Upload
palupi-darmanti
View
218
Download
0
Embed Size (px)
Citation preview
7/30/2019 Phosphorescence ErythrosinPMMA
1/10
Fluorescence and Phosphorescence of Erythrosin
Purpose
The fluorescence, delayed fluorescence and
phosphorescence of erythrosin immobilized in
polymethylmethacrylate (PMMA) will be observed. The dye
is immobilized to promote phosphorescence. Lifetimes of
delayed fluorescence and phosphorescence will be
measured. Temperature dependence of the intensities of
delayed fluorescence and phosphorescence will give the
energy gap between the lowest triplet state and the first
excited singlet state of erythrosin. Quantum-chemical
calculations will give the singlet-singlet absorption energy and wavelength.
Introduction
Fluorescence and phosphorescence are both examples of luminescence.1 Erythrosin is a highly
colored molecule that absorbs light near 500 nm and emits longer wavelengths. Fluorescence is
fast, occurring on the order of nanoseconds. Phosphorescence occurs more slowly, in about a
millisecond. Phosphorescence is usually observed at low temperatures, but certain conditions
favor observation of room temperature phosphorescence (RTP) of a dye:2
elimination of solvent molecules from the immediate vicinity of the dye
a rigid matrix surrounding the dye molecule
For this experiment erythrosin is dissolved in polymethylmethacrylate (PMMA).
RTP of erythrosin is accompanied by "delayed fluorescence,"2,3 which is normal fluorescence
except that the fluorescent state is populated by transfer of molecules from the phosphorescent
state. That transfer is temperature dependent. Its activation energy Ea is approximately the energy
gap between the triplet and excited singlet states. Delayed fluorescence and room temperature
phosphorescence spectra of erythrosin B have been published for silica gel (reference 4, Figure 1),
for aqueous solutions (reference 5, Figure 2), for various plastics (reference 6), and for sucrose
films (reference 7, Figure 1). Wavelengths for erythrosin in PMMA have also been published. 5
erythrosinPMMA.odt 1
Figure 1: erythrosin
http://www.d.umn.edu/~psiders/courses/chem4644/labinstructions/erythrosinPMMA.pdf
7/30/2019 Phosphorescence ErythrosinPMMA
2/10
The diagram shows absorption from the singlet ground state S0 to excited vibrational levels of the
first excited singlet state, S1. The initial absorption is
followed by rapid relaxation to the ground vibrational
level of S0. Ordinary fluorescence corresponds to the
emission of a photon as the system returns rapidly
from S1 to S0. Some molecules, rather than
fluorescing, make a transition from S1 to the excited
triplet state T1. That transfer is "intersystem crossing."
T1 is initially formed in an excited vibration level that
rapidly relaxes. Phosphorescence occurs as T1 returns
to S0, a slow process because an electron spin flip is
required. Some of the molecules in T1 acquire enough
vibrational energy to back-intersystem-cross to S1
rather than phosphorescing. From S1 they then fluoresce. This fluorescence occurs long after the
initial fluorescence is finished so it is called "delayed fluorescence."
Ideally, delayed fluorescence occurs at the same wavelength as fluorescence but with the same
lifetime as room temperature phosphorescence. Also, delayed fluorescence intensity depends on
temperature because an activation barrier slows transitions from T1 to S1.
Kinetic scheme
Fluorescence is much faster than DF and RTP so it is regarded as instantaneous. The initial
concentration of molecules in T1 is established before fluorescence is complete, so at t=0. Let
[T1]0 be the initial concentration of molecules in T1. For simplicity, neglect nonradiative decay and
quenching of T1 and S1.
d[T1]/dt = -kP[T1] kTS [T1] (1)
where kTS is the rate constant for the back-isc T1 to S1 transfer, a first-order process. The first-
order rate constant for phosphorescence is kP. Solving for [T1],
[T1] = [T1]0 e
-(kP+kTS)t
= [T1]0 e-t /RTP
(2)
where RTP is the lifetime for room temperature phosphorescence. RTP will be measured directly in
this experiment. In terms of the present kinetic scheme, RTP=1/(kP+kTS). The net rate of formation
of the excited singlet is
erythrosinPMMA.odt 2
Figure 2: Jablonski diagram
7/30/2019 Phosphorescence ErythrosinPMMA
3/10
d[S1]/dt = kTS[T1] kF[S1] (3)
where kF is the rate constant for fluorescence; kF= 1/ F is large so d[S1]/dt0 at all times.
