Upload
others
View
16
Download
0
Embed Size (px)
Citation preview
Practical Aspects of Medical Physics Surveys of Mammography Equipment
and FacilitiesMelissa Martin, M.S., FAAPM, FACR, FACMP
AAPM Annual Meeting - PhiladelphiaJuly 19, 2010
MO-B-204C-1
Educational Objectives
To cover the annual tests to be performed by the medical physicist for both film/screen and digital mammography units
To discuss efficient methods of performing these required tests
To discuss the required tests for laser printers and monitors or processors and view boxes
Film/Screen MQSA Required Tests
Follow the tests outlined in the ACR 1999 QC Manual for Film/Screen Mammography
Physicists must use the Summary Forms from the ACR Website: www.ACR.org
Download forms for both Equipment Evaluations and Annual Surveys
FFDM Equipment Evaluations and Surveys
At the present time, the FDA requires the physicist to use the FFDM Manufacturer’s QC Manual and Procedures.
Physicist must download the Summary Forms for both Equipment Evaluations and Annual Surveys from the ACR website for the particular manufacturer’s unit tested.
FFDM Equipment Evaluations and Surveys
Physicist must follow the latest version of the Manufacturers QC Manual procedures for the unit tested (may not be the latest version available from that particular manufacturer if the unit has not had the latest upgrade/update installed)
Test results must meet manufacturer’s performance standards
FFDM Equipment Evaluations and Surveys
Failures must be fixed before use on patients or within 30 days as specified for each particular test
GE, Fuji, Hologic and Siemens have received FDA Approval for alternative standards for some QC test results to allow 30 days to achieve compliance with manufacturer’s specifications
ACR’s Current FFDM QC Requirements
same as FDA’s ACR suggests using Manufacturer’s data
forms Medical Physicists MUST complete the
ACR’s Summary FormsIf the medical physicist passes a test, ACR accepts it.If the medical physicist fails a test, ACR requires corrective action documentation.If the medical physicist says NA or “see comments”, the ACR requests further clarification.
LORAD Selenia QC TestsQC Manual tests: Medical Physicist Annually
Radiologic Technologist Daily Weekly Biweekly Monthly Quarterly Semiannually
Medical Physics Annual Tests Mammographic Unit Assembly Evaluation Collimation Assessment Artifact Evaluation kVp Accuracy and Reproducibility HVL Measurement Evaluation of System Resolution Breast Entrance Exposure and Glandular Average
Dose Radiation Output Rate Phantom Image Quality Evaluation Signal-To-Noise and Contrast-To-Noise
Measurements Viewbox Luminance and Room Illuminance Softcopy Workstation QC
13
ACR Required Tests
The ACR does require specific tests, common to all vendors:
Weekly: Viewboxes and Viewing Conditions
Monthly: Visual ChecklistSemi-Annual: Compression force
Taken from the ACR Mammography Quality Control Manual 1999
14
Test GE Fuji CR Siemens Hologic Fischer
Monitor Cleaning Daily
Laser Image Quality Test Weekly Daily
Daily and Weekly QC Tests(per Vendor)
Flat Field Calibration Weekly Weekly Weekly Weekly
Artifact Evaluation Weekly Weekly
Phantom Image Quality Weekly Weekly Daily Weekly Weekly
Phantom Acquisition Test Weekly
Signal-to-noise-ratio Weekly Weekly
Contrast-to-noise ratio Weekly Weekly Weekly Weekly
Viewbox and Viewing Conditions Weekly Weekly
Compression Thickness Indicator Test Bi-weekly
15
Monthly and Quarterly QC TestsTest GE Fuji CR Siemens Hologic Fischer
MTF MeasurementDS/Essential –weekly
2000D - monthly
AOP Mode and SNR Monthly SNR Weekly SNR Weekly
System Resolution Monthly
System Operation Monthly
Visual Checklist Monthly Monthly Monthly
Repeat Analysis Quarterly Quarterly Quarterly Quarterly Quarterly
Compression Force Semi-annuallySemi-annually Semi-annually Semi-annually
Semi-annually
Soft Copy Workstation QC FDA guidance
FDA guidance FDA guidance FDA guidance FDA guidance
Laser Film Printer QC Per printer mfgPer printer mfg Daily Weekly
Mobile Unit QC At every move Unknown At every move At every moveAt every move
Image Plate FogSemi-annually
Laser Printer Quality Control SMPTE Pattern from Acquisition
Workstation Measure 10%, 40% and 90% Density
Patches Daily Performance40% density tracks within ±15% (40% - 10%) density tracks within ±15%90% density tracks within ±15%
NOTE: Printer uses linear LUT
Kodak daily sensitometry
Laser Processor QC
Base + Fog
Density Difference – OD closest to 2.20 minus OD closest but not less than 0.45
Mid-density – step closest to but not less than 1.20
Dmax
Action Limits:
MD & DD + 0.15 OD
B+Fog = 0.03
Dmax + 0.25
FDA Required Laser Printer QC GE 2000D & Senographe DS: Follow the laser
printer manufacturer’s QC manual Fischer Senoscan: Follow the laser printer
manufacturer’s QC manual Lorad Selenia: Follow the Lorad Selenia QC Manual Siemens Mammomat Novation DR: Follow the laser
printer manufacturer’s QC manual but conduct QC every day that images are printed.
