42
Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Embed Size (px)

Citation preview

Page 1: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Orbitrap Technology: New Frontiers in Pesticide Residue Analysis

Michal Godula

Special Solutions Center Europe

Thermo Scientific

Page 2: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

2

Premier Brands

• 50,000 employees in 50 countries

• $17 billion in annual revenues

• Unparalleled commercial reach

Unmatched Depth

• Leading innovative technologies

• Applications expertise

• Lab productivity partner

Global Scale

We Are The World Leader in Serving Science

Page 3: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

3

• No compromise MS, MS/MS and MSn

• Speed, sensitivity, ultrahigh resolution, precise mass accuracy, and dynamic range - at the same time

Orbitrap: The Performance Leader since 2006

What Orbitrap provides

Unmatched ultrahigh resolution, accurate mass performance

Which applications are optimized• Virtually all molecular analyses:

accurate identification, structural analysis, and quantification of organic molecules, lipids, carbohydrates, peptides & proteins in complex mixtures

• Proteomics, drug discovery, toxicology, forensics, environmental, food safety and many more

Page 4: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

4

Orbitrap FTMS

Makarov A. Anal. Chem. 2000, 72, 1156-1162.

Page 5: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

5

Orbitrap Mass Analyzer Features

• Fundamental difference to other HRAM instruments• Parameter measured is frequency, not time/voltage/current• Resolution allows more accurate m/z determination• Less prone to ambient conditions changes• Usually stable within <2 ppm during several days• No need for lock mass in “routine work”• Small footprint• Easy to setup

Page 6: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

How accurate is your mass?

Mass accuracy

610/

true

truemeas

m

mmzm

• Quadrupole MS

• Orbitrap MSTOF MS

ppmzm 20010500

0.5001.500/ 6

ppmzm 21010314.500

10214.50010314.500/ 6

Page 7: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Selectivity increases with higher mass accuracy

7.2 7.4 7.6 7.8 8.0 8.2 8.4 8.6 8.8 9.0 9.2 9.4 9.6 9.8 10.0 10.2 10.4 10.6Time (min)

0

10

20

30

40

50

60

70

80

90

100

Re

lativ

e A

bund

ance

RT: 8.45

9.298.728.678.99

8.789.118.217.85 8.838.137.627.59 7.81 8.10 8.357.38

7.337.27 9.46 10.6110.509.50 10.3410.169.64 10.09

100 ppm

20 ppm

2 ppm

7.2 7.4 7.6 7.8 8.0 8.2 8.4 8.6 8.8 9.0 9.2 9.4 9.6 9.8 10.0 10.2 10.4 10.6Time (min)

0

10

20

30

40

50

60

70

80

90

100

Re

lativ

e A

bund

ance

RT: 8.45

8.217.62 7.857.598.257.407.37 7.81 7.92 7.99

8.617.30 10.617.26 10.508.88 10.318.67 9.11 10.1610.099.46

7.2 7.4 7.6 7.8 8.0 8.2 8.4 8.6 8.8 9.0 9.2 9.4 9.6 9.8 10.0 10.2 10.4 10.6Time (min)

0

10

20

30

40

50

60

70

80

90

100

Re

lativ

e A

bund

ance

RT: 8.45

8.557.997.49 7.917.76 10.617.417.26 10.5010.3110.167.67 9.46

NL: 1.20E5m/z= 484.25313-484.25507 F: FTMS {1,0} + p APCI Full ms [300.00-850.00] MS pivo_5_UH

*Zachariasova M et al, Anal. Chim. Acta

Page 8: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Resolution (resolving power)

• Resolution

m

mR

m (FWHM)

m 10%

8336.0

500

m

mR• Quadrupole MS

• Orbitrap MS 100000005.0

500

m

mR

Page 9: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Does resolution matter?

