An Acne Technology For Immediate Results
Operation Guide
TheraClear
© Theravant Corp, 2014. Do not distribute without permission
Treatment Modes
Two Mechanisms of Action
Vacuum & Light Mechanism
Uses Regular or Small Inserts
Acne Surgery Mechanism
Uses Plastic Insert with Needle
© Theravant Corp, 2014. Do not distribute without permission
Pneumatics deep cleans pores by extracting build up of sebaceous material
Obstruction in the pilosebaceous apparatus is removed and skin normalizes
TheraClear hand piece makes full contact with treatment area
Along with targeting heating of the dermis, the endogenous effect of light activates porphryns to destroy P. acnes and reduces sebum production
TheraClear Treats the Causes of Acne.
© Theravant Corp, 2014. Do not distribute without permission
Pneumatics pull acne lesion into needle, opening access to lesion
Sebum is forced out by the pressure difference .
Removal of Sebum immediately reduces size of lesion.
TheraClear hand piece makes full contact with treatment area
New Extraction Insert Enhances and Speeds Results
© Theravant Corp, 2014. Do not distribute without permission
This first section illustrates the following features.
Turning the system on
Entering the pass code
Setting treatment parameters
Treatment Mode
Turning System Off
TheraClear Basic Operation
© Theravant Corp, 2014. Do not distribute without permission
TURN ON POWER
Turn the power switch located on the rear panel to the ON position.
The Keypad screen, pictured on the next slide, will appear after a few seconds.
Starting the TheraClear
© Theravant Corp, 2014. Do not distribute without permission
ENTER PASS CODE
Next enter pass code 1410
Press the enter key at lower right of Keypad.
Use the back arrow at lower left if you make a mistake.
Starting the TheraClear
© Theravant Corp, 2014. Do not distribute without permission
Starting the TheraClear
Standby Mode
System enters Standby Mode.
© Theravant Corp, 2014. Do not distribute without permission
Standby Mode
Set Treatment Parameters
Skin Types I to IV
Energy Level 6
Vacuum Level 2
Double Pulse
750 Between Pulses
Use Arrow Keys to Increase/Decrease Parameters.
Press on Orange Area For Double Pulse
Press Space Between Pulses to Change Time Between Pulses.
Decreases Parameter Increases Parameter
Adjusts Time Between Pulses
Set Double Pulse
© Theravant Corp, 2014. Do not distribute without permission
Standby Mode
Set Treatment Parameters
Skin Type V
Energy Level 4
Vacuum Level 2
Double Pulse
750 Between Pulses
Use Arrow Keys to Increase/Decrease Parameters.
Press on Orange Area For Double Pulse
Press Space Between Pulses to Change Time Between Pulses.
Decreases Parameter Increases Parameter
Set Double Pulse
Adjusts Time Between Pulses
© Theravant Corp, 2014. Do not distribute without permission
Standby Mode
Set Treatment Parameters
Skin Type VI
Energy Level 2
Vacuum Level 2
Single Pulse
Use Arrow Keys to Increase/Decrease Parameters.
Decreases Parameter Increases Parameter
© Theravant Corp, 2014. Do not distribute without permission
Treatment Mode
Hand Switch
Treatment Mode
Remove hand piece from the hand piece holder.
Enter Treatment mode by pressing and releasing the hand switch.
Approximately two seconds after pressing the hand switch, the light on the hand piece will change from blue to green.
The Treatment indicator (light bulb icon) will appear on the touch screen.
System is in Treatment Mode.
© Theravant Corp, 2014. Do not distribute without permission
Treatment Mode
Hand Switch
Treatment Mode
Remove hand piece from the hand piece holder.
Enter Treatment mode by pressing and releasing the hand switch.
Approximately two seconds after pressing the hand switch, the light on the hand piece will change from blue to green.
The Treatment indicator (light bulb icon) will appear on the touch screen.
System is in Treatment Mode.
Light Bulb Icon
© Theravant Corp, 2014. Do not distribute without permission
Soft Vacuum Settings
Hand Switch
Soft Vacuum Settings
Activate by Pressing on Vacuum Number.
Letter “S” appears in Front of Number
Press Again to Return to Regular Vacuum
Use on Forehead.
Use on Sensitive Skin Areas
Soft Vacuum Button
© Theravant Corp, 2014. Do not distribute without permission
Return to Standby Mode
Hand Switch
Return to Standby Mode
Return to Standby Mode by pressing and releasing the hand switch.
The light on the hand piece will change from green to blue.
The Treatment indicator (light bulb icon) will disappear on the touch screen.
Repalce hand piece in the hand piece holder.
System is in Standby Mode.
© Theravant Corp, 2014. Do not distribute without permission
Vacuum Extraction
Treatment Modes
Light & Vacuum
© Theravant Corp, 2014. Do not distribute without permission
Procedure • Treat the Lesion with the acne needle insert.
• Vacuum Extraction set at 3 seconds. • Needs the Time Mode RFID
• Treat the Lesion with Vacuum Only. (Optional)
• Vacuum Extraction set at 3 seconds
• Treat the Lesion with Vacuum & Light • Double Pulse Mode • Energy level 6 • Vacuum Level 2
• Treat the Entire Face with Vacuum & Light (Optional)
• Double Pulse Mode • Energy level 6 (Skin Type Dependent) • Vacuum Level 2
Acne Surgery Protocol
© Theravant Corp, 2014. Do not distribute without permission
Shutting Off System
Hand Switch
Sleep Mode
Press the TheraClear logo button (the Sleep Button) located at the bottom of the screen.
Pressing this Button places the system in a state in which a pass code needs to be entered before the system can be used again.
Next press the sleep icon.
System is in Sleep Mode.
© Theravant Corp, 2014. Do not distribute without permission
TURN OFF POWER
Turn power switch located on the rear panel to the Off position.
System can be powered off from Standby Mode, Treatment Mode or Sleep Mode.
Powering Off the TheraClear
© Theravant Corp, 2014. Do not distribute without permission
This section illustrates the following features.
Installing the Inserts
Installing the Acne Surgery inserts
Using the RFID for installing Pulses
Using the RFID for installing Time Segments
TheraClear Inserts
© Theravant Corp, 2014. Do not distribute without permission
Regular Size for Majority of the Face
Smaller Size for Noise and Above Lip
Acne Surgery Insert
The regular and small insert sell for $15 a piece. Sold in boxes of 10. Should last for 10 or more treatments
Acne Surgery Insert sells for $9.90 Plastic part can be autoclaved Needle is single patient use.
Treatment Inserts
© Theravant Corp, 2014. Do not distribute without permission
Treatment Inserts
Installing Inserts
Press Insert into Aperture
Insert Should be Snug
Remove by Pulling Insert out of Aperture
© Theravant Corp, 2014. Do not distribute without permission
Inserts are Used Per Patient and Cleaned Between Uses
Place system in standby or OFF mode. Remove the Insert from the handpiece.
Clean all sides and surfaces of the Insert with a Sani Cloth ™
A Q-tip may be used to better reach the insides of the Insert
© Theravant Corp, 2014. Do not distribute without permission
Debris Remains on the Lower Edge of Insert For Easy Cleaning
Skin does not touch the uppermost part of insert keeping it free of evacuated debris Skin debris remains
in lower part of insert
Skin Elevated into Treatment Insert
© Theravant Corp, 2014. Do not distribute without permission
Acne Surgery Insert
© Theravant Corp, 2014. Do not distribute without permission
Easy to Install Tip
© Theravant Corp, 2014. Do not distribute without permission
Treatment
© Theravant Corp, 2014. Do not distribute without permission