"1 Report SAM-TR-78-38 "A' SELECTRICAL PARAMETER VALUES OF SOME HUMAN TISSUES IN THE RADIOFREQUENCY RADIATION RANGE Joir;. W. Penn, B.A. Earl L. Bell, M.S. CD U L.L- C.. December 1978 Interim Report: for Period October 1977 - February 1978 [-Approved for public release; distribution unlimited. USAF SCHOOL OF AEROSPACE MEDICINE ,xe Aerospace Medical Division (AFSC) Brooks Air Forte Bas* Texas 78235

PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

"1 Report SAM-TR-78-38

"A'

SELECTRICAL PARAMETER VALUES OF SOME HUMAN TISSUESIN THE RADIOFREQUENCY RADIATION RANGE

Joir;. W. Penn, B.A.

Earl L. Bell, M.S.

CD

U

L.L-

C.. December 1978

Interim Report: for Period October 1977 - February 1978

[-Approved for public release; distribution unlimited.

USAF SCHOOL OF AEROSPACE MEDICINE ,xe

Aerospace Medical Division (AFSC)Brooks Air Forte Bas* Texas 78235

Page 2: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

NOTICES

This interim report was submitted by personnel of the Biomathemat-ics Modeling Branch, Data Sciences Division, USAF School of Aerospace Med-icine, Aerospace Medical Division, AFSC, Brooks Air Force Base, Texas,under job order 7757-01-69.

When U.S. Government drawings, specifications, or other data areused for any purpose other than a definitely related Government pro-curement operation, the Government thereby incurs no responsibility norany obligation whatsoever; and the fact that the Government may haveformulated, furnished, or in any way supplied the said drawings, specifi-cations, or other data is not to be regarded by implication or otherwise,as in any manner licensing the holder or any other person or corporation,or conveying any rights or permission to manufacture, use, or sell anypatented invention that may in any way be related thereto.

This report has been reviewed by the Information Office (01) andis releasable to the National Technical Information Service (NTIS).At NTIS, it will be available to the general public, including foreignnations.

This technical report has been reviewed and is approved for publi-,cation.

_/JOHN W. PENN, B.A. RICHARD A. ALBANESE, M.D.Project Scientist Supervisor

I AW. C . ENDERSColonel, USAF, MCCotnan !"r

Page 3: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

UNCLASSIFIEDSECURITY LAS','IrICATION OF THIS PAGE (Whien Date Rnltered)

REPORTDOCUMNTATIO PAGEREAD INSTR CTONEPOR DOCMENTYIONPAGEBEFORE COMPLETING FORM

R kP')4IhLa&" 12 GOVT ACCESSION N -. RCIPIENT'S CATALOG NUMBER

j/SAM-TR-78-I_________ __

9. -- TIL _-n _uIfe - 11 YPE o v . .,.O ERED

j.~ETRICAL ZARAMETER VALUES OF jOME WIJMAN InSUS Otem W RepFb 297... II HE gADIOFREQUENCY'RADIATION RANGE, TISE Oc W77-Fe I -

L -6. PERFORMING ORG.TPF U ER

7AUT HOR(a) 6. CONTRACT OR GRANT NUMBER(e)

John W./enn, B.A.IEdrl L./Bell., M.S.

9. PERFORMING ORGANIZATION NAME AND ADDRESS I0. PROGRAM ELEMENT. PROJECT. TASKAREA a WORK UNIT NUMBERS

UJSAF School of Aerospace Medic.inev(BR) 622Aerospace Medical Division (AFSC)- . ..7

Brooks Air Force Base. Texas 78235 =F______________

11. CONTROLLING OFFICE NAME AND ADDRESS .IOR

USAF School of Aerospace Medicine (RiP) cm= 7Aerospace Medical Division (AFSC) I UBRO AE

Brooks Air Force Base, Texas 78235 214. MONITORING AGENCY NAME a ADORESS(il diferent from Controlling Office) 15. SECURITY CLAS- o eot

Unci assi fled

15 . DECL ASSI FIC ATION/D-OWN GRADINGSCHECULE

16. DISTRIBUTION STATEMENT (of this Report)

Approved for public release; distribution unlimited.

