point
Error contact =»open«Automatic acknow-ledgment of thedisplay;Error contact =»closed«
BG 805 096 BE (9907) DualGauge.om Appendix 83
& [1] www.bi.balzers.com/library/Operating manualCompact Pirani Gauge TPR 250BG 803 005 BE / DBalzers Instruments, FL–9496 Balzers,Liechtenstein
& [2] www.bi.balzers.com/library/Operating manualCompact Pirani Gauge TPR 260 / 261BG 803 015 BE / ABalzers Instruments, FL–9496 Balzers,Liechtenstein
& [3] www. pfeiffer-vacuum.deOperating manualCompact Pirani Gauge TPR 261BG 805 175 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
& [4] www.pfeiffer-vacuum.deOperating manualCompact Pirani Gauge TPR 265BG 805 174 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
& [5] www.bi.balzers.com/library/Operating manualCompact Cold Cathode Gauge IKR 250BG 803 006 BE / EBalzers Instruments, FL–9496 Balzers,Liechtenstein
& [6] www.pfeiffer-vacuum.deOperating manualCompact Cold Cathode Gauge IKR 251BG 803 151 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
& [7] www.bi.balzers.com/library/Operating manualCompact Cold Cathode Gauge IKR 260BG 803 007 BE / EBalzers Instruments, FL–9496 Balzers,Liechtenstein
D: Literature
84 Appendix BG 805 096 BE (9907) DualGauge.om
& [8] www.pfeiffer-vacuum.deOperating manualCompact Cold Cathode Gauge IKR 261BG 805 153 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
& [9] www.pfeiffer-vacuum.deOperating manualCompact Cold Cathode Gauge IKR 270BG 805 008 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
& [10] www.bi.balzers.com/library/Operating manualCompact FullRange™ Gauge PKR 250BG 803 016 BE / CBalzers Instruments, FL–9496 Balzers,Liechtenstein
& [11] www.pfeiffer-vacuum.deOperating manualCompact FullRange™ CC Gauge PKR 251BG 805 155 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
& [12] www.bi.balzers.com/library/Operating manualCompact FullRange™ Gauge PKR 260BG 803 017 BE / ABalzers Instruments, FL–9496 Balzers,Liechtenstein
& [13] www.pfeiffer-vacuum.deOperating manualCompact FullRange™ CC Gauge PKR 261BG 805 157 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
& [14] www.pfeiffer-vacuum.deOperating manualCompact Process Ion Gauge IMR 260BG 805 038 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
BG 805 096 BE (9907) DualGauge.om Appendix 85
& [15] Operating manualCompact Capacitance Gauge ACR 261,ACR 262, ACR 263, ACR 274BG 803 011 BN / BBalzers Instruments, FL–9496 Balzers,Liechtenstein
& [16] www.pfeiffer-vacuum.deOperating manualCompact Capacitance Gauge CMR 261,CMR 262, CMR 263, CMR 264, CMR 271,CMR 272, CMR 273, CMR 274, CMR 275BG 805 161 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
& [17] www.pfeiffer-vacuum.deOperating manualCompact Piezo Gauge APR 250, APR 260,APR 262, APR 265, APR 266, APR 267BG 805 035 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
& [18] www.pfeiffer-vacuum.deCommunication protocolDualGauge™BG 805 098 BEPfeiffer Vacuum GmbH, D-35614 Asslar,Deutschland
86 Appendix BG 805 096 BE (9907) DualGauge.om
bar mbar µbar Pa kPa Torr mTorr 1) psi
bar 1 103 106 105 102 750 750×103 14.5mbar 10-3 1 103 102 0.1 0.75 750 14.5×10-3
µbar 10-6 10-3 1 0.1 10-4 7.5×10-4 0.75 14.5×10-6
Pa 10-5 10-2 10 1 10-3 7.5×10-3 7.5 14.5×10-5
kPa 10-2 10 104 103 1 7.5 7.5×103 14.5×10-2
Torr 1.33×10-6 1.33 1.33×103 133 0.133 1 1000 19.3×10-3
mTorr 1.33×10-6 1.33×10-3 1.33 0.133 1.33×10-4 10-3 1 19.3×10-6
psi 6.89×10-2 68.9 68.9×103 6890 6.89 51.7 51.7×103 1
1) mTorr = micron = µ
E: Conversion ofpressure units
BG 805 096 BE (9907) DualGauge.om Declaration of conformity 87
Declaration of conformity
Vacuum measurement and control unit for CompactGauges
We herewith declare that the above product complieswith the provisions of the listed guidelines.
Guidelines, harmonized standards, national standards inlanguages and specifications which have been applied:
72/23/EEC (7/93) .....................................................
89/336/EEC (7/93) ...................................................
EN 61010-1: 1993 ....................................................
EN 50081-1: 1992 ....................................................
EN 50082-2: 1995 ....................................................
Asslar, 13.7.99
Wolfgang DondorfManaging director
Product
EU Declaration ofconformity as defined by thelisted guidelines
Signature
Pfeiffer Vacuum GmbHEmmeliusstrasse 33D–35614 AsslarDeutschlandTel +49 (0) 6441 802-0Fax +49 (0) 6441 802-202
Original: German BG 805 096 BE (9907) [email protected]
www.pfeiffer-vacuum.de