35
Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Embed Size (px)

Citation preview

Page 1: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Pixel Ladder QA manual

Hardware

R. Akimoto, A. Shaver. July 2010

Page 2: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Check List 1• Interlock

– Temperature of NOVECCheck if the difference from the value in the RIKEN QA is less than 5° C

– Humidity inside test boxCheck if the difference from the value in the RIKEN QA is less than 5%

Relative Humidity (R.H.)• Power supply

– Voltage and current of the power supply for the SPIRO and LADDER are at the operational values.SPIRO: Check that the difference of the current from the value of the

same SPIRO is less than 0.1 ALADDER: Check the difference of the current from the value in the

RIKEN QA is less than 0.1 A– Bias voltage and leakage current at 5 V increments

Check the difference in the leakage current from the value in the RIKEN QA is less than 0.3 μA

2

Page 3: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Check List 2

• DAC value– SPIRO S/N

Confirm that the SPD FEM is reading the correct S/N value

– Reference voltage (DAC_REF_VDD, DAC_MID, and GTL_REF) Check the difference from the value of the same SPIRO is less than 2

– Threshold value for each chip Check the difference from the value in the RIKEN QA is less than 10

– Mask pixel Check if there are differences from the results in the RIKEN QA

• β source testing Check the difference in the dead area against the RIKEN QA

3

Page 4: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

SETUP FOR QA TESTING

4

Page 5: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Connected SPIRO board (with test adapter)

5

Page 6: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

(Left) Ladder connection and Setup

6

Page 7: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Supporting the Ladder fixture

We used this book to support ladders coming out of and going into the test box.7

Page 8: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Supporting the Ladder in fixture with Hex (Allen) wrench

Note the position of the wrench supporting the stave

Page 9: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Useful ToolsMinor divisions are ~ 1mm

Plastic-Sheathed Eyeglass Screwdriver:Safely lifts data connection from ladder

5/64” Hex (Allen) Wrench:Supports Ladder for NOVEC tube attachment/detachment

9

Page 10: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Test Box Setup 1• Setup procedures must be done very carefully1. Remove the top cover and bottom acrylic plate– It is helpful to place across the test box something to

support the stave above the box until you’re ready to lower the stave into the box. This is also useful during disassembly (See book in the previous slide)

2. Attach tubes for NOVEC– Silicon grease is used to smoothly attach and detach the

tubes. It is applied to the inside of the tubes– Use a ~ 5/64” or 2 mm hex (Allen) wrench to support the

Ladder in its fixture (see picture)– Secure the tube to the white plastic attachment point

underneath the ladder• Be gentle in this connection, but make sure that the connection

is well made.

10

Page 11: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Test Box Setup 23. Attach power connections– Attach the Red 3.5 VDD leads from the Ladder to the VDD

connection on the bus– Attach the Black GND leads from the Ladder to the GND

connection on the bus– Attach the (white taped) Bias leads from the bus to the Bias

connection on the Ladder.– Each of the leads can fit into multiple positions within the

associated connection. The specific position doesn’t matter for VDD and GND.

– The position does matter for Bias. • On the bus, the two connections closest to the data connectors have

lines going to the SPIRO board for Bias voltage. Connect the wire to the connectors closest to the data connections. The other two connectors are connected to GND.

• On the Ladder, the two connections closest to the data connectors are usually soldered together with a wire to GND. The other connections are the Bias connections. Connect the wire to the connectors furthest from the data connections. 11

Page 12: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Test Box Setup 34. Attach Data connections– Use canned/compressed air to blow out dust on the connectors

• Block or otherwise direct the air such that it doesn’t blow across the sensor chips (so dust doesn’t land on them)

– Line up the connectors on the Bus and Ladder– Be Careful! When properly aligned, firmly press the connectors

together. There may be some feeling of connectors snapping together. Make sure a good connection is formed, but remember that these parts are very fragile!

5. Lower the ladder into the test box. – Consider holding the corners of the aluminum ladder support

and lowering straight down.6. Close up the box, flow dry N2 and begin Interlock sequence

12

Page 13: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Ladder Connections

13

Data Connectors

Ladder VDD

GND

BiasGND

Page 14: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Bus (Extender)

14

Notice the two bias lines going to the BIAS connection. They connect to the two connectors closest to the data connections. The BIAS supply must be connected to one of these two connectors. The other two are connected to ground (GND).