Therefore,
[S1] (kTS/kF) [T1] (4)
Let iDF denote the intensity of delayed fluorescence and iRTP the intensity of room temperature
phosphorescence.
iDF = kF[S1] = kTS[T1] (5)
iRTP = kP [T1] (6)
The concentration of triplet cancels from the ratio of DF to phosphorescence intensities.
iDF/iRTP = kTS / kP (7)
The triplet-to-singlet back intersystem crossing rate constant is activated. That is,
kTS = ATS e- Ea/(RT) (8)
where ATS is an Arrhenius prefactor, T is the absolute temperature and E a is the activation energy.
Therefore,
` ln(iDF/iRTP ) = ln(ATS/kP) (Ea/R) (1/T) (9)
The activation energy can be measured by graphing ln(iDF/iRTP) versus 1/T, with T in Kelvin.
Quantum-chemical calculations will complement the spectroscopic measurements. Energies of thetwo singlet states, S0 and S1 will be calculated. Challenging aspects of the calculation are the four
iodine atoms in erythrosin and the need for an excited-state energy. The difficulty with iodine is
that heavy atoms have many electrons, large electron correlation energy, and likely relativistic
effects. Pseudopotentials will be used, which replace iodine core electrons with potential-energy
functions and include relativistic corrections. The difficulty with excited states is that the usual
density-function and Hartree-Fock calculations calculate only ground states. We will use time
dependent density functional theory to calculate the energy of the first excited singlet state. The
excited-state energy will be calculated at the ground-state geometry, which is consistent with the
Franck-Condon principle. After energy difference between S1 and S0 is calculated and convertedto convenient units such as Joules, one may calculate the wavelength simply from = hc/E.
erythrosinPMMA.odt 3
7/30/2019 Phosphorescence ErythrosinPMMA
4/10
Materials and Reagents
Solid solution of erythrosin B in PMMA. The dye erythrosin B has the formula C20H6I4Na2O5,
formula mass 880 g/mol. Solid solutions of erythrosin in PMMA were prepared as described
by Lettinga, Zuilhof and van Zandvoort.
6
Choose a samplefrom the drawer below the fluorescence instrument.
Samples E6 and E7 are convenient. Should you choose
sample BT8 note that it is cut to fit the sample
compartment directly, not in a cuvette.
The drawing of erythrosin at right8 is planar, for simplicity.
In three dimensions, the iodine-bearing rings are planar and
are perpendicular to the carboxylate group and its phenyl
ring.
A plastic cuvette. Use a four-side-clear cuvette for fluorescence. A cuvette that is frosted or
ribbed on two sides is not suitable for fluorescence measurements.
Recommended: something to read while waiting for temperature equilibration and
phosphorescence scans.
erythrosinPMMA.odt 4
Figure 3: planar
7/30/2019 Phosphorescence ErythrosinPMMA
5/10
Procedure
Record the fluorescence spectrum
Use the Varian Cary Eclipse instrument. Choose SCAN mode. Locate a four-sided
plastic or glass cuvette. Zero the instrument with the water-filled cuvette. Under
Setup, set the excitation wavelength to 500 nm. Select 5 nm emission slit width and
10 nm excitation slit width. The detector sensitivity is set using the photomultiplier
tube (PMT) voltage on the Options tab. I suggest setting PMT to medium. Then one
may switch to Low or High as necessary to bring the peak signal on-scale. Once
could also increase intensity by choosing a larger (e.g., 20 nm) excitation slit.
