● 6 Selectable Voltage Ranges:
DC Digital Power Analyzer (100Vp, 12Adc / 60Ap)
● 0.375mA minimum current range & 0.01uA Current resolution that meets the measuring the current of IoT device.● 3.5-inch color LCD digital numeral and graphic (TFT) display● Data Logger mode:
● Built-in power switch to control
● Inrush Current and Voltage measurement.● Support external shunt measurement function for expanding the measurement of Current range. ● Support external shunt measurement function : Can be used with Prodigit 7550A, 1000A and 2000A to expand the higher measurement current and power integration Whr, Ahr measurement function demand● Optional Interface :GPIB、RS232、USB、LAN、9943 Measuring Fixture Connection
Features
Description
• Up to 256 records for VDC, ADC, Watt measurements• External PC for unlimited number of records for long-term quality monitoring
• Test period and repetition times up to 9999 times• Repeat test period can up to more than 138days.
● The 4017 is a new generation digital power analyzer designed specifically for DC power measurement. The 3.5” TFTLCD display screen provides graphics display and digital display. It can be providing highly accurate and convenient power measurements.● Suitable for any type of battery including dry cell, button cell battery, lithium cell lithium battery, alkaline cell and rechargeable battery. Especially 2.5V and 5V range can provide 0.1mV resolution.● The 4017 digital power analyzer offers a wide measurement range. It supports 18 selectable current ranges from 0.375mA to 12Adc and 6 selectable voltage ranges up to 100Vdc. For an even wider range of current and voltage measurements, it can also be combined with an external shunt (such as the Prodigit 7550A,1000A Precision Current Shunt ) and set to the required current scaling. With this setup, it can support current measurements up to 250A, 1000A and 2000A.● The 490.6KHz V/A high synchronous sampling rate and DSP accumulation of data supports no gap analysis over long periods of time. This feature is particular important for IoT devices, smartwatch, Bluetooth device current and power measurements... etc. While the current of these types of UUTs keeps changing according to the operating modes, the 4017 can accurate measure key data like average current Aav, Ahr, and Whr. during the entire measured period.● The Built-in battery charge and discharge mode accurately measures battery capacity in % for charge and discharge operation. It can be used to verify the current for battery power meter when the battery works on regular/irregular and charge / discharge <Gauge Meter>.● In order to understand the stability of the UUT (unit under test), the 4017 provides the Data Logger function. The Data Logger can store 256 states each for VDC, ADC, Watt, and ITHD. If a PC is available, there is no limit to the stored number of state . This provides a convenient and accurate power measurement of UUT stability over time.● To understand the effect of the UUT (unit under test) on long-term, repeated ON/OFF power cycles, the 4017 has a built-in a power switch that can control the ON/OFF angle of the DC input power, test period and repetition times up to 9999 times. For example, when running a turn ON and turn OFF every 10 minutes continuously, the longest observation test period can be longer than 138 days.● For remote operation, the 4017 digital power analyzer provides 4 optional interfaces GPIB / RS232 / USB / LAN for data capture and storage.● Available 9943 test kits: USB type A, USB Type C, Micro-USB, red/black connector cables. these test kits provide the necessary connections between an UUT and the 4017 DC analyzer.
0.375 ADC / 0.01mA0.75 ADC / 0.1mA1.5 ADC / 0.1mA3 ADC / 0.1mA6 ADC / 0.1mA12 ADC / 0.001A15 ADC / 0.001A〈800ms〉30 ADC / 0.001A〈400ms〉60 ADC / 0.001A〈200ms〉
2.5 VDC / 0.1mV,5 VDC / 0.1mV,10 VDC / 1mV25 VDC / 0.001V,50 VDC / 0.001V,100 VDC / 0.01V
0.375m ADC / 0.01uA0.75m ADC / 0.1uA1.5m ADC / 0.1uA3.75m ADC / 0.1uA7.5m ADC / 0.001mA15mA ADC / 0.001mA0.0375 ADC / 0.001mA0.075 ADC / 0.01mA0.15 ADC / 0.01mA
Digital Power Analyzer
01All specifications are subject to change without notice.
● Embedded high-speed DSP, 16 bits Analog / Digital converters to provide continuous gapless measurement with max sampling rate up to 409.6kHz● Input Range to 100 VDC / 12 ADC and provide the 3 Current ranges up to 15Apeak, 30Apeak, 60Apeak that the highest inrush current.