Page 15: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Bus Connections

Data Connection(Under-side)

Power connectionsTop Left: GND (ground)Bottom Left: VDD (Ladder 3.5V)Right: Bias (0-50V)

15

Page 16: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Nitrogen Panel

16

Valve to open N2 Flow

Flow meter to Control Rate of N2

Page 17: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Nitrogen Flow• To open flow, turn black knob on upper right of panel (near

yellow tube) counter-clockwise. Then turn the left flow meter knob (in the lower left-hand side of the panel) counterclockwise to flow nitrogen out to the test box.

• To close flow, turn flow meter knob clockwise, then turn the black knob on the upper right of the panel clockwise completely.

• Note: When you are initially purging the air from test box to get down to 0% Rel. Humidity, it is useful to open the flow up fully (8-10 on the flow meter). After the humidity check has passed, you should turn down the flow to 4-5 to conserve nitrogen.

17

Page 18: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Valve Controls

18

Supply and Return lines to:

Clean tent

Work Bench

Main Valve:

Arrow pointing up is On

Arrow pointing down will purge lines with N2

Position shown is OFF

Page 19: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Interlock System• Before powering the Ladder and Bias, you must pass Interlock system

checks. The interlock system will test the humidity inside the test bench, the NOVEC temperature, and the NOVEC flow rate. It also includes valve controls and monitoring for the NOVEC flow.

• Colors correspond to the diagram on the following slide1. Turn on the Power to the Interlock System. 2. With the N2 flowing, the humidity in the box should be decreasing.

– We need to prevent condensation when the ladder is chilled, so the humidity must be sufficiently low

3. After the humidity reaches its threshold, 0% Relative Humidity (R.H.) and the checkpoint is passed, start the pump and refrigeration– The LED on the Interlock will change from red to green to indicate – For each check you must confirm that the LED has changed.

4. Open the valves on the chiller and open the valve on the Interlock Control; Turn on pumps and refrigeration. (Chiller instructions on slides 16, 17)– Confirm that the interlock system is measuring a flow with the ‘FLOW’ LED

5. When the temperature reaches the threshold temperature (17.5° C for testing) begin to power-up the low voltage equipment

19

Page 20: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Interlock system

Valve control:Open

Close

Interlock power

Relative Humidity:measuredthreshhold

Temperature reading (here, of the NOVEC, as no temp. sensor is on the staff)

System Checks: green light indicates check passed, red light indicates check failedValve, Output, Humidity, Temp, Flow

Humidity sensor reading

20

Page 21: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Chiller

Chiller description:•The switch on the left is the main power switch.•The ‘Pump’ button toggles the pump on or off, and the status is indicated by the ‘Pump’ LED.•The ‘Run Stop’ button turns the refrigeration control on or off. The status of this control will be indicated by the LED beside °C. If that LED is blinking, the thermostat is on and the unit will maintain the set temperature. If it is steadily on, the unit is not going to cool the liquid. The ‘Refrigerator’ LED merely indicates whether the refrigerator is actively running at that moment.•Holding ‘Set Clear’ and using the arrows allows you to change the temperature the chiller will try to maintain.

Page 22: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Chiller operation

• To start, turn on the main power switch, press ‘Pump’ and press ‘Run Stop’– The machine stores temperature settings, so if

you need to adjust the temperature, see the previous slide.

• To turn off, press ‘Pump’ and press ‘Run Stop’ and then turn off the main power switch

22

Page 23: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Low Voltage • After interlock checks are all passed (LEDs are green) begin

powering voltages to SPIRO board then to the ALICE1LHCB chips (Ladder / VDD). After providing VDD begin the procedure in the software manual to set DACS. Then provide voltage to the Bias.

1. SPIRO: Turn on power supply, output 5 V to the SPIRO board– Should be 5V, and approximately 1.8 A

2. VDD:1. Turn on power supply, confirm that the minimum voltage is set2. Output the minimum voltage to the VDD3. At a rate of approximately 0.1 [V/sec], increase the voltage to 3.5 V

• The current may rise steeply between approximately 1.5 V and 3V• Do not allow the current to exceed 5 A• If the current is quickly approaching 5 A, pause for a moment (~5 sec), and

the system will equilibrate somewhat, and then continue at a slower pace.• The current will be approximately 4.4-4.6A when the voltage reaches 3.5 V.