Transfer a block of erythrosin in PMMA into the cuvette, pushing out water so that the sample is
surrounded by water. The water will give good thermal contact. Record the fluorescence emission
spectrum from 530 to 760 nm. If the spectrum is off scale, go back to Setup, choose Options, andreduce the photomultiplier tube (PMT) voltage and or the excitation slit width.
Record the wavelength, F, of maximum fluorescence intensity.
Record the DF and RTP spectrum
Under Setup on the SCAN application, select Phosphorescence. Set the following:
Table 1. phosphorescence settings.
excitation wavelength 500 nm
start scan 530 nmstop scan 760 nm
excitation slit 10 nm wavelength range for the exciting light
emission slit 10 nm wavelength range for a single emissionintensity
tavg 0.1 seconds time over which to average the emissionintensity
4 nm interval between wavelengths in the spectrum
PMT medium "high" gives greatest intensity but may take
the signal off scale (i.e., above 1000).Turn on the spectrophotometer's temperature control unit. Turn it on with the switch on its back.Start the water pump after making sure it is submerged. Verify that water is circulating throughthe hoses. The circulating water does not heat or cool the sample directly; rather, it is to cool thePeltier temperature unit that controls the sample temperature. If water does not circulate turn offthe temperature control unit and solve the circulation problem.
erythrosinPMMA.odt 5
Figure 4:erythrosin in
cuvette
7/30/2019 Phosphorescence ErythrosinPMMA
6/10
Set the temperature to 40C. You can set this manually just by depressing the up- or down- arrow
key on the top of the temperature control unit. Wait until the temperature display reaches 40C,
then wait another 15 minutes. (This might be a good opportunity to use item three in the
Materials and Reagents list.) Then use distilled water to zero the instrument.
Put your sample in the sample holder. Click START
to start the scan. A phosphorescence scan takes longer
than a fluorescence scan so be patient. You may need
to autoscale the Y axis during the run to see the
spectrum if its intensity is low. You should see two
peaks. The longer-wavelength peak is room
temperature phosphorescence (RTP), the shorter is
delayed fluorescence (DF). Note max for each. That
for DF isDF; that for RTP is RTP. You should find that DF is the nearly the same as F, the max
for fluorescence.
Measure lifetimes
Choose the Varian/Cary Eclipse "lifetimes" application. Set the following:
Table 2. Cary lifetime settings
excitation wavelength 500 nmemission wavelength RTP. to measure RTPflashes 25 25 flashes of the excitation lamp
slits 20 nm both slits wide to increase intensity
delay time 0.2 ms This waits until the lamp is off (about 0.1 ms) be-fore collecting emission data
gate time 0.05 ms time between measurementstotal time 3 ms
number of cycles 50 averaging over many cycles improves signal tonoise ratio
on the "Options" tab set PMT voltage "high" for maximum intensity, reduce tomedium if signals is off scale.
on the "analyze" tab stop = total time, single exponential, check "auto calculate"and check "Lifetimes"
Click on START. After all 50 cycles, the lifetime ("tau") will be calculated and printed in the text
window. The fitted line should also be drawn through the data points. If the fit is not displayed on
the graph, go to "Trace Preferences" and check the "SingleExpFit" trace.
Repeat the procedure at DF to measure the lifetime of the delayed fluorescence, DF. One expects
to find DF RTP .
erythrosinPMMA.odt 6
Figure 5: schematic spectrum
7/30/2019 Phosphorescence ErythrosinPMMA
7/10
Temperature Dependence of DF and RTP intensities
Return to the SCAN application. Check that the parameters are reset to the values you used to
record the DF and RTP spectra. In order to display multiple spectra on a single graph you may
check the "Overlay Traces" option under Setup. Zero the instrument with a cuvette of distilled
water.
Record DF and RTP spectra at the following
temperatures: 40, 50, 60, and 70C. Changing
temperature 10 takes about 15 minutes, based on
trials run with a 3-mL sample of liquid water. The
graph at right shows the instruments displayed
temperature and the actual sample temperatures when
a sample initially at 20C was heated or cooled by 10.