17. DISTRIBUTION STATEMENT (of the abs tract entered in Rlock 20. If different from Report)

I0. SUPPLEMENTARY NOTES

19. KEY WORDS (Continue on revrse side if necessary end Identify by block number)

Electrical parameter values Dielectric constantElectrical parameter curves Electrical propertiesElectrical properties of human tissue Radiofrequency radiationConductivity

1$ADSTRACT (Continlue an reverse side if necessary and identify hy block momber)

his report is an attempt to consolidate electrical parameter values of somehuman body tissues for the radiofrequency radiation range: 10 - 10,000 MHz. Thesources of the data were widely distributed in scientific journals, books. andtechnical reports and included data points secured by experimental techniques,by computations (Debye or modified Debye equations), and by curve fittingmethods. Values extracted from the conductivity and dielectric constant curvesprovided in the report give reasonable estimates of the interrelationships

D I AN7 73 OIONO NV1S001? UNCLASSIFIED

§1.7 SCURITY CL ASSIVICATION OP THIS 04,1 ""b~ "t, 0",o

Page 4: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

UNCLASSIFIEDCURITY CLASSIFICATION OF THIS PAGIE(Whanw Data Entered)

0. ABSTRACT (continued)between various tissues over the frequency range of interest. These values areimportant in the study of effects of radiofrequency radiation on livingorganisms.

UNCLASSIFIED116CIFI8? CLAUWICA?g@ OF TOMS PA$E(~a, Dat EntonE)

Page 5: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

I

TABLE OF CONTENTS

PageINTRODUCTION ........ .. ........................... 3

ELECTRICAL PARAMETER DATA ON VARIOUS HUMAN BODY TISSUES ......... 3

CONCLUSIONS ........ ................. .... . ...... 6

REFERENCES ........ ... ............................ 7

List of Illustrations

Figure

I Conductivity of muscle as function of frequency. Tem-

perature 370C. Symbols B, C, F, and H refer to Refer-

ences 2, 3, 6, and 8, respectively ...... ............. 9

2 Dielectric constant of muscle as function of frequency.

Temperature 370 C. Symbols B, C, E, H, and K refer to

References 2, 3, 5, 8, and 11, respectively .... ......... 10

3 Conductivity of skin-dura as function of frequency. Temperature

370C. Symbols B, F, G, J, and L refer to References 2, 6, 7,

10, and 12, respectively ....... .................. 11

4 Dielectric constant of skin-dura as function of frequency.

Temperature 370C. Symbols B, G, J, K, and L refer to

References 2, 7, 10, 11, and 12, respectively ............ 12

5 Conductivity of brain as function of frequency. Temperature

370C. Symbols B, G, H, and L refer to References 2, 7, 8,

and 12, respectively ........ .................... 13

6 Dielectric constant of brain as function of frequency. Tem-

perature 370C. Symbols E, G, H, and L refer to References

5, 7, 8, and 12, respectively ................... .... 14

woo

Page 6: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

I

TABLE OF CONTENTS (continued)