After applying the DAC settings later, the current will drop to approximately 3.3-3.5 A

23

Page 24: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Low Voltage SetupRight Half-Ladder Left Half-Ladder

SPIRO Power Supplies

RHL VDDPower Supply

LHL VDDPower Supply

24

Page 25: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Bias Voltage1. Turn key on crate to “local.”2. On computer monitor: when booted, press enter, then enter user for both

the user name and password.3. Select ‘Menu,’ then ‘Channels.’ move down to the HV-1 Channel, then right

to ‘Off.’ Press the space bar to turn the channel on. Repeat for HV-2– Grey options indicate settings that the user can set. White are generally

monitoring data.– The screen is partially cut-off at the edges, so maybe only V-1 will appear on

screen.4. Move left to the voltage column. Apply 5 Volts (type 5 and press enter) to

both HV-1 and 2.5. Wait for the system to come to voltage, and wait about 10 seconds for the

voltage to equilibrate.6. Record current.7. Increase the voltage in 5 volt increments to 50 V, or until the current

continues to grow with each voltage increase.8. To turn off, decrease the voltage in 5 volt steps until 0 Volts. Change column

to off. Turn key to “off.”• Note there are columns RUp and RDwn. These columns set the rate at which

the voltage will increase. I find that 1 [V\sec] limits the current to ~1.3μA25

Page 26: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Installing Beta Source• Keep metal shields covering source as often as possible

in installing and removing the beta source.1. Turn source mechanism upside-down. Remove plate

from chamber.2. While leaving protective shields over source place

source into chamber. As quickly as possible, remove shield, place the plate back down, and place the shield back over the source.

3. Screw plate back on.4. Turning the mechanism so that the source points away

from you, turn it upright and place it on the posts on the test box.

26

Page 27: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Beta Source Pictures

27

Step 1 Step 2

Page 28: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

DISASSEMBLY

28

Page 29: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Disassembly notes 11. In general, proceed through the assembly procedures in

reverse2. Power-Down

1. Gradually decrease Bias voltage at approximately .1 [V/sec]2. Turn off output of Bias power supplies, and power-off Bias power

supplies.3. Decrease VDD voltage at ~.1 [V/sec], turn off power supply

output, and power-off VDD power supplies4. Turn off SPIRO power supply output, and power-off SPIRO power

supplies3. Thermalize

1. Turn off pump and refrigeration2. Close the valves on the chiller and close the valve on the Interlock

system3. Wait until the temperature reading rises to the 17.5° C threshold

to prevent condensation when you open the test box29

Page 30: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Disassembly notes 24. Open the test box, lift out the Ladder and support it5. Carefully disconnect the NOVEC tubes:

4. Use the Hex(Allen) wrench to support the ladder in its fixture again.

5. Twisting back and forth, slowly pull off the NOVEC tube6. Using a plastic sheathed eyeglass screwdriver (as pictured in

the useful tools slide #9),slide the screwdriver end between the two connections on the data connector and carefully pry the bus data connection off the ladder.4. See following picture5. Plastic sheathing is important to prevent damage to the ladder or

bus. 7. Disconnect the VDD and GND wires from the bus; Disconnect

the Bias from the Ladder.8. Tilt the ladder to drain the remaining NOVEC9. Return Ladder fixture to case and secure the case.

30

Page 31: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Removing Data Connector

Page 32: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

APPENDIX

32

Page 33: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Bus and SPIRO connection• It is convenient to just keep the bus connected to a

SPIRO board throughout QA testing procedure, but if you need to attach/detach:

• Attaching the Bus to the SPIRO– Line up the bus connection to the SPIRO or SPIRO test

board, and press together– NOTE: these are pinned connections, be very careful about

bending pins on the SPIRO connection. The connection should go on fairly smoothly without excessive resistance

• Detaching the bus and SPIRO– Rock the connection back and forth while gently pulling

the bus off of the SPIRO board

33

Page 34: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Spiro connection on Bus

Page 35: Pixel Ladder QA manual Hardware R. Akimoto, A. Shaver. July 2010

Bus connection on SPIRO Test Boards