Based on these data, a 20-minute wait is suggestedafter calling for a 10 temperature change.
On each temperatures spectrum, measure iDF and iRTP. Measure these intensities at DF and RTP,
the same maximum-intensity wavelengths that you identified when you first recorded DF and RTP
spectra at 40C.
Print the spectra.
When you are done with the fluorescence instrument, please
Remove the erythrosin/PMMA sample from the fluorescence instrument, being careful
because the final temperature of 70C is uncomfortably hot.
Turn off the temperature-control unit (switch is on the back)
Turn off the water pump
Turn off the spectrophotometer
Calculate the activation energy for T1 to S1 back intersystem crossing.
erythrosinPMMA.odt 7
Figure 6: temperature equilibration
7/30/2019 Phosphorescence ErythrosinPMMA
8/10
Quantum-Chemical Calculations
For this lab, calculations are conveniently done using GAMESS, which is installed on several
computers in room 338 and can also easily be installed on a personal computer. The calculations
require several hours. You may want to start calculations and then let them run overnight. Sharingcalculations with your lab partner could also be good.
Step-by-step instructions follow.
1. Draw the erythrosin molecule in the form of the di-sodium
salt but with H atoms in place of the Na atoms. Start with H in
place of Na to make the initial optimization fast.
2. Optimize using the semi-empirical PM3 method.
$CONTRL SCFTYP=RHF RUNTYP=OPTIMIZE $END$BASIS GBASIS=PM3 $END$STATPT OPTTOL=0.0001 NSTEP=400 $END
You may find energy approximately -188 Hartree, the heat
of formation about 50 kJ/mol.
3. Replace two H atoms with Na. Make the O-Na bond length 2.05 Angstroms, which is the O-
Na distance in diatomic NaO. Do not optimize this structure with PM3, because PM3 handles the
sodium atoms poorly. Rather, set up an equilibrium-
geometry calculation using Hartree-Fock Theory, the Hay-
Wadt basis set (with d polarization functions), and Hay-
Wadt pseudopotentials.
$CONTRL SCFTYP=RHF RUNTYP=OPTIMIZEMAXIT=30 MULT=1 PP=HW $END
$SYSTEM MWORDS=100 $END$BASIS GBASIS=HW NDFUNC=1 $END
$SCF DIRSCF=.TRUE. $END$STATPT OPTTOL=0.001 NSTEP=200 $END
It is the command PP-HW that selects Hay-Wadt
pseudopotentials to replace core electrons. The command
OPTTOL=0.001 relaxes the default tolerance for geometry
optimization. Submit the job. Expect about 6 hours to optimize the geometry.
Use the optimized geometry (i.e., the coordinates below EQUILIBRIUM GEOMETRY in the
output file) for the TDDFT calculation below.
erythrosinPMMA.odt 8
Figure 7: Erythrosin B.R. T. Bailey, et al.,
Analytica Chimica Acta, 2003
Figure 8.
7/30/2019 Phosphorescence ErythrosinPMMA
9/10
4. Calculate the excitation energy using TDDFT. Calculations involving excited states should
include electron correlation, so we will use the B3LYP density functional. Correlated
calculations, especially those for an excited state, require larger basis sets. We will use the MCP-
DZP basis set and its associated MCP pseudopotentials. Coordinates of the atoms should be
taken from the previous geometry optimization.
$CONTRL SCFTYP=RHF RUNTYP=ENERGY MAXIT=50TDDFT=EXCITE DFTTYP=B3LYP PP=MCP ISPHER=1 $END
$BASIS GBASIS=MCP-DZP $END$SCF DIRSCF=.TRUE. $END$SYSTEM MWORDS=100 $END
The command ISPHER=1 tells GAMESS to use spherical-harmonic functions for angle-
dependence of the basis functions, rather than the Cartesian functions (e.g., x for a p x orbital) that
are used by default. The directive MWORDS=100 sets aside nearly a gigabyte of RAM for the
calculation.