● 18 Selectable Current Ranges:
Features
● Power-On display and Main Menu
Key Functions
System Setup
Digital Display Mode Graphics Display Mode
Graphics Display Mode : V&I waveform
Meter Mode
● There are two modes for Display types: Digital display and Graphics display.● Press the “Graph” key to switch between Digital display and Graph display.● Press the Up and Down keys to scroll the displayed readings up ordown.
Prodigit Electronics Co., Ltd.02
● Average:1~1024 Default for Average is 10 measurement cycles. We recommend when measureming dynamic DC power and current, the average setting is increased according to rate of change of the measured signals, The width of each individual measurement is 10ms● 50kHz Filter : ON / OFF Suggest set to “ OFF ” for when you use with Linear Power source Suggest set to “ ON” for when you use with Switching Power Source.● Graph Mode : AVG, Peak AVG is averaging mode Peak is integration mode Suggest set to “Peak” when you want to check the noise variation range. Suggest Set to “AVG” when you want to view a stable graph.● Shunt : Int ’ , Ext “ Int ” is using the internal Shunt “ Ext ” is using an external Shunt● Ext Scale : 1~100,000. Sets the A / V scaling for the externalcurrent shunt used.
Graphics Display Mode : V & I & W Waveform
Graphics Display Mode : V&W waveform
● The 4017 Display updated rate is 200ms. ● Parameters Measurement : The unit measurement value is obtained at a fixed sampling rate of 4096 points in the sampling cycle, and the measurement of each unit is performed continuously with no gaps.● Time of Measurement is one divided by 409.6K/s sampling rate of the A/D converter for 4096 points in the 16-Bits/per cycle. The DSP processes and displays the Watt value at the time and performs integral accumulation operation to achieve performed continuously gapless Standby Power accumulation and Pav value calculations.
Voltage
Current
Sampling
Sampling Frequency = 409600 Hz
Compute Compute Compute Compute
Accumulate result 〈Whr, Ahr, Pav calculate〉
Real time display
200 ms
Compute
10ms
03All specifications are subject to change without notice.
● The 4017 Provide a 1~64 times/Measurement cycle averaging function : When you measure the dynamic DC power or DC Current, increase the Average parameter ( For example, the average measurement cycle is 16 times or more ) so the 4017 can obtain stable Vdc, Adc, Watt measurement values.
* Display value= value 1+value 2 ….. value N / N Average (N Average setup up by user)= ComputeC
C
Average 10 〈N=1~64〉
Real time display
200ms
Voltage
Current
Sampling
Sampling Frequency = 409600
C C C C C C C C C C
Average 10 〈N=1~64〉
C C C C C C C C C C C C C C
Average 30 〈N=1~64〉Average 30 〈N=1~64〉
Real time display
200 ms
Voltage
Current
Sampling
Sampling Frequency = 409600
Real time display
200 ms
C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C
Power ONOscilloscope display for V & A waveform
Power ON4017 display for V & A waveform
Inrush Current
● Press “Edit” key could edit the range of window and range - Setup the time for 0~100ms
- V Range - I Range
( For example, setup 50ms that the display will show the graph from 50ms)
04 Prodigit Electronics Co., Ltd.
Battery Charge Analysis
● Start value is the start voltage for start charge● Stop value is the stop voltage for stop charge● Max value is the max current during charge period.● Aav value is Ahr/Hr ● Watt value is Watt in time● Pav Value is Whr/Hr
● Graphics display mode provides reference information for voltage, current, and power variations during charging.
● Press “Edit” key for Setup the ranges, capacity, protection :
- V Range- Current Range- Auto Up- Capacity- Start Capacity- Stop Current- Over Voltage Protect- Under Voltage Protect- Over Current Protect
Digital Display Mode Graphics Display Mode
05All specifications are subject to change without notice.
● Press “Edit” key for Setup the ranges, capacity, protection :
Battery Discharge Analysis
- V Range- Current Range- Auto Up- Capacity- Start Capacity- Stop Current- Over Voltage Protect- Under Voltage Protect- Over Current Protect
Digital Display Mode Graphics Display Mode
● Start value is the start voltage for start discharge● Stop value is the stop voltage for stop discharge● Max value is the max current during discharge period.● Aav value is Ahr/Hr ● Watt value is Watt in time● Pav Value is Whr/Hr
● Graphics display mode provides reference information for voltage, current, and power variations during discharging.
