Wednesday, December 07, 2016

PA-100M

Report on the inspection for compliance with

Temperature Range Operation

- Low Temperature Test -

Name Date Signature

Conducted by: Valentin Ladizhensky 05/12/2016

Approved by: Dimitry Ginzburg 06/12/2016

Page 2: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 2 of 12 RD-4-16-137 Ver.1.0

Page 2 of 12

Contents

1. Scope………….……………………………………………………….4

2. Reference………………………………………………………………5

3. Test Instrumentation…………………………………………………….5

4. Test procedures ………….…………………………………………5

5. Results………………………………………………………………….9

Page 3: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 3 of 12 RD-4-16-137 Ver.1.0

Page 3 of 12

This Page Is Left Blank

Page 4: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 4 of 12 RD-4-16-137 Ver.1.0

Page 4 of 12

1. Scope

Purpose.

- These tests are conducted in addition to the inspection for compliance with United

States Military Standards MIL-STD-810E (Meth 501.3/502.3 Procedure II.)

- The reason for additional test is due to a fail at low temperatures (-10°C). Details are

described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering

Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST

REPORT_001_DG” by Rotem Industries Ltd.

1.1. Equipment under test:

- PA-100M – Alpha Detector contains atmospheric cells made of aluminium containing

a thin anode wire. The surface of the cell is a membrane of aluminized Mylar, protected

by a fine stainless steel wire mesh.

- RAM R-200 - This Monitor connect with the two above-mentioned detectors (PA-100

and RP-10). RAM-R-200 is portable multifunctional, rugged survey meter designed for

measuring wide range gamma radiation fields and contamination. The RAM R-200

along with its internal detectors is waterproof to a depth of 1 meter.

- RAM R-200 Detector Cable: This is a special cable for connection between the

Monitor and Detectors.

1.2. Specification of Detector PA-100M:

- Temperature Range Operation: -10°C to +50°C

- Accuracy ± 15%

Page 5: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 5 of 12 RD-4-16-137 Ver.1.0

Page 5 of 12

Table 1. Detectors under Test

Note: UUT-1 using the PA-100M after repair, with the old version software. Registered

instability in operation under normal conditions.

Page 6: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 6 of 12 RD-4-16-137 Ver.1.0

Page 6 of 12

2. References

- Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental

& Mechanical Lab”

- Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG” by Rotem

Industries Ltd.

3. Test Instrumentation

TEMPERATURE CHAMBER

- Programmable H/L Temp.Tester Chamber

Model: FC-60P, -20/+150[°C], S/N: 200808001, manufacture “MRC”

4. Test Procedure

UUT-1 and UUT-2 tested inside the Chamber with source Torium-232

Fig.4.1. Temperature profile during Temperature Operation Test

Page 7: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 7 of 12 RD-4-16-137 Ver.1.0

Page 7 of 12

Table.4.1. Temperature profile

- After stabilization at each temperature step three readings were recorded.

- The average of these readings was calculated and presented in the results section.

Page 8: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 8 of 12 RD-4-16-137 Ver.1.0

Page 8 of 12

Pic.4.1.1. UUT inside the Chamber.

Page 9: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 9 of 12 RD-4-16-137 Ver.1.0

Page 9 of 12

Pic.4.1.2. Programmable H/L Temp.Tester Chamber. UUT inside Chamber.

Page 10: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 10 of 12 RD-4-16-137 Ver.1.0

Page 10 of 12

5. Results

5.1. Checking prior to the test.

All instruments have been tested in the chamber in a stable geometry before starting

Temperature testing at room temperature +24°C.

Table. 5.1. Pre-Test (Before the start of stage of tests Temperature Operation)

Page 11: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 11 of 12 RD-4-16-137 Ver.1.0

Page 11 of 12

5.2. Operation Temperature

Table 5.2. Results of Temperature Operation Test

- UUT-1 (PA-100M S/N: n/a) – on Temperature 5°C Accuracy is unacceptable, the

deviation continues to increase in proportion to the decrease in temperature. Normal

operation on temperature greater than 10°С.

- UUT-2 (PA-100M S/N: 9615-006) – on Temperature -5°C Accuracy is 20%.

The deviation continues to increase in proportion to the decrease in temperature.

Specificated operation on temperature greater than -5°С.

6. Conclusion

According to the obtained results it is suggested to modify the Operation

Temperature range of the PA-100M Detector to be the following:

-5°C to +50°C (20% Temperature dependence)

Page 12: PA-100M … · described in Reports No: 20160918-1605 by “QualiTech Compliance Engineering Environmental & Mechanical Lab”., and Report “RAM R-200&PA-100&RP-10 TEST REPORT_001_DG”

Date: December,

2016

Report on the inspection for compliance with

Temperature Range Operation

R&D Department

Edition: 001 PA-100M

Page 12 of 12 RD-4-16-137 Ver.1.0

Page 12 of 12

END OF REPORT