    U Medical Systems Inc

    PENN PET Model 240 H

    WHOLE BODY POSITRON SCANNER BASED ON

    LARGE AREA POSITION SENSITIVE DETECTORS

  • 8/9/2019 PENN PET 240 H

    2/8

    WHY PENN PET

    The

    PENN PET

    scanner was designed with clinical imaging

    in

    mind. It therefore offers

    a combination of features not commonly found

    in

    PET instruments.

    Equal resolution all 3 directions

    combined with fine axial sampling

    allows reslicing into coronal, sagittal

    and oblique sections.

    Large axial field of view

    12 8

    cm

    and no gantry motion, such as wobbling,

    permits gated cardiac imaging and fast

    dynamic studies without sampling

    problems.

    Reslicing along major axis of the heart

    64

    transverse slices and 2 mm spacing

    gives superior quantitative accuracy by

    eliminating partial volume effect.

    Equal resolution

    in

    XYZ and 2 mm sampling

    Superior energy resolution

    of sodium iodide detector material

    allows use of large acceptance angle

    without septa for hig h sensitivity and

    low scatter fraction.

    6 crystals and

    240

    photomultipliers

    result in a scanner with clinically important

    features at an affordable price.

    400

    r - - - - - - - - - - - - - - -

    300

    >

    o

    200

    u

    100

    200 300 400 500

    60 0

    ENERGY keV)

    10 energy resolution

    SPHERE CENTERED SLICE

    SPHERE LOCATED BETWEEN SLICES

    10000

    10000

    2 mm slices PENNPET

    8000

    8000

    6 mm slices

    >

    6000

    6000

    u

    lOOO

    2

    mm

    slices

    4000

    PENN.PET

    2000

    2000

    ,

    ,

    ,

    0

    _ _ _ _ I

    0

    40

    20 30

    40

    AXIAL DISTANCE mm)

    Axial profile through 10 mm sphere with 2mm samples PENN-PET) and with a system having 6 mm slice

    separation demonstrating better quantitative accuracy with 2

    mm

    slices.

  • 8/9/2019 PENN PET 240 H

    3/8

    SYSTEM DES RIPTION

    DETE TORS

    6 large-area posit ion-sensit ive

    Nal TI) detectors

    high spatial resolution along both

    axes 5-6 mm)

    high energy resolution 10 FWHM

    at5 keV

    good coincidence timing

    nanosec

    FWHM)

    pulse shortening and local centroid

    calculation for high count rates

    S NNER

    large patient port suitable for body

    imaging 50 cm diameter)

    12.8 cm axial field of view with

    continuous axial sampling

    completely stationary gantry - no

    need for complicated wobbling

    mechanism

    compact design to minimize space

    requirements

    ,

    DATA ACQUISITION SYSTEM

    separate terminal for acquisit ion

    nearscanner

    acquisition is controlled by a

    pre-processor - frees host

    computer during data collection

    high speed sorting into project

    ion matrices

    provisions for dynamic or gated

    data acquisition

    compact electronics rack - no

    need for separate equipment

    room

    IM GE DISPL Y

    Unix-based display and data

    analysis software

    transverse, sagit tal, coronal and

    oblique sections

    variable image format, multiple

    windows

    2-D and 3-D image display and

    analysis using regions or

    volumes-of-i nterest

  • 8/9/2019 PENN PET 240 H

    4/8

    U

    Medical Systems Inc.

    UGM Medical Systems is dedicated to providing its customers with PET instrumentation

    tai lored to clinical use. UGM operates under the direction of Dr Gerd Muehllehner, one

    of the leading physicists in PET with over 20 years of experience in Nuclear Medicine

    instrumentation. The efforts of Dr.

    Muehllehner

    and his associates

    over

    the last 15

    years have been aimed at bringing the advantages of continuous position-sensitive

    detectors to PET, first through research based at the University of Pennsylvania and

    now at UGM Medical Systems.

    Sodium Iodide Nal TI Detectors

    P T

    Until now, commercially available PET scanners have used detectors composed of many

    discrete

    or

    block-type Bismuth Germanate BGO) crystals. While BGO crystals have

    certain attractive properties,

    their

    scintillation light output and

    therefore

    energy

    resolution) is insufficient for use in large-area continuous detectors. To gain the

    advantage of

    continuous

    scintil lation detectors, UGM has chosen to use sodium iodide,

    Nal TI), one of

    the

    most luminescent scintil lators known, and the mainstay of

    single-photon

    Nuclear

    Medicine

    in

    the form of the

    Anger

    camera. By using Nal TI)

    detectors, the P NN P T can not only provide continuous axial sampling, but can also

    obtain the energy resolution necessary for good scatter rejection without extensive

    detector

    shielding, thereby leading to higher sensitivity.

