1 Pressure drop, mass transfer, haemolysis: a difficult relationship Filip De Somer, Ph.D. University Hospital Gent Belgium Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow [L/min] Sarns Turbo Cobe Optima Cobe Duo 1 Cobe Duo 2 * 5 LPM Pressure drop Diameter 3/8 inch (9.65 cm) L = 38.7 m Diameter 2 mm L = 7.5 cm P = 300 mmHg Q b = 5000 mL/min 7.2 10 5 erg/cm³/min 1.6 10 7 erg/cm³/min

Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

  • Upload
    others

  • View
    11

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

1

Pressure drop, mass transfer, haemolysis: a difficult

relationship

Filip De Somer, Ph.D.University Hospital Gent

Belgium

Oxygenator pressure drop

050

100150200250300350

Pres

sure

dro

p [m

mHg]

2 4 6

Blood flow [L/min]

Sarns Turbo

Cobe Optima

Cobe Duo 1

Cobe Duo 2

* 5 LPM

Pressure drop

• Diameter 3/8 inch (9.65 cm)• L = 38.7 m• Diameter 2 mm• L = 7.5 cm

P = 300 mmHgQb = 5000 mL/min

7.2 105

erg/cm³/min1.6 107

erg/cm³/min

Page 2: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

2

RBC hemolysis

Chambers 1996

Pressure [mmHg] Test sample Paired control Test sample Paired control-710 3.6 2.3 110 15-680 10 7 58 21-550 4 2.6 43 25-300 12 10 23 11-100 8 7.3 39 24350 3.7 2.8 17 11

1000 4.1 2.4 18 16

Absolute plasma free Hb values [mg/dL]Blood-air interfaceBlood-mineral oil interface

1. Laminar Flow Model

VelocityProfilePressure

Forces ShearForces

L

r

wP )

2( L )

( r )

2. Force BalanceP ) ( r²) = ( w r ) ( L )

Shear stress

Shear example

Page 3: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

3

20

10

0

40

30

200 400 600 800

Shear stress (dynes / cm2)

Sero

tinin

Rel

ease

(%)

Shear stress and plateletsSerotonin release

Hellums 1994

141 cm-1

81 cm-1

57 cm-1

T: 5 min

Shear related cellular activation

> 75 dyne/cm²Leukocyte activation

> 100 dyne/cm²Platelet activation

> 2000 dyne/cm²Rupture of the erythrocyte membrane

Shear stress oxygenator

• Complex geometry• Open area for flow: hydraulic radius

• Pressure drop = P membrane + P HE

Page 4: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

4

Current Technology:Cross Flow Fiber Bundles

Frontal area, Af

Blood flow path length, LBlood Flow

Hollow fibers ofdiameter d

Aspect Ratio, AR, =

Fiber bundle withvoid fraction Vf

LAf /

Cobe Optima XP

Length

Do Di

Af

Lb

Qb

Medtronic Affinity

Do

Di

Af

Lb

Leng

th

Qb

Page 5: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

5

Af

Lb

Qb

Quadrox

Relationship shear stress - time

Threshold for shear induced hemolysis

Mockros 2001

Relationship shear stress - time

Threshold for shear-induced platelet release of 5-HT, ADP, ATP, and -TG

Threshold for shearInduced trauma

Page 6: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

6

Relationship shear stress - time

“Safe” zone

Minimum recommendedfor material-inducedtrauma

Threshold for shearInduced trauma

Funakubo 2003

Oxygenator pressure drop

Max

ima

PRF

Fort

e

Aff

inity

Aff

inity

NT

Qua

ntum

Opt

ima

Capiox

1.8

Qua

drox

Hilite

0

50

100

150

200

250

Pres

sure

dro

p [m

mHg]

Segers 2001

Page 7: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

7

Relationship shear stress - time

“The high pressure gradients in some of the devices did not result in more haemolysis.”

Segers 2001

1.6 ± 1.882 ± 1553 ± 14Hilite

1.3 ± 1.379 ± 5944 ± 22Quadrox

3.5 ± 1.866 ± 1840 ± 9Optima

8.4 ± 3.366 ± 1539 ± 7Capiox 1.8

4.0 ± 2.567 ± 1638 ± 11Quantum

1.3 ± 0.775 ± 2349 ± 14Affinity NT

5.3 ± 4.562 ± 843 ± 5Affinity

4.6 ± 3.565 ± 2140 ± 18Forté

3.0 ± 1.684 ± 3750 ± 24Maxima

Plasma free Hb[mg/dL]

CPB time[minutes]

Ao clamp[minutes]

Pulsatile flow, P, shear stress

• Quadrox vs Capiox 1.8• Low vs high P• Average pulsatile blood flow 5 L/min

– Peak flow: approx. 10 L/min– Base flow: 28% (1.4 L/min)– Frequency 72 beats/min

Simons 2010

Page 8: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

8

Relationship shear stress - time

Simons 2010

Relationship shear stress - time

Hemolysis

PHigh P Oxygenator

Low P oxygenator

< 0.001838 ± 38648 ± 25

Systolic pulse pressure [mmHg]

< 0.001827 ± 53619 ± 28Systolic inlet pressure [mmHg]

0.40.104 ± 0.0880.088 ± 0.074NIH [g/100 L]

Simons 2010

Page 9: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

9

“Optimal window”

Quadrox

Avant

Cannula

• DLP ref 71424– Length: 25 cm (tapered)– Diameter: 24 Fr, 7.9 mm– P 4 L/min: 46 mmHg– P 6 L/min: 105 mmHg– Shear stress 4 L/min: 485 dyne/cm²– Shear stress 6 L/min: 1109 dyne/cm²

Cannula

Page 10: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

10

Window of a given oxygenatorFlow vs Tau & Reynolds

0

5

10

15

20

25

1 2 3 4 5 6 7 8

Blood flow [LPM]

Rey

nold

s nu

mbe

r [ ]

0

10

20

30

40

50

60

70

Tau

[dyn

e/cm

²]Tau AvantTau MonolythNre AvantNre Monolyth

Reed-Thurston presented in Dubai 2011

O2 transfer and shear stress

R = 0.61

Segers 2001

Page 11: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

11

O2 transfer and shear stress

AvantApexEosPrimo2xMonolythOptima XP

R=0.88

O2 transfer and shear stress

R=0.95

LpkUpku s

DARCY’S LAW

POROUS MEDIUM

2

fs

m kkty Permeabili

sm

AQUVelocity lSuperficia

Pa ppp Drop Pressure

s

m Q Rate Flow

m DDLA Area Frontal

VVPorosity

outin

3

2iof

bed

voids

QQ/LQ/L/2

DiDo LLb

Pin

Pout

Page 12: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

12

Q

P

STROOMVERDELINGexperimenteel

NRe

1 10

NSh

.NSc

-1/3

0.1

1

10NRe > 6.5; NPe > 3200NRe < 6.5; NPe < 3200

Sarns Turbo 440Cobe OptimaDideco Compactflo

NSh·NSc-1/3= ' ·NRe

NSh·NSc-1/3= ' ·NRe

m'

“Ideal” oxygenator

Surface AreaBlood interaction

Shear stressHemolysis (RBC,

WBC, PLts)

Mass transferO2, CO2, heat

Idealoxygenator

Page 13: Oxygenator pressure drop - Scansectscansect.org/wp-content/uploads/2015/04/Pressure.pdf · Oxygenator pressure drop 0 50 100 150 200 250 300 350 Pressure drop [mmHg] 2 4 6 Blood flow

13

Conclusions

• DP is not an independent variable of hemolysis

• Shear causes hemolysis– Time– Absolute value– CAVE: local shear