7 Conductivity of blood as function of frequency. Tem-

perature 370C. Symbols B and F refer to References 2

and 6,respectively ...... .. ....................... 15

8 Dielectric constant of blood as function of frequency.

Temperature 370C. Symbols B, E, H, and K refer to

References 2, 5, 8, and 11,respectively ................ .16

9 Conductivity of cerebrospinal fluid (CSF) as function of

frequency. Temperature 370C. Symbols G and L refer to

References 7 and 12, respectively ... ............... ... 17

10 Dielectric constant of cerebrospinal fluid (CSF) as function

of frequency. Temperature 370 C. Symbols G and L refer to

References 7 and 12, respectively ... ............... ... 18

11 Conductivity of fat-bone-yellow bone marrow as function of

frequency. Temperature 370C. Symbols B, C, E, F, H, and

L refer to References 2, 3, 5, 6, 8, and 12, respectively . . . . 19

12 Dielectric constant of fat-bone-yellow bone marrow as

function of frequency. Temperature 370C. Symbols B, C,

F, H, K, and L refer to References 2, 3, 6, 8, 11, and 12,

respectively ...... ... .......................... 20

Table

1 Values of electrical parameters of human tissues ex-

tracted from appropriate curves .... ................. .21

2

Page 7: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

ElFCTRICAL PARAMETER VALUES OF SOME HUMAN TISSUES IN THE

RADIOFREQUENCY RADIATION RANGE

INTRODUCTION

Radiofrequency radiation can be absorbed by living organisms. Thepatterns of absorbed power distribution and the associated buildup ofhot spots within living objects vary greatly with the frequency, direc-tion, and strength of the impinging waves and with the size, shape, tem-perature, and internal structure of the body irradiated.

Researchers in the field of electromagnetic bloeffects are con-fronted with the problem of data scarcity on electrical parameters ofliving tissues. The conductivity (a) and complex dielectric constant(E) of each type of tissue are complicated functions of temperature,percent water content, electrolyte concentration, large protein moleculecontent, cell structure and cell wall properties of the tissue. Theoriesto explain the variation of o and F with frequency are well known(6, 8). In tissue, three S-shaped bends in the conductivity and dielec-tric curves can be seen. These occur in the low frequency, radiofre-quency, and superhigh frequency ranges and are called a-, a-, and y-dispersions, respectively (6).

In the study of human effects of radiofrequency radiation, the con-ductivity and the real part of the complex dielectric constant at 3 70Cfor various human tissues are required. An attempt to determine theseproperties in the frequency range 10-10,000 MHz produced the resultsdescribed in this report. The 8-dispersion causes rapid changes in aand r at the low end of this frequency range, and the y-dispersioncauses rapid changes at the high end with relatively small changes inbetween.

The knowledge to be gleaned from this effort is directly relatedto the research effort of the Radiation Sciences Division at the Schoolof Aerospace Medicine. Briefly, here studies are being currently con-ducted: (1) to determine radlofrequency radiation-induced effects inbiological specimens, (2) to accurately determine the distribution ofenergy in the whole biological body or a particular organ, (3) toextrapolate response to radiation from the test animal to man in ameaningful manner, and (4) to contribute in the design of realisticsafety standards with a sound scientific foundation.

ELECTRICAL PARAMETER DATA ON VARIOUS HUMAN BODY TISSUES

Schwan (8, 9, 10) has published much of the available data in thisfield. These data are published in various forms--in tables, in graphs

3

4 r v

Page 8: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

I

drawn through available laboratory measurements, and in plots that appearto be based on equations derived from available data. On some graphsSchwan (8) shows the source of these measurements to be various studiesdating from 1937 through 1956. Johnson and Guy (3) have published tablesof values for (a) muscle, skin, and tissues of high water content and(b) fat, bone, and tissues of low water content, and state that tissuesof intermediate water content will lie between the two listed groups.Other authors who print tables of values have used much of Schwan's datain addition to available new data from frequently unspecified sources.

Three methods are commonly used to obtain values of a and cfor various tissues, as a function of frequency, from available data.

First, the available points can be plotted and a smooth curve drawnthrough the points. A table of values is then extracted from the curveand used for computing interpolated values at any frequency. This methodyields the most accurate results, but for nearly all tissues investigated,the available points leave room for speculation in at least some partsof the frequency range.

A second method is to fit the available points to an equation. Theequations utilized are derived from the work of Debye (1):

F +b(f/f ) 21

1+(f/f )2

Fd+(f/f )2

C1+(f/f0 )2

where

f = frequency in GHz,

fo = frequency in GHz associated with the highfrequency inflection of the curve,

a, b, c, and d = constants for the tissue type.