    To maintain the countrate capability, the large-area detectors

    in

    the P NN P T use

    pulse-clipping to shorten the scintillation response of the detectors, thereby increasing

    the rate at which the detectors can handle photon interactions. Pulse-clipping has

    enabled the

    PENN-PET

    detectors to operate successfully at count rates of over 2 million

    counts per second, a factor of 10 improvement over the best comparable single-photon

    device. Furthermore, instead of calculat ing the event using all photomult ip liers in a

    crystal, the

    P NN P T

    detectors only use the photomultipliers in the immediate vicinity

    of the scintil lation to posit ion the event, cal led the local centroid . This avoids

    interference of signals from other scintillations which may occur in the same crystal

    nearly simultaneously. The result is a continuous detector with all the features that

    have made it the detector of choice in Nuclear Medicine, but with the count rate

    capability necessary for PET.

    REFERENCES:

    The PENN-PET Model 240 H is based on the design described

    in

    the following publications:

    1

    Karp JS, Muehllehner G Mankoff DA, et al: Continuous slice PENN-PET: A positron tomograph with

    volume imaging capabil ity. J Nucl Med to be published).

    2 Muehllehner G Karp JS, Mankoff DA, et al: Design and performance of a new positron tomograph.

    IEEE Trans Nucl Science, 35:670-674, 1988.

    3

    Mankoff DA, Muehllehner

    G

    Karp JS: The high count rate performance of a two-dimensionally

    position-sensitive detector for positron emission tomography. Phys Med Bioi,

    34:437-456 1989.

  • 8/9/2019 PENN PET 240 H

    5/8

    PROVEN CLINIC L PERFORM NCE

    The

    PENN PET s nner

    is well suited for a wide variety of clinical procedures. High

    resolution equil ibrium images of the brain using O have been used in the study of

    Alzheimer s disease tumor detection and staging and the assessment of the severity

    of stroke. Ca rd ia c PET has been used in the evaluation of tissue viabili ty for choosing

    treatment options and assessing treatment eff icacy. The fine axial resolution and 2mm

    slice spacing is particularly effective in neuroreceptor studies.

    Normal FDG brain scan

    FDG cardiac scan of patient with myocardial

    infarction. Papillary muscles are visualized

    on these diastolic gated images.

    FDG scan of patient with Alzheimers disease

    Dynamic study using F-18 spiperone in a

    baboon. Each row shows several slices

    different rows represent different times.

  • 8/9/2019 PENN PET 240 H

    6/8

  • 8/9/2019 PENN PET 240 H

    7/8

    WESTERN

    BIOMEDICAL RESEARCH

    4815

    Mr.

    Pt.EASANT DR

    FLAGSTAFF,

    AZ

    86004

    602

    527-9428

    . v l

    PERFORM NCE

    SPECIFICATIONS

    PENN-PET

    MODEL

    240-H

    G NTRY

    Pa t i en t aper tu re :

    Gantry

    t lt

    motor ized

    t r ans l a t i on

    50 cm

    /- 15 degrees

    for

    pa t i en t couch

    .

    .

    DETECTORS

    mate r i a l :

    s i ze :

    usefu l

    ax i a l f i e ld of view

    length

    th ickness

    Number

    of

    de tec to r s

    Photomul t ip l iers per

    de tec to r

    Detec to r

    separa t i on

    In t r in s i c

    s pa t i a l r e so lu t ion FWHM):

    t ransverse

    ax i a l

    Detec to r t iming r e so lu t ion FWHM):

    SYSTEM

    Sl ice spac ing in

    a l l t h ree di rec t ions :

    Sl ice t h i ckness increment

    se l e c t ab l e

    a f t e r

    data acqu i s i t ion

    Sl ice

    separa t i on : over lap o r c on tig uo us

    Number

    of

    s l i c e s :

    Fie ld of view:

    t r ansver se

    axia l

    Spa t i a l Resolu t ion FWHM):

    t r ansver se

    a t 0

    from center

    t r ansver se a t 10 cm

    ax i a l a t 0 from cen te r

    axia l

    a t 10 cm

    NaI Tl

    14 cm

    50

    cm

    2.5 cm

    6

    40

    85 cm

    5.2

    mm

    6 .2

    mm

    3 nsec

    d

    2

    mm

    2

    mm

    64

    50 cm

    12.8

    cm

    5.5 mm

    6.5 mm

    5.8 mm

    8 .5

    mm

    Volume

    s en s i t i v i t y :

    rand om s an d

    s c a t t e r sub t rac t ed

    photopeak only

    200

    keY th reshold

    Energy r e sou l t ion FWHM @ 5l lkeV

    S c att er f ra c tio n :

    18

    cm

    dia . /photopeak

    18

    cm dia/250 keV

    th reshold

    70,000 CPS/uC i/ cc

    140 ,000 CPS/uC i/ cc

    10

    15

    25