06 Prodigit Electronics Co., Ltd.
Battery Cycling Analysis
● Press “Edit” key for Setup the ranges, capacity, protection:
- V Range- Current Range- Auto Up- Capacity- Start Capacity- Stop Current- Over Voltage Protect- Under Voltage Protect- Over Current Protect
Digital Display Mode Graphics Display Mode
● Value of Charge_1/Discharge_1 are the times for charge and discharge.● Start value is the start voltage for start charge/discharge● Stop value is the stop voltage for stop charge/discharge● Max value is the max current during charge/discharge period.● Aav value is Ahr/Hr ● Watt value is Watt in time● Pav Value is Whr/Hr
● Graphics display mode provides reference information for voltage, current, and power variations during charge/discharge.
07All specifications are subject to change without notice.
Ahr/Whr Accmulator
● Press “Edit” key for Setup the ranges, capacity, protection : - Voltage Range - Current Range - Auto Up - Capacity - CountMode - Stop Current● Press “Graph” key can show the average for Watt and Current variations.
● Watt value is Watt at time● Pav vale is Whr/hr● Aav value is Ahr/hr
● Pav、Aav Graphics display provides the best answer ( the best time setting of accumulate time) for Watt and Current. When the Pav, Aav reaches its lowest and most stable value, that means it is the optimal accumulate time.
Data LoggerData Logger Sett ing
● Press “Edit” key for Setup the ranges, capacity, protection: - V Range - Current Range - Auto Up - Capaci ty - Update rate● Press “←”or “ →” key can switch the display for Vdc/ Adc/ Watt record graph
Accumulator Sett ing
CH1:DC input Vol tageCH2:DC output Current
08 Prodigit Electronics Co., Ltd.
Digital Display Mode Graphics Display Mode
Oscilloscope display 4017 display
Cycling Setting
ON/OFF Cycling
Display Peak Vol tage
Display Peak Current
CH1:DC input Vol tageCH2:DC output Current
● Press “ Edit ” key for Setup the ranges, capacity, protection: - ON Time - OFF Time - Times for Repeat - Voltage Range - Current Rang● Press “Select” key to switch the Graphics for Vpeak or Ipeak.● Press “←” or “ →” key to switch the Current Value for Vpeak or Ipeak.● Press “ ↑ “ or “ ↓ “ to switch the Average Value for Vpeak or Ipeak.
09All specifications are subject to change without notice.
Oscilloscope display 4017 Display
DCSource
OrBattery
LOAD
CT
4017 DC Digital Power Analyzer
Ext. I +Ext. I -Vsense -Source -
Vsense+Source +
Electronic LoadOr
Real Device
The circui t d iagram for External CT and 4017 connect ion
The Devices connect ion for External CT and 4017
CT
VR<=2.5VR
4017
DC power supply
DC Electronic Load
10 Prodigit Electronics Co., Ltd.
External CT connection
DCSource
OrBattery
DCLOAD
R
4017 DC Digital Power Analyzer
Ext. I +Ext. I -Vsense-Source-
Vsense+Source +
Electronic LoadOr
Real Device
CT
VR<=2.5VR
4017
11All specifications are subject to change without notice.
External Shunt connection
The circui t d iagram for External 7550A and 4017 connect ion
DC power supply
DC Electronic Load
The Devices connect ion for External Resistor shunt and 4017
DCSource
OrBattery
DCLOAD
7550A
4017 DC Digital Power Analyzer
Ext. I+Ext. I -Vsense -Source -
Vsense+Source +
+-
+- Vout
Electronic LoadOr
Real Device
4017
The Devices connect ion for External 7550A and 4017
The circui t d iagram for External 7550A and 4017 connect ion
12 Prodigit Electronics Co., Ltd.
External 7550A Shunt Connection
DC power supply
7550A Precision Current Shunt
DC Electronic Load
USB Type A
uUSBUSB
Type CUSB
Type C
4017 DC Digital Power Analyzer
Sense SourceLoad+ - + - + -
9943 DC Test Kit
Other data signal connection
紅 黑
接線柱
紅 黑
接線柱
Load Source
The circuit diagram for 9943 measuring fixture ( DC test Kit ) and 4017 connection
The Devices connection for 9943 measuring fixture and 4017
13All specifications are subject to change without notice.
9943 Measuring Fixture connection
Measuring connect ion for Smartwatch charge/discharge ( Red / Black cable )
Measur ing connect ion for Smartphone charge ( USB type A )
132
Application connection
14 Prodigit Electronics Co., Ltd.
Measuring connect ion for Smartphone charge ( USB Type C )
Measur ing connect ion for Power Bank discharge ( USB Type C )
15All specifications are subject to change without notice.