4

Page 9: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

I

Weil (12), Joines and Spiegel (4), some plots of Schwan (10), and localefforts show several disadvantages of this method. These equationsattempt to fit data in the y-dispersion area, and points in the fre-quency range 10-30 GHz are not available. The equations do not fit theobservations well, but they are still useful in a narrow frequency range.From our experience with available data, this method seems to over-emphasize the differences between tissues.

When only a few points in a narrow frequency range are availablefor a given tissue, a third method is used. A tissue for which valuesare available over the desired frequency range is selected as a base.The ratio of values for the base tissue to those for the tissue with fewvalues is calculated for the frequencies measured. This ratio and thebase curve are then used to calculate data at other frequencies. Inour experience this method underestimates the differences between tissuetypes. Johnson and Guy (3) and Lin et al. (5) use this technique.

Data for the following discussion of electrical properties weregathered from the References and, after eliminating duplicated values,coded by the number of the source in the Reference List. In most casesthe originator of the measurements is unknown. Some of the data wereread from plots. Where a range of measured values, from one or severalstudies, was found in one reference, the midpoint of the range was re-corded.

Primarily due to water content of the samples tested, the variationsin measurements are less than 10% for most tissues; fat is a notableexception.

Muscle is the tissue with the most recorded data and the leastvariable data. The measured points and estimated mean curves for aand c of muscle are shown in Figures 1 and 2 respectively. Thesecurves probably furnish the best base curves for estimating the electri-cal property values of other tissues with medium to high water contentwhen sufficient laboratory measurements are missing.

Figures 3 and 4 show the measured points for skin. Johnson and Guy(3) print one table for muscle, skin, and tissue of high water content.Data from other sources indicated that a and E for skin are lowerthan for muscle. The curves drawn on these plots are 0.9 times valueson the curves for muscle. Tinga and Nelson (11) show values for skinfrom different parts of the human anatomy. It is alleged that dura hasproperties similar to skin (7, 12), but no values for dura alone werefound.

The available values for brain tissue in Figures 5 and 6 are too fewin number and too scattered to define an adequate curve. Lin et al.(5) print new data for E of the brain that are lower in value than the

5

Page 10: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

older data found. The plotted curves are for 0.658 times muscle values.This is consistent with the newer values of c for brain tissues. Mathe-matical models of the human head show that small changes in these brainparameters cause large changes in the hot spots calculated. Schwan (9)states that red bone marrow has electrical properties similar to thoseof the brain, but no data for this tissue were ever found.

The electrical properties of whole blood, shown in Figures 7 and 8,are quite high since whole blood has a high water content. These prop-erties vary with the cell counts for the blood sample tested. The curvesplotted are 1.2 times muscle values for a and 1.1 times muscle valuesfor c. The frequency response curves for blood are obviously differentfrom those for muscle; however, the available points do not clearly de-fine the correct curve at the ends of the frequency range.

Cerebrospinal fluid (CSF), Figures 9 and 10, is about 99% water andtherefore has very high values for any frequency. The only values foundare those published by Shapiro et al. (7) at 3,000 MHz and by Weil (12)at 1,000 MHz. CSF should have electrical properties similar to thoseof a physiological saline solution and those of the vitreous humor;the scanty data available for these substances do not help in establishingcurves for CSF.

Of human tissues with low water content, fat and bone are of thegreatest interest. Bone samples vary very little in water content andin values of a and E. Fat samples vary greatly in water content andtherefore in electrical properties. The conductivity of fat varies muchmore than the dielectric constant--up to 5 to 1 ratio for differentsamples. Average values for fat and bone are very similar. Most refer-ences print one set of values for the two tissue types. Figures 11 and12 show the values found and curves drawn through these points. Thesecurves are well defined except at the low end of the frequency range.Schwan (10) states that yellow bone marrow has the lowest values c and cof all body tissues. The few values found do not differ significantlyfrom the values for bone and fat.