Expect the calculation to take about 7 hours. In the output file, near the end, should be a small
table of results. E will be given in eV.
SUMMARY OF TDDFT RESULTSSTATE ENERGY EXCITATION TRANSITION DIPOLE, A.U. OSCILLATOR
HARTREE EV X Y Z STRENGTH0 A -751.XXXXXXXXXX 0.0001 A -751.XXXXXXXXXX X.XXX -0.0290 -0.0304 -0.1943 0.002
SELECTING EXCITED STATE IROOT= 1 AT E= -751.XXXXXXXXXX AS THE STATE OF INTEREST.
5. Convert E to in nm.
erythrosinPMMA.odt 9
7/30/2019 Phosphorescence ErythrosinPMMA
10/10
Lab Report
Include all spectra.
Compare your fluorescence, DF and RTP wavelengths to literature values (which may be
for matrices other than PMMA).
Compare your DF and RTP lifetimes to each other.
Compare Ea to a literature value for erythrosin in some solid matrix.
Report your DFT B3LYP energies of the ground and excited states, your E, and your
wavelength. Compare your calculated wavelength to an experimental or literature max.
References
1. Engel, Thomas Quantum Chemistry and Spectroscopy, Pearson Benjamin-Cummings, San Francisco,
2006, Sections 15.5-15.8.Silbey, R. J.; Alberty, R. A.; Bawendi, M. G.Physical Chemistry, 4th ed., John Wiley & Sons, Inc.:
New York, 2005; Section 14.8.
2. Wayne, R. P.Principles and Applications of Photochemistry; Oxford University Press: Oxford, 1988;
pages 91-92.
3. Levy, D.; Avnir, D. Room temperature phosphorescence and delayed fluorescence of organic molecules
trapped in silica sol-gel glasses. J. Photochem. Photobiol. A: Chem.1991, 57, 41-63.
4. Lam, S. K.; Lo, D. Time-resolved spectroscopic study of phosphorescence and delayed fluorescence ofdyes in silica-gel glasses. Chemical Physics Letters1997, 281, 35-43. Here is a link to the article.
doi:10.1016/j.physletb.2003.10.071 (UMD username and password may be required. If that does not
work, see the link under CHEM 4644 lab instructions.) Pertinent values from the paper are listed here.
All values are for erythrosin B in sol-gel silica.
F = DF = 5613 nm. RTP = 6833 nm. E = 0.40 eV (Ea)
F = 1.1 ns DF = 230100 s RTP = 23010 s on page 3701.
5. Duchowicz, R.; Ferrer, M. L.; Acuna, A. U. Kinetic spectroscopy of erythrosin phosphorescence and
delayed fluorescence in aqueous solution at room temperature. Photochemistry and Photobiology1998,
68(4), 494-501. Table 1 give RTP in PMMA.
6. Letinga, Minne Paul; Zuilhof, Han; van Zandvoort, Marc A. M. Phosphorescence and fluorescence
characterization of fluorescein derivatives immobilized in various polymer matrices. Physical
Chemistry Chemical Physics2000, 2, 3697-3707. PMMA preparation is briefly described.
7. Pravinata, Linda C.; You, Yumin, Ludescher, Richard D. Erythrosin B phosphorescence monitors
molecular mobility and dynamic site heterogeneity in amorphous sucrose. Biophysical Journal, 2005,
88, 3551-3561.
8. Bailey, R.T.; Cruickshank, F.R.; Deans, G.; Gillanders, R.N.; Tedford, M.C. Characterization of afluorescent sol-gel encapsulated erythrosin B dissolved oxygen sensor.Analytica Chimica Acta, 2003,
487, 101-108.
erythrosinPMMA.odt 10
http://dx.doi.org/10.1016/S0009-2614(97)01172-Xhttp://dx.doi.org/10.1016/S0009-2614(97)01172-Xhttp://dx.doi.org/10.1016/S0009-2614(97)01172-X