Measuring connect ion for Battery discharge ( Red/Black cable )
4017 DC Digital Power Analyzer
(100Vp, 12Adc / 60Ap)
Optional interface:
Option:9943 DC Test Fixture
1 2 3 4
Order Information
GPIB Card RS232 Card USB Card LAN Card
16 Prodigit Electronics Co., Ltd.
4017MODEL
DCVVDC,
Vpk+/Vpk-,Vmax/Vmin
RangeMax. Input
Accuracy
10VDC
±0.1% of (Reading + Range)±0.5% of (Reading + Range, For Peak)
DC Power Watt±0.2% of (Reading + Range)
Vrange*ArangeAccuracyRange
Range
RangeMax. Input
Max. Input
Shunt 0.015A(75Ω)
Input Resistance ≧400KΩ
RangeMax. Input 100VDC
Input Resistance ≧4MΩ
Shunt 0.15A(1Ω)
Shunt 1.5A(0.08Ω)
Shunt 12A(0.01Ω)
Max. Input
Max. Input
Accuracy
Range
15mADC continuous0.0375ADC / 0.001mA 0.075ADC / 0.01mA 0.15ADC / 0.01mA
0.15ADC continuous
1.5ADC continuous3ADC / 0.1mA6ADC / 0.1mA
12ADC / 0.001A60Apeak per 200ms / 12ADC continuous
10KΩ 0~+/-2.5VpeakExt. Input
Scaling
Input impedanceInput Range
1.00~100000.00
±0.5% of (Reading + Range,For Peak)±0.1% of (Reading + Range)
0.375mADC / 0.01uA0.75mADC / 0.1uA
1.5mADC / 0.1uA3.75mADC / 0.1uA
7.5mADC / 0.001mA15mADC / 0.001mA
Range 0.375ADC / 0.01mA 0.75ADC / 0.1mA 1.5ADC / 0.1mA
Inrush V/A
Range
AccuracyMax. Input
Range
Voltage
CurrentShunt 12A (0.01Ω)
Same as DCV
±2% of (Reading + Range)
AccuracyMeasurement Wide 100mS
±2% of (Reading + Range)Max. Input
3A~60A
0D0H0M0S ~ 9999D23H59M59S
DCADCA,
Apk+/Apk-, Amax/Amin
DC Ahr /Whr
Calculator
Accumulated Time
AHr
CounterAccuracy
0.000000 nWHr~999.999999 WHrWHr 1.000000~999.9999999 KWHr
1.000000~9999.9999999 MWHr0.000000 uAhr~999.999999 Ahr
1.000000~999.999999 KAHr1.000000~9999.999999 MAHr
0H0M0S~99H59M59S
±0.2% of (Reading + Range)
Low Pass Filter(V & A)Interface (Option)
50KHzRS-232, GPIB, USB, Ethernet
ItemUpdata RateData Logger
Vrms、Arms、Watt0.2、0.5、1、2、5、10 Second
ON timeOFF Time
ON / OFFCycling
0M0.500S ~ 10M0S
0M0.500S ~ 10M0S
Repeat Cycle 0~9999
+ or - Peak Value (+/-Vpk, +/-Apk)Max.or Min Value
(Vmax/Vmin, Amax/Amin, Wmax/Wmin)
Max [ Value(t) ] or Min [ Value(t) ]
Max [ Value ] or Min [ Value ]
>400k HzSampling Rate<2.5usInrush Sampling
Dual 16-Bit, 500KSPS ADC with DSPV/A ADC110 / 220V 50/60HzPower Input
50VAConsumption
OperatingTheory
2.5VDC / 0.1mV
25VDC / 0.001V
15Apeak / 0.001A〈800ms〉30Apeak / 0.001A〈400ms〉60Apeak / 0.001A〈200ms〉
3.6x11mm 250Vac 0.2A FastShunt 0.015A (75Ω)3.6x11mm 250Vac 0.2A FastShunt 0.15A (1Ω)
6*30mm 250V/10ASwitch3.5" TFT LCD, 320 x RGB x 240 Display
3.5 KgWeight -20 ºC to +60 ºC (-4 °F to 140 °F)Storage temperature 0 ºC to 40 ºC (32 °F to 104 °F)Operating temperature
2000 M (6562 ft)Maximum operating altitude80% for temperatures up to 31 °C (88 °F) decreasing linearly
to 50 % relative humidity at 40 °C (104 °F)Maximum relative humidity
Protection(fuse)
99.4 mm with feetHeight213 mmWidth304 mmDepth
Dimensions
5VDC / 0.1mV 10VDC / 1mV
50VDC / 0.001V 100VDC / 0.01V
Specifications
All specifications are subject to change without notice. 17