Available values for the electrical properties of heart muscle,liver, spleen, kidney, and lung were examined and found to be insufficientfor the development of the frequency response curves over the range dis-cussed.

CONCLUSIONS

In summary, a study of the open literature in search of electricalparameter values of various body tissues yielded a few values that hadto undergo careful scrutiny. The useful data gleaned was employed asinput to a computer plotting routine. After many trials of data

6

Page 11: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

processing, a set of graphs, consisting of two graphs per human bodytissue providing conductivities and dielectric constants, eventuallyemerged. The range of the frequency covered is 10-100,000 MHz. Becauseof the scarcity of measured values, particularly at the high and lowends of the freluency range where a and c vary rapidly, the curvesfor tissues of medium or high water content were drawn proportional tothe curves for muscle, for which relatively extensive data exists.Values extracted from these curves give reasonable estimates of theinterrelationships between various tissues over the frequency range ofinterest. Table I is an example of tissue difference at 800 MHz. Asadditional measurements become available, the electrical property tissuecurves will be refined.

REFERENCES

1. Debye, P. Polar molecules. New York: The Chemical Catalog Co., Inc.,1929.

2. Durney, C. H., et al. Comparison of theoretical and experimentalabsorption of radio frequency power, Report No. 77AFQ3, ThirdQuarterly Report, for Brooks Air Force Base, by University ofUtah, Salt Lake City, Utah, 1977.

3. Johnson, C. C., and A. W. Guy. Nonionizing electromagnetic waveeffects in biological materials and systems. Proceedings ofIEEE, vol. 60, pp. 692-718, 1972.

4. Joines, W. T., and R. J. Spiegel. Resonance absorption of micro-waves in the human skull. IEEE Trans Biomed Eng, vol. BME-21,pp. 46-48, 1974.

5. Lin, J. C., et al. Power distribution in a spherical model of manexposed to 1-20 MHz electromagnetic fields. IEEE Trans MicrowaveTheory Techniques, vol. MTT-21, pp. 179-797, 1973.

6. Presman, A. S. Electronic fields and life, pp. 33-44. New York:Plenum Press, 1970.

7. Shapiro, A. R., et al. Induced fields and heating within a cranialstructure irradiated by an electromagnetic plane wave. IEEETrans Microwave Theory Techniques, vol. MTT-18, pp. 187-196, 1971.

7

b

I @.

Page 12: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

8. Schwan, H. P. In J. W. Lawrence and C. A. Tobias (eds.). Biologicaland medical physics, vol. V, pp. 147-209. New York: AcademicPress, 1957.

9. Schwan, H. P. In E. C. Okress (ed.). Microwave power engineering,vol. II, pp. 215-231. New York: Academic Press, 1958.

10. Schwan, H. P. Survey of microwave absorption characteristics ofbody tissues, Proceedings 2nd Tri-Service Conference ofBiological Effects of Microwave Energy, pp. 126-145, 1958.

11. Tinga, W. R., and S. D. Nelson. Dielectric properties of materialsfor microwave processing-tabulated. J Microwave Power 8(l):23-65(1973).

12. Weil, C. M. Absorption characteristics of multilayered sphere modelsexposed to UHF/microwave radiation. IEEE Trans Biomed Eng vol.BME-22, pp. 468-476, 1975.

8

-" - - '" - -- - _ _ J,. , ..

Page 13: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

i 4

12

1o

CD/

101010 00

FRQENYH

Fiue1 Cnutvtyo ucea ucto ffeuny

Teprtr 70C ybo$B ,F n eetoRfrne 2 ,6 n , epciey

Page 14: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

280

240

200

I--

1601

EI,--

120

80

40

F

0 I

10 100 1000 10000FREQUENCY (MHz)

Figure 2. Dielectric constant of mus8le as function offrequency. Temperature 37 C. Symbols B, C,E, H, and K refer to References 2, 3, 5, 8,and 11, respectively.

10

V OEM

Page 15: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

i4

i0

E6

4

2

010100 1000 10000

FREQUENCY (M)

Figure 3. Conductivity of skin-dugs as function of fre-quency. Temperature 37 C. Symbols B, F, G,J, and L refer to References 2, 6, 7, 10, and12, respectively.

T1

Page 16: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

i40

120

100

cc80

I-JC5

~60

40

20

0 I

10 100 1000 10000FREQUENCT(M z)

Figure 4. Dielectric constant of skin-dora as functionof frequency. Temperature 37 C. $yubols 8,G, 3, K, and L refer to References 2, 7, 10,11, and 12, respectively.

12

I77

Page 17: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

14

12

10

4e

U

4

2

0 1p

10 100 1000 10000FREQIJENCY(NHz)

Figure 5. Conductivity of brain a8I function of fre-quency. Temperature 37 C. Symbols 3, G,H, and L refer to References 2, 7, 8, and12, respectively.

13

Page 18: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

140

120

80 HCS, L

60.L

46O

40 D

E20

0 p ! ,

10 100 1000 10000FREQUENCY (MHz)

Figure 6. Dielectric constant of brain as function offrequency. Temperature 37 C. Symbols Z,G, f& and L refer to References 5, 7, 8, and12, respectively.

14

-vi

Page 19: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

14

12

1 0-

F

E

2 B

0

10 100 FEUNYf) 1000 10000

Figure 7. Conductivity of blood to function of fre-quency. Temperature 37 C. Symbol@ B andF refer to References 2 and 6, respectively.

15

1T

Page 20: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

280

240

200

80

420

K K K

0aI10 100 1000 10000

FRCqUENCY(Ofz)

7TIZUr 8. Dielectric constanttof blogd as function offrequency. Tempersture 37 C. Symbols 3, 3,R, and K refer to References 2, 5, 8, and 11,respectively.

16

Page 21: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

I I I

4f

'4

12

01L_)

10 100 1000 10000FREQUENCY(NHz)

Figure 9. Conductivity of cerebrospinal fluid (CI7) asfunction of frequency. Temperature 37 C.Synbols G end L refer to References 7 and 12,respectively.

17

mom

Page 22: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

280

240

200

S160

00

40

80

10 100 1000 10000FREQUENCY (MH)

Figure 10. Dielectric constant of cerebrospinal fluig (CSF)a. function of frequency. Temperature 37 C.Symbols G and L refer to References 7 and 12,re-spectively.

18

Page 23: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

1.4

i2

o0.8

F

0.46

0.2

10 100 1000 10000FREQUENCY (MHZ)

Figure 11. Conductivity of fat-bone-yellow bone mg rrow asfunction of frequency. Temperature 37 C.Symbols B, C, E, F, H, and L refer to Refer-ences 2, 3, 5, 6, 8, and 12, respectively.

19

Page 24: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

56

48

40

S32z

I,-

L_, 24LL,

16

8B8 B 61 B K E

KK K

10 100 1000 10000FREQUENCY(lNz)

Figure 12. Dielectric constant of fat-bone yellow bone marrowas function of frequency. Temperature 37°C.Symbols B, C, F, H, K, and L refer to References 2,3, 6, 8, 11, and 12, respectively.

20

ib

Page 25: PARAMETER VALUES OF SOME HUMAN TISSUES IN THE ... › dtic › tr › fulltext › u2 › a064202.pdf · "1 report sam-tr-78-38 "a' selectrical parameter values of some human tissues

TABLE 1. VALUES OF ELECTRICAL PARAMETERS OF HUMAN TISSUESEXTRACTED FROM APPROPRIATE CURVES

Tissue Conductivity* Dielectric Constant

a

(mho/m)

Brain 0.96 33.76

CSF 1.74 79.47

Dura 1.23 45.64

Bone 0.096 5.61

Fat 0.096 5.61

Skin 1.23 45.64

Muscle 1.45 51.27

Blood 1.60 61.52

Yellow bone 0.096 5.61marrow

At frequency 800 MHz and temperature 370C.