PACS SVR 22/23 June 2006
PACS Thermal Behaviour 1
PACS Thermal Behaviour Tests
T. Müller, H. Feuchtgruber, A. Contursi,Cs. Kiss, Sz. Csizmadia
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 2
Available Reports
• Thermal Behaviour Tests During EQM IMT: SPEC & PHOT , PACS-ME-TR-067, issue 1.0• Blue Spectrometer Temperature Sensor Variation – EQM IMT PACS-ME-TR-066, issue 1.0• Comparison of the nominal and the PACS/SPIRE parallel mode cooler Recycling procedures • Cooler Holdtimes of the Photometer Focal Plane Unit, PICC-KO-TR-004, Draft 0.1
As part of the PACS ILT report: • Cooler Recycling Procedure of the Photometer Focal Plane Unit AVM/CQM ILT• Thermal Behaviour Tests during CQM ILT: SPEC & PHOT
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 3
PACS EQM Cryostat Environment
To be compared with IST
and TV/TB Environment
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 4
FM IST Cryostat Environment
T101 T102
E201
P501
SV521
V503V501
S501
N511 N512
V103
V106
T112T111
P101SV123
H101 H102
H103 H104
A101 - A1XX
PPS111
RD124
E421 E441 E461
HTT
HOT
CVV
HS1 HS3
SV922
S101
N513
V504
T702, 703
T106,107
H502/H503
Legend: A: adsorberE: heat exchangerH: heaterHS: heat shield L: liquid level probeN: exhaust nozzleP: pressure sensorPPS: passive phase separatorRD: rupture discS: filterSV: safety valveT: temperature sensorV: valveVG: vacuum gauge
SV723
RD724
P701
SV921
T501
T504T506
HS2
SV121
E601
Filling port
OBA
T105 T104
V505
V506
T507
T505
L101
L701
T701, 704
L702
L102
T103
H701 H702
V105V701
V104
V102
V702
VG901
VG902
H501
T502
V502
Non-CCU components
P502
Closed !(drifting)
Tilt may affect Red Ge:Ga strap
Full
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 5
FM TV/TB Cryostat Environment
T101 T102
E201
P501
SV521
V503V501
S501
N511 N512
V103
V106
T112T111
P101SV123
H101 H102
H103 H104
A101 - A1XX
PPS111
RD124
E421 E441 E461
HTT
HOT
CVV
HS1 HS3
SV922
S101
N513
V504
T702, 703
T106,107
H502/H503
Legend: A: adsorberE: heat exchangerH: heaterHS: heat shield L: liquid level probeN: exhaust nozzleP: pressure sensorPPS: passive phase separatorRD: rupture discS: filterSV: safety valveT: temperature sensorV: valveVG: vacuum gauge
SV723
RD724
P701
SV921
T501
T504T506
HS2
SV121
E601
Filling port
OBA
T105 T104
V505
V506
T507
T505
L101
L701
T701, 704
L702
L102
T103
H701 H702
V105V701
V104
V102
V702
VG901
VG902
H501
T502
V502
Non-CCU components
P502
Empty Tilt may affect Red Ge:Ga strap
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 6
In-Flight Cryostat Environment
T101 T102
E201
P501
SV521
V503V501
S501
N511 N512
V103
V106
T112T111
P101SV123
H101 H102
H103 H104
A101 - A1XX
PPS111
RD124
E421 E441 E461
HTT
HOT
CVV
HS1 HS3
SV922
S101
N513
V504
T702, 703
T106,107
H502/H503
Legend: A: adsorberE: heat exchangerH: heaterHS: heat shield L: liquid level probeN: exhaust nozzleP: pressure sensorPPS: passive phase separatorRD: rupture discS: filterSV: safety valveT: temperature sensorV: valveVG: vacuum gauge
SV723
RD724
P701
SV921
T501
T504T506
HS2
SV121
E601
Filling port
OBA
T105 T104
V505
V506
T507
T505
L101
L701
T701, 704
L702
L102
T103
H701 H702
V105V701
V104
V102
V702
VG901
VG902
H501
T502
V502
Non-CCU components
P502
Empty
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 7
Thermal Behaviour Test: Spectroscopy (1)
• Problems:- He leak- Bugs in the script- Blue detector temperature sensors where not functional
• Remarks: - Two cover heat-up activities took place during the test, causing some impact on L1 temperatures - General upward drift of L1 temperatures, probably due to cryostat operations - Significant increase of Red detector temperature in second half of the test probably due to the He-leak effects and combined with the cryostat operations. - The internal calibration sources were not switched-on due to a bug in the command script
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 10
Blue Detector Heating in Spectroscopy
Iheater=200µA
Iheater=100µA
After Recovery of Temperature Sensor:DECB_TS_2_3
Input to thermal shimming for FM
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 11
Thermal Behaviour Test: Photometry (1)
• Problems:- He leak- Bugs in the script- excessive dissipation by bolometer group 5 (PACS-ME-NCR-163) - Timing FPGA parameters problem DMC (PACS-ME-NCR-159)- DPU CFM generated corrupted data
• Remarks: - Executed 4 times (2 with leak, 1 without, 1 parallel mode) - Ideally, more time would be required to see stabilization of CSs
• The PACS/SPIRE parallel mode worked from the thermal point of view, at the He-mass flow given during EQM IMT.
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 13
Thermal Behaviour Test: Photometry (3)
with Bias group 5 problemwithout Bias group 5 problem,
during parallel mode
PACS-ME-NCR-163
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 14
Cooler hold Time with He-leak
End of 2h10mRecycling Procedure
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 15
Cooler hold Time without He-leak
only little detector operation duringcooler hold time !
~40 hours
End of 2h10mRecycling Procedure
PACS SVR 22/23 June 2006
PACS Thermal Behaviour 16
Thermal Behaviour Conclusions• Do not run a test with a He-Leak. Consequence: Non-Nominal He- flow scheme due to main tank leak • Absolute L1 He-flow more than a factor 10 larger than specified for FM [spec. (IID-A) = 2.1 mg/s ; actual >25 mg/s] • L1 temperatures became stable only towards the end of the test phase• Photometry: bias group 5 problem caused anomalous heat dissipation• Overall thermal behaviour of PACS was within expectations and comparable to ILT• Cooler hold time too short, even with little bolometer operations: Recy. Opt. + improved cooler filling; Cover-T ? L1-T ? Straps ? Leak ? • Cryo-cover, L2, L1 and L0 temperatures should be kept constant during thermal behaviour tests• The malfunction of at least one blue spectroscopy detector could be recovered to obtain the required FM design information.• Stabilized value of DM_CAL_SRC temperature not visible due to time constraint• Steep learning curve for implementation of related commanding scripts