RAFA 2013: Łukasz Rajski, María del Mar Gómez Ramos, Amadeo R. Fernández-Alba; EURL for Pesticide Residues in Fruits and Vegetables. Pesticide Residue Research Group. University of Almeria, Spain. e-mail: [email protected]

Page 10: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Q Exactive MS - a 3D view

Hyperbolic Quadrupole Mass filter

HCD Cell

Orbitrap Mass Analyzer

C-Trap

RF-Lens

Page 11: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Q Exactive FS-ddMS2 screening process

Inclusion List

Ion Accumulation

Full Scan Orbitrap Analysis

Precursor Selection

Precursor accumulation

Fragmentation in HCD

MS2 Orbitrap Analysis

Full Scan MS Data: Screening/Quantitation

Full Scan MS/MS Data:Confirmation

Page 12: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Multi-residue screening by Q Exactive LC-MS

• Complete solution for quantitiative analysis of LC amenable compounds• Implementation method for most relevant pesticides/mycotoxins and

pyrrolizidine alkaloids covering different chemistries• In 3 relevant matrices• Critical evaluation of established method performance criteria vs. current

regulation

strawberry leek wheat flour

Page 13: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Quechers extraction

10 g of homogenized sample is weight into Quechers extraction tube

+ 20 mL of water

+ 10 mL of ACN

Centrifugation 5 min @ 5000 rpm

Analysis

shaking 10 min

Ext

ract

ion

Page 14: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

UHPLC instrumental method

Thermo ScientificTM UltiMate® 3000 RSLC : • column: Accucore aQ 100 mm x 2.1 mm x 2.6 µm • Column oven T: 25 °C • Injection Volume: 1 µL• Mobile phase A: Water/MeOH (98:2) 5mM Ammonium

formate, 0.1% FA• Mobile phase B: Water/MeOH (2:98) 5mM Ammonium

formate, 0.1% FA• Flow rate: 300 µL/min

Page 15: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Q Exactive Conditions – ddMS2 experiment

• Q-ExactiveTM MS conditions: Spray voltage 3800V, Capillary temperature 295°C, sheath gas 32 au, Aux gas 7 au, S-lens RF level 55

• Full MS/data dependent MS2

• full MS:resolution 70,000 FWHM

scan range 120-1200 m/z

• ddMS2: resolution 35,000 FWHM

Inclusion list on

isolation 1 m/z

exclusion 5 s

List of 556 target analytes

1234567891011121314151617181920212223242526 … … …556

Page 16: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Isotopes

Fragments

LibrarySpectrum

Chromatogram

Target analytes

TraceFinderTM Screening Data Review

Page 17: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Thermo EFS Screening Database: 1729 substances

• Targeted screening in “known RT” compounds• Untargeted in the rest of database (RT not known)

Page 18: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Screening Example – Acetamiprid in honey sample

Page 19: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Acetamiprid confirmation

Isotopic Pattern

+ Main fragment

Page 20: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Acetamiprid details

Standard 50 ug/kg

Sample – determined concentration 5.3 ug/kg

SampleMS2 spectra

Standard MS2 spectra

Page 21: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

Isotopic pattern fitting is influenced by resolution

Page 22: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

22

Q Exactive Focus: Introducing vDIA

• Variable Data independent acquisition (vDIA) acquires MS2 spectra systematically and independent of previously acquired scans

• You obtain:• complete record of data

• very good reproducibility

• high throughput comprehensive quantification acquisition methods with quantitative information

• no requirement for detailed sample knowledge prior to data acquisition

• Your routinely acquired data is ready to go for any non-target screening approach

Page 23: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

23

Variable DIA – alternative to All Ions Fragmentation

Presented by P Zomer, RIKILT at EPRW 2014:

Traditional AIF

Variable Data Independent Analysis

Page 24: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

24

AIF vs. vDIA

AIF

vDIA

Oxacillin in Pig Kidney @ 2 ppb background

Page 25: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

25

Running 380 Compounds with the Q Exactive Focus

Data Dependent-MS2 vDIA

Quan/Qual Quantitation Targeted Screening

# of targeted compounds

Total: 380345 Pesticides, 35 Mycotoxins and Alkaloids

Selectivity Retention time, fragment ions, ion ratios, isotope ratio, library match

Linearity – working range [ppb]

1-500 (6 levels and 2 replicates)

Matrix Black Tea/Honey

Sample Prep Buffered QuEChERS method clean-up made by MgSO4, PSA and C18

Page 26: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

27

vDIA sensitivity in honey

80%

20%

Number of compounds with <1ppb sensitivity

Peaks in honey sample spiked to 1ppb

Page 27: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

28

345 target pesticides in tea matrix

Data still being processed

Page 28: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

29

A New Chapter in GC-MS

Thermo Scientific Q Exactive GC Hybrid Quadrupole-Orbitrap GC-MS/MS System

Offering Unprecedented Depth in GC-MS Analysis

• Hugely powerful HR/AM performance for GC-MS

• Uncompromised quantitative and qualitative capability

• Reliable, robust and easy to operate

Page 29: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

30

Bringing GC and Orbitrap Technology Together

C-TRAP

HCD Cell

AQT Quadrupole

Orbitrap Mass Analyzer

BentFlatapole

ExtractaBrite™ Ion Source

Page 30: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

31

Breakthrough in GC-MS Performance

Highest selectivity and confidence with high resolving power

Unrivalled Resolving Power

• Up to 120k @ m/z 200

• Allows excellent selectivity in complex matrix, maintaining confidence in detection and identification

• Drives highly resolved data sets and high capacity peak detection

• Fast scanning Orbitrap for well defined, reproducible GC peaks

Page 31: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

32

Breakthrough in GC-MS Performance

Highest selectivity and confidence with high resolving power

Chlorprophram

• 10 ng/g QuEChERS extract of leek

• Full scan with resolving power of >30k (FWHM @ m/z 200) provides interference free detection

• Excellent mass accuracy for confident identification

Page 32: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

33

Pesticides screening

High resolving power crucial to avoid false negatives

False negatives at 15,000 RP

Secure screening

• 5 ppm identification threshold

• Higher resolving power allows confident detection and identification

• Mid- resolution of 15k causes false negatives

• Q Exactive GC routinely operates at 60k (FWHM @ m/z 200)

Page 33: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

34

Breakthrough in GC-MS Performance

Highest selectivity and confidence with high resolving power

XIC of dieldrin in baby food (m/z 262.85642) 100ppb

12

3

4

5

6

7

89

10

11

12

13

14

1516 17

60,000 RP (m/z 200) 2.4 s peak

Fast acquisitions

• Important for accurate profiling of narrow GC peaks

• Full scan with resolving power of 60 (FWHM @ m/z 200) generates 17 scans

• Fast enough for GC!

17 scans

Page 34: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

35

Scan to-scan

XIC of dieldrin in baby food (m/z 262.85642) 100ppb

60,000 RP (m/z 200) 2.4 s peak

Breakthrough in GC-MS Performance

Consistently excellent mass accuracy

< 1ppm (internal)

< 3ppm (external)

• Scan-to-scan

• Low level in matrix

• Over full mass range

• Over full concentration range

• No need for continuously calibrating in sequence

0.70.7

0.6

0.5

0.3

0.6

0.5

0.3

0.3

0.3

0.5

0.6

0.6

0.70.5

0.5 ppm RMS

Page 35: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

36

Breakthrough in GC-MS Performance

Consistently excellent mass accuracy

Mass accuracy over >6 orders

• Hexachlorobenzene

• 0.2 – 250,000 ng/mL

• Mass accuracy within 1 ppm

• Orbitrap spectral fidelity independent of concentration

Over full concentration range

Data courtesy of Hans Mol and Marc Tienstra, RIKILT

Page 36: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

37

Is it sensitive enough? Q Exactive GC vs. Triple Q

Page 37: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

38

Breakthrough in GC-MS Performance

Unprecedented quantitation power for HR/AM GC-MS

SRMFull Scan

Page 38: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

39

Compound Discovery and Identification

Tracefinder: automated compound detection and identification

1. Peak detection

2. Spectral deconvolution

3. Library searching (e.g. NIST, Wiley, custom)

4. % TIC explained using elemental composition and accurate mass filtering

Page 39: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

40

Automated Workflow for Accurate Mass Confirmation

Unknown Identification Workflow

Spectral Deconvolution on EI HRAM

Data

Spectral Library SearchScoring

AdditionalHigh

Resolution Scoring

m/z = 275.11298

C11H23O4Si2m/z = 275.11294

m/z = 275.11298

Page 40: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

41

Pesticides screening

Horse feed detection limit

100 % detected at MRL levelMajority at ½ MRL

Page 41: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

42

Pesticides screening

Quantitation

Fenpropimorph calibration curve in leek matrix.

• For all pesticides, the coefficient of determination (R2) was >0.99 with an average value of R2 = 0.997

Curve from 0.5 to 50 ng/g (triplicate at each level)

R2=0.9999

Page 42: Orbitrap Technology: New Frontiers in Pesticide Residue Analysis Michal Godula Special Solutions Center Europe Thermo Scientific

43

Complete solution for total pesticide analysis

Q Exactive GC Hybrid Quadrupole-Orbitrap GC-MS/MS System

Q Exactive FocusHybrid Quadrupole-Orbitrap LC-MS/MS System