Fuji MammoCR System: Follow the laser printer manufacturer’s QC manual
Viewing Conditions for Monitors
From the ACR QC Manual, 1999, the room illuminance must be below 50 lux.
From the February 2006 version of the Medical Physicist Equipment Evaluation forms from ACR, room illuminance for soft copy displays must be below 20 lux.
23
Hologic QC tests on SecurView
Required testing is determined by monitor type CRT LCD (Flat panel)
CRT Manually perform QC tests using photometer on each
monitor to document the testing LCD
QC tests are set up to run automatically at night Technologist must check results and record
24
Hologic QC tests on SecurView
CRT: Cathode Ray Tube
Weekly Calibration Settings
Check Measure Display White Measure Display Black
Monthly Measure Display Black Measure Display White
Quarterly Uniformity Measurement
LCD: Liquid Crystal Display
Weekly Automatic I-Guard Check Calibration Settings
Check Measure Display White
Monthly Measure Quality Level
1 3 . D i a g n o s t i c R e v i e w W o r k s t a t i o n Q C
Date: 18-May-10
Review Station Model: SVDX-00400
Review Station SN: 29808071104
Left Monitor Right Monitor Comments
Monitor Serial Number 1890097709 1890097714 Barco 5K Monitors
White Level Performance 599.57 599.58 Target = 600 cd/sq m
Black Level Performance NA - LCD Monitors NA - LCD Monitors
Quality Level Performance OK OK Within 3% of 100% target
Uniformity Performance NA - LCD Monitors NA - LCD Monitors
Action Limit: The white level shall not be different from the monitor
calibration level by more than 6%. The quality level performance
shall not exceed 10%.
Phantom Image - Hologic Selenia FFDM Units
Review Weekly Phantom Image ResultsMeasured film densitiesPhantom objects seen (printer,
monitors)5 fibers4 speck groups4 masses
SNR and CNR MeasurementsACR Accreditation Phantom with
Disc (Weekly)Measure SNR and CNR from
ImageSelenia FFDM Units
SNR and CNR Measurements
Passing Criteria - Selenia FFDM UnitsSNR at least equal or greater than
40CNR should stay within ±15% of
measurement obtained during acceptance testing of system
Artifacts
Physicist must test all Target/Filter combinations and both sizes of focal spots for artifacts
GE Senographe Essential
Large and Small Fields Require Collimation Assessment of all three positions of small field plus large field for both target materials
GE Senographe Essential
Measure Entrance Skin Exposurewith Ion Chamber at 5 cm from outside edge of detector area.
Tabs to position IQST with respect to chest wall edge of image receptor.
Rails to position the IQST left-to-right
Uniform area for noise power spectrum measurement
Step wedge for contrast measurement
Edge object for MTF measurement
Mesh for resolution uniformity measurement
Rulers for measuring distance of detector from chest wall edge of Bucky
QA Phantom OverviewGE Senographe DS &
Essential
Fuji CRm System - CNR Test
CNR PER OBJECT THICKNESS
Image Mode: Auto Time
Focal Spot: 3
Density Control: 0
AEC Mode: Auto Time
Phantom Thickness Target Filter kVp mAs (Auto)
mAs (Manual) CNR
CNR (relative to
4cm PMMA)
Acceptable level
2 cm Mo Mo 25.0 27.2 28.0 17.197 114.4% >110%
4 cm Mo Mo 25.0 135.0 130.0 15.030 100.0% 100%
6 cm Mo Rh 30.0 282.0 280.0 13.765 91.6% >90%
PASS
Exposure Unite ID: Lorad M-III
AEC-mode
Target Filter kVp
Exposure conditionsAuto Time Mo Mo 25
Group: Small Cassettes
Cassette IDmAs
Acceptable
SNR
Acceptable
A456, #3 132 Yes70.946 Yes
A234, #2 132 Yes70.952 Yes
A123, #1 132 Yes70.152 Yes
A222, #4 132 Yes70.166 Yes
A111, #5 132 Yes69.521 Yes
A777, #6 133 Yes69.662 Yes
mAs limit is plus or minus 10 % from mean.SNR limit is plus or minus 15 % from mean.
Inter Plate Consistency
Fuji CRm - Monitor TestsSite: Med. Img Ctr of So Cal Date of Survey: 4/19/2007
Room #: Reading Room Date of Installation: Apr-07
DOME©E5
Target Value = 500 cd/sq-m
499.2
Serial Numbers
Luminance Value Setting (cd/sq-m)
Left Monitor Right Monitor
703PNKN00028
500.88
703PNKN00011
Is the CXtra icon present in thetaskbar and shown as a green
check mark?
Manual Conformance report forboth displays attached?
Is DICOM calibration Graph forboth displays attached?
NoYes
Yes
Yes
No
No
Thank You !!!
Melissa C. Martin, M.S., FACRTherapy Physics Inc.
879 West 190 St., Ste 419Gardena, CA 90248
Office Phone: 310-217-